Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
RESEARCH ARTICLE
Picosecond CARS measurements of nitrogen rotational/translational and vibrational temperature in a nonequilibriumMach 5 flow
A. Montello • M. Nishihara • J. W. Rich •
I. V. Adamovich • W. R. Lempert
Received: 12 June 2012 / Revised: 20 November 2012 / Accepted: 7 December 2012
� Springer-Verlag Berlin Heidelberg 2012
Abstract Picosecond Unstable-resonator Spatially Enhanced
Detection Coherent Anti-Stokes Raman Scattering (USED-
CARS) is used for the measurement of nitrogen Q-branch
(DJ = 0) spectra in the subsonic plenum and supersonic
flow of a highly nonequilibrium Mach 5 wind tunnel.
Spectra are processed to infer rotational/translational (Trot)
and first-level vibrational (Tvib) temperatures in the
200–370 torr plenum simultaneously. Operation of the
nominally high reduced electric field (E/npeak * 500 Td),
nsec pulsed discharge alone results in fairly significant
vibrational loading, Tvib * 720 K/Trot * 380 K; addition
of an orthogonal low E/n (*10 Td) DC sustainer discharge
produces substantial vibrational loading, Tvib * 2,000 K/
Trot * 450 K. Effects of injection of CO2, NO, and H2
downstream of the pulser–sustainer discharge are examined,
which result in vibrational relaxation accompanied
by simultaneous gas heating, Tvib * 800–1,000 K/Trot *600 K. CARSk measurements within very low-density
flows in the Mach 5 expansion nozzle are also performed,
with Tvib measured in both the supersonic free-stream and
downstream of a bow shock created by a 5-mm-diameter
cylindrical test object in the Mach 5 flow. Measurements
within 300 lm of the cylinder leading edge show that for
pure N2, or N2 with 0.25 torr CO2 injection, no vibrational
relaxation is observed behind the bow shock.
1 Introduction
The ability to tailor nonequilibrium hypersonic flows by
control of the loading of internal degrees of freedom
(vibrational and electronic states), as well as by control of
dissociation and ionization fraction, is a topic of much
current interest. For example, turbulent transition delay in a
Mach 5 flow over a 5� cone by means of injection of carbon
dioxide into nitrogen or air flow has been recently dem-
onstrated (Leyva et al. 2009). Kinetic modeling calcula-
tions (Wagnild et al. 2010) suggest that transition delay is
caused by absorption of acoustic perturbations in the
boundary layer by vibrational energy modes of CO2, which
may also result in CO2 dissociation. Another example is
relaxation of energy stored in internal degrees of freedom
of molecules behind a bow shock, which may significantly
increase shock stand-off distance. Additionally, the pres-
ence of vibrationally and electronically excited species in a
hypersonic flow may strongly affect the emission signature
from the shock layer. Finally, short-pulse electric dis-
charges efficiently loading electronic energy levels of
nitrogen and oxygen in air are currently being explored as a
means of hypersonic flow control, by producing repetitive
localized pressure perturbations in the flow (Nishihara et al.
2011).
Since the cost of obtaining full-scale hypersonic flight
test data or operating large-scale ground test facilities is
extremely high, alternative methods of data production are
necessary. This provides a significant incentive for the
development and use of small-scale test facilities which are
capable of recreating environments seen in the flow con-
ditions of interest, and which lend themselves to the
development of optical diagnostics of nonequilibrium
hypersonic flows. Our approach to generating a nonequi-
librium hypersonic flow is to use a high-pressure, low-
A. Montello � M. Nishihara � J. W. Rich �I. V. Adamovich � W. R. Lempert (&)
Michael A. Chaszeyka Nonequilibrium Thermodynamics
Laboratories, Department of Mechanical and Aerospace
Engineering, The Ohio State University,
Columbus, OH 43210, USA
e-mail: [email protected]
123
Exp Fluids (2013) 54:1422
DOI 10.1007/s00348-012-1422-1
temperature, diffuse electric discharge sustained in the
plenum of a small-scale Mach 5 wind tunnel to load
internal energy modes of nitrogen and oxygen molecules
(Nishihara et al. 2012).
Similar to our previous work (Nishihara et al. 2012),
target parameters for nonequilibrium gas flow include run
time for steady-state conditions of 5–10 s, with plenum
pressures in the range of P0 * 200–370 torr. As our pre-
vious results have shown, the pulser–sustainer electric
discharge present in the wind tunnel plenum is capable of
producing highly nonequilibrium flow conditions, with
translational/rotational temperature inferred by nitrogen
UV emission spectroscopy near Trot * 350–400 K, and
vibrational temperature up to Tvib * 2,000 K (Montello
et al. 2012). Additionally, as these results demonstrated,
the flow internal energy mode nonequilibrium can be fur-
ther controlled by injection of ‘‘rapid relaxer’’ species
(carbon dioxide, nitric oxide, or hydrogen) into the sub-
sonic nonequilibrium flow, after the discharge section but
upstream of the nozzle throat. Qualitatively, the results of
the study demonstrated effects similar to what vibrational
energy transfer rates predict; however, these rates are
highly sensitive to the gas (rotational/translational) tem-
perature. To compare experimental vibrational energy
relaxation effects with energy exchange rate predictions,
simultaneous direct measurement of the gas temperature is
necessary.
This paper provides new experimental data character-
izing both the vibrational energy loading and gas heating of
molecular nitrogen in a nonequilibrium flow excited by a
non-self-sustained electric discharge in the plenum of a
Mach 5 wind tunnel. Volume ionization, at plenum pres-
sures in the range of P0 = 200–370 torr, is accomplished
using a high peak reduced electric field (E/n)peak *500 Td, where (E/n)peak is the peak reduced electric field
inferred from the estimated peak electric field, 5 nsec
duration pulse discharge, operating at a pulse repetition
rate of 100 kHz. An orthogonal DC sustainer discharge
(E/n * 10–30 Td), which accounts for approximately
80 % of the total power loading into the flow, efficiently
excites the N2 vibrational energy mode. Theoretical pre-
dictions suggest that as much as 80–90 % of the DC sus-
tainer input power goes into nitrogen vibrational excitation
by electron impact at these reduced electric field values.
Temporally and spatially resolved ground- and first-level
vibrational populations of nitrogen, N2 (X1Rg, v = 0–1),
yielding first-level vibrational temperature in the pulser–
sustainer discharge, up to Tvib (N2) = 2,000 K, are mea-
sured using picosecond Coherent Anti-Stokes Raman
Scattering (CARS) spectroscopy (Eckbreth 1996; Roy et al.
2010).
Measurements using the nsec pulsed discharge alone
(i.e., without the DC sustainer) are also presented, the
results of which demonstrate that significant vibrational
loading occurs, Tvib * 720 K at 300 torr of pure N2, even
at the relatively high nominal reduced electric field,
E/npeak * 500 Td. Our quasi-one-dimensional nanosecond
pulse discharge model (Adamovich et al. 2009) is used to
compare predicted versus measured energy coupling to the
flow, as well as post-breakdown energy loading to nitrogen
vibrational states, with good agreement to the values
measured by CARS. Modifications to the CARS system
allow for the capture of high-resolution Q-branch spectra,
with partial resolution of the rotational structure. These are
fit with the Sandia CARSFT least squares fitting code
(Palmer 1989), yielding best fit rotational temperatures of
Trot * 380 K for the pulser operating alone and
Trot * 450 K for the pulser–sustainer discharge (sustainer
voltage VPS = 4.5 kV).
Additionally, the effect on nitrogen Tvib and Trot of
injection of CO2, NO, and H2, between the discharge
section and the CARS measurement location, which results
in partial vibrational relaxation of nitrogen accompanied by
gas heating, is also examined. This demonstrates quanti-
tatively the ability to control and tailor the vibrational
energy content of the flow prior to its expansion through
the Mach 5 nozzle of the wind tunnel. Conversely, addition
of oxygen or nitrogen (not excited in the discharge) to the
vibrationally excited nitrogen flow did not result in sig-
nificant vibrational temperature change, due to slow
vibration–vibration energy transfer from N2 to O2 for the
case of oxygen addition and extremely slow vibration-
translation relaxation at these temperatures for either O2 or
N2 injection.
Finally, also presented are the results of measurement of
nitrogen Q-branch CARS spectra captured within the
extremely low-density Mach 5 supersonic flow. Measure-
ments are made within the free-stream as well as at mul-
tiple locations behind the bow shock produced by a 5-mm-
diameter quartz cylinder mounted across the flow. As will
be discussed, the utilization of micrometer adjustment
translation stages along with pitch/catch optical arms
allows for measurement location adjustment in the vertical
and streamwise directions within the flow. CARS spectra
are captured as near as 300 lm from the cylinder leading
edge, for the nonequilibrium flow produced by the pulser–
sustainer discharge, for no relaxer injection, as well as the
slightly equilibrated case of 0.25 torr injection of CO2. In
both cases, the flow remains highly nonequilibrium, with
vibrational temperatures found similar to those measured in
the subsonic plenum for the same discharge conditions.
The integrated spectra are also used to estimate local
number density, leading to an estimated shock stand-off
distance measured to 1.0 mm, very close to the 1.2 mm
measured previously by Schlieren imaging and NO PLIF
(Nishihara et al. 2012).
Page 2 of 13 Exp Fluids (2013) 54:1422
123
2 Experimental
2.1 Mach 5 wind tunnel
The experiments conducted for this study were performed
in the Mach 5 nonequilibrium flow tunnel previously
developed and described in detail (Nishihara et al. 2012).
Briefly, the laboratory scale wind tunnel, a schematic of
which is given in Fig. 1, operates at plenum pressures from
0.25 to 1 atm with nitrogen or air supplied from high-
pressure cylinders. The tunnel, constructed from transpar-
ent acrylic plastic, is capable of producing steady-state
nonequilibrium supersonic flow by utilizing a high-pres-
sure, diffuse, nanosecond pulser/DC sustainer electric dis-
charge operating in the plenum section. The discharge can
be tailored, to load internal (vibrational and electronic)
modes of nitrogen, while maintaining low translational–
rotational temperatures. Previous measurements by nitro-
gen UV emission spectroscopy in a similar discharge
arrangement yielded Trot * 350–400 K (Nishihara et al.
2012). Downstream of the discharge section, the gas flows
through a choked flow injector with 20 injector holes 1 mm
in diameter, arranged in a single row in both the top and
bottom channel walls, also shown in Fig. 1. Gases inducing
vibrational relaxation of nitrogen, such as carbon dioxide,
nitric oxide, hydrogen, can be injected into the flow, which
allows further control of the energy distribution among the
internal molecular modes in the flow. After the injector, the
flow traverses the nozzle throat, with a height of 1.6 mm,
and passes through the aerodynamically contoured Mach 5
expansion section to the nozzle exit. There is a diffuser
section just upstream of the 8-inch-diameter exit pipe
leading to a 110-ft3 vacuum tank and 200-cfm pump.
Throughout the length of the flow tunnel, several quartz
windows are flush mounted in the side and top/bottom
walls, providing ample optical access to the flow.
The pulser–sustainer electric discharge, a cross-sectional
schematic of which is given in Fig. 2, is comprised of two
fully overlapping discharges, (a) a repetitively pulsed,
high-voltage nanosecond discharge and (b) a transverse DC
sustainer discharge. As shown in Fig. 2, the pulser elec-
trodes (3 cm 9 4 cm copper plates) are flush mounted in
the top and bottom walls of the discharge section (nozzle
plenum), covered by alumina ceramic dielectric plates
1/16 inch thick, with a gap of 0.5 cm in between. The
sustainer electrodes (4.0 cm 9 0.5 cm copper plates) are
placed along the side walls of the discharge section and are
removable. In the present experiments, the sustainer elec-
trodes are separated by 3.0 cm. Both the pulser and sus-
tainer electrodes are rounded along the edges to reduce the
electric field nonuniformity. The acrylic plastic walls of the
channel downstream of the discharge section are covered
with alumina ceramic plates, attached with silicone rubber
adhesive. These protect the plastic walls from overheating
in the possibility of instability development or arcing in the
discharge. The main benefit from using the pulser–sus-
tainer discharge is the ability to generate stable nonequi-
librium plasmas at high pressures and high discharge
energy loading.
The repetitive nanosecond pulse discharge operates
using a high peak voltage (up to 30 kV), short-pulse dura-
tion (5 nsec), high-pulse repetition rate (up to 100 kHz)
voltage waveform, produced by a high-voltage nanosecond
pulse generator (FID GmbH 60-100MC4). Figure 3 shows
typical pulse voltage and current traces (top), as well as
instantaneous power and coupled pulse energy (i.e., tem-
porally integrated power) traces (bottom), measured in a
repetitively pulsed discharge in nitrogen at P0 = 300 torr
and m = 100 kHz, with the DC electrodes present in the test
section, but not powered (Yin et al. 2012). The waveforms
shown are for a pulse generated 0.1 s after the beginning of
the pulse burst (i.e., for pulse #10,000). The pulser is
operated using an external trigger/function generator. Each
high-voltage pulse generates volume ionization in the dis-
charge section, and the high voltage is then turned off
before any ionization/heating instability (Raizer 1991) has
time to develop. The high frequency of pulse repetition
prevents the plasma from fully decaying before the next
ionizing pulse occurs, thus providing pulse-periodic, spa-
tially uniform ionization in the discharge section (Nishihara
et al. 2005, 2006).
Between the ionizing pulses, energy is coupled to the
flow by the DC discharge. The DC electrodes are con-
nected to a Glassman 5 kV, 2 A power supply, operated in
Model
Injector(NO, H2,CO2, etc.)
Pulserelectrodes
DC electrodes
Main flow(N2)
Nozzle throat Nozzle exit
Diffuser
To vacuum
Optical access(CARS measurement locations)
8 cm 9 cm 10 cm
Fig. 1 Schematic side view of
Mach 5 tunnel
Exp Fluids (2013) 54:1422 Page 3 of 13
123
a voltage stabilized mode, in series with a 1.5 kX ballast
resistor. The DC voltage is deliberately kept below
breakdown threshold, that is, typically below 4–5 kV, to
preclude the development of self-sustained DC discharge
(i.e., independent of pulsed ionization) in the high pressure
flow, which would result in instability development and
arcing. The power coupled to the flow by the DC discharge
is significantly higher than the repetitively pulsed discharge
power. Previously, this approach has been used in our work
to sustain high-power discharges in a Mach 3 MHD wind
tunnel (Nishihara et al. 2005, 2006), and in an electrically
excited gas dynamic oxygen–iodine laser (Bruzzese et al.
2010). In the present experiments, the repetitively pulsed
discharge is operated for 0.6 s, and the DC discharge for
0.55 s. The DC power supply is triggered by an externally
generated rectangular-shaped trigger pulse, the rising edge
of which also triggers the function generator that produces
a burst of 60,000 trigger pulses for the high-voltage
nanosecond pulse generator at m = 100 kHz pulse repeti-
tion rate. The estimated reduced electric fields in the two
discharges are significantly different, (E/n)peak * 500 Td
in the nanosecond pulsed discharge and E/n * 10 Td in
the DC discharge (1 Td = 10-17 V cm2). At these condi-
tions, a significant fraction of input power in the pulsed
discharge is spent on electronic excitation, dissociation,
and ionization of nitrogen, while nearly all input power in
the DC discharge (up to *80–90 % according to theory
(Raizer 1991)) is stored in the vibrational energy mode of
nitrogen, with little power going to translational/rotational
modes, that is, to heat. Due to a very long N2 vibrational
self-relaxation time at near room temperature, *1 atm sec
(Gordiets et al. 1988), this approach can create essentially
vibrationally frozen nitrogen and air flows in the super-
sonic test section, with vibrational temperature greatly
exceeding the translational/rotational mode temperature.
During the experiment, both the main flow through the
discharge and the injection flows are controlled using
solenoid valves. The main flow rate is calculated using a
1-D choked flow equation, based on the plenum pressure
and the nozzle throat area. The injection flow rate has been
both measured directly using a mass flow controller and
calculated from the choked flow equation, with both
methods giving similar results. At the baseline conditions,
nitrogen at P0 = 200–370 torr, the mass flow rate through
the tunnel is 3.9–7.2 g/sec and the steady-state run time at
constant static pressure in the supersonic test section is
5–10 s; runs can be repeated every few minutes.
2.2 Picosecond CARS diagnostic system
Picosecond CARS spectroscopy has been used for the
direct measurement of the ro-vibrational distribution
functions of nitrogen. CARS is a four-wave mixing spec-
troscopic technique which has been used extensively for
thermometry and species concentration measurements of
combustion and other gas phase reacting and nonreacting
flows (Eckbreth 1996; Roy et al. 2010). Simply stated,
CARS involves the interaction of three input photons,
termed pump, Stokes, and probe, with a molecule, resulting
in the generation of a signal photon. If the pump/Stokes
photons have a frequency difference corresponding to an
internal degree of freedom (generally rotational or vibra-
tional), a strong coherent (laser-like), ‘‘resonant’’ signal is
produced in a direction determined by what is known as the
‘‘phase-matching’’ criteria. All data presented in this work
used vibrational Q-branch (DJ = 0) transitions.
The CARS signal intensity, ICARS, has a linear depen-
dence on the Stokes laser intensity, IStokes, while scaling
quadratically with the magnitude of the CARS suscepti-
bility (proportional to number density, vCARSj j / N), that
is, ICARS / IStokes � N2. These effects are accounted for in
the data processing. Normalizing the CARS intensity by a
nonresonant background (NRB) spectrum captured in
argon accounts for the variation in the intensity profile of
the Stokes laser. It was experimentally determined that the
average NRB profile was very stable over the course of a
⊗Flow into the page
Pulsed electrodes
Sustainer DC electrodes (removable)
Alumina ceramic plates
0.5 cm3.0 cm
Fig. 2 Schematic diagram of pulser–sustainer discharge section
0 50 100 150 200 250 300 350
-10
0
10
20
Vo
ltag
e [k
V]
-40
0
40
80
Cu
rren
t [A
]
0 50 100 150 200 250 300 350
-0.50
0.51
1.5
Time [nsec]
Po
wer
[M
W]
01.534.56
En
erg
y [m
J]
Fig. 3 Nsec pulse voltage, current (upper), power, and coupled pulse
energy (lower) measurements. Nitrogen, P0 = 300 torr, m = 100 kHz
Page 4 of 13 Exp Fluids (2013) 54:1422
123
day, so one 200-shot average NRB spectrum was used to
normalize the data captured each day. After the data nor-
malization, the square root of the intensity is taken to
account for the quadratic dependence on number density;
all plots of spectral data below, are of the square root of
normalized intensity. For measurement of Tv, the CARS
cross section must also be accounted for, which for low
vibrational levels of nitrogen scales approximately as
(v’’ ? 1)2 (Lempert 2005) Since the square root of the data
has already been taken, dividing the integral of each peak
by (v’’ ? 1) yields the relative vibrational population in
each level.
For this work, the Unstable-resonator Spatially Enhanced
Detection (USED) phase-matching geometry (Marko and
Rimai 1979; Klick et al. 1981) has been employed. For this
geometry, the single 532 nm pump/probe beam is enlarged
with a telescope and the center portion of the beam is
removed, creating an annulus. This beam is then combined
coaxially with the Stokes beam by a dichroic mirror,
similar to a collinear alignment except that there is no
spatial overlap because the Stokes beam occupies the void
created within the pump/probe beam. After the beams pass
the focusing lens, the two regions of the annular pattern
remain spatially separated until they arrive very near to the
focal region; it is within this small overlap volume that
CARS signal generation occurs. According to the phase-
matching criteria, the generated CARS signal appears as a
ring outside of the pump/probe beam. This phase-matching
scheme is primarily chosen due to concerns about potential
beam steering that would occur for measurements per-
formed in the vicinity of the characteristic bow shock
produced by test objects in the Mach 5 section of the flow,
as well as a desire to perform measurements as close as
possible to the test object surface. There is a small sacrifice
in spatial resolution, relative to the more common folded
BoxCARS geometry, but this is not significant for the
measurements reported here. The transverse spatial
resolution for our system is on the order of 50 lm; a
measurement of the longitudinal resolution, performed by
scanning a glass flat in the vicinity of the CARS
measurement volume and observing the nonresonant
background signal, indicates that [95 % of the signal
generated comes within a 3 mm length at the beam focus,
as compared to *1 mm or less for BoxCARS, and 1–2 cm
for collinear CARS (determined in the same manner).
The broadband Stokes beam is generated with a home-
built dye laser, patterned after Roy et al. (2005), which
employs side-pumped oscillator and pre-amplifier cells,
followed by an end-pumped final amplifier cell (note there
is no output coupler). The combination of a half-wave plate
and thin film polarizer allows for adjustment of the ratio
between energy pumping the dye versus the energy in the
CARS pump/probe beam. Dye laser energy efficiency is as
high as 10 %. For this work, mixtures of Rhodamine 640
(R640) and Kiton Red 620 (KR620) were used. The
oscillator/pre-amplifier cells used a mixture of 34 mg R640
and 18 mg KR620 in 750 mL of methanol, while the
amplifier cell solution was 20 mg R640 and 18 mg KR620
in 750 mL methanol. The dye laser output is centered near
604 nm, with a full width at half maximum (FWHM) of
approximately 5–6 nm (Montello et al. 2012).
The dye laser is pumped by an Ekspla SL-333 Nd:YAG
laser, as can be seen in Fig. 4, which shows a schematic of
the CARS experimental arrangement. The nearly transform
limited YAG outputs pulses of approximately 150 pico-
seconds in duration, with variable energy output of up to
120 mJ/pulse at 532 nm. The choice to use a picosecond
(psec) system is motivated by several reasons, including
enabling time-resolved measurements at nsec and sub-nsec
time scales, as well as lowering necessary pulse energies,
reducing the risk of window damage. After the dichroic
mirror, which creates the 532/607 nm annulus, the beams
are focused into the test section with a 250-mm focal length
lens. After the focal point, located in the center of the wind
tunnel, the beams are re-collimated using a 100-mm focal
length lens.
After the collimating lens, a series of long-wavelength
passing dichroic mirrors reflect the *473 nm CARS beam
while dumping the pump/probe and Stokes beams. Finally,
the CARS signal passes through a short-pass filter, with a
cutoff wavelength of 503 nm, and is focused by a 100-mm
lens onto the slit of a 0.75 m And or Shamrock 750 spec-
trometer. At the exit plane of the spectrometer, a relay lens
magnification system, comprised of two Nikon F-mount
lenses, a 35-mm lens attached to the spectrometer and an
80- to 200-mm telephoto lens attached to the camera, pro-
vides variable spectral resolution magnification. For the
0.75 m Spectrometer
Broadband Dye Laser
Short-Pass Filter100mm Lens
Beam Dump
Dichroic Mirrors
Nd:YAG
Camera
Relay Lens Magnification
System
Flo
w
Wind Tunnel
Fig. 4 Experimental schematic diagram
Exp Fluids (2013) 54:1422 Page 5 of 13
123
work presented, the lens was set to give a *2.39 magni-
fication, which resulted in a spectral resolution of
*0.4 cm-1 when used with a 3,600 lpmm grating. This is
sufficient to partially resolve the rotational structure in room
temperature Q-branch spectra of nitrogen, as will be dem-
onstrated in the next section. The camera used is an Andor
Newton EM-CCD; the 1,600 by 400 pixel sensor array is
cooled to -90 �C, with the EM gain set to 150. The camera
and spectrometer are interfaced to a laboratory computer for
data recording. The entire picosecond CARS system is
placed on a custom-built cart, allowing the entire setup to be
easily transported between experimental facilities.
2.3 Mach 5 flow measurement setup
In the supersonic nozzle exit, as previously described, a
5-mm-diameter quartz cylinder model is mounted across
the hypersonic flow, which creates a bow shock, as
depicted in the schematic in Fig. 5a. The position of the
model relative to the entire tunnel can be seen in Fig. 1
above. For this work, a 7.5-mm-long section of 5-mm-
diameter quartz tube was secured to the leading edge of a
2-mm-diameter stainless steel rod, with the rod mounted to
the tunnel walls, as shown in Fig. 5b. The use of the short
rod section allows for closer access by the laser beams,
which propagate in the same direction as the rod axis, than
a full-length 5-mm-diameter rod would allow. Previous
Schlieren imaging (Nishihara et al. 2012) has shown that
the bow shock stand-off distance for this arrangement,
*1.2 mm for a plenum pressure of 300 torr, is uniform in
the spanwise direction over a distance of *6 mm, nearly
double the longitudinal spatial resolution of the system,
which as previously mentioned is *3 mm.
An experimental arrangement using optical pitch/catch
arms, along with two manual translation stages, allows for
simple adjustment of the CARS measurement location. A
3 inch by 18 inch optical breadboard section, which holds
the necessary turning prisms (for the vertical pitch/catch)
and CARS focusing and collimating lenses, is attached to
the vertical adjustment stage by a right-angle adapter plate.
The base of this stage is mounted to a 6 inch by 18 inch
breadboard, which also holds a second set of turning
prisms, comprising the horizontal pitch/catch. This larger
breadboard is mounted onto a horizontally positioned
translation stage, providing measurement location adjust-
ment down the length of the tunnel. Both stages are fitted
with fine adjustment micrometers, providing spatial reso-
lution of 10 lm per division in each dimension.
3 Results and discussions
3.1 Pure nitrogen flow, no discharge
This work focuses on both translational/rotational temper-
ature, Trot, as well as the ‘‘first-level’’ vibrational temper-
ature, Tvib, defined as:
Tvib ¼hvib
ln½n0=n1�ð1Þ
where hvib = 3,353 K is the energy difference between
vibrational levels v = 0 and v = 1 in temperature units. A
sample single-shot spectrum of 370 torr pure N2 flowing
through the tunnel plenum with no discharge is shown
~1.1 mm
5 mm
Bow Shock
(a) (b)
Flow
7.5 mm
Fig. 5 Schematic of a bow shock created by a quartz cylinder present
in supersonic flow, along with b a perspective view highlighting the
2-mm stainless steel support rod
2320 2324 2328 23320
20
40
60
80
100
-1]
Sq
rt(I
nt.
) [a
.u.]
2320 2324 2328 2332
Raman Shift [cm Raman Shift [cm-1]
DataCarsfit
(a) (b)
Fig. 6 Pure N2 CARS spectra
at 370 torr with no discharge,
a single-shot and b 20-shot
average, along with Carsfit best
synthetic spectra; Tfit = 322 K
Page 6 of 13 Exp Fluids (2013) 54:1422
123
below in Fig. 6a. While under low resolution the spectral
output of the Stokes beam from the modeless dye laser
appears very similar from one shot to the next, higher
resolution reveals significant shot-to-shot spectral profile
variation. To partially mitigate this, 20 single-shot spectra
are averaged together, an example of which can be seen
below in Fig. 6b, taken at P0 = 370 torr and nominal room
temperature, T * 300 K. For rotational temperature
inference, after the averaging is performed, the spectrum is
fitted with the Sandia CARSFT least squares fitting code
(Palmer 1989). The resulting nitrogen best fit rotational
temperature for the spectrum shown in Fig. 6b is Trot
(N2) = 322 K, with precision equal to ±20 K, determined
as discussed below. For vibrational temperature inference,
the v = 0 and v = 1 bands, which as can be seen in
Fig. 12, are well isolated spectrally and are numerically
integrated, which upon division by v’’ ? 1 to correct for
the cross section (for assumed harmonic potential), yield
the v = 1 and 1 level populations. Note that the baseline
observed in Fig. 12, which is due to a combination of
nonresonant background and data acquisition offset, is
subtracted prior to the numerical integration.
To quantify measurement precision and accuracy, 60
20-shot (for rotation) and 100-shot (for vibration) average
spectra are obtained, identically to that of Figs. 6b and 8a,
respectively, for rotational and vibrational temperature. For
rotational temperature, the set of spectra were individually
fit with the CARSFT code. The results are shown in the
histogram plot in Fig. 7, along with a Gaussian curve with
the same mean temperature, Trot = 316.6 K, and standard
deviation, r = 10.0 K. The system precision is demon-
strated to be quite good, with a 95 % confidence interval of
±20 K, despite the difficulty for this low temperature
condition (due to the small number of rotational levels
populated and the narrow spectral spacing of the low-lying
levels). The system accuracy is comparable, as the true gas
temperature for these measurements is *300 K. Note that
while not measured explicitly for this work, the absolute
accuracy of Q-branch CARS thermometry is typically
found to improve with increasing temperature due to the
population of higher rotational levels, which are more
readily resolved in the CARS spectra.
3.2 Pure nitrogen flow, nsec pulser alone
The first discharge condition for consideration is the case
of 300 torr pure nitrogen flowing through the discharge
with the nsec pulser in operation alone, that is, no DC
sustainer discharge. A typical spectrum is shown in Fig. 8a,
from which a value of Tvib = 761 ± 74 K is inferred,
significantly exceeding the gas temperature, Trot = 379 ±
15 K. Uncertainties reported are for 95 % confidence
interval, based on collection of 60 individual spectra, as
discussed in Sect. 3.1 above. (Note that no measureable
signal for the v = 1 transition is observed when the dis-
charge is not operated, indicating that the appearance of the
280 300 320 3400
2
4
6
8
10
12
14
Fit Temperature [K]
# C
ou
nts
Fig. 7 Histogram of CARS best fit rotational/translational tempera-
ture obtained from Carsfit; nominally room temperature N2,
P0 = 370 torr
2290 2300 2310 2320 23300
10
20
30
40
v = 0 →
← v = 1
Tvib
= 761 +/- 74 K
Trot
(v = 0) = 379 +/- 15 K
Sq
rt(In
t.) [a
.u.]
(a) Raman Shift [cm-1]
2320 2324 2328 2332
(b) Raman Shift [cm-1]
CarsfitData
Fig. 8 Sample 20-shot average
spectra, pulser alone,
300 torr N2 showing
a Nonequilibrium v = 0 and
v = 1 peaks, and b v = 0 peak
with Carsfit synthetic spectra
Exp Fluids (2013) 54:1422 Page 7 of 13
123
vibrational ‘‘hot’’ band is completely due to electron
impact excitation from the plasma, rather than potential
optical pumping due to stimulated Raman scattering).
Figure 8b plots the same data, zoomed on the ground
vibrational band, along with the CARSFT synthetic spec-
trum. Ten of these 20-shot averaged spectra have been
collected and processed for this condition, as well as for the
conditions of 200 and 370 torr nitrogen flowing through the
nsec pulser discharge. The results of these runs have been
plotted and are shown below in Fig. 9, along with error
bars showing the 95 % confidence bounds.
Since the nanosecond pulse peak voltage exceeds 20 kV
(see Fig. 3), the estimated peak reduced electric field in the
5-mm gap between the pulser electrodes is very high,
E/npeak * 300–400 Td at P0 = 200–400 torr. At these
high reduced electric fields, the discharge energy fraction
going to vibrational excitation of nitrogen by electron
impact is predicted to be very low, less than 3 % (Raizer
1991). However, kinetics of energy coupling to the plasma
in a nanosecond pulse discharge is more complex. Although
the electric field before breakdown is very high, field in the
plasma after breakdown is reduced considerably due to
charge separation in the sheath and plasma shielding. Since
the applied voltage continues to vary after breakdown, this
generates displacement current in the external circuit, equal
to the conduction current in the plasma between the elec-
trodes (Adamovich et al. 2009). This causes additional
energy coupling to the plasma at relatively low reduced
electric fields, that is, at the conditions when a significant
fraction of input energy may be loaded directly into the
vibrational energy mode of nitrogen by electron impact.
To evaluate the discharge pulse energy fraction stored in
nitrogen vibrations, we used an analytic model of nano-
second pulse discharge in quasi-one-dimensional plane
geometry, with both electrodes covered by dielectric layers
(Adamovich et al. 2009). Briefly, the model incorporates
key effects of pulsed breakdown, plasma shielding, and
sheath development on nanosecond time scale. The model
predicts pulse energy coupled to the plasma during and
after breakdown versus discharge geometry, pressure,
temperature, and pulse voltage waveform parameters. In
the model, the experimental pulse voltage waveform,
shown in Fig. 3 (Yin et al. 2012), was approximated by a
superposition of Gaussian shape pulses, as shown in
Fig. 10. Multiple voltage peaks in the experimental
waveform are due to pulse reflections off the load (the
discharge) and the pulse generator. The reflected pulses
couple additional energy to the plasma generated during
the incident pulse. It can be shown (Adamovich et al. 2009)
that after breakdown and sheath formation, the field in the
quasi-neutral plasma, Ep(t), which controls the energy
loading kinetics, is given as:
EpðtÞ ¼1
1þ els2l
ee0
2lenele
dVappðtÞdt
� ee0
2lenele
dVappðtÞdt
ð2Þ
where Vapp(t) is the applied voltage, le and ne are electron
mobility and quasi-steady-state electron density reached
after breakdown, l and e are the thickness of the dielectric
layers and the dielectric constant, and ls is the sheath
thickness. The analytic model is in very good agreement
with numerical modeling of nanosecond pulse discharge in
nitrogen (Adamovich et al. 2009).
The calculations have been done for nitrogen at T =
300 K and P0 = 200–370 torr, for the discharge gap of
L = 5 mm, electrode surface area of A = 3.0 x 3.6 cm =
10.8 cm2, and alumina ceramic dielectric thickness of
l = 1.5 mm (e = 9.1). The electron mobility was assumed
to be constant, le = 400 [760/P(torr)] cm2/V s, and the
Townsend ionization coefficient in nitrogen was taken to
be the same as in our previous work (Adamovich et al.
150 200 250 300 350 400
300
400
500
600
700
800
900
1000
Discharge Pressure [Torr]
Tem
per
atu
re [
K]
Tvib
Trot
Fig. 9 Rotational and first-level vibrational temperature for nsec
pulsed N2 discharge alone versus discharge (plenum) pressure
Fig. 10 Applied electric field (experimental and Gaussian fit), field
in the plasma, and electron density in the plasma; nitrogen,
P0 = 300 torr
Page 8 of 13 Exp Fluids (2013) 54:1422
123
2009). The discharge energy fraction going directly into
vibrational excitation by electron impact versus reduced
electric field was taken from the numerical solution of the
Boltzmann equation for plasma electrons (Raizer 1991).
The initial electron density before the pulse was assumed to
be very low, ne0 = 108/cm3. It was shown previously that
pre-ionization starts affecting the coupled pulse energy at
initial electron densities above ne0 = 1011/cm3 (Adamo-
vich et al. 2009).
Figure 10 plots the electron density and the field in the
plasma during and after breakdown at P0 = 300 torr. It can
be seen that after breakdown, the field in the plasma falls
rapidly and becomes about an order of magnitude lower
than the applied field, as discussed previously. During
breakdown, electron density increases by several orders of
magnitude and reaches steady state when the field in the
plasma becomes too low to produce additional ionization.
Figure 11 plots the reduced electric field in the plasma
during and after breakdown at the conditions of Fig. 10, as
well as energy coupled to the plasma, energy loaded into
nitrogen vibrational mode, and total energy in the load
(which includes energy coupled to the plasma and elec-
trostatic energy stored in the capacitive load) versus time.
These results are compared to the total energy in the load
measured in the experiment (Yin et al. 2012).
From Fig. 11, it can be seen that energy coupled to the
plasma by the incident pulse and two reflected pulses is
close to the experimentally measured value (in the exper-
iment, the coupled energy increases by an additional
*10 % due to additional pulse reflections not shown in
Figs. 10 and 11). It is also apparent that vibrational energy
loading occurs primarily after breakdown (during the latter
portion of the incident pulse and during reflected pulses),
when the reduced electric field in the plasma does not
exceed 30 Td (see Fig. 11). The net energy fraction loaded
into nitrogen vibrations, predicted by the model, is fairly
significant, approximately 33 %. This is in fairly good
agreement with the experimental nonequilibrium vibra-
tional and rotational temperatures inferred from the CARS
spectra, *15–30 % at P0 = 200–300 torr (see Fig. 9).
Thus, comparison of the results of CARS measurements
with kinetic model demonstrates that fairly significant
vibrational excitation can be achieved in repetitive nano-
second pulse discharges, due to energy coupling to the
plasma after breakdown.
3.3 Pulser–sustainer discharge, with relaxer injection
Figure 12 shows a pair of N2 CARS spectra obtained from
operation of the pulser–sustainer discharge, with a DC
power supply voltage (VPS) of 4.5 kV. The dash-dotted
black curve corresponds to a baseline case of pure nitrogen
flow at 300 torr, and the solid blue curve shows the spec-
trum when 1 torr partial pressure of CO2 is injected
downstream of the discharge, approximately 9 cm (*2 ms)
upstream of the CARS measurement location, as shown in
Fig. 1. It is clear that there is a significant difference
between the two conditions. For no relaxer injection, the
flow is extremely nonequilibrium, with a vibrational tem-
perature of nearly 2,000 K, while the inferred gas rota-
tional/translation temperature is Trot * 450 K. With only
1 torr partial pressure of carbon dioxide injected, nearly all
the vibrational energy has been removed from the nitrogen,
evidenced by the nearly equilibrated Tvib = 815 K and
Trot = 630 K.
Fig. 11 Reduced electric field, coupled energy, energy loaded into
vibrational mode of nitrogen, and total energy in the load (experi-
mental and predicted); nitrogen, P0 = 300 torr
2290 2300 2310 2320 23300
20
40
60
80
100
Raman Shift [cm-1]
Sq
rt(I
nt.
) [a
.u.]
no inj: Tvib
= 1980K, Trot
= 450K
1 Torr CO2: T
vib = 815K, T
rot = 630K
Fig. 12 CARS spectra for pulser–sustainer discharge at 300 torr total
pressure, in pure nitrogen (i.e., no relaxer injection, black dash-dot)and 1 torr CO2 partial pressure injection (blue solid)
Exp Fluids (2013) 54:1422 Page 9 of 13
123
Five different species (carbon dioxide, nitric oxide,
hydrogen, oxygen, and nitrogen) were injected over a range
of partial pressures. These gases were chosen because their
rates for nitrogen vibrational relaxation vary by several
orders of magnitude. Figure 13 plots both the nitrogen
vibrational and rotational/translational temperatures mea-
sured in these mixtures versus the partial pressure of the
injected species. It can be seen that the addition of oxygen
to the discharge-excited nitrogen (solid red curve), up to
20 % mole fraction (nearly ‘‘synthetic air’’), does not cause
a significant change in either the vibrational or rotational
gas temperature, due to the low vibration–vibration (V–V)
energy transfer rate coefficient, N2 (v = 1) ? O2 (v = 0)
? N2 (v = 0) ? O2 (v = 1), kVV = 3 9 10-17 cm3/s at
Trot = 450 K (Taylor and Bitterman 1969). For injection of
60 torr oxygen, the characteristic time for nitrogen V–V
relaxation, sVV * 1/kVVnO2*17 ms, is significantly
longer than the *2 ms flow residence time between the
injection location and the CARS measurement region. The
characteristic time for V–T relaxation for N2–O2 is even
longer.
As can also be seen in Fig. 13, carbon dioxide (magenta
dash-dot curve), which is known to be an extremely rapid
relaxer of nitrogen vibrations, exhibits the most rapid rate
of relaxation. Even at the lowest injection partial pressure,
PCO2= 0.25 torr, a substantial reduction in nitrogen
vibrational temperature, by more than 450 K, is observed,
as well as a corresponding increase in nitrogen rotational
temperature by 50 K. The rate coefficient for V–V energy
transfer, N2 (v = 1) ? CO2 (000) ? N2 (v = 0) ? CO2
(001), at Trot = 450 K, is very high, kVV = 4.5 9 10-13
cm3/sec (Taylor and Bitterman 1969) (sVV * 1/kVVnO2
*70 lsec for CO2 partial pressure of only 1 torr). CO2
vibrationally excited in collisions with nitrogen relaxes via
rapid intramolecular V–V energy transfer to m2 (010) and m1
(100) modes, with subsequent rapid V–T relaxation.
Injection of nitric oxide and hydrogen exhibits similar
behavior to carbon dioxide, although on a slower time
scale. The rate coefficient of V–V energy transfer from
nitrogen to nitric oxide, N2 (v = 1) ? NO (v = 0) ? N2
(v = 0) ? NO (v = 1), is kVV = 3.4 9 10-15 cm3/sec
at 450 K (Doyennette and Margottin-Maclou 1986)
(sVV * 1/kVVnNO *2 ms for NO partial pressure of 5 torr,
comparable to the flow residence time). The V–T relaxa-
tion rate coefficient for N2–H2, N2 (v = 1) ? H2 ? N2
(v = 0) ? H2, is kVT = 6.1 9 10-16 cm3/sec at 450 K
(Allen et al. 1980) (sVT * 1/kVTnH2*6 ms for H2 partial
pressure of 10 torr). Using the temperature dependence
suggested in the work by Capitelli et al. (2001), this rate
increases to kVT = 1.3 9 10-15 cm3/s at T = 550 K
(sVT * 3 ms). Similar to N2–O2, the V–V energy transfer
for N2–H2 is extremely slow, due to a significant difference
in vibrational quanta (Bott 1976).
Injection of nonexcited nitrogen was also performed;
although the V–V energy transfer rate coefficient for
nitrogen, N2 (v = 1) ? N2 (v = 0) ? N2 (v = 0) ? N2
(v = 1), is relatively high, kVV = 1.5 9 10-14 cm3/sec at
room temperature (Ahn et al. 2004) (sVV * 1/kVVnN2
*0.5 ms for injection N2 partial pressure of 5 torr), the
resonant energy transfer process simply results in the
redistribution of N2 vibrational energy among the ‘‘dis-
charge-loaded’’ and ‘‘cold-injected’’ molecules, but the
energy remains ‘‘locked’’ in the nitrogen vibrational mode.
As the results of Fig. 13 indicate, energy extraction from
the nitrogen vibrational mode results in an increase in the
gas temperature, that is, thermalization of the vibrational
energy as the discharge-loaded flow is equilibrated.
Figure 14 plots the average total nitrogen translational/
rotational ? vibrational energy per molecule as a function
of partial pressure of the injected species for all of the
conditions shown previously in Fig. 13. As the figure
demonstrates, within the experimental uncertainty, the total
.2 .3 .5 .7 1 2 3 5 7 10 20 30 50 70 100
400
600
800
1000
1200
1400
1600
1800
2000
2200
Injection Partial Pressure [Torr]
Tem
per
atu
re [
K]
Tv no inj
Tr no inj
CO2
NOH
2O
2"cold" N
2
Fig. 13 Pulser–sustainer discharge, VPS = 4.5 kV, various injected
species, 300 torr total mixture pressure
0 .2 .3 .5 .7 1 2 3 5 7 10 20 30 50 700
20
40
60
80
100
120
140
Injection Partial Pressure [Torr]
Avg
. En
erg
y L
oad
ing
/mo
lec
[meV
]
no injCO
2NOH
2O
2cold N
2
Fig. 14 Sum of nitrogen translational ? rotational ? vibrational
energy as a function of relaxant species partial pressure
Page 10 of 13 Exp Fluids (2013) 54:1422
123
nitrogen rotational, translational, and vibrational energy is
conserved. This indicates that any inter-species V–V
transfer is followed by rapid V–T relaxation, with a result
that there is negligible energy storage in the vibrational
modes of injected species.
3.4 Measurements in a supersonic flow
Finally, measurements of vibrational temperature in the
supersonic section have also been performed. These mea-
surements are especially challenging due to very low free-
stream static pressure, P0 = 1.2 torr, measured using a
wall pressure tap at the end of the nozzle. As mentioned, a
5-mm-diameter quartz cylinder is positioned in the super-
sonic flow, which creates a bow shock as shown previously
in Fig. 5a. The CARS spectra in Fig. 15a are eight-shot
averages, collected both in the supersonic free-stream and
behind the bow shock for 300 torr pure N2 in the plenum
with no discharge. The difference in signal-level results
from the different number densities present in the two
measurement locations; the spectra also demonstrate typi-
cal signal-to-noise levels for this arrangement. The plots in
Fig. 15b, c show ten-shot average spectra, collected in the
supersonic free-stream and behind the bow shock, respec-
tively, for 300 torr N2 in the plenum, with the pulser–
sustainer discharge in operation (sustainer VPS = 4.5 kV).
Electromagnetic interference (EMI) caused by the nsec
pulser can be clearly seen in both spectra and becomes
increasingly problematic as signal levels decrease. The
significant reduction in the ground vibrational level signal
strengths for both of these spectra (compared to the ‘‘no-
discharge’’ data acquired at each location) can be attributed
to both a reduction in total number density, due to gas
heating from the discharge, and significant population loss
to excited vibrational states. The ratio of the integrated
square root of peak intensities is used to determine the
first-level nitrogen vibrational temperatures, as previously
described.
Due to the quadratic scaling of CARS signal intensity to
the density, as previously mentioned, the integrated square
root of the CARS signal gives an estimate of local number
density. This can be used to determine whether the mea-
surement volume is upstream or downstream of the shock
location, due to the significant rise in density as the flow
traverses the shock. Spectra captured without the discharge
in operation, such as those seen in Fig. 15a, are collected at
various locations upstream of the cylinder, and the square
root of the spectral peaks is integrated to give an estimate
of local number density. The results are shown in Fig. 16a,
where the black x’s are for pulser–sustainer discharge
operating in 300 torr pure N2 flow, and the red ?’s are
recorded with 0.25 torr CO2 injection, as previously
described. As the plot indicates, a sharp rise in density
occurs 1.0 mm upstream of the cylinder surface. The
cluster of data points captured at x = -1.00 mm (number
density *10 in arbitrary units) represent the free-stream
position directly adjacent to the shock front. Moving
downstream, measurements captured at x = -0.99, -0.98,
and -0.95 mm, shown as the next three black x’s on the
plot, demonstrate much higher densities, indicating the
measurement region has moved beyond the shock location.
As mentioned above, shock stand-off distance detected
in the present work, 1.0 mm, is somewhat smaller than
0
2
4(b)
Sq
rt(I
nt)
[a.
u.]
2300 2310 2320 23300
5
10(c)
Sq
rt(I
nt)
[a.
u.]
Raman Shift [cm-1]
0
30
60(a)
Sq
rt(I
nt)
[a.
u.]
Free-streamBehind Shock
Fig. 15 Sample CARS spectra collected in the supersonic flow;
a eight-shot average spectra collected in the free-stream and behind
the shock layer with no discharge present; ten-shot average spectra
with the pulser–sustainer discharge operating, collected in the
b supersonic free-stream and c behind the bow shock
-2 -1.8 -1.6 -1.4 -1.2 -1 -0.8 -0.6 -0.4 -0.2 0300
600
900
1200
1500
1800
2100
Tvi
b [K
]
Distance from cylinder [mm]
Cyl. Surface→
(b)
-2 -1.8 -1.6 -1.4 -1.2 -1 -0.8 -0.6 -0.4 -0.2 0
10
20
30
40
~ n
um
ber
den
sity
[a.
u.]
Bow Shock→(a)
pure N2
CO2 inj.
Fig. 16 Inferred a estimated number density, and b vibrational
temperature in the supersonic flow free-stream and behind the bow
shock
Exp Fluids (2013) 54:1422 Page 11 of 13
123
measured in our previous experiments using a 4-cm-long,
5-mm-diameter cylinder, supported at the tunnel sidewalls,
1.2 mm (Nishihara et al. 2012). This may be explained by
the effect of finite cylinder length, which has been studied
in a Mach 4 flow in a shock tube experiment (Kim 1956).
In the work by Kim (1956), it was shown that the ratio of
stand-off distance to the cylinder diameter, d/D, tends to
decrease as the cylinder length-to-diameter ratio, l/D, is
reduced below l/D = 4. This effect is primarily due to flow
three-dimensionality, that is, spanwise flow behind the bow
shock. At the present conditions, l/D = 7.5/5 mm = 1.25.
The number densities measured behind the shock are
nominally 3–4 times higher than those measured in the
free-stream, somewhat less than the bow shock density
jump of 5.1, predicted by a 3-D compressible Navier–
Stokes flow code for a cylinder model extending wall-to-
wall (Nishihara et al. 2012). The density ratio across the
shock may be also affected by the three-dimensionality of
the flow over the short cylinder model used in the present
work, as well as possible beam steering due to the index
gradient in the vicinity of the shock. Since density mea-
surement was not a focus of this work, beam steering was
not examined in detail, other than confirmation that the
apparent density increased smoothly through the shock with
no large increase in shot-to-shot fluctuations in the signal.
As mentioned previously, the Unstable-resonator Spatially
Enhanced Detection Coherent Anti-Stokes Raman Scatter-
ing geometry was specifically used in order to minimize
potential beam steering (compared to folded BOXCARS),
although the effective of this strategy was not examined in
any detail.
The plot in Fig. 16b shows the inferred Tvib (N2) for
several locations both in the free-stream and behind the
shock. These data are taken with the pulser–sustainer dis-
charge operating in 300 torr N2 in plenum, with and
without 0.25 torr CO2 injection. Both with CO2 injection
and without injection, vibrational temperatures inferred
behind the shock are very similar to those observed in the
free-stream. As the recovery pressure behind the shock is
significantly lower than the plenum pressure, vibrational
relaxation beyond that present in the relatively high density
subsonic flow does not occur. The vibrational temperature
inferred for the case of 0.25 torr CO2 injection is close to
the value inferred in plenum at the same conditions,
Tvib * 1,450 K. The vibrational temperature for pure N2
without injection is *150 K less than the value measured
in plenum at the same conditions, Tvib = 1,900 K, a dif-
ference of less than 10 %. Significant spread in the data is
observed, primarily due to the rather low signal-to-noise
levels present due to the EMI effects from the nsec pulser
previously mentioned, as well as the general difficulty of
very low CARS signal levels in these extremely low-den-
sity flows.
4 Conclusions
Picosecond Unstable-resonator Spatially Enhanced Detec-
tion Coherent Anti-Stokes Raman Scattering Spectroscopy
(USED-CARS) has been used to determine N2 rotational/
translational and first-level vibrational temperature simul-
taneously, in a nonequilibrium flow excited by a nsec pulser/
DC sustainer electric discharge in the plenum of a Mach 5
wind tunnel at pressures in the range 200–370 torr, and ini-
tial temperature of *300 K. It is shown that operation of the
nominally high reduced electric field, (E/n)peak * 500 Td,
nsec pulsed discharge alone results in fairly significant
vibrational loading of pure N2, with Tvib * 720 K at
300 torr. The accompanying rise in translational/rotational
temperature is much smaller, with Trot * 380 ± 20 K.
Modeling predictions based on a recently published 1-D
pulsed nsec discharge model suggest that energy coupling
after discharge breakdown results in significant coupled
energy to the plasma at relatively low E/n (\30 Td), due to
sheath formation and plasma shielding. The addition of an
orthogonal DC sustainer discharge (E/n * 10–30 Td),
which accounts for approximately 80 % of the total power
loading into the flow, is shown to efficiently excite the N2
vibrational energy mode, leading to first-level vibrational
temperatures up to Tvib (N2) = 2,000 K and corresponding
rotational temperatures of Trot * 450 ± 20 K.
Measurements are also performed quantifying the effect
of relaxant gas species on nitrogen rotational/translational
and vibrational temperatures. Specifically, injection of
CO2, NO, and H2 results in partial vibrational relaxation of
nitrogen accompanied by gas heating, demonstrating the
ability to control and tailor the vibrational energy content
of the flow prior to its expansion through the Mach 5
nozzle of the wind tunnel. Determination of average
total nitrogen translational/rotational ? vibrational energy
demonstrates, to within the experimental uncertainty, that
total nitrogen rotational, translational, and vibrational
energy is conserved, indicating that any inter-species V–V
energy transfer occurring is followed by rapid V–T relax-
ation, such that energy storage in molecules other than
nitrogen is negligible.
Finally, the ability to capture vibrational CARS spectra
within very low-density flows present in a Mach 5
expansion nozzle has been demonstrated. Measurements of
first-level vibrational temperature in the supersonic free-
stream and downstream of the bow shock created by
insertion of a cylinder model in the Mach 5 flow have been
performed. Vibrational nonequilibrium persists through the
nozzle throat and expansion, with highly nonequilibrium
gas arriving at the bow shock. Measurements within
300 lm of the cylinder surface reveal that in pure N2
without relaxant species injection, as well as in N2
with 0.25 torr of CO2 injected, no detectable vibrational
Page 12 of 13 Exp Fluids (2013) 54:1422
123
relaxation occurs behind the bow shock, as the recovery
pressure and the corresponding number density are too low.
Acknowledgments The authors would like to thank Dr. Sukesh Roy
for much help and many useful discussions regarding fabrication of
the picosecond broadband dye laser as well as picosecond CARS in
general. Additionally, many thanks are given to Mr. Zhiyao Yin and
Dr. Keisuke Takashima for their work measuring the nsec pulse
waveforms. This research is supported by the AFOSR Hypersonics
program (Technical Manager Dr. John Schmisseur), and the DOE
Low Temperature Plasma Physics Plasma Science Center, Depart-
ment of Energy, Office of Fusion Energy Science Contract DE-
SC0001939.
References
Adamovich IV, Nishihara M, Choi I, Uddi M, Lempert WR (2009)
Energy coupling to the plasma in repetitive nanosecond pulse
discharges. Phys of Plasmas 16:113505
Ahn T, Adamovich IV, Lempert WR (2004) Determination of
nitrogen V–V transfer rates by stimulated Raman pumping.
Chem Phys 298:233–240
Allen DC, Chandler ET, Gregory EA, Siddles RM, Simpson CJSM
(1980) Vibrational deactivation of N2 (v = 1) by n-H2 and by
p-H2. Chem Phys Lett 76:347–353
Bott JF (1976) Vibrational energy exchange between H2 (v = 1) and
D2, N2, HCl, and CO2. J Chem Phys 65:3921–3928
Bruzzese JR, Hicks A, Erofeev A, Cole AC, Nishihara M, Adamovich
IV (2010) Gain and output power measurements in an electri-
cally excited oxygen-iodine laser with a scaled discharge. J Phys
D Appl Phys 43:015201
Capitelli M, Ferreira CM, Gordiets BF, Osipov AI (2001) Plasma
kinetics in atmospheric gases, Chap. 7. Springer, Berlin, p 107
Doyennette L, Margottin-Maclou M (1986) Vibrational relaxation of
NO(v = 1) by NO, N2, CO., HCl, CO2, and N2O from 300 to
600 K. J Chem Phys 84(12):6668–6678
Eckbreth AC (1996) Laser diagnostics for combustion temperature
and species. Gordon and Breach, Amsterdam
Gordiets BF, Osipov VA, Shelepin LA (1988) Kinetic processes in
gases and molecular lasers. Gordon and Breach, London
Kim C-S (1956) Experimental studies of supersonic flow past a
circular cylinder. J Phys Soc Japan 11(4):439–445
Klick D, Marko KA, Rimai L (1981) Broadband single-pulse CARS
spectra in a fired internal combustion engine. Appl Opt 20(7):1178
Lempert WR (2005) Diagnostics. In: Becker K, Kogelschatz U,
Schoenbach K, Barker R (eds) Non-equilibrium air plasmas at
atmospheric pressure. IOP Publishing, Bristol, UK
Leyva IA, Laurence S, War-Kei Beierholm A, Hornung HG, Wagnild
R, Candler G (2009) Transition delay in hypervelocity boundary
layers by means of CO2/acoustic instability interactions. AIAA
Paper 2009-1287
Marko KA, Rimai L (1979) Space- and time-resolved coherent anti-
stokes Raman spectroscopy for combustion diagnostics. Opt Lett
4(7):211
Montello A, Nishihara M, Rich JW, Adamovich IV, Lempert WR
(2012) Nitrogen vibrational population measurements in the
plenum of a hypersonic wind tunnel. AIAA J 50(6):1367
Nishihara M, Jiang N, Rich JW, Lempert WR, Adamovich IV,
Gogineni S (2005) Low-temperature supersonic boundary layer
control using repetitively pulsed MHD forcing. Phys Fluids
17(10):106102
Nishihara M, Rich JW, Lempert WR, Adamovich IV, Gogineni S
(2006) Low-temperature M = 3 flow deceleration by Lorentz
Force. Phys Fluids 18(8):086101
Nishihara M, Takashima K, Rich JW, Adamovich IV (2011) Mach 5
bow shock control by a nanosecond pulse surface dielectric
barrier discharge. Phys Fluids 23:066101
Nishihara M, Takashima K, Jiang N, Lempert WR, Adamovich IV,
Rich JW, Doraiswamy S, Candler GV (2012) Development of a
mach 5 nonequilibrium flow wind tunnel. AIAA J 50:2255–2267
Palmer RE (1989) The CARSFT computer code for calculating
coherent anti-stokes Raman spectra: user and programmer
information. Sandia National Labs., Rept SAND89-8206,
Livermore, CA
Raizer YuP (1991) Gas discharge physics. Springer, Berlin
Roy S, Meyer TR, Gord JR (2005) Broadband coherent anti-stokes
Raman scattering spectroscopy of nitrogen using a picosecond
modeless dye laser. Opt Lett 3:3222
Roy S, Gord JR, Patnaik AK (2010) Recent advances in coherent anti-
stokes Raman scattering spectroscopy: fundamental develop-
ments and applications in reacting flows. Prog Energ Combust
Sci 36:280–306
Taylor RL, Bitterman S (1969) Survey of vibrational relaxation data
for processes important in the CO2-N2 laser system. Rev Mod
Phys 41:26–47
Wagnild R, Candler GV, Leyva IA, Jewell JS, Hornung HG (2010)
Carbon dioxide injection for hypervelocity boundary layer
stability. AIAA Paper 2010-1244
Yin Z, Takashima K, Adamovich IV (2012) Measurements and
kinetic modeling analysis of energy coupling in nanosecond
pulse dielectric barrier discharges. AIAA Paper 2012-3093, 43rd
AIAA plasmadynamics and lasers conference, New Orleans, LA,
25–28 June 2012; accepted for publication in Plasma Sources
Science and Technology, 2012
Exp Fluids (2013) 54:1422 Page 13 of 13
123