Upload
nguyenhuong
View
223
Download
0
Embed Size (px)
Citation preview
PASSIVE SAMPLING FUNDAMENTALS
AND APPLICATION FOR MONITORING
VOC’S IN VAPOR INTRUSION
INVESTIGATIONS
2nd October 2014
Atlanta
Nicola Watson
Agenda
• Diffusive sampler theory
• Factors that influence sampler performance
• Sampling complex atmospheres
• Comparison studies
• Real World applications
Diffusive Sampler Theory
• Diffusion is a molecular transport property. It is the process by which matter progresses along a concentration gradient until there is an equalisation of concentration within a single phase.
• Thus, as with any chemical system, a diffusing gas is spontaneously adopting its most probable energy distribution in its quest for an even, maximum dispersion and thus maximum entropy.
• The rate of this migration property is measured by its flux (J), which is the quantity of matter passing through a reference surface area per unit time.
• As Jx is the component of a vector and as matter flows down the concentration gradient away from its source, the coefficient of proportionality, D (the diffusion coefficient) in the matter flux expression must be negative.
Fick’s 1st Law
dx
dCJ x
where:
Jx - rate of diffusion (molcm-2s-1);
dC/dx - concentration gradient (molcm-2).
D - diffusion coefficient (cm2s-1).
dx
dCDJ x
therefore,
Application of Fick's First Law to passive samplers
• A passive sampler is essentially a collection medium, either
– solid sorbent,
– liquid sorbent,
– or chemically impregnated inert support,
which is separated from the atmosphere of interest by a zone of still air.
Axial type samplers
The driving force for
matter flux across the
still-air gap is the
induced concentration
gradient formed
between the sampler
opening and the air-
adsorbent interface
where vapour phase
matter is scavenged.
Application of Fick's First Law to passive samplers
• Ambient concentration of the analyte at the surface of the monitor (Camb) i.e. does not take matter from its surrounding environment faster than it can be replaced
• Zero concentration of the analyte at the surface of the sorbent, i.e. the adsorbent is a zero sink and therefore there is no saturation of the adsorbent (Cads = 0)
• A linear concentration gradient between the two. steady state conditions always exist
Axial type samplers
For application of Fick's First Law to a diffusive sampler several simplifying
assumptions are necessary:
Calculation of ambient concentrations
When these conditions apply, Fick’s First Law of Diffusion applies, and
analytes will migrate to the surface of the sorbent (uptake rate Uideal) at a rate
that is dependent on:
• the distance between the sorbent surface and the monitor surface (L)
• the cross-sectional area of the sampler (A)
• the diffusion coefficient of the analyte through air (D)
Knowing the,
• the time of exposure (t)
• the mass collected on tube (mt)
Will allow the ambient concentration (Camb) to be calculated
tU
mC
ideal
tamb
L
DAU ideal
Application of Fick's First Law to passive samplers
• The previously derived expression for the quantity of mass sampled by a
passive device holds only for axially configured devices.
• The simplified equation, still holds But Uideal is now,
Radial type samplers
tU
mC
ideal
tamb
)/(
2
ao
idealrrIn
hDU
Factors that influence sampler performance
The uptake rate is directly proportional to adsorbent bed surface area and
inversely proportional to diffusion path length. Commercially available passive
samplers fall into two main categories
• low uptake rate tube-type devices,
• high uptake rate badge-type devices.
Sample geometry
Factors that influence sampler performance
• Temperature
– A 20°C difference in temperature represents a five percent difference in sampled mass.
– However increased temperature does not lead to faster matter transfer and thus a greater quantity of analyte being up taken, but the opposite, which is why sampling devices should be protected from heat sources, such as direct sunlight, in the field.
• Relative humidity
– This is more influential when using polar adsorbents. At high RH, water molecules may preferentially occupy adsorbent sites, leading to premature saturation. Secondly,
– When the compounds of interest are water sensitive, hydrolysis or solution may take place on the adsorbent surface.
– When sampling for prolonged durations from ambient atmospheres, where high relative humidities are commonly encountered, a hydrophobic adsorbent is preferable.
Environmental factors
Factors that influence sampler performance
• Zone A: Represents stagnant, or slowly moving air. The main process by which this
matter is replaced is by diffusion and it has been shown that under these conditions
the replenishment rate is too low.
• Zone C: At the other end of the scale there is non-ideal behaviour through excessive
face velocity. Here due to turbulence in the diffusion air gap, the effective diffusion
path length is decreased and consequently the uptake rate increases.
Environmental factors - Face velocity (air speed)
Recommended minimum face
air velocities are:
Badge type samplers 0.05 to
0.10 ms-1.
Axial tube type samplers
0.001 ms-1.
Factors that influence sampler performance
A poor choice of adsorbent can lead to unpredictable sampler behaviour and unusable results that no amount of subsequent calculation or careful
laboratory procedure can correct.
Considerations when selecting a sorbent/compound pair:
• Ensure that the sorbent is strong enough for the compound(s) of interest and also for the duration of the sampling event i.e. the sorbent must behave as a zero sink. This ensures a constant sampling rate and no loss of compounds prior to analysis.
• If under extreme adsorbent loading conditions
– extended exposure to high concentration atmospheres
– misuse of high uptake rate devices
– poor matching of an adsorbent to the target species
a build-up in the vapour phase concentration at the adsorbent surface may be observed. This can lead to deviation from theoretical uptake rates and even sample loss back to the atmosphere
Sorbent selection
Factors that influence sampler performance
• For air monitoring situations, where concentrations of targets are typically high, the most critical factor governing the suitability of an adsorbent for a given compound is its adsorption capacity. An adsorbent’s capacity may be determined from its isotherm.
Sorbent selection – high concentration atmospheres
Isotherm "a" is favourable for
passive sampling.
The linear isotherm and
isotherm "b" are unfavourable
as they predict a rapid build
up in the vapour phase
concentration after only a
minimal uptake of adsorbate.
Factors that influence sampler performance
• When monitoring low concentrations in ambient atmospheres, sampler
blanks and adsorbent inertness to interference from humidity and reactive
atmospheric species also become governing factors for adsorbent choice
Sorbent selection – low concentration atmospheres
Sorbent name Volatility range
Tenax TA C7 – C30
Carbograph 2TD C8 – C20
Carbograph 1TD C5/6 – C14
Carbograph 5TD C3/4 – C6/7
SulfiCarb C3 – C8
Carboxen 1003 C2 – C5
Water retention
Effects of poor sorbent/compound pairings
• This occurs due to the inversion of the concentration gradient within the
sampler air-gap. It occurs when the vapour phase concentration within the
sampler becomes greater than the vapour phase concentration in the
surrounding air. This is particularly acute when sampling poorly retained
compounds from transient atmospheres.
Sample losses i.e. back diffusion
Sampling complex atmospheres
• Three VOC atmospheres –
with the concentration a) static
(blue), b) falling (red) and c)
increasing (green) over the
monitoring period – but with
the same time weighted
average concentration overall
Why is it important to ensure a constant sampling rate?
• Three more VOC atmospheres
with an emission event
occurring
a) early-, b) mid- and
c) late- in the monitoring cycle,
but with the same time weighted
average concentration overall.
T=0 T= t
T=0 T= t
A B C
Sampling complex atmospheres
The result, if sampling rate is not constant e.g. slowly dropping
T=0 T= t
Model does not take in to account back diffusion, which would further skew these results
Sampling considerations
• When designing a passive sampling study, the target compounds and typical
concentrations should be known.
– This will help define
• the type of sampler to use
• the sorbent
• the exposure period
• As a general rule, concentration ranges are,
Compound target concentration
8 h Four week
Axial 2 μg m–3 to 10 mg m–3 0.3 μg m–3 to 300 μg m–3
• Sampling over shorter periods is possible (30 minutes) for higher concentrations, as
is sampling over longer periods (six weeks) for much lower atmospheric
concentrations (sub-ppb), providing the correct sorbent is used.
• However, when atmospheric concentrations are very low (<1 ppb) and when a result
is required within a few hours, in these instances high uptake rate samplers provide a
better solution.
Performance Evaluation
• Assessments covered should include,
– the repeatability and reproducibility of replicate measurements;
– experimentally observed uptake rates at different exposure
doses to identify any deviation from ideal behaviour;
– the influence of environmental factors including temperature,
pressure, humidity and wind velocity on sampler performance;
– the effects of potential interferents on the adsorbent and
adsorbed compounds
• Many publications exist on validation of sorbent based samplers and
international standards are available
Laboratory and field
Performance Evaluation
• ASTM D6246 - 08(2013) Standard Practice for Evaluating the Performance of Diffusive Samplers
• ASTM D4597 – 10, Standard Practice for Sampling Workplace Atmospheres to Collect Gases or Vapors with Solid Sorbent Diffusive Samplers
• ASTM D6306 – 10, Standard Guide for Placement and Use of Diffusion Controlled Passive Monitors for Gaseous Pollutants in Indoor Air
• Protocol for assessing the performance of a diffusive sampler (Method for the Determination of Hazardous Substances No. 27), UK Health & Safety Executive.
• CEN EN 838: Workplace atmospheres – Requirements and test methods for diffusive samplers for the determination of gases and vapours.
• EN ISO 16017: Air quality – Sampling and analysis of volatile organic compounds in ambient air, indoor air and workplace air by sorbent tube/thermal desorption/ capillary gas chromatography. Part 2: Diffusive sampling.
• MDHS 80: Volatile organic compounds in air. Laboratory method using diffusive solid sorbent tubes, thermal desorption and gas chromatography (August 1995).
• E.R. Kennedy et al., Protocol for the evaluation of passive monitors, US National Institute for Occupational Safety and Health (NIOSH), published in: ‘Diffusive sampling: An alternative approach to workplace air monitoring’ (CEC Publication No. 10555EN), 1986.
Publications/Methods
Comparison Data
Sample Location
Compound
TO-17
(14 day)TO-15 RPD
TO-17
(14 day)TO-15 RPD
TO-17
(14 day)TO-15 RPD
TO-17
(14 day)TO-15 RPD
Tetrachloroethene 3.4 3.4 0% 3.6 3.6 0% 0.45 J 0.43 J 5% 0.82 J 0.73 J 12%
Trichloroethene 1.3 1.2 8% 1.8 1.3 32% 0.11 U 0.46 U NA 16 13 21%
Samples collected in June 2012
Units in micrograms/cubic meter (ug/m3)
AI103 AI103P AI106 AI109
Strong correlation between passive diffusion sorbent tubes and
summa canister data for the two primary compounds of concern
(PCE and TCE)
Passive diffusion samplers allow you to sample over several days
or weeks to measure organic compounds in indoor and ambient
air providing a sample that may be more representative of average
concentrations
Passive Indoor Air Sampling – NAS JAX
Duplicate Sample Results
Duplicate samples
suspended side by side
show excellent
correlation with a 14-day
sampling period.
Sample Location
Methylene Chloride 0.60 0.59 1.7%
Trichloroethene 0.85 0.87 2.3%
Toluene 34.0 33.8 0.4%
Tetrachloroethene 306.7 305.6 0.4%
Ethylbenzene 4.11 4.01 2.5%
p & m-Xylene 10.98 10.59 3.6%
o-Xylene 2.66 2.64 0.8%
Sample Location
1,1,1-Trichloroethane 1.45 1.39 4.2%
Trichloroethene 2.44 2.58 5.6%
Tetrachloroethene 29.24 31.27 6.7%
Ethylbenzene 1.15 1.07 7.2%
Indoor #1
Beacon Passive
Diffusion Sampler
(14 Days)
Beacon Passive
Diffusion Sampler
(14 Days)
RPD
Compound
Compound
Indoor #2
Beacon Passive
Diffusion Sampler
(14 Days)
Beacon Passive
Diffusion Sampler
(14 Days)
RPD
Results in µg/m3
Passive Indoor Air Sampling – TN Sites
Real World applications of
passive sampling
Using Sorbent Tubes for Diffusive Sampling of
Outdoor Air
• 100 sampling sites - black
dots
• Yellow dots = pollution
hot spots
Mapping urban pollution concentrations with low-cost diffusive
(passive) sampling1
1 TDTS 10 Use of diffusive sampling with TD-GC for ambient air monitoring
Passive sampling for air monitoring in industry
Industrial perimeter
monitoring
Workplace air monitoring
Monitoring guidelines (sorbents, uptake
rates, sampling volumes, etc.) available from
standard methods
(Occupational hygiene)
In situ monitoring of
underground contamination
Sampling in confined atmospheres e.g.
Sub slab soil gas sampling
• Undesirable effect – starvation depends on two factors
– At low air velocities, a static layer of air external to the sampler will cause ‘starvation’ or an extension of the diffusive path and hence a flattening of the concentration gradient.
– If the air to which the sampler is exposed is stagnant, then the sampler may remove VOC vapors from the air faster than they are replenished, hence disturbing the equilibrium between the soil/water/vapour phases, also the sampler itself will impose a reduction in the concentrations it is intended to measure.
Ambient
concentration
Camb
Concentration
at adsorbent
surface
Cads
Diffusion
Path-length L
Air Velocity
Ambient conc. of
vapours in the
environment
Zero conc. of
vapours at the
sorbent surface
Air g
ap
Uptake rates can be reduced by:
• Making the diffusive path length longer
• Reducing the surface area of absorbent exposed
• Or a combination of both
Combating the Starvation Effect
Indoor Air and Personal Exposure
• TD-GC/MS is used for
several applications relating to
poor IAQ and ‘sick building
syndrome’.
• In this case residents were
complaining of poor air quality in
their home.
– Diffusive sampling with ‘axial’ sorbent
tubes was used to monitor indoor &
outdoor air quality at the house and to
monitor the personal exposure of
residents.
1st floor
2nd floor
3rd floor
Garage
Living
Quarters
Summary
• Many types of sorbent diffusive samplers (i.e. different geometries) have
been designed since the introduction and validation of sorbent based
samplers years ago.
• How do we decide which samplers should be included in standards, what
needs to be shown before we consider them? What properties MUST they
have
– Constant sampling rate
– Zero sink
• Passive sampling is a useful tool for VI monitoring and studies have
shown good correlation between these and other sampling devices.
• Applications could include, not only indoor air sampling, but personal and
soil gas (Sub slab).
Take home messages
Questions?
We would like to
acknowledge the work of
Beacon Environmental
Services
Contact us/Learn More
1-866-483-5684
www.markes.com
@MarkesInt
Markes International
Google plus
TheMarkes1