There are 3 fundamental calibrations

• Flow rate • Ionization efficiency • Particle size calibration

µg

m3

mass

volume

From Mass Spectrometer Ionization Efficiency calibration

From volumetric flow rate

Particle Mass Loading

µg

m3

µg

m3

mass

volume

mass

volume

From Mass Spectrometer Ionization Efficiency calibration

From volumetric flow rate

Particle Mass LoadingParticle Size pTOF to Dva

Page 3: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Flow Rate Calibration

Page 4: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Flow Meter

Low pressure drop flow meter

Gilibrator™ or equivalent soap film

meter 10 Torr lens

pressure gauge

P

Aerodynamic Lens

Fine metering valve

AMS Flow Rate Calibration Setup

Poiseuille equation relates volumetric laminar flow (F) to a pressure drop (ΔP) across the tube

Flow into vacuum

Replace 100 um pin hole with fine metering valve

SS-SS4.

Page 5: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Flow Calibration Depends on Pressure and Temperature

Always document ambient pressure and temperature

100 µm leak

Page 6: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Flow Rate Calibration Summary

• In principal you only need to do this once as long as you document ambient pressure and temperature condition when you do the calibration (also assumes you are always sampling air).

• What you are measuring/calibrating here is a property of the aerodynamic.

• When reporting µg m-3, always reference the T&P conditions.

Page 7: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Velocity Calibration

Page 8: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Particle Velocity Calibration

Daero = Dgeo * ρ* Shape Fac

Sample known size particles and calculate a velocity… Velocity = flight path / TOF PSLs and NH4NO3

Page 9: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

What’s wrong with this calibration?

Page 10: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Velocity calibration data should span maximum measurable range

Extrapolation of fit beyond data points can introduce errors

Critical data points

Page 11: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not
Page 12: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Slow vaporization and bounce Extrapolation of size calibration curve fit

Do not “over-interpret” small and large sizes.

Page 13: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Size Calibration Summary • Work hard at getting data at the small and large

sizes. • Use caution extrapolating velocity curve. • Document lens pressure (flow) during a calibration. • Watch for a lens pressure change when connecting

the AMS inlet to a sampling system. • Our calibration is based around determining velocity

but we don’t use velocity for anything… • Recommend using peak of mass distribution to find

pToF. Advanced discussion topic

Page 14: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

IE Calibration • We have two approaches

– Single particle (requires DMA) – Total mass based method (requires DMA and CPC)

• In principal the SP method is probably better.

• In practice there are problems with both

approaches so it is recommended to do both.

Page 15: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Single Particle Based IE Calibration

Time

Sig

nal

Single particle pulses

Particle threshold set above single ion level

Single ions above electronic noise level

Time

Am

ps (C

oulo

mbs

/tim

e)

Average single ion pulse

Average single particle pulse = Ions per particle (IPP)

Ionization Efficiency = IPP/Molecules per Particle

Page 16: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

AMS DMA

CPC

Atomizer

Aerosol Diluter

Setup for CPC-Mass Based IE Determination

By-pass

Input Mass = ρ x Volume(size) x Number Instrument Response

filter

Plot Instrument Response vs Input Mass

Drier

Mixing Tube

Very Important! Very Important!!

Page 17: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

NH4NO3 Ionization Efficiency Calibration

NO3 = NO+ + NO2+ NH4 = NH+ + NH2

+ + NH3+

NO3 NH4

NO

3 Si

gnal

NH

4 Si

gnal

• Multi-point calibration, tests linearity of system • Ratio of slopes directly measures RIE NH4

Page 18: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

IE Calibration • Issues with mass based method:

– DMA-multiple charged diameters. • Use pTOF to determine contribution of larger diameters.

– Lens transmission limitations. – Requires a CPC. – Accuracy limited by CPC.

Page 19: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

CPC-Mass Based IE Determination

Lens transmission limitation

Multiple charged diameters N

itrat

e Si

gnal

Calculated Mass

Lets discuss the units on this plot…

Page 20: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Minimizing multiple charged mobility diameters by decreasing NH4NO3 atomizer concentration

5 mM solution recommended

Page 21: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

BFSP - IE Calibration

• Issues with single particle approach: – IPP can be low, not much bigger than a single ion,

use larger diameter particles (>400 nm). – Need faster acquisition to time resolve entire

vaporization event, increase TOF extraction rate. – Its only a single point calibration.

Could look at different sizes, but limited by DMA – Selection of threshold is critical.

We use a default of 10, should be ok…

Page 22: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Last year we proposed the CPC-mass based method to be integrated with

ToF DAQ

Now in V4.x ToF AMS DAQ

This will help simplify and standardize IE calibrations.

Page 23: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

“Ionization Efficiency” - Units

Sign

al in

Ions

/ se

c

the AMS detects ions which we want to relate to particle mass

molecules / sec pico-grams / sec

Page 24: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

New proposed units for a “mass based Ionization Efficiency”

mIE

Ions / pg

• Has a more intuitive meaning.

• Can be more easily related to other instrument.

Page 25: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Typical Values IENO3 (ions/molec.) mIENO3 (ions/pg)

Q-AMS C-TOF-AMS 10-6 104

V-mode (HR-AMS) 10-7 1000

W-mode (HR-AMS) 10-8

100

ACSM 10-7 1000

ccTOF AMS ? ?

Order of magnitude relationship between IE and mIE for different

mass spec systems.

Page 26: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

Lens Alignment

• This is often not checked. • Generally, this does not need to be adjusted

but if its off and not checked your data will likely be invalid.

• Recommend checking this after shipping the AMS.

Page 27: PowerPoint Presentationcires.colorado.edu/jimenez-group/UsrMtgs/UsersMtg... · • This is often not checked. • Generally, this does not need to be adjusted but if its off and not

END PRESENTATION