User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed below. 1. Applicable User’s Manual and Page IM No. (Edition) Page Revised contents IM 01C50T01-02EN(3) YTA610 and YTA710 Temperature Transmitters (HART Protocol) Add CAUTION. Add *2 and *3. 3-15 3.5.3 Burnout Function 3.5.9 Burst Mode Table 4.1 List of Alarms Table 4.1 List of Alarms Table 4.2 Output Operation 4.3 Diagnostics Function Applicable part 4-3 4-4 Add *2 and *3. Change Terminal Sensor Fail in Table4.2. Add NOTE. 4-2 3-18 Add NOTE. Add CAUTION.

Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

  • Upload
    others

  • View
    1

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

User’sManual

Manual Change No. 18-0028-E

Oct. 19, 2018

Please use this manual change with the user’s manuals listed below.

1. Applicable User’s Manual and Page

IM No. (Edition) Page Revised contents

IM 01C50T01-02EN(3)

YTA610 and YTA710 Temperature Transmitters (HART Protocol)

Add CAUTION.

Add *2 and *3.

3-15 3.5.3 Burnout Function

3.5.9 Burst ModeTable 4.1 List of Alarms Table 4.1 List of Alarms Table 4.2 Output Operation 4.3 Diagnostics Function

Applicable part

4-34-4

Add *2 and *3.Change Terminal Sensor Fail in Table4.2.

Add NOTE.

4-23-18

Add NOTE. Add CAUTION.

Page 2: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

<3. Parameter Setting > 3-15

IM 01C50T01-02EN

(2) Auto Release Time

ThisparametercanbeusedtosettheautoreleasetimeofLoopTestandSimulation(See3.5.14).• ProceduretocallupthedisplayDevicesetup→Diag/Service→TestDevice→AutoReleaseTimeSelectfrom10minutes,30minutes,60minutes,3hours,6hoursor12hours.Defaultsettingis10minutes.

NOTEIfthetransmitterisequippedwiththeintegralindicator,theLCDdisplaysF.O.

3.5.3 Burnout Function

(1) Sensor burnout

Configuretheburnoutmodeinthecaseofsensorfailureordisconnection.• ProceduretocallupthedisplayDevicesetup→Basicsetup→AOBasicSetup→AOSensorBODir:SelectHigh,Low,UserConfigorOff.→AOSensorBOUnit:SelectmAor%→AOSnsBOUsrVal:Outputsusersetvaluewithin3.6to21.6(mA)or-2.5to110(%)Whenthesensorfailureisdetected,thetransmitteroutputsoneofthefollowingvalues.Low:Outputs3.6mAHigh:Outputs21.6mAUser(mA):OutputsusersetvalueinmA.

Settablewithin3.6to21.6mA.User(%):Outputsusersetvaluein%.

Settablewithin-2.5to110%.Off:TheburnoutoutputisNOTdefined

Duringsensorburnouttime,anyofAL.09,AL.10to13orAL.22alarmmessageisgenerated.(SeeChapter4fordetails.)

(2) Burnout in hardware failure

TheoutputstatusofthetransmitterinhardwarefailureissetusingaslideswitchontheMAINassembly.Thecurrentsettingcanbecheckedinparameter“PVAlrmtype”.• Proceduretocallupthe“AOAlrmtypeparameter”displayDevicesetup→Basicsetup→AOBasicSetup→AOAlrmtypHi:Morethan110%output(21.6mA)Lo:Lessthan-5%output(3.2mA)

3.5.4 Integral Indicator Display ModeWhenanintegralindicatorisspecifiedwiththetransmitter,thedisplayitemsanddisplaycyclecanbeconfigured.

(1) Process Variable Display

ProcessvariablesthatcanbedisplayedinYTAareshownintheTable3.1.Acycleofuptofourdisplays(DispOut1,DispOut2,Disp%,DispmA)canbeshownbyassigningvariablestotheparameters.Indicatevaluesofprocessvariableswiththeindicationlimits–99999to99999.

F0305.ai

P SP T F

%

Table 3.1 Process Variable Display

Process variable Disp Out1 Disp Out2Sensor1 Sensor1−Terminal Terminal Sensor2 Sensor2−Terminal Sensor1−Sensor2 Sensor2−Sensor1 Averege SensorBackup PV SV TV QV Notused ─

NOTEEven if abnormality or disconnection of the terminal block temperature sensor occurs, AL22 may not be output and AL04 may be generated.

CAUTIONThere is a time lag between sensor failure and sensor abnormality detection.Since the failed measurement value is output, the current output also becomes undefined.When an error is detected, it becomes the output set with this parameter.However, when sensor burnout is set to OFF, it may be held with indefinite measurement value, so please understanding this point.

Page 3: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

<3. Parameter Setting > 3-18

IM 01C50T01-02EN

operation,eveniftheSensor1recovers,thetransmitterwillcontinuetouseSensor2untilthebackupoperationisresetbyparameterorpoweristurnedoff.IfSensor2failsduringbackupoperation,thetransmitterwillsend“BackupSns2Fail”totheintegralindicatorandHARTcommunicatorandoutputthe“Sensorburnout”value.

Buttwosensorsshouldbesametype.Sensorvalue:DisplayinBackup(1) ToenableSensorbackupfunctiona) Setthesensortype,wireandunitforSensor1and2.(See3.4.1)AnySensortypeexcept“Non-connection”canbeselected.

b) MaptheSensorBackupasPV.c) SetthesensorburnouttypetoHigh,Loworusersettingvalue.Ifthesensorburnouttypeissetto“off”,thesensorfailurewillnotbedetectedandthebackupfunctionwillnotact.

Table 3.2 Burst Parameters

Command Parameter Burst Command Burst Msg Trigger

ModeBurst Trigger

Source Burst Trigger Units

PV Cmd1:PV Continuous ─Window PV Dependonthe

assignedvariableto PV

RisingFallingOn-change

LoopCurrentandPercentRange

Cmd2:%range/current Continuous ─Window %range %RisingFallingOn-change

PV,SV,TV,QV Cmd3:Dynvars/current Continuous ─Window PV Dependonthe

assignedvariableto PV

RisingFallingOn-change

DeviceVariable Cmd33:Devicevars Continuous ─ ─Window TopofBurstDevice

VariableDependsonmappingRising

FallingOn-change

DeviceVariablewithstatus

Cmd9:Devicevarsw/status Continuous ─ ─Window TopofBurstDevice

VariableDependsonmappingRising

FallingOn-change

AdditionalDevicestatus

Cmd48:ReadAdditionalDeviceStatus

Continuous ─ ─On-change Allstatus ─

(2) TomakethetransmitterreturntousingSensor1inBackupOperation

b) ProceduretocallupthedisplayDevisesetup→Detailedsetup→SensorSetup→SensorBackup→recoverSensor

c) ConfirmthatSensorBackupStateis“Sensor1 Active”.

3.5.9 Burst ModeWhentheBurstmodeisenabled,thetransmittercontinuouslysendsuptothreedatalistedinTable3.2.WhentheBurstmodeissetto“ON”,transmittercontinuouslysendsalarmsignalalso.WhenchangingthesettingofBurstmode,set“Off”totheBurstmode.Defaultsettingis“Off”.

CAUTIONThere is a time lag between sensor failure and sensor abnormality detection.Since the failed measurement value is output, the current output also becomes undefined.For this reason, until failure of Sensor 1 switches to Sensor 2, or ”Output state at sensor burnout” due to failure of Sensor 2 The output becomes undefined until switching to the output.When an abnormality is detected, it is output according to the measured value of Sensor 2 or "Output state at sensor burnout".

a) ConfirmthattheSensor1isrecovered

• SpecificationsSpecifications:DualinputonlySensorType:Anytypeavailable

Page 4: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

<4. Diagnostics > 4-2

IM 01C50T01-02EN

• ProceduretocallupthedisplayDevicesetup→Processvariables→DynamicVariable→PV→ PDQ:PVdataquality→ LS:PVlimitstatusDevicesetup→Processvariables→DeviceVariables→Sensor1→ Sns1PDQ:Sensor1dataquality→ Sns1LS:Sensor1limitstatusAboveprocedureisthesameaboutothervariables.

Table 4.1 List of Alarms

Indicator HART display Cause Output operation during alarm Action

AL.00 CPUFail MainCPUfailed. AccordingtotheBurnoutinHardwarefailureoutputCommunicationdisabled

Replacetheamplifier.

AL.01 SensorNVFail Sensornon-volatilememoryverifiesalarm.

AccordingtotheBurnoutinHardwarefailureoutputCommunicationenabledAL.02 TempNVFail Non-volatilememoryof

TEMPASSYverifiesalarm.AL.03 ADConvFail Inputcircuithardwarefailed.AL.04*2 MainRvrsCalcFail MAINASSYmemoryfailed.AL.05 TempRvrsCalcFail TEMPASSYmemoryfailed.AL.06 TempVoltageFail TEMPASSYvoltagefailed.AL.07 CommNVFail Communicationnon-volatile

memoryverifiesalarm.AL.08 TempNVWarning Non-volatilememoryofthe

TEMPASSYverifiesalarm.Continuetooperateandoutput

AL.09 IntCommFail Internalcommunicationfailed.

AccordingtotheSensorBurnoutoutputCommunicationenabled

AL.10 Sensor1Failure Sensor1failedordisconnectedfromterminalblock.

ReferTable4.2. • Checksensorfordamage.

• Checkterminalconnection.AL.11 Sensor2Failure Sensor2failedor

disconnectedfromterminalblock.

AL.12*1 Sensor1Short Sensor1short-circuited.AL.13*1 Sensor2Short Sensor2short-circuited.AL.14*1 Sensor1 Corrosion Sensor 1 corroded. Continuetooperateand

outputChecksensor,terminal,andcable.AL.15*1 Sensor2 Corrosion Sensor 2 corroded.

AL.20 Sensor1SignalError Sensor1inputisoutofmeasurablerange.

• Checksensorconnection.

• Checksensortypeselected.

AL.21 Sensor2SignalError Sensor2inputisoutofmeasurablerange.

AL.22*3 TerminalSensorFail Terminalblocktemperatureisabnormal.Orterminalblocktemperaturesensorfailed.

ReferTable4.2 Replacetheamplifier.

AL.23 BackupSensor1Fail Duringsensorbackupoperation,Sensor1fails,ithasoutputSensor2.

Operatingtothebackupside.Whenthebackupsidealsofails,outputisaccordingtoburnoutsetting.

Checksensor1fordamage.

AL.24 BackupSensor2Fail Duringsensorbackupoperation,Sensor2fails.

Continuetooperateandoutput

Checksensor2fordamage.

Page 5: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

<4. Diagnostics > 4-3

IM 01C50T01-02EN

Indicator HART display Cause Output operation during alarm Action

AL.25 Sensor Drift Sensor drift Continuetooperateandoutput

Checksensorfordamage.

AL.26*1 Sensor1TempCycle TemperaturecyclingtimesofSensor1exceedsthethreshold

Resettempcycleofsensor1.

AL.27*1 Sensor2TempCycle TemperaturecyclingtimesofSensor2exceedsthethreshold.

Resettempcycleofsensor2.

AL.30 OutputTooLow PVvalueisbelowtherangelimitsetting.

Lowerlimit3.68mA(-2%) ChecktheLRVsettingandadjust.

AL.31 OutputTooHigh PVvalueisabovetherangelimitsetting.

Upperlimit20.8mA(105%) ChecktheURVsettingandadjust.

AL.40 Sensor1TempLow Measuredtemperatureofsensor1istolow.

Continuetooperateandoutput

Checksensortypeselected.

AL.41 Sensor1TempHigh Measuredtemperatureofsensor1istohigh.

AL.42 Sensor2TempLow Measuredtemperatureofsensor2istolow.

AL.43 Sensor2TempHigh Measuredtemperatureofsensor2istohigh.

AL.44 AmbTempLow Ambienttemperatureisbelow-40degreeC.

Useaheatertoraisetheambienttemperature,orresettoahigherlevel.

AL.45 AmbTempHigh Ambienttemperatureisabove85degreeC.

Takeenoughdistancefromthesourceofheat,orresettoalowerlevel.

AL.50 LRVTooLow LRVsettingisbelowthesensoroperatingtemperaturerange.

ChecktheLRVsetting.

AL.51 LRVTooHigh LRVsettingisabovethesensoroperatingtemperaturerange.

AL.52 URVTooLow SettingtheURVisbelowthesensoroperatingtemperaturerange.

ChecktheURVsetting.

AL.53 URVTooHigh SettingtheURVisabovethesensoroperatingtemperaturerange.

AL.54 SpanTooSmall Spanisbelowtherecommendedminimumspan.

ChecktheLRVandURVsetting.

AL.60 IllegalPVCfg ThereisasettingalarminthesensorthatismappedtothePV

HoldtheoutputofthepreviousalarmWhenitoccursatstartup,holdat4mA

CheckthePVmapping.

AL.61 IllegalSensor1Cfg Thereisafalsesettosensor1.

Continuetooperateandoutput

Checkthesensor1setting.

AL.62 IllegalSensor2Cfg Thereisafalsesettosensor2.

Checkthesensor2setting.

AL.70 OutputManualMode Underlooptestmode Settingvalueoflooptest Checklooptestmode.

*1: ApplicableonlyforYTA710.*2: AL04 may be generated even when abnormality or disconnection of the terminal block temperature sensor occurs.*3: Even if abnormality or disconnection of the terminal block temperature sensor occurs, AL22 may not be output and AL04 may be generated.

Page 6: Please use this manual change with the user’s manuals ... · User’s Manual Manual Change No. 18-0028-E Oct. 19, 2018 Please use this manual change with the user’s manuals listed

<4. Diagnostics > 4-4

IM 01C50T01-02EN

Table 4.2 Output Operation

Current output mapping Sensor1 Failure Sensor2 Failure Sensor1 Short*3 Sensor2 Short *3 Terminal

Sensor FailSENS.1 SensorBurnout *1 SensorBurnout *1 Output Hold*4

S.1-TER SensorBurnout *1 SensorBurnout *1TERM *1 *1 *1 *1SENS.2 *1 SensorBurnout *1 SensorBurnoutS.2-TER *1 SensorBurnout *1 SensorBurnoutS.1-S.2 SensorBurnout SensorBurnout SensorBurnout SensorBurnoutS.2-S.1 SensorBurnout SensorBurnout SensorBurnout SensorBurnoutAVG SensorBurnout SensorBurnout SensorBurnout SensorBurnoutBACKUP *2 *2 *2 *2

*1: Continuetooperateandoutput.*2: Whenbothsensor1andsensor2occurerror,outputisburnout.*3: ApplicableonlyforYTA710.*4: Fixed output value just before the alarm occurred.

4.2 Logging FunctionsTheYTAserieshavethecapabilitytostorethedatausefulfordiagnosisoftheproblems.

(1) Alarm log

Uptofouralarmhistoriesarestoredinthetransmittermemory,andcanbeseenusingthefollowingprocedures.Callupthedisplay:Devicesetup→Diag/Service→StatusLog

(2) Max/Min log

MinimumvalueandMaximumvalueoftheprocessvariablesarestoredinthetransmittermemoryandcanbechecked.a) Proceduretocallupthedisplay

Devicesetup→Diag/Service→SensorMax/MinLog

b) TocleartheloggeddataforProcessvariablesexceptTermvariables,selectresetSnsMax/Min.

NOTEThesevaluesmaybecomebigbydisconnectionofthesensor,etc.Pleaseresetinthiscase.

(3) Operation Time

Transmitter’soperationtimefromthefirstpower-oniscounted.Callupthe“OperationTime”displaytocheckthetime.Devicesetup→Detailedsetup→Deviceinformation→Date&Time→OperationTime

NOTEOperationTimeissavedevery15minutes.

4.3 Diagnostics FunctionTheYTAhasthefollowingdiagnosticfunctions.

(1) Hardware Failure

ThisfunctiondetectsFailuresofCPU,ADconverter,memory,etc.

(2) Sensor Failure

Thisfunctiondetectsanydisconnectionofsensors.Whenthecurrentisappliedandtheresistanceofasensorisabovethethreshold,thefunctionoutputsadisconnectionalarm.Parametersetting:NotnecessaryAlarmoutput:Whentheresistanceisabovethethreshold,thefunctionoutputsthealarmofSensor1Failure(AL.10)orSensor2Failure(AL.11).

(3) Terminal block Failure

ThisfunctiondetectsanydisconnectionatCJC.WhenitdetectsthattheresistanceofCJCisabovethethreshold,thefunctionoutputsanalarm.Parametersetting:NotnecessaryAlarmoutput:Whentheresistanceisabovethethreshold,thefunctionoutputsthealarmofTerminalSensorFail(AL.22).

NOTEEven if abnormality or disconnection of the terminal block temperature sensor occurs, AL22 may not be output and AL04 may be generated.

Output Hold*4

Output Hold*4

Output Hold*4

Output Hold*4

Output Hold*4

Output Hold*4

Output Hold*4

Output Hold*4