1 Cardiac Pacing Sami Viskin Sami Viskin Tel Aviv, 1998. Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate modulation Programmable parameters in single chamber pacing. Lower rate limit Sensitivity Output Voltage Pulse width Sensitivity mV 5 2 8 Undersensing Good sensing Oversensing High Low Sensitivity. Temporary pacemaker Demand Asynch. Demand (less sensitive) Programmable parameters Pacing output. Voltage Pulse width

Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

  • Upload
    others

  • View
    12

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

1

Cardiac PacingSami ViskinSami Viskin

Tel Aviv, 1998.Tel Aviv, 1998.

Pacing code.Chamber

PacedChamber

SensedResponse

VAD

VAD

ITD

PR

O O O

ProgramRate

modulation

Programmable parameters in single chamber pacing.

• Lower rate limit• Sensitivity• Output

VoltagePulse width

SensitivitymV

5

2

8 Undersensing

Good sensing

OversensingHigh

Low

Sensitivity.Temporary pacemaker

Demand Asynch.Demand(less sensitive)

Programmable parametersPacing output.

Vol

tage

Pulse width

Page 2: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

2

87654321

0.2 0.4 0.6 0.8 1.0 1.4 1.6Pulse width (msec)

Volts

Threshold

2X

3X

Pacing threshold.Strength-duration curve.

Basic timing cycles: VVI pacing.

Lower rate limit

V:V R V

Basic timing cycles: VVI pacing.

V:Refractory

Oversensing.T wave oversensing

V:Myopotential oversensing

V:

Ventricular undersensing.

V:

Pseudo-undersensing.

Page 3: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

3

Failure to pace.No output

No capture

Failure to pace

No output:• Pacemaker failure• Total battery

depletion• Lead fracture• Lead disconnection• Oversensing

No capture• Partial battery

depletion.• Lead dislodgment• Lead insulation

brake• Threshold rise

Undersensing / oversensing.• Lead dislodgment• Poor lead position• Lead insulation break• Inappropriate

programming

Basic timing cycles: DDD pacing.

AV interval

VA interval

A

V

Lower rate limit = AV interval + VA intervalLower rate limit = AV interval + VA interval

Pacemaker-mediated tachycardia (DDD mode)

Tracking of a fast atrial rhythm.

End-less loop tachycardia

Initiation of end-less loop tachycardia.

Retrograde P

Retrograde P

PVC

Atrial non-capture

Page 4: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

4

Timing cycles (DDD) Refractory

AV

PVARP

Blanking periodSafety pacing

Page 5: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

5

Referred to emergency room because of this electrocardiog

What to do?

Page 6: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

6

P

QT=680 msec

Same Holter

Page 7: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

7

V-rate 38/min R-R=1600 msec; QT = 600 msec

Magnet-rate 98/min QT = 7200 msec

R-R=620 msec R-R=1320 msec

PPP PP PPP A

720 600

*

P P P P P PP PP P

A

A

P P

A woman (80 years old) with femur fracture in need for surgery

Page 8: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

8

Soon after surgery……….

Toran called because of pacemaker malfunction

What to do?

Toran places magnet on pacemaker………………..

Everything is fine after the magnet… o’Kus po’Kus?

Soon after surgery……….

Page 9: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

9

QT = 600 msec

QT = 480 msec

A

V

Stored electrogram in a dual chamber ICD: Is this VT?

A

V

A

V

A

V

Page 10: Pacing code. Cardiac Pacing - IHS FILESERVER · Cardiac Pacing Sami Viskin Tel Aviv, 1998. Pacing code. Chamber Paced Chamber Sensed Response V A D V A D I T D P R O O O Program Rate

10

Appropriate detection of VT 180/min

VT detection rate > 180/min

Antitachycardia pacing

VT = 179/min not detected