Upload
takashi-suzuki
View
213
Download
0
Embed Size (px)
Citation preview
Nuclear Instruments and Methods in Physics Research B 223–224 (2004) 87–91
www.elsevier.com/locate/nimb
Performance of the new iodine-129 beamline at JAERI-AMS
Takashi Suzuki *, Takafumi Aramaki, Toshikatsu Kitamura, Orihiko Togawa
Marine Research Laboratory, Mutsu Establishment, Japan Atomic Energy Research Institute, Mutsu, Aomori 035-0064, Japan
Abstract
Anticipating the application of 129I as an oceanographic tracer in marine environment studies in the North Pacific, a
new beamline has been set up at the AMS facility of the Japan Atomic Energy Research Institute for the measurement
of iodine isotopic ratios (129I/127I). It is demonstrated that iodine isotopic ratios can be measured with a precision below
2%, and that the detection limit is substantially better than 10�13. The method of preparing AgI sample material from
seawater is described.
� 2004 Elsevier B.V. All rights reserved.
PACS: 82.80.Ms; 07.75.+h
Keywords: Accelerator mass spectrometry; AMS; Iodine-129; Seawater; Time-of-flight; TOF
1. Introduction
A new beamline has been set up at the Accel-
erator Mass Spectrometry (AMS) facility of the
Japan Atomic Energy Research Institute (JAERI-
AMS). The AMS system, manufactured by High
Voltage Engineering Europa (HVEE), features a3 MV Tandetron and two independent beamlines
[1]. One is used for high-precision 14C-AMS [2] and
the other has been added for AMS with a variety of
other long-lived radiometric nuclides, particularly129I. In the following, we concentrate on129I-AMS
and refer to the associated system components as
the iodine beamline. In Section 2, we give an over-
view of this beamline, repeating the description and
* Corresponding author. Tel.: +81-175-28-2616; fax: +81-
175-22-4213.
E-mail address: [email protected] (T. Suzuki).
0168-583X/$ - see front matter � 2004 Elsevier B.V. All rights reser
doi:10.1016/j.nimb.2004.04.021
the 129I/127I acceptance test given in [3]. In Section
3, we discuss the performance of this beamline
(reproducibility, precision and detection limit). In
Section 4, we present our chemistry method for the
extraction of iodine from seawater and subsequent
preparation of ion source target material. We con-
clude with a brief summary and research outlook ofour laboratory.
2. Description of the iodine line
A diagram of the JAERI-AMS system is shown
in Fig. 1. The ion source is a HVEE model 846B
Cs sputtering source for negative ions and has acarousel which can hold up to 59 target holders.
Samples are inserted and measured under com-
puter control. A combination of 54� electro-
static analyzer (calculated energy resolution: 400)
with a 90� bouncer magnet (calculated mass
ved.
TOF detector
65o ESA
115o MagnetTandem accelerator90o Bouncer Magnet
54o ESA
Ion Source
127I
129I
Start MCPCarbon foil
Stop MCP
Q-snout
Q-pole
Stripper Canal
Iodine lineCarbon line
Stop plate
Slit
Fig. 1. The whole JAERI-AMS system. Bold lines denote the new beam line which is optimized to measure iodine isotopic ratio (129I/127I). Dotted lines denote 14C beam line.
88 T. Suzuki et al. / Nucl. Instr. and Meth. in Phys. Res. B 223–224 (2004) 87–91
resolution: 400) provides mass analysis with dou-
ble-focussing properties. Injection of specific
isotope beams into the accelerator is donesequentially by high-voltage biasing the magnet
chamber. Injection times for 129I and 127I are cho-
sen to be 8 and 2 ms, respectively. A Faraday,
located on the image point of the bouncer magnet,
allow the measurement of the stable isotope (127I)
intensity during injection of the radioisotope (129I)
into the accelerator. A full description of the
Tandetron accelerator, recirculating gas stripper,and associated beam optics can be found elsewhere
[4]. The maximum terminal voltage is �3 MV. The
measurement condition at the terminal is 2.5 MV
and charge state is selected 5+ because of no in-
terferencing peak of molecular fragments [5].
In the high-energy section, mass analysis is
performed by a 115� magnet (q ¼ 1200 mm,
m=Dm ¼ 2000). The 127I current is measured by anoff-axis Faraday cup located near the magnet’s
image point. Next, a 65 � electrostatic analyzer
(ESA) (q ¼ 1700 mm, E=DE ¼ 1000) provides the
required dispersion to separate out isotopic inter-
ferences (such as charge-exchanged 127I) from 129I.
Two electrostatic quadrupole lenses, located near
the ESA’s entrance and exit, ensure proper focus-
ing into the detector.
Particle detection and identification of 129I isperformed via time-of-flight (TOF). A detailed
layout of the TOF system is shown in Fig. 2. With
each incoming ion passing through a carbon foil
(20 lg/cm2) secondary electrons are emitted. These
electrons are directed by a permanent magnet onto
a micro channel plate (MCP: Comstock model CP-
640C/50F) for the generation of the start signal.
Subsequently, the ions are stopped on a sphericalaluminum stop plate with a diameter of 400 mm
located 1.5 m behind the carbon foil. The emitted
secondary electrons on the stop plate are directed
to the stop MCP, to create the stop signal. A
magnetic lens is implemented to focus the electrons
on the stop MCP. This lens is placed outside of the
TOF detector and can be adjusted in position and
rotation for optimal detection efficiency. Radio-isotope is discriminated from remaining unwanted
particles by the flight time from start signal to stop
signal. The time resolution of this TOF system is 1
ns. The difference between the expected flight time
of 129I and 127I is 5 ns under the condition that the
momentum is same.
ionsSecondary electrons
Stop MCP
Start MCPStart foilMagnetic lens
Stop plate
Grid
Stop bend magnet
Start bend magnet
(Carbon foil)
Slit
Fig. 2. Principal layout of TOF detector. Bold lines denote incoming ions. Dotted lines denote secondary electrons which are emitted
from carbon foil and stop plate.
8.00
7.80
7.60
7.40
7.20
7.00
Target #1 Target #2 Target #3 Target #4 Average
129 I
/127
I (x
10-1
1 )
1 2 3 4 5 1 2 3 4 5 1 2 3 4 5 1 2 3 4 5
Run Number
Fig. 3. Measured 129I/127I ratios form the acceptance test on 17
July 2000. Samples (AgI) were prepared at IsoTrace Labora-
tory. In target #1–5 cells, error bars attached to all measured
points are calculated from counting statistics. In the average
cell, the large error bar (1r) was determined by all the mea-
surements and the small error bar (1r) gave the reproducibility(uncertainty) in the mean sample value.
T. Suzuki et al. / Nucl. Instr. and Meth. in Phys. Res. B 223–224 (2004) 87–91 89
3. Performance
3.1. Reproducibility and precision
The acceptance test for the iodine line was car-
ried out on two consecutive days (July 17 and 18,
2000). AgI (129I/127I ¼ 1.1 · 10�10) standard sample
was prepared by IsoTrace (University of Toronto).
After sputter cleaning for 5 to 10 min., four sam-ples were measured during several runs. One run
(about 10 min) was injected 60,000 times for 129I
and 127I. The resulting machine ratios on the first
day are shown in Fig. 3. A reproducibility was
calculated from the mean value per target and a
precision was calculated from 20 measurements (4
targets· 5 runs). On the first day, the resulting ratio
and standard deviation of the reproducibility andthe precision are (7.50± 0.12) · 10�11 (1.6%) and
(7.50 ± 0.15) · 10�11 (2.0%), respectively. On the
second day, they are (7.21± 0.05) · 10�11 (0.7%)
and (7.21 0.11) · 10�11 (1.5%), respectively [3]. This
demonstrates that this new 129I beamline has the
good reproducibility and precision. The difference
between standard and measured machine ratio
suggests that the transmission efficiency from the115� magnet to the TOF detector is about 70%,
which is probably dominated by the efficiency for
electron detection of MCP in TOF detector.
3.2. Detection limit
The detection limit of the new beamline for 129I/127I measurements was investigated with commer-cially available silver iodide (ACROS Co. Ltd.
lot# 13564/1), instead of the Woodward iodine
used elsewhere [6,7]. The measured ratio, uncor-
rected for any detector background or processingblank, is 2.3 · 10�13 which was normalized to the
IsoTrace standard and a TOF spectrum from these
measurements is shown in Fig. 4. This order is
consistent with 129I/127I values reported by other
AMS laboratories on samples that they processed
from commercially available potassium iodide
[6,8]. Since our measured TOF-detector spectrum
does not exhibit a background underneath the 129Ipeak, we conclude that the detection limit must be
100
80
60
40
20
0590 595 600 605 610 615 620
Channel
Cou
nts
5 ns
Fig. 4. TOF spectrum on commercial AgI material. One
channel corresponds to 1 ns. Measurement time to obtain this
spectrum is about 50 min.
Seawater Sample
I-+IO3-
Iodine Carrier (2mg)
CCl4NaNO2
Organic Phase(CCl4)
H2SO4NaHSO3
Inorganic Phase(Water)
NH3
AgNO3
Organic Phase(CCl4)
repeat (x5-6)
Inorganic Phase(Seawater)
I-
I2
AgI
I-
H2SO4+NaHSO3
Fig. 5. Flow chart of extraction method from seawater (see text
for details).
90 T. Suzuki et al. / Nucl. Instr. and Meth. in Phys. Res. B 223–224 (2004) 87–91
at least one order of magnitude lower than the
measured ratio. In the near future, this measure-
ment will be repeated using the Woodward iodine
instead, in order to compare with previously re-ported values in the range (4–9) · 10�14 [6,7].
3.3. Status
After the acceptance test, we encountered a
variety of problems with the new beamline. The
first was a crack in the stop MCP plate of the TOF
detector, and the second fault was an unexpectedclosing of the slit located in front of the carbon
foil. After repairing the slit from the outside and
replacing the start and stop MCP plates with new
ones, 129I was observed again. Despite this,
counting rates of the TOF detector have consid-
erably decreased some point in time after the
acceptance test. A probable cause is an earthquake
having affected the alignment of the high-energysection beamline behind the analysing magnet,
particularly that of the TOF detector. It is ex-
pected that realignment of that part of the beam-
line will solve this problem.
4. Sample preparation
Preparation of 129I-AMS targets from seawater
is performed by an extraction method [9], see the
flow chart given in Fig. 5. First, iodine carrier (2
mg) is added to the sample, about 1 L of seawater.
Iodate is reduced to iodide with sulfuric acid and
sodium hydrogen sulfite, and the iodide is oxidized
to I2 by the addition of sodium nitrite. The I2 isextracted with carbon tetrachloride and then back-
extracted into sodium bisulfite solution by reduc-
tion of I2. These extraction and back-extraction
steps are repeated until the color of carbon tetra-
chloride disappears with concentrating and puri-
fying the iodine. This usually requires repeating
this step 5–6 times. The concentrated iodine is
precipitated as silver iodide with silver nitrate andammonia in a dark room. The silver iodide is
rinsed by ammonia and distilled water. After
drying, the AgI is mixed with Nb (AgI:Nb¼ 1:2.5)
and pressed in a target holder.
T. Suzuki et al. / Nucl. Instr. and Meth. in Phys. Res. B 223–224 (2004) 87–91 91
5. Summary and outlook
The new beamline was optimized to measure129I/127I isotopic ratios, featuring sequential injec-
tion into the 3 MV Tandetron and particle detec-
tion by time-of-flight. Iodine isotopic ratios can be
measured with a precision better than 2%, and
inspection of the TOF detector spectrum suggests
that the detection limit is well below 10�13. The
high-energy section of this new beamline has had a
few problems since the acceptance tests, most ofwhich have been solved. It is expected that
realignment of that section will recover some or all
of the transmission that was lossed probably due
to an earthquake. A method has been described to
prepare 129I-AMS targets from seawater samples.
Two nuclear fuel reprocessing plants, at La
Hague (France) and Sellafield (England), have
released large quantities of 129I into seawater. Ithas been used as an oceanographic tracer in the
north Atlantic and Arctic oceans [9,10]. In Japan,
a new nuclear fuel reprocessing plant will be
operated at Rokkasho in the near future. We in-
tend to investigate the potential of using 129I as an
oceanographic tracer in the North Pacific Ocean,
using the emission of 129I from this Rokkasho
facility. Past dumping of radioactive waste in theJapan Sea by the Russian Federation and the
former Soviet Union [11] may complicate a
description of the 129I tracer source function in
that area. Sampling prior to the anticipated release
by the Rokkasho facility is expected to provide a
proper baseline. Hence, we consider the arrival ofthe Rokkasho plant as an opportunity and our
high-sensitivity 129I-AMS beamline as the right
tool to mount new marine environmental research.
References
[1] A. Gottdang, D.J.W. Mous, Nucl. Instr. and Meth. B 123
(1997) 163.
[2] T. Aramaki et al., Nucl. Instr. and Meth. B 172 (2000) 18.
[3] A. Gottdang et al., Appl. Accelerators Res. Ind. AIP CP
576 (2000) 403.
[4] D.J.W. Mous, R. Koudijs, P. Dubbelman, H.A.P. van
Oosterhout, Nucl. Instr. and Meth. B 62 (1992) 421.
[5] L.R. Kilius, X.-L. Zhao, A.E. Litherland, K.H. Purser,
Nucl. Instr. and Meth. B 123 (1997) 10.
[6] E. Boaretto, D. Berkovits, G. Hollos, M. Paul, Nucl. Instr.
and Meth. B 50 (1990) 280.
[7] M.L. Roberts, M.W. Caffee, I.D. Proctor, Nucl. Instr. and
Meth. B 123 (1997) 367.
[8] S. Hatori, M. Ohseki, H. Nawata, H. Matsuzaki, T.
Misawa, K. Kobayashi, Nucl. Instr. and Meth. B 172
(2000) 299.
[9] J.N. Smith, K.M. Ellis, L.R. Kilius, Deep-Sea Res. I 45
(1998) 959.
[10] G.M. Raisbeck, F. Yiou, Z.Q. Zhou, L.R. Kilius, J.
Marine Syst. 6 (1995) 561.
[11] P.P. Povinec, B. Oregioni, A.J.T. Jull, W.E. Kieser, X.-L.
Zhao, Nucl. Instr. and Meth. B 172 (2000) 672.