RUN SWAB PLANS REPORTS SETTINGS PLAN LOCATION LIMIT RESULT SURFACE OPERATOR1 OPERATOR2 MISC PASS OPERATOR1 NONE FILLER NOZZLE LINE 1 STAINLESS STEEL 0 1000 LOCATION PLAN SURFACE MISC OPERATOR1 OPERATOR2 LINE 1 TEST 1 STAINLESS STEEL FILLER NOZZLE OPERATOR1 NONE LOCATION FILLER NOZZLE LINE 1 LINE 2 LINE 3 RUN RUN RUN PLANS RE-TEST DELETE RE-TEST BACK 659 Andover Street, Lawrence, MA 01843-1032, USA T +1.978.687.9200 | F +1.978.687.9216 | E [email protected] | www.charm.com © 2016 Charm Sciences, Inc. Charm, novaLUM and PocketSwab are registered trademarks of Charm Sciences, Inc. MRK-1056-001 May-2016 Select RUN. Note: Swabs can be held for up to 6 hours at room temperature before activation. Select PLANS button. 1 Select PLANS. PocketSwab ® Procedural Flow Chart Refer to Current Charm ® Operator’s Manual for Complete Test Procedure Withdraw the swab by gently pulling and twisting the swab handle. Use Left ( )or Right ( ) arrows to pick location or use List View ( ) to scroll to find correct location. Swab an area 100 cm 2 (16 in 2 ). Rotate the handle to achieve full swab contact with surface. Reinsert the swab and twist the handle down. Gently shake side to side 3 times. White tablet will break apart. Insert the PocketSwab Plus swab fully into the chamber of the novaLUM ® II system in an upright position. Press to start count. Results in 5 seconds. The result of “Pass” or “Fail” is based on established limit. 2 3 5 6 8 4 7 9 PLAN SURFACE MISC OPERATOR1 OPERATOR2 LINE 1 TEST 1 STAINLESS STEEL OPERATOR1 NONE Refer to Current Charm ® Operator’s Manual for Complete Test Procedure

PocketSwab Procedural Flow Chart Refer to Current Charm ... · Refer to Current Charm® Operator’s Manual for Complete Test Procedure pick location or use List View ( ) to Withdraw

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Page 1: PocketSwab Procedural Flow Chart Refer to Current Charm ... · Refer to Current Charm® Operator’s Manual for Complete Test Procedure pick location or use List View ( ) to Withdraw

RUN SWAB PLANS REPORTS SETTINGS

PLAN

LOCATION

LIMIT

RESULT

SURFACE

OPERATOR1

OPERATOR2

MISC

PASS

OPERATOR1

NONE

FILLER NOZZLE

LINE 1

STAINLESS STEEL

0

1000

LOCATION

PLAN

SURFACE

MISC

OPERATOR1

OPERATOR2

LINE 1

TEST 1

STAINLESS STEEL

FILLER NOZZLE

OPERATOR1

NONE

LOCATION FILLER NOZZLE

LINE 1

LINE 2

LINE 3

RUN

RUN

RUN

PLANS RE-TEST

DELETE RE-TEST

BACK

659 Andover Street, Lawrence, MA 01843-1032, USAT +1.978.687.9200 | F +1.978.687.9216 | E [email protected] | www.charm.com© 2016 Charm Sciences, Inc. Charm, novaLUM and PocketSwab are registered trademarks of Charm Sciences, Inc.

MRK-1056-001 May-2016

Select RUN.

Note: Swabs can be held for up to 6 hours at room temperature before act ivat ion.

Select PLANS button.

1

Select PLANS.

PocketSwab® Procedural Flow Chart

Refer to Current Charm ® Operator’s Manual for Complete Test Procedure

Withdraw the swab by gently pul l ing andtwist ing the swab handle.

Use Left ( )or Right ( ) arrows to pick location or use List View ( ) to scrol l to f ind correct location.

Swab an area 100 cm2 (16 in2). Rotatethe handle to achieve ful l swab contactwith surface.

Reinsert the swab and twist the handledown. Gently shake side to side 3 t imes.White tablet wi l l break apart.

Insert the PocketSwab Plus swab ful ly into the chamber of the novaLUM® II system in an upright posit ion. Press to start count.

Results in 5 seconds. The result of “Pass”or “Fai l” is based on establ ished l imit.

2 3

5 6

8

4

7 9

PLAN

SURFACE

MISC

OPERATOR1

OPERATOR2

LINE 1

TEST 1

STAINLESS STEEL

OPERATOR1

NONE

Refer to Current Charm® Operator’s Manual for Complete Test Procedure