PMT test setup in Erlangen
Oleg Kalekin
University of Erlangen
Pylos, 16.04.2007
Meeting
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
2
PMT Test SetupCharge response
Pulser (custom made):• Rate 10 kHz• FWHM 10ns• Variable amplitude
Dark Box:• Copper shielding• Various diameter PMT
holders
ADC LeCroy 2249A• No. of channels 12• No. of Bits 10• Full scale 256 pC
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
3
PMT Test SetupCharge response
Pulser (custom made):• Rate 10 kHz• FWHM 10ns• Variable amplitude
Dark Box:• Copper shielding• Various diameter PMT
holders• Diaphragms
ADC LeCroy 2249A• No. of channels 12• No. of Bits 10• Full scale 256 pC
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
4
PMT Test SetupCharge response
Pulser (custom made):• Rate 10 kHz• FWHM 10ns• Variable amplitude
Dark Box:• Copper shielding• Various diameter PMT
holders• Diaphragms
ADC LeCroy 2249A• No. of channels 12• No. of Bits 10• Full scale 256 pC
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
5
PMT Test SetupCharge response
Reconstruction of single-/multiple- photoelectron responseModel(*):
• 1st dynode: if mean number of ph.e., probability to detect n ph.e. = P(n, ) - Poisson function
• Anode: single ph.e. charge = G(Q1,1) – Gaussian
n ph.e. charge = G(Qn,n), Qn=n•Q1 , 2n= n•2
n
• PMT charge response: P(i,)•C(i)•exp(-(x-Qi)2/(22i)), i=1,2,…
Qi=i•Q1, 2i= i•2
1, C()=1/ 2
• 0 ph.e. (pedestal): P(0, )=exp(- ), if N – total number of event, number of pedestal events N0=N•exp(- ), and then =-ln(N0/N) can be used as first estimation of mean number of ph.e.
• Fit method: 2-minimizing with , Q1 , 1 as fit parameters
• Software: ROOT Version 5.08/00
(*) E.H. Bellamy, et al., NIM A 339 (1994) 468-476
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
6
PMT Test Setup Charge response
Reconstruction of single-/ multiple- photoelectron response: simulations
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
7
PMT Test Setup Charge response
Reconstruction of single-/ multiple- photoelectron response
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
8
PMT Test SetupAbsolute and Spectral Calibration
• Monochromator
LOT-Oriel MS301
fully automated
350-900 nm• Tungsten lamp• PMT bases for QE
measurements
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
9
PMT Test SetupAbsolute and Spectral Calibration
• Monochromator
LOT-Oriel MS301
fully automated
350-900 nm• Tungsten lamp• PMT bases for QE
measurements
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
10
PMT Test SetupAbsolute and Spectral Calibration
• Monochromator
LOT-Oriel MS301
fully automated
350-900 nm• Tungsten lamp• PMT bases for QE
measurements
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
11
Photonis PMTs
• 10" PMT ordered, expected to be delivered end of April
• 3" PMT will be provided by NIKHEF
Diam/Areainch/cm2
9/380
5/120
10/600
9/380
8/400
12/670
Euro/cm2
3.72
7.48
2.72
3.54
3.26
3.20
Oleg Kalekin University Erlangen
16.04.2007 KM3NeT Meeting, Pylos
12
Summary
• PMT test setup in Erlangen works
• Further development is in progress to cover all important PMT characteristics
• Tests of Photonis PMTs will start soon
• Hamamatsu PMTs will be evaluated to order
• Methods, codes, results will be posted on the KM3NeT Portal