PIXEL CP SERIES X-RAY GENERATOR

OPERATOR’S MANUAL

Address any questions regarding X-ray generator operation to:

ITALRAY srl

Via del Parlamento Europeo 9/D 50018 SCANDICCI FIRENZE – ITALY –

Tel + 39 055 7228511 Fax + 39 055 7228512

www.italray.it [email protected]

Page 2: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 1

INTRODUCTION 1

OPERATOR’S

SAFETY AND SPECIFICATIONS 2

MANUAL CONSOLE CONTROLS 3

P.N. 740985 PROGRAMMING THE CONSOLE 4

ERROR CODES AND MESSAGES 5

GENERATOR EXPOSURE TABLES 6

MAINTENANCE SCHEDULE 7

X-RAY TUBE DATA 8

The original version of this manual (15 January 2002) has been drafted in the English language.

Page 3: Pixel cp operators manual

Page 2 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 4: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 3

INTRODUCTION 1 This high frequency X-ray generator is a component of a radiographic X-ray system used in hospitals and clinics. It is used with film-based systems or with DR flat panel detectors along with an imaging system for diagnostic radiographic imaging. It is a non-invasive device, and is designed to image the major systems of the body: Skull, shoulder, thorax, upper arm, lower arm, abdomen, pelvis, femur, knee, tibula-fibula, foot.

The generator features state-of-the-art computer-based control to ensure minimum patient dose, excellent reproducibility, and superior image contrast. The operator control functions are designed to be simple and user-friendly.

The APR (Anatomical Programming) and the optional AEC (Automatic Exposure Control) features give you controlled exposure factors, automatically optimized for the radiological study selected.

MAIN FEATURES

• Output power levels 32 to 80 kW. • Output capability 125 kV / 150 kV, dependant on model. • Smaller, lighter, modular design. • Constant dose output due to kV and mA regulation during exposures. • User-friendly controls. • Large LCD panel for APR and technique display. • User-friendly system configuration. • Programmed for APR operation, with manual override of technique factors. • Extensive self-diagnostics with operator prompt messages. • Time and mA / mAs selections are to IEC standards.

OPTIONS • Remote exposure hand switch. • AEC (Automatic Exposure Control). • Tomography. • Dual Speed Starter (400 / 480 VAC models only). • Various digital interfaces, dependent on model. • Dose-Area Product (DAP).

PATIENT POPULATION • Age: Infant to geriatric. • Weight:

∗ APR mode: Small, medium, large patients. ∗ Manual mode: Exposure factors are set manually, will allow weights from newborn to obese adult.

• Health: Patients requiring an X-ray may have conditions ranging from mild trauma to chronic, life threatening illnesses.

FREQUENCY OF USE

• Typical range 20 exposures / day to 1500 exposures / day (1 day = 24 hr period).

Page 5: Pixel cp operators manual

1 Introduction

Page 4 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

SAFETY NOTICE: This manual contains important safety information. An understanding of this information is critical to the safe operation of your equipment. Please ensure that you read the warning notices before using the equipment.

GENERATOR DUTY CYCLE LIMIT

NOTE: THE FOLLOWING SECTION CONTAINS IMPORTANT INFORMATION. PLEASE READ AND UNDERSTAND THIS MATERIAL BEFORE CONTINUING.

Internal X-ray generator components will heat up during normal use of the generator. This is similar to X-ray tube heating during normal generator operation. The amount of heat produced is proportional to the product of kV, mA, and time. Modern X-ray generators are designed to operate with the majority of X-ray tubes over their rated power ranges. They are designed for operating duty cycles consistent with practical patient examination routines that allow for reasonable cooling intervals between X-ray exposures. This X-ray generator has internal duty cycle monitoring to warn of excessive heat build-up. If the generator calculates that the next exposure will exceed the rated generator duty cycle limit, a warning message “GEN DUTY WARNING” will be displayed. Further exposures are inhibited at this point, and the generator must be allowed to cool sufficiently such that this message is no longer displayed. The generator also monitors the X-ray tube’s thermal switch, and will inhibit exposures when the tube reaches its thermal limit. It is the responsibility of the installer to implement and verify this interlock.

DAILY X-RAY TUBE WARM-UP PROCEDURE

CAUTION: THE FOLLOWING PROCEDURE PRODUCES X-RAYS. OBSERVE ALL SAFETY PRECAUTIONS TO PROTECT PERSONNEL.

Use this procedure when the generator is first turned on for the day, or when it has not been used for several hours. This procedure provides for exposures at medium power before the tube is used at maximum mA or kV values. This will reduce the possibility of damaging the anode and high voltage components. No test setup is required.

For maximum stability and reliability, use the following techniques at start up: Select the following: • Large focal spot. • 80 kV. • Normal 50/60 Hz anode rotation.

For a 300 kHU to a 400 kHU tube, use approximately 200 mAs per exposure. For a 200 kHU to 300 kHU tube use approximately 150 mAs per exposure. • Depending on the X-ray tube power rating, select either 100 mA or 200 mA. • Make one (1) to three (3) exposures (depending on tube loading) at 30-second intervals.

Page 6: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 5

SAFETY AND SPECIFICATIONS 2 Keep this operator’s manual with the equipment at all times, and periodically review the operating and safety instructions.

SAFETY / WARNING SYMBOLS

Warning symbol used to indicate a potential hazard to operators, service personnel or to the equipment. It indicates a requirement to refer to the accompanying documentation for details.

Radiation exposure symbol used on operator console. Lights to indicate that an exposure is in progress. This is accompanied by an audible tone from the console.

WARNING THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS, OPERATING INSTRUCTIONS AND MAINTENANCE SCHEDULES ARE OBSERVED.

Radiation warning label on console. Never allow unqualified personnel to operate the X-ray generator.

WARNING: PROPER USE AND SAFE OPERATING PRACTICES WITH RESPECT TO X-RAY GENERATORS ARE THE RESPONSIBILITY OF THE USERS OF SUCH GENERATORS. THE MANUFACTURER PROVIDES INFORMATION ON ITS PRODUCTS AND ASSOCIATED HAZARDS, BUT ASSUMES NO RESPONSIBILITIES FOR AFTER-SALE OPERATING AND SAFETY PRACTICES.

THE MANUFACTURER ACCEPTS NO RESPONSIBILITY FOR ANY GENERATOR NOT

MAINTAINED OR SERVICED ACCORDING TO THE SERVICE AND INSTALLATION MANUAL OR ANY GENERATOR THAT HAS BEEN MODIFIED IN ANY WAY.

THE MANUFACTURER ALSO ASSUMES NO RESPONSIBILITY FOR X-RAY RADIATION

OVEREXPOSURE OF PATIENTS OR PERSONNEL RESULTING FROM POOR OPERATING TECHNIQUES OR PROCEDURES.

WARNING: THIS X-RAY UNIT MAY BE DANGEROUS TO PATIENT AND OPERATOR UNLESS SAFE EXPOSURE FACTORS, OPERATING INSTRUCTIONS AND MAINTENANCE SCHEDULES ARE OBSERVED.

CAUTION: DO NOT EXCEED THE TUBE MAXIMUM OPERATING LIMITS SHOWN IN THE X-RAY TUBE DATA SECTION AT THE END OF THE OPERATOR’S MANUAL. INTENDED LIFE AND RELIABILITY WILL NOT BE OBTAINED UNLESS GENERATORS ARE OPERATED WITHIN PUBLISHED SPECIFICATIONS.

Page 7: Pixel cp operators manual

2 Safety and Specifications

Page 6 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

X-ray radiation exposure may be damaging to health, with some effects being cumulative and extending over periods of many months or even years. X-ray operators should avoid any exposure to the primary beam and take protective measures to safeguard against scatter radiation. Scatter radiation is caused by any object in the path of the primary beam and may be of equal or less intensity than the primary beam that exposes the film. No practical design can incorporate complete protection for operators or service personnel who do not take adequate safety precautions. Only authorized and properly trained service and operating personnel should be allowed to work with this X-ray generator equipment. The appropriate personnel must be made aware of the inherent dangers associated with the servicing of high voltage equipment and the danger of excessive exposure to X-ray radiation during system operation. • Operators should be trained on the use of the equipment within the facility. They should have the knowledge

in radiology, and the skills, attitude, and judgment to safely and effectively operate the equipment.

• Operators are trained on the use of the X-ray generator by the installer of the X-ray generator. The operator’s manual may be used as a training aid.

• Wear protective clothing. Protective aprons with an equivalent of a minimum of 1/64” (0.35 mm) of lead are recommended.

• To protect the patient against radiation, always use radiation protection accessories in addition to devices which are fitted to the X-ray equipment.

• Keep as large a distance as possible away from the object being exposed and the X-ray tube assembly.

• Never operate this X-ray equipment in areas where there is a risk of explosion. Detergents and disinfectants, including those used on patients, may create explosive mixtures of gases. Please observe the relevant regulations.

• The operator console, or anything electrically connected to it, must never be used within 6 ft (1.8 m) of the patient environment.

• Do not place liquids (coffee, beverages, flowers, etc) on the control console or generator main cabinet.

• Always ensure adequate ventilation around the control console and generator main cabinet. Do not operate the equipment near curtains, drapes, etc which may block the ventilation slots.

• The control console must be located inside an X-ray shielded control booth within the X-ray room, or outside the X-ray room.

• The control console is intended for fixed mounting. It is not a portable device.

• Do not operate the console or generator main cabinet in direct sunlight or near any heat sources.

• Do not operate the console near strong magnetic fields (microwave ovens, speakers, etc), and avoid routing the console cables near these devices.

• The console and generator main cabinet must be operated in locations that are clean (free of excess dust, dirt, debris, etc), stable (free of vibration), and secure such that the console cannot slip or tip.

• Only trained maintenance staff may remove the covers of the generator cabinet and the control console.

Do not connect unapproved equipment to the rear of the console. J3 is for connection of an external hand switch, J4 is a serial port for use by an external computer, and J8 is for the interconnect cable to the main cabinet. INCORRECT CONNECTIONS OR USE OF UNAPPROVED EQUIPMENT MAY RESULT IN INJURY OR EQUIPMENT DAMAGE.

Page 8: Pixel cp operators manual

Safety and Specifications 2

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 7

APPLICABLE STANDARDS

This series of X-ray generators comply with the regulatory requirements and design standards in this section as follows: • VZW2555 series: Only the standards marked with an asterisk * under SAFETY.

• VZW2556 series: All standards in this section.

A) SAFETY

• * FDA Center for Devices & Radiological Health (CDRH) - 21 CFR subchapter J (USA).

• * Radiation Emitting Devices Act - C34 (Canada).

• Medical Devices Regulations (Canada). • EC Directive 93/42/EEC concerning Medical Devices (European Community).

• * EN 60601-1/IEC 60601-1, EN 60601-2-7/IEC 60601-2-7, CSA 601.1, UL60601.1. -Type of protection against electric shock: Class I equipment. -Degree of protection against electric shock: Not classified. -Degree of protection against harmful ingress of water: Ordinary equipment. -Mode of operation: Continuous operation with intermittent loading (standby - exposure). -Equipment not suitable for use in presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide.

• EN 60601-1-4/IEC 60601-1-4, ISO 14971, EN 60601-1-6/IEC 60601-1-6. NOTE: All referenced standards are considered to be at the latest adopted revision.

The CE Mark is a declaration by the manufacturer that the product complies with the requirements of the applicable European Union (EU) medical device directive and that the product has been subject to conformity assessment procedures as provided in that directive.

A CSA mark with the indicators “C” and “US” means that product is certified for both the U.S. and Canadian markets, to the applicable U.S. and Canadian standards.

Page 9: Pixel cp operators manual

2 Safety and Specifications

Page 8 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

B) EMC (EN 60601-1-2:2001/IEC 60601-1-2:2001) Guidance and manufacturer’s declaration – electromagnetic emissions

The VZW2556 series of X-ray generators are intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2556 series should assure that it is used in such an environment. Emissions test Compliance Electromagnetic environment - guidance

RF emissions CISPR 11

Group 1 The VZW2556 series of X-ray generators use RF energy only for their internal functions. Therefore, the RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.

RF emissions CISPR 11

Class A (The VZW2556 series of X-ray generators in

combination with shielded location)

Harmonic emissions

IEC 61000-3-2

Not Applicable

Voltage fluctuations/

flicker emissions IEC 61000-3-3

Not Applicable

The VZW2556 series of X-ray generators must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that exits the shielded location, a minimum RF filter attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230 MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the minimum at 230 MHz is 47dB.) The VZW2556 series is suitable for use in all establishments other than domestic and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.

NOTE It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specifications.

Page 10: Pixel cp operators manual

Safety and Specifications 2

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 9

Guidance and manufacturer’s declaration – electromagnetic immunity

The VZW2556 series of X-ray generators are intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment – guidance

Electrostatic discharge (ESD) IEC 61000-4-2

± 6 kV contact ± 8 kV air

± 6 kV contact ± 8 kV air

Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30%.

Electrical fast transient/burst IEC 61000-4-4

± 2 kV for power supply lines ± 1 kV for

input/output lines

± 2 kV for power supply lines ± 1 kV for

input/output lines

Mains power quality should be that of a typical commercial or hospital environment.

Surge IEC 61000-4-5

± 1 kV differential mode

± 2 kV common mode

± 1 kV differential mode

± 2 kV common mode

Mains power quality should be that of a typical commercial or hospital environment.

Voltage dips, short interruption,

and voltage variations on power supply

input lines IEC 61000-4-11

< 5 % UT

(> 95 % dip in UT) for 0.5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT)

< 5 % UT (> 95 % dip in UT)

for 5 s

< 5 % UT

(> 95 % dip in UT)for 0.5 cycle

40 % UT

(60 % dip in UT) for 5 cycles

70 % UT

(30 % dip in UT)

< 5 % UT (> 95 % dip in UT)

for 5 s

Mains power quality should be that of a typical commercial or hospital environment. If the user of the VZW2556 series X-ray generator requires continued operation during power mains interruptions, it is recommended that the X-ray generator be powered from an uninterruptible power supply or battery.

Power frequency (50/60 Hz)

IEC 61000-4-8

3 A/m 3 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment

NOTE: UT is the A.C. mains voltage prior to application of the test level.

Page 11: Pixel cp operators manual

2 Safety and Specifications

Page 10 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

Guidance and manufacturer’s declaration – electromagnetic immunity

The VZW2556 series of X-ray generators is intended for use in the electromagnetic environment specified below. The customer or the user of the VZW2556 series should assure that it is used in such an environment.

Immunity test

IEC 60601 test level

Compliance level

Electromagnetic environment - guidance

Conducted RF IEC 61000-4-6

3 Vrms 150 kHz to

80MHz

3 Vrms 150 kHz to

80MHz

The VZW2556 series of X-ray generators must be used only in a shielded location with a minimum RF shielding effectiveness and, for each cable that enters the shielded location, a minimum RF filter attenuation of 40dB from 30 MHz to 230 MHz and 47dB from 230 MHz to 1 GHz. (The minimum at 30 MHz is 40dB and the minimum at 230 MHz is 47dB.)

Radiated RF IEC 61000-4-3

3 V/m 80MHz to 2.5

GHz

3 V/m 80MHz to 2.5

GHz

Field strengths outside the shielded location from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than 3 V/m.a

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 These guidelines may not apply all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. NOTE 2 It is essential that the actual shielding effectiveness and filter attenuation of the shielded location be verified to assure that they meet the minimum specification. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and

land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the VZW2556 series of X-ray generators is used exceeds the applicable RF compliance level above, the X-ray generator should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the X-ray generator.

Page 12: Pixel cp operators manual

Safety and Specifications 2

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 11

ELECTROMAGNETIC COMPATIBILITY (EMC) In accordance with the intended use, some models of this series of X-ray generators comply with the European Council Directive concerning Medical Devices. The CE marking affixed to compliant products signifies this. One of the harmonized standards of this Directive defines the permitted levels of electromagnetic emission from this equipment and its required immunity from the electromagnetic emissions of other devices. It is not possible, however, to exclude with absolute certainty the possibility that other high frequency electronic equipment, which is fully compliant to the EMC regulations, will not adversely affect the operation of this generator. If the other equipment has a comparatively high level of transmission power and is in close proximity to the generator, these EMC concerns (the risk of interference) may be more pronounced. It is therefore recommended that the operation of equipment of this type such as mobile telephones, cordless microphones and other similar mobile radio equipment be restricted from the vicinity of this X-ray generator. OUTPUT PARAMETERS

kV range: 40 to 125 kV, 40 to 150 kV some models kV accuracy: ± (5 % + 1) over the full kV range. Time range: 1.0 to 6300 milliseconds. mAs range: 0.1 to 500 mAs (Pixel CP 450/ Pixel CP 550 32/40 kW)

0.1 to 630 mAs (Pixel CP 650 50 kW) 0.1 to 800 mAs (Pixel CP 850 65 kW) 0.1 to 1000 mAs (Pixel CP 1050 80 kW)

mAs accuracy: ± (10 % + 0.2) mAs. mA range: 10 to 400 mA (Pixel CP 450 32 kW)

10 to 500 mA (Pixel CP 550 40 kW) 10 to 630 mA (Pixel CP 650 50 kW) 10 to 800 mA (Pixel CP 850 65 kW) 10 to 1000 mA (Pixel CP 1050 80 kW)

Coefficient of linearity: 0.05 (station to station) mAs. Coefficient of reproducibility: kV, mAs ≤ 0.05. Lowest current – time product: 0.1 mAs. Loading factors which will allow lowest current - time product:

Refer to table 1 in section 6.

Page 13: Pixel cp operators manual

2 Safety and Specifications

Page 12 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

OUTPUT PARAMETER GENERATOR SERIES LOADING FACTOR

Maximum X-ray tube voltage and highest X-ray tube current at

that voltage

Pixel CP 450 32 kW Pixel CP 550 40 kW Pixel CP 650 50 kW Pixel CP 850 65 kW

Pixel CP 1050 80 kW

125 kV, 250 mA / 150 kV, 200 mA 125 kV, 320 mA / 150 kV, 250 mA

150 kV, 320 mA 150 kV, 400 mA 150 kV, 500 mA

Maximum X-ray tube current and highest X-ray tube voltage at

that current

Pixel CP 450 32 kW Pixel CP 550 40 kW Pixel CP 650 50 kW Pixel CP 850 65 kW

Pixel CP 1050 80 kW

400 mA, 80 kV 500 mA, 80 kV 630 mA, 80 kV 800 mA, 81 kV 1000 mA, 80 kV

Combination of X-ray tube current and X-ray tube voltage

resulting in highest output power

Pixel CP 450 32 kW Pixel CP 550 40 kW Pixel CP 650 50 kW Pixel CP 850 65 kW

Pixel CP 1050 80 kW

320 mA, 100 kV 400 mA, 100 kV 500 mA, 100 kV 630 mA, 103 kV 800 mA, 100 kV

Highest constant output power at 100 kV, 0.1 sec

Pixel CP 450 32 kW

Pixel CP 550 40 kW

Pixel CP 650 50 kW

Pixel CP 850 65 kW

Pixel CP 1050 80 kW

32 kW (320 mA, 100 kV, 0.1 s)

40 kW (400 mA, 100 kV, 0.1 s)

50 kW (500 mA, 100 kV, 0.1 s)

63 kW (630 mA, 100 kV, 0.1 s)

80 kW (800 mA, 100 kV, 0.1 s)

Nominal shortest irradiation time (AEC exposures)

All models 10 ms

ENVIRONMENTAL SPECIFICATIONS

OPERATING Ambient temperature range 10 to 40 °C (50 to 104 °F). Relative humidity 20 to 80%, non-condensing. Altitude -700 to 3000 meters (1100 to 700 hPa).

This information is provided to help you establish safe operating conditions for both you and your X-ray generator. Do not operate this X-ray generator except in accordance with information included in this section, and any additional information provided by the X-ray generator manufacturer and / or competent safety authorities.

Page 14: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 13

CONSOLE CONTROLS 3 OPERATOR CONTROL PANEL

I II III SMALL LARGEAEC mAs mA/mS

PREPAUX 1 AUX 2WALLBUCKY

NON-BUCKY

TABLEBUCKY

OFF ON

EXPOSE

MODE FOCALSPOT

AEC FIELD

MENU

ON

TEXT VERSIONCMP200_002B.CDR

I II IIImAs mA/mS

AUX 1 AUX 2

SYMBOL VERSION

1 5 2

34 2

1 5 2

243

1. Power ON and OFF buttons 2. PREP and X-RAY EXPOSE buttons, and PREP and X-RAY EXPOSURE indicators 3. Radiography controls and display 4. Image receptor buttons 5. Anatomical programming controls and display, BACK / FORWARD buttons, and

MENU (reset) button

Page 15: Pixel cp operators manual

3 Console Controls

Page 14 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

POWER, PREP, AND X-RAY EXPOSURE CONTROLS Power On, Power Off

OFF ON

OR

Press ON or to switch the X-ray generator on. The console will light up and a brief self - check will be performed. All front panel LEDs, including the X-ray exposure LED, should light briefly. This is an LED self - test and does NOT indicate the presence of X-RAYS. The speaker should briefly beep near the end of the console self-test. Press the MENU (reset) button to continue if any error messages are presented.

Press OFF or to switch the generator off. Prep, X-Ray Exposure and Exposure Indicator

PREP EXPOSE

OR

Press and hold the PREP or button to spin the rotor. The prep indicator LED will light when ready to make an exposure.

While pressing the prep or button, press and hold the EXPOSE or button to make an X-ray exposure.

The X-ray exposure indicator will light when an X-ray exposure is being taken.

Pressing the EXPOSE or button only will cycle the generator through prep and then exposure.

When the prep button is pressed, SPINNING

ROTOR will be displayed in the LCD display window.

When the prep cycle is complete, X-RAY READY

will be displayed in the LCD display window. During the X-ray exposure, X-RAY ON will be

displayed in the LCD display window.

Page 16: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 15

HAND SWITCH (Optional): Press the hand switch halfway to the PREP position. This will spin the rotor. Fully depress and hold the hand switch in the EXPOSE position to make an X-ray exposure. RADIOGRAPHY CONTROLS AND DISPLAY

1. Technique select button and indicators. 2. Field select buttons and indicators. 3. Film / Screen select button and indicators. 4. Focus select button and indicators. 5. kV up / down buttons and display (non-CM thickness mode).

CM thickness up / down buttons and display (CM thickness mode). 6. mA up / down buttons and display (non-CM thickness mode).

kV up / down buttons and display (CM thickness mode). 7. Density / ms / mAs up / down buttons and display. 8. ms / mAs readout. 9. AEC - ON indicator / post-mAs display.

Page 17: Pixel cp operators manual

3 Console Controls

Page 16 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

Technique Select Button and Indicators

AEC mAs mA/mS

MODE

OR

mAs mA/mS

This function may be disabled in programming. Press the MODE button until the appropriate LED lights to indicate the desired technique.

• AEC or for Automatic Exposure Control (1 point technique).

• mAs for mAs technique (2 point technique). • mA/mS for mA and ms technique (3 point technique).

PLEASE NOTE THE FOLLOWING WHEN IN CM-THICKNESS MODE:

• If you are in CM-thickness (AEC mode), and then select CM-thickness (mAs mode) via the technique selector button, the previously set backup mAs will be displayed for that CM thickness. Press the CM - or + buttons to update the display and use the normal mAs, not the backup mAs that was carried forward.

• If you are in CM-thickness (mAs mode), and then select CM-thickness (AEC mode) via the technique selector button, the displayed backup ms will be based on the programmed mAs for that CM thickness. Press the CM - or + buttons to set the backup ms display to use the appropriate backup ms, not the value that was carried forward.

Field Select Buttons and Indicators (for AEC only)

AEC FIELD

OR

Press the appropriate field select button(s) to select the desired field or combination of fields. The adjacent LED(s) will light to indicate the selected field(s). All three fields may not be deselected. Operates only when AEC is selected.

Caution: The exposures parameters, particularly the mA and ms, must be confirmed before making an exposure. These parameters may change when switching between AEC and non-AEC modes, depending on the AEC backup mode or generator programming.

Page 18: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 17

Film / Screen Select Button and Indicators. (Non-DR models only)

I II III

OR

I II III

In AEC mode, press the FILM SCREEN or button to select the desired film / screen combination. The appropriate LED will light to indicate the desired selection.

• I for film / screen I. • II for film / screen II. • III for film / screen III. This adjusts the dose to maintain the correct optical density. In non-AEC mode, indicates which film / screen has been programmed for the selected APR. The corresponding film / screen combination should then be used for that exposure.

Focus Select Button and Indicators

SMALL LARGE

FOCALSPOT

OR

Press the FOCAL SPOT button until the appropriate LED lights to indicate the desired focal spot.

• SMALL or for small focus.

• LARGE or for large focus. This function may be programmed for auto focal spot selection.

Page 19: Pixel cp operators manual

3 Console Controls

Page 18 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

Note: The rate of scrolling for the kV, mA, and ms display increases if the buttons are pressed continuously. To return to the slowest scroll rate, release the button(s) and start again.

kV up / down buttons and display (CM up / down buttons and display in CM-thickness mode)

To increase kV / CM thickness, press +. To decrease kV / CM thickness, press -. Displays demanded kVp of the exposure. Displays selected CM thickness when CM-thickness function is active.

mA up / down buttons and display (kV up / down buttons and display in CM-thickness mode)

To increase mA / kV, press +. To decrease mA / kV, press -. Displays demanded mA of the exposure. This is active for AEC in non CM-thickness mode or mA/ms techniques only. Displays kV when the CM-thickness function is active.

Page 20: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 19

Density / ms / mAs up / down buttons and display For non CM-thickness mode:

To increase density, ms, mAs, or backup ms/mAs press +. To decrease density, ms, mAs, or backup ms/mAs press -.

IN AEC MODE: • Displays DENS (density) when “fixed” AEC backup mode

is selected. Displays DENS and backup ms when “variable ms” AEC backup mode is selected. Displays DENS and backup mAs when “variable mAs” AEC backup mode is selected. DENS will be displayed above or below the backup ms / mAs in these modes, depending on setting. The + / - buttons only adjust the parameter displayed on top. To swap the density and the backup ms or mAs displays, press both the - and + density buttons. Refer to Setting AEC Backup ms or mAs for further details. The density range is typically -5 to +5. (Installer programmable to have a range of up to -8 to +8). This varies the optical density by changing the dose. The % dose change per density step change is installer programmable.

• Displays mAs when mAs is selected.

• Displays time (ms) when mA/ms is selected.

For CM thickness mode:

To increase density, ms, or mAs press +. To decrease density, ms, or mAs press -. • Displays DENS (density) when AEC is selected.

The density range is typically -5 to +5. (Installer programmable to have a range of up to -8 to +8). This varies the optical density by changing the dose. The % dose change per density step change is installer programmable.

• Displays mAs when mAs is selected.

• Displays time (ms) when mA/ms is selected.

Page 21: Pixel cp operators manual

3 Console Controls

Page 20 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

ms / mAs readout The significance of the ms/mAs readout varies, depending on mode selected: • In AEC mode, with fixed backup time selected, the AEC backup time is displayed. • In AEC mode, with either variable ms or variable mAs backup time, the backup mAs

or ms value is displayed. • In mAs mode, the calculated exposure ms will be displayed (based on the mAs

selected). • In mA/ms mode, the calculated mAs of the exposure will be displayed (based on the

mA and ms selected). AEC - ON indicator / post-mAs display • Displays AEC when AEC is selected. • Displays Post mAs after an AEC exposure.

Page 22: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 21

Setting AEC backup ms or mAs (applies to non CM-thickness mode only) The AEC backup mode is installer programmable for each receptor. If FIXED AEC backup time has been programmed for the selected receptor, the backup ms and mAs cannot be adjusted. If the selected receptor is programmed for FIXED mode, density (DENS) only is displayed, as in the top figure, below. If ms or mAs AEC backup has been selected, the AEC backup ms or mAs respectively may be operator adjusted. Follow the steps below. 1. If variable ms AEC backup has been programmed for the selected receptor, press the + or

- buttons shown below to change the AEC backup time.

2. If variable mAs AEC backup has been programmed for the selected receptor, press the + or - buttons shown below to change the AEC backup mAs.

3. To select density, press both the - and + density buttons. This will swap the density and AEC backup adjustments. Pressing both buttons again will swap the displays to allow further adjustment of AEC backup ms or mAs.

Setting AEC backup mAs (CM-thickness mode only) The AEC backup mAs in CM-thickness mode is set as described in section 4, under PROGRAMMING CM THICKNESS.

Page 23: Pixel cp operators manual

3 Console Controls

Page 22 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

IMAGE RECEPTOR BUTTONS AND INDICATORS

TEXT VERSION SYMBOL VERSION

AUX 1 AUX 2WALLBUCKY

NON-BUCKY

TABLEBUCKY

AUX 1 AUX 2

1. Table, with Bucky, image receptor select. 2. Table, non Bucky, image receptor select. 3. Wall Bucky image receptor select. 4. Auxiliary 1 image receptor select. 5. Auxiliary 2 image receptor select.

Note: Image Receptors (1-5) are depicted as the console is shipped from the factory. The receptors may have been reprogrammed during installation. As a result, the receptors may contain installer - applied icons that differ from those shown in this manual.

Table, with Bucky, Image Receptor

TABLEBUCKY

OR

Press this button to select the table Bucky image receptor. The adjacent indicator will light.

Page 24: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 23

Table, no Bucky, Image Receptor

NON-BUCKY

OR

Press this button to select the table non - Bucky image receptor. The adjacent indicator will light.

Wall Bucky Image receptor

WALLBUCKY

OR

Press this button to select the wall Bucky image receptor. The adjacent indicator will light.

Aux 1 Image Receptor

AUX 1

If CM thickness mode is enabled, this is used for CM thickness programming. Refer to section 4, Programming the Console, for details. If CM thickness mode is disabled, this receptor is active, and may be programmed by the installer.

Aux 2 Image Receptor

AUX 2

If CM thickness mode is disabled, this receptor is active, and may be programmed by the installer.

Page 25: Pixel cp operators manual

3 Console Controls

Page 24 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

LINEAR TOMOGRAPHY (OPTIONAL) If the tomography option is installed, any image receptor may be configured by the installer for tomo operation. For tomo operation, the tomo table supplies a start and stop command to start and stop the tomo exposure. The generator will terminate the exposure if the tomo backup time is exceeded. ANATOMICAL PROGRAMMING CONTROLS AND DISPLAY

CMP200_004B.CDR

OR

SYMBOLVERSION

1 1

2 3 4

5

6A 6B

7

THE ABOVE IS A TYPICAL DISPLAY. YOUR UNIT MAY DIFFER DEPENDING ON MODEL, SET-UP, AND PROGRAMMING

For non CM-thickness mode:

1(a) Anatomical section and anatomical view selection buttons. To select an APR technique: • From the main APR menu (shown in the above figure), press the button next to the

desired anatomical region. The APR display will show anatomical views within that region. Pressing button 6B (forward) if >> is displayed will show additional anatomical views within that region. Pressing button 6A (back) if << is displayed will return to the previous anatomical views.

• Press the button for the required anatomical view. The selected view will be highlighted and the image receptor symbol will be replaced with a normal size patient symbol.

• Press the function button again to toggle between normal, large and small patient sizes.

Page 26: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 25

• The LCD display will show the programmed X-ray technique factors. • You may override the programmed techniques by pressing the appropriate selection

button (the ability to override the AEC / mAs / mA/ms function may be disabled in programming). When any of the default techniques have been changed, the highlight will be replaced with a rectangular outline around the anatomical view. Simply overriding the programmed parameters will not change the programmed parameters. To make permanent changes to the APR, see CHAPTER 4: PROGRAMMING THE CONSOLE.

For CM thickness mode:

1(b) Anatomical section and anatomical view selection buttons. NOTE: CM thickness mode is only available if it has been enabled in programming.

This must be done by your service representative. CM thickness with AEC is only available on image receptors that are

programmed for “variable mAs” AEC backup. CM thickness (mAs) is available with all AEC backup configurations.

CM thickness is only available when an anatomical view is selected. If an anatomical view has not been selected, the technique selector button selects mA/ms-mAs-AEC in non CM-thickness mode. When an anatomical view has been selected, the technique selector button selects mA/ms-mAs-AEC in a combination of CM thickness and non CM-thickness modes as depicted below.

To select an APR technique: • From the main APR menu (shown in the figure on the previous page), press the

button next to the desired anatomical region. The APR display will show anatomical views within that region. Pressing button 6B (forward) if >> is displayed will show additional anatomical views within that region. Pressing button 6A (back) if << is displayed will return to the previous anatomical views.

Page 27: Pixel cp operators manual

3 Console Controls

Page 26 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

• Press the button for the required anatomical view. The selected view will be highlighted and the image receptor symbol will be replaced with a normal size patient symbol. The patient size cannot be changed in CM thickness mode.

• The LCD display will show the programmed X-ray technique factors. • The mid value of the programmed CM range will initially be displayed. Pressing + or

- will scroll through the available CM range for that anatomical view. • The kV and / or mAs is varied automatically as the CM is changed. • You may override the programmed techniques by pressing the appropriate selection

button (the ability to override the AEC / mAs / mA/ms function may be disabled in programming). When any of the default techniques have been changed, the highlight will be replaced with a rectangular outline around the anatomical view. Simply overriding the programmed parameters will not change the programmed parameters. To make permanent changes to the APR, see CHAPTER 4: PROGRAMMING THE CONSOLE.

2. Tube 1 display.

This indicator will display TUBE 1 at all times. 3. Anode heat indicator (%HU) readout.

Indicates the tube anode heat for the selected X-ray tube. An anode heat-warning message will be displayed at an installer-programmable level, typically 80 % of the maximum tube H.U. rating. Exposures that will exceed a value of typically 90 % of the tube H.U. rating will be inhibited (this is also installer programmable).

4. Time readout - 24 hour mode.

The current time is displayed in 24 hour format. 5. Status and operator message display area.

This area shows status, warning, and error messages. 6. Back (<<) and forward (>>) selection buttons 6A and 6B.

Press button 6A (BACK) and button 6B (FORWARD) to scroll forward or backward through the display. The symbols << or >> will be displayed when these buttons are active.

7. MENU button.

• Press the MENU button to return to the main menu when in one of the APR submenus. RESET will be displayed below the MENU button when this is active.

• Pressing this button will clear error messages.

NOTE: THE FACTORY-DEFAULT X-RAY TECHNIQUE FACTORS MUST BE CONFIRMED AS BEING SUITABLE FOR THE APPLICATION. THESE SHOULD BE EDITED AS REQUIRED, WITH THE UPDATED TECHNIQUE FACTORS SAVED AS PER CHAPTER 4.

Page 28: Pixel cp operators manual

Console Controls 3

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 27

DAP - DOSE AREA PRODUCT (OPTIONAL) The DAP function is only available if the optional DAP device is installed in the X-ray system, and if the DAP function has been enabled in programming. DAP Overview

• The DAP device requires a brief “settling” period after the generator is switched on.

The console will display the start-up screen during the DAP settling period. If the DAP does not stabilize within the normal settling time, the console will display the message DAP NOT READY, and will then proceed to the normal operating mode without the DAP display. If this happens, it indicates that the DAP did not stabilize in the normal time. You may wish to switch the generator off and then on again in an attempt to let the DAP stabilize. If the DAP is still not ready after a second attempt, you may choose to proceed without DAP and then contact your service representative.

• The DAP display is reset to zero when the generator is switched off, and then on

again.

Note: A reading of 0mGycm2 indicates that the DAP display has been reset at power-on or manually reset as described later in this section, and that no exposures have been taken since it was reset.

Note: Always check local regulations to determine how frequently the DAP device must be tested. The procedure to test the DAP follows later in this section.

DAP Display • Before proceeding with initial dose-area product measurements, ensure that the DAP

display is 0mGycm2. The DAP display may be manually reset to zero as described in DAP Reset / Test / Print, following.

• The DAP measures and displays cumulative dose-area product. • The maximum reading on the DAP display is 9999999. An error message will be

presented when this limit is reached.

Page 29: Pixel cp operators manual

3 Console Controls

Page 28 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

DAP Reset / Test / Print The DAP display may be reset to zero, a manual DAP functional test may be performed when desired, and a label may be printed with the DAP and patient information as described below.

1. Press the MENU button when DAP is displayed at the top center of the LCD screen at the main APR menu. This will present the DAP reset / test / print menu. • Press RESET to reset the DAP display to zero. The display will reset, and the

console will return to the main APR menu.

• Press TEST to test the DAP. The message “DAP TEST: …OK” will be presented if the DAP has passed its test. The console will return to the main APR menu.

• Press PRINT to print the DAP label. Printing…..will briefly be displayed at the

bottom of the screen while data is being sent to the printer. The printer will print the date and time near the top of the label, and the accumulated Dose-Area Product (mGycm2) near the middle of the label. This information is retrieved from the generator at the time the label is printed. Several headings are printed on the labels; the corresponding information may be written in the adjacent spaces using indelible ink:

NAME: Enter the patient’s name. M / F: Check (Male or Female). BORN: Date of Birth. Use the format shown (d / m / y). I.D. #: Enter the patient’s and / or operator’s identification number. TECH: Enter techniques as appropriate for the record.

• Press the MENU button to exit when finished.

NOTE: THE X-RAY GENERATOR MUST BE CONFIGURED FOR THE SPECIFIC PRINTER BY THE INSTALLER OR SERVICE ENGINEER (SEIKO INSTRUMENTS DPU-414 OR SLP-200 PRINTER). THE GENERATOR WILL BE COMPATIBLE ONLY WITH THE SELECTED PRINTER, THEREFORE ONLY THAT PRINTER MODEL MUST BE USED IN THIS INSTALLATION.

NOTE: THE PAPER OR LABELS USED IN THE PRINTER MUST MEET ALL APPLICABLE REGULATIONS. MEDICAL GRADE PAPER OR LABELS, APPROVED FOR MEDICAL RECORDS, MUST NORMALLY BE USED.

Refer to the printer manual for the printer operating instructions, paper feeding, printer troubleshooting, and related information. During the console start-up routine, after the generator has been switched on, the paper in the printer is automatically positioned by the RESET PRINTER function

Page 30: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 29

PROGRAMMING THE CONSOLE 4 MAKING CHANGES TO PRE-PROGRAMMED APR TECHNIQUES The default techniques associated with the anatomical views in the APR display window may be edited if desired. To do this, follow the steps in this chapter. ENTERING INTO PROGRAMMING MODE 1. Start with the generator switched OFF.

2. Press and hold the MENU button while pressing the power ON or button on the operator console.

3. When the console finishes its power-on sequence, and the ENTER PASSWORD prompt is displayed, press the MENU button 4 times to access the following menu:

* GENERATOR SETUP * UTILITY APR EDITOR: DISABLED EXIT SETUP

Access is now provided to the following groups of functions. These will be discussed in sequence. UTILITY • Set Time & Date.

• Error Log. • Statistics. • Console.

APR EDITOR • Enables/disables changes to the APR techniques.

EXIT SETUP • Returns to normal operation i.e. non programming mode.

Page 31: Pixel cp operators manual

4 Programming the Console

Page 30 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

UTILITY MENU To select the UTILITY menu, press the selection button adjacent to UTILITY in the APR display window. This allows access to the following functions:

* UTILITY * SET TIME & DATE

ERROR LOG

STATISTICS

CONSOLE

EXIT SET TIME & DATE • Allows setting of the time and date. ERROR LOG • Allows the error log to be reviewed. STATISTICS • Allows review of the tube exposure

counter, and the generator accumulated exposure counter.

CONSOLE • Allows console parameters to be set. EXIT • Returns to the GENERATOR SETUP

menu (previous page). Set Time and Date

* SET TIME & DATE * YEAR: 2002 HOUR: 1 MONTH: 1 MIN: 35 DAY: 1 + - EXIT

Page 32: Pixel cp operators manual

Programming the Console 4

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 31

Use these steps to set the time and date. 1. From the UTILITY menu, select SET TIME & DATE. 2. Select the parameter to change by pressing the associated button. Press the buttons next

to the + or - symbols on the LCD display to set the desired value. Time is set in 24 hour format.

3. Select EXIT to return to the UTILITY menu. Error Log

* ERROR LOG * ERROR # 10 OF 100 DATE: 09-12-2001 TIME: 13:09ERROR CODE: 30 +ERROR MESSAGE: ROTOR FAULT -

kVp: 80 mA: 12.5 ms: 500 EXIT

Use these steps to view the error log.

1. From the UTILITY menu select ERROR LOG. 2. Select ERROR # and press the + or - buttons to scroll through the error log. The error

code, error message, date and time of the error, and exposure factors will be displayed in the APR display window, while control settings will be indicated via the appropriate indicator LEDs on the console.

3. Select EXIT to return to the UTILITY menu.

Page 33: Pixel cp operators manual

4 Programming the Console

Page 32 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

Statistics

* STATISTICS * TUBE 1 EXP: 0 TOTAL EXP: 0 EXIT

The statistics menu shows the X-ray tube exposure count, and the accumulated generator exposure count.

1. From the UTILITY menu select STATISTICS. 2. The statistics (the number of exposures on the X-ray tube, and the total generator

exposures) may now be viewed.

3. Select EXIT to return to the UTILITY menu.

Page 34: Pixel cp operators manual

Programming the Console 4

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 33

Console

* CONSOLE * SLOW KEY REPEAT: 200MS LCD SCREENMED. KEY REPEAT: 150MS APR MODE: NOFAST KEY REPEAT: 75MS +SPEAKER VOLUME: 15 - EXIT

The CONSOLE menu displays console parameters that may be changed to suit operator preferences.

1. From the UTILITY menu select CONSOLE. 2. Select SLOW KEY REPEAT. Press the buttons next to the + or - symbols on the LCD

display to set the speed at which the displays change for the first 5 counts while a button is pressed.

3. Select MED KEY REPEAT. Press the + or - button to select the speed at which displays change while the button is pressed for the next 5 counts.

4. Select FAST KEY REPEAT. Press the + or - button to select the speed at which displays change while the button is pressed after 10 counts.

5. Select SPEAKER VOLUME. Press the + or - button to set the speaker volume in the range 1 to 15.

6. Select LCD SCREEN. Press button to toggle between normal and reverse display for the LCD display screen.

7. Select APR MODE. Press the selection button to toggle between NO and YES. Selecting YES disallows changes to the AEC / mAs / mA/ms selections when in normal operating mode. If manual technique changes are disabled via this utility, AEC / mAs / mA/ms changes can only be made by selecting an APR technique that has been programmed to the desired technique. APR MODE must be set to NO to allow switching between AEC / mAs / mA/ms in normal operating mode.

8. Select EXIT to return to the UTILITY menu.

9. Select EXIT again to return to the GENERATOR SETUP menu.

Page 35: Pixel cp operators manual

4 Programming the Console

Page 34 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

APR EDITOR

* GENERATOR SETUP * UTILITY APR EDITOR: DISABLED EXIT SETUP

Use these steps to edit the programmed APR techniques.

NOTE: BEFORE MAKING CHANGES TO THE DEFAULT APR TECHNIQUES, PLEASE RECORD ALL CURRENT DEFAULTS. THESE DEFAULTS MAY BE RECORDED IN COPIES OF THE APR PROGRAMMING WORKSHEET AT THE END OF THIS CHAPTER. 1. From the GENERATOR SETUP menu select APR EDITOR.

2. Press the APR EDITOR button to toggle between ENABLED and DISABLED. Set the APR EDITOR to ENABLED.

For reference, DISABLED does not allow editing and then saving the changed APR technique(s). ENABLED allows the operator to change the default APR techniques, and then save the changes. The APR will then default to the newly saved techniques.

The generator stores the last APR editor setting (ENABLED and DISABLED) prior to being switched OFF. If the APR editor was previously ENABLED, APR changes may subsequently be made and then saved in normal operating mode without the need to manually set the APR editor to ENABLED. To disable this feature, the APR editor must be set to DISABLED.

3. Return to normal operation by pressing EXIT SETUP.

4. The following menu is displayed. This is a typical menu, actual anatomical views may have been reprogrammed, and the screen may differ slightly depending on generator model.

Page 36: Pixel cp operators manual

Programming the Console 4

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 35

TUBE1 0%HU 1:00

*APR TABLES* SKULL UPPER EXTREMITY THORAX LOWER EXTREMITY ABDOMEN G.I. PELVIS SPINE 2 mAs

kVp: 70 mA: 100 ms: 20

5. Perform the APR technique change as outlined below. 6. Select the desired APR menu to be edited; for example SKULL. 7. Select the desired APR view to edit, for example SKULL AP/PA, table Bucky image

receptor. 8. Select the patient size to edit within that view, for example medium. All techniques with the exception of kV will be common to all patient sizes. The

common techniques for all patient sizes within that view should be programmed first, then the desired kV for the selected patient size. After these techniques are saved per the remainder of this section, kV values may be programmed if desired for the other patient sizes in the same APR view.

9. The default selections are shown on the console, for example: 70 kV, 100 mA, 20 ms, technique = mA/ms, large focus. 10. Change the technique to the desired value(s). Parameters that may be edited are: kV,

mA/mAs, time (ms), technique select (AEC / mAs / mA/ms), focus, image receptor, film screen, plus density and field selection in AEC mode.

The APR view being edited will be highlighted with a rectangular outline as soon as any parameter is changed within that view. Also the word SAVE will appear in the display window below the MENU button.

11. Continue making all desired APR changes within that view.

12. When finished, press the MENU button. This is the first step necessary to save the changes. This also allows the option of canceling the changes in the next step.

13. If you do not wish to save the changes, select EXIT (BACK button) to abort the SAVE function. To save the changes, select SAVE (FORWARD button). The message SAVING DATA... will briefly appear.

14. Press the MENU button to return to the main APR screen.

Page 37: Pixel cp operators manual

4 Programming the Console

Page 36 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

15. You may now make changes to other APR techniques if desired by repeating steps 6 to 14.

16. It is suggested that the APR editor be set to DISABLED as described earlier in this section, in order to avoid unintended APR changes in the future.

PROGRAMMING CM THICKNESS 1. Follow steps 1 to 3 of the previous section (APR EDITOR). 2. Select the desired APR to edit. 3. Select a CM thickness (mAs) or CM thickness (AEC) mode. Select the desired image

receptor, film screen, focal spot, and AEC fields (if applicable) for the selected anatomical view.

4. The AUX 1 button is used for CM thickness programming. In programming mode (with the APR EDITOR set to ENABLED), pressing AUX 1 scrolls through CM START → CM STEP → CM as required to program these parameters.

CM START The minimum CM for the selected technique (CM START range is 1 to 14). To program, refer to step 5.

CM STEP The step size used to establish the CM vs kVp and mAs curve. The CM STEP range is 1 to 5. To program, refer to step 5.

CM The steps on the CM vs kV and mAs curve. To program, refer to step 5.

5. (a) Determine the minimum and maximum thickness in cm for the selected anatomical

view. Thickness outside of this range cannot be selected later in normal operating mode, therefore these dimensions should be chosen carefully.

(b) Determine the cm step size by the following formula: CM Step Size = 8

Maximum thickness (cm) - Minimum thickness (cm)

The calculated step size must be rounded up to the nearest integer, i.e. 1.66 rounds up to a step size of 2.

(c) Select CM START, then enter the minimum thickness determined in step (a).

(d) Select CM STEP, then enter the cm step size calculated in step (b), maximum step size = 5.

(e) Select CM. Press the CM - and + buttons to confirm that the minimum and maximum thickness limits are acceptable. Repeat the above steps if necessary to optimize the minimum and maximum thickness limits. Set CM to minimum thickness before attempting to toggle back to CM START, or else an error message GOTO 1st CM will be presented.

Page 38: Pixel cp operators manual

Programming the Console 4

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 37

(f) Press CM - to select the minimum thickness. Set the kV and mAs or kV and backup mAs as appropriate for the selected thickness. (In CM thickness (mAs) mode, kV and mAs may be programmed; in CM thickness (AEC) mode, kV and backup mAs may be programmed).

Refer to Programming the mAs / backup mAs, below for the recommended procedure to set this parameter.

(g) Press CM + to select the next thickness step. Set the kVp and mAs / backup mAs for that thickness as per 5 (f). Nine (9) thickness steps are available, with the step size as previously set. The desired kV and mAs / backup mAs may be set for each of those 9 thickness steps. In the normal operating mode, the thickness steps will be 1 cm, regardless of the step size set in APR edit mode. The console will interpolate between the step sizes set up in APR edit mode to determine the correct kV and mAs / backup mAs for each 1 cm thickness step in normal operating mode.

(h) Repeat 5 (g) for the remaining CM steps.

6. When finished editing the APR, press the MENU button. This is the first step necessary to save the changes. This also allows the option of canceling the changes in the next step.

7. If you do not wish to save the changes, select EXIT (BACK button) to abort the SAVE function. To save the changes, select SAVE (FORWARD button). The message SAVING DATA... will briefly appear.

8. Repeat this procedure to edit other anatomical views.

9. Set the APR EDITOR to DISABLED to return to normal operating mode and prevent inadvertent programming changes.

Page 39: Pixel cp operators manual

4 Programming the Console

Page 38 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

Programming the mAs / backup mAs 1. When programming the mAs (for CM-thickness, mAs mode), or the backup mAs (for CM-

thickness, AEC mode), use the proper mAs (that will produce the correct non-AEC exposure) for each CM thickness step. This will ensure proper CM-thickness programming:

• In CM-thickness (mAs mode), the correct mAs will be assigned to each CM thickness step.

• In CM-thickness (AEC mode), the backup mAs will be nominally two times the programmed mAs, i.e. if 11 mAs is programmed for a certain CM thickness, the backup mAs will be nominally 22 mAs. This “doubling” of the programmed mAs to arrive at the backup mAs is done automatically by the generator when in CM-thickness, AEC mode.

2. The mAs values that are entered during CM-thickness programming are saved in memory after performing the SAVE function as described above.

• If you are in CM-thickness (AEC mode), and the CM - or + buttons are pressed to review the programmed mAs values, the mAs display will update to show the “two times” backup mAs.

• If you are in CM-thickness (AEC mode), and then select CM-thickness (mAs mode) via the technique selector button, the “two times” backup mAs value will be displayed for that CM thickness. Press the CM - or + buttons to update the display and use the normal mAs, not the doubled backup mAs that was carried forward.

• If you are in CM-thickness (mAs mode), and then select CM-thickness (AEC mode) via the technique selector button, the normal mAs value will be displayed for that CM thickness. Press the CM - or + buttons to update the display and use the “two times” backup mAs, not the normal mAs that was carried forward.

MAKING CHANGES TO THE APR TEXT The anatomical sections and views as displayed in the APR window may be altered to suit your requirements, but you will need to contact your service representative to have this done. Use copies of the worksheet on the next page to record your desired changes. One sheet should be used per APR view; the master copy should be photocopied as many times as necessary to record the required changes.

Page 40: Pixel cp operators manual

Programming the Console 4

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 39

APR PROGRAMMING WORKSHEET (THIS IS A MASTER, MAKE WORKING COPIES AS NEEDED)

kV Table Bucky Receptor

Tech Select

Patient Size Small

Focus

FieldSelect

Non Bucky Receptor

mAs

Medium

Wall Bucky Receptor

Aux 1

Aux 2

mA/ms

Large

Film Screen

mA/mAs

ms

DENSITY

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

__________________

MENU 1APR TABLESMAX 16 CHARACTERS

MENU 2ANATOMICAL VIEWMAX 16 CHARACTERS

AEC Small

Large

CMP200_005B.CDR

Page 41: Pixel cp operators manual

4 Programming the Console

Page 40 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 42: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 41

ERROR CODES AND MESSAGES 5 The generator console will display status messages on the APR display during normal and abnormal operation of the generator. This chapter contains tables of those messages and suggests actions to be taken should any malfunctions occur. Messages may be cleared by pressing the MENU button. Operator Messages

These messages indicate the status of the generator. No action is required.

MESSAGE DESCRIPTION INITIALIZATION Displayed during power up sequence. SPINNING ROTOR Displayed when prep state is active. X-RAY READY Displayed when generator is ready to expose X-RAY ON Displayed during an X-ray exposure. DAP NOT READY The optional DAP is in its “warm up” state, and not ready to make DAP

measurements. Limit Messages

These messages indicate that an exposure has been requested that exceeds one or more limits.

MESSAGE PROBLEM ACTION AEC DENSITY LIM. Requested density not programmed. Select another

density or request those steps be programmed. Consult your service representative for this.

ANODE HEAT WARN. Anode has exceeded programmed warning level.

Wait for anode to cool.

CAL LIMIT Requested parameter not calibrated. Consult your service representative.

GEN DUTY WARNING The X-ray generator has reached its duty warning limit.

Re-evaluate technique factors. Further exposures will be inhibited. The generator must be allowed to cool sufficiently.

Page 43: Pixel cp operators manual

5 Error Codes and Messages

Page 42 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

MESSAGE PROBLEM ACTION

GEN. JOULE LIMIT The requested exposure will exceed the generators kilo joule heat limit.

Allow the generator to cool sufficiently to allow the requested exposure.

GEN. KV LIMIT Requested kV not allowed as generator kV limit has been reached.

None.

GEN. KW LIMIT Requested parameter not allowed as generator kW limit has been reached.

None.

GEN. MA LIMIT Requested mA not allowed as generator mA limit has been reached.

None.

GEN. MAS LIMIT Requested mAs not allowed as generator mAs limit has been reached.

None.

GEN. MS LIMIT 1. Requested ms not allowed as generator ms limit has been reached.

2. Maximum ms reached at requested power. Reduce mA or kV.

None.

GOTO 1st CM Displayed if attempting to exit CM if the CM setting is not at minimum.

Set CM to minimum value before continuing.

HOUSE HEAT WARN. Tube housing has exceeded housing warning level.

Wait for housing to cool.

INVALID PARAM. Generator detected invalid parameter within received message, message ignored.

Select valid parameter.

LABEL JAMMED The labels are jammed. Clear the paper jam. OUT OF LABELS The printer is out of labels. Load more labels. PLATEN OPEN The paper platen is not positioned

properly. Check the platen position.

PRINTER COMM ERROR The printer has reported a communication error.

Try printing again.

PRINTER ERROR The printer self-diagnostics have reported a printer problem.

Refer to the printer manual.

PRINTER OFF-LINE The printer is off-line. Put the printer on-line.

TUBE KV LIMIT Requested kV not allowed as tube kV limit has been reached.

None.

TUBE KW LIMIT Requested parameter not allowed as tube kW limit has been reached.

None.

Page 44: Pixel cp operators manual

Error Codes and Messages 5

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 43

TUBE MA LIMIT Requested mA not allowed as tube mA

limit has been reached. None.

TUBE MAS LIMIT Requested mAs not allowed as tube mAs limit has been reached.

None.

These messages indicate that an error has occurred.

ERROR CODE MESSAGE PROBLEM ACTION E001 GEN EPROM ERR. Consult your service

representative. E002 GEN EEPROM ERR. Consult your service

representative. E003 GEN NVRAM ERR. Consult your service

representative. E004 GEN RTC ERROR Generator CPU real time

clock is not functioning. Reset time and date. Consult your service representative.

E005 PS CONTACT ERR. Consult your service representative.

E006 ROTOR FAULT Power unit off and retry rotor start. Consult your service representative.

E007 FILAMENT FAULT Consult your service representative.

E008 KV/MA FAULT Consult your service representative.

E009 PS NOT READY Power supply is not ready to make an exposure.

Retry exposure. Consult your service representative.

E010 LOW KV FAULT Consult your service representative.

E011 HIGH MA FAULT Consult your service representative.

E012 LOW MA FAULT Consult your service representative.

E013 MANUAL TERMIN. Operator released exposure switch during exposure.

Re-take exposure if necessary. Consult your service representative.

Page 45: Pixel cp operators manual

5 Error Codes and Messages

Page 44 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

ERROR CODE MESSAGE PROBLEM ACTION

E014 AEC BUT ERROR AEC exposure exceeded allowed back up time.

Check exposure technique settings. Consult your service representative.

E015 AEC BU MAS ERR. AEC exposure exceeded allowed back up mAs.

Check exposure technique settings. Consult your service representative.

E016 TOMO BUT ERROR Tomo exposure exceeded back-up time.

Check exposure technique settings.

E017 NOT CALIBRATED Consult your service representative.

E018 PREP TIMEOUT Generator has been in prep state too long.

Reduce length of time in prep state.

E019 ANODE HEAT LIMIT Selected parameters will cause X-ray tube to exceed its programmed anode heat limit.

Reduce parameters or wait for tube to cool.

E020 THERMAL INT #1 X-ray tube # 1 too hot and its thermal switch has opened.

Wait for X-ray tube # 1 to cool.

E021 THERMAL INT #2 X-ray tube # 2 too hot and its thermal switch has opened.

Wait for X-ray tube # 2 to cool.

E022 DOOR INTERLOCK Door is open. Close door. E023 COLLIMATOR ERR. Collimator is not ready. Check collimator. E024 CASSETTE ERROR Cassette is not ready. Check cassette. E026 SPARE INT. Spare input is not ready. Consult your service

representative. E028 PREP SW CLOSED Consult your service

representative. E029 XRAY SW CLOSED Consult your service

representative. E032 CONSOLE COMM ERR Generator has detected

error in communication to console.

Turn power off and then on to reset generator. Consult your service representative.

E033 GEN BATTERY LOW Consult your service representative.

E034 +12VDC ERROR Consult your service representative.

E035 -12VDC ERROR Consult your service representative.

Page 46: Pixel cp operators manual

Error Codes and Messages 5

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 45

ERROR CODE MESSAGE PROBLEM ACTION

E036 +15VDC ERROR Consult your service representative.

E037 -15VDC ERROR Consult your service representative.

E038 CAL DATA ERROR Consult your service representative.

E039 AEC DATA ERROR Consult your service representative.

E041 REC DATA ERROR Consult your service representative.

E042 TUBE DATA ERR. Consult your service representative..

E043 KV ERROR Switch OFF generator. Prevent further use of generator, then. consult your service representative.

E044 COMM ERROR Received communication message not valid and ignored.

Reset error.

E045 NOT SUPPORTED Received message is valid, but not supported by this system.

Reset error.

E046 MODE INHIBITED Attempting to change CM thickness in AEC mode if the receptor is not programmed for mAs backup. Received message is valid, but not allowed during present state.

Consult your service representative. Reset error.

E048 FOCUS MISMATCH Consult your service representative.

E049 NOT ENABLED Requested function has not been programmed to be enabled.

Consult your service representative.

E050 GEN DATA ERROR Consult your service representative.

E051 AEC DEVICE ERR Generator has detected no feedback signal from AEC device.

Check that X-ray tube is pointing at correct AEC device. Consult your service representative.

E052 HIGH SF CURRENT Consult your service representative.

E053 HIGH LF CURRENT Consult your service representative.

Page 47: Pixel cp operators manual

5 Error Codes and Messages

Page 46 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

ERROR CODE MESSAGE PROBLEM ACTION

E054 AEC OUT OF RANGE Consult your service representative.

E055 NO FIELDS ACTIVE AEC enabled but no fields are selected.

Select AEC field(s).

E056 NO TUBE SELECTED Consult your service representative.

E057 AEC STOP ERROR Consult your service representative.

E058 CONSOLE BUT ERR. Consult your service representative.

E059 HOUSE HEAT LIMIT X-ray tube housing has exceeded limit.

Wait for tube to cool.

E060 EXP. KV HIGH Consult your service representative.

E061 EXP. KV LOW Consult your service representative.

E062 EXP_ SW ERROR Consult your service representative.

E063 FACTORY DEFAULTS Consult your service representative.

E065 TOMO DEVICE ERR. Consult your service representative.

E067 PS DUTY LIMIT Generator duty limit reached.

Allow the generator to cool before continuing.

E069 MAS OVERTIME ERR Consult your service representative.

E070 SOFTWARE KEY ERR Consult your service representative.

E071 DAP DOSE OVERFLW The accumulated DAP value exceeds the display limit.

Reset the DAP.

E072 DAP DEVICE ERR. The DAP device is not functional.

Consult your service representative.

E073 DAP DATA ERROR Consult your service representative.

E074 INVERTER 1 ERROR Consult your service representative.

E075 INVERTER 2 ERROR Consult your service representative.

E076 INVERTER 3 ERROR Consult your service representative.

E077 RES. CIRCUIT ERR Consult your service representative.

E078 BUCKY1 INTERLOCK Consult your service representative.

E079 BUCKY2 INTERLOCK Consult your service representative.

Page 48: Pixel cp operators manual

Error Codes and Messages 5

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 47

ERROR CODE MESSAGE PROBLEM ACTION

E080 INTERLOCK 1 OPEN Consult your service representative.

E081 INTERLOCK 2 OPEN Consult your service representative.

E082 KV OVER VOLTAGE Consult your service representative.

E083 ANODE MA FAULT Consult your service representative.

E084 CATHODE MA FAULT Consult your service representative.

E085 ROTOR MAIN FAULT Consult your service representative.

E086 ROTOR SHFT FAULT Consult your service representative.

E150 TABLE COMM ERR A communication error has occurred with the table.

Ensure that the table is powered on. Consult your service representative.

Page 49: Pixel cp operators manual

5 Error Codes and Messages

Page 48 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 50: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 49

GENERATOR EXPOSURE 6 TABLES

Table 1 following shows the nominal mAs resulting from preselected ms and mA values. This table also shows the range and interrelation of these loading factors. For example, if 20 mAs is selected at 200 mA, it can be seen that the exposure time will be approximately 100 ms. This is determined by reading down the 200 mA column to 20 mAs; then by reading the nominal exposure time 100 ms as shown at the left side of the table, along the 20 mAs row. An extra copy of this table is provided on pages 61 to 70 that may be photocopied as required, or removed from this manual and placed in a suitable location as per local requirements.

Page 51: Pixel cp operators manual

6 Generator Exposure Tables

Page 50 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 52: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 51

GENERATOR TECHNIQUE SELECTION (3 Point Selection: mA/ms & mAs)

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

1.0 1.1 1.2 1.4 1.6 0.10 1.8 0.10 0.11 2.0 0.10 0.11 0.125 2.2 0.10 0.11 0.125 0.14 2.5 0.10 0.11 0.125 0.14 0.16 2.8 0.10 0.11 0.125 0.14 0.16 0.18 3.2 0.10 0.11 0.125 0.14 0.16 0.18 0.20 3.6 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 4.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 4.5 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 5.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 5.6 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 6.3 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 7.1 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 8.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 9.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 10 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 11 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 12.5 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 14 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 16 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 18 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 20 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 22 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4

Table 1: mAs values vs. mA & time selected Table 1 continued on next page

Page 53: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 52

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

1.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 1.1 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 1.2 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 1.4 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 1.6 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 1.8 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 2.0 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 2.2 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 2.5 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 2.8 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 3.2 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 3.6 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 4.0 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 4.5 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 5.0 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 5.6 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 6.3 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 7.1 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 8.0 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 9.0 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 10 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 11 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 12.5 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 14 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 16 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 18 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 20 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 22 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 54: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 53

Time mA Selected (ms) 500 560 630 710 800 900 1000

1.0 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.2 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 5.0 5.6 6.3 7.1 8.0 9.0 10 11 5.6 6.3 7.1 8.0 9.0 10 11 12.5 6.3 7.1 8.0 9.0 10 11 12.5 14 7.1 8.0 9.0 10 11 12.5 14 16 8.0 9.0 10 11 12.5 14 16 18 9.0 10 11 12.5 14 16 18 20 10 11 12.5 14 16 18 20 22 11 12.5 14 16 18 20 22

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 55: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 54

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

25 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 28 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 32 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 36 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 40 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 45 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 50 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 56 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 63 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 71 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 80 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 90 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 100 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 110 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 125 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 140 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 160 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 180 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 200 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 220 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 250 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 280 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 320 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 360 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 400 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 450 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 500 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 56: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 55

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

25 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 28 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 32 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 36 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 40 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 45 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 50 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 56 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 63 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 71 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 80 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 90 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 100 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 110 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 125 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 140 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 160 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 180 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 200 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 220 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 250 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 280 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 320 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 360 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 400 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 450 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 500 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 57: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 56

Time mA Selected (ms) 500 560 630 710 800 900 1000

25 12.5 14 16 18 20 22 25 28 14 16 18 20 22 25 28 32 16 18 20 22 25 28 32 36 18 20 22 25 28 32 36 40 20 22 25 28 32 36 40 45 22 25 28 32 36 40 45 50 25 28 32 36 40 45 50 56 28 32 36 40 45 50 56 63 32 36 40 45 50 56 63 71 36 40 45 50 56 63 71 80 40 45 50 56 63 71 80 90 45 50 56 63 71 80 90 100 50 56 63 71 80 90 100 110 56 63 71 80 90 100 110 125 63 71 80 90 100 110 125 140 71 80 90 100 110 125 140 160 80 90 100 110 125 140 160 180 90 100 110 125 140 160 180 200 100 110 125 140 160 180 200 220 110 125 140 160 180 200 220 250 125 140 160 180 200 220 250 280 140 160 180 200 220 250 280 320 160 180 200 220 250 280 320 360 180 200 220 250 280 320 360 400 200 220 250 280 320 360 400 450 225 250 280 320 360 400 450 500 250 280 320 360 400 450 500

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 58: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 57

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

560 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 630 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 710 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 800 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 900 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 1000 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 1100 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 1250 12.5 14 16 18 20 22 25 28 32. 36 40 45 50 56 63 71 80 1400 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 1600 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 1800 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 2000 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 2200 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 2500 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 2800 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 3200 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 3600 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 4000 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 4500 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 5000 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 5600 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 6300 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 59: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 58

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

560 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 630 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 710 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 800 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 900 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 1000 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 1100 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 1250 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 1400 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 1600 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 1800 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 2000 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 2200 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 2500 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 2800 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 3200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 3600 250 280 320 360 400 450 500 560 630 710 800 900 1000 4000 280 320 360 400 450 500 560 630 710 800 900 1000 4500 320 360 400 450 500 560 630 710 800 900 1000 5000 360 400 450 500 560 630 710 800 900 1000 5600 400 450 500 560 630 710 800 900 1000 6300 450 500 560 630 710 800 900 1000

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 60: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 59

Time mA Selected (ms) 500 560 630 710 800 900 1000

560 280 320 360 400 450 500 560 630 320 360 400 450 500 560 630 710 360 400 450 500 560 630 710 800 400 450 500 560 630 710 800 900 450 500 560 630 710 800 900 1000 500 560 630 710 800 900 1000 1100 550 630 710 800 900 1000 1250 630 710 800 900 1000 1400 710 800 900 1000 1600 800 900 1000 1800 900 1000 2000 1000 2200 2500 2800 3200 3600 4000 4500 5000 5600 6300

Table 1 (Cont): mAs values vs. mA & time selected Continued on next page

Page 61: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 60

Note: PIXEL CP 450 32 kW Generators Maximum mA = 400 Maximum mAs = 500 PIXEL CP 550 40 kW Generators Maximum mA = 500 Maximum mAs = 500 PIXEL CP 650 50 kW Generators Maximum mA = 630 Maximum mAs = 630 PIXEL CP 850 65 kW Generators Maximum mA = 800 Maximum mAs = 800 PIXEL CP 1050 80 kW Generators Maximum mA = 1000 Maximum mAs = 1000

Note: 1) mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at higher kV selections.

2) The maximum power output of this family of generators is 32 kW, 40 kW, 50 kW, 65 kW or 80 kW respectively. The generator

software does not allow a kV and mA combination to be selected that exceeds these maximum values.

Page 62: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 61

GENERATOR TECHNIQUE SELECTION (3 Point Selection: mA/ms & mAs)

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

1.0 1.1 1.2 1.4 1.6 0.10 1.8 0.10 0.11 2.0 0.10 0.11 0.125 2.2 0.10 0.11 0.125 0.14 2.5 0.10 0.11 0.125 0.14 0.16 2.8 0.10 0.11 0.125 0.14 0.16 0.18 3.2 0.10 0.11 0.125 0.14 0.16 0.18 0.20 3.6 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 4.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 4.5 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 5.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 5.6 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 6.3 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 7.1 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 8.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 9.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 10 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 11 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 12.5 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 14 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 16 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 18 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 20 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 22 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4

Table 1: mAs values vs. mA & time selected Table 1 continued on next page

Page 63: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 62

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

1.0 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 1.1 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 1.2 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 1.4 0.10 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 1.6 0.11 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 1.8 0.125 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 2.0 0.14 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 2.2 0.16 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 2.5 0.18 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 2.8 0.20 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 3.2 0.22 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 3.6 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 4.0 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 4.5 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 5.0 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 5.6 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 6.3 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 7.1 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 8.0 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 9.0 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 10 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 11 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 12.5 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 14 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 16 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 18 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 20 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 22 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 64: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 63

Time mA Selected (ms) 500 560 630 710 800 900 1000

1.0 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.2 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 5.0 5.6 6.3 7.1 8.0 9.0 10 11 5.6 6.3 7.1 8.0 9.0 10 11 12.5 6.3 7.1 8.0 9.0 10 11 12.5 14 7.1 8.0 9.0 10 11 12.5 14 16 8.0 9.0 10 11 12.5 14 16 18 9.0 10 11 12.5 14 16 18 20 10 11 12.5 14 16 18 20 22 11 12.5 14 16 18 20 22

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 65: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 64

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

25 0.25 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 28 0.28 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 32 0.32 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 36 0.36 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 40 0.40 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 45 0.45 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 50 0.50 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 56 0.56 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 63 0.63 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 71 0.71 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 80 0.80 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 90 0.90 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 100 1.0 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 110 1.1 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 125 1.25 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 140 1.4 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 160 1.6 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 180 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 200 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 220 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 250 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 280 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 320 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 360 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 400 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 450 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 500 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 66: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 65

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

25 1.8 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 28 2.0 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 32 2.2 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 36 2.5 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 40 2.8 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 45 3.2 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 50 3.6 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 56 4.0 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 63 4.5 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 71 5.0 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 80 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 90 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 100 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 110 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 125 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 140 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 160 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 180 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 200 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 220 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 250 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 280 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 320 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 360 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 400 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 450 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 500 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 67: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 66

Time mA Selected (ms) 500 560 630 710 800 900 1000

25 12.5 14 16 18 20 22 25 28 14 16 18 20 22 25 28 32 16 18 20 22 25 28 32 36 18 20 22 25 28 32 36 40 20 22 25 28 32 36 40 45 22 25 28 32 36 40 45 50 25 28 32 36 40 45 50 56 28 32 36 40 45 50 56 63 32 36 40 45 50 56 63 71 36 40 45 50 56 63 71 80 40 45 50 56 63 71 80 90 45 50 56 63 71 80 90 100 50 56 63 71 80 90 100 110 56 63 71 80 90 100 110 125 63 71 80 90 100 110 125 140 71 80 90 100 110 125 140 160 80 90 100 110 125 140 160 180 90 100 110 125 140 160 180 200 100 110 125 140 160 180 200 220 110 125 140 160 180 200 220 250 125 140 160 180 200 220 250 280 140 160 180 200 220 250 280 320 160 180 200 220 250 280 320 360 180 200 220 250 280 320 360 400 200 220 250 280 320 360 400 450 225 250 280 320 360 400 450 500 250 280 320 360 400 450 500

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 68: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 67

Time mA Selected (ms) 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63

560 5.6 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 630 6.3 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 710 7.1 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 800 8.0 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 900 9.0 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 1000 10 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 1100 11 12.5 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 1250 12.5 14 16 18 20 22 25 28 32. 36 40 45 50 56 63 71 80 1400 14 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 1600 16 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 1800 18 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 2000 20 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 2200 22 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 2500 25 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 2800 28 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 3200 32 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 3600 36 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 4000 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 4500 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 5000 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 5600 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 6300 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 69: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 68

Time mA Selected (ms) 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450

560 40 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 630 45 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 710 50 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 800 56 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 900 63 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 1000 71 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 1100 80 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 1250 90 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 1400 100 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 1600 110 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 1800 125 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 2000 140 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 2200 160 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 2500 180 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 2800 200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 3200 220 250 280 320 360 400 450 500 560 630 710 800 900 1000 3600 250 280 320 360 400 450 500 560 630 710 800 900 1000 4000 280 320 360 400 450 500 560 630 710 800 900 1000 4500 320 360 400 450 500 560 630 710 800 900 1000 5000 360 400 450 500 560 630 710 800 900 1000 5600 400 450 500 560 630 710 800 900 1000 6300 450 500 560 630 710 800 900 1000

Table 1 (Cont): mAs values vs. mA & time selected Table 1 continued on next page

Page 70: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 69

Time mA Selected (ms) 500 560 630 710 800 900 1000

560 280 320 360 400 450 500 560 630 320 360 400 450 500 560 630 710 360 400 450 500 560 630 710 800 400 450 500 560 630 710 800 900 450 500 560 630 710 800 900 1000 500 560 630 710 800 900 1000 1100 550 630 710 800 900 1000 1250 630 710 800 900 1000 1400 710 800 900 1000 1600 800 900 1000 1800 900 1000 2000 1000 2200 2500 2800 3200 3600 4000 4500 5000 5600 6300

Table 1 (Cont): mAs values vs. mA & time selected Continued on next page

Page 71: Pixel cp operators manual

Generator Exposure Tables 6

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 70

Note: PIXEL CP 450 32 kW Generators Maximum mA = 400 Maximum mAs = 500 PIXEL CP 550 40 kW Generators Maximum mA = 500 Maximum mAs = 500 PIXEL CP 650 50 kW Generators Maximum mA = 630 Maximum mAs = 630 PIXEL CP 850 65 kW Generators Maximum mA = 800 Maximum mAs = 800 PIXEL CP 1050 80 kW Generators Maximum mA = 1000 Maximum mAs = 1000

Note: 1) mA/TIME values are tube rating dependent. For certain tubes, some mA/TIME selections are not available at higher kV selections.

2) The maximum power output of this family of generators is 32 kW, 40 kW, 50 kW, 65 kW or 80 kW respectively. The generator

software does not allow a kV and mA combination to be selected that exceeds these maximum values.

Page 72: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 71

MAINTENANCE SCHEDULE 7 This section is organized into two parts. The first part of this section details tests by which the operator can verify that the AEC circuits are functioning, and that there are no apparent defects in the console or associated equipment. These tests should be done monthly as a minimum, or more frequently if required by local regulations.

CAUTION: THE AEC VERIFICATION PROCEDURE REQUIRES THE PRODUCTION OF X-RAYS. OBSERVE CORRECT OPERATING PROCEDURES, AND TAKE APPROPRIATE PRECAUTIONS AGAINST X-RADIATION.

The second part of this section is the recommended maintenance schedule for your generator.

WARNING: REFER ALL SERVICING TO APPROPRIATE SERVICE PERSONNEL WHO ARE TRAINED TO SERVICE THIS EQUIPMENT AND WHO ARE FAMILIAR WITH THE POTENTIAL HAZARDS ASSOCIATED WITH THIS EQUIPMENT.

NOTE: MAINTENANCE SCHEDULE FREQUENCY MAY BE DETERMINED BY CERTAIN REGULATORY REQUIREMENTS OF THE COUNTRY OR STATE IN WHICH THE INSTALLATION IS LOCATED. ALWAYS CHECK THE LOCAL CODES AND REGULATIONS WHEN DETERMINING A MAINTENANCE SCHEDULE.

Page 73: Pixel cp operators manual

7 Maintenance Schedule

Page 72 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

AEC FUNCTIONAL CHECK The following procedure may be used to verify that the AEC circuits are functioning on generators equipped with AEC (automatic exposure control). 1. Switch the generator ON, and select an appropriate radiographic image receptor. 2. Align the X-ray tube and the selected image receptor such that the central ray is directly

over the center field of the AEC pickup device. Set the focal spot to film plane distance to 40 in. (1 m).

3. Select AEC mode of operation. Select center field, large focus.

MINIMUM EXPOSURE TIME: 4. With no object in the radiation field, adjust the collimator or beam limiting device to project

a 10 in. X 10 in. (24 cm X 24 cm) field at the image receptor. 5. Select 80 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not

operator selectable, the default AEC backup settings must be used. Refer to the note at the end of this page to determine the AEC backup mode that has been programmed.

6. Make an exposure and verify that the POST mAs reading is ≤ 2 mAs.

MAXIMUM EXPOSURE TIME: 7. Close the collimator or beam limiting device completely. Place a folded lead apron over

the image receptor. 8. Select 60 kVp, 100 mA, and a backup mAs of 50 if it is operator selectable. If this is not

operator selectable, the default backup settings must be used. Refer to the note at the end of this page to determine the AEC backup mode that has been programmed.

9. Make an exposure and verify that the error message AEC DEVICE ERR, AEC BUT ERROR or AEC BU MAS ERR is indicated in the APR display window after the exposure has terminated. This confirms that the exposure has continued until it was terminated by the AEC backup circuits in the generator.

NOTE: The AEC backup mode is installer programmable. Three modes are available, FIXED, MAS, and MS. Only the MAS mode allows the operator to set the backup mAs for an AEC exposure. FIXED: The generator will determine the maximum AEC backup time, not to exceed preset AEC backup mAs/ms values or system limits. The characters AEC will be displayed in the time window of the LED display during AEC operation. MAS: Allows the operator to adjust the AEC backup mAs, not to exceed preset AEC backup mAs/ms values or system limits. The mAs value will be displayed in the time window of the LED display during AEC operation. MS: Allows the operator to adjust the AEC backup ms, not to exceed preset AEC backup mAs/ms values or system limits. The ms value will be displayed in the time window of the LED display during AEC operation.

Page 74: Pixel cp operators manual

Maintenance Schedule 7

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 73

CONSOLE AND MISCELLANEOUS GENERATOR CHECKS The user must routinely check the X-ray equipment for apparent defects or damage. Report any apparent defects or irregular operation of any equipment to service personnel immediately, and discontinue use of the suspected faulty equipment until repairs are made. Continuing operation with faulty equipment may present various safety hazards, including risk of increased radiation. 1. Check all visual displays (warning and status lights, technique displays and indicators,

etc) for normal operation. 2. Check all audible indicators (speakers, beepers, etc) for normal operation, and check that

the loudness settings are adequate for the environment. 3. Check all interconnect cables and connectors for damage.

Page 75: Pixel cp operators manual

7 Maintenance Schedule

Page 74 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

MAINTENANCE ACTIVITIES IN THE TABLE BELOW ARE TO BE PERFORMED BY AUTHORIZED SERVICE PERSONNEL ONLY. HOWEVER THE OWNER OR OPERATOR OF THE EQUIPMENT MUST SCHEDULE THE SUGGESTED MAINTENANCE ACTIVITY WHEN REQUIRED BY CONTACTING THEIR LOCAL SERVICE AGENCY.

Maintenance Frequency

Description of Preventative Maintenance

Every 6 Months AND whenever a related certifiable X-ray component is replaced:

1. Clean and re-grease all HV connections using vapor proof compound.

2. Check that all HV connections are tight. 3. Clean the control console and main cabinet as needed.

REFER TO NEXT PAGE, CLEANING, BEFORE PROCEEDING.

4. Ensure that there are no obstructions blocking the ventilation holes or louvers on the generator cabinet.

5. Perform the X-ray tube auto calibration routine. 6. Verify the calibration of the generator. 7. Perform any additional tests required by laws governing

this installation. Every 12 months: 1. Examine the following for any visible damage and replace

any damaged components:

• The exterior of the control console, including the membrane switch assembly.

• The cable between the control console and the generator main cabinet.

• The hand switch (if used) and the cable connecting this to the console.

2. Open the generator cabinet and examine the unit for any

visible damage: missing or loose ground connections, oil leaks, damaged cables etc.

Every 3 years Replace the cooling fan in the generator main cabinet. Every 5 years: Replace the batteries on the console board in the control

console and the on the generator control board in the main cabinet.

Page 76: Pixel cp operators manual

Maintenance Schedule 7

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 75

CLEANING • To clean the console surfaces (without disinfecting) use soap and water. This should be

applied using a lint-free cloth as described below. • If console surfaces need to be disinfected, use Actichlor or equivalent with a Hypochlorite

concentration of 3 to 5%. ∗ Dilute the disinfectant 50/50 with water. ∗ Spray or dampen a lint-free cloth with the diluted disinfectant. The cloth must be damp

only, not wet. ∗ Gently wipe the console surfaces with the damp cloth.

• Never use any corrosive, solvent or abrasive detergents or polishes. • Ensure that no water or other liquid can enter any equipment. This precaution prevents

short circuits and corrosion forming on components. • Disinfection by spraying is not recommended because the disinfectant may enter the X-ray

equipment. • Methods of disinfection used must conform to legal regulations and guidelines regarding

disinfection and explosion protection. • If disinfectants which form explosive mixtures of gases are used in the room, these gases

must have dissipated before switching on the equipment again. • If room disinfection is done with an atomizer, it is recommended that the equipment be

switched OFF, allowed to cool down and covered with a plastic sheet. When the disinfectant mist has subsided, the plastic sheet may be removed and the equipment be disinfected by wiping.

Page 77: Pixel cp operators manual

7 Maintenance Schedule

Page 76 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 78: Pixel cp operators manual

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 77

X-RAY TUBE DATA 8

The following pages contain data for the X-ray tube(s) used with your X-ray generator at the time of installation.

Page 79: Pixel cp operators manual

8 X-Ray Tube Data

Page 78 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)

Page 80: Pixel cp operators manual

X-Ray Tube Data 8

Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N Page 79

NOTE TO INSTALLER:

INSERT TUBE DATA HERE.

Page 81: Pixel cp operators manual

8 X-Ray Tube Data

Page 80 Pixel CP series X-Ray Generator Operator’s Manual 740985 Rev. N

(This page intentionally left blank)