Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
Method validation of particle size distribution using static image analysis
Test & Measurements conference
Fortunate Modiba
Material characterisation
17 September 2019
Outline
INTRODUCTION
EXPERIMENTAL
RESULTS
Introduction
Particle size distribution
The particle-size distribution (PSD) of a powder, or granular material, or particlesdispersed in fluid, is a list of values or a mathematical function that defines therelative amount, typically by mass, of particles present according to size [1].
[1] Ujam, A. J., & Enebe, K. O. Experimental Analysis of Particle Size Distribution using Electromagnetic Sieve.http://www.ajer.org/papers/v2(10)/H02107785.pdf
Optical microscope
Optical Microscope is a technique that has
been used for many years within the NMISA,
to determine the shape, morphology and
size of particles
• Resolution: 250 μm
ISO 13322 :2014 the analysis of images for
the purpose of determining particle size
distributions [2].
[2] www.iso.org/standard/51257.html
Fish-Bone diagram
PSD
Calibration
ThresholdSample preparation
Experimental
As-received samples for PSD analysis
Sample preparation
• ISO 14887:2000, sample preparation can be done by following dispersing procedures for
powders in liquids [3].
• The sample is dispersed with the selected dispersant by first adding a small amount of the
sample on the glass slides and four drops of the dispersant (preferably isopropanol) are
added on the sample .
Sample dispersion
[3] https://www.iso.org/standard/25861.html
Obeservations
Agglomerated Crushed Well dispersed
Results
10 slides 10 areas on 1 slide
Analyst Mean (µm) %RSD Mean (µm) %RSD
1 2.2 ± 0.2 9.24 2.3 ± 0.4 17.94
2 2.7 ± 0.5 15.5 2.3 ± 0.2 6.83
Table 1: Sample preparation precision.
❖ Sample preparation
10 slides 10 areas on 1 slide
AnalystMean (µm) %RSD Mean (µm) %RSD
2.3 ± 0.35 6.13 2.3 ± 0.4 7.05
• F-test: standard deviations (SD) for the 10 slides for both analysts do not differ significantly.
• This implies that the mean and standard deviation are significantly comparable which makes the method reproducible.
• Comparing the two experimental means on the 10 areas SDs differs significantly.
Reproducibility
Repeatability
Observation
Effect of pixel size
pixel decreases
CalibrationTo ensure SI traceability of measurement results, the microscope was calibrated by using a certified calibration artefact from NIST.
NIST RM 8820 magnification artefact at 100x (10x lens).
Y
X
• Theoretical calibration based on the objective (magnification), camera pixel size and camera adopter which is setup in the microscope calibration.
• As no optical element can be produced the same therefore it’s required to create a manual calibration using a stage micrometre to ensure that the theoretical calibration it’s still within range.
Results
MagnificationTheoretica
l ratio (µm)
X1 %ErrorX1 X2 %ErrorXMean (X1X2)
%Error(X1X2)
50x 0.62 0.62 0.62 0.62 0.62 0.62 0.62
100x 0.31 0.31 0.61 0.32 3.84 0.31 0.61
200x 0.16 0.15 0.00 0.16 0.00 0.16 0.00
500x 0.06 0.06 1.24 0.06 0.55 0.06 1.24
1000x 0.03 0.03 0.95 0.03 0.20 0.03 0.95
MagnificationTheoretical ratio(µm)
Y1 %ErrorY1 Y2 %Error YMean (Y1Y2)
%Error(Y1Y2)
50x 0.62 0.62 0.62 0.62 0.69 0.62 0.62
100x 0.31 0.31 0.61 0.31 0.45 0.31 0.61
200x 0.16 0.16 0.00 0.15 0.00 0.16 0.00
500x 0.06 0.06 1.24 0.06 0.55 0.06 1.24
1000x 0.03 0.03 0.95 0.03 0.20 0.03 0.95
Table 3: Calibration
❖ Calibration
Reproducibility
The x and y are the co-ordinates used in the software for calibration. The X1 and Y1 represent the results for analyst 1 and X2, Y2 for analyst 2.
Results
MagnificationTheoretical
(µm)Experimental
(µm)%Bias %RSD
50x 0.63 0.62 0.58 0.07100x 0.31 0.31 0.42 0.07200x 0.16 0.16 0.00 0.07500x 0.06 0.06 0.16 0.09
1000x 0.03 0.03 0.33 0.04
MagnificationTheoretical
(µm)Experimental
(µm)%Bias %RSD
50x 0.63 0.62 0.62 0.07100x 0.31 0.31 0.35 0.09200x 0.16 0.16 0.00 2.51500x 0.06 0.06 0.16 0.08
1000x 0.03 0.03 0.33 0.04
Table 4: Calibration results on the x-direction
Table 5: Calibration results on the y-direction
• The %RSD was calculated from the experimental data was observed to the of < 3%.
• The y -direction in this case is found to be more stable and precise as compared to the x-direction.
Threshold
Polystyrene microspheres sphere Over threshold Good threshold
Results
10x mag 20x mag
Analyst Mean (µm) %RSD Mean (µm) %RSD
1 2.54 ± 0.05 2.15 1.51 ± 0.06 3.90
2 2.50 ± 0.03 1.38 1.47 ± 0.04 2.54
Table 2: Threshold precision.
• The F-test was used to compare the standard deviation (SD) and precisions of the two analysts.
• The test indicated that the two SDs do not differ significantly which implies that there is no significant difference in precision (<5%).
• The means are significantly accurate according to the t-test.
❖ Threshold
Reproducibility
Repeatability
10x mag 20x mag
Analyst Mean (µm) %RSD Mean (µm) %RSD
1 2.54 ± 0.03 2.15 1.50 ± 0.05 2.08
Repeatability
How accurate?
Results
Polystyrene microspheres3 µm 10 µm 30 µm
Experimental. mean 3.17 9.97 30.88
Experimental. Std. dev. 0.29 0.80 0.66Certified mean 3.01 9.71 30.39
Certified standard deviation 0.02 0.02 0.10%Bias 5.22 2.65 1.60
Degrees of freedom n-1 = 9tcalc 2.00 1.00 2.00tcrit 2.26 2.26 2.26
Decision rulestcalc<tcrit, the
method is accurate
The largest bias was observed at the smallest particle size. This may in part be related to
working at close to the method’s detection capability. Bias of 5% would still be fit for purpose.
The t-test with the value of t for a confidence interval of 95% (critical value of |t| for p value of
0.05).
Table 8: Bias results.
❖ Bias
Limit of detection and
quantification
Results
CRM size (µm)
Magnification
5x 10x 20x 50x 100x
3.00 3.34 2.75 3.39 3.36 3.51
1.50 2.20 2.32 2.30 1.70 2.01
1.00 2.71 1.83 1.77 1.99 2.47
0.70 2.38 1.45 1.45 1.50 1.32
0.50 3.30 1.73 1.40 1.45 0.00
The 3.0 µm results were observed to be comparable to the reference value, with the resultsfor the smaller particles not agreeing with the reference values.The SD for the 3.0 µm CRM (0,30 µm) was therefore used to determine the LOD and LOQ forthe methods:LOD = 0.86 µmLOQ = 2.90 µm
❖ Limit of detection/quantification
Table 9: CRM results for LOD/LOQ determination
What does it mean?
Results
SRW (%) Sr (%) %uCRM %Bias
5.00 16.00 8.96 0.05
• The NORT-test determined the Rw to be 17% and the %uBiasto be 9%.
• The method was observed to have the precision of 19%.
Table 10: NORT-test results
❖ Method precision using NORT-test
Conclusion
❖ The method proves to be accurate within 5% and precise within 19%.
❖ The particle size analysis can be conducted on powders with good accuracy and
±19% precision.
❖ The future work will be on evaluation of uncertainty on the sample preparation.
Acknowledgement
NMISA Green Economies
&
Particle Size cross-cutter project