Page 1 of 14
ATI
PANEL MOUNT VERSION
1) Panel Mount Enclosure Material:Aluminum, Black Anodized
2) Nema-4X Encloure Material:Base: Molded ABS, (tan in color)Cover: Clear Polycarbonate with
dual hinge.
3) Nema-4X Rating:
NEMA rated types 4,4XIEC 529 rated IP66Flammibility Rating UL94 HB
NOTESNEMA-4X VERSION
3.54"89.9mm
6.10"154.9mm
5.46"138.7mm.34"
8.6mm
3.78"96mm
3.78"96mm
Enclosure Dimensions
ATI-0124
ATI-010
Model A15/76 Turbidity Monitor -
SS-A1576, Rev-E (8/11)
Page 2 of 14
ATI
NOTESJUNCTION BOX
1) Nema-4X Encloure Material:Molded FRP (Fiberglas reinforcedpolyester),gray in color.
2) Nema-4X Rating:
UL listed per std. 508CSA certified per std. C22.2 No. 94NEMA rated types 4,4X,6,6P,12,13CEMA rated types 2,3,4 and 5IEC 529 rated IP 67-11
3) Mounting Dimensions:
3.11" [79mm] x 3.11" [79mm]
4) Part No. 00-0726.
ATI-0241
Optional Equipment DimensionsModel A15/76 Turbidity Monitor -
RE (8/11)
Page 3 of 14
ATI
NOTES
1) Front of panel is NEMA 1, general purposeonly.
2) Panel mounting clips supplied with monitor.
3.60"91.4mm
3.60"91.4mm
MOUNTING CLIPS(2 PER UNIT)PANEL CUT-OUT
WALL MOUNT - NEMA-4X
HANDRAIL MOUNT - NEMA-4X
1) Wall mounting requires customer suppliedscrews or bolts.
1) All listed hardware is supplied as part ofhandrail mounting kit # 00-0139.
PANEL MOUNT
Installation Diagrams
ATI-0125
ATI-0126
ATI-043
Model A15/76 Turbidity Monitor -
RE (8/11)
Page 4 of 14
ATI
A15/76 TURBIDITY MONITOR (VAC) ELECTRICAL CONNECTIONS
A15/76 TURBIDITY MONITOR(BACK VIEW)
ATI-0443
Terminal Block DesignationsModel A15/76 Turbidity Monitor -
RE (8/11)
Page 5 of 14
ATI
A15/76 TURBIDITY MONITOR (VDC) ELECTRICAL CONNECTIONS
A15/76 TURBIDITY MONITOR(BACK VIEW)
RE (8/11)
Page 6 of 14
ATI
SYSTEM WIRING DIAGRAM w/OPTIONAL JUNCTION BOX
ATI-0445
Model A15/76 Turbidity Monitor -
RE (8/11)
Wiring Diagrams
Page 7 of 14
ATI
FLOW SENSOR DIMENSIONS
Sensor Dimensions and Installation
ATI-0434
FLOWCELL DIMENSIONS
ATI-0435
Model A15/76 Turbidity Monitor -
RE (8/11)
Page 8 of 14
ATI
FLOWSENSOR/FLOWCELL INSTALLATION DRAWING
Sensor Dimensions and Installation
ATI-0444
Installation of flowcell must be done Horizontally as shown above.
Model A15/76 Turbidity Monitor -
RE (8/11)
Page 9 of 14
ATI Sensor Installation
TURBIDITY SENSOR 1½ FLOW TEE ASSEMBLY
VERTICAL CONFIGURATION IS RECOMMENDEDBUT HORIZONTAL CONFIGURATION IS ACCEPTABLE,PROVIDED SENSOR IS MOUNTED IN THE 3:00 OR 9:00POSITION.
ATI-0542
Model A15/76 Turbidity Monitor -
RE (8/11)
Page 10 of 14
ATI Sensor Installation
FLOW "T" RECOMMENDED PIPING
ATI-0513
Model A15/76 Turbidity Monitor -
RE (8/11)
Page 11 of 14
ATI
RE (8/11)
D15/76 AUTO-CLEAN TURBIDITY MONITOR
1) Nema-4X Encloure Material:Base: Molded ABS, Tan in ColorCover: Clear Lexan w/Dual Hinge
2) Nema-4X Rating:Nema rated types 4, 4XIEC 529 rated IP66
3) Sensor Material:Noryl EN265 & Stainless Steel
NOTES
AUTO-CLEAN TURBIDITY SENSOR
Dimensions
ATI-0553
ATI-0473
Model D15-76 Auto-Clean Turbidity Monitor -
Page 12 of 14
ATI
NOTESMONITOR HANDRAIL INSTALLATION
1) Always mount sensor cleaner within 25 ft. ofsensor if possible. Maximum air line distanceis 50 ft.
2) Sun shade over monitor is recommended inhot climates.
3) Monitor Handrail Mounting Kit # 00-0930supplies all hardware listed for mountingmonitor.
4) Wall mounting requires customer suppliedscrews or bolts.
5) Sensor Pipe Mounting Kit # 00-0624 suppliesmounting plate with hardware and PVC pipemounting adaptor.
6) Sensor Pipe Mounting Kit with Rag Shedder# 00-0629 supplies mounting plate withhardware and PVC rag shedder.
SENSOR HANDRAIL INSTALLATION
RE (8/11)
Installation Diagrams
ATI-0490
Model D15-76 Auto-Clean Turbidity Monitor -
Page 13 of 14
ATI
STANDARD SYSTEM WIRING DIAGRAM
RE (8/11)
System Interconnections
ATI-0534
Model D15-76 Auto-Clean Turbidity Monitor -
Page 14 of 14
ATI
REMOTE MONITOR WIRING DIAGRAM
RE (8/11)
System Interconnection
ATI-0536
Model D15-76 Auto-Clean Turbidity Monitor -