33
Model PM-3514 Two-Color Pyrometer Operator’s Manual Revision 2.1 Logika Technologies Inc. 2-2857 Sherwood Heights Dr. Oakville ON, CanadaL6J 7J9 Tel: 905-829-5841 www.logikatech.com

PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

  • Upload
    others

  • View
    123

  • Download
    25

Embed Size (px)

Citation preview

Page 1: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Model PM-3514 Two-Color Pyrometer

Operator’s Manual

Revision 2.1

Logika Technologies Inc.

2-2857 Sherwood Heights Dr.

Oakville ON, CanadaL6J 7J9

Tel: 905-829-5841

www.logikatech.com

Page 2: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 1

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev 2.1

November 13

1 INTRODUCTION .................................................................................................. 2

2 DESCRIPTION ...................................................................................................... 2

2.1 MODEL NOMENCLATURE ................................................................................................................. 2

2.2 OPERATING PRINCIPLE ..................................................................................................................... 2

2.3 SPECIFICATIONS.............................................................................................................................. 3

3 LOCATION AND MOUNTING ............................................................................... 5

3.1 PYROMETER LOCATION .................................................................................................................... 5

3.2 PYROMETER MOUNTING .................................................................................................................. 7

4 UTILITY CONNECTIONS ....................................................................................... 7

4.1 ELECTRICAL CONNECTIONS ............................................................................................................... 7

4.2 WATER COOLING CONNECTIONS ....................................................................................................... 9

4.3 AIR PURGE CONNECTION ............................................................................................................... 10

5 OPERATION ...................................................................................................... 10

5.1 CONTROLS .................................................................................................................................. 10

5.2 SET MENU FUNCTIONS AND PROGRAMMING ..................................................................................... 10

Set Menu 1- Emissivity/Slope Adjustment ................................................................................ 11

Set Menu 2- Display Mode: ....................................................................................................... 16

Set Menu 3– Output Signal Integration Time: .......................................................................... 18

Set Menu 4-, 5-, 6- Process Alarm Settings: .............................................................................. 19

Set Menu 7-, 8- Current/Voltage Analog Output Span Settings: .............................................. 19

Set Menu 9- Aiming Laser Setup: .............................................................................................. 20

Display Menu 10- Enclosure Temperature: .............................................................................. 20

Display Menu 11- Two Color Ratio ............................................................................................. 21

Set Menu 12- Zero Compensation Gain ................................................................................... 21

Set Menu 13- Two-color mode signal attenuation ratio ........................................................... 21

5.3 DEBUGGING ................................................................................................................................ 27

6 MAINTENANCE ................................................................................................. 29

6.1 REGULAR MAINTENANCE ............................................................................................................... 29

6.2 INSTRUMENT RETURN ................................................................................................................... 29

7 APPENDIX ......................................................................................................... 30

Page 3: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 2

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

1 Introduction

PM-3514 two-color infrared pyrometer is primarily used in the metallurgical industry to

measure temperature of steel or molten iron, wire rod, hot rolled steel, as well as cement kilns.

The pyrometer can “look” into a process and measure the target temperature from a distance,

providing an accurate reading while eliminating problems associated with contact measuring

devices. Dual infrared wavelengths input signals overcome application challenges such as

small targets, variable emissivity, dust, water or other interfering media between the target

and the pyrometer. Accurate temperatures are obtained even when the target is 95%

obstructed.

2 Description

2.1 Model Nomenclature

PM- 3514 - 024

Logika two color infrared pyrometer

Lower limit temperature 350°C

Upper limit temperature 1400°C

Input power supply 24VDC

Figure 2.1 PM-3514 Model Nomenclature

2.2 Operating Principle

Logika PM-3514 pyrometer uses a state-of-the-art photodiode to sense the infrared energy

from the target. Our pyrometer measures infrared energy emitted by a target at two

wavelengths and calculates the ratio of these measurements to determine the target

temperature. Since the emissivity of the two wavelengths is identical in same environment, it

does not influence the accuracy of the temperature measurement.

Page 4: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 3

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

The primary advantage of PM-3514 over thermocouple temperature measurement is that the

pyrometer accurately measures the target temperature by way of non contact means and

responses much faster.

Compared with single wavelength pyrometers, the advantages of the Logika PM-3514 two-

color pyrometer are:

1. Overcomes optical fouling or interfering media between the target and the pyrometer,

such as dust, dirt, smoke, and water vapor, to yield accurate temperatures.

2. Automatically compensates for changes in target emissivity or varying distances between

the target and pyrometer.

3. Almost no restrictions to pyrometer’s field of view, leaving more flexibility for pyrometer

mounting. The PM-3514 can accurately measure the target temperature even with as little

as 5% of the field of view.

4. PM-3514 can also switch to single color mode for special field applications.

The PM-3514 also has built-in temperature slope adjustment for calibrating the pyrometer

to the specific characteristics of the target metallurgy, and a visible target aiming laser for easy

installation. See Section 5.2.

2.3 Specifications

Power Input 24 VDC ±20% @ 210 mA

Power loss

protection

Reinstates all previous setup parameters when power is restored

Wavelength

One-color mode: 0.9-1.7μm

Two-color mode: 0.9-1.7μm and 1.7μm

Response time Adjustable – the fastest response time is 20 ms

Accuracy 0.5%

Repeatability ±2°C

Operating

temperature

No water cooling: -20˚C to +55˚C (-4˚F to +130˚F).

With water cooling: -20˚ C to 120˚C (-4˚F to 250˚F)

Page 5: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 4

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Humidity Relative humidity range: 10-95%

Measuring Range

Single color mode:350-1400˚C (662˚F to 2552˚F)

Two color mode:350-1400˚C (662˚F to 2552˚F)

Field of View

One-color mode: D/S=50.

*D/S = 100 is available as special order

D is the distance between the infrared pyrometer and the target

S is the target diameter. See Figure 3.1 for more details.

Two-color mode: The target size can be much smaller. If

temperature is greater than 600°C, the maximum signal

attenuation ratio can be 95%.

Display mode: 4 digital LEDs. Display peak, valley, average or real time options

Focus range 0.5m- ∞

Calibration

coefficient

Two color mode: Temperature slope, 0.850 - 1.150. Adjustable in

0.001 increments (0.001↑results in 1.4°C ↓).

One color mode: Emissivity rate, 0.100-1.000 adjustable in 0.001

increments (0.01↑ results in 1-2.0°C ↓).

Alarms output:

High or low process temperature alarm, rated 0.2A at 220 VAC,

0.5A at 30 VDC, MOS relay output.

Ambient temperature high or low alarm

Analog output

Linear 4-20 mA (or 0-20mA) proportional to temperature

Linear 0-10VDC proportional to temperature

Aiming method

Visible aiming laser

wavelength 630nm±10%, Power≤3mW, safety rating IIa

Warm up time 10 minutes

Housing

Hermetically sealed stainless steel enclosure rated IP66, includes

purge air and cooling water connections.

Optional fiber optic lens assembly includes air purge fitting.

Page 6: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 5

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Size 370mm L X 100mm W X 190mm H (14.5” L X 4” W X 7.5” H)

Weight 4 kg (8.8 Lbs)

3 Location and Mounting

3.1 Pyrometer Location

Basic installation considerations:

• Pyrometer should not be installed above heat sources, also avoid directly exposure to

water vapor, excessive dust or smoke. If harsh ambient environment is unavoidable,

connect clean, dry air to the air purge fitting on the pyrometer enclosure. Heat

shielding or fan can also be applied to dissipate water vapor or smoke.

• If conditions allow, mount the pyrometer as close to the target as possible. This will

supply enough infrared energy for the pyrometer provides strong signal. At the same

time, the pyrometer must be kept far away enough to protect the electronics inside the

enclosure from excessive heat.

• Installation distance: dependent on one-color or two-color mode.

o One-color mode: target should be covered by the pyrometer’s field of view.

Check Figure 3.1 and Figure 3.2 for reference.

o Two-color mode: target can be 95% obstructed. See Figure 5.7 for reference.

For example:

If the target diameter is 20mm, pyrometer Displacement Coefficient is 50:1

(D/S=50:1).

In one-color mode, according to Figure3.1 and Figure3.2, the pyrometer has to be

installed less than 1 meter away from the target.

In two-color mode, it is not necessary for the target to be totally covered by the field

of view; the pyrometer can be installed more than 1 meter away from the target.

• Maximum ambient temperature for the pyrometer

o 50oC (122

oF) without water cooling

o 120oC (250

oF) with water cooling connected. See Figure 4.2

Page 7: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 6

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Note: If target to pyrometer distance is less than 1000mm, Pyrometer field of view is more than

calculation value.

Note: If target to pyrometer distance is less than 1500mm, Pyrometer field of view is more than

calculation value.

Figure 3.1: One color mode Pyrometer Field of View

Page 8: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 7

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Target is larger Target is equal Target is smaller

Ideal Acceptable Not Acceptable

PM Series

Field ofView

Figure 3.2 One-Color Mode Pyrometer Ideal Target Size

3.2 Pyrometer Mounting

Installation – PM-3514 has an adjustable mounting foot which is secured using a M10 bolt.

Aiming - Use the aiming laser to properly align the pyrometer with target.

4 Utility Connections

4.1 Electrical Connections

All electrical connections (power and signal) are made via a 12-pin connector and mating cable

for the junction box connections.

Cable Specification

Standard cable length is 2.0 meters; minimum bent radius for cable is 30mm. Other cable

lengths are available from Logika Technologies Inc. upon request.

Page 9: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 8

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Figure 4.1 Pyrometer connector and Cable assignment

Junction Box Wiring

Please see the Table 4.1, Appendix Figure 7.4 and Figure 7.5 for pyrometer wiring terminal

description. The supplied cable should be connected directly to the terminals of a junction box.

Junction box is shielded and connected to ground to eliminate electrical noise, preventing

triggering false detection. Any unused output wires must have insulation protection or be

connected to a free terminal in junction box.

For Junction Box with AC-DC converter: AC Input 85V-265VAC, connect to L, N, E

Cable No. Color Function Junction Box

Number

Junction Box

Name

1 Brown +24VDC 1 +24V

2 Red 0V 2 0V

3 Orange Analog 0-10V Output 3 0-10V

4 Yellow Analog 4-20mA Output 4 mA

5 Green Analog Output Ground 5 AGND

6 Blue PNP output (Upper limit )alarm) 6 PNP

7 Purple Upper limit alarm closed=Alarm 7 UPPER1

8 Gray Upper limit alarm Common 8 UPPER2

9 White Lower limit alarm closed=Alarm 9 LOWER1

10 Black Lower limit alarm Common 10 LOWER2

11 Pink RS485 A 11

12 Sky blue RS485 B 12

Table 4.1 Cable Terminal Wiring Description

Page 10: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 9

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

4.2 Water Cooling Connections

General description

Water cooling is required when the ambient temperature is higher than 50oC (122

oF) to

protect electronics inside pyrometer tube from overheating, the connectors are two Φ10mm

OD barbed fittings on the enclosure. Stainless steel heat exchanger tubing is embedded in

the housing for efficient heat removal using circulating water.

Cooling Water Requirements

• Clean industrial water

• Maximum water inlet temperature 25oC (77

oF)

• Maximum water pressure of 4 bars (60 PSI).

• Water flow rate - 1 to 2 L/min (0.26 to 0.53 gal/min)

Connections

• Use Φ10mm ID or 3/8” ID hose for connections

• One fitting is connected to plant’s water supply

• The other one is connected to drain

• Both fittings should be secured with hose clamps

• Setup the cooling water flow rate by Figure 4.2

Am

bie

nt

Te

mp

era

atu

re

60

80

100

120

140

160

°C

Water0.3 0.5 1.0 1.5 2.0 L/M

8 °C

20 °C

25°C

Figure 4.2 Ambient Temperature vs. Water Flow Rate

Page 11: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 10

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

4.3 Air Purge Connection

Description- Compressed plant air can be used to protect the lens from dust and vapor. Purge

air cleans the lens, reduces maintenance times, and prolongs sensor life where dust or

corrosive vapor is present.

Purge Air Requirements

• Must be clean, dry instrument air no oil, dust and contaminants included. Poor air quality

will result in dirty lens and decrease detector performance. If the quality of compressed air

can’t guaranteed, it is better not connect it to the sensor to avoid the lens contaminated

and affect the sensor’s performance.

• Air filtration prior to the inlet fitting is recommended to improve the quality of purge air

• Air Pressure range from 50 to 200 g/cm2

(0.7 psi to 2.8 psi)

• Air Flow rate from 4 to 16 L/min (0.14 to 0.56 ft3/min)

Connection- Connect the plant air source to the purge air barbed fitting on the sensor’s lens

shroud with Φ10mm ID or 3/8” ID hose. Purge air flows out of the sensor shroud and

dissipates into the ambient environment.

5 Operation

5.1 Controls

The PM-3514 pyrometer’s control panel is located at the rear of the enclosure. The various

functions are described in greater detail below.

To view and adjust pyrometer menus:

• Use M3 Allen key to remove the back cover from unit

• Unthread the glass cover on the pyrometer body by turning it counterclockwise.

!!!!Note:

o The pyrometer tube may slide out of water jacket if angled downwards. Keep unit level

when disassembling for menu adjustment.

o The pyrometer tube is sealed during manufacturing to retain accuracy and quality.

Warranty is void if pyrometer tube is dismantled for any reason.

5.2 Set Menu Functions and Programming

In display mode:

Press to switch between one-color and two-color mode

Page 12: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 11

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Press ▲ to switch temperature display between °F and °C

1 2 3 4 5 6 7

89101112131415

Figure5.1 Keypad setup sequence

Press SET repeatedly to choose setup parameters

Press ENTER to display current value

Use the arrow key to change the value

Press ENTER again to save the new value and EXIT setup

Set Menu 1- Emissivity/Slope Adjustment

One-color mode: Emissivity setup value

Emissivity: The relative ability of target measured surface to emit energy by radiation. It is the

ratio of energy radiated by a particular material to energy radiated by a black body at the same

temperature. A true black body would have an emissivity = 1 while any real object would have

emissivity < 1.

By adjusting the emissivity, the pyrometer can accurately measure target temperatures. If the

target’s emissivity is setup higher than its true value, the measured temperature by pyrometer

will be lower than the actual temperature.

The emissivity data in the table below is for reference only. Material surface finish can cause

varying emissivity. If the metal surface is slightly oxidized and dirty, its emissivity will drop

significantly.

Page 13: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 12

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Table 5.1 One-color emissivity table for different materials

Material and status

Emissivity

(wavelength 1.6ìm)

Aluminum

Not oxidized

Oxidized

Aluminum Alloy A3003, Oxidized

Rough surface

Grinded surface

0.02-0.2

0.4

0.4

0.2-0.6

0.02-0.1

Brass

Grinded surface

Polished surface

Oxidized

0.01-0.05

0.6

Chromium 0.4

Copper

Grinded surface

Rough surface

Oxidized

0.03

0.05-0.2

0.2-0.9

Gold

Hastelloy

Inconel Alloy

Oxidized

Sandblast

0.6-0.9

0.6-0.9

0.3-0.6

Page 14: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 13

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Electro-polished 0.25

Iron

Not Oxidized

Oxidized

Rusted

Molten

0.1-0.3

0.5-0.8

0.6-0.9

0.4-0.6

Cast iron

Not Oxidized

Oxidized

Molten

0.3

0.7-0.9

0.3-0.4

Wrought iron

Dark color

0.9

Lead

Grinded surface

Rough surface

Oxidized

0.05-0.2

0.6

0.3-0.7

Magnesium 0.05-0.3

Mercury 0.05-0.15

Molybdenum

Page 15: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 14

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Oxidized

Not Oxidized

0.4-0.9

0.1-0.35

The Monel Copper-Nickel Alloy 0.2-0.6

Nickel

Oxidized

Electrolysis

0.4-0.7

0.1-0.3

Platinum black 0.95

Silver 0.02

Steel

Cold-rolled steel Coil

Polished steel plate

Molten Steel

Oxidized

Stainless Steel

0.8-0.9

0.25

0.25-0.4

0.8-0.9

0.2-0.9

Tin (Not oxidized) 0.1-0.3

Titanium

Grinded surface

Oxidized

0.3-0.5

0.6-0.8

Wolfram

Grinded surface

0.1-0.3

Zinc

Oxidized

Grinded surface

0.15

0.05

Page 16: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 15

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Procedure to adjust the emissivity in one-color mode:

1. Set the pyrometer to work in one-color mode (Mode LED off).

2. Determine actual target temperature with a contact RTD, thermocouple or other

appropriate instrument.

3. Press SET button until the display shows “1-“(flashing), then press ENTER to adjust

the emissivity.

4. Use the “up” and “right” arrow buttons to adjust the value.

5. The emissivity adjustment range is from 0.100 to 1.000; in 0.01 increments. For

0.01↑, results in 1-2.0°C ↓, factory default value is 0.99.

6. Press ENTER to save the adjustment and return to the temperature display mode.

Two-color mode: Temperature Slope Adjustment (Calibration)

Temperature slope is a coefficient to compensate for the two wavelengths emissivity

difference.

Note: The listed ratios are approximate values and can change with alloy surface condition and

oxidization.

For these oxide metals, the temperature slope setup value is 1.000

Cobalt Nickel Iron Stainless steel Steel

For these metals, the temperature slope value is 1.030

Cobalt Iron Stainless Steel

Rhodium Tungsten Molybdenum

Platinum Steel Nickel

Tantalum Cast Iron

Slope adjustment allows the user to refine the accuracy of the pyrometer for specific

applications. This feature can be used to calibrate the pyrometer.

1. Use a reliable thermometer to measure the surface temperature of the target. If

temperature readings vary in different measured areas, use the average reading.

2. Set the pyrometer to work in two-color mode (Mode LED ON)

3. Press SET button until the display shows “1-“(flashing), then press ENTER to adjust the

temperature slope.

4. Use the “up” and “right” arrow buttons to adjust the value.

5. The temperature slope adjustment range is 0.850 to 1.150 in 0.001 increments. For

0.001↑, results in 1.4°C ↓. Factory default setup value is 1.000.

6. Press ENTER to accept the slope adjustment and return to the temperature display mode.

Page 17: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 16

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Set Menu 2- Sample Mode:

This feature allows the user to select different temperature sample modes; Peak temperature;

Valley temperature; or Average temperature values.

Peak: Response time setup range from 0-999.9s. If response time is set greater than 999.9

seconds, the current sample mode will remain unchanged.

Applications: Best for moving target temperature measurement (steel plate, sheet). This

function can also be used for measuring surface temperature of liquid metal when highest

temperature measurement is required.

Valley: Response time setup range from 0-999.9s. If response time is set greater than 999.9

seconds, the current sample mode will remain unchanged.

Applications: Best for measuring surface temperature of liquid metals when lowest

temperature measurement is required.

Average: Response time setup range from 0-60.0 seconds.

Applications: Most field applications. For example, if there is scale on steel plate or bar,

average temperature sample mode should be selected.

Real time sample: Set response time to 0.0s. In this mode, the sampling time is set 20ms by

pyrometer and display changes each time new data is collected.

1. Press SET button until the display shows “2-“(flashing).

2. Press ENTER button and use the “up” arrow button to select sample mode. Select (1) for

Peak, (2) for Valley or (3) for Average. The factory default value is Average value (indicated

by the number 3).

3. Press ENTER to return to the temperature display mode.

Page 18: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 17

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Temperature

TimeΔt 2Δt 3Δt 4Δt0

Peak Display

Actual Temperature

Figure 5.2 Peak Sample Mode

Temperature

TimeΔt 2Δt 3Δt 4Δt0

Valley Display

Actual Temperature

Figure 5.3 Valley Sample Mode

Page 19: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 18

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Temperature

TimeΔt 2Δt 3Δt 4Δt0

Average display

Actual temperature

Figure 5.4 Average Sample Mode

Temperature

Time(ms)20ms 40ms 60ms 80ms0

Real timedisplay

Actualtemperature

Figure 5.5 Real Time Sample Mode

Set Menu 3– Output Signal Integration Time:

This feature allows the user to average (integrate) the output signal over a set time for

control purposes and to avoid wide temperature signal swings.

Page 20: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 19

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Press SET until the display shows “3“ (flashing), then press ENTER to adjust the signal

integration time. Use the “up” and “right” arrow buttons to adjust the integration time.

The time adjustment range is 0 to 999 seconds. The factory default setting is 0 seconds.

Press ENTER to return to the temperature display mode.

For peak and valley temperature detection, the time adjustment range is 0-999s. For

average temperature detection, the time adjustment range is 0-60s.

Set Menu 4-, 5-, 6- Process Alarm Settings:

Use these menus to set the upper and lower alarms and the alarm deadband (to prevent

alarm relay contacts “chatter”, this is the temperature range in which the alarm turns off

after being triggered).

Menu “-4”: Upper Limit Alarm, default 1400°C

Menu “-5”: Lower Limit Alarm, default 350°C

Menu “-6”: Alarm Deadband coefficient, default 0.001

If temperature is lower than 350°C, lower alarm relay is triggered, “Lower” alarm LED

indicator turns ON

If temperature is higher than 1400°C, upper alarm relay is triggered. PNP output high

(24VDC) “upper” alarm LED indicator turns ON.

Note: Lower alarm may be set higher than the upper alarm

Deadband coefficient range 0 - 0.500

Lower alarm deadband= ±deadband coefficient*lower alarm setting

Upper alarm deadband= ±deadband coefficient*upper alarm setting

1. Use the “up” and “right” arrow buttons to adjust the alarm settings. Alarm settings are °C

(or °F).

2. Press ENTER to return to the temperature display mode.

Set Menu 7-, 8- Current/Voltage Analog Output Span Settings:

Set the linear temperature span to correspond with the current or voltage output of the

pyrometer.

Menu 7- : Lower Temperature for the output span, default 350°C

Menu 8- : Upper Temperature for the output span, default 1400°C

1. Use the “up” and “right” arrow buttons to adjust the span settings

2. Press ENTER to return to the temperature display mode.

Page 21: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 20

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Use the following guidelines to calculate the current/voltage output that corresponds

to temperature:

Voltage output (0-10 V) Current output (4-20mA)

T= present measured target temperature

TL = output span lower limit temperature

TH = output span upper limit temperature

Example:

Measured temperature is 943oC; output span is 4 to 20 mADC proportional to 700 to 1400

oC.

Calculated current output at measured temperature is

Set Menu 9- Aiming Laser Setup:

Aiming laser will shut off when the internal sensor tip temperature is more than 70 o

C.

Four setups available: OFFd, OFFn, ON d and ON n. See below for details. This function selects

whether the laser switch function is open or closed when the temperature limit is exceeded.

OFFd: Turns the aiming laser off. It will automatically turn back on when the target

temperature is greater than the lower limit setup value.

OFFn: Turns aiming laser off.

ON d: Turns aiming laser on. After 30 minutes, the laser will turn off if the target temperature

is below the lower limit setup value.

ON n: Turns aiming laser on. After 30 minutes, laser will turn off.

1. Use SET and ENTER buttons to select the Aiming Laser Switch (Menu “9-“).

2. Use the “up” and “right” arrow buttons to change the setting. The factory default is “OFFd”.

3. Press ENTER to return to the temperature display mode.

Display Menu 10- Enclosure Temperature:

This feature only displays the current enclosure temperature for diagnostic purpose.

If the enclosure temperature is more than 70 oC, the display shows “EEE1”.

Page 22: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 21

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

If the enclosure temperature is lower than -10 oC, the display shows “EEE2”.

Either of these conditions exists, steps must be taken to adjust the ambient temperature.

Display Menu 11- Two Color Ratio

This function only displays the ratio of the current two wavelength signals for service

diagnostic purposes.

Set Menu 12- Zero Compensation Gain

This feature is used to adjust the detector gain to compensate for targets with a starting

temperature below the specified measurement limits of the Logika PM-3514 pyrometer.

Use SET and ENTER buttons to select the Zero Compensation Coefficient (“Menu 12-“)

The four digit number shows the coefficient; the first number represents the coefficient

polarity (1 = positive and 0 = negative). The remaining three numbers show the coefficient

value.

Use the “up” and “right” arrow buttons to change the Zero Compensation Coefficient. This

value is always set below 1.00. The factory default for the Zero Compensation Coefficient code

is 10.10 which equals +0.10.

Press ENTER to return to the temperature detection mode.

No UsePolarity1: Positive0:Negative

Coefficient

Value X.XX

Figure 5.6 Zero Coefficient Gain setup

Set Menu 13- Two-color mode signal attenuation ratio

Attenuation ratio is a threshold for verifying the validation of the measured temperature.

Check Figure 5.7 for the term of attenuation ratio in this section. After the target temperature

is calculated by the pyrometer, it calculates each wavelength’s emissivity at this temperature.

For each wavelength, the attenuation ratio (100% – Emissivity value) should not be greater

than the setup value. If either calculated attenuation ratio is higher than the setup value,

means the pyrometer sensor didn’t receive enough radiation energy at this wavelength and

Page 23: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 22

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

results final temperature not accurate, the LEDs display the lower limit temperature of the

pyrometer.

This function is only used in two-color mode. The setup range is 20%-100% and the default

setup value is 95%. If the attenuation ratio is set greater than 95%, this function will shut

down; and again, the lower limit temperature will be displayed. Next table gives reference for

different attenuation ratio setup values vs. different sizes of target.

Target size(Diameter) Accurate measurement

attenuation ratio

Simplified measurement

attenuation ratio

2mm-10mm 95% 95%

10mm-30mm 90% 95%

30mm-100mm 80% 95%

>100 mm 70% 70%

Table 5.2 Attenuation Ratio Setup Value

The values above are for reference only when setting up the attenuation ratio. The value

should be adjusted depending on application environment and target size.

1. Use SET and ENTER buttons to select the signal attenuation ratio setup (“Menu 13-“)

2. Use the “up” and “right” arrow buttons to change the value.

3. Press ENTER to return to the temperature display mode.

6001112

8001472

9001652

10001832

11002012

12002192

13002372

14002552

10

20

30

40

50

60

70

80

90

100

Attenuation ratio can beset at 95% (target can beobstructed to 95%)

Ma

x a

llo

we

d a

tte

nu

ati

on

ra

tio

(%

)

°C°F

7001292

350-400662-752

1

2

5

10

20 Attenuation FactorAttenuation FactorAttenuation FactorAttenuation Factor

500932

Target Temperature

Figure 5.7 Two-color mode signal attenuation ratio

Page 24: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 23

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

In Figure 5.7

Attenuation Ratio=1-Emissivity Attenuation Factor= �

��������

Check the term Emissivity in Chapter 5.2 Set Menu 1 Emissivity/Slope adjustment

Set Menu 14- RS485 COM Port Address

1. Use SET and ENTER buttons to select the Menu “14-“

2. The pyrometer RS485 COM port address is displayed from 0-25.

3. Use the “up” and “right” arrow buttons to change the setting.

4. Press ENTER to return to the temperature display mode.

5.

Set Menu 15- RS485 COM Port Baud Rate

1. Use SET and ENTER buttons to select the Menu “15-“

2. The pyrometer RS485 COM port Baud Rate is displayed.

0-4800

1-9600

2-19200

3-38400

4-57600

5-115200

3. Use the “up” and “right” arrow buttons to change the setting.

4. Press ENTER to return to the temperature display mode.

5.3 RS485 Command and Protocol

Send Command: Sent from Computer or other control units, 10 ASCII bytes.

<Start byte> <Address bytes> <Command 2 bytes> <Data 5bytes> <CR>

Byte 1 Byte 2 Byte 3,4 Byte 5-9 Byte 10

* A-Z

Start Address Command Data 1-5 CR

Note: For the Address 0-25 displayed on pyrometer, uses capital letter A-Z to represent the

address instead for easy command input, see the next table for details.

Page 25: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 24

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Pyrometer Address 00 01 02 03 04 05 06 07 08 09 10 11 12

Command Address A B C D E F G H I J K L M

Pyrometer Address 13 14 15 16 17 18 19 20 21 22 23 24 25

Command Address N O P Q R S T U V W X Y Z

CR is verifying code:

CR=byte2^ byte3^ byte4^ byte5^ byte6^ byte7^ byte8^ byte9

For example:

If send *ADT00000<CR>, answer is *ADT0350C<CR>

<CR>= 0x41^ 0x44^0x54 ^ 0x30^0x30 ^0x30 ^0x30 ^ 0x30=0x61,

<CR>=a。

In the next table, the COM port address is 0 (A).

Command Description

SD One/two color selection

*ASDx0000<CR>

x=1 One color; x=2 Two Color

SL Two color Slope

*ASLxxxxx<CR> xxxxx: 0.850~1.150

SE One color Emissivity

*ASExxxxx<CR> xxxxx: 0.100~1.000

SM Sample mode setup

*ASMA0000<CR> ;Average sample mode

*ASMP0000<CR> ;Peak sample mode

*ASMV0000<CR> ;Valley sample mode

RT Output signal integration time setup

*ARTxxxxx<CR>

xxxxx: For peak and valley range 000.0-999.0;

Average:000.0-060.0;

DM Temperature Unit Display Mode

*ADMC0000<CR> ;Display Celsius

*ADMF0000<CR> ;Display Farhenheit

Page 26: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 25

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

LU Upper Limit Alarm Temperature

*ALUxxxxx<CR>

xxxxx: 0350C(0662F)-1400C(2552F);

LD Lower Limit Alarm Temperature

*ALDxxxxx<CR>

xxxxx: 0350C(0662F)-1400C(2552F);

AS Analog span start temperature

*AASxxxxx<CR>

xxxxx: 0350C(0662F)-1400C(2552F);

AE Analog span end temperature

*AAExxxxx<CR>

xxxxx: 0350C(0662F)-1400C(2552F);

LS Aiming laser state setup

*ALS10000<CR> ;ON d

*ALS20000<CR> ;ON n

*ALS30000<CR> ;OFF d

*ALS40000<CR> ;OFF n

For ON d, ON n, OFF d, OFF n see 5.2 set menu 9 for details.

DB Alarm Deadband setup

*ADBxxxxx<CR>

xxxxx: 0.000-0.500;

SF One color full scale adjustment

*ASF0xxxx<CR>

xxxx: 0.30-3.00;

SO

One color zero point compensation adjustment

*ASOxxxxx<CR>

xxxxx: -1.00-+1.00;

DF Two color full scale adjustment

*ADFxxxxx<CR>

xxxxx: 0.990-1.010;

DT Display current measured temperature

*ADT00000<CR>

Answer

*ADTxxxxC(F)

xxxx: 0350-1400(0662-2512)

DA Display current ambient temperature

Page 27: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 26

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

*ADA00000<CR>

Answer

ADAxxxxC(F)

xxxx: -100----+100

VR Display current software version

*AVR00000<CR>

Answer

AVRxxx00

xxx: 1.0

BR

Baud Rate setup

*ABRx0000<CR>

x=0-5

0---4800;

1---9600;

2---19200;

3---38400;

4---57600;

5---115200;

RS485 Com protocol: 8 bit, 1 stop, no parity

PA Display setup parameter

*APA00000<CR>

Answer:

*ASDx0000<CR>

*ASEX.XXX<CR>

*ASLX.XXX<CR>

*ASMX0000<CR>

*ART00X.X<CR>

*AMC0000<CR>

*ALU1400C<CR>

*ALD0350C<CR>

*ADB0.001h<CR>

*AAS0350C<CR>

*AAE1400C<CR>

*ALSX0000<CR>

*ASF0X.XX<CR>

Page 28: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 27

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

*ASO-X.XX<CR>

*ADF1.000<CR>

*ABRX0000<CR>

*AVR1.000j

5.4 Debugging

Adjust emissivity according to material properties

(a) Use a thermocouple to measure the real temperature of the object, and then adjust

the slope until the test result is the same as the thermocouple.

(b) If the object is painted black, the emissivity can reach 0.98. Test the adjacent area,

and adjust the slope until the test result is the same as the thermocouple. Now the

slope has been properly adjusted.

Page 29: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 28

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Table 5.2 PM-3514 Keypad Operation

Function Description Default Key

SET Setup Parameters Display parameters from 1- to 13- each time button is pressed.

Press ENTER to Exit N/A SETUP

1- Slope/Emissivity

Adjustment

Single Color: Emissivity Rate, Range 0.10-1.00

Two Color: Temperature Slope: 0.850-1.150

0.95

1.000

2- Display/Sample

Mode

1 = Peak 2 = Valley 3 = Average 3

3- Response Time 0.0-999.9s adjustable. 0s=20ms, internal setup value, real time

display 0.5

4- Upper Limit Alarm Modify the Upper limit alarm temperature value 1400°C

5- Lower Limit Alarm Modify the Lower limit alarm temperature value 350°C

6- Alarm Deadband Modify alarm deadband value 0.001(0.1%)

7- Analog Output

Span Lower Limit

Temperature corresponds to linear analog current/voltage

output span lower value 350°C

8- Analog Output

Span Upper Limit

Temperature corresponds to linear analog current/voltage

output span upper value 1400°C

9- Aiming Laser 4 setups available: OFFd, OFFn, ONd, ONn. See setup menu 9-

for details. OFFd

10- Enclosure

Temperature

Setup constant at 40°C. If display is higher than 50°C, the

pyrometer should be cooled down N/A

Display

only

11- Two Color Signal

Ratio

Only used for pyrometer diagnostic purposes. In two color

mode, select 3(average) in setup 3-, the display value is the

ratio of the two wavelength signals

N/A Display

only

12- Zero

compensation

gain

Adjusts the pyrometer gain to compensate for targets with a

starting temperature below specified measurement limits. First

digit on left 1=positive, 0=negative. The remaining 3 numbers

are the coefficient value. The value should always be set below

0.30

10.10 (+0.1)

13- Two Color Signal

Attenuation Ratio

Only used in two color mode. If setup value is greater than

95%, this function is off. See setup menu 13- for details. 95%

14- RS485 COM

Address

Used for pyrometer communicates by RS485 0

15- Baud Rate 0-4800; 1-9600; 2-19200; 3-38400; 4-57600; 5-115200 1

Page 30: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 29

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

6 Maintenance

6.1 Regular Maintenance

Regular attention to the following will ensure accurate and prolonged operation of the

pyrometer:

Lens Cleaning - Routinely check the lens glass for dust or oil residue. If necessary, clean glass

with alcohol and lens paper or soft cloth.

Calibration - Occasional calibration using the temperature slope adjustment (see “Set Menu 1“)

will ensure that maximum accuracy is maintained.

6.2 Instrument Return

Contact us at 1-888-856-4452 (1-888-8LOGIKA) with the Serial Number of your pyrometer

before you return our product. If the problem cannot be resolved by telephone or email, we

will provide you with a return authorization number.

Please return to: Logika Technologies Inc.

2-2857 Sherwood Heights Drive,

Oakville, ON, L6J 7J9

Canada

Do not return the pyrometer without a return authorization number. If the product is out of

warranty, we will provide a repair estimate and then complete the repairs upon your approval.

Page 31: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 30

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

7 Appendix

Figure7.1 PM-3514-024 Dimension

Figure7.2 Junction Box Mechanical Dimension

Page 32: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 31

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

+24V OUT

0V OUT

0-10V

mA

GND

PNP

LOWER

LOWER

UPPER

UPPER

J1

J2

J3

Filter

RS485 A

RS485 B

0-10V

mA

GND

PNP

LOWER

LOWER

UPPER

UPPER

+

-

E

+24V IN

EARTH

0V IN

RS485 A

RS485 B

Figure7.3 24V DC-DC Junction Box

J3

L

EARTH

N

J1

+24V OUT

0V OUT

0-10V

mA

GND

PNP

LOWER

LOWER

UPPER

UPPER

J2

+24V OUT

0V OUT

0-10V

mA

GND

PNP

LOWER

LOWER

UPPER

UPPER

RS485 B

RS485 A

85-260VAC

24VDCConverter

DC Filter

RS485 B

RS485 A

Figure7.4 AC-24VDC Junction Box with

Page 33: PM-3514 Manual 2 - Logika Techlogikatech.com/wp-content/uploads/downloads/2013/12/PM-3514-Manual-2.1.pdf · Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev

Web: www.logikatech.com Phone: 905-829-5841 Fax: 905-829-8787 Email: [email protected]

P a g e | 32

Logika Technologies PM-3514 Two-color Pyrometer Operator’s Manual Rev. November 13

Figure7.5 PM-3514 control panel layout

LOGIKAUpper LimitAlarmLower LimitAlarm △ △TechnologiesFront

Rear Pin Connection

1

2

3

4

5

6

7

8

9

10

11

12

13

Voltage In

Current In +

-

+

Signal Ground

L

E

N

DC +24V Out

0V Out

Figure7.6 Universal Temperature Display mechanical dimension