16
P/N:8024313 VersionA Date2018-11-26 Installation Manual of Orthodontic Mobile Cabinet http://www.adsequip.com/

PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

P/N:8024313 Version:A Date:2018-11-26

Installation Manual of

Orthodontic Mobile Cabinet

http://www.adsequip.com/

Page 2: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

CONTENTS

1 Tools recommended…………………………………………………1

2 Dimensions…………………………………………………………2

3 Installation……………………………………………………………4

4 Water & air diagrams…………………………………………………8

5 Functions………………………………………………………………9

Installation Manual of Orthodontic Mobile Cabinet

Page 3: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

1

1 Tools recommended

The tools and instruments need to prepare before installation.

P/N:8024313A

Electrical drill

Pincers

Hammer

Scissors

Leveling

instrument

Flat-bladed

Screw driver

Phillip screw

driver

Allen wrench

Spanner

Hexagon

socket wrench

Installation Manual of Orthodontic Mobile Cabinet

Page 4: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

2 Dimensions

2 P/N:8024313A

OC-2

Installation Manual of Orthodontic Mobile Cabinet

Page 5: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

3 P/N:8024313A

OC-3

Installation Manual of Orthodontic Mobile Cabinet

Page 6: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

4 P/N:8024313A

3 Installation

1. Route the tubes through the

corrugated pipe as the figure shown.

This step is not necessary for OC-3 cabinet

2. Let one end of the corrugated

pipe pass through the hole

and reach inside the cabinet,

fasten the pipe clip by a

ST4*16 cross recessed pan

head tapping screw to fix the

corrugated pipe.

Installation Manual of Orthodontic Mobile Cabinet

This step is not necessary for OC-3 cabinet.

Page 7: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

5 P/N:8024313A

4. Connect the tubes in corresponding colors.

3. Pass the other end of the corrugated pipe through a metal buckle, fasten the

buckle when the pipe is connected with the frame of junction box.

Installation Manual of Orthodontic Mobile Cabinet

Page 8: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

6 P/N:8024313A

Use 2pcs nuts to lock the display

turntable onto the mounting seat.

Use 4pcs screws to lock

through the keyboard bracket

and the display.

5. Installation of monitor bracket

Use 4pcs screws to mount the

monitor seat lock on the back

panel.

③ ②

Installation Manual of Orthodontic Mobile Cabinet

Page 9: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

7 P/N:8024313A

The tubes and wires are in the tubing zone or inside the external junction box,

their places depend on your chair configurations.

1). Open the valves to exhaust water and air, clean the tubes before connecting.

2). Open the cover of junction box,connect the air-in,water-in and water-out .

3). Close the cover after making sure they are well-performed.

6. Install junction box.

Suction

Air compressor

City water

Installation Manual of Orthodontic Mobile Cabinet

Page 10: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

8 P/N:8024313A

4 Water & air diagrams

1/8" blue, water tube

1/8" yellow, air tube

1/8"orange, air On/Off,

flow in

1/8"black, air On/Off, flow

out

1/8"Grey, foot control tube

Installation Manual of Orthodontic Mobile Cabinet

Page 11: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

9 P/N:8024313A

5 Functions

Air Coolant Master ON/Off Handpiece Flow Controls

5.1 Control panel

Air coolant : This valve controls the amount of air coolant flow to all the

handpieces.Rotate counterclockwise to increase flow,clockwise to decrease flow.

Master On/Off :This toggle switch activates the air and water shut-off

valves,which control the air and water supplies to the unit.

Handpiece flow controls :These valves adjust the amount of water coolant

supplied to its corresponding handpiece.Rotate counterclockwise to increase

water coolant flow, clockwise to decrease the flow.

Air adjustment of water bottle

Toggle switch of water bottle

Pressure Gauge

Installation Manual of Orthodontic Mobile Cabinet

Page 12: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

10 P/N:8024313A

5.2 Foot control

Turn off mist spray

Turn on mist spray

NO

OK

Note:

Adjust the air pressure to meet the requirements of handpiece

manufacturers. Refer to your handpiece documentation to learn

more about the requirements of handpiece air pressure.

Installation Manual of Orthodontic Mobile Cabinet

Page 13: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

11 P/N:8024313A

5.3 Three-way syringe adjustment

Installation Manual of Orthodontic Mobile Cabinet

Page 14: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

12 P/N:8024313A

6 Maintenance and sterilization instructions

Flush the tubings independently or together. After each treatment :

1.Disconnect the handpieces before flushing the tubings.

2. Gather handpiece tubings which use water coolant and hold them

over a sink, or a basin.

3. Hold the tubings to let the water flow into the container.

4. Flush the tubings continuously about 20~30 min .

1.Suck a cup of fresh water to clean after every two hours during operation.

2.Suck a litre of water to flush after finishing every day's work.

6.1 Handpiece tubings cleaning & maintenance

6.2 Suction handles cleaning

Notice :

Discharge air and water of all tubes for about 20–30

seconds after each treatment.

Installation Manual of Orthodontic Mobile Cabinet

Page 15: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

13 P/N:8024313A

Keep weekly maintenance in regular use and increase maintenance frequency with

frequent use. To clean the dirty collector:

1. Unscrew the dirty collector counterclockwise and remove it;

2. Discard the used cotton.

3. Put fresh cotton into dirty collector;

4. Tighten the dirty collector clockwise.

6.3 Dirty collector cleaning

The filter prevents solids from entering the suction

system. Clean the filter screen at least twice a

week to ensure the suction system works properly.

To clean the filter:

1.Close the suction system.

2.Remove the lid, take out the filter screen and

clean the solids.

3.Reinstall the filter.

6.4 Filter cleaning

Installation Manual of Orthodontic Mobile Cabinet

Page 16: PowerPoint PresentationInstallation Manual of Orthodontic ......Spanner Hexagon socket wrench Installation Manual of Orthodontic Mobile Cabinet 2 Dimensions 2 P/N:8024313A OC-2 Installation

ADS Dental System Inc.

Add: 1590 S Milliken Ave., unit A, Ontario, CA 91761 Tel: (626) 620 0456 E-mail: [email protected] http://www.adsequip.com/

We reserve the right to make any alterations which may be due to technical

improvements.