Upload
anne-claire
View
214
Download
2
Embed Size (px)
Citation preview
Dynamic Article LinksC<JAAS
Cite this: J. Anal. At. Spectrom., 2011, 26, 1474
www.rsc.org/jaas PAPER
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online / Journal Homepage / Table of Contents for this issue
Polyatomic interferences in plutonium determination in the femtogram rangeby double-focusing sector-field ICP-MS
Fabien Pointurier,* Am�elie Hubert, Anne-Laure Faur�e, Philippe H�emet and Anne-Claire Pottin
Received 30th July 2010, Accepted 7th March 2011
DOI: 10.1039/c0ja00097c
The determination of very low concentrations of plutonium, in the femtogram per millilitre range, in
environmental samples using ICP-MS can be interfered by polyatomic species containing heavy
elements, from hafnium to bismuth. Mass-to-charge ratios of these polyatomic species are very close to
the ones of the plutonium isotopes and cannot be separated from them in the low resolution mode
currently used for trace analysis. In this paper, we present an evaluation of the extent and impact of
these interferences on the background at masses of plutonium isotopes 239Pu and 240Pu, based on
measurement of environmental samples for three years using a double-focusing sector-field ICP-MS.
We demonstrate that these molecular interferences, especially the ones involving lead, mercury, and
iridium through species PbO2+, ArHg+, IrO3
+, must be considered as the additional background
induced by these interferences (on average for all interferences respectively �1.2 and �1.9 counts s�1 at
239 and 240 atomic mass units) is often higher than the instrumental background measured with
deionised water acidified at 2% by ultra-pure grade HNO3 (�0.4 counts s�1), and are generally of the
same order as other usual sources of background for low-level plutonium analysis by ICP-MS, i.e. 242Pu
tracer isotopic impurities, 238U hydrides and peak tail. Thus, if not corrected, these polyatomic
interferences may lead to false detection of femtogram amounts of plutonium or overestimation of
results. The extents of formation of the main polyatomic species are highly variable from one analysis
to the other, over one or two orders of magnitude, depending on instrumental settings. Chemical
purification procedures are very efficient in eliminating the greater part of heavy elements present in the
initial samples. However, these procedures also bring heavy elements into the final sample solutions
through contamination by the atmosphere, glassware and reagents.
Introduction
Numerous examples can be found in the literature about ICP-MS
with best detection limits for plutonium isotopes in the femtogram
range (10�15 g) for quadrupole-based instruments1–4 and, more
often, for sector-field instruments.5–25 However, these instru-
mental detection limits are obtained with synthetic solutions and
the analytical detection limits obtained with real-life samples may
be considerably higher, generally because of a loss of sensitivity
due tomatrix effects or tomemory effects.10With the exception of238U hydrides atmass 239, the possible occurrence of interferences
due to polyatomic species with total masses equal to masses of
actinide isotopes is rarely considered. At the high end of the mass
range, the actinides have typically been considered as far removed
from the notorious polyatomic interferences (PIs) that tend to
plague the lighter elements. Nevertheless, some authors
mentioned and, in some cases studied, possible occurrence of
some PIs in the actinide mass range6,11,13,17,22,23,26–31 containing
CEA, DAM, DIF, F-91297 Arpajon Cedex, France. E-mail: [email protected]; Fax: +33 1 69 26 70 65; Tel: +33 1 69 26 49 17
1474 | J. Anal. At. Spectrom., 2011, 26, 1474–1480
a heavy element like platinum,28 lead,7,22,23,26,28 lanthanides,23,30,31
mercury22 or even of organic nature in samples with significant
organic content.13,29 Nygren and co-workers31 carried out an
extensive study on lanthanide phosphate molecular species which
interfere with plutonium isotopes extracted from soil and sedi-
ment samples. However, influence of PIs is regarded as very low,
significant only in special cases and a high confidence is put in the
chemical preparation to remove the interfering species. We
observed that with a sector-field ICP-MS which exhibits a very
high sensitivity (typically a few million counts s�1 per ng ml�1 of
plutonium), despite a neat chemical purification with many
precautions to limit contamination, pure background values
measured in the actinide mass range with real-life samples
(1 counts s�1 or more) are systematically higher than the instru-
mental background (IB) recorded with deionised water and ultra-
pure grade nitric acid (�0.4 counts s�1). We established that this
additional background is due to polyatomic species that are
combinations of one heavy element (HE) like Pb, Hg, W, Ir, Pt,
etc., with some of the most abundant atoms in the plasma (O, N,
H, Cl, Ar).32 In a previous paper, we proposed a method to esti-
mate and to correct these interferences,32 which, if not identified
This journal is ª The Royal Society of Chemistry 2011
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
and corrected, may lead to false detection of femtograms of
plutonium with high sensitivity ICP-MS. This method has been
now applied to real samples for about three years. So experience
has been gained on the degree of influence of polyatomic species
on the background in the actinide mass range.
The purpose of this paper is to present an evaluation of the
extent and impact of PIs on the background at masses of
plutonium isotopes 239Pu and 240Pu. This evaluation is based on
the measurement of plutonium in environmental samples for the
years 2007, 2008, and 2009, following the method developed in
the laboratory. For all measured samples, the sum of the back-
ground due to PIs at mass 239 and 240 is systematically
compared to other known sources of background (electronic
background, 238U hydrides, 238U peak tailing, tracer impurities,
etc.). Average values and variations of the extents of formation
of the most relevant polyatomic species are given. We also
present an evaluation of the efficiency of the chemical purifica-
tion procedure in the case of cotton wipe samples. This work
demonstrates that, for many of the analyses, total count rates
arising from polyatomic interfering species are higher than the
instrumental background and cannot be neglected for other
sources of background. The intention of this study is to show
that when appropriate conditions are encountered (very sensitive
instrument, samples with ultra-low plutonium contents), the
correction of polyatomic interference must be considered to
avoid false detections of plutonium at femtogram level.
Experimental
Instrumentation
For trace analyses of long-lived actinides, we routinely use
a double-focusing sector field ICP-MS (‘‘Axiom SC’’, VG
Elemental,Winsford, Cheshire, UK).Operation and optimisation
of this instrument for measurement of low amounts of actinides
had already been thoroughly described.9,12Briefly, this instrument
is equippedwith nickel cones, a platinum guard electrode andwith
the S-option, an additional primary pump that decreases vacuum
in the interface andprovides an increase in sensitivity by a factor of
three compared to a non-S option. A Teflon micro-concentric
nebuliser (‘‘PFA 100’’, Elemental Scientific, Inc., Omaha, NE,
USA) in the self-aspiration mode (�100 ml min�1) is used for
sample introduction. The nebuliser is routinely operated in
combination with two on-line spray chambers (‘‘cyclonic’’ and
Table 1 Instrument and data acquisition settings for the double focusingsector-field ICP-MS VG ‘‘Axiom’’
Instrument settingsForward power 1300 WCool gas 14 L min�1
Auxiliary gas 0.85–0.95 L min�1
Nebuliser gas 0.8–0.9 L min�1
Resolution 360Data acquisition settingsAcquisition mode E-scan, peak jumpingMonitored masses 233–247 (15 masses)Number of points per peak 1Dwell time per mass 100 msNumber of sweeps 100Number of repetitions 5Total integration time per mass 50 sTotal analysis time per sample 15 minutes
This journal is ª The Royal Society of Chemistry 2011
‘‘impact bead’’, Thermo-Fisher, LesUlis, France) cooled to 12 �C.Performance is daily optimised with respect to sensitivity, short
term stability (10 minutes), and background. The instrument is
operated in electric scanning by varying the accelerating voltage,
with the intensity of the magnetic field set to a fixed value. Typical
instrument and data acquisition settings for the acquisition of
plutonium isotopes are given in Table 1. Typical performance of
the instrument is given in Table 2. The instrumental detection limit
for plutonium isotopes calculated according to the 3s criteria
based on the background obtainedwith deionisedwater and ultra-
pure grade HNO3 at 2% is about 0.25 to 0.5 fg ml�1.
Sample preparation and purification procedure
All the acids used for sample preparation and purification are of
ultrapure grade (Merck, Darmstadt, Germany). 233U and 242Pu
are used as isotopic dilution (ID) tracers for uranium and
plutonium quantitative analysis. The standard solutions were
diluted (by weight) to obtain stock solutions between 0.2 and
2 ng ml�1. Heavy element (HE) solutions used for the determi-
nation of PI extents of formation and efficiency of the chemical
purification are mono-elemental certified solutions
(Spex, Longjumeau, France).
For each analysis, a preparative chemistry based on ion-
exchange chromatography is carried out on the samples. The
procedure in the case of cotton wipe samples, used to sample dust
inside nuclear facilities, is described hereafter. Samples are
transferred to Pyrex beakers, and 10 ml of concentrated HNO3
are added. The mixture is boiled on a hot plate at 120–150 �C) forat least 2 hours. During heating, the beaker is covered with
a watch glass to prevent significant evaporation. The mixture is
evaporated to dryness and the sample is then reduced to ash at
550 �C for 12 hours in an electric furnace to decompose organic
matter. Dissolution of the ashes is accomplished by repetition of
acid digestions: 20 ml of aqua regia, evaporation to dryness, 10
ml of concentrated HNO3 + 1 ml of H2O2, evaporation to
dryness, 10 ml of concentrated HCl and evaporation to dryness.
Isotopic dilution (ID) tracers 233U and 242Pu are added in
appropriate quantities to the sample solutions before separa-
tions. Spiked samples are evaporated to dryness and 20 ml of
concentrated nitric acid are added. Adjustment and redox state
stabilisation of plutonium are performed by the addition of
nitrite followed by evaporation. 20 ml of 8 N HNO3 are added
and the solution is passed through the conditioned ion-exchange
Table 2 Typical instrumental performance of the double-focusing sector-field ICP-MS VG ‘‘Axiom’’. Background count rates are evaluated fordeionised water acidified with HNO3 at 2%. Instrumental detection limitfor the plutonium isotopes are defined as 3 times the standard deviationover the instrumental blanks
Instrumental performance
Solution introduction rate �100 ml min�1
Sensitivity for 238U 2–4 � 106 counts s�1 (ng ml�1)�1
Background at masses 240–242 0.1–0.5 counts s�1
U hydride ratio 2–4 � 10�5
U oxide ratio 2–3%238U peak tail at mass 237 25–35 � 10�6
238U peak tail at mass 236 4–6 � 10�6
Detection limit for 239Pu and 240Pu 0.25–0.5 fg ml�1
J. Anal. At. Spectrom., 2011, 26, 1474–1480 | 1475
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
resin columns. The first plutonium purification is performed with
a 20 ml column filled with Dowex AG1X8 anion-exchange resin
(10 ml of 50/100 mesh resin at the bottom and 10 ml of 100/200
mesh resin on the top). The plutonium fraction is eluted with
HCl/NH4I. Further U–Pu purification is achieved by using a 2 ml
column filled with Dowex AG1X4 anion-exchange resin for Pu
or AG1X8 100/200 mesh for U. These resins are washed with 8 N
HNO3 (U fraction), 10 N HCl (Th fraction) and, finally, NH4I
(1.5%)–12 N HCl solution is added to the eluted Pu fraction. The
final solution is evaporated to dryness and recovered with 5 ml of
deionised water acidified by HNO3 at 2%. In addition, one or
more process blanks are prepared in the same conditions as the
samples for each series of samples.
Fig. 1 Example of comparison between the instrumental background
(IB), estimated from instrumental blanks made with 2% ultra-pure grade
nitric acid in deionised water, and the background induced by mercury-
containing (ArHg+) and lead-containing (PbO2+) polyatomic species. PB1
and PB2 are process blanks and S1, S2, S3, S4 are environmental samples
(cotton wipes).
Method of analysis including correction of polyatomic
interferences
Briefly, themethodnowused for over three years in the laboratory
for correction of PIs32 requires first the determination of the count
rates of at least one isotope of each HE present in the sample
solutions, from hafnium to bismuth. This allows identification of
the heavy elements whose concentrations in the sample solutions
may result in a significant contribution to the background at
masses of the plutonium isotopes, knowing the order of magni-
tudes of the extents of formation of the inducedbackground.Then
the extents of formation of the background induced by poly-
atomic species made with the HEs identified in the first step must
be measured as accurately as possible with mono-elemental
standard solutions using the same settings of the instrument and
the same analytical conditions, ideally just after the measurement
of the sample solutions. The background induced by polyatomic
interferences including heavy element HEj at mass i is given by:
ni;pa ¼Xj
fi;HEj$nHEj
(1)
where fi,HEjis the extent of formation of heavy element HEj at
mass i, and nHEjis the count rate of the heavy element HEj in the
sample solution.
Therefore, the net count rate at mass 239 is given by:
n0239 ¼ n239 � n239,IB � n239,pa � sUH + AS n238 � simp.239 n242(2)
where n239, n238, n242 are respectively the count rates at masses 239,
238 and 242 measured in the sample solutions, n239,IB the back-
ground at mass 239 in the instrumental blank (instrumental back-
ground), sUH + AS the extent of formation of 238U hydrides and the
peak tail of 238U atmass 239 (abundance sensitivity) and simp.239 the
impurity ratio of 239Pu of the 242Pu isotopic dilution tracer.
The uncertainty on the net count rate is:
sn0239
¼ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi�sn239
�2þ�sn239;IB
�2þ�sUHþAS$sn238
�2þðn238$ssUHþASÞ2þ�
simp:239$sn242�2þ�
n242$ssimp:239
�2þXj
�f239;HEj
$snHEj
�2
þXj
�nHEj
$sf239;HEj
�2s
and the detection limit according to the 3s criteria is:
DL239 ¼ 3$
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi�sn239;IB
�2þ�sUHþAS$sn238
�2þðn238$ssUHþASÞ2þ�
simp:239$sn2
s
1476 | J. Anal. At. Spectrom., 2011, 26, 1474–1480
A few practical precautions must be taken: (i) careful peak
centering during performance optimisation, for actinides but also
for heavy elements, to avoid bias of count rates; (ii) as stated
before, the same instrument settings or as close as possible when
the extents of formation are measured and during analysis of the
samples, to obtain relevant extents of formation; (iii) as HE
concentrations may be very high in order to estimate precisely the
extents of formation, it is necessary to consider a bias introduced
by saturation of the detector or acquisition carried out with
another detector (Faraday cup instead of ion counter) or another
counting mode (so-called ‘‘analog mode’’). It should also be
noted that the additional background induced by PIs can be in
some cases higher for the process blanks (PB) than for the
samples themselves. This unfortunately precludes correction by
a simple subtraction of the background count rates of the PB,
because it might lead to missing detection by overcorrecting the
background. An example is shown in Fig. 1: the additional
background at mass 239 due mainly in this case to lead- and
mercury-containing species is higher for one PB (‘‘PB2’’) than for
the samples. Therefore, in order to determine trace amounts of
plutonium in samples and in the PB, the most accurate method is
to correct from PIs, for the samples as well as for the PB.
Results and discussion
The extents of formation of interfering species obviously vary
from one analytical procedure to another, although no imme-
diate correlation was found with a specific parameter. First, we
examine the extents of formation of these interferences and their
(3)
ffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi42
�2þ�n242$ssimp:239
�2þXj
�f239;HEj
$snHEj
�2
þXj
�nHEj
$sf239;HEj
�2
(4)
This journal is ª The Royal Society of Chemistry 2011
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
range of variability. Starting from all the results obtained with
the above-mentioned analytical procedure for years 2007 to
2009, i.e. 137 low-level plutonium measurements, we compare
the interference induced background with the IB and other
sources of background (ID tracer impurities, 238U hydrides, 238U
peak tail), and we discuss the importance and benefits of the
correction of PIs. Lastly, we give some results relative to the
efficiency of the chemical purification procedure for the elimi-
nation of HEs.
Fig. 2 Range of the extents of formation for the background induced by
polyatomic species made with heavy elements fromHf to Bi at masses 239
(a) and 240 (b) recorded for the years 2007, 2008, and 2009.
The extents of formation of the background induced by
polyatomic interferences
The extents of formation are evaluated using mono-elemental
standard solutions containing only a given HE in appropriate
concentration. For instance, extent of formation of the back-
ground at mass 239, induced by the polyatomic species including
the heavy element E is defined by:
f239;E ¼ n239;std � n239;IB
nmE;std � nmE;IB
(5)
where n239,std and n239,IB are the count rates at mass 239 in the
standard mono-elemental solution and the instrumental blanks
(deionised water acidified at 2% by nitric acid), nmE,stdand nmE,IB
are the count rates recorded for the selected isotope of mass m of
the heavy element E.
We plot in Fig. 2 the range of the extents of formation at
masses 239 and 240 recorded for polyatomic species composed of
each HE, from hafnium to bismuth, from all analysis carried out
during the years 2007–2009. We observe that formation rates
vary widely—for some elements over several orders of magni-
tude—from one analysis to another. Obviously, instrumental
settings and plasma conditions are different and this may affect
recombination of atoms/ions in the plasma.
However, most of the extents of formation (for hafnium,
tantalum, tungsten, rhenium, platinum, gold, thallium, bismuth)
are very low, around 10�8 or even below. This means that the
background induced by these elements is negligible except if the
count rates of the HEs reach at least 107 counts s�1 meaning that
their concentrations are hundreds of mg l�1 or higher in the
sample solutions. This is highly unlikely after a chemical
purification.
However, the extents of formation of the background induced
at mass 239 by iridium (measured from isotope 193Ir) and
mercury (measured from 202Hg) can respectively be higher than
10�6 and 10�5. The involved polyatomic species are 191Ir16O3+ and
199Hg40Ar+. Similarly, the background induced at mass 240 by
mercury (measured from 202Hg) is between 10�5 and 10�4. In this
case, the corresponding polyatomic species is 200Hg40Ar+.
Consequently, concentrations in the mg l�1 range for these two
elements are enough to lead to a significant increase of the
background, at least in the order of one count s�1. Now, these
concentrations are occasionally really encountered in sample
solutions, even after chemical purification. Lead is a special case.
The extents of formation of the background induced by lead
(measured from the 208Pb isotope) at masses 239 and 240
respectively through the species 207Pb16O2+ and 208Pb16O2
+ are
quite low, ranging roughly from 10�9 to 2 � 10�7. However,
contrary to other HEs, lead is a ubiquitous element, relatively
This journal is ª The Royal Society of Chemistry 2011
abundant in some samples, in the atmosphere, and in glassware
and reagents used for chemical purifications as well. Its
concentrations in the sample solutions can thus reach tens of
mg l�1 or even hundreds of mg l�1, possibly giving rise to signifi-
cant background at masses 239 and 240. We report in Fig. 3 all
the extents of formation of background at mass 239 induced by
species made with mercury, iridium, and lead, measured for the
years 2007–2009. We observe that the extents of formation of the
background induced by these elements vary respectively from
6 � 10�6 to 6 � 10�5 for mercury, from 2 � 10�7 to 4 � 10�6 for
iridium, and from 7� 10�10 to 2� 10�7 for lead. These variations
of one or more than two orders of magnitude from one analysis
to another show the need to measure up-to-date extents of
formation valid for the current analysis. It should be noted that
the extents of formation certainly depend on the introduction
system: use of a desolvating device (membrane or heating/
condensing systems) could reduce the extent of formation of
some molecular ions, especially oxides. Such a system (‘‘Apex’’,
ESI, Omaha, NE, USA) will be implemented soon in the labo-
ratory and the extents of formation will be compared with the
ones obtained with the current conventional introduction
system.
Levels of background induced by polyatomic species and
comparison to other sources of background
The average, maximum, and minimum contributions of PIs
made with iridium, mercury and lead on the background at
masses 239 and 240 for plutonium measurements carried out for
the years 2007, 2008, and 2009 (i.e. 137 measurements) are
J. Anal. At. Spectrom., 2011, 26, 1474–1480 | 1477
Fig. 3 Various extents of formation of the background induced at mass
239 by polyatomic species made with mercury, iridium, and lead. The
extents of formation are ranked by increasing values and measured for
the years 2007, 2008, and 2009.
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
reported in Table 3. The contributions of the other HEs (Hf, Ta,
W, Re, Pt, Au, Tl, Bi) are negligible (<0.01 counts s�1) and are
not reported. We also report in the table the average, maximum,
and minimum values of the IB, obtained from the instrumental
blanks. For both masses 239 and 240, sums of polyatomic species
induced-background are on average higher than average IB (1.23
counts s�1 versus 0.41 counts s�1 at mass 239, and 1.95 counts s�1
versus 0.43 counts s�1 at mass 240). The additional polyatomic
species induced-background is equivalent to plutonium concen-
trations of roughly 0.5 fg$ml�1.
In addition to IB and to PI induced-background, other sources
of background are, on the one hand isotopic impurities of the
isotopic dilution (ID) tracer 242Pu, which contains small amounts
of 239Pu and 240Pu, and on the other, peak tail of 238U at masses
239 and 240 plus hydrides of 238U at mass 239. For each of the
measured samples, we determined the percentage of the back-
ground count rate due to each source of background. Then from
all these values we calculated the average percentages, plotted in
Fig. 4. At mass 239, the sum of PIs is one of the largest sources of
background (�22%) with impurities of the ID tracer 242Pu
(�24%) and sum of 238U hydrides and peak tail (�37%). IB at
mass 239 only accounted for �17% of the total background. At
mass 240, the sum of PIs represents the most important contri-
bution to the total background (�50%) well above IB, ID tracer
Table 3 Average, maximum and minimum contributions of polyatomicinterferences made with iridium, mercury and lead to the background atmasses 239 and 240 for plutoniummeasurements carried out for the years2007–2009 (137 measurements); comparison with the instrumentalbackground, noted ‘‘IB’’. The sum of all polyatomic interferences is noted‘‘Total’’. All values are given in counts s�1
Heavy element
Mass 239 Mass 240
AverageRangeof variation Average
Rangeof variation
Ir 0.16 0.00–3.78 0.03 0.00–0.80Hg 0.40 0.01–3.66 0.59 0.00–8.18Pb 0.67 0.00–14.05 1.33 0.00–29.68Total 1.23 0.01–14.34 1.95 0.01–30.10IB 0.41 0.00–1.67 0.43 0.00–1.75
1478 | J. Anal. At. Spectrom., 2011, 26, 1474–1480
impurities and 238U peak tail. By calculating the theoretical limit
of detection for each measurement only based on electronic
background, ID tracer impurities, 238U peak tail and 238U
hydrides, without taking into account the PIs, we show that false
detections of 239Pu and 240Pu may have occurred for respectively
15% and 22% of the samples.
Validation of this method is not easy, as we do not have
a certified reference material with very low amounts of pluto-
nium—in the femtogram range—at our disposal. However, a test
was done on a cotton wipe sample spiked with 10 fg of pluto-
nium. This test sample was prepared by another institution
(Khlopin Research Institute, St Petersburg, Russia). Calcula-
tions were made with and without correction of PIs (see Fig. 5).
The result obtained with correction of PIs is in better agreement
with the target value, whereas the result obtained without
correction of PIs is slightly overestimated.
Origin of heavy elements in the sample solution
Heavy elements responsible for PIs are largely eliminated during
sample preparation; this was also observed by other authors.22 In
fact, according to theoretical retention coefficients, with the
chemical media used in our purification procedures, a large part
of all the investigated HEs must be eliminated. To check this, we
applied the chemical purification procedure previously described
Fig. 4 Breakdown of the composition of the total background budget at
masses 239 (a) and 240 (b), average percentages over all plutonium
measurements carried out for the years 2007–2009 (137 measurements).
This journal is ª The Royal Society of Chemistry 2011
Fig. 5 Comparison of the results of the analysis of 3 process blanks
(PB1, PB2 and PB3) and of one quality control sample (‘‘S’’) spiked with
10 fg of 239Pu + 240Pu, using two calculation methods: without correction
of PIs (empty circles) and with correction of PIs (filled circles). Uncer-
tainties for the experimental results are given with a coverage factor of 2.
No uncertainty was provided for the target value.
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
in section 2.1 for the analysis of cotton wipe samples on three
solutions spiked with relatively large amounts of certified mono-
elemental standard solutions of Hf, Ta, W, Re, Os, Ir, Pt, Au,
Hg, Pb, Tl, and Bi. The mass of each element added to the
solutions ranged from 120 ng to 10 mg. The efficiency of each
separation step was evaluated through measurement of the
remaining masses of each HE eluted. We give here the overall
efficiency of the purification procedure for plutonium analysis,
defined as the ratio of the masses of each HE in the final sample
solutions to the masses introduced in the initial solutions
(see Table 4). We give an upper value of the overall efficiency
when the element was not detected in the sample solutions. We
thus demonstrate that all HEs are more or less thoroughly
eliminated. However, as shown by the PB, the purification
procedure can also bring significant and variable amounts of
some HEs, through impurities in glassware, reagents and atmo-
spheric contamination. In the case of lead, efficiency of the
purification is probably better than the one given here, as a large
part of the lead measured in the sample solution is surely
Table 4 Overall efficiency of the elimination of HEs by the purificationprocedure for plutonium analysis starting from cotton wipe sample (seesection 2.1). Detection limits for the overall efficiency are defined as 3times the standard deviation over the process blanks. Uncertainties aregiven with a coverage factor of 2
Heavy elementOverallelimination efficiency
Hf <3 � 10�8
Ta <4 � 10�5
W <10�7
Re (8 � 2) � 10�9
Os <1.5 � 10�5
Ir (3 � 1) � 10�5
Pt <3 � 10�7
Au (6 � 3) � 10�9
Hg <10�4
Pb <10�3
Tl <5 � 10�6
Bi (8 � 4) � 10�4
This journal is ª The Royal Society of Chemistry 2011
introduced by the purification procedure itself. In particular,
impurities in reagents used for the last steps of the purification
procedure, like NH4I for the final plutonium elution, will
certainly be found in the final sample solutions. Investigations
are in progress to identify the main source of Pb, Ir and Hg in the
sample solutions. There is strong evidence that NH4I may be
responsible for a significant part of these contaminations. Use of
another reducing agent, like HBr or HI, is considered.
Conclusions
For three years we applied an original method for low-level
plutonium measurement using a very sensitive double focusing
sector-field ICP-MS (typically 2–4 million counts s�1 per ng ml�1
of plutonium). This method includes correction of polyatomic
interferences containing heavy elements from hafnium to
bismuth, which interfere at the masses of long-lived plutonium
isotopes 239Pu and 240Pu. We pointed out that the extents of
formation of these molecular species are highly variable, strongly
dependent on instrumental settings, and can vary over one or two
orders of magnitude from one analysis to another. Such extents
of formation are very low for most of the heavy elements and
should not lead to significant additional background in the
actinide mass range. Exceptions are mercury and iridium, which
show higher extents of formation for ArHg+ and IrO3+, respec-
tively of the order of 10�5 and 10�6, and may produce significant
additional background for limited concentrations—in the mg l�1
range—of the elements in sample solutions. The extents of
formation of PbO2+ molecular species are low, in the order of
10�9, but lead concentrations in sample solutions can be rela-
tively high, up to the mg l�1 range. Finally, we observed that the
corresponding additional background induced by these molec-
ular species is generally of the same order as, and in many cases
higher than, other usual sources of background for low-level
plutonium analysis by ICP-MS, i.e. electronic background, 242Pu
tracer isotopic impurities, 238U hydrides and peak tail. Thus, if
not corrected, these polyatomic interferences may lead to false
detection of femtogram amounts of plutonium or overestimation
of results.
Correction of these polyatomic interferences using the specific
methodology described in this paper serves to improve accuracy
and avoid false detections. However, corrections are not
a panacea as these corrections are tainted with analytical error
and result in an increase in the detection limits. A better way is of
course to minimise the concentrations of heavy elements in
sample solutions and/or the extents of formation of molecular
species. For this, the instrument should be optimised not only for
maximum analyte signal, but also for minimum oxide ratio to at
least minimise oxide interferences (PbO2+, IrO3
+, etc.) during
plutonium analysis. More generally, a thorough study of the
instrumental settings that minimise specific polyatomic interfer-
ences should be undertaken. In this respect, a desolvating
introduction system which in addition to enhancing the sensi-
tivity may decrease the oxide ratio, will be tested soon. Another
way is to use the high resolution capability of our instrument to
separate plutonium isotopes from PbO2+, ArHg+, IrO3
+ molec-
ular species (a moderate resolution of about 3500 is enough for
this) combined with a high-efficiency desolvating introduction
system to compensate at least in part for the loss of transmission
J. Anal. At. Spectrom., 2011, 26, 1474–1480 | 1479
Publ
ishe
d on
08
Apr
il 20
11. D
ownl
oade
d by
Uni
vers
ity o
f Il
linoi
s at
Chi
cago
on
31/1
0/20
14 1
9:53
:58.
View Article Online
in the medium-resolution mode. Lastly, our purification proce-
dures based on anion-exchange chromatography provide a very
efficient elimination of most of the heavy elements. However,
some of them are apparently introduced by the atmosphere, the
reagents and the glassware during the chemical purification.
Efforts must therefore be made to improve the chemical proce-
dures by using lower quantities of reagents, selecting purer
reagents and protecting the samples from atmospheric contam-
ination through use of clean-room-like conditions.
References
1 Y. Muramatsu, T. Hamilton, S. Uchida, K. Tagami, S. Yochida andW. Robison, Sci. Total Environ., 2001, 278, 151–159.
2 S. Tonouchi, H. Habuki, K. Katoh, K. Yamazaki and T. Hashimoto,J. Radioanal. Nucl. Chem., 2002, 252, 367–371.
3 Z. Wang and M. Yamada, Earth Planet. Sci. Lett., 2005, 233, 441–453.
4 M. Liezers, S. A. Lehn, K. B. Olsen, O. T. Farmer, III andD. C. Duckworth, J. Radioanal. Nucl. Chem., 2009, 282, 299–304.
5 I. Rodushkin, P. Lindahl, E. Holm and P. Roos, Nucl. Instrum.Methods Phys. Res., Sect. A, 1999, 423, 472–479.
6 J. B. Truscott, P. Jones, B. E. Fairman and E. H. Evans, Anal. Chim.Acta, 2001, 433, 245–253.
7 R. N. Taylor, T. Warneke, J. A. Milton, I. W. Croudace,P. E. Warwick and R. W. Nesbitt, J. Anal. At. Spectrom., 2001, 16,279–284.
8 T. C. Kenna, J. Anal. At. Spectrom., 2002, 17, 1471–1479.9 F. Pointurier, N. Baglan and P. H�emet, Appl. Radiat. Isot., 2003, 60,561–566.
10 Y. Muramatsu, S. Yochida and A. Tanaka, J. Radioanal. Nucl.Chem., 2003, 255, 477–480.
11 B. G. Ting, R. S. Pappas and D. C. Paschal, J. Anal. At. Spectrom.,2003, 18, 795–797.
12 N. Baglan, P. H�emet, F. Pointurier and R. Chiappini, J. Radioanal.Nucl. Chem., 2004, 261, 609–617.
13 R. S. Pappas, B. G. Ting and D. C. Paschal, J. Anal. At. Spectrom.,2004, 19, 762–766.
1480 | J. Anal. At. Spectrom., 2011, 26, 1474–1480
14 J. S. Becker, M. Zoriy, L. Halicz, N. Teplyakov, C. M€uller, I. Segal,C. Pickhardt and I. T. Platzner, J. Anal. At. Spectrom., 2004, 19,1257–1261.
15 M. E. Ketterer, K. M. Hafer, C. L. Link, D. Kolwaite, J. Wilson andJ. W. Mietelski, J. Anal. At. Spectrom., 2004, 19, 241–245.
16 C. S. Kim, C. K. Kim and K. J. Lee, J. Anal. At. Spectrom., 2004, 19,743–750.
17 M. V. Zoriy, C. Pickhardt, P. Ostapzuk, R. Hille and J. S. Becker, Int.J. Mass Spectrom., 2004, 232, 217–224.
18 D. Schauml€offel, P. Giusti, M. V. Zoriy, C. Pickhardt, J. Szpunar,R. Lobinski and J. S. Becker, J. Anal. At. Spectrom., 2005, 20, 17–21.
19 J. Zheng and M. Yamada, J. Radioanal. Nucl. Chem., 2006, 267, 73–83.
20 J. Zheng and M. Yamada, Talanta, 2006, 69, 1246–1253.21 K. Norisuye, K. Okamura, Y. Sohrin, H. Hasegawa and
T. Nakanishi, J. Radioanal. Nucl. Chem., 2006, 267, 183–193.22 Z. Varga, G. Suranyi, N. Vajda and Z. Stefanka,Microchem. J., 2007,
85, 39–45.23 H. Liao, J. Zheng, F. Wu, M. Yamada, M. Tan and J. Chen, Appl.
Radiat. Isot., 2008, 66, 1138–1145.24 O. T. Farmer, III, K. B. Olsen, M. L. Thomas and S. J. Garofoli, J.
Radioanal. Nucl. Chem., 2008, 276, 489–492.25 D. Larivi�ere, K. Benkhedda, S. Kiser, S. Johnson and R. J. Cornett,
Anal. Methods, 2010, 2, 259–267.26 J. S. Becker and H. J. Dietze, J. Anal. At. Spectrom., 1997, 12, 881–
889.27 S. H. Lee, J. Gastaud, J. J. La Rosa, L. Liong Wee Kwong,
P. P. Povinec, E. Wyse, L. K. Fifield, P. A. Hausladen, L. M. DiTada and G. M. Santos, J. Radioanal. Nucl. Chem., 2001, 248, 757–764.
28 E. J. Wyse, S. H. Lee, J. La Rosa, P. Povinec and S. J. De Mora, J.Anal. At. Spectrom., 2001, 16, 1107–1111.
29 C.-C. Shen, R. L. Edwards, H. Cheng, J. A. Dorale, P. B. Thomas,S. B. Moran, S. E. Weinstein and H. N. Edmonds, Chem. Geol.,2002, 185, 165–178.
30 U. Nygren, I. Rodushkin, C. Nillson and D. C. Baxter, J. Anal. At.Spectrom., 2003, 18, 1426–1434.
31 U. Nygren, H. Rameb€ack, D. C. Baxter and C. Nilsson, J. Anal. At.Spectrom., 2005, 20, 529–534.
32 F. Pointurier, P. Hemet and A. Hubert, J. Anal. At. Spectrom., 2007,22, 1–9.
This journal is ª The Royal Society of Chemistry 2011