User manual M07716 Issue 1 October 2017 Clip SGD Portable single Gas Detector

Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

  • Upload
    donhu

  • View
    215

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

User manual

M07716Issue 1 October 2017

Clip SGD

Portable single Gas Detector

Page 2: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

⚠ WARNING• ClipSGDisahazardousareacertifiedgasdetectorandassuchmustbeoperatedandmaintainedinstrictaccordancewiththeinstructions,warningsandlabelinformationincludedinthemanual.

• Beforeoperatingthisdevice,pleasereadcarefullyandunderstandallinstructionsinthemanual.

• Understandthescreendisplayandalarmwarningspriortouse.• Beforeuse,pleasechecktheactivationdate,andiftheactivationdatehasexpiredpleasedonotusethedevice.

• Beforeuseensurethatthedeviceisingoodconditionandtheenclosure,thedisplaywindow,alarmLEDwindowsandsounderapertureareundamagedandclearfromcontaminants.

• Beforeuseperformtheself-testoftheLED’s,sounderandvibrator.• Bumptestthedeviceregularlybyapplyingagasconcentrationexceedingthealarmsetpoint.Itisrecommendedthatabumptestisperformedatleastevery90days.

• FortheO2detectorperformacalibrationevery30daysinafreshairenvironment.• Ifthedevicefailsabumptest,calibrationorself-testpleasedonotusethedevice.• Donotdisassembleorsubstitutecomponents,oroperatebeyondtheintendedrangeofuseasthismayimpairintrinsicsafety,invalidatesafetycertificationandwillvoidthewarranty.

• DonotattempttoreplacethebatteryorsensorastheClipSGDisdesignedtobemaintenancefreeanddisposable.Changingthebatteryorsensormayimpairintrinsicsafety,invalidatesafetycertificationandwillvoidthewarranty.

• UseIRcommunicationsonlyinasafeareathatisfreefromhazardousgases• TopreventthegenerationofstaticelectricitydonotcleanorrubtheLCDscreenofthedevicewithadryclothorhandsinahazardousareaenvironment

• Performcleaningandmaintenanceofthedeviceinasafeareaonlythatisfreefromhazardousgases.

• TheClipSGDisshippedwithdefaultalarmlevelsettings.Itistheusersresponsibilitytoensurethealarmlevelsettingsareappropriateforthespecificapplication;thesecanbechangedasrequiredviatheIR_Link_PCSoftware.

• Gasconcentrationmeasurementsbythesensorcanvarybasedontheenvironment(temperature,pressureandhumidity).ThereforecalibrationofClipSGDshouldbeperformedinthesame(orsimilar)environmentofthedevice’sactualuse.

• Severevibrationorshocktothedevicemaycauseasuddengasreadingchange.PleaseuseClipSGDafterthevalueofgasconcentrationhasstabilized.

• ExcessiveshocktoClipSGDcancausethedeviceand/orsensortomalfunction.• Ifthetemperaturechangessharplyduringuseofthedevice(forexamplemovingfromindoorstooutdoors),thevalueofthemeasuredgasconcentrationcansuddenlychange.PleaseusetheClipSGDafterthegasconcentrationvaluehasstabilized.

• Thedeviceisintendedforuseinnormalatmosphericconditions,pressure80kPa(0.8bar)to110kPa(1.1bar);andairwithnormaloxygencontent,typically21%v/v(volume/volume).

Page 3: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

⚠ CAUTION• Thisdeviceisnotameasurementdevice,butagasdetector.• Cleandetectorswithasoftclothanddonotusechemicalsubstancesforcleaning.• Toensurea24monthsbatterylifeisattaineditisimportanttoavoidthefollowingactivitiesasthiswilldepletethebatterylifetimetolessthan24months.1. Allowing the detector to be in gas alarm frequently or for prolonged periods (Normal Alarm Use: No more than 4 minutes alarm time per day)2. Operating the detector user button frequently.3. Connecting the detector with the Clip SGD IR Link frequently, except for regular bump

testing. • Viewaserialnumberonthelabelatthebacksideofthedevice.(ex,20170101)

1. The serial number indicates below.

ex)SG0101001 2017(Year) 01(Months) 01(Day) 001(ManufactureOrder)

Page 4: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols
Page 5: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

Table of Contents1. DiagramandLCD............................................... 6

2. Activation........................................................... 7

3. Mode.................................................................. 8

3.1 GasMeasurementMode........................... 8

3.2 DisplayDeviseSettings.............................. 8

3.3 AlarmsandAlerts...................................... 8

4. EventLog......................................................... 11

5. Calibration........................................................ 12

5.1 FreshAirCalibration................................ 12

5.2 StandardGasCalibration........................ 13

5.3 ReturntoGasMeasurementMode......... 14

6. Specification..................................................... 15

7. SelfTestandBumpTest.................................... 16

7.1 Selftest................................................... 16

7.2 BumpTest............................................... 16

8. Certificates....................................................... 18

Appendix1ClipSGDDockingStation................... 19

Appendix2IRLinkUserGuide............................... 26

LimitedWarranty.................................................... 29

Page 6: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

6

Clip SGD

1. Diagram and LCD

1. IRPort2. LCDdisplay3. AlarmLED4. Buzzer5. Button6. Gassensor7. GasType

LCD display symbols

ALARM Alarmcondition M Monthsremaining

1Lowalarm D Daysremaining

2Highalarm T Timeremaining(hours)

v Testsuccess MAX Maximumpeakvalue

x Testfailure M NI Minimumpeakvalue

FreshaircalibrationLEL

PPMVOL

%

%MeasurementUnit

StandardgascalibrationLowbatterylife(Lessthan30days)

1

2

45

6

7

3

3

3

MAX

LELPPMVOL

%

%

M NI

ALARM 1 V

M

D

T

X2

Page 7: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

7

Clip SGD

2. Activation

Only activate in a safe environment.Pressthebuttonfor3secs;gastypeandfirmwareversion(e.g.v2.2)willbedisplayed.Followingthe10secscountdown,thedevicewillbefullyactivated,the V iconwillflashonthedisplayscreenandthedevicewillbeinGasMeasurementmode.Afterturningonthedevice,theremaininglifeisdisplayed;23monthsoriginally.Aftertheremaininglifereaches1month,30daysremainingisdisplayed.When1dayisleft,24hours(T)remainingisdisplayed.

⚠ CAUTIONBeforeuse,checktheactivatebydateonthebox.Iftheactivationdateispast,donotactivatethedevice.ShelfLifewithin6monthsfrommanufacture.

V

%

D

X

MAX

LELPPMVOL

%

%

M NI

ALARM 1 V

M

D

T

X2

%

T

⚠ CAUTIONAppropriatecalibrationofthedeviceisrequiredpriortooperation.Alwaysensurethatthedevicerespondscorrectlytotherelevantgas.Verifythatforeignmaterialsthatcouldinterferewiththedetectionofgasarenotblockingthesensor.

Intheeventthatactivationofthedevicefails,the X iconwillappearonthedisplayandGasMeasurementmodewillnotbeentered.Inthiscase,performcalibrationorcontactauthorizedresellerorCrowconDetectionInstrumentscustomersupportat+44(0)1235557711forrepair/returninformation.

Page 8: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

8

Clip SGD

3. Mode

3.1 Gas Measurement Mode

FollowingactivationthedevicewilloperateinGasMeasurementMode.Theremainingsensorlife(defaultsetting)orgasconcentration(configurableoption)willbedisplayedonthescreen.Oxygenconcentrationisdisplayedinpercentbyvolume(%Vol)andtoxicconcentrationisdisplayedinpartspermillion(PPM).

3.2 Display Device Settings

InGasMeasurementModetoviewthedevicesettingspressthebuttontocyclethroughthesettings.Thesettingswillbedisplayedinthefollowingorderuponeachbuttonpress.

VOL% VOL%

M NIMAX

VOL%

MAX

VOL%

M NI

DetectionmodeMinlevelseen(onlyforoxygen)

MaxlevelseenClearmin/max

levels

M

VOL%

1

VOL%

2V

VOL%

Remaininglife( T , D , M )

1stalarmsetpoint

2ndalarmsetpoint

Firmwareversion

Calibrationconcentration

AtthelaststepthedevicewillreturntoGasMeasurementMode.

3.3 Alarms and Alerts

Whenagasconcentrationexceedstheconfiguredalarmsetpoints,1or 2 willbe

displayedonthescreen,thedevicewillvibrate,alarmLEDswillflash,andthebuzzerwillsound.Thealarmswillstopwhenthedeviceismovedtoacleanairenvironment.

%

M

Page 9: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

9

Clip SGD

Thetablebelowshowsthe‘alarms’and‘alerts’thatmaybeshownonthedisplayduringoperation.

Alarm AlarmStandard ShownonLCDDisplay

AlarmandVibrationDisplay

LowAlarm Exceeds1stAlarmsetpoint

1icon&

concentration

Buzzer,LED

Vibration

HighAlarm Exceeds2ndAlarmsetpoint 2 icon&

concentration

Buzzer,LED

Vibration

Remaininglife Below30daysicon

Lifetimeexpiration

Past24monthsServicelifeisover(ReplacewithanewClipSGD)

Testfailure Failureofsensortest/calibration

Display X iconwithbeeping

Batterytest Lowpower

Bumptest Bumptestperiod

V X Pressthepushbuttontoturnoffthealarm

Calibration Calibrationperiod

Pressthepushbuttontoturnoffthealarm

⚠ CAUTIONDefaultalarmsetpointsarebasedoncommoninternationalstandards.Itisuserresponsibilitytoensuretheconfiguredalarmsetpointsareinaccordancewithlocalregulationsandsiteprocedures.

Page 10: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

10

Clip SGD

Default Alarm Set Point

Gas Type CO H2S O2

Part Number

CL-C-25 CL-C-30 CL-C-35 CL-H-2 CL-H-5 CL-H-10 CL-O-19 CL-O-19.5

Lower Alarm Level

25ppm 30ppm 35ppm 2ppm 5ppm 10ppm 19% 19.5%

Upper Alarm Level

50ppm 100ppm 50ppm 5ppm 10ppm 15ppm 23.5% 23.5%

Page 11: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

11

Clip SGD

4. Event LogThelast30eventsarestoredonadevice.Once30eventsarestored,thelogeventsareremovedautomaticallyintheorderstartingatEvent1.ThestoredlogeventscanbedownloadedusingtheIRLinkortheDockingStation(refertoAppendix1and2).Eachalarmeventrecordsthefollowing:• Alarmtype(1stor2nd)• Alarmconcentrationinppmor%• Peakconcentration

Event log

Page 12: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

12

Clip SGD

5. Calibration

5.1 Fresh Air Calibration

ToperformaFreshAirCalibrationpressthebuttontonavigatetoCalibrationMode,thecalibrationiconwillbedisplayedonthescreen .• Pressandholdthebuttonfor5seconds,‘CAL’andthefreshaircalibrationiconwillbeseenonthescreen .

• Pressandholdthebuttonfor3secondstoinitiatetheFreshAirCalibration.Whenthecalibrationbegins,acountdown(startingat10)willbedisplayedonthescreen.

VOL%

V

%

PressESCtoshowGasMeasuremntmode.Oncecompleted, V iconwillappearontheLCD.

%

Fresh Air Calibration

Standard Gas Calibration

⚠ CAUTIONForO2unitsonly.Initialcalibrationisperformedonalldevicespriortoshipment.Oncereceived,calibrationshouldbeperformedmonthly(orquarterly)dependingonfrequencyofuse.

⚠ CAUTIONCalibrationshouldbeperformedinafresh-airenvironmentthatisfreefromtheinfluenceofother

gases(calibrationisassumedtobeperformedinanenvironmentwithaconcentrationof20.9%oxygen).Itisalsorecommendedthatcalibrationshouldnotbeperformedinaconfinedspace.

Page 13: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

13

Clip SGD

%

X

%

X Ifcalibrationfails, X iconwillappearonthedisplay.Ifthiscontinues,pleasecontactthesalesrepresentativeorCrowconcustomersupport.

5.2 Standard Gas Calibration

ToperformaStandardGasCalibration,usingthesuppliedcalibrationcapconnecttheappropriategascylindertothedevice.• PressthebuttontonavigatetoCalibrationMode,thecalibrationiconwillbeshown, .• Pressandholdthebuttonfor5seconds,‘CAL’andthefreshaircalibrationiconwillbedisplayedonthescreen, .

• Pressthebuttononceuntilthecalibrationiconisshown, .• Pressandholdthebuttonfor3secondstoinitiatetheStandardCalibration.Whenthecalibrationbegins,acountdown(60secondsormoredependingonthegastype)willappearonthescreen.

% VOL%

VOL%

V

%

V

%

Oncecompleted, V iconwillappearforseveralsecondsonthedisplay.ThedevicewillthenreturntoGasMeasurementmode.

%

X

%

X Ifcalibrationfails, X iconwillappearonthedisplay.Ifthiscontinues,pleasecontacttheauthorisedresellerortheCrowconcustomersupportteam.

Page 14: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

14

Clip SGD

5.3 Return to Gas Measurement Mode

ToreturntoGasMeasurementModefromCalibrationMode,pressthebuttontocyclethroughtheFreshAirCalibration andStandardCalibration screensuntil‘ESC’isshownonthedisplay.Whilst‘ESC’isshownonthedisplaypressandholdthebuttonfor3seconds,thedevicewillreturntotheGasMeasurementscreen.

VOL%

%

Calibration concentration.

Gas O2 CO H2S

Concentration 18%Vol 100ppm 25ppm

Page 15: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

15

Clip SGD

6. Specification

Model ClipSGD

Measure Gas O2 CO H2S

Sensor range 0-30%Vol 0-300ppm 0-100ppm

Sensor Type Electrochemical

Measurement Diffusiontype

Display LCDdisplay

Audible 90dBat10cm

Warning Lamp RedFlashingLEDs(Light-EmittingDiode)

Vibration VibrationAlarm

Battery Manufacture:Vitzrocell/P/N:SB-AA02(P)/System:LithiumPrimaryBatteryNominalvoltage:3.6V/Nominalcapacity:1.2Ah

Temperature & Humidity

-40°Cto+50°C(forToxic)/-35°Cto+50°C(forO2)5%to95%RH(non-condensing)

Ingress protection IP65andIP67

Case RubberEnclosure

Accessories Calibrationcap,manual,calibrationandtestcertificate

Option IRLink,DockingStation

Size & Weight Size:54mm(W)x91mm(H)x32mm(D)Weight:93g(Toxic),104g(O2)(Battery,clipincluded)

Operating Life 24months,basedon4minutesofalarmperday(>2900Alarmminutes)

Event Log Recent30alarms

Approval SeeCertificationonpage16

Page 16: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

16

Clip SGD

7. Self Test and Bump Test

7.1 Self Test

TheClipSGDprovidesaSelfTestfunction.ThedefaultintervaloftheSelfTestreminderis20hrs.UsetheIRLinkorDockingStationtoconfiguretheintervalbetween1hrand20hrs,orsetto“off”.WhentheconfiguredSelfTestisdue,“STS”willflashonthescreen(themessagewillcontinuetoflashuntiltheuserperformsasuccessfulSelfTest).

MAX

LELPPMVOL

%

%

M NI

ALARM 1 V

M

D

T

X2

PPM

1

PPM

2

• PressthebuttontoinitiatetheSelfTestfunction.Thedevicewillperformaselftestbyoperatingthebuzzer,LED’s,vibratorandLCD.Itwillalsoshowtheconfiguredlowandhighalarmsetpointsonthedisplay.Afterthetestiscompleted,“END”willbedisplayedonthescreen.(Theuserisrequiredtocheckthattheitemsoperatedintheselftest,performedasexpected).

7.2 Bump Test

ThedevicecanbeconfiguredtoprovideanonscreenBumpTestreminder.ThedefaultsettingfortheBumpTestreminderis“off”,noBumpTestreminderwillbedisplayed.IfaBumpTestreminderintervalisrequireditcanbeconfiguredusingtheIRLinkorDockingStation.Theintervalcanbeconfiguredfrom1dayto365days.WhenconfiguredandaBumpTestisdue,the“bts”messagebedisplayedandwillflashonthescreen.

Page 17: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

17

Clip SGD

ALARM 1 V2

VOL%

ALARM 12

VOL%

ALARM 12

VOL%

• Pressandholdthebuttonfor3seconds,andthe“tst”messagewillbedisplayedfor45seconds(tocancel,pressthebuttonforonesecond).

• Applythegasusingthecalibrationcapandgastubewithinthe45secondperiod.IfaBumpTestisnotundertakenwithinthisperiodthe“bts”messagewillflashuntilasuccessfultestisconducted.

Aftertheselectedgasisappliedandifthetestissuccessful,the“SUC”messagewiththe V iconwillbedisplayedfor3secondsthenimmediatelyreturntoGasMeasurementMode.Thecalibrationcapandgastubecanthenberemoved.

Ifthetestfails,the“FA”messagewith X iconwillbedisplayedand“bts”messagewillflashuntilasuccessfultestisconducted.

X

Page 18: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

18

Clip SGD

8. CertificatesTheSGDClipmeetsthefollowingcertificationsandstandards

IECEx: ExiaIICT4Ga➀ ➁ ➂ ➃ ➄

1:Explosionprotected2:ProtectionConcept3:GasGroup4:TemperatureClassification5:Equipmentprotectionlevel

ExiaIICT4GaClassI,Zone0,AExiaIICT4GaClassI,Division1,GroupsA,B,C,D,T4

ATEX: II1GExiaIICT4GaKRH17ATEX0013Directive2014/34/EU

Standards Theelectricalapparatusandanyacceptablevariationstoitspecifiedinthescheduleofthiscertificateandtheidentifieddocuments,wasfoundtocomplywiththefollowingstandards:

IEC60079-0:2011Ed6IEC60079-11:2011Ed6

UL61010-1,Ed.3UL913,Ed.8UL60079-0,Ed.6UL60079-11,Ed.6

C22.2No.60079-0:2015C22.2No.60079-11:2014C22.2No.61010-1-12:2012

EN60079-0:2012+A11:2013EN60079-11:2012

Compliance ElectromagneticCompatibilityDirective2014/30/EUManufacturingApproval:

ThedetectormanufactureriscertifiedcompliantwithISO9001:2000provisions

Page 19: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

19

Clip SGD

Appendix 1 Clip SGD Docking Station User Manual

Clip SGD Docking Station ConfigurationUserscanchangesettings,upgradefirmware,anddownloadlogeventsthroughtheDockingStationIRlink(seeAppendix2).

Page 20: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

20

Clip SGD

Figure 1: Clip SGD Dock PC Application

IR / USB DiskChoseeitherIRLinkorUSBSticktotransferconfigurationtoandfromDockingStation.• Click‘IR’➀toconfigureDockingStationviatheClipSGDIRLink.• Click‘USBDisk’➁toselectUSBStickconnectedtoPCandtransferconfigurationtotheUSBStick–thiscanthenbeinsertedintotheDockingStationandtheconfigurationuploaded.

• InfoRead/Write:Selectto‘Read’or‘Write’configuration➂toDockingStationviaIRlinkortoUSBStick

Docking Station ThelayoutoftheClipSGDDockPCApplicationisshownaboveFigure1.• TointeractwithDockingStationviatheClipSGDIRLink,itmustbeturnedon.• EnsurethePChastheClipSGDDockPCApplicationinstalledandconnecttheClipSGDIRLinktothePC.

• PlacethetopsideoftheClipSGDIRLinkfacedownindetectorbaylabelled“Unit1”.N.B.:TheIRlinkonlyinteractswithdetectorbaylabelled“Unit1”.

3

1 2

Page 21: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

21

Clip SGD

Figure 2

Docking Info• SerialS/N:Factoryserialnumber• Version:DockingStationfirmwareversion• AutoPower-Off:DockingStationautomaticpower-offtime(Max3600seconds)• MaximumBumpTime:Bumptesttime(seconds)–Default:30/Min:20/Max:120• Location:Configuretheinstalledlocation• GasExpirationDate:Entertheexpirationdateofthegascylinder.• GasBottleLot#:Enterthelotnumberofthegascylinder.• InfoRead/Write:RetrieveorwritesettingsfromDockingStation

Enable/Disable Function Checkboxtoenableordisablefunctions;atickmeansenabled.• Calibrate:enables/disablesDockingStation‘Calibration’button.• Bumptest:enables/disablesDockingStation‘BumpTest’button.• UpdateSGDConfiguration:enables/disablesClipSGDconfigurationupdatesduringbumptestorcalibration.

• UpdateSGDFirmware:enables/disablesClipSGDfirmwareupdateduringbumptestorcalibration.

Docking Firmware Upgrade• AllowsDockingStationfirmwaretobeupgraded–click‘Browse’tonavigatetofile.• Click‘Write’todownloadfile.

Page 22: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

22

Clip SGD

Figure 3

SGD Configuration OptionsAllowstheconfigurationoftheClipSGDtobemodifiedwhilsttheDockingStationperformsabumptestoracalibration.• UserID:EnterauserIDifrequired• CalibrationInterval(0-365days):Settheintervalofthecalibrationalert• BumpInterval(0-365days):Settheintervalofthebumptestalert• SelfTestInterval(0-24hours):Settheintervaloftheselftestreminder.• LowAlarmandHighAlarm:Settherequiredlowandhighlevelgasalarms.• Display:ChooseforClipSGDtodisplayeithergasreading(“Sensor”)orinstrumentliferemaining(“Life”)

SGD Options Info Read/Write• Read:RetrievethestoredsettingsfromClipSGDDockingStation• Write:WritetheadjustedsettingstotheClipSGDDockingStation

Page 23: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

23

Clip SGD

Figure 4

Gas ConcentrationAllowsthegaslevelsinthecylinderconnectedtotheClipSGDDockingStationtobeentered• Choosegasconcentrationrequiredforcalibration,defaultsettingsareasfollows:

H2S:25ppm,CO:100ppm,O2:18%Vol

SGD Firmware Upgrade• AllowstheClipSGDfirmwareupgradefiletobespecified–click‘Browse’tonavigatetofile.

• Click‘Write’todownloadfile.

IR / USB DiskChoseeitherIRLinkorUSBSticktotransferconfigurationtoandfromDockingstation• Click‘IR’toconfiguredockingstationviatheClipSGDIRLink.• Click‘USBDisk’toselectUSBStickconnectedtoPCandtransferconfigurationtotheUSBStick–thiscanthenbeinsertedintotheDockingStationandtheconfigurationuploaded.

• SGDOptionInfoRead/Write:Selectto‘Read’or‘Write’configurationtoDockingStationviaIRlinkortoUSBStick

Page 24: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

24

Clip SGD

Figure 5

Logs• Refresh:Viewavailablelogfilesfromdockingstation• SelectAll:Selectalllogfilesavailableinthedockingstation• LocalStore:SaveselectedlogfilestoPC–thelogfilenamewillbesavedbasedontheserialnumber

Page 25: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

25

Clip SGD

Figure 6 Docking Station Log

Figure 7 Clip SGD Event Log

Page 26: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

26

Clip SGD

Appendix 2IR Link User Guide

Use IR Link to change settings, upgrade firmware, download log events, and perform calibration

Getting startedTointeractwithClipSGD,thesoftwaremustbeinstalledonyourcomputer.PlugintheIRLinkcable.Then,placethetopsideoftheSGDfacedowntotheIRLinkLCD.

4

2

1

3

Page 27: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

27

Clip SGD

Software overview

1 Tab menusFile• Load:Opentheinstalledsettings.(*cfg)• Save:Savethecurrentsettings.(*cfg)• Exit:Closetheprogram.

Tools• Calibration:PerformtheZERO/SPANcalibration.• LogRead:Retrievethelogevents.• LogErase:Eraselogevents.• SelfTest:TestifLED,LCD,Buzzer,vibration,batteryandtemperatureworkproperly.• FirmwareUpgrade:Forthelatestfirmwareversion,gotowww.crowcon.comorcontactyourlocalserviceagent.Afterafirmwareupdate,thedetectorMUSTberecalibrated.

• FactoryDefault:Removealldataandrestoretheinitialfactorysettings.Afterfactorysettingsarerestored,thedetectormustberecalibrated.

Help• About:Displayabriefcompanyprofile.

2 Tool Interface• READ:Retrievesettings• WRITE:Updateandadjustsettings.• LOG:Displaythesavedlogevents.• CALIBRATION:PerformtheZEROandSPANcalibration• FIRMWAREUPGRADE:Upgradefirmware.

3 IR Link Interface• InordertointeractwiththeClipSGDusingtheIRLink,click“OPEN”.WhentheClipSGDsettingshavebeensuccessfullyretrieveduserswillsee“ReadComplete”.

4 Clip SGD Interface• SerialNumber:Displayapresetserialnumber.• FirmwareVersion:Displayafirmwareversion.• UserID/Message:Userscanprogramadesiredmessage/ID.• Display:Userscanchoosetodisplayeithertherealtime“SensorReading”or“SensorLifeRemaining”

• Stealth(Option):Userscanturnoffthebuzzer,LED,andvibration.

Page 28: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

28

Clip SGD

• SelfTest(Hour):SetanintervaloftheSelfTestalert.• SensorType:Displaythesensortypeofthedevice.• LowAlarm:Adjustthelowalarmsetpointincompliancewithinternationalorlocalstandards.

• HighAlarm:Adjustthehighalarmsetpointincompliancewithinternationalorlocalstandards

• CalConcentration:Setastandardcalibrationvalue.• CalTime(sec):Setastandardcalibrationtime.• CalInterval(day):Setanintervalofthecalibrationalert.• BumpInterval(day):Setanintervalofthebumptestalert.

Page 29: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

29

Clip SGD

Warranty

1.1. Limited Warranty

CrowconwarrantsthisproducttobefreefromdefectsinmaterialandworkmanshipundernormaluseandserviceforaperiodoftwoyearsbeginninguponthedateofactivationforallClipSGDproducts.Thiswarrantyisvalidonlyifthepersonalmonitorisactivatedbythe‘activateby’dateontheexternalpackaging.Thiswarrantyextendsonlytothesaleofnewandunusedproductstotheoriginalbuyer.Shouldthebatterydepletepriortothe24monthsofactivation,thiswilldemonstratethatthemonitor(s)haveexperienced>2900alarmminutesandwillthereforebeclassedasoutsidewarranty.

Warranty ProcedureTofacilitateefficientprocessingofanyclaim,contactyourlocalCrowconagent/distributor,aCrowconregionalofficeorourglobalcustomersupportteam(Englishworkinglanguage)on+44(0)1235557711orcustomersupport@crowcon.comtoobtainareturnsformforidentificationandtraceabilitypurposes.Thisformmaybedownloadedfromourwebsite‘crowconsupport.com’andrequiresthefollowinginformation:

•Yourcompanyname,contactname,phonenumberandemailaddress.•Descriptionandquantityofgoodsbeingreturned,includinganyaccessories.•Instrumentserialnumber(s).•Reasonforreturn.

ProductswillnotbeacceptedforwarrantywithoutaCrowconReturnsNumber(CRN).Itisessentialthattheaddresslabelissecurelyattachedtotheouterpackagingofthereturnedgoods.Theguaranteewillberenderedinvalidiftheinstrumentisfoundtohavebeenaltered,incorrectlystored,modified,dismantled,tamperedwith,orhasnotusedCrowconsparesforreplacementpartsincludingsensors)orhasbeenservicedorrepairedbyanypartynotauthorisedandcertifiedbyCrowcontodoso.Thewarrantydoesnotcovermisuseorabuseoftheunitincludinguseoutsideofspecifiedlimits.

1.2. Warranty Disclaimer

Crowconacceptnoliabilityforconsequentialorindirectlossordamagehowsoeverarising(includinganylossordamagearisingoutoftheuseoftheinstrument)andallliabilityinrespectofanythirdpartyisexpresslyexcluded.Thiswarrantydoesnotcovertheaccuracyofthecalibrationoftheunitorthecosmeticfinishoftheproduct.Theunitmustbemaintainedinaccordancewiththeinstructionsinthismanual.Thewarrantyonreplacementconsumableitemssuppliedunderwarrantytoreplacefaultyitems,willbelimitedtotheunexpiredwarrantyoftheoriginalsupplieditem.Crowconreservestherighttodetermineareducedwarrantyperiod,ordeclineawarrantyperiodforanysensorsuppliedforuseinanenvironmentorforanapplicationknowntocarryriskofdegradationordamagetothe

Page 30: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

30

Clip SGD

sensor.Ourliabilityinrespectofdefectiveequipmentshallbelimitedtotheobligationssetoutintheguaranteeandanyextendedwarranty,conditionorstatement,expressorimpliedstatutoryorotherwiseastothemerchantablequalityofourequipmentoritsfitnessforanyparticularpurposeisexcludedexceptasprohibitedbystatute.Thisguaranteeshallnotaffectacustomer’sstatutoryrights.

Crowconreservestherighttoapplyahandlingandcarriagechargewherebyunitsreturnedasfaulty,arefoundtorequireonlynormalcalibrationorservicing,whichthecustomerthendeclinestoproceedwith.

For warranty and technical support enquiries please contact:Customer SupportTel: +44 (0) 1235 557711Fax: +44 (0) 1235 557722Email: [email protected]

Page 31: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols
Page 32: Portable single Gas Detector - crowcon.com · 6 Clip SGD 1. Diagram and LCD 1. IR Port 2. LCD display 3. Alarm LED 4. Buzzer 5. Button 6. Gas sensor 7. Gas Type LCD display symbols

Contact/www.crowcon.com/[email protected]/44(0)1235-557711