Upload
ali-soleimani
View
212
Download
0
Embed Size (px)
Citation preview
SCIENTIFIC PAPER
Photon activation therapy: a Monte Carlo study on doseenhancement by various sources and activation media
Mahdi Bakhshabadi • Mahdi Ghorbani •
Ali Soleimani Meigooni
Received: 13 November 2012 / Accepted: 28 July 2013 / Published online: 11 August 2013
� Australasian College of Physical Scientists and Engineers in Medicine 2013
Abstract In the present study, a number of brachytherapy
sources and activation media were simulated using
MCNPX code and the results were analyzed based on the
dose enhancement factor values. Furthermore, two new
brachytherapy sources (131Cs and a hypothetical 170Tm)
were evaluated for their application in photon activation
therapy (PAT). 125I, 103Pd, 131Cs and hypothetical 170Tm
brachytherapy sources were simulated in water and their
dose rate constant and the radial dose functions were
compared with previously published data. The sources
were then simulated in a soft tissue phantom which was
composed of Ag, I, Pt or Au as activation media uniformly
distributed in the tumour volume. These simulations were
performed using the MCNPX code, and dose enhancement
factor (DEF) was obtained for 7, 18 and 30 mg/ml con-
centrations of the activation media. Each source, activation
medium and concentration was evaluated separately in a
separate simulation. The calculated dose rate constant and
radial dose functions were in agreement with the published
data for the aforementioned sources. The maximum DEF
was found to be 5.58 for a combination of the 170Tm source
with 30 mg/ml concentration of I. The DEFs for 131Cs and170Tm sources for all the four activation media were higher
than those for other sources and activation media. From
this point of view, these two sources can be more useful in
photon activation therapy with photon emitter sources.
Furthermore, 131Cs and 170Tm brachytherapy sources can
be proposed as new options for use in the field of PAT.
Keywords Brachytherapy � Photon activation
therapy � Dose enhancement � TG-43 parameters
Introduction
In photon activation therapy (PAT), the energy deposition
within a tumour may be increased by provoking the
emission of Auger electrons through photoelectric effect,
by addition of high Z elements [1]. The binding energy of
the activation element within the tumour is closely matched
with the incident photon energy to maximize the produc-
tion of free electrons and, subsequently, enhancement of
the absorption energy [2].
When an atom is ionized, the excited atom returns to
stable state via radiative and nonradiative processes. The
radiative processes result in emission of characteristic
X-ray photons, while the nonradiative transitions result in
ejection of atomic electrons, via Auger, Coster–Kronig and
super Coster–Kronig processes. The ejected electrons
through nonradiative processes, are denoted as Auger
electrons. Furthermore, physical decay of the radionuclide
by either orbital electron capture or internal conversion will
also result in inner atomic shell ionization. Such radio-
nuclides are known as Auger emitters, of which 125I is one
that is commonly used in brachytherapy [3].
For photon activation therapy, it is feasible to use var-
ious sources and activation media. There are studies in
which the synchrotron X-rays were used for the purpose of
photon activation therapy [4–6]. The possibility of tuning
M. Bakhshabadi
North Khorasan University of Medical Sciences, Bojnurd, Iran
M. Ghorbani (&)
Medical Physics Department, Faculty of Medicine, Mashhad
University of Medical Sciences, Pardis-e-Daneshgah, Vakil
Abad Blvd., 9177948564 Mashhad, Iran
e-mail: [email protected]
A. S. Meigooni
Comprehensive Cancer Centers of Nevada, Las Vegas, NV, USA
123
Australas Phys Eng Sci Med (2013) 36:301–311
DOI 10.1007/s13246-013-0214-0
the energy of X-rays to the K-edge energy make a syn-
chrotron a great tool for photon activation therapy. How-
ever, synchrotrons are not as widely available as other
radiation sources [7]. Brachytherapy sources may be an
alternative for photon activation therapy. Since brachy-
therapy provides much greater localization of radiation
dose to the tumour, compared to external beam therapy
sources, use of brachytherapy sources in photon activation
therapy can decrease normal tissue radiation dose, and
therefore improve the therapeutic outcomes compared to
external radiotherapy sources [1].
In a study by Nath et al. [8], the dependence of IUdR
radiosensitization on photon energy was investigated
through an in vitro study irradiating Chinese hamster cells at
dose rate of 0.72 Gy/h. IUdR radiosensitization was com-
pared for 226Ra, 241Am and 125I sources. The results have
indicated that IUdR produces significant radiosensitization
with all three sources at 0.72 Gy/h irradiation dose rate. This
is due to an increase in IUdR radiosensitization just above the
K-edge energy of iodine atoms with photon induced Auger
electrons. Greater radiosensitization was observed for 241Am
photons compared to 226Ra on the higher energy range and to125I on the lower energy range. The results suggested that a
combination of IUdR with low energy emitting sources may
have a clinical potential for photon activation therapy.
In a photon activation therapy study by Young et al. [2], a
dose enhancement by silver tetraphenyl sulfonato porphyrin
(AgTPPS4) agent with an 125I brachytherapy source was
quantified. The dose in the tumour volume was calculated
for various concentrations of AgTPPS4, ranging
0–20 mmol/kg, using the EGS4 Monte Carlo code. There
was a linear increase in tumour dose with corresponding
increase in AgTPPS4 concentration. In vitro toxicity studies
have shown that TPPS4 porphyrin derivatives were cyto-
toxic at concentration levels needed to have significant
brachytherapy dose enhancement. However, the toxicity
effect was not due to presence of silver atoms in the
structure of AgTPPS4 Ranjbar et al. [9], have investigated
the dose enhancement by gold nanoparticles when the tissue
is irradiated by a monoenergetic photon beam. MCNPX
code was used for this purpose and a phantom model was
developed and verified by comparison of the obtained depth
dose data with corresponding published values. Following
this verification, four various concentrations of gold nano-
particles were introduced in the tumour inside a phantom.
Monoenergetic synchrotron X-rays were simulated as the
irradiation source. The optimum energy for dose enhance-
ment was found to be around 83–90 keV for all concen-
trations. Dose enhancement factor was increased linearly
with concentration of the activation elements. In a study by
Karnas et al. [10], DNA damage due to Auger electrons
from 127I atoms, when irradiated by external X-rays was
investigated. In this study DNA double-strand breaks
(DSBs) were evaluated by Monte Carlo simulations and the
results were compared with measured data. The results of
Monte Carlo simulations were in good agreement with
measurements using comet assay. The results have shown
that the Auger electrons produced approximately 0.03
DSBs per vacancy in N-shell and 0.3 DSBs per vacancy in
K-shell or L-shell. It was concluded that Auger electrons
from iodine produced a modest increase in the number of
strand breaks. The order of the increase was 10 % but the
DSBs were very difficult to be repaired or potentially prone
to misrepair because of their complex nature. Based on the
results of this study, it was proposed that the DNA damage
may have consequences in cell survival and will be
exploitable in photon activation therapy of tumours sensi-
tized with iodine using kilovoltage X-rays. In a different
study by Moiseenko et al. [11], nucleosomal DNA damage
from Auger electrons due to the incorporation of IUdR was
modeled by Monte Carlo simulations. DSB production
following vacancy in various atomic shells of iodine was
estimated. The results have also indicated that the proba-
bility of an Auger electron cascade leading to at least one
DSB depends on the shell of initial vacancy production. The
probability was approximately equal to 0.35 for K and L
shells and 0.02 for the N shell. Furthermore, DSBs were
predominantly produced in a nucleosome containing iodine.
Approximately 14 % of DSBs were produced due to a
vacancy in the L1 orbital. The results provide a basis for
photon activation therapy using kilovoltage X-rays when
the tumour is sensitized by IUdR. In another study, as a Ph.
D. thesis research by Stephens [12], radioactive 161Ho has
been successfully created and characterized for photon
activation therapy. This radioisotope emits K-edge X-rays
of 45 and 52 keV which are just above the K-edge of iodine
(33 keV). In this research the survival of human colorectal
cancer (HT-29) cells with and without presence of IUdR
was evaluated following irradiation of the cells with pho-
tons emitted by 137Cs, 125I and 161Ho radioactive sources
and 300 kVp X-rays. The results have indicated that full
efficacy of a combination of using 161Ho source and IUdR
activation medium, which was not exploited for many
years, can be realized. Furthermore, in vivo and clinical
studies on this combination have shown the potential to
save lives and thus are of importance for consideration. In a
study by Young et al. [13], the tumour dose was enhanced
while minimizing the absorbed dose by the surrounding
normal tissues. To achieve this goal a 125I brachytherapy
source was used. The number of photoelectric interactions
within the tumour was increased through the introduction of
a silver compound to take advantage of the K-edge effect.
X-ray fluorescence excitation spectroscopy method and
clinically accepted calculation techniques were utilized to
estimate the absorbed dose enhancement by the presence of
7.5 mM of silver in a water phantom. There was excellent
302 Australas Phys Eng Sci Med (2013) 36:301–311
123
agreement between the Monte Carlo and experimental
results. The results of this study have indicated that the
K-edge enhancement effect in brachytherapy with 125I
sources is possible with further development of nontoxic
silver compounds.
As previously mentioned, there are various Monte Carlo
and cellular studies on photon activation therapy by various
radiation sources and different activation media. However,
to the best of our knowledge, in most of these studies, a single
source and no activation medium was evaluated and, there is
no consensus study comparing various combinations of
radiation sources and activation media. In the present study,
various brachytherapy sources and activation media com-
binations were simulated by MCNPX code and the results of
dose enhancement factors were compared with the aim of
selecting the best source and media combination. Further-
more, in the process 131Cs and 170Tm sources were added to
the list of the sources used in photon activation therapy.
Materials and methods
Radioactive sources
Three brachytherapy sources and a hypothetical source
were evaluated in the present study: 125I, 103Pd, 131Cs and
hypothetical 170Tm source. 125I and 103Pd nuclides have
been previously used in photon activation therapy studies
[1, 8]. However, the 131Cs and 170Tm sources are presented
in this study as new options for photon activation therapy.
The physical characteristics and the source models used in
this study are listed in Table 1.
A schematic diagram of the 125I, 103Pd, 131Cs and
hypothetical 170Tm sources used in this study are illustrated
in Fig. 1. In this figure, the dimensions of various sources’
parts including the active core, encapsulation and source
cable were mentioned.
We have used an IsoAid Advantage 125I source (model
IA1-125A) in our simulations. This source model includes
a cylindrical silver X-ray marker with 3 mm length and
0.5 mm diameter which is coated with AgI within
0.001 mm thick. The AgI coating includes 125I isotope
uniformly distributed on the cylindrical surface of the
marker as well as two ends. The marker and coating
assembly is encapsulation within 0.05 mm thickness tita-
nium. The overall length and diameter of the source is 4.5
and 0.8 mm respectively [14, 18].
The IsoAid Advantage 103Pd source (model IAPd-103A)
is composed of a cylindrical silver X-ray marker with
1.25 mm length and 0.5 mm diameter. Two spherical
pellets made of polystyrene with 0.5 mm diameters and
mass density of 1.2 g/cm3 are placed at each end of the
Table 1 Characteristics of the
four source models used in this
study
Source type Half-life (days) Source model Reference
125I 59.40 IsoAid Advantages (model IA1-125A) Meigooni et al. [14]103Pd 16.991 IsoAid Advantages (model IAPd-103A) Meigooni et al. [15]131Cs 9.7 IsoRay (model CS-1) Murphy et al. [16]170Tm 128.6 Flexisource (hypothetical) Ballester et al. [17]
Fig. 1 Geometry of the four source models used in this study: a IsoAid Advantage 125I source [14], b IsoAid Advantage 103Pd source [15],
c IsoRay 131Cs source [16], d hypothetical Flexisource 170Tm source [17]
Australas Phys Eng Sci Med (2013) 36:301–311 303
123
marker. 103Pd isotope is absorbed uniformly throughout the
volume of the polystyrene pellets. The source has a tita-
nium capsule with 4.5 mm length and 0.8 mm outer
diameter. The wall thickness of the capsule is 0.05 mm.
The thickness of the two end caps is 0.35 mm. The
effective active length of the source, Leff, was taken as
3.61 mm. This length is the maximum distance between
the proximal and distal ends of the active pellets (Fig. 1)
[15, 19].
In the IsoRay 131Cs source (model CS-1) the capsule is a
titanium tube with outer and inner diameters of 0.8 and
0.7 mm, respectively. There is also a central X-ray gold
marker with 0.25 mm diameter. The gold marker is sur-
rounded by a glass/ceramic tube with 0.4 and 0.65 mm
inside and outside diameters, respectively. The active 131Cs
source is uniformly distributed throughout the outer
0.11 mm thickness of the glass/ceramic complex. The
overall length of the source complex is 4.7 mm, including a
4.5 mm long tube and two 0.1 mm thick caps on the ends
(Fig. 1) [16].
Since there were no commercial 120 Tm brachytherapy
sources available, a hypothetical one was simulated and
used for the photon activation study. Our selection of this
hypothetical source was based on a study by Ballester
et al. on 170Tm radionuclide proposed for use in medical
application [17]. The source geometry is similar to the
commercially available Flexisource 192Ir HDR source
[20]. However, in the hypothetical 170Tm source, the
active core is composed of 170Tm radionuclide. The active
core of the hypothetical 170Tm source is a pure 170Tm
cylinder (density = 9.3 g/cm3). The core’s active length
is 3.5 mm and its diameter is 0.6 mm. The active core has
a 304-type stainless steel capsule (density: 8.0 g/cm3).
The outer diameter of the capsule is 0.85 mm, having a
total length of 4.6 mm. The source cable was simulated as
a 304 stainless steel cylinder with 5 mm length and
0.5 mm diameter (Fig. 1) [17]. The gamma and X-rays of
the 170Tm source were defined in the simulations, but the
beta emissions by 170Tm radionuclide were ignored in this
study.
Activation media used
In photon activation therapy, the Auger electrons have a
dominant role in transferring the radiation energy to the
DNA molecules. Production of these electrons originates
from vacancies in the atomic K, L shells. In this study, four
activation media were used in the target volume: Ag, I, Pt
and Au. These elements were selected based on previous
studies in the field of photon activation therapy [1, 8, 9,
21]. With the purpose of a comparison, the orbital energy
corresponding to K and L shells for Ag, I, Pt and Au ele-
ments are listed in Table 2. The data presented in this table
will be useful in the comparison of dose enhancements
related to these elements, with the purpose of evaluation of
the dependence of dose enhancement on the type of acti-
vation medium.
Monte Carlo simulations of the sources
Dosimetric parameters of the sources were determined
according to the TG-43U1 protocol [23]. Following this
protocol, the dose rate distribution around a brachytherapy
source is determined from the following formalism:
_Dðr; hÞ ¼ SkKGðr; hÞ
Gðr0; h0ÞgðrÞFðr; hÞ ð1Þ
where Sk is the air kerma strength of the source; K is the
dose rate constant at reference point of (1 cm, p/2); G(r, h)
is the geometry function; g(r) is the radial dose function;
and F(r, h) is the anisotropy function of the source. Details
on the definition of the above quantities can be found in the
update of task group No. 43 report (TG-43U1) by Ameri-
can Association of Physicists in Medicine (AAPM) [23]
and previous publications on TG-43 parameter determina-
tions for various brachytherapy sources [24–26], therefore
they are not repeated here. MCNPX (version 2.4.0) Monte
Carlo code was used. This code is a general purpose Monte
Carlo code which is used for transport of electrons, pho-
tons, neutrons and other particles in various radiation
problems. It includes a tool for geometry modeling and
various tallies to score particle flux, particle current, energy
deposition and other radiation physics quantities [27]. The
dose rate constant and radial dose function were calculated
through simulations of the source and were compared with
the previously published data. For this purpose, the source
geometries, including before active core, capsule, cable,
etc. were defined in the simulations and the dose values
were obtained in a water phantom.
In the calculation of the air kerma strength, the source
was positioned in vacuum and a number of toroid cells
containing dry air were defined at distances ranging from
5–10 cm from the source. The thickness of each toroid cell
Table 2 K-edge and L-edge energy (keV) for the four activation
media used in this study
Activation
media
Atomic
number
K-edge L-edge
L-I L-II L-III
Ag 47 25.5140 3.8058 3.5237 3.3511
I 53 33.1694 5.1881 4.8521 4.5571
Pt 78 78.3948 13.8799 13.2726 11.5637
Au 79 80.7249 14.3528 13.7336 11.9187
The data were adopted from [22]
304 Australas Phys Eng Sci Med (2013) 36:301–311
123
was 0.1 cm. Air kerma was scored in the tally cells using
F6 tally following scoring 7 9 107 photons.
The dose rate constant was calculated by dividing dose
rate at reference point (r0 = 1, h0 = 908) by air kerma
strength. Dose rate at the reference point was obtained
using *F8 tally in a water phantom with radius of 15 cm.
The input file was run for 107 photons and the related error
in the tally cell in the reference point was less than 3.32 %
for the four sources.
The radial dose function was obtained at radial distances
0.2–10 cm from the source in a water phantom with 15 cm
radius. For this purpose, *F4 tally was calculated in toroid
cells defined at these distances. The thickness of the toroid
cells was 0.1 cm and they had not overlapping with each
other. *F4 tally scores photon fluence. The fluence values
were then converted to absorbed dose by multiplication of
fluence values to the corresponding mass energy absorption
coefficients for water extracted from NIST database [28].
Since the mass energy absorption coefficient depends on
the photon energy, photon fluence in various energy bins
were obtained and then multiplied by the corresponding
mass energy absorption coefficient at that energy bin. The
total dose then was calculated by summing the dose frac-
tions in these energy bins. It was possible to use *F8 tally
to calculate the energy deposition directly, but *F4 tally
was applied to minimize the running time needed for the
simulations. Maximum Monte Carlo statistical error in the
tally cells was 2.59 % following transport of 1.5 9 108
primary photons.
The photon spectra including the emitted photon ener-
gies and the related probabilities used in the definition of
the four sources are listed in Table 3. The data in this table
were adopted from [17, 23, 29, 30].
Dose enhancement calculations
The dose enhancement in the tumour was obtained by
calculation of dose enhancement factor (DEF). Dose
enhancement factor was calculated from the following
definition:
DEF values were obtained on a transverse plane inside
the tumour volume and the maximum and averaged values
were reported as DEF for that specific source and activa-
tion media. The source was positioned inside a spherical
phantom with 15 cm radius. The phantom contained soft
tissue. The composition for soft tissue was defined as
outlined by ICRU report, 44 [31]. Based on this report the
soft tissue composition was: 76.2 % oxygen, 10.1 %
Table 3 Energy spectrum data for the 125I, 103Pd, 131Cs and 170Tm radionuclides used in the present study
125I (TG-43U1 [23]) 103Pd (Rivard [29]) 131Cs (Rivard [30]) 170Tm (Ballester et al. [17])
Energy
(keV)
Photons per
disintegration
Energy
(keV)
Photons per
disintegration
Energy
(keV)
Photons per
disintegration
Energy
(keV)
Photons per
disintegration
27.202 0.406 20.074 0.224 4.11 0.086 6.95 0.000276
27.472 0.757 20.216 0.423 29.461 0.211 7.42 0.0292
30.98 0.202 22.717 0.104 29.782 0.389 48.22 0.000291
31.71 0.0439 23.312 0.0194 33.562 0.0363 49.13 0.000513
35.492 0.0668 39.755 0.000683 33.624 0.0702 51.35 0.0097
62.51 0.0000104 34.419 0.0213 52.39 0.0169
294.95 0.000028 55.48 0.0000546
357.46 0.000221 55.67 0.0001055
497.054 0.0000401 57.14 0.0000354
59.16 0.00185
59.38 0.00356
60.96 0.001205
78.7 0.000035
84.26 0.0248
DEF ¼ Dose in a point with presence of activation media
Dose in the same point without presence of activation mediað2Þ
Australas Phys Eng Sci Med (2013) 36:301–311 305
123
hydrogen, 11.1 % carbon and 2.6 % nitrogen. The central
part of the phantom was defined as the tumour. The tumour
was a sphere with 1.5 cm radius. The compositions of soft
tissue phantom and tumour were the same, but with the
difference that the composition of activation media was
added to that of soft tissue in the case of tumour containing
activation media. Dose values were then calculated in the
absence and presence of activation media inside the
tumour. DEF for the four activation media (Ag, I, Pt and
Au) with concentrations of 7, 18 and 30 mg/ml tumour was
obtained. Each source, activation medium and concentra-
tion was evaluated separately in a separate simulation. A
schematic diagram illustrating the position of source,
tumour and phantom can be seen in Fig. 2. Energy depo-
sition in tally cells (MeV) was calculated using *F8 tally.
For this purpose, this tally type was scored for electrons
before and after loading of activation media in the tumour
in the simulations. Since the mass of tally cells were dif-
ferent for various concentrations and activation media, the
energy deposition was then divided by the corresponding
mass to obtain energy deposition in term of MeV/g. Energy
cut off of 1–10 keV was defined for both photons and
electrons in the simulations for obtaining dose enhance-
ments. The energy cut offs for the sources were the same as
those values defined in simulation of the sources for
obtaining TG-43 parameters. Furthermore, the effect of
energy cut off on the value of dose enhancement was
evaluated for a number of situations. For this purpose, dose
enhancement for 103Pd and 170Tm sources in presence of
Ag in the tumour with various concentrations was com-
pared for 10 and 1 keV cases. Toroid cells with 0.1 cm
thickness were defined as the tally cells at radial distances
of 0.4–6.4 cm from the source center on the transverse
plane. This amount of thickness was defined to avoid
overlapping of the toruses. The number of photons trans-
ported in dose enhancement calculations in each input file
was 1.5 9 108 and the maximum statistical error in the
tally cells was 18.12 %. However, the maximum MC error
was on average 3.43 %. With respect to these values it
should be mentioned that, since there were four sources,
four activation media, and three concentrations, the total
number of forty-eight simulations were performed in DE
calculations. The maximum MC error in each simulation
was noted and the maximum (18.12 %) and average
(3.43 %) of these maximum values were reported.
Results
Dose rate constant and radial dose function
As it is mentioned in the materials and methods section,
the dose rate constant and radial dose function of the 125I,103Pd, 131Cs and hypothetical 170Tm sources were
obtained and compared with the previously published
data for the source models. The results of dose rate
constants are presented in Table 4. The corresponding
previously published data can be considered in the table
as well. Additionally, the percentage differences of our
calculated values and the published data at each specific
point were listed in this table. The radial dose function
values from the calculations, previously reported data and
the percentage differences between the two data sets are
listed in Table 5.
As is evident from the data in Table 4, the maximum
discrepancy between two sets of dose rate constants is
related to 131Cs source which is equal 9.91 %.
From this table it is evident that there are a number of
points with negative percentage values. Since the percentage
Fig. 2 A schematic geometry illustrating position of source, tumour
and soft tissue phantom and the related dimensions
Table 4 Dose rate constant
(cGy h-1 U-1) for the four
brachytherapy sources obtained
in this study and from the other
studies
Source type This study Other studies Reference Difference (%)
125I 1.03 0.98 Meigooni et al. [14] 4.78103Pd 0.701 0.709 Sowards [32] -1.16131Cs 1.006 0.915 Murphy et al. [16] 9.91
Hypothetical 170Tm 1.20 1.23 Ballester et al. [17] -2.8
306 Australas Phys Eng Sci Med (2013) 36:301–311
123
difference between our values and those from other
studies was calculated, these data points are related to the
points in which the data from this study was less than that
from the other study. As it is evident from the data in
Table 5, the maximum difference between the radial dose
function values from the simulations and accepted values
amounts to 13.99 %, which is related to the distance of
6 cm from the 170Tm source. The reason for this level of
discrepancy for 170Tm source can be related to the fact
that we have ignored the beta emissions by 170Tm source,
while they were considered in the previous study on this
source.
Dose enhancement
The maximum dose enhancement factors in the transverse
plane in the tumour for the mentioned sources and
activation media are presented in Table 6. This table also
includes the averaged DEFs in the transverse plane in the
tumour.
As it can be considered from the values of Table 6, the
maximum dose enhancement factor is related to 170Tm
source and I media combination which amounts to 5.58.
The average value of dose enhancement is highest for the
source 170Tm and I combination. However, DEF for170Tm source for other materials and for 131Cs source for
Pt and Au are close to this value. DEF for 125I and 103Pd
sources with Pt and Au are relatively higher than those for
Ag and I.
The evaluation of the effect of energy cut off on dose
enhancement in the case of 103Pd and 170Tm sources in the
presence of Ag shows that the dose enhancement was
unchanged when 10 or 1 keV energy cutoffs were used for
the particles.
Table 5 Radial dose functions for the four brachytherapy source models compared with other studies
r (cm) 125I 103Pd 131Cs Hypothetical 170Tm
Present
study
Meigooni
et al. [14]
Diff.
(%)
Present
study
Sowards
[32]
Diff.
(%)
Present
study
Murphy
et al. [16]
Diff.
(%)
Present
study
Ballester (Personal
communication)
Diff.
(%)
0.2 1.070 1.066 0.38 1.189 1.234 -3.68 0.957 – – – – –
0.3 1.069 1.065 0.36 1.272 1.296 -1.88 0.978 – – – – –
0.4 1.065 1.056 0.86 1.273 1.290 -1.30 0.989 – – 0.896 0.968 -7.44
0.5 1.055 1.048 0.69 1.244 1.260 -1.28 0.996 1.003 -0.71 0.917 – –
0.6 1.047 1.041 0.61 1.201 1.213 -0.95 1.001 – – 0.935 0.972 -3.76
0.7 1.038 1.042 -0.41 1.154 1.160 -0.51 1.004 – – 0.952 – –
0.8 1.027 1.027 -0.01 1.102 1.106 -0.38 1.003 – – 0.968 0.985 -1.73
0.9 1.014 1.013 0.09 1.050 1.053 -0.28 1.003 – – 0.985 – –
1 1.000 1.000 0.00 1.000 1.000 0.00 1.000 1.000 0.00 1.000 1.000 0.00
1.5 0.925 0.923 0.18 0.766 0.768 -0.31 0.974 – – 1.071 1.034 3.58
2 0.841 0.834 0.86 0.576 0.576 -0.02 0.930 0.923 0.74 1.130 1.065 6.15
2.5 0.756 0.750 0.85 0.428 0.429 -0.25 0.875 – – 1.178 – –
3 0.675 0.669 0.89 0.317 0.318 -0.47 0.813 0.806 0.87 1.218 1.108 9.89
3.5 0.598 0.592 1.02 0.233 0.233 0.07 0.751 – – 1.251 – –
4 0.528 0.523 1.01 0.171 0.173 -1.10 0.687 0.679 1.19 1.273 1.131 12.59
4.5 0.462 0.462 -0.10 0.126 0.127 -0.82 0.627 – – 1.290 – –
5 0.406 0.399 1.63 0.093 0.092 0.68 0.568 0.558 1.73 1.295 1.138 13.78
5.5 0.355 0.353 0.54 0.067 0.069 -2.75 0.512 – – 1.292 – –
6 0.309 0.305 1.28 0.050 0.050 -0.83 0.460 0.454 1.36 1.288 1.130 13.99
6.5 0.272 0.269 0.99 0.036 0.037 -2.16 0.411 – – 1.272 – –
7 0.234 0.222 5.52 0.027 0.028 -4.11 0.369 0.361 2.27 1.253 1.112 12.68
7.5 0.202 0.189 6.97 0.021 0.020 2.67 0.326 – – 1.228 – –
8 0.175 0.163 7.31 0.015 0.015 -2.04 0.291 – – 1.203 1.082 11.15
8.5 0.152 0.138 10.34 0.011 0.011 1.87 0.256 – – – – –
9 0.131 0.126 3.81 0.009 0.008 8.19 0.229 – – – – –
9.5 0.113 0.105 7.80 0.006 0.006 3.43 0.202 – – – – –
10 0.097 0.090 7.93 0.005 0.005 0.22 0.179 – – – – –
Australas Phys Eng Sci Med (2013) 36:301–311 307
123
Plot of DEF and dose rate (cGy mCi-1 h-1) versus
radial distance from the source for Au with 7, 18 and
30 mg/ml concentrations in the case of the four sources
used in this study was illustrated in Fig. 3. The minimum
value of dose enhancement factor in the cells outside the
tumour was 0.04. The value is related to 103Pd source when
Au element with 30 mg/ml is used as the activation media.
The average value of DEF outside the tumour for the 125I,103Pd, 131Cs and hypothetical 170Tm sources for all media
are 0.54, 0.43, 0.59 and 0.88 respectively.
Discussions and conclusions
Three brachytherapy sources (125I, 103Pd and 131Cs) and a
hypothetical 170Tm source were simulated and dose rate
constant and radial dose function were obtained and com-
pared with the published data for the sources. The dosi-
metric values from the present study are in agreement with
the previously published data for these source models,
validating the simulations.
The brachytherapy sources with four activation agents
(Ag, I, Pt and Au) were studied to determine the dose
enhancement in photon activation therapy. Two brachy-
therapy sources, 131Cs and 170Tm, were added to the list of
the sources and were evaluated for use in photon activation
therapy applications. From all the possible situations for
combinations of these sources and media (Table 6), the
hypothetical 170Tm source when used with I medium loa-
ded in the tumour had the highest dose enhancement factor
of 5.58. The 131Cs source shows a DEF value (4.82) close
to this value which is related to 30 mg/ml concentration of
Au activation media.
By comparing various sources from a DEF point of view
(Table 6), DEF for 131Cs and hypothetical 170Tm sources
are higher than those for 125I and 103Pd sources. Although
there are minor differences between the DEFs for 131Cs and170Tm sources, these differences are not high and can be
relatively considered equal. When comparing various
media (Table 6), DEF values were higher for Pt and Au
media with 125I and 103Pd sources, and for Ag, Pt and Au
with 131Cs source. The DEF values for various media for
hypothetical 170Tm source were relatively equal. By
keeping in mind the sources’ energies, atomic numbers and
orbital edge energies of the activation media, it is not easy
to find simple rules for description of increasing trends of
DEFs for various cases. However, performing a study to
determine the spectrum of electrons and photons at the
surface of the source capsule and at various depths in the
phantom may be useful.
As it can be seen from the data in Table 6, neglecting
some exceptions, DEF increases with concentration of the
activation media. This fact can be justified with consider-
ing that with increased concentration, the number of acti-
vation element atoms present in the limited tumour volume
increases. Thus, the probability of interaction of source
photons increases with activation element concentration
leading to a higher dose enhancement.
From the data in Fig. 3 it can be noticed that both DEF
and dose rate have decreasing trends inside the tumour.
However, dose rate (cGy mCi-1 h-1) decreases with
steeper trends inside the tumour for all the four sources in
presence of Au media in the tumour. The decreasing trend
in DEF in the tumour can be due to the fact that a number
of low energy photons are absorbed at first few millimeters
in the tumour. Therefore, the photon beams will have
Table 6 Maximum and average dose enhancement factor in tumour for 125I, 103Pd, 131Cs and hypothetical 170Tm sources and various con-
centrations of Ag, I, Pt and Au media
125I 103Pd 131Cs Hypothetical170Tm
Concentration (mg/ml) 7 18 30 7 18 30 7 18 30 7 18 30
Maximum dose enhancement factor
Activation media
Ag 1.73 2.63 3.35 1.18 1.44 1.67 2.05 3.38 4.44 2.02 3.50 5.02
I 1.35 1.85 2.32 1.26 1.60 1.88 1.47 2.12 2.74 2.17 3.85 5.58
Pt 2.01 3.18 4.03 1.72 2.31 2.48 2.14 3.57 4.73 1.97 3.42 4.89
Au 2.04 3.24 4.08 1.73 2.32 2.49 2.17 3.64 4.82 2.00 3.48 4.98
Average dose enhancement factor
Activation media
Ag 1.59 2.08 2.30 1.12 1.24 1.31 1.88 2.65 3.03 2.01 3.41 4.69
I 1.29 1.64 1.91 1.16 1.30 1.35 1.40 1.88 2.26 2.14 3.69 5.11
Pt 1.74 2.29 2.45 1.35 1.33 1.10 1.91 2.75 3.17 1.94 3.26 4.49
Au 1.76 2.31 2.46 1.35 1.32 1.09 1.94 2.79 3.20 1.97 3.32 4.56
The tumour was a 1.5 radius sphere with the source located at its center
308 Australas Phys Eng Sci Med (2013) 36:301–311
123
higher average energy at steeper distances and because of
inverse dependence of probability of photoelectric effect
with photon energy, the DEF decreases with distance in the
tumour. In the present study, we have only calculated DEF
in a transverse plane relative to longitudinal axes of the
sources. In this case, the angle theta will be equal to 908. It
is predicted that DEF may show angular dependence. This
effect was not evaluated here but can be a subject for
further studies in this field. It should be noticed from the
data in Fig. 3 that while the dose rate was plotted for dose
rate of 0.00–18.00 cGy mCi-1 h-1 range on vertical axis
for 125I, 103Pd and 131Cs sources, the dose rate range on
vertical axis for hypothetical 170Tm source is
0.00–1.00 cGy mCi-1 h-1. The reason is that for the same
activity and time for the four sources, the dose rate values
for 170Tm source were considerably less than those for
other sources. This can be related to the low photon yields
for 170Tm radionuclide. This figure was only plotted for
Au, but it is predicted that similar trends will be seen for
other media. Dose rate value in tumour is higher for higher
concentrations for all four sources. Higher dose enhance-
ment by 170Tm source is an advantage of this source over
the other ones, but when comparing dose rate values, this
source shows lower dose rates and thus this effect can be a
disadvantage of this source compared to 125I, 103Pd and131Cs sources. In other words, with the same activity and
time used for the four sources, 125I, 103Pd and 131Cs sources
shows higher dose rate values in presence of the activation
media. This can be an advantage of these sources when
compared with 170Tm.
The results of dose enhancement factors outside the
tumour, which are less than unity, indicate that the absor-
bed dose is reduced in the tissue outside the tumour when
the tumour is loaded by the activation media. The dose
reduction amounts to a value of 96 % in some cases (DEF
of 0.04). This is because of the shielding effect of the
activation media in the tumour which absorbs the source
radiation, thereby reducing the number of photons that
escape the tumour. This fact is considered as an advantage
of photon activation therapy, since the presence of acti-
vation media increases the dose inside the tumour, and
reduces the dose to tissues outside the tumour. However,
this effect has been studied in the present work only for the
points on a transverse line crossing the tumour. Calculation
Fig. 3 Dose enhancement factor and dose rate (cGy mCi-1 h-1) for 7, 18, 30 mg/ml concentrations of Au: a 125I source, b 103Pd source, c 131Cs
source, d hypothetical 170Tm source
Australas Phys Eng Sci Med (2013) 36:301–311 309
123
of the dose reduction on other parts of the tissues outside
tumour is a subject of future study. Furthermore, in a real
situation the activation media will be absorbed in some
extent by the normal tissues. This will result to enhance-
ment of dose in these tissues. In fact the dose in normal
tissues will be affected by both the shielding effect of
tumour and the dose enhancement in these normal tissues.
The 170Tm source has a longer half-life (128.6 days)
compared to the other sources. This may be an advantage in
photon activation therapy because a source with longer half-
life will need to fewer source exchanges, calibrations and
commissioning. Less source exchanges will be more cost-
effective for brachytherapy departments. On the other hand,
considering the low yield values of useful photons with 170Tm
source (Table 3), it will be a need for a source with higher
activity to compensate the low yields of this radioisotope.
As a future study, obtaining the DEF values for 145Sm,241Am and 161Ho sources and the four activation media of
Ag, I, Pt and Au will be useful. However, it should be
noticed that some of these radionuclides are not currently
available as standard brachytherapy sources. In the present
study, only physical effect (DEF) of presence of activation
media in tumour was evaluated in photon activation therapy
with various brachytherapy sources. Since the biological
response of the cells for a specific source and activation
media differs from the physical response (DEF), it is also
suggested to evaluate the response of the human cells to
various activation media and brachytherapy source combi-
nations in future studies on photon activation therapy.
Acknowledgments The authors are thankful to Dr. Facundo Bal-
lester for his help to review the manuscript. The authors would also
like to thank North Khorasan University of Medical Sciences for
funding this work.
References
1. Laster BH, Dixon DW, Novick S, Feldman JP, Seror V, Goldbart
ZI et al (2009) Photon activation therapy and brachytherapy.
Brachytherapy 8(3):324–330
2. Young LA, Kalet IJ, Rasey JS, Nelson JA (1998) 125I brachy-
therapy k-edge dose enhancement with AgTPPS4. Med Phys
25(5):709–718
3. Sastry KSR (1992) Biological effects of the Auger emitter iodine-
125: a review. Report No. 1 of AAPM Nuclear Medicine Task
Group No. 6. Med Phys 19(6):1361–1370
4. Corde S, Biston MC, Elleaume H, Esteve F, Charvet AM, Joubert
A et al (2002) Lack of cell death enhancement after irradiation
with monochromatic synchrotron X-rays at the K-shell edge of
platinum incorporated in living SQ20B human cells as cis-di-
amminedichloroplatinum (II). Radiat Res 158(6):763–770
5. Ferrer S (2005) Radiation therapy with synchrotron radiation and
cisplatin-based chemotherapy as a treatment of gliomas. Med
Clin (Barc) 124(7):271–273
6. Laster BH, Thomlinson WC, Fairchild RG (1993) Photon acti-
vation of iododeoxyuridine: biological efficacy of Auger elec-
trons. Radiat Res 133(2):219–224
7. Suortti P, Thomlinson W (2003) Medical applications of syn-
chrotron radiation. Phys Med Biol 48(13):1–35
8. Nath R, Bongiorni P, Rossi PI, Rockwell S (1990) Enhanced
IUdR radiosensitization by 241Am photons relative to 226Ra and125I photons at 0.72 Gy/h. Int J Radiat Oncol Biol Phys
18(6):1377–1385
9. Ranjbar H, Shamsaei M, Ghasemi MR (2010) Investigation of the
dose enhancement factor of high intensity low mono-energetic
X-ray radiation with labeled tissues by gold nanoparticles. Nuk-
leonika 55(3):307–312
10. Karnas SJ, Moiseenko VV, Yu E, Truong P, Battista JJ (2001)
Monte Carlo simulations and measurement of DNA damage from
X-ray-triggered auger cascades in iododeoxyuridine (IUdR).
Radiat Environ Biophys 40(3):199–206
11. Moiseenko VV, Karnas SJ, Yu E, Battista JJ (2002) Monte Carlo
simulations of DNA damage from incorporated cold iodine fol-
lowing photoelectrically induced Auger electron cascades. Radiat
Prot Dosimetry 99(1–4):113–116
12. Stephens BJ (2010) 161Ho?IUdR: optimized photon activation
therapy. Ph. D. Thesis in Physics, http://etd.library.vanderbilt.
edu/available/etd-05122010-153611/unrestricted/Stephens_Bryan_
dissertation.pdf, Accessed Sept 10, 2012, Vanderbilt University
13. Young LA, Phillips MH, Nelson JA (1999) Validation of K-edge125I brachytherapy enhancement with silver compounds. Phys
Med Biol 44(8):1921–1935
14. Meigooni AS, Hayes JL, Zhang H, Sowards K (2002) Experi-
mental and theoretical determination of dosimetric characteristics
of IsoAid ADVANTAGE 125I brachytherapy source. Med Phys
29(9):2152–2158
15. Meigooni AS, Dini SA, Awan SB, Dou K, Koona RA (2006)
Theoretical and experimental determination of dosimetric char-
acteristics for ADVANTAGE Pd-103 brachytherapy source. Appl
Radiat Isot 64(8):881–887
16. Murphy MK, Piper RK, Greenwood LR, Mitch MG, Lamperti PJ,
Seltzer SM et al (2004) Evaluation of the new cesium-131 seed for
use in low-energy X-ray brachytherapy. Med Phys 31(6):1529–1538
17. Ballester F, Granero D, Perez-Calatayud J, Venselaar JL, Rivard
MJ (2010) Study of encapsulated 170Tm sources for their poten-
tial use in brachytherapy. Med Phys 37(4):1629–1637
18. http://www.physics.carleton.ca/clrp/seed_database/I125/Advantage_
IA1-125A, Accessed September 10, 2012, Carleton University
19. http://www.physics.carleton.ca/clrp/seed_database/Pd103/
Advantage_IAPd_103A, Accessed September 10, 2012, Carleton
University
20. Granero D, Perez-Calatayud J, Casal E, Ballester F, Venselaar J
(2006) A dosimetric study on the Ir-192 high dose rate flexi-
source. Med Phys 33(12):4578–4582
21. Young LA, Phillips MH, Nelson JA (1999) Validation of K-edge125I brachytherapy enhancement with silver compounds. Phys
Med Biol 44(8):1921–1935
22. University of Washington. http://skuld.bmsc.washington.edu/
scatter/AS_periodic.html, Accessed September 10, 2012
23. Rivard MJ, Coursey BM, DeWerd LA, Hanson WF et al (2004)
Update of AAPM Task Group No. 43 Report: a revised AAPM
protocol for brachytherapy dose calculations. Med Phys
31(3):633–674
24. Bahreyni Toossi MT, Ghorbani M, Mowlavi AA, Soleimani
Meigooni A (2012) Dosimetric characterizations of GZP6 60Co
high dose rate brachytherapy sources: application of superimpo-
sition method. Radiol Oncol 46(2):170–178
25. Slate LJ, Elson HR, Lamba MA, Kassing WM, Soldano M,
Barrett WL (2004) A Monte Carlo brachytherapy study for dose
distribution prediction in an inhomogeneous medium. Med Do-
sim 29(4):271–278
26. Mowlavi AA, Binesh A, Moslehitabar H (2006) Dose distribution
and dosimetry parameters calculation of MED3633 palladium-103
310 Australas Phys Eng Sci Med (2013) 36:301–311
123
source in water phantom using MCNP. Iran J Radiat Res 4(1):
15–119
27. Waters LS, MCNPX User’s Manual, Version 2.4.0., Report LA-
CP-02-408 2000, Los Alamos National Laboratory
28. NIST. http://physics.nist.gov/PhysRefData/XrayMassCoef/
ComTab/water.html, Accessed Sept 10, 2012
29. Rivard MJ (2001) A discretized approach to determining TG-43
brachytherapy dosimetry parameters: case study using Monte
Carlo calculations for the MED3633 103Pd source. Appl Radiat
Isot 55(6):775–782
30. Rivard MJ (2007) Brachytherapy dosimetry parameters calcu-
lated for a 131Cs source. Med Phys 34(2):754–762
31. ICRU (1989) Tissue substitutes in radiation dosimetry and mea-
surement. ICRU Report no. 44. International Commission on
Radiation Units and Measurements, Bethesda, MD
32. Sowards KT (2007) Monte Carlo dosimetric characterization of
the IsoAid ADVANTAGE 103Pd brachytherapy source. J Appl
Clin Med Phys 8(2):18–25
Australas Phys Eng Sci Med (2013) 36:301–311 311
123