Technical Catalogue
Paddle Wheel Flow Meters & Instruments
Updated
January 2011
PDF Published January 2, 2011
Instruments
Flow Sensors
2 Digiflow® FlowX3 Technical Catalogue
Installation Fittings
Type F3.80Oval GearFlow Sensors
Type F3.00Remote Flow Sensors
Type F3.10Mini Flow Sensors
Type F3.01Direct Mount Flow Sensors
Type F111Hot Tap FlowSensors
Type ULFUltra LowFlow Sensors
Type F3.30 BlindTransmitters
Type ULF3.30 Ultra Low FlowBlind Transmitters
Type F3.15AdjustableFlow Switches
Type ULF3.15Adjustable UltraLow Flow Switches
• Tees
• Bolt-On Saddles
• Metal Strap-On Saddles
• Wafer Fittings
• Weld-On Adaptors
• Hot-Tap Assemblies
Type F9.00, .02, .03Flow Monitor/Transmitters
Type F9.50, F9.51Batch Controllers
Type F9.20Battery Powered Flow Monitor
Type F3.05No-FlowSwitches
©Chemline Plastics Limited 2011
Type F9.6Electromagnetic FlowMonitors/Transmitters
Type F3.6Electromagnetic FlowBlind TransmittersType F3.61 Hot-Tapversion is available
PDF Published January 2, 2011
Product Line
FlowX3 Product Line 2
Table of Contents 3
A Complete Range of Flow Monitoring Equipment 4-5
Building a FlowX3 Flow Monitoring System 6-7
Flow Sensors – F3.00 / F3.01 / F3.10 8-12
Ultra Low Flow Sensors – ULF 13-14
Oval Gear Flow Sensors – F3.80 15-16
Hot-Tap Flow Sensors – F111 17-18
No-Flow Switches – F3.05 19
Adjustable Flow Switches – F3.15 / ULF3.15 20-21
Blind Transmitters – F3.30 / ULF3.30 22-23
Item Numbers – Flow Switches, Blind Transmitters 24
Instruments and Mounting Kits 25
Flow Monitor/Transmitters – F9.00 / F9.02 26-27
Dual Input Flow Monitor/Transmitter – F9.03 28-29
Batch Controller – F9.50 / F9.51 30-31
Battery Powered Flow Monitor – F9.20 32
Instruments – Technical & Dimensions 33
Electromagnetic Flow Transmitters – F3.60/F3.63/F3.61 34-36
Electromagnetic Flow Monitor/Transmitters – F9.60/F9.63 37-38
Installation Fittings 39-43
Working Pressure vs. Temperature 43
Installation Guidelines 44
Calibration Factors & Flow Ranges 45-46
Selection Questionnaire 47
A Cost Effectiveand Complete Rangeof Flow Monitoring
Equipment
Digiflow® FlowX3 Technical Catalogue 3
Contents Page
PDF Published January 2, 2011
Instruments
Sensors
4 Digiflow® FlowX3 Technical Catalogue
Direct Mount Panel Mount Wall Mount
■ Designed for Corrosive ServicesFlowX3 paddle wheel flow meters are ofcorrosion resistant solid plastic construction. Theyare recommended for services with up to 10%solids*. Bearings and sensor body near the rotor aredesigned to be self-cleaning and small magnets inthe rotor blade lessen the chance of fouling due toaccumulation of magnetic particles in the process.
■ A Selection of Plastic or Metal Sensor MaterialsThe standard thermoplastic flow sensors of CPVCor PVDF are compatible with a wide range ofplastic piping materials. Brass and 316L stainlesssteel flow sensors allow application in metal pipingsystems and for fluids at higher temperatures. Lowflow sensors are available in PP, POM, ECTFE (Halar®) and 316L stainless steel.
■ Submersible Sensors AvailableNEMA 6, 6P (IP68) models are available foroutdoor or submersible installations.
* Particle size not exceeding 0.5 mm cross section or length.
The Digiflow® FlowX3 line includes a complete rangeof paddle wheel flow sensors, instruments and sensorinstallation fittings. Also included are insertion typeelectromagnetic flow transmitters.
FlowX3 products provide high quality andstate-of-the-art technology. All products aremanufactured under ISO 9001:2000 and conformto the highest CE standards.
Chemline backs the FlowX3 line with service and support.• Expert technical assistance• After sales service• Customer training• Fast deliveries from Chemline's inventory
■ High Accuracy± 0.75% of full scale with standard K factor± 1% of reading value with field calibration
■ Excellent Low Flow MeasurementAccurately measures flow velocities down to0.15 m/s (0.5 ft./sec.)
■ Large Flow Velocity RangeThe frequency output is linearly proportional toflow rate from 0.15 to 8 m/s (0.5 to 25 ft./sec.).Approximately 45 Hz/m/s.
■ Easily Installed into all Types and Sizes of PipingSensor installation fittings are available in a largeselection of types, materials and pipe sizes up to24" and larger.
■ Up-to-Date Electronics – Many OutputsDigiflow® FlowX3 instruments were first introducedin 2003. The line of instruments has since expanded.A large offering of control outputs includes 4 to20 mA, open collector (pulse) and relays. All unitshave 3 line LCD displays and LEDs for indicationof output status.
■ Modular Design – 1/4 DIN SizeOnly with Digiflow® FlowX3 can the sameinstrument be mounted in 3 different ways: directlyto the flow sensor, or remotely in a panel or on awall.
PDF Published January 2, 2011
A Complete Range of Flow
Removable Cover
5 Button SiliconeRubber Keypad
1/4 DIN sizeMonolithicEnclosure
Integrated SolidState Electronics•Encapsulated in epoxy
for corrosion resistance
5 Blade ECTFE(Halar®) Rotor•Open channel design
avoids cavitation•Five small magnets within
rotor blades produce fivepulses per revolution
Sensor Body•Available in CPVC,
PVDF, 316L StainlessSteel or Brass
•2 sensor lengthsfor 1/2" to 24" pipes
Ceramic Shaft & Bearings•High chemical resistance•Long Life – 20 years/m/s•Self cleaning design
Rotor Bearing (2)
Shaft
Digiflow® FlowX3 Technical Catalogue 5
■ Long Service LifeFlowX3 have a ceramic rotor shaft and bearings.On aggressive chemicals and services containinggrit, FlowX3 sensors have outlasted other typeswith metal shafts and no bearings.
Plug-in RemovableTerminals• exclusive to FlowX3
EpoxyEncapsulatedElectronics
3 line LCD Display• two 12–digit Alphanumeric lines
• Flow Rate and Total Flow together
• one Output Status Icon line
• backlit displays
Status Displays,up to 3 LEDs
■ Easy Set-upSetting up the instrument is easy using the keypadand self-explanatory menus. Plug-in terminalsmake instrument connection and removal easy.Auto calibration provides automatic calculation ofK factors.
■ Designed for Heavy Duty IndustrialApplicationsInstruments have a NEMA 4, 4X monolithic clearpolycarbonate enclosure in 1/4 DIN size, epoxyencapsulated electronics and a 5 button siliconerubber keypad.
New
PDF Published January 2, 2011
Monitoring Equipment Approved
Instruments
Flow Sensors
InstallationFittings
F9.50, F9.51Batch Controllersp.30-31
F9.02, F9.03Flow Monitor andTransmitters (3/4 wire)p.26-29
F9.00Flow Monitor andTransmitter (2 wire)p.26-27
Hot-TapMetal StrapSaddles p.42
3" to 18"
F3.10 MiniFlow MeterTee p.12
1/2" to 1-1/2"
Hot-Tap 316LStainless SteelWeld-OnAdaptors p.42
16" plus
F9.20Battery PoweredFlow Monitor p.32
F111.C
6 Digiflow ® FlowX3 Technical Catalogue
Hot-Tap FlowSensors p.17-18
Hall Effect Sensors p.8-12
F3.01.HNEMA 4, 4X
F3.00.HNEMA 6, 6P
F3.00.HNEMA 4, 4X
F3.10.HNEMA 4, 4X
Saddles p.39
2" to 12"Tees p.39
1/2" to 1-1/2"Metal Strap-OnSaddles p.40
3" to 18"
Hall CoilF3.61MHot-Tap
ElectromagneticFlow Transmittersp.34-38
F3.6MF9.6M F111.H
PDF Published January 2, 2011
It’s easy to build a flow monitoring systemFollow the lines to determine which specific Flow Sensors,
F3.05 F3.15
AdjustableFlow Switches p.20-21
Modular DesignInstrumentsThe same instrumentmay be mounted inthree different waysusing mounting kits.p.25
BlindTransmitters p.22-23
ULF3.15 F3.30 ULF3.30
ULF.RReed
ULF.HHall
F3.80Hall
F3.00.CNEMA 6, 6P
F3.01.CNEMA 4, 4X
Coil Effect Sensors p.8-10
WaferFittings p.40
10" and 12"
Weld-OnAdaptors p.41
1-1/2" to 24"
316L Stainless SteelThreaded Tees p.41
3/4" to 1-1/4"
Digiflow® FlowX3 Technical Catalogue 7
Panel MountingDirect Mountingto flow sensor
Wall Mounting
F3.00.CNEMA 4, 4X
Ultra LowFlow Sensorsp.13-14
Oval GearFlow Sensorsp.15-16
No-FlowSwitches p.19
PDF Published January 2, 2011
using Digiflow® FlowX3 componentsInstruments and Installation Fittings may be connected
The Type F3.00/F3.01 Paddle Wheel Flow Sensors are thecore items in the Digiflow® FlowX3 line. A square waveoutput signal is generated with frequency proportional torate of rotor rotation and flow velocity. This pulse outputis normally fed to a FlowX3 flow monitor/transmitter,blind transmitter, batch controller or adjustable flowswitch. It can also be fed to other brand instruments orPLC’s.
Three types of sensors are available, Hall Effect whichrequires a 5 to 24 VDC power supply, Coil Effect whichoperates with less power, 3 to 5 VDC and Push-pullsensors for safe connection to any NPN or PNP inputs.Coil is required with the battery powered flow monitor.Hall Effect signals may be transmitted up to 300 meters (984ft.) without the need for conditioning whereas Coil Effectsignals may be transmitted up to 16 m (52.5 ft.) withoutconditioning.
Body Materials: CPVC, PVDF, 316L Stainless Steel, Brass
Rotor: ECTFE (Halar®)
Shaft & Bearings: Ceramic
Seals: EPDM, Viton®
Pipe Sizes: 1/2"– 24" in two sensor lengths, L0 or L1See Installation Fittings (pages 39-41)
Flow Ranges: See page 45
■ Features
• Ceramic Shaft and Bearings – For longer life on servicescontaining grit
• Self Cleaning Design – Lower maintenance
• Submersible Sensors Available – NEMA 6, 6P (IP68)models are available for outdoor or submersibleinstallation
BrownGreenWhite
FlowX3Instruments
NEMA 4, 4X (IP65)Sensor
NEMA 6, 6P (IP68)Sensor
Other BrandInstruments
Other BrandInstruments
FlowX3Instruments
■ WiringSensor Connections to Instruments
■ Connectable FlowX3 InstrumentsFlowX3Instruments*
SensorType
SensorNo.
InstrumentMounting
Hall F3.01.H Direct F9.00, F9.02,F3.00.H Panel or Wall F9.03, F9.50, F9.51
Coil F3.01.C Direct F9.20F3.00.C Panel or Wall
Push-pull F3.00.P – NPN or PNP Inputs
* 10 kW pull-up resistor may be required when Hallsensors are connected to other brand instruments.Resistor is not required for the F3.00.P sensors.
*
*
* Power supply is normally fed from FlowX3 instruments.
8 Digiflow® FlowX3 Technical Catalogue
PDF Published March 18,2011
Type F3.00/F3.01Flow Sensors
New
* F3.01 sensor is NEMA 6, 6P but when instrumentis attached, unit becomes NEMA 4, 4X.
† Required sensor length (L0 or L1) dependson choice of installation fittings.See pages 39 to 41.
F3.00For Remote Connection to Instruments
NEMA 4, 4X (IP65) NEMA 6, 6P (IP68)
8 m (26.4 ft.)long cablesupplied
17 cm (7") longcable supplied
L0 = 68.3 mm (2.69")L1 = 98.5 mm (3.88")
■ Technical – GeneralOutput Signal: Square wave (pulse)
Output Frequency: 45 Hz per m/s nominal(13.7 Hz per ft./sec.)
Electrical Class: NEMA 6, 6P (IP68) – F3.00 onlyNEMA 4, 4X (IP65) – F3.00 and F3.01
Accuracy: < ± 1% of reading value after fieldcalibration or ± 0.75% of full scale
Repeatability: ± 0.5% of full scale
Velocity Range: 0.15 to 8 m/s (0.5 to 25 ft./sec.).See page 46 for corresponding flowranges.
Viscosity Range: 0.5 to 20 centistokes. Field calibrationis required if outside this range, up to40 centistokes maximum.
Maximum % Solids: 10% with particle size not exceeding0.5 mm cross section or length
Max. OperatingPressure/Temperature: See chart on page 45
Cable (where supplied): 22 AWG, 3 conductors
■ Technical – F3.00.H and F3.01.H Hall SensorsSupply Voltage*: 5 to 24 VDC regulated
Supply Current: < 30 mA @ 24 VDC
Output Type: Transistor NPN open collector
Output Current: 10 mA max.
Max. Cable Length: Max. 300 m (984 ft.) recommendedwithout signal conditioning
■ Technical – F3.00.C and F3.01.C Coil SensorPower Supply: Normally 2 x 3.6 V Lithium batteries
located in the F9.20 flow monitor or3 to 5 VDC regulated
Supply Current: < 10 mA
Min. Input Impedance: 100 kWMax. Cable Length: Max. 16 m (52.5 ft.) recommended
without signal conditioning
■ Technical – F3.00.P Push-pull SensorsSupply Voltage: 12 to 24 VDC regulated
Supply Current: < 30 mA @ 24 VDC
Output Type: Push-pull for connection to NPN and PNP inputs
Output Current: 20 mA max.
Max. Cable Length: Max. 300 m (984 ft.) recommendedwithout signal conditioning
* Supply voltage is normally fed from FlowX3 instruments.
■ Installation• See page 44 for guidelines on installation in piping systems
• See pages 39 to 41 for installation fittings Digiflow® FlowX3 Technical Catalogue 9
3 pole DINconnector(cable notsupplied)
F3.01For Direct Connection to Instruments
NEMA 4, 4X (IP65)*
†
††
PDF Published January 2, 2011
Type F3.00/F3.01Flow Sensors
PDF Published March 18, 2011
10 Digiflow® FlowX3 Technical Catalogue
NEMA 4, 4X(IP65)
ElectricalClass O-Ring Brass
SensorLength† PVDF 316L SSCPVC
■ Item Numbers
NEMA 6, 6P(IP68)
NEMA 4, 4X(IP65)
ElectricalClass O-Ring Brass
SensorLength† PVDF 316L SS
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
L0
L0
L1
L1
F3.00.H.05
F3.00.H.06
F3.00.H.07
F3.00.H.08
F3.00.H.17
F3.00.H.18
F3.00.H.19
F3.00.H.20
CPVC
F3.00.H.01
F3.00.H.02
F3.00.H.03
F3.00.H.04
F3.00.H.13
F3.00.H.14
F3.00.H.15
F3.00.H.16
F3.00.H.09
F3.00.H.10
F3.00.H.11
F3.00.H.12
F3.00.H.21
F3.00.H.22
F3.00.H.23
F3.00.H.24
F3.00.H.25
F3.00.H.26
F3.00.H.27
F3.00.H.28
F3.00.H.29
F3.00.H.30
F3.00.H.31
F3.00.H.32
For Direct Mounting to Instruments
Coil Effect Sensors – For Connection to F9.20 FlowX3 Instrument
Hall Effect Sensors – For Connection to F9.0 and F9.5 Series FlowX3 Instruments
For Connection to Panel or Wall Mount Instruments
NEMA 4, 4X(IP65)
ElectricalClass O-Ring Brass
SensorLength† PVDF 316L SS
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
F3.01.C.05
F3.01.C.06
F3.01.C.07
F3.01.C.08
CPVC
F3.01.C.01
F3.01.C.02
F3.01.C.03
F3.01.C.04
F3.01.C.09
F3.01.C.10
F3.01.C.11
F3.01.C.12
F3.01.C.25
F3.01.C.26
F3.01.C.27
F3.01.C.28
For Direct Mounting to Instruments
NEMA 6, 6P(IP68)
NEMA 4, 4X(IP65)
ElectricalClass O-Ring Brass
SensorLength† PVDF 316L SS
Body Material
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
L0
L0
L1
L1
F3.00.C.05
F3.00.C.06
F3.00.C.07
F3.00.C.08
F3.00.C.17
F3.00.C.18
F3.00.C.19
F3.00.C.20
CPVC
F3.00.C.01
F3.00.C.02
F3.00.C.03
F3.00.C.04
F3.00.C.13
F3.00.C.14
F3.00.C.15
F3.00.C.16
F3.00.C.09
F3.00.C.10
F3.00.C.11
F3.00.C.12
F3.00.C.21
F3.00.C.22
F3.00.C.23
F3.00.C.24
F3.00.C.25
F3.00.C.26
F3.00.C.27
F3.00.C.28
F3.00.C.29
F3.00.C.30
F3.00.C.31
F3.00.C.32
For Connection to Panel or Wall Mount Instruments
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
F3.01.H.01
F3.01.H.02
F3.01.H.03
F3.01.H.04
F3.01.H.05
F3.01.H.06
F3.01.H.07
F3.01.H.08
F3.01.H.09
F3.01.H.10
F3.01.H.11
F3.01.H.12
F3.01.H.25
F3.01.H.26
F3.01.H.27
F3.01.H.28
Body Material
Body Material
Body Material
Push-pull Sensors – For Connection to NPN or PNP inputs
† Required sensor length (L0 or L1) depends on choice of installation fittings. See pages 39 to 41.
NEMA 6, 6P(IP68)
NEMA 4, 4X(IP65)
ElectricalClass O-Ring Brass
SensorLength† PVDF 316L SS
Body Material
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
L0
L0
L1
L1
F3.00.P.05
F3.00.P.06
F3.00.P.07
F3.00.P.08
F3.00.P.17
F3.00.P.18
F3.00.P.19
F3.00.P.20
CPVC
F3.00.P.01
F3.00.P.02
F3.00.P.03
F3.00.P.04
F3.00.P.13
F3.00.P.14
F3.00.P.15
F3.00.P.16
F3.00.P.09
F3.00.P.10
F3.00.P.11
F3.00.P.12
F3.00.P.21
F3.00.P.22
F3.00.P.23
F3.00.P.24
F3.00.P.25
F3.00.P.26
F3.00.P.27
F3.00.P.28
F3.00.P.29
F3.00.P.30
F3.00.P.31
F3.00.P.32
For Connection to Panel or Wall Mount Instruments
PDF Published March 18, 2011
Type F3.00/F3.01Flow Sensors
Digiflow® FlowX3 Technical Catalogue 11
The Type F3.10 Mini Flow Sensor features compact sizeand low cost. It can measure flow from 0.25 to 4 m/sproducing a highly repeatable frequency output signal.Rugged construction and proven Hall Effect sensortechnology provides exceptional performance with littleor no maintenance. A 4-blade paddle wheel withoutbearings generates a square wave output signal with afrequency of 15 Hz per m/s (4.6 Hz per ft./sec.).
■ Features
• Low Cost – Ideal for water treatment equipment
• Compact Design
• Submersible Rating – NEMA 6, 6P (IP68) enclosure ratingallows for outdoor or submersible applications
• Connects to the Full Range of FlowX3 InstrumentsAll models of the Flow Monitor/Transmitters and BatchController may be connected remotely, either panel orwall mounted.
Brown
Green
White
F3.10 Sensor Connection toFlowX3 Instruments
Brown
Green
White
F3.10 Sensor Connection toOther Brand Instruments
■ Wiring
-
+
Body Materials: ABS, PVDF
Rotor: ABS, PVDF
Shaft: 316L SS
Seals: EPDM, Viton®
Pipe Sizes: 1/2"– 1-1/4"
Flow Ranges: 0.25 to 4 m/s (0.8 to 12.5 ft.sec.)Monodirectional flow
■ Connectable FlowX3 Instruments
FlowX3Instruments*
SensorType
SensorNo.
InstrumentMounting
Hall F3.10.H Direct F9.00, F9.02, F9.03,F9.50, F9.51
* Power supply is normally fed from FlowX3 instruments.
10 KW
2 KW to 10KW pull-up resistor may be required.
PDF Published January 2, 2011
Type F3.10Mini Flow Sensors
■ Technical – GeneralOutput Signal: Square wave (pulse)
Output Frequency: 15 Hz per m/s nominal(4.6 Hz per ft./sec.)
Output Type: NPN open collector
Output Current: 10 mA max.
Pulses per Litre: See page 45
Electrical Class: NEMA 6, 6P (IP68)
Accuracy: < ± 2% of reading value after fieldcalibration or ± 1.0% of full scale
Repeatability: ± 0.5% of full scale
Viscosity Range: 1 to 10 centistokes
Applications: Clean non-corrosive services only,no solids recommended
Working Temperature: –10 to 80oC (14 to 176oF)
Working Pressure: 5 bar (70 psi) max. @ 25oC (77oF)
Cable: 22 AWG, 3 conductors, 2 m (6.6 ft.)length supplied standard
Max. Cable Length: Max. 300 m (990 ft.) recommendedwithout signal conditioning
Supply Voltage*: 5 to 24 VDC regulated
Supply Current: < 30 mA @ 24 VDC
* Supply voltage is normally fed from FlowX3 instruments.
Installation Fittingfor Mini Flow Sensors
TeesSizes: 1/2" to 1-1/4"Maximum Pressure:70 psi @ 25oC (77oF)
PipeSize B
PVC / EPDM / SocketE Item No.H
1/2"
3/4"
1"
1-1/4"
1.10
1.18
1.28
1.50
d0.787
0.984
1.260
1.575
1.12
1.30
1.61
1.96
TEA005ES-310
TEA007ES-310
TEA010ES-310
TEA012ES-310
2.20
2.60
3.11
3.78
DIMENSIONS INCHES
1.6"
0.8"
G 3/4"
12 Digiflow® FlowX3 Technical Catalogue
2 m (6.5 ft.) longcable supplied
PDF Published January 2, 2011
Type F3.10Mini Flow Sensors
Digiflow® FlowX3 Technical Catalogue 13
The Type ULF Ultra Low Flow Sensors are available intwo flow ranges. They contain a transducer and a fourblade paddle wheel rotor with a permanent magnet insideeach blade. A square wave output signal is generated withfrequency proportional to rate of rotor rotation and flowvelocity. This pulse output is normally fed to a FlowX3flow monitor/transmitter, blind transmitter, batchcontroller or adjustable flow switch. It can also be fed toother brand instruments or PLC’s.
Two types of sensors are available, Hall Effect whichrequires a 5 to 24 VDC power supply and Reed Effectwhich have a volt-free contact. Reed Effect is requiredwith the battery powered flow monitor. Hall Effect signalsmay be transmitted up to 300 meters (984 ft.) without theneed for conditioning whereas Reed Effect signals may betransmitted up to 16 m (52.5 ft.) without conditioning.
■ Features
• Connects to the Full Range of FlowX3 InstrumentsAll models of the Flow Monitor/Transmitters and BatchController may be connected remotely, either panel orwall mounted.
† Corepoint® is a steel alloy.
Polyoxymethylene(POM) Version
ECTFEVersion
Sensor Body: POM ECTFE (Halar®)
Rotor: POM ECTFE (Halar®)
Shaft: Corepoint®† Sapphire
Bearings: – Sapphire
O-Ring: Viton® Viton® or Kalrez®
Connections: Inline 1/4" BSP male threaded
Flow Ranges:ULF01 Model: 1.5 to 100 l/h (0.0066 to 0.44 gpm)ULF03 Model: 6 to 250 l/h (0.0264 to 1.10 gpm)
Brown
White
Green
ULF.R Sensor Connection to FlowX3and Other Brand Instruments
ULF.H Sensor Connection toFlowX3 Instruments
Brown
Green
White
ULF.H Sensor Connection toOther Brand Instruments
■ Wiring
■ Connectable FlowX3 Instruments
FlowX3 Instruments*InstrumentMounting
SensorNo.
SensorType
Panel or Wall ULF.H Hall F9.02, F9.03, F9.50, F9.51
Panel or Wall ULF.R Reed F9.00, F9.20
* Power supply is normally fed from FlowX3 instruments.
Brown
White
Green
-
+
PDF Published January 2, 2011
Type ULFUltra Low Flow Sensors
Connections:1/4" BSP male thread
■ Technical – GeneralOutput Signal: Square wave (pulse)
Pulses per Litre: See page 45
Electrical Class: NEMA 4, 4X (IP65)
Accuracy: < ± 2% of reading value after fieldcalibration or ± 1.0% of full scale
Repeatability: ± 0.5% of full scale
Viscosity Range: 1 to 10 centistokes
Maximum % Solids: Clean services only, no solidsrecommended
Working Temperature: –10 to 80oC (14 to 176oF)
Working Pressure: 5 bar (70 psi) max. @ 22oC (72oF)
Cable: 22 AWG, 3 conductors
■ Technical – ULF.H (Hall Effect) SensorSupply Voltage*: 5 to 24 VDC regulated
Supply Current: < 15 mA @ 24 VDC
Max. Cable Length: Max. 300 m (984 ft.) recommendedwithout signal conditioning
■ Technical – ULF.R (Reed Effect) SensorOutput Type: Volt free Reed contact
Max. Cable Length: Max. 16 m (52.8 ft.) recommendedwithout signal conditioning
* Supply voltage is normally fed from FlowX3 instruments.
■ Installation Guidelines
• The sensor can be installed in any position, horizontally or vertically,although horizontal is preferred. A non horizontal installation maycause a greater error in the lower flow measurement range.
• Install the sensor with the arrow pointing in the flow direction.
• Maximize the distance between sensor and pump. Do not install thesensor immediately downstream of valves, elbows or any obstacles.15 cm (6") of straight pipe is suggested before and after the sensor.
14 Digiflow® FlowX3 Technical Catalogue
NEMA 4, 4X(IP65)
ElectricalClass O-RingBody l/hr USGPM Hall Reed
Item No.Flow RateMaterials
POM
ECTFE
Viton®
Viton®
Kalrez®
1.5 – 100
6.0 – 250
1.5 – 100
6.0 – 250
1.5 – 100
6.0 – 250
0.0066 – 0.44
0.0264 – 1.10
0.0066 – 0.44
0.0264 – 1.10
0.0066 – 0.44
0.0264 – 1.10
ULF01.H.0
ULF03.H.0
ULF01.H.2
ULF03.H.2
ULF01.H.3
ULF03.H.3
ULF01.R.0
ULF03.R.0
ULF01.R.2
ULF03.R.2
ULF01.R.3
ULF03.R.3
■ Item Numbers
2 m (6.5 ft.) longcable supplied
PDF Published January 2, 2011
Type ULFUltra Low Flow Sensors
The Type F3.80 Oval Gear Flow Sensors are positivedisplacement flow meters suitable for high viscosity fluidsup to 1,000 cP. Accuracy and repeatability is high.
Inside is a transducer and two oval gears oriented at 90degrees to each other with embedded permanent magnets.A square wave output signal is generated with frequencyproportional to rate of rotor rotation and flow velocity.This pulse output is normally fed to a FlowX3 flowmonitor/transmitter or batch controller. It can also be fedto other brand instruments or PLC’s.
PPVersion
ECTFEVersion
SSVersion
Sensor Body: PP ECTFE 316L SS
Oval Gears: ECTFE ECTFE ECTFE
Shaft: Zirconium Zirconium 316L SS
O-Ring: Viton® Viton® Viton®
Connections: Inline 1/4" BSP female threaded
Flow Ranges:F3.81.H Model: 10 to 100 l/h (0.044 to 0.44 gpm)F3.82.H Model: 25 to 150 l/h (0.11 to 0.66 gpm)
■ Connectable FlowX3 Instruments
FlowX3 Instruments*InstrumentMounting
Panel or Wall F9.02, F9.03, F9.50, F9.51
* Supply voltage is normally fed from FlowX3 instruments.
Special versions with 4 to 20 mA output or alarmoutput available on request.
F3.80 Sensor Connection to FlowX3Instruments
F3.80 Sensor Connection to Other BrandInstruments
■ Wiring
Brown
Green
White
Brown
Green
White
-
+
Digiflow® FlowX3 Technical Catalogue 15
■ Features
• High Accuracy and Repeatability
• Excellent choice for low flows of high viscosity fluids
• Suitable for pulsating flows
PDF Published January 2, 2011
Type F3.80Oval Gear Flow Sensors
■ Installation Guidelines
• The sensor can be installed in any position, horizontally or vertically,although horizontal is preferred. A non horizontal installation maycause a greater error in the lower flow measurement range.
• Install the sensor with the arrow pointing in the flow direction.
• Maximize the distance between sensor and pump. Do not installthe sensor immediately downstream of valves, elbows or anyobstacles.15 cm (6") of straight pipe is suggested before and afterthe sensor.
■ TechnicalSupply Voltage*: 5 to 24 VDC regulated
Supply Current: < 15 mA @ 24 VDC
Output Signal: Square wave (pulse)
Electrical Class: NEMA 4, 4X (IP65)
Accuracy: ± 1% of reading value
Repeatability: < ± 0.3% of reading value
Max. Fluid Viscosity: 1,000 centipoise
Maximum % Solids: Clean services only, no solidsrecommended
Working Temperature: –10 to 60oC (14 to 140oF)
Maximum Working Pressure:PP Body: 6 bar (87 psi) @ 25oC (77oF)
3 bar (43 psi) @ 60oC (140oF)ECTFE Body: 8 bar (116 psi) @ 25oC (77oF)
5 bar (72 psi) @ 60oC (140oF)SS Body: 8 bar (116 psi) @ 60oC (140oF)
Cable: 22 AWG, 3 conductors
Max. Cable Length: Max. 300 m (984 ft.) recommendedwithout signal conditioning
* Supply voltage is normally fed from FlowX3 instruments.
NEMA 4, 4X(IP65)
ElectricalClass O-RingBody Gears l/hr USGPM Item No.
Flow RateMaterials
PP
ECTFE
316L SS
ECTFE
ECTFE
ECTFE
Viton®
Viton®
Viton®
10 – 100
25 – 150
10 – 100
25 – 150
10 – 100
25 – 150
0.044 – 0.44
0.11 – 0.66
0.044 – 0.44
0.11 – 0.66
0.044 – 0.44
0.11 – 0.66
F3.81.H.01
F3.82.H.01
F3.81.H.02
F3.82.H.02
F3.81.H.03
F3.82.H.03
■ Item Numbers
16 Digiflow® FlowX3 Technical Catalogue
2 m (6.5 ft.) longcable supplied
PDF Published January 2, 2011
Type F3.80Oval Gear Flow Sensors
Digiflow® FlowX3 Technical Catalogue 17
The Type F111 Hot-Tap Flow Sensors may be removedand reinstalled while the pipe is under pressure. A squarewave output signal is generated with frequencyproportional to rate of rotor rotation and flow velocity.This pulse output is normally fed to a FlowX3 flowmonitor/transmitter, blind transmitter, batch controller oradjustable flow switch. It can also be fed to other brandinstruments or PLC’s.
Two types of sensors are available, Hall Effect whichrequires a 5 to 24 VDC power supply and Coil Effectwhich operates with less power, 3 to 5 VDC. Coil isrequired with the battery powered flow monitor. HallEffect signals may be transmitted up to 300 meters (984 ft.)without the need for conditioning whereas Coil Effectsignals may be transmitted up to 16 meters (52.5 ft.) withoutconditioning.
■ Features
• Ceramic Shaft and Bearings – Provide long life on servicescontaining grit
• Heavy Duty Industrial Design
• Side Pressure Port – This port is used to equalize pressure,assisting sensor insertion into the line under pressure. It canbe used as a pressure monitoring port.
F111.H and F111.C Sensor Connection toFlowX3 Instruments
F111.H Sensor Connection to OtherBrand Instruments
■ Wiring
F111.C Sensor Connection to OtherBrand Instruments
BrownGreenWhite
BrownGreenWhite
BrownGreenWhite
Body Material: 304 Stainless Steel
Rotor: ECTFE (Halar®)
Shaft & Bearings: Ceramic
Seals: Viton®
Pipe Sizes: 3" – 24"Special versions available for other sizesSee Installation Fittings (page 42)
*10 kW Pull-up resistor may be required whenHall sensors are connected to other brandinstruments.
*
■ Connectable FlowX3 Instruments
FlowX3 Instruments*InstrumentMounting
SensorNo.
SensorType
Panel or Wall F111.H Hall F9.00, F9.02, F9.03,F9.50, F9.51
Panel or Wall F111.C Coil F9.20
* Power supply is normally fed from FlowX3 instruments.
Side Pressure PortBall valve andsaddle not shown.
PDF Published January 2, 2011
Type F111Hot-Tap Flow Sensors
18 Digiflow® FlowX3 Technical Catalogue
■ Technical – GeneralOutput Signal: Square wave (pulse)
Output Frequency: 45 Hz per m/s nominal(13.7 Hz per ft./sec.)
Electrical Class: NEMA 6, 6P (IP68)
Accuracy: < ± 1% of reading value after fieldcalibration or ± 0.75% of full scale
Repeatability: ± 0.5% of full scale
Velocity Range: 0.15 to 8 m/s (0.5 to 25 ft./sec.)
Viscosity Range: 0.5 to 20 centistokes. Field calibrationis required if outside this range, up to40 centistokes maximum.
Maximum % Solids: 10% with particle size not exceeding0.5 mm cross section or length
Connection to Fitting: 2" Male BSP (GAS) thread
Side Pressure Port: Quick Connect 3/8"
Max. OperatingPressure/Temperature: 150 psi at 80oC (176oF)†
Cable: 22 AWG, 3 conductors
■ Technical – F111.H Hall Effect Sensor Supply Voltage*: 5 to 24 VDC regulated
Supply Current: < 30 mA @ 24 VDC
Current Consumption: 12 to 30 mA
Output: Transistor NPN open collector
Output Current: 10 mA max.
Max. Cable Length: Max. 300 m (984 ft.) recommendedwithout signal conditioning
■ Technical – F111.C Coil Effect Sensor Power Supply: Normally 2 x 3.6 V Lithium batteries
located in the F9.20 flow monitor or3 to 5 VDC regulated
Supply Current: < 10 mA
Min. Input Impedance: 100 kW
Max. Cable Length: Max. 16 m (52.5 ft.) recommended without signal conditioning
* Supply voltage is normally fed from FlowX3 instruments.† Maximum pressure is limited to 150 psi by the Hot-Tap saddle and
ball valve illustrated. The Hot-Tap flow sensor itself is rated to 290 psi.
■ Installation Fittings
• The Hot-Tap Sensor must be installed with a Hot-Tap Saddle orWeld-On Adaptor, complete with ball valve shown above andon page 42. Other Hot-Tap Saddles (including ball valves) maybe used, providing they have a 2" BSP (GAS) female connection.
8 m (26.4 ft.)long cablesupplied
F111Sensor
Hot-TapSaddlewith BallValve(page 42)
PDF Published January 2, 2011
Type F111Hot-Tap Flow Sensors
Digiflow® FlowX3 Technical Catalogue 19
The Type F3.05 No-Flow Switches offer reliable pumpprotection. They open an electrical circuit when flowvelocity decreases below 0.15 m/s (0.5 ft./sec.) and arenot adjustable. The unique Halar® paddle wheel design isnot available from other manufacturers.
Body Materials: CPVC, PVDF, 316L Stainless Steel, Brass
Rotor: ECTFE (Halar®)
Shaft & Bearings: Ceramic
Seals: EPDM, Viton®
Pipe Sizes: 1/2"– 24" in two sensor lengths, L0 or L1See Installation Fittings (pages 39–41)
Velocity Trip Point:0.15 m/s (0.5 ft./sec.) not adjustable. The normallyopen contact closes after a delay of 2 seconds.
Visual Flow Bicolour LEDIndication: Red = No Flow, Green = Flow
Relay Output: Mechanical SPST contact,No Flow = open contact1A @ 24 VDC, 0.5A @ 125 VAC,0.1A @ 230 VAC
■ Features• Reliable Pump Protection – Typical applications are to
protect a pump from running dry or pumping against aclosed valve in the main pipe.
• Visual Flow Indication – Red/green bicolour LED
■ TechnicalSupply Voltage: 12 to 24 VDC, regulated
Current Consumption: < 50 mA
Electrical Class: NEMA 4, 4X (IP65) 4-pole DIN 43650
Maximum % Solids: 10% with particle size notexceeding 0.5 mm cross sectionor length
Max. OperatingPressure/Temperature: See chart on page 43
■ Installation Guidelines• See page 44 for preferred installation positions.
No minimum length of straight pipe is required.
■ Item Numbers• See page 24
■ Wiring
+12/24VDC
Relay
Relay
Ground
L0 = 68.3 mm (2.69")L1 = 98.5 mm (3.88")
Bicolour LEDred = no flowgreen = flow
DIN Connector4 pole(cable not supplied)
PDF Published January 2, 2011
Type F3.05No-Flow Switches
The Type F3.15 and ULF3.15 Adjustable Flow Switchestrip an electrical relay when flow velocity decreases belowa preset trip point. This unique unit doubles as a flowsensor. The pulse output (with frequency proportional toflow velocity) may be fed to a FlowX3 flowmonitor/transmitter or batch controller.
Body Materials: CPVC, PVDF, 316L Stainless Steel, Brass
Rotor: ECTFE (Halar®)
Shaft & Bearings: Ceramic
Seals: EPDM, Viton®
Pipe Sizes: 1/2"– 24" in two sensor lengths, L0 or L1See Installation Fittings (pages 39–41)
Velocity Trip Point: 0.15 to 1 m/s (0.5 to 3 ft./sec.),freely adjustable
■ Type F3.15
■ Type ULF3.15
Sensor Body: POM ECTFE (Halar®)
Rotor: POM ECTFE (Halar®)
Shaft: Corepoint®† Sapphire
Bearings: – Sapphire
O-Ring: Viton® Viton® or Kalrez®
Connection: Inline 1/4" BSP male threaded
Flow Trip Point: Freely adjustableULF01 Model: 1.5 to 21 l/hr (0.0066 to 0.09 USGPM)ULF03 Model: 6 to 53 l/hr (0.0264 to 0.23 USGPM)
† Corepoint® is a steel alloy.
Polyoxymethylene(POM) Version
ECTFEVersion
Bicolour LEDred = no flowgreen = flow
Bicolour LEDred = no flowgreen = flow
F3.15
ULF3.15
20 Digiflow® FlowX3 Technical Catalogue
■ Connectable FlowX3 Instruments
FlowX3 Instruments*Instrument Mounting
Panel or Wall F9.02, F9,03, F9.50, F9.51
* Power supply can be fed from FlowX3 instruments.
■ Features
• Reliable Pump Protection – Typicalapplications are to protect a pump fromrunning dry or pumping against a closedvalve in the main pipe.
• SPDT Relay with Adjustable Trip Point
• Includes Open Collector (frequency)Pulse Output – Useable for continuous flowrate monitoring
■ Item Numbers• See page 24
PDF Published January 2, 2011
Type F3.15and ULF3.15AdjustableFlow Switches
Power Supply
Sensor
RelayOutput
Calibration Trimmer for MIN alarm set point1Switch 1 switches between calibration (OFF) and operating (ON)Switch 2 is used to select the switch delay (3 or 10 sec.)
2
■ Wiring & Operation (Cable not supplied)
■ Technical – GeneralSupply Voltage: 12 to 24 VDC, regulated
Current Consumption: 10 mA max.
Output Signal: Square wave (pulse)
Output Frequency: 45 Hz per m/s nominal(13.7 Hz per ft./sec.)
Transistor Output: NPN open collector
Relay Output: Mechanical SPDT contact,3A @ 30 VDC, 3A @ 250 VACresistive load
Switch Delay: 3 or 10 seconds, selectable
Enclosure: NEMA 4, 4X (IP65) PVC housing
Ambient Temperature: 0 to 60oC (32 to 140oF)
■ Technical – Type F3.15Viscosity Range: 0.5 to 20 centistokes. Factory
calibration is required if outsidethis range, up to 40 centistokesmaximum.
Maximum % Solids: 10% with particle size notexceeding 0.5 mm cross sectionor length
Max. OperatingPressure/Temperature: See chart on page 43
■ Technical – Type ULF3.15Viscosity Range: 1 to 10 centistokes
Maximum % Solids: Clean services only, no solidsare recommended
Working Temperature: –10 to 80oC (14 to 176oF)
Working Pressure: 5 bar (70 psi) max. @ 22oC (72oF)
F3.15
ULF3.15
L0 = 68.3 mm (2.69")L1 = 98.5 mm (3.88")
Digiflow® FlowX3 Technical Catalogue 21
■ Item Numbers• See page 24
(See InstructionManual for detailedwiring information)
PDF Published January 2, 2011
Type F3.15and ULF3.15AdjustableFlow Switches
The Type F3.30 and ULF3.30 are 3-wire flowtransmitters with a 4 to 20 mA analog output. Forapplications not requiring visual flow indication, they arean economical alternative to an F9.00 instrument/sensordirect mount assembly.
■ Features
• Large Selection of Sensor Materials – Includingstainless steel and brass.
• Corrosion Resistant PVC Electronics Housing
• Ceramic Shaft and Bearings – Provide long life onservices containing grit
• Self Cleaning Design – Lower maintenance
Body Materials: CPVC, PVDF, 316L Stainless Steel, Brass
Rotor: ECTFE (Halar®)
Shaft & Bearings: Ceramic
Seals: EPDM, Viton®
Pipe Sizes: 1/2"– 24" in two sensor lengths, L0 or L1See Installation Fittings (pages 39–41)
Flow Ranges: See page 45
Sensor Body: POM ECTFE (Halar®)
Rotor: POM ECTFE (Halar®)
Shaft: Corepoint®† Sapphire
Bearings: – Sapphire
O-Ring: Viton® Viton® or Kalrez®
Connection: Inline 1/4" BSP male threaded
Flow Range:ULF01 Model: 1.5 to 100 l/h (0.0066 to 0.44 gpm)ULF03 Model: 6 to 250 l/h (0.0264 to 1.10 gpm)
† Corepoint® is a steel alloy.
PolyoxymethyleneVersion (POM)
ECTFEVersion
■ Type F3.30
■ Type ULF3.30
F3.30
ULF3.30
22 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Type F3.30and ULF3.30Blind Transmitters
■ Item Numbers• See page 24
Power Supply
Sensor
Calibration SwitchThe full scale range, corresponding to 4 to 20 mA output is setby means of switches 2, 3 and 4.Switch 1 is used to set a filter on the current output.
1
■ Wiring & Operation (Cable not supplied)
F3.30
ULF3.30
L0 = 68.3 mm (2.69")L1 = 98.5 mm (3.88")
Digiflow® FlowX3 Technical Catalogue 23
■ Technical – GeneralSupply Voltage: 12 to 24 VDC regulated
Current Consumption: < 50 mA
Max. Loop Impedance: 800W @ 24 VDC, 300W @ 12 VDC
Enclosure: NEMA 4, 4X (IP65) PVC housing
Repeatability: ± 0.5% of full scale
Output: 4 to 20 mA, 3 wire
■ Technical – Type F3.30 Accuracy: < ± 1% of reading value after field
calibration or ± 0.75% of full scale
Viscosity Range: 0.5 to 20 centistokes. Factorycalibration is required if outsidethis range, up to 40 centistokes maximum.
Maximum % Solids: 10% with particle size not exceeding0.5 mm cross section or length
Ambient Temperature: 0 to 60oC (32 to 140oF)
Max. OperatingPressure/Temperature: See chart on page 43
■ Technical – Type ULF3.30Accuracy: < ± 2% of reading value after field
calibration or ± 1% of full scale
Viscosity Range: 1 to 10 centistokes
Maximum % Solids: Clean service only, no solidsrecommended
Ambient Temperature: 0 to 60oC (32 to 140oF)
Working Temperature: –10 to 80oC (14 to 176oF)
Working Pressure: 5 bar (70 psi) max. @ 22oC (72oF)
(See Instruction Manual for detailed wiring information)
PDF Published January 2, 2011
Type F3.30and ULF3.30Blind Transmitters
NEMA 4, 4X(IP65)
ElectricalClass
SensorLengthO-Ring CPVC PVDF 316L SS
Item No.
Type ULF3.15
Type F3.15
Type F3.05
24 Digiflow® FlowX3 Technical Catalogue
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
F3.05.H.01
F3.05.H.02
F3.05.H.03
F3.05.H.04
F3.05.H.05
F3.05.H.06
F3.05.H.07
F3.05.H.08
F3.05.H.09
F3.05.H.10
F3.05.H.11
F3.05.H.12
NEMA 4, 4X(IP65)
ElectricalClass
SensorLengthO-Ring CPVC PVDF 316L SS
Item No.
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
F3.15.H.01
F3.15.H.02
F3.15.H.03
F3.15.H.04
F3.15.H.05
F3.15.H.06
F3.15.H.07
F3.15.H.08
F3.15.H.09
F3.15.H.10
F3.15.H.11
F3.15.H.12
Brass
F3.15.H.13
F3.15.H.14
F3.15.H.15
F3.15.H.16
NEMA 4, 4X(IP65)
ElectricalClass Flow Trip Point (l/hr) POM† / Viton® ECTFE* / Viton® ECTFE* / Kalrez®
1.5 – 21
6 – 53
ULF3.15.01.0
ULF3.15.03.0
ULF3.15.01.2
ULF3.15.03.2
ULF3.15.01.3
ULF3.15.03.3
Type F3.30
Type ULF3.30
NEMA 4, 4X(IP65)
ElectricalClass
SensorLengthO-Ring CPVC PVDF 316L SS
Item No.
EPDM
Viton®
EPDM
Viton®
L0
L0
L1
L1
F3.30.H.01
F3.30.H.02
F3.30.H.03
F3.30.H.04
F3.30.H.05
F3.30.H.06
F3.30.H.07
F3.30.H.08
F3.30.H.09
F3.30.H.10
F3.30.H.11
F3.30.H.12
Brass
F3.30.H.13
F3.30.H.14
F3.30.H.15
F3.30.H.16
NEMA 4, 4X(IP65)
ElectricalClass Flow Range (l/hr) POM† / Viton® ECTFE* / Viton® ECTFE* / Kalrez®
1.5 – 100
6 – 250
ULF3.30.01.0
ULF3.30.03.0
ULF3.30.01.2
ULF3.30.03.2
ULF3.30.01.3
ULF3.30.03.3
† POM = "Polyoxymethylene". * ECTFE is also referred to as "Halar®".
Item No.
■ No-Flow Switches See page 19
■ Adjustable Flow Switches See pages 20 and 21
■ Flow Transmitters See pages 22 and 23
Item No.
PDF Published January 2, 2011
Item NumbersFlow Switches andBlind Transmitters
Direct Mount package consistsof a flow instrument, flowsensor and direct mounting kit.
The K9.KC1 kit includes an ABSenclosure with EPDM gasket forNEMA 4, 4X watertightinstallation, a cap and lockingring for connecting the sensorto the enclosure, plus fourscrews to attach the instrument.
DirectMountingKit
F3.01 Flow Sensor(Hall or Coil Effect)
DirectMountedFlowInstrument
Panel Mount instrument consistsof a flow instrument and thepanel mounting kit.
The K9.KP1 kit includes a metalmounting clip and a neoprenegasket for NEMA 4, 4Xwatertight installation. Themonitor fits a standard 1/4 DIN(96 mm x 96 mm or 3.78" x3.78") panel cutout.
Panel Mounting
Wall Mount instrument consistsof a flow instrument and thewall mounting kit.
The F9.KW1 kit includes an ABSplastic NEMA 4, 4X surfacemount enclosure, EPDM gasketfor NEMA 4, 4X watertightinstallation, plus screws forattaching the instrument.
The F9.KW2 includes in additiona 110/230 VAC to 24 VDC powersupply mounted inside the ABSenclosure to provide aregulated low voltage to theflow monitor.
Wall Mounting
PanelMountingKit
PanelMountFlowInstrument
WallMountingKit
WallMountFlowInstrument
Direct Mountingto Flow Sensor
Removable Cover
3 line LCD Display
• Two 12–digit Alphanumeric lines• Both Flow Rate and Total Flow
appear together
• Third line shows Output Status Icons• Backlit Display is available for
F9.00, F9.02, F9.03, F9.50, F9.51
Up to 3 LEDs
• For output status indication
1/4 DIN sizeMonolithicEnclosure
5 Button SiliconeRubber Keypad
The same instrument may be mountedthree different ways using mounting kits
Digiflow® FlowX3 Technical Catalogue 25
PDF Published January 2, 2011
Instruments andMounting Kits
■ Features
• Easy Set-up – Setting up the instrument iseasy using the keypad and self-explanatorymenus
• Plug-In Removable Terminals – Makesinstrument connection and removal easy
• Epoxy Encapsulated Electronics– For durability and reliability
• Auto-calibration – Automatic calculation ofK-factors
• Auto Systematic Error Compensation (ASEC)– For increased linearity and accuracy
• Output Simulation – For system testing
■ Technical
Supply Voltage:• 12 to 24 VDC ± 10% regulated• 110/230 VAC with F9.KW2 Wall Mount Kit
Sensor Input (Frequency):• Sensor Power: 3.8 to 5 VDC @ < 30 mA• Range: 0.5 to 1,000 Hz• Optically isolated from current loop• Short circuit protected
Enclosure:• NEMA 4, 4X (IP65) front• 1/4 DIN Size• Monolithic clear polycarbonate plastic with
silicone rubber keypad
For dimensions and technical specifications,see page 33.
Type F9.00 and F9.02 Flow Monitors/ Transmitters aredesigned to process the pulse signals from all FlowX3Hall Effect flow sensors. All units have a number ofcontrol outputs including 4 to 20mA, Solid State relaysand SPDT relays. They have 3 line LCD displays andLED indicators for control output status (see page 27).The instruments supply voltage to the flow sensor. Thesame instrument may be mounted in three differentways: directly to the flow sensor or remotely eitherpanel or wall mount.
■ Connectable FlowX3 Sensors
Sensor No.MountingInstrument
F9.00 F3.00.H, F3.01.H, F3.10.H, ULF.R, F111.H
F3.00.H, F3.01.H, F3.10.H, F3.15.H,F9.02 ULF.H, ULF3.15.H, F3.80, F111.H,
F3.60M, F3.63M
■ Output Signals
Wiring4 – 20 mARelay(SPDT)*
ItemNo.
F9.00 1 2 Wire 1 – 1
F9.02 1 3/4 Wire 2 1 3
† User selectable as MIN alarm, MAX alarm, Pulse Out, WindowFrequency Out or Off.
* User selectable as MIN alarm, MAX alarm, Pulse Out, Window or Off.
††Backlit display not available for F9.00 if 4 – 20 output is connected
No. ofLEDs
F9.02
26 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Type F9.00& F9.02Flow Monitor/Transmitters
Solid StateRelay†
■ Backlit Displays††
Flow Rate – 5 digitsResettable Total – 6 digitsNon-Resettable Total – 10 digits
■ Output Specifications
F9.00 – 2 Wire:
4 to 20 mA Output: Isolated, fully adjustable and reversible• Not available when backlit configuration is on• Maximum Loop Impedance: 150W @ 12 VDC,
330W @ 18 VDC, 600W @ 24 VDC
One Solid State Relay (S.S.R.) Output with LED display:• User selectable as MIN alarm, MAX alarm, Pulse Out,
Frequency Out, Window or Off• Optically isolated, 50 mA maximum sink, 24 VDC maximum
pull-up voltage• Maximum pulses per minute: 300• Hysteresis: Adjustable
F9.02 – 3/4 Wire:
4 to 20 mA Output: Isolated, fully adjustable and reversible• Maximum Loop Impedance: 150W @ 12 VDC,
330W @ 18 VDC, 600W @ 24 VDC
Two Solid State Relay (S.S.R.) Outputs with LED display:• User selectable as MIN alarm, MAX alarm, Pulse Out,
Frequency Out, Window or Off• Optically isolated, 50 mA maximum sink, 24 VDC maximum
pull-up voltage• Maximum pulses per minute: 300• Hysteresis: Adjustable
One Relay Outputs with LED display:• User selectable MIN alarm, MAX alarm, Pulse Out, Window
or Off• Mechanical SPDT contact• Rated maximum: 3A @ 30 VDC, 3A @ 250 VAC resistive load• Maximum pulses per minute: 300• Hysteresis: Adjustable
Power Supply
Sensor
Solid StateRelay Output
Power Supply
Sensor
RelayOUT 1
Solid StateRelay Output
Solid StateRelay Output
F9.02
F9.00
■ Wiring – Rear Terminal View(See Instruction Manual for detailed wiring information)
Digiflow® FlowX3 Technical Catalogue 27
PDF Published January 2, 2011
Type F9.00& F9.02
■ Technical
Supply Voltage:• 12 to 24 VDC ± 10% regulated• 110/230 VAC with F9.KW2 Wall Mount Kit
Sensor Input (Frequency):• Sensor Power: 3.8 to 5 VDC @ < 30 mA• Range: 0.5 to 1,000 Hz• Optically isolated from current loop• Short circuit protected
Enclosure:• NEMA 4, 4X (IP65) front• 1/4 DIN Size• Monolithic clear polycarbonate plastic with
silicone rubber keypad
For dimensions and more technicalspecifications, see page 33.
Type F9.03 Flow Monitor/Transmitter is designed toprocess the pulse signals from two FlowX3 Hall Effectflow sensors or two Type F3.60.M Electromagnetic FlowTransmitters at the same time. Control outputs are two 4to 20 mA, one open collector (pulse) and two relays. Twoanalog signals and relay outputs can correspond to any twoof three flow measurements: Flow A (sensor A), Flow B(sensor B) or Delta Flow (flow difference between flowsensor A and B). One open collector output can be usedfor one of the three flow measurements. The instrumentsupplies voltage to the flow sensors.
The unit has a 3 line LCD display and LED indicators forthe control output status (see page 27). Mounting may beeither direct, panel or wall mount using different kits.
■ Displays
Flow Rate – 5 digitsResettable Total – 6 digitsNon-Resettable Total – 10 digits
■ Connectable FlowX3 Sensors
Sensor No.InstrumentMounting
Direct F3.01.H
F3.00.H, F3.15.H, ULF3.15HPanel or Wall F3.10.H, F3.80, F111.H, ULF.H,
F3.60.M, F3.63M
■ Output Signals‡
4 – 20 mAOpenCollector†
Relay(SPDT)*
ItemNo.
F9.03 2 1 2 3
‡ Associated to Flow A, Flow B or Differential Flow.† User selectable as MIN alarm, MAX alarm, Pulse Out,
Frequency Out or Off.* User selectable as MIN alarm, MAX alarm, Pulse Out or Off.
No. ofLEDs
■ Features
• Easy Set-up – Setting up the instrument iseasy using the keypad and self-explanatorymenus
• Plug-In Removable Terminals – Makesinstrument connection and removal easy
• Epoxy Encapsulated Electronics– For durability and reliability
• Auto-calibration – Automatic calculation ofK-factors
• Auto Systematic Error Compensation (ASEC)– For increased linearity and accuracy
• Output Simulation – For system testing
28 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Type F9.03Dual Input FlowMonitor/Transmitter
■ Output Specifications
Two 4 to 20 mA Loop Outputs correspond to (2) ofFlow A, Flow B or �Flow: Isolated, fully adjustableand reversible• Maximum Loop Impedance: 150W @ 12 VDC,
330W @ 18 VDC, 600W @ 24 VDC
One Open Collector Output with LED display correspondsto (1) of Flow A, Flow B or �Flow:• User selectable as MIN alarm, MAX alarm, Pulse Out,
Frequency Out, Window or Off• Optically isolated, 50 mA maximum sink, 24 VDC maximum
pull-up voltage• Maximum pulses per minute: 300• Hysteresis: Adjustable
Two Relay Outputs with LED display correspond to (2) ofFlow A, Flow B or �Flow:• User selectable MIN alarm, MAX alarm, Pulse Out, Window
or Off• Mechanical SPDT contact• Rated maximum: 3A @ 30 VDC, 3A @ 250 VAC resistive load• Maximum pulses per minute: 300• Hysteresis: Adjustable
Power Supply A
B
C
D
E
F
Sensor A
Sensor B
OpenCollector
Output
RelayOUT 1
RelayOUT 2
■ Wiring – Rear Terminal View(See Instruction Manual for detailed wiring information)
Digiflow® FlowX3 Technical Catalogue 29
PDF Published January 2, 2011
Type F9.03Dual Input FlowMonitor/Transmitter
The Type F9.50 Batch Controller is designed foraccurate and reliable batching or blending of liquids. Itaccepts a pulse input from all FlowX3 Hall Effect flowsensors.
The instrument offers complex control capability. It iseasy to calibrate and operate using self explanatorymenus. A Simple and Advanced mode are availablewhich allow the choice of quick calibration and basicoutputs or a more customized and detailed controlset-up. Four separate control outputs (two relays, anopen collector and an analog output) allow the operatorto set-up for accurate batch operation. State-of-the-artelectronic design ensures performance and reliability.
Flow sensor supply voltage is provided by the controller.The same instrument may be mounted in three differentways: directly to the flow sensor or remotely eitherpanel or wall mount.
The Type F9.51 Batch Controller stores up to 10 batchvalues.
■ Features
• Modular Design – The same instrumentmay be mounted in three different waysusing mounting kits. See page 25.
• Simple or Advanced Operating Modes
• External Start, Stop and Resume
• Two-Stage Shutdown Control (F9.50)
• Permanent and Resettable Totalizer
• Auto-Calibration
• Auto Systematic Error Compensation (ASEC)– For increased linearity and accuracy
• Automatic or Manual Overrun Compensation (F9.50)
• Overrun Alarm
• Missing Signal Alarm
• Count-up or Count-down BatchIndication (F9.50)
• End of Batch Pulse Output
• Output Simulation for System Testing
• Advanced Valve Control (F9.50)
• Password Protection
• Stores up to 10 Batches (F9.51)
■ Connectable FlowX3 Sensors
SensorInstrumentMounting
Direct F3.01.H
Panel or Wall F3.00.H, F3.15.H, ULF.H, ULF3.15HF111.H, F3.10, F3.80, F3.60M, F3.63M
LEDs
■ Output Signals
Wiring4–20mA*
Solid StateRelayOutput†
Relay(SPDT)††
No. ofBatches
ItemNo.
F9.50 1 3/4 Wire 1 2 3 1
F9.51 – 3/4 Wire – 2 3 10
† User selectable as START Batch, END Batch or Off.†† OUT1 – Batch: Batch in Progress indication
OUT2 – Option: User selectable Two-Stage Shutdown, Endof Batch, Overrun Alarm or Missing Signal Alarm.
* User selectable as Valve Control or Degree of Batch Completion.
Direct Panel Wall
30 Digiflow® FlowX3 Technical Catalogue
■ Backlit Displays
Batch in Progress – 6 digitsFlow Rate – 5 digitsResettable Total – 6 digitsNon-Resettable Total – 10 digits
PDF Published January 2, 2011
Type F9.50/F9.51Batch Controllers
Power Supply
Sensor
Solid StateRelay
Output
RelayOUT 1
RelayOUT 2
RemoteControl
■ Wiring■ Technical
Supply Voltage:• 12 to 24 VDC ± 10% regulated• 110/230 VAC with F9.KW2 Wall Mount Kit
Sensor Input (Frequency):• Sensor Power: 3.8 to 5 VDC @ < 30 mA• Range: 0.5 to 1,000 Hz• Optically isolated from current loop• Short circuit protected
Enclosure:• NEMA 4, 4X (IP65) front• 1/4 DIN Size• Monolithic clear polycarbonate plastic with silicone
rubber keypad
Operating Temperature:• –10 to 70oC (14 to 158oF)
Storage Temperature:• –15 to 80oC (5 to 176oF)
For dimensions and more technical specifications, see page 33.
■ Output Specifications
4 to 20 mA Output (F9.50):• Isolated, fully adjustable and reversible• Maximum Loop Impedance: 150W @ 12 VDC,
330W @ 18 VDC, 600W @ 24 VDC• User Selectable as Valve Control or Batch Completion (F9.50)
Solid State Relay (F9.50) Output with LED display:• User selectable as START Batch, END Batch or Off• Optically isolated, 50 mA maximum sink,
24 VDC maximum pull-up voltage• Maximum pulses per minute: 300
Two Relay Outputs with LED display:• OUT1 – Batch: Batch in progress indication• OUT2 – Option: User selectable Two-Stage Shutdown (F9.50),
End of Batch, Overrun Alarm or Missing Signal Alarm• Mechanical SPDT contact• Expected Mechanical Life (minimum operations): 107
• Expected Electrical Life (minimum operations): 105
NO/NC switching capacity8A @ 240 VAC
• Rated maximum: 3A @ 30 VDC or3A @ 250 VAC resistive load
• Maximum pulses per minute: 300• Hysteresis: Adjustable
Additional Functions:• External Start, Stop and Resume
REAR TERMINAL VIEW(See Instruction Manual for detailed wiring information)
Digiflow® FlowX3 Technical Catalogue 31
PDF Published January 2, 2011
Type F9.50/F9.51Batch Controllers
The Type F9.20 Battery Powered Flow Monitor isdesigned to process the pulse signals from all FlowX3Coil Effect and ULF Reed Effect flow sensors. The longlife Lithium AA batteries make this unit ideal for remotelocations or for use as a portable meter. The two-linealphanumeric 12 digit display shows flow rate and totalflow together. The instrument supplies voltage to theflow sensor.
■ Features• Battery Operated
Five year estimated battery lifeBattery status indicatorBattery can be exchanged without lossof calibration data
• Ideal for Remote Locations – Or for use asa portable flow meter
• Auto-Calibration – Automatic calculation ofK-factors
• Modular Design – The same instrument maybe mounted in three different ways usingmounting kits. See page 25.■ Technical
Power Supply:• (2) 3.6 V Lithium Batteries• AA standard size (0.58" diameter x 2" long)
Sensor Input (Frequency):• Sensor Power: 3.6 Volt• Range: 1 to 500 Hz
Enclosure:• NEMA 4, 4X (IP65) front• 1/4 DIN Size• Monolithic clear polycarbonate plastic with
silicone rubber keypad
Dimensions:• See page 33
More technical specifications on page 33
Sensor
■ Connectable FlowX3 Sensors
Sensor No.InstrumentMounting
Direct F3.01.C
Panel or Wall F3.00.C, ULF.R, F111.C
Direct Panel Wall
■ Displays
Flow Rate – 5 digitsResettable Total – 6 digitsNon-Resettable Total – 10 digits
32 Digiflow® FlowX3 Technical Catalogue
■ Rear Terminal View(See Instruction Manual for detailed wiring information)
PDF Published January 2, 2011
Type F9.20Battery PoweredFlow Monitor
Technical – F9.00, .02, .03, .20, .50 and .51 Instruments
The monitor fits into a standard 1/4 DIN (96 mm x 96 mm or 3.78" x 3.78") panel cutout.
Display:• 3 Line LCD: 2 x 12 alphanumeric characters
and 1 icon line• Update rate: 1 sec.• Contrast: User adjustable with 5 levels
Enclosure:• NEMA 4, 4X (IP65) front• 1/4 DIN size• Clear polycarbonate plastic
Relative Humidity:• 0 to +95% non condensing
Operating Temperature:• –10 to 70oC (14 to 158oF)• Minimum recommended operating
temperature for instruments is –10oC (14oF).They will tolerate short term drops down toas low as –30oC (–22oF) but are not operableat these temperatures.
Storage Temperature:• –15 to 80oC (5 to 176oF)
Dimensions – Instruments with Mountings
Digiflow® FlowX3 Technical Catalogue 33
■ Direct Mount ■ Wall Mount
■ Panel Mount
PDF Published January 2, 2011
InstrumentsTechnical & Dimensions
Types F3.60M, F3.63M and F3.61M are insertion typeblind transmitters which generate a 4 to 20 mA analogoutput signal and a digital open collector output signalproportional to flow velocity. These units will handlesolids, ferrous particles and higher viscosity fluids. Theyare not affected by ferrous particles which can foul themagnetic rotors of the paddle wheel type. The TypeF3.60M and F3.60M are the standard installation versions.F3.61M is the Hot-Tap version of the F3.60M. Hot-Tapfunction allows sensor insertion and removal while the pipeis pressurized. The units may be connected to FlowX3instruments.
Body Materials: 316L SS, PVDF CuNi Alloy, PVDF 316L SS, PEEK†
Electrodes: 316L SS CuNi Alloy 316L SS
Seals: EPDM, Viton® EPDM, Viton® Viton®
Pipe Sizes: 1/2" to 24" (consult Chemline for larger pipe sizes)
Installation Fittings: See pages 41 to 42
■ Connectable FlowX3 Instruments
FlowX3 InstrumentsInstrumentMounting
Panel or Wall F9.02, F9.03, F9.50, F9.51
F3.60M/F3.63M
F3.61M Hot-Tap Version
34 Digiflow® FlowX3 Technical Catalogue
■ Unique Features
Velocity0.05 to 8 m/s 0.15 to 8m/s 0.05 to 8 m/s
Measurement(0.15 to 25 ft/sec.) (0.5 to 25 ft/sec.) (0.15 to 25 ft/sec.)
Range
Full Scale 5 m/s (16.4 ft/sec.) standardRange (others available upon request)
DirectionalFlow Bidirectional Monodirectional BidirectionalMeasurement
Installation Fittings
Tees: 1/2" to 1-1/2" 1/2" to 1-1/2" –PVC, PP, PVDF PVC, PP, PVDF
Bolt-OnSaddles: 2" to 12" 2" to 12" –
Strap-OnSaddles: 3" to 18" 3" to 18" 3" to 18"
Weld-OnAdaptors: 1-1/2" to 24" 1-1/2" to 24" 16" to 24"
F3.60MF3.61MHot-Tap VersionF3.63M
Side Pressure PortBall valve andsaddle not shown.
■ Common Features• Low maintenance – No moving parts, no
wear• Will handle slurries and high viscosity
conductive and homogeneous fluids• Easier to install and remove compared to
full port magmeters• Less costly than full port magmeters• For both plastic and metal pipes
PDF Published January 2, 2011
Type F3.60M/F3.63M & F3.61MElectromagneticFlow Transmitters
StandardSea WaterApplications
HighTemperatureApplications
† PEEK = Polyether Ether Ketone
NewNew
■ TechnicalPower Supply: 12 to 24 VDC ± 10% regulated
Maximum CurrentConsumption: 150 mA
Analog Output Signal: 4 to 20 mAMaximum loop resistance (load): 600W@ 24 VDCPositive or negative flow indication
Frequency Output: Open Collector NPNFrequency: 0 to 500 HzMaximum Current: 50 mAMaximum Voltage: 24 VDCFrequency output can be fed to FlowX3F9.02, F9.03, F9.50 or F9.51 instruments
Digital Output Signal: Open Collector NPN(for F3.60M & F3.61M) Maximum Current: 50 mA
Maximum Voltage: 24 VDCFrequency: 0 to 500 HzFlow Direction: 0 VDC arrow-wise
+ VDC anti arrow-wise
Enclosure: NEMA 4, 4X (IP65), PC and PVC materials
Accuracy/Linearity: 1% of measured value + flow rate valueat +0.01m/s (0.03 ft/sec.)
Repeatability: 0.5% of measured value
Velocity Range: F3.60 & F3.61: 0.05 to 8 m/s (0.15 to 25 ft./sec.)F3.63: 0.15 to 8 m/s (0.5 to 25 ft./sec.)
Standard Full Scale Setting: 5 m/s (16 ft./sec.) factory settings up to
8 m/s AWS available
Liquid: Conductive and homogeneous liquidsor slurries
Conductivity of Fluids: Minimum 20 mS (microSiemens)
Ambient Temperature: 0 to 60oC (32 to 140oF)
Working Temperature: PVDF bottom (standard & sea water) versions: –10 to 60oC (14 to 140oF)PEEK High Temperature Version:–10 to 150oC (14 to 302oF)
Maximum Pressure: 232 psi at 77oF (16 bar at 25oC)124 psi at 140oF (8.6 bar at 60oC)
Ground Protection: Low impedance ground required– below 10W
Standards & Approvals: Manufactured under ISO 9001 (Quality),ISO 14001 (Environmental), CE
F3.61MFlowTransmitter
Hot-TapSaddlewith BallValve(page 42)
■ Dimensions
L0 = 68.5 mm (2.70")L1 = 98.5 mm (3.88")
F3.60M/F3.63M
F3.61M
Digiflow® FlowX3 Technical Catalogue 35
PDF Published January 2, 2011
Type F3.60M/F3.63M & F3.61MElectromagneticFlow Transmitters
1 + VDC2 + LOOP3 – LOOP4 – VDC
Power Supply
24 VDC +24 VDC –
Connection Only Frequency Output Version
A
■ Power/Loop Wiring Diagrams
■ Open Collector Wiring Diagrams
FlowX3MonitorTerminals
Internal PLCConnection
Connection to a PLC/Instrument with ONE Separate Power Supply
Connection to FlowX3 Instruments (Only withFrequency Output)
Power Supply
24 VDC +
24 VDC –
Sensor
GND 5
IN 6
V+ 7
PCL/Instrument
4-20mA Loop –4-20mA Loop +
1 + VDC2 + LOOP3 – LOOP4 – VDC
A
Connection with a built-in Power Supply (3-Wire Connection)
PCL/Instrument
Power Supply +Power Supply–4-20mA Loop–4-20mA Loop +
Internal PLCConnection
Imax = 50 mA
PCL/Instrument
1 + VDC2 + LOOP3 – LOOP4 – VDC
A
5 N.C.6 N.C.7 Dir +8 Dir –9 Freq +10 Freq –CB
FlowX3MonitorTerminals
Connection to PLC NPN Open Collector Input
Connection to PLC/Instrument Digital Inputwith Separate Power Supply
5 N.C. 6 N.C.7 Dir +8 Dir –9 Freq +10 Freq –C
Imax = 50 mA
5 N.C. 6 N.C.7 Dir +8 Dir –9 Freq +10 Freq –C
Connection to a PLC/Instrument with TWO Separate Power SuppliesPower Supply
24 VDC +
24 VDC –
PCL/Instrument
4-20mA Loop –4-20mA Loop +
1 + VDC2 + LOOP3 – LOOP4 – VDC
A
Power Supply
24 VDC +
24 VDC –
Power Supply
24 VDC +
24 VDC –
PCL/Instrument
Digital INPUT –Digital INPUT +
36 Digiflow® FlowX3 Technical Catalogue
■ Principle of OperationTwo magnetic coils in the body of the instrumentgenerate a magnetic field perpendicular to the flowdirection. The magnetic field B and the flowvelocity V induce a voltage E between the twoelectrodes. The voltage E is directly proportional tothe flow velocity V:
E = K x B x VK = instrument constantB = intensity of magnetic fieldV = flow velocity of fluid
The voltage across the electrodeterminals is transmitted to anintegral converter and convertedinto an output signal of 4 to 20mA or frequency output signal.
Power Supply A
C
SensorConnections
SensorConnections
OpenCollectorOutputs
■ Rear Terminal View
For
F3.6
0 &
F3.
61 o
nly
Power Supply +Power Supply –O.C. IN +O.C. IN –
N.C. = No Connection
PDF Published January 2, 2011
Type F3.60M/F3.63M & F3.61MElectromagneticFlow Transmitters
Types F9.60M and F9.63M are insertion typemonitors/transmitters. They have 3 line LCD displays andLED indicators for control output status. Control outputsare 4 to 20 mA, open collector and SPDT relays. Theseunits will handle solids and higher viscosity fluids. Theyare not affected by ferrous particles which can foul themagnetic rotors of the paddle wheel type.
■ Connectable FlowX3 Instruments
FlowX3 InstrumentsInstrumentMounting
Panel or Wall F9.02, F9.03, F9.50, F9.51
New
F9.60M/F9.63M
■ Unique Features
Velocity Measurement 0.05 to 8 m/s 0.15 to 8m/sRange (0.15 to 25 ft/sec.) (0.5 to 25 ft/sec.)
Full Scale Range 5 m/s (16.4 ft/sec.) standard(others available upon request)
Directional FlowBidirectional MonodirectionalMeasurement
Installation Fittings
Tees: 1/2" to 1-1/2" 1/2" to 1-1/2"PVC, PP, PVDF PVC, PP, PVDF
Bolt-On Saddles: 2" to 12" 2" to 12"
Strap-On Saddles: 3" to 18" 3" to 18"
Weld-On Adaptors: 1-1/2" to 24" 1-1/2" to 24"
(Consult Chemline for larger pipe sizes.)
F9.60M F9.63M
■ Principle of OperationTwo magnetic coils in the body of the instrumentgenerate a magnetic field perpendicular to the flowdirection. The magnetic field B and the flowvelocity V induce a voltage E between the twoelectrodes. The voltage E is directly proportionalto the flow velocity V:
E = K x B x VK = instrument constantB = intensity of magnetic fieldV = flow velocity of fluid
The voltage across the electrode terminals istransmitted to an integral converter and convertedinto an output signal of 4 to 20 mA or frequencyoutput signal.
Digiflow® FlowX3 Technical Catalogue 37
■ Common Features• Low maintenance – No moving parts, no
wear• Will handle slurries and high viscosity
conductive and homogeneous fluids• Easier to install and remove compared to
full port magmeters• Less costly than full port magmeters• For both plastic and metal pipes
PDF Published January 2, 2011
Type F9.60M/F9.63M Electromagnetic FlowMonitor/Transmitters
Body Materials: 316L SS, PVDF CuNi Alloy, PVDF 316L SS, PEEK†
Electrodes: 316L SS CuNi Alloy 316L SS
Seals: EPDM, Viton® EPDM, Viton® Viton®
Pipe Sizes: 1/2" to 24" (consult Chemline for larger pipe sizes)
Installation Fittings: See pages 41 to 42
StandardSea WaterApplications
HighTemperatureApplications
† PEEK = Polyether Ether Ketone
New New
■ TechnicalEnclosure: NEMA 4, 4X (IP65), PC case and EPDM gasketKeypad: 5 button silicone rubberLinearity: F9.60: ± 1% of measured value + 1.0 cm/s
(0.03 ft./sec.)F9.63: ± 2% of measured value + 1.0 cm/s(0.03 ft./sec.)
Repeatability: 0.5% of measured valueVelocity Range: F9.60: 0.05 to 8 m/s (0.15 to 25 ft./sec.)
F9.63: 0.15 to 8 m/s (0.5 to 25 ft./sec.)Liquid: Conductive and homogeneous liquids
or slurriesConductivity of Fluids: Minimum 20 mS (microSiemens)Ambient Temperature: –10 to 70oC (14 to 158oF)Working Temperature: PVDF bottom (standard & sea water) versions:
–10 to 60oC (14 to 140oF)PEEK High Temperature Version:–10 to 150oC (14 to 302oF)
Maximum Pressure: 230 psi at 77oF (16 bar at 25oC)124 psi at 140oF (8.6 bar at 60oC)
Standards & Approvals: Manufactured under ISO 9001 (Quality),ISO 14001 (Environmental), CE
Power Supply: 12 – 24 VDC ± 10% regulatedMaximum CurrentConsumption: 300 mAGround Protection: Low impedance ground required – below 10WAnalog Output Signal: 4 to 20 mA, isolated, fully adjustable
Maximum loop impedance: 600W@ 24 VDCPositive, negative or bi-directional flowindication
Open Collector Output: Type: Solid state NPN, optically isolated(OPT) Output Modes: Off, MIN, MAX, Window (F9.63)
Volumetric PulseMaximum Fequency: 500 HzMaximum Pull-up Voltage: 24 VDCMaximum Current: 50 mA, current limitedHysteresis: User selectable plus adjustable timer delay
Relay Output : Type: Mechanical SPDT contact(OUT1 and OUT2) Output Modes: Off, MIN, MAX, Volumetric,
Window (F9.63)PulseMaximum Voltage Rating: 3A @ 30 VDC,3A @ 250 VAC resistive loadHysteresis: User selectable plus adjustable timerdelayExpected Mechanical Life (min. operations): 107
Expected Electrical Life (min. operations):1. N.O. switching capacity 5A @ 250 VAC: 5 x 104
2. N.C. switching capacity 2A @ 250 VAC: 2 x 105
■ Dimensions
L0 = 68.5 mm (2.70")L1 = 98.5 mm (3.88")
F9.60M/F9.63M
F9.60M/F9.63M
Power Supply A
E
C
D
SensorConnections
SensorConnections
Open CollectorOutputs
Out 1 Relay
Out 2 Relay
■ Rear Terminal View
38 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Type F9.60M/F9.63MElectromagnetic FlowMonitor/Transmitters
Digiflow® FlowX3 Technical Catalogue 39
DIMENSIONS INCHES
Size d E h HHole Diameter
in PipeSensorLength
MaterialsSaddle / Insert / Seal
PipeSize B
PVC / EPDM / SocketE Item No. H Z L
1/2"
3/4"
1"
1-1/4"
1-1/2"
2.87
3.15
3.19
3.31
3.25
d0.85
1.06
1.33
1.67
1.91
2.09
2.44
2.80
3.31
3.86
TEA005ES
TEA007ES
TEA010ES
TEA012ES
TEA015ES
4.92
5.51
6.04
7.34
8.15
3.15
3.50
3.78
4.80
5.39
0.89
1.00
1.13
1.26
1.38
L0
L0
L0
L0
L0
PP / EPDM / ThreadedItem No. H Z L
TEB005ET
TEB007ET
TEB010ET
TEB012ET
TEB015ET
4.45
4.96
5.49
6.69
7.83
3.05
3.54
3.71
4.84
5.59
0.70
0.71
0.89
0.93
1.12
PVDF / Viton® / SocketItem No. H Z L
TEK005VS
TEK007VS
TEK010VS
TEK012VS
TEK015VS
4.37
4.74
5.26
6.36
7.62
3.23
3.48
3.82
4.74
5.77
0.57
0.63
0.71
0.81
0.93
SensorLength
TeesSizes: 1/2" to 1-1/2"Maximum Pressure:150 psi @ 20oC (70oF).See page 43 forworking pressures athigher temperatures.
SACC020E
SACC025E
SACC030E
SACC040E
SACC060E
SACC080E
SABC100N
SABC120N
CPVC / CPVC / EPDM†
CPVC / CPVC / EPDM†
CPVC / CPVC / EPDM†
CPVC / CPVC / EPDM†
CPVC / CPVC / EPDM†
CPVC / CPVC / EPDM†
PP / CPVC / Nitrile
PP / CPVC / Nitrile
Item No.
2"
2-1/2"
3"
4"
6"
8"
10"
12"
MaterialsSaddle / Insert / Seal
SACK020V
SACK025V
SACK030V
SACK040V
SACK060V
SACK080V
SABK100N
SABK120N
CPVC / PVDF / Viton®
CPVC / PVDF / Viton®
CPVC / PVDF / Viton®
CPVC / PVDF / Viton®
CPVC / PVDF / Viton®
CPVC / PVDF / Viton®
PP / PVDF / Nitrile
PP / PVDF / Nitrile
Item No.
L0
L0
L0
L0
L1
L1
L1
L1
4.13
4.13
4.13
4.13
4.72
4.92
3.46
3.46
3.3
3.4
3.6
4.0
5.9
6.6
8.35
8.66
4.57
5.28
5.98
6.93
9.37
13.11
15.15
15.15
2.375
2.875
3.500
4.500
6.625
8.625
10.750
12.750
1-3/8"
1-3/8"
1-9/16"
1-9/16"
1-9/16"
1-9/16"
1-9/16"
1-9/16"
Bolts, nuts and washers are carbon steel.† Viton® seals are also available.
PVC ends are ANSI socket. PP tees are female NPT and PVDF tees have DIN socket ends. Other ends available are socket, threaded or flangedin all materials, butt fusion ends (B) for PP and PVDF, and CPVC socket ends for installing PP tees in CPVC pipes.Standard elastomer seals are EPDM or Viton®. Other materials are available.
DIMENSIONS INCHES
Bolt-On SaddlesSizes: 2" to 12"Maximum Pressure:150 psi @ 20oC (70oF).Use the insert materialfor determination ofworking pressures athigher temperatures.See page 43.
PDF Published January 2, 2011
Installation Fittings for F3.0 Sensors& Magmeters
Metal Strap-On SaddlesPipe Sizes: 3" to 18"Body Materials: Cast Iron/Stainless SteelInsert Materials: CPVCNut & Washer Material: Carbon SteelSeals: Nitrile (NBR)Maximum Pressure: 150 psi @ 20oC (70oF). See page 43.
PVC / EPDMBody / Seal Material
DIMENSIONS PipeSize Min.
OD (mm) OD (inches)Max. Min. Max.
DrillingHole
ItemNo.
88
112
140
168
218
272
322
356
406
457
104
126
154
184
244
295
354
414
472
534
3.5
4.4
5.5
6.6
8.6
10.7
12.7
14.0
16.0
18.0
4.1
5.0
6.1
7.2
9.6
11.6
13.9
16.3
18.6
21.0
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
1-5/8"
h(inches)
4.64
4.92
5.15
5.63
7.56
8.26
9.01
9.72
11.22
11.97
L0
L0
L0
L0
L1
L1
L1
L1
L1
L1
SensorLength
10"
12"
11.0
12.4
15.9
18.7
4.5
4.5
10.6
12.9
PipeSize d H Z
1.0
1.1
f
L0
L0
SensorLength B1
8.3
9.6
B2
13.2
15.4
A Min.
14.3
17.0
A Max.
WFA100E
WFA120E
PVC / Viton®
WFA100V
WFA120V
PPG / EPDM
WFB100E
WFB120E
PPG / Viton®
WFB100V
WFB120V
3"
4"
5"
6"
8"
10"
12"
14"
16"
18"
MSS030N
MSS040N
MSS050N
MSS060N
MSS080N
MSS100N
MSS120N
MSS140N
MSS160N
MSS180N
Wafer FittingsPipe Sizes: 10" and 12"Body Materials: PVC, PPG (Glass Reinforced PP)Insert Material: CPVCSeals: EPDM, Viton®
Maximum Pressure @ 20oC (70oF): 10"–150 psi, 12"–115 psi
DIMENSIONS INCHES
40 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Installation Fittings for F3.0 Sensors& Magmeters
Digiflow® FlowX3 Technical Catalogue 41
Weld-On AdaptorsPipe Sizes: 1-1/2" to 24"Body Materials: 316 Stainless Steel, Carbon Steel, Brass, PE.
Other materials are available upon request.Maximum Pressure: 365 psi with metal sensor. See page 43.
3/4"
1"
1-1/4"
TES007T–NPT
TES010T–NPT
TES012T–NPT
3/4"
1"
1-1/4"
3.74
4.13
4.72
0.79
0.88
0.81
L0
L0
L0
DIMENSIONS INCHES
Pipe Size Item No. d H L
3.20
3.20
3.30
B
1.65
1.65
2.12
ESensorLength
316 Stainless Steel Threaded TeesSizes: 3/4" to 1-1/4" NPT female thread.Maximum Pressure: 365 psi with metal sensor. See page 43.
WOSS–L0
WOSS–L1
1-1/2" to 8"
10" to 24"
2.70
3.88
1.724
1.724
DIMENSIONS INCHESItem No.Pipe Size L D
L
G 1-1/4"
1.724"
ID (Inside Diameter)
W(Wall Thickness)
H = 0.12 x ID + 0.14"+ W
Weld-On Adaptors must be installedat height H level with pipe outsidediameter. H = 0.12 times the pipeinside diameter plus 0.14" plus pipewall thickness.
L0
L1
Sensor Length
PDF Published January 2, 2011
Installation Fittings for F3.0 Sensors& Magmeters
Hot-Tap Metal Strap-on SaddlesIncludes 2" brass ball valve with BSP female threaded ends toconnect directly to the F111 flow sensor. See pages 17 and 18.
Pipe Sizes: 3" to 18"Body Materials: Cast Iron/Stainless SteelSeals: Nitrile (NBR)Nut Materials: Carbon SteelMaximum Pressure: 150 psiMaximum Temperature: 80oC (176oF), same as the sensor
Hot-Tap 316L Stainless SteelWeld-On Adaptors2" ball valve is not included but is required. Consult Chemline.
DIMENSIONS
ItemNo.
PipeSize
3"
4"
5"
6"
8"
10"
12"
14"
16"
18"
MSS030N–HT
MSS040N–HT
MSS050N–HT
MSS060N–HT
MSS080N–HT
MSS100N–HT
MSS120N–HT
MSS140N–HT
MSS160N–HT
MSS180N–HT
DIMENSIONS INCHESItem No.Pipe Size Min. OD Min. Connection Size (BSP)
2.00"15.7516" WOSS–HT
Min.OD (mm) OD (inches)
Max. Min. Max.104
126
154
184
244
295
354
414
472
534
3.5
4.4
5.5
6.6
8.6
10.7
12.7
14.0
16.0
18.0
4.1
5.0
6.1
7.2
9.6
11.6
13.9
16.3
18.6
21.0
h(inches)
7.28
7.83
8.62
9.13
10.31
11.22
12.32
13.30
14.01
15.35
88
112
140
168
218
272
322
356
406
457
42 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
Installation Fittingsfor Hot-Tap F111 Sensor& F3.61M Magmeters
Digiflow® FlowX3 Technical Catalogue 43
Item No. Body Material O-Ring
PVC
PVC
CPVC
CPVC
PP
PP
PVDF
EPDM
Viton®
EPDM
Viton®
EPDM
Viton®
Viton®
PLUGAE
PLUGAV
PLUGCE
PLUGCV
PLUGBE
PLUGBV
PLUGKV
1.40"
0.50"
3.54"
1.05"
Plugs for 3.0/3.3 Flow SensorsPlugs are used to cap off tees or saddles (see pages 40-42) after asensor is removed for relocation, repair or pipeline maintenance.
Materials: PVC, CPVC, PP, PVDFSeals: EPDM, Viton®
Maximum Pressure:150 psi @ 20oC (70oF). See page 43 for working pressures at highertemperatures.
MATERIALS
For Paddle WheelFlow Sensors andInstallation FittingsTo determine maximumrecommended pressure at a giventemperature check maximumpressure ratings of the sensor as wellas installation fittings selected. Thelower pressure rating applies.
-20 -10 0 10 20 30 40 50 60 70 80 90 100 110 120 130
0
14.5
29.0
43.5
58.0
101.5
116.0
130.5
145.0
365.0
2662482302121941761581401221048668503214-4
25
10
9
8
7
6
5
4
3
2
1
0
SS and Brass
CPVC
CPVCPVDF
PVDF
PVC
Co
il Ef
fect
Sen
sors
Hal
l Eff
ect
Sen
sors
87.0
72.5
PP
PP
PVC
-30
-22
-40
-40
Temperature (oF)
Temperature (oC)
Pres
sure
(b
ar)
Pres
sure
(p
si)
■ Temperature RangesTemperature RangeMaterial
PVC: Polyvinyl Chloride 5 to 60oC (41 to 140oF)
CPVC: Chlorinated Polyvinyl Chloride 5 to 80oC (41 to 176oF)
PP: Polypropylene –20 to 80oC (–4 to 176oF)
PVDF: Polyvinylidene Fluoride –30 to 100oC (–22 to 212oF)
316L SS & Brass: Allowable temperature range is limited by the sensor’s electronicsHall Sensors: –30 to 120oC (–22 to 248oF)Coil Sensors: –30 to 100oC (–22 to 212oF)
PDF Published January 2, 2011
Working Pressure vs. Temperature
Developed Turbulent Flow
VelocityProfile
Digiflow® FlowX3 flow meters measure liquids only. No airbubbles should be present and the pipe must always be full.The sensors will not work properly in laminar or transitionalflow applications. Minimum Reynolds number required is4500.
For accurate flow measurement there must be a developedturbulent velocity profile at the sensor location. This requiresa straight run pipe with a minimum number of pipediameters distance upstream and downstream of the flowsensor. These distances depend on the type of piping element(i.e. valves, elbows, reducers etc.) causing the disturbance.
To ensure maximum accuracy, the following guidelines shouldbe observed when installing.
Flow
Figure 2Figure 1 Figure 3
OK if no sedimentsare present
OK if no air bubbles arepresent
Preferred installation if sediments* orair bubbles may be present
* Maximum % Solids: 10% with particle size not exceeding 0.5 mm cross section or length.
Inlet Outlet Inlet Outlet Inlet Outlet
5xID50xID
Valve
5xID15xID
Reducer
5xID10xID
Socket Flange
Inlet Outlet
20xID 5xID
90o Elbow
2 x 90o Elbow
Outlet
5xID25xID
Inlet Inlet
40xID
2 x 90o Elbow - 3 Dimensional
Outlet
5xID
■ Installation Positions
44 Digiflow® FlowX3 Technical Catalogue
PDF Published January 2, 2011
InstallationGuidelines
Digiflow® FlowX3 Technical Catalogue 45
Min. Flow withoutField Calibration
ULFH Hall EffectPulses per Litre
(K Factor)l/h USGPM l/h USGPM l/h USGPM Flow Range
Ends
0.035
0.066
8
15
1.5
6.0
Min. Flow withField Calibration
0.0066
0.0264
1.5–100
6–250
0.0066–0.44
0.0264–1.10
1/4"
1/4"
8,431
3,394
ULFR Reed EffectPulses per Litre
(K Factor)
2,108
848
ULF Sensors
PVDF 316L SS Minimum MaximumPPPVC
K FactorsFittingSize
0.42
0.75
1.17
1.91
2.99
22.4
39.8
62.2
102.0
159.3
212.17
135.32
89.36
48.94
42.10
225.06
139.38
94.66
51.37
43.07
–
157.06
92.84
51.52
–
Flow Range – USGPMMinimum Maximum
0.09
0.17
0.26
0.43
0.68
5.1
9.0
14.1
23.2
36.2
Flow Range – m3/h
1/2"
3/4"
1"
1-1/4"
1-1/2"
235.45
142.46
91.53
51.57
42.89
K Factors for Tee Fittings
Calibration (K) Factors & Flow RangesCalibration (K) factors are the numbers that must be programmed into the instruments. They convert the pulse output fromthe sensor (pulses per litre) to instrument readings and depend on the inside diameter of the pipe and specific fitting.Flow range is the minimum and maximum which may be measured within stated accuracy. This corresponds to the flowvelocity range of 0.15 to 8 m/s (0.5 to 25 ft./sec.).
Installation Weld-On Adaptor for Metal PipeFor Weld-On Adaptors on metal pipe, where the pipeinside diameter (ID1) is different from the valuesshown in the chart to the left, use the followingcorrection equation to calculate the new K factor (K1).
K1 = K ID 2
ID1
ID, K – Values from the chart to the leftID1 – Inside diameter of pipe
(different from sch 40 or sch 80)
K1 – New K factor to be inputted intoinstruments
( )
K FactorK Factor
Schedule 40 Schedule 80PipeSize
2"
2-1/2"
3"
4"
5"
6"
8"
10"
12"
14"
16"
18"
20"
24"
21.93
15.26
10.30
5.06
3.19
2.13
1.21
0.73
0.51
0.48
0.36
0.27
0.22
0.15
K Factors for Metal Clamp Saddlesand Weld-On Adaptors on CS or Stainless Steel Pipe
25.11
17.40
11.62
5.64
3.53
2.38
1.33
0.81
0.57
0.53
0.39
0.29
0.24
0.17
ID
1.913
2.290
2.864
3.786
4.768
5.709
7.565
9.493
11.294
12.412
14.224
16.014
17.814
21.418
ID
2.047
2.445
3.042
3.998
5.016
6.031
7.941
9.976
11.888
13.072
14.936
16.809
18.743
22.544
ID K FactorUSGPM
Flow Range
K Factor
Schedule 40 Schedule 80
FittingSize
Flow Range
2"
2-1/2"
3"
4"
5"
6"
8"
25.97
17.76
10.92
5.79
3.61
2.39
1.30
ID
2.047
2.445
3.042
3.998
5.016
6.031
7.941
m3/h
1.15 – 61
1.6 – 87
2.5 – 135
4.4 – 233
6.9 – 367
9.9 – 531
17.2 – 920
K Factors for Plastic Saddles installed on PVC Pipe
m3/h USGPM
5 – 270
7 – 380
11 – 590
19 – 1,030
30 – 1,600
44 – 2,340
76 – 4,050
1.0 – 53
1.4 – 77
2.2 – 120
3.9 – 209
6.2 – 332
8.9 – 476
15.7 – 835
4.4 – 235
6.3 – 340
9.9 – 530
17.3 – 920
27.4 – 1,460
39.3 – 2,090
68.9 – 3,680
1.913
2.290
2.864
3.786
4.768
5.709
7.565
29.74
20.25
12.36
6.47
4.00
2.68
1.46
PDF Published January 2, 2011
Calibration Factorsfor F3.0/3.3 Flow Sensors
46 Digiflow® FlowX3 Technical Catalogue
K
PVC PPFittingSize
1/2"
3/4"
1"
1-1/4"
1-1/2"
462.04
272.89
157.86
101.60
63.72
l/s
1.08
1.83
3.17
4.92
7.85
USGPM
17.12
29.00
50.24
77.98
124.42
K
510.01
294.29
178.60
105.74
67.60
F.S. l/s
0.98
1.56
2.59
4.02
6.31
USGPM
15.53
24.73
41.05
63.72
100.01
PVDF
K
510.01
294.29
178.60
105.74
67.60
F.S. l/s
0.98
1.70
2.80
4.73
7.40
USGPM
15.53
26.94
44.38
74.97
117.29
316L SS
K
–
308.21
177.84
88.85
–
F.S. l/s
–
1.62
2.81
5.63
–
USGPM
–
25.68
44.54
89.24
–
K Factors for Tee Fittings
Calibration (K) Factors & Flow RangesCalibration (K) factors are the numbers that must be programmed into the instruments. They convert the pulse output fromthe sensor (pulses per litre) to instrument readings and depend on the inside diameter of the pipe and specific fitting.F3.60M and F9.60M magmeters are factory calibrated for 5 m/s velocity. F.S. is the Flow Rate value (l/s and USGPM) at 5 m/s.
K Factor
Schedule 80PipeSize
1-1/2"
2"
2-1/2"
3"
4"
5"
6"
8"
10"
12"
14"
16"
18"
20"
24"
K Factors for Metal Clamp Saddles and Weld-On Adaptorson CS or Stainless Steel Pipe
53.74
43.60
26.22
19.02
11.28
7.43
5.10
3.00
1.91
1.35
1.12
0.83
0.62
0.52
0.36
F.S. l/s
9.30
11.47
19.07
26.29
44.32
67.30
98.13
166.92
261.85
369.82
447.32
600.96
801.28
961.54
1,373.63
USGPM
147.41
181.80
302.26
416.70
702.48
1,066.73
1,555.39
2,645.74
4,150.41
5,861.77
7,088.74
9,525.41
12,700.55
15,240.72
21,772.48
ID
1.500
1.913
2.290
2.864
3.786
4.768
5.709
7.565
9.493
11.294
12.412
14.224
16.014
17.814
21.418
ID K FactorK Factor
Schedule 40 Schedule 80FittingSize
2"
2-1/2"
3"
4"
5"
6"
8"
47.10
33.05
21.33
16.00
12.44
5.43
3.13
F.S. l/s
10.6
15.1
23.4
40.5
63.7
92.1
159.7
F.S. USGPM
168.2
240.0
371.5
641.7
1,010.1
1,460.3
2,531.7
ID
2.047
2.445
3.042
3.998
5.016
6.031
7.941
K Factors for Plastic Saddles installed on PVC Pipe
1.913
2.290
2.864
3.786
4.768
5.709
7.565
53.93
37.67
24.06
17.84
13.77
6.06
3.45
F.S. l/s
9.3
13.3
20.8
36.3
57.6
82.6
145.0
F.S. USGPM
146.9
210.3
329.4
575.5
912.7
1,308.8
2,298.3
Installation Weld-On Adaptor for Metal Pipe
For Weld-On Adaptors on metal pipe, where thepipe inside diameter (ID1) is different from thevalues shown in the chart to the left, use thefollowing correction equation to calculate thenew K factor (K1).
K1 = K ID 2
ID1
ID, K – Values from the chart to the leftID1 – Inside diameter of pipe
(different from sch 40 or sch 80)
K1 – New K factor to be inputted intoinstruments
( )
PDF Published January 2, 2011
Calibration Factorsfor ElectromagneticFlow Transmitters
Reference Notes
1. Digiflow® FlowX3 flow meters measure liquids only. No air bubbles should be present and the pipe must always be full.
2. Check chemical resistance of the fluid to verify if wetted parts are acceptable.
3. Check that the service is within the pressure/temperature rating of installation fitting.
4. Viscosity range is 0.5 to 20 cSt (centistokes). Factor calibrationis required if outside this range, up to 40 cSt maximum. Maximum viscosity for Oval Gear flow sensor is 1,000 cP. Insertion electromagnetic flow transmitters have no viscositylimitations. See Units of Viscosity below.
5. Maximum percentage solids content for paddle wheel sensors is 10%, with particles not to exceed 0.5 mm in length or diameter (0% solids for ULF flow sensors). The insertion typeelectromagnetic flow transmitters can handle homogeneoussolids easily. Particle size is not an issue.
6. Paddle wheel flow meters cannot be used on services containing ferrous particles i.e., those which will be attracted to the magnetic rotor blades. The insertion type electromagnetic flow transmitters is only sensitive to metallicparticles if they are magnetic.
7. Electromagnetic flow transmitters require liquid to haveminimum conductivity at 20 microSiemens.
8. For accurate flow measurement there must be a developedturbulent velocity profile at the sensor location. This requires astraight run pipe with a minimum number of pipe diametersdistance upstream and downstream of the flow sensor. Thesedistances depend on the type of piping element (i.e. valves, elbows, reducers etc.) causing the disturbance. See page 36.
9. Hall Effect signals may be transmitted up to 300 meters(984 ft.) without the need for conditioning whereas Coil orReed Effect signals may be transmitted up to 16 m (52.5 ft.) without conditioning. Coil Effect sensors are normally used with the battery powered flow monitor.
Units of ViscosityAbsolute Viscosity (m): 1 poise (P) = 1 dyne = 1 gm
cm2 cm ● sec.1 centipoise (cP) = 0.01 poiseViscosity of water at 20oC is 1.002 cP
Kinematic Viscosity (n) = Absolute ViscosityDensity of Fluid
1 stoke (St) = 1 m2
sec.1 centistoke (cSt) = 0.01 stoke
Viscosity is temperature dependent. Liquid viscosities drop withincreased temperature.
Concentration:
Viscosity:
Fluid:
Temperature: oC
psi
cSt
%
Units:/
mm dia.
%
(Notes 1, 2)
(Note 2)
(Notes 2, 3)
(Note 3)
(Note 4)
(Note 5)
(Note 5)
(Note 6)
(Note 7)
Pressure:
Ferrous Particles in Fluid?:
Electromagnetic flow meters only:conductivity of fluid greaterthan 20 microSiemens?:
Particle Size:
❏ Yes ❏ No
❏ Yes ❏ No
Maximum Solids:
Maximum/minimum measured flow rate:
Pipe Size:
Sensor Body: ❏ CPVC ❏ 316L SS❏ PVDF ❏ Brass
Sensor O-Rings: ❏ EPDM ❏ Viton®
❏ Other
Installation Fittings: ❏ Tee ❏ Saddle❏ Weld-On Adaptor
Power Supply Available:
Distance of straight pipe upstream/downstream
of flow sensor: /
Type of piping element upstream/downstreamof flow sensor (i.e. valve, elbow, reducer, etc.):
/
Maximum distance from flow sensor to remote
instrument:
Pipe Material:
Digiflow® FlowX3 Technical Catalogue 47
(Note 8)
(Note 9)
(Note 8)
PDF Published January 2, 2011
SelectionQuestionnaire
The Heart of the SystemFive-blade ECTFE (Halar®)rotors with ceramic shaft
and bearings
Complete Product Range• Paddle wheel flow sensors• Sensor installation fittings
• Instruments
Digital OutputPulse output fromflow sensor providesfor high accuracyat low flows
A square wave (pulse)output signal from thesensor is proportional
to the flow velocity. This is fed to a
FlowX3 flowmonitor/transmitter.
A Cost Effectiveand Complete Rangeof Flow Monitoring
Equipment
55 Guardsman Road, ThornhillOntario, Canada, L3T 6L2
Your Pipeline To Quality Valves, Piping, Flow Meters and Controls
FX3C01-11©Chemline Plastics Limited 2011Chemline is a registered trade mark of Chemline Plastics Ltd.
Tel: 905-889-7890 [email protected]: 905-889-8553 www.chemline.com