Series 1000 and 2000 Transmitterswith MVD™ Technology

Product Data SheetPS-00400July 2001

Page 2: PDS_1000-2000 MVD_PS-00400-2001-07_EN

2 Series 1000 and 2000 Transmitters

Micro Motion® Series 1000 and 2000 Transmitterswith MVD™ TechnologyThe new approach tosensor electronics

Only Micro Motion® combines new MVD™

Technology with a modular architecture that

redefines sensor electronics. That means

multivariable digital processing that’s scalable

for any flow application. MVD Technology gets

your most basic – or most complex –

application up and running quicker, easier and

more cost effectively than ever before.

MVD Technology

MVD Technology makes your Micro Motion

flowmeter work smarter. Front-end digital

processing dramatically reduces signal noise

and gives you faster response time, compared

to analog devices. Innovative MVD Technology

also enables multiple variable measurement

and diagnostics never before possible. And this

is just the beginning.

Only MVD Technology allows you to:

• Measure multiple variables• Choose integral or remote mounting with

a standard twisted, shielded 4-wire signalcable

• Identify and resolve problems easily withbuilt-in smart diagnostics

• Choose transmitter capabilities based onyour application’s needs

• Upgrade transmitter functionalityas needed

What’s the bottom line ofMVD Technology?

Reducing costs in your bottom line through

improved process consistency and

maximized uptime.

Only four wires

Approved forhazardous areas

Scalable architecture

You asked for it and Micro Motion has

delivered. Series 1000 and 2000 transmitters

allow you to choose the functionality you want.

Series 1000 transmitters are perfect for

applications that require single variable

measurement. For more demanding

applications, Series 2000 transmitters measure

multiple variables simultaneously, and have

more output and digital communication options.

What happens when you put Micro Motion’sMVD Technology together with the Series1000 and 2000 transmitters?

Page 3: PDS_1000-2000 MVD_PS-00400-2001-07_EN

Series 1000 and 2000 Transmitters 3

Clean, noise-free,digital signalsthat improvemeasurementperformance

Delivering a suite of power-packedstandard features

All Series 1000 and 2000 transmitters offer:

• Class I, Division 1 / Zone 1 local operator

interface to:

• View process variables

• View meter status at a glance

• View and acknowledge alarms

• Start, stop, and reset transmitter totalizers

• Zero flowmeter

• Perform output simulation tests

• Change measurement units

• Assign variables to outputs

• Scale outputs

• Set RS-485 communications options

• Interface functions can be customized and

password protected

• Compact, integral mounting to sensor with

360 degrees of rotation

• Cost-effective, hassle-free, 4-wire remote

mounting to sensor

• Simple start-up with virtually no special

programming requirements

• Digital communications

• Easy to access diagnostics: meter status,

process issues, and more

Series 1000 flow measurementtransmitter

For applications requiring onlymass flow or volume flow measurement

Series 1000 transmitters are ideal for flow

applications where only a single variable is

needed at any given time. Series 1000

transmitters feature a milliamp and a

frequency/pulse output, and HART® or

Modbus® digital communications.

Series 1000 transmitters can output any oneof the following variables:

• Mass flow rate• Volume flow rate• Total flow

Series 2000 Multivariable Transmitter

For applications requiring simultaneousmonitoring of multiple flow variables

Series 2000 transmitters are designed

specifically for applications where multiple

variables are needed simultaneously.

Series 2000 transmitters feature a milliamp and

a frequency/pulse output, plus HART, Modbus,

FOUNDATION™ fieldbus, and Profibus PA

digital communications.

Series 2000 transmitters can simultaneouslyoutput:

• Mass flow rate• Volume flow rate• Total flow• Density• Temperature• Drive gain

Page 4: PDS_1000-2000 MVD_PS-00400-2001-07_EN

4 Series 1000 and 2000 Transmitters

Series 1000 and 2000 functional specifications

Electrical connectionsInput and output connections

Three pairs of wiring terminals for transmitter outputs.

Screw terminals accept one or two solid conductors, 14 to 12 AWG (2.5 to 4 mm2); or one or two stranded conductors, 22 to 14 AWG (0.34 to 2.5 mm2).

Power connectionOne pair of wiring terminals accepts either AC or DC power.

One internal ground lug for power-supply ground wiring.

Screw terminals accept one or two solid conductors, 14 to 12 AWG (2.5 to 4 mm2); or one or two stranded conductors, 22 to 14 AWG (0.34 to 2.5 mm2).

Service port connectionTwo clips for temporary connection to the service port.

Input/output signalsAll transmitters

One 4-wire sensor signal input connection with ground, intrinsically safe.

Non-Intrinsically Safe output HART/Modbus transmitters (output option code A)

One active 4-20 mA outputNot intrinsically safeIsolated to ±50 VDC from all other outputs and earth

groundMaximum load limit, 600 ohmsSeries 1000 can report mass flow or volume flow; Series

2000 can report mass flow, volume flow, density, temperature, or drive gain

Output is linear with process from 3.8 to 20.5 mA , per NAMUR NE43 (June 1994)

One active frequency/pulse outputNot intrinsically safeCan report mass flow or volume flow, which can be used

to indicate flow rate or totalFor Series 1000, output is dependent on mA output; for

Series 2000, output is independentScalable to 10,000 HzMaximum output of 30 VDC max., 24 VDC typicalInternal 2.2 kohm pull-up, sinking up to 500 mA at 30

VDC maximumOutput is linear with flow rate to 12,500 Hz

Intrinsically Safe transmitters (output option code D)One intrinsically safe passive 4–20mA output (two with Series 2000)

HART communication on one 4–20mA outputMaximum input voltage, 30 VDC, 1 watt maximumMaximum current 300 mAMaximum load limits, see chart belowSeries 1000 can report mass flow or volume flow; Series

2000 can report mass flow, volume flow, density, or temperature

Output is linear with process from 3.8 to 20.5 mA, per NAMUR NE43 (June, 1994)

One intrinsically safe passive frequency/pulse/discrete output

Maximum input voltage, 30 VDC, 0.75 watt maximumMaximum current 100 mAMaximum load limit, see chart belowCan report mass flow or volume flow, which can be used

to indicate flow rate or totalFor Series 1000, output is dependent on mA output; for

Series 2000, output is independentScalable to 10,000 HzOutput is linear with flow rate to 12,500 Hz

Fieldbus and Profibus PA transmittersOne FOUNDATION fieldbus H1 or Profibus PA output

FOUNDATION fieldbus and Profibus PA wiring is intrinsically safe with an intrinsically safe power supply

Manchester-encoded digital signal conforms to IEC 1158-2.

0

100

200

300

400

500

600

700

800

900

1000

12 14 16 18 20 22 24 26 28 30

mA Output Load Resistance ValueRmax = (Vsupply - 12)/0.0023*

*If communicating with HART a minimum of 250 ohms and 17.75V supply is needed

Supply voltage (volts)

Ext

erna

l res

isto

r (o

hms)

OperatingRegion

0

1000

2000

3000

4000

5000

6000

7000

8000

9000

10000

5 7 9 11 13 15 17 19 21 23 25 27 29

Frequency Output Load Resistance ValueRmax = (Vsupply - 4)/0.003

*Rmin = (Vsupply - 25)/0.006*Absolute minimum = 100 ohms for Vsupply < 25.6 volts

Supply voltage (volts)

Ext

erna

l res

isto

r (o

hms)

OperatingRegion

Page 5: PDS_1000-2000 MVD_PS-00400-2001-07_EN

Series 1000 and 2000 Transmitters 5

Series 1000 and 2000 functional specifications continued

Digital communicationsAll transmitters

One service port can be used for temporary connection only.

Uses RS-485 Modbus signal, baud rate of 38.4 kilobaud, one stop bit, no parity

HART or Intrinsically Safe HART output (output option codes A or D)

HART Bell 202 signal is superimposed on the primary milliamp output, and is available for host system interface.

Frequency 1.2 and 2.2 kHzAmplitude 0.8 V peak-to-peak1200 baudRequires 250 to 600 ohms load resistance

HART/Modbus transmitter (output option Code A)One RS-485 output can be used for direct connection to a HART or Modbus host system. Accepts baud rates between 1200 baud and 38.4 kilobaud.

Fieldbus transmittersTransmitters are registered with the Fieldbus Foundation, and conform to the Foundation fieldbus H1 protocol specification.

Input frequency from sensorMass flow 20 HzVolume flow 20 HzDensity 10 HzTemperature1 Hz

Analog Input Function BlocksCycle time host dependentUpdate rate 50 millisecondsRefresh ratehost dependent

Power supplySelf-switching AC/DC input, automatically recognizes supply voltage.

Complies with low voltage directive 73/23/EEC per IEC 1010-1 with amendment 2

Installation (Overvoltage) Category II, Pollution Degree 2

The transmitter fieldbus circuit is passive, and draws its power from the fieldbus segment. Current draw from the fieldbus segment is 11.5 mA.

AC power85 to 265 VAC, 50/60 Hz, 5 watts typical, 8 watts maximum

DC power18 to 100 VDC, 5 watts typical, 8 watts maximumMinimum 22 VDC with 1000 feet of 18 AWG (300 meters of 0.8mm2) power-supply cable

FuseIEC 127-1.25 fuse, slowblow

Environmental limitsAmbient temperature limits

Operating -40 to 140°F (-40 to 60°C)Storage -40 to 140°F (-40 to 60°C)

Note: Display responsiveness decreases, and display may become difficult to read, below -4°F (-20°C). Above 131°F (55°C), some darkening of display might occur.

Humidity limits5 to 95% relative humidity, non-condensing at 140°F (60°C)

Vibration limitsMeets IEC 68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1.0 g.

Environmental effectsEMI effects

Series 1000 and 2000 transmitters (except IS I/O; option code D) conform to NAMUR NE21 (August 1998 German and May 1999 English).

Series 1000 and 2000 transmitters meet EMC directive 89/336/EEC per EN 50081-2 (August 1993) and EN 50082-2 (March 1995), and EN 61326 Industrial.

Ambient temperature effectOn analog outputs±0.005% of span per °C

Hazardous area classificationUL and CSATransmitter

Class I, Div. 1, Groups C and D. Class II, Div. 1, Groups E, F, and G explosion proof (when installed with approved conduit seals). Otherwise, Class I, Div. 2, Groups A, B, C, and D.

OutputsProvides nonincendive sensor outputs for use in Class I, Div. 2, Groups A, B, C, and D; or intrinsically safe sensor outputs for use in Class I, Div. 1, Groups C and D or Class II, Div. 1, Groups E, F, and G.

CENELECHART/Modbus transmitters

Flameproof when installed with approved cable glands:with display EEx d [ib] IIB+H2 T5without display EEx d [ib] IIC T5

Increased safety when installed with approved cable glands:

with display EEx de [ib] IIB+H2 T5without display EEx de [ib] IIC T5

Fieldbus, Profibus PA, and IS output transmittersFlameproof when installed with approved cable glands:

with display EEx d [ia/ib] IIB+H2 T5without display EEx d [ia/ib] IIC T5

Increased safety when installed with approved cable glands:

with display EEx de [ia/ib] IIB+H2 T5without display EEx de [ia/ib] IIC T5

Page 6: PDS_1000-2000 MVD_PS-00400-2001-07_EN

6 Series 1000 and 2000 Transmitters

Series 2000 with FOUNDATION™ fieldbus

Series 2000 fieldbus software functionalitySeries 2000 Foundation fieldbus software is designed to permit remote testing and configuration of the transmitter using the DeltaV™ Fieldbus Configuration Tool, or other Foundation fieldbus compliant hosts. The illustration below shows how the Coriolis sensor signal is channelled through the flowmeter to the control room and the Foundation fieldbus configuration device.

Transducer blockThe transducer block holds the data from the Coriolis sensor. It includes information about the sensor type, sensor configuration, engineering units, calibration, damping, and diagnostics.

Resource blockThe resource block contains physical device information, including available memory, manufacturer identification, type of device, and features.

Function blocksThe Analog Input (AI) function block processes the measurement and makes it available to other function blocks. It also allows filtering, alarming, and engineering unit change. The four Series 2000 AI blocks process mass flow, volume flow, density, and temperature signals from the Coriolis sensor.

Integrator blockThe integrator block provides functionality for the transmitter totalizers. The flow variable (mass or volume) can be selected and reset.

Diagnostics and serviceSeries 2000 transmitters automatically perform continuous self diagnostics. Using the transducer block, the user can perform on-line testing of the transmitter and sensor. Diagnostics are event driven and do not require polling for access.

FOUNDATION fieldbus compliantcommunications stack

Resource blockPhysical device

information

Function blocksAI1, AI2, AI3, AI4

Integrator blockFlow totalizers

Transducer blockCoriolis sensor signal processing

(equations, diagnostics)

Block diagram for Series 2000 transmitters with FOUNDATION fieldbus

Page 7: PDS_1000-2000 MVD_PS-00400-2001-07_EN

Series 1000 and 2000 Transmitters 7

Series 1000/2000 performance specifications

Sensor Process Variable Accuracy Repeatability

ELITE flow(1)

liquids and slurriesgas(2)

(1)Flow accuracy includes the combined effects of repeatability, linearlity, and hysteresis. All specifications for liquids are based on reference conditions of 68 to 77°F (20 to 25°C) and 15 to 30 psig. Specifications for gas are based on air at 60°F (15.5°C) and 1000 psig (70 bar), unless otherwise noted.

(2)CMF400 sensors are not rated for gas.

±0.10% of rate(3)

±0.35% of rate(3)

(3)When flow rate is less than [zero stability/(base accuracy percentage/100)], accuracy equals ±[(zero stability/flow rate) x 100]% of rate.

±0.05% of rate(4)

±0.20% of rate(4)

(4)When flow rate is less than [zero stability/(base accuracy percentage/100)], repeatability equals ±[½(zero stability/flow rate) x 100]% of rate.

densityCMF010Pothers

±0.002 g/cc (±2.0 kg/m3) ±0.0005 g/cc (±0.5 kg/m3)

±0.001 g/cc (±1.0 kg/m3)±0.0002 g/cc (±2.0 kg/m3)

temperature ±1°C ± 0.5% of reading in °C ±0.2°C

T-Series flow (liquid only)(1) ±0.15% of rate(3) ±0.05% of rate(4)

density (liquid) ±0.002 g/cc (±2.0 kg/m3) ±0.001 g/cc (±1.0 kg/m3)

temperature ±1°C ± 0.5% of reading in °C ±0.2°C

F-Series flow(1)

liquids and slurriesgas

±0.20% of rate(3)

±0.70% of rate(3)

±0.10% of rate(4)

±0.35% of rate(4)

density (liquid) ±0.002 g/cc (±2.0 kg/m3) ±0.001 g/cc (±1.0 kg/m3)

temperature ±1°C ± 0.5% of reading in °C ±0.2°C

R-Series flow(1)

liquidgas(5)

(5)R200 sensors are not rated for gas.

±0.5% of rate(3)

±1.0% of rate(3)

±0.25% of rate(4)

±0.50% of rate(4)

density not rated for density

temperature ±1°C ± 0.5% of reading in °C ±0.2°C

Page 8: PDS_1000-2000 MVD_PS-00400-2001-07_EN

8 Series 1000 and 2000 Transmitters

Model 1700 and 2700 physical specifications

Field-mount housingNEMA 4X (IP67) epoxy painted cast aluminum housing.

Terminal compartment contains output terminals, power terminals and service-port terminals. The output terminals are physically separated from the power- and service-port terminals.

The electronics compartment contains all electronics and the standard display.

The sensor compartment contains the wiring terminals for connection to the core processor on the sensor.

Screw-terminal on housing for chassis ground.

Cable gland entrances are either 1/2”-14 NPT or M20 x 1.5 female conduit ports.

MountingModel 1700 and 2700 field-mount transmitters are available integrally mounted to Micro Motion T-Series sensors, or in a remote-mount configuration.

Remote-mount transmitters include a mounting bracket, and require standard 4-wire twisted, shielded signal cable, up to 1000 feet (300 meters) in length, between the sensor and the transmitter. Hardware for installing the transmitter on the mounting bracket is included.

Remote-mount transmitters with a 9-wire signal cable between the sensor and the transmitter have a maximum cable length of 60 feet (20 meters). Hardware for installing the transmitter on the supplied mounting bracket is supplied.

The transmitter can be rotated on the sensor or the mounting bracket, 360 degrees, in 90-degree increments.

Interface/displaySegmented 2-line display with LCD screen with optical controls and flowmeter-status LED is standard. Suitable for hazardous area installation.

To facilitate various mounting orientations, the display can rotate on transmitter, 360 degrees, in 90-degree increments.

LCD line 1 lists the process variable, line 2 lists engineering unit of measure. Non-glare tempered glass lens.

Display controls feature optical switches that are operated through the glass with a red LED visual-feedback to confirm when a "button" is pressed.

Display functionsOperational

View process variables; start, stop, and reset totalizers; view and acknowledge alarms.

Off-line (specific functions where applicable, depending on option board)

Zero flowmeter, simulate outputs, change measurement units, configure outputs, and set RS-485 communications options.

Status lightThree-color LED status light on display panel indicates flowmeter condition at a glance. Green, yellow, or red, either continuously on or blinking, status light immediately indicates flowmeter status.

Weight4-wire remote mount transmitter 8 lb (3.6 kg)9-wire remote transmitter 13 lb (5.9 kg)

For weight of integrally mounted transmitter and sensor, refer to sensor specifications.

Page 9: PDS_1000-2000 MVD_PS-00400-2001-07_EN

Series 1000 and 2000 Transmitters 9

Model 1700 and 2700 physical specifications continued

Dimensions in inches(mm)

2 7/16(62)

ø4 7/8(124)

2 3/8(60)

(to 1/2 NPTor M20)

4 3/4(120)

4 13/16(122)

4 1/2(114)

6 7/8(174)

3 15/16(99)

2 11/16(69)

1 7/8(48)

1 3/4(45)

8 7/16(214)

wall mount

9 5/8(244)

to centerline of 2” pipepipe mount

4 5/16(110)

2 1/4(58)

1(26)

3x 1/2-14 NPTor M20 x 1.5

4-wire remote mount Series 1000/2000 transmitter

6 7/8(174)

3 15/16(99)

3x 1/2-14 NPTor M20 x 1.5

2 11/16(69)

9 5/8(244)

to centerline of 2” pipepipe mount

8 7/16(214)

wall mount

4 5/16(110)

4 3/8(111)

2 9/16(65)

13/16(21)

3 1/16(78)

7/8(22)

8 3/16(207)

3 11/16(93)2 13/16

(71)

4x ø3/8(10)

2 13/16(71)

9-wire remote Series 1000/2000 transmitter

For dimensions of integrally mounted transmitter and sensor, refer to sensor specifications.

Page 10: PDS_1000-2000 MVD_PS-00400-2001-07_EN

10 Series 1000 and 2000 Transmitters

Series 1000 model number matrixModel Product Description

1700 Micro Motion Coriolis MVD™ single variable flow transmitter

Code Mounting

R 4-wire remote mount transmitter

I Integral mount transmitter

C(1)

(1)9-wire remote transmitter with integral core processor available for new CENELEC applications only.

9-wire remote transmitter (requires J-box sensor)

Code Power

1 18 to 30 VDC and 85 to 265 VAC; self switching

Code Display

1 Dual line display for process variables and totalizer reset (standard)

3 No display

Code Output

A One mA; one frequency; RS485

D Intrinsically safe outputs: one mA, one frequency

Code Conduit Connections

B 1/2-inch NPT - no gland

C 1/2-inch NPT with brass nickel cable gland

D 1/2-inch NPT with stainless steel cable gland

E M20 - no gland

F M20 with brass nickel cable gland

G M20 with stainless steel cable gland

Code Approvals

M Micro Motion Standard (no approval)

U UL

C CSA (Canada only)

A CSA (US and Canada)

Z CENELEC - Increased safety

F CENELEC - Flameproof

Code Language

A Danish quick reference guide and English manual

D Dutch quick reference guide and English manual

E English quick reference guide and English manual

F French quick reference guide and French manual

G German quick reference guide and German manual

H Finnish quick reference guide and English manual

I Italian quick reference guide and English manual

J Japanese quick reference guide and English manual

M Chinese quick reference guide and English manual

N Norwegian quick reference guide and English manual

O Polish quick reference guide and English manual

P Portuguese quick reference guide and English manual

R Russian quick reference guide and English manual

S Spanish quick reference guide and Spanish manual

W Swedish quick reference guide and English manual

Code Software Options 1

ZZ Reserved for future use

Code Factory OptionsZ Standard product

X CEQ hardware

Typical Model Number: 1700 I 1 1 A D M E ZZ Z

Page 11: PDS_1000-2000 MVD_PS-00400-2001-07_EN

Series 1000 and 2000 Transmitters 11

Series 2000 model number matrixModel Product Description

2700 Micro Motion Coriolis MVD™ multivariable flow and density transmitter

Code Mounting

R 4-wire remote mount transmitter

I Integral mount transmitter

C(1)

(1)9-wire remote transmitter with integral core processor available for new CENELEC applications only.

9-wire remote transmitter (requires J-box sensor)

Code Power

1 18 to 30 VDC and 85 to 265 VAC; self switching

Code Display

1 Dual line display for process variables and totalizer reset (standard)

3 No display

Code Output

A One mA; one frequency; RS485

D Intrinsically safe outputs: two mA, one frequency

E FOUNDATION fieldbus H1

G Profibus PA

Code Conduit Connections

B 1/2-inch NPT - no gland

C 1/2-inch NPT with brass nickel cable gland

D 1/2-inch NPT with stainless steel cable gland

E M20 - no gland

F M20 with brass nickel cable gland

G M20 with stainless steel cable gland

Code Approvals

M Micro Motion Standard (no approval)

U UL

C CSA (Canada only)

A CSA (US and Canada)

Z CENELEC - Increased safety

F CENELEC - Flameproof

Code Language

A Danish quick reference guide and English manual

D Dutch quick reference guide and English manual

E English quick reference guide and English manual

F French quick reference guide and French manual

G German quick reference guide and German manual

H Finnish quick reference guide and English manual

I Italian quick reference guide and English manual

J Japanese quick reference guide and English manual

M Chinese quick reference guide and English manual

N Norwegian quick reference guide and English manual

O Polish quick reference guide and English manual

P Portuguese quick reference guide and English manual

R Russian quick reference guide and English manual

S Spanish quick reference guide and Spanish manual

W Swedish quick reference guide and English manual

Code Software Options 1

Z Flow & density variables (standard)

X CEQ software option 1

Code Software Options 2

Z No software options 2

A Standard fieldbus function blocks

X CEQ software option 2

Code Factory Options

Z Standard product

X CEQ hardware

Typical Model Number: 2700 I 1 1 A D M E Z Z Z

Page 12: PDS_1000-2000 MVD_PS-00400-2001-07_EN

© 2001, Micro Motion, Inc.All rights reservedPS-00400 (7/01)

Micro Motion Inc. USAWorldwide Headquarters7070 Winchester CircleBoulder, Colorado 80301Tel 303-530-8400

800-522-6277Fax 303-530-8459

Micro Motion Asia1 Pandan CrescentSingapore 128461Republic of SingaporeTel (65) 777-8211Fax (65) 770-8003

Micro Motion EuropeGroeneveldselaan 83903 AZ VeenendaalThe NetherlandsTel +31 (0) 318 549 549Fax +31 (0) 318 549 559

For the latest Micro Motion product specifications, view thePRODUCTS section of our Web site at www.micromotion.com

Due to Micro Motion's commitment to continuous improvement of our products, all specifications are subject to change without notice. Micro Motion is a registeredtrademark, and MVD and ProLink II are trademarks of Micro Motion, Inc., Boulder, Colorado. PlantWeb is a registered trademark, and Fisher-Rosemount andDeltaV are trademarks of Fisher-Rosemount, Clayton, Missouri. HART is a registered trademark of the HART Communication Foundation, Austin, Texas. Modbus isa registered trademark of Modicon, Inc., North Andover, Massachusetts. FOUNDATION is a trademark of the Fieldbus Foundation, Austin, Texas.

Micro Motion JapanShinagawa NF Bldg. 5F1-2-5, Higashi ShinagawaShinagawa-kuTokyo 140-0002 JapanTel (81) 3 5769-6803Fax (81) 3 5769-6843