P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical. System Design Review. The Team. Team Members: Pattie Schiotis – Team Manager (ME) Shane Reardon – Lead Engineer (ME) Dana Kjolner (EE) Robert Ellsworth (EE) Sam Hosig (CE) John Williams (CE) Faculty Guide: Dr. DeBartolo. Agenda. - PowerPoint PPT Presentation

Citation preview

Page 1: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

System Design Review

Page 2: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Team Members:◦ Pattie Schiotis – Team Manager (ME)◦ Shane Reardon – Lead Engineer (ME)◦ Dana Kjolner (EE)◦ Robert Ellsworth (EE)◦ Sam Hosig (CE)◦ John Williams (CE)

•Faculty Guide: Dr. DeBartolo

The Team

Page 3: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Introduction Work Breakdown Structure Customer Needs Engineering Specifications Functional Decomposition Concepts Component Benchmarking Risk Assessment MSD I Project Plan

Agenda

Page 4: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Lasting side effect of a stroke: foot drop◦ Inability to dorsiflex the foot

Ankle Foot Orthotics (AFOs) currently used to aid dorsi-flexion.◦ Passive devices don’t allow for movement when

walking on ramps and stairs Foot is always pointed upwards

Project Background

Page 5: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

User will have no ability to either plantar-flex or dorsi-flex their foot

Side to side stability of the foot will be ignored

Worst case will be analyzed:◦ 95 percentile male having heavy foot.◦ Fast walker – gait cycle less than 1 second.

Device may not use air muscles as an actuation source

Assumptions & Constraints

Page 6: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Work Breakdown Structure

Page 7: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Key Customer NeedsPrimary Needs: Secondary Needs: Safety Portable

◦ Lasts all day without charging/refueling

◦ Lightweight◦ Tolerable to wear all day

Reliable Accommodates Flat Terrain Accommodates Special

Terrain◦ Stairs◦ Ramps◦ Obstacles

Comfortable◦ Aesthetically Pleasing

Durable◦ Water Resistant◦ Corrosion Resistant

Salt & Environment Biocompatibility

Convenient◦ Easy to put on and take

off

Page 8: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Key Engineering Specifications

Engineering Specification

Number

Engineering Specification Description

Units of Measure

Preferred Direction

Nominal Value

Method of Validation

Stems From Customer Need

s1 torque on Foot N-m Up ≥±3.0 Test FT1,2,4,ST1,5

s2 system response time (sensing terrain to actuating device) ms down <400 Test ST3

s4 predicts step down yes/no - yes Test ST1,2,4s5 predict flat yes/no - yes Test FT1,ST5s7 predicts ramp down yes/no - yes Test ST1,2,4

s10 allowable range of motion between foot and shin degrees range 70 to 135 Test FT1,3,CF8,9,ST1

s12 untethered usage time hrs/steps up 8 hrs or 3000 steps   P1,2,D1

s17 force to secure constraints N down < 80 Test C4s18 force to remove constraints N down < 80 Test C3s23 radius of edges/corners on AFO mm up 0.5 - S4,CF1,2s24 weight of entire device kg down ≤3 Test FT6

s28 Operates in environment temperature range °C range -17.8 to

37.8Component

Ratings D2

s31 Minimum life until failure steps up 5.5 million test D1

Page 9: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Functional Decomposition

Page 10: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Concept Generation Table

Page 11: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Mechanical Locking MethodUses a solenoid to unlock the heel which allows the foot to drop.

Design Option 1

Page 12: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Mechanical Locking Method◦ Mechanically restrict the foot from dorsi/plantar

flexing while in the air.◦ Use gravity to help the foot plantar-flex as

needed (stairs/ramps)◦ Methods:

1. Use a solenoid to unlock the heel which allows the foot to drop.

2. Use a brake to unlock the heel which allows the foot to drop. 3. Ratcheting Device attached to ankle

Design Option 1

Page 13: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Design Option 2 Active Actuation

◦ Uses a linear actuator to force the foot into the proper position.

◦ Methods: Solenoid Piezoelectric Power Screw Hybrid

Power Screw with electric motor

Page 14: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Hard StopTwo preset distances can be moved back and forth usinga servo motor and screw.

Design Option 3

Page 15: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Preliminary Calculations

42%

0.152H

Fw

NmmkgsmHFwankleM 636.1)868.1)(152.0)(42.0)(4.1(/8.9)152.0%)(42(/ 2

lbfNcmNm

lTF 25.1150

3.3636.1

If our actuator is 1.5” (3.3 cm) behind ankle joint:

Anthropometric Data for male, 95 percentile:•Stature=1.868 m•Foot Weight=1.4 kg

Page 16: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Calculations Continued

θ=20°cmly 28.1)20sin(3.3sin

Dorsi-flexion:

Plantar-flexion:cmly 33.2)45sin(3.3sin

Total stroke=3.61 cmIn 0.4s, we need 9 cm/s

Page 17: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Active Actuation SourcesSolenoid Electro-Mechanical Piezoelectric

Actuation Force 30 lbs 22 lbs 100NStroke 1.1 cm >4.4 cm micro-metersSpeed <100 ms 10 cm/s -Power Consumption 92V, 7.2 A (moving), 0.08 A (holding) 12VDC, 5 A 100 V

Page 18: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Pros and ConsPros Cons

Mechanical Locking

• Low Power Consumption• Can hold the foot up in the

event of an electrical failure

• Uses Gravity to position the foot

• Can only hold the foot in a position that it has been

• Needs a Large voltage spike to trigger the solenoid

Active Actuation

• Can move the foot to any location needed

• Can be designed to hold the foot up under failure

• Large amount of energy required

• Slow response time• Heavy

Hard Stop • Low Power Consumption• Capable of counter-acting

large amounts of plantar-flexion force

• Difficult to create failsafe

• Slow response time• Does not reset it self

when the toes need to be pointed up

Page 19: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Concept Selection Matrix    Concepts  A B C D  Mechanical

LockingActive

Actuation Hard Stop Passive Device 

Selection Criteria Weight Rating Weighted Score Rating Weighted

Score Rating Weighted Score Rating Weighted

Score

Safety 20% 2 0.4 1 0.2 1 0.2 1 0.2Portable 15% 2 0.3 0 0 3 0.45 3 0.45Reliable 15% 2 0.3 3 0.45 1 0.15 2 0.3Accommodates Flat Terrain 15% 2 0.3 1 0.15 1 0.15 1 0.15Accommodates Special Terrain 15% 3 0.45 1 0.15 2 0.3 0 0Comfortable 8% 1 0.08 0 0 2 0.16 1 0.08Durable 8% 1 0.08 1 0.08 2 0.16 2 0.16Removability 4% 2 0.08 2 0.08 3 0.12 2 0.08

  Total Score 1.99 1.11 1.69 1.42  Rank 1 4 3 2  Continue? Yes No No N/A

Page 20: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Bounces infrared light off terrain to determine distance

10 cm to 80 cm range Worst case power consumption

per sensor is 0.00176 kWhr Output from -0.3 to +0.3 volts Highly accurate within

operational ranges Low cost (~$15)

https://www.sparkfun.com/products/242

System Component: Sensors

Page 21: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

  A Far A Normal A Close

B Far

Mid Stride, Down Slope/ Stairs

Mid Stride, Normal Terrain

Mid Stride,

Up Slope/ Stairs

B Close

Foot Planted, Down Slope/ Stairs

Foot Planted, Normal Terrain

Foot Planted,

Up Slope/ Stairs

IR Distance and Terrain Type

Sensor A

Sensor B

Terrain

Page 22: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Sensor Calibration

Nonlinear response Reflective ratio of

materials is mostly irrelevant

Static position

Concerns about irregular terrain types

https://www.sparkfun.com/products/242

Page 23: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Predicts what type of terrain we are walking on

Calculates where in the gait cycle we are Has some modeling that improves

performance of predictions

NEED TO CHANGE Need to be able to fully implement in C Error checking for invalid states Nothing is done at run-time

Current Code: Developed by Chappy

Page 24: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

System Component: Micro Controller

Page 25: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Micro controller needs:◦ Interface with two IR sensors and possible angle

senor ADC

◦ Control actuation method PMW or Digital I/O

◦ Other considerations Must run on battery power for at least 8 hours Must be able to simulate system in “real time” Must be able to fit on orthotic Must be able to export data to sd card if needed

System Component: Micro Controller

Page 26: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Power ConsumptionPower From Micro Controller and Sensors

DeviceTime

(Hours)Current

(A)Voltage

(V)Power

(mWhr) Total Power (mWhr)IR Sensors 8 0.03 5 1200 2400Micro C. 8 3.00E-04 5 12 12

Total Power 241280% Efficiency 2894

Solenoid Power Usage for Option 1Watts per step Time per step mWhr / step

40 0.5 5.5

Page 27: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Risk Assessment

ID Risk Item Effect Cause

Likelihood

Severity

Importance Action to Minimize Risk Owner

1Foot failing in down position User would trip and fall Sensor unable to detect terrain

or send faulty readings 2 3 6Software fail safe, if no data is sensed will fail in natural position

Engineering Lead

2Actuator malfunction

1 2 2System free to move if actuator breaks

Engineering Lead

3

Structural failure, orthotic unable to support equipment 1 3 3

Large enough factor of safety Engineering Lead

4Spring Yielding/Buckling

1 2 2Structural modification, mechanical prevented action

Engineering Lead

5

Power supply

3 2 6

Before system runs out of power, lock in normal state. Warning signal included

Electrical Engineers

6

Scheduled deadlines not met Timeline falls behind, other deadlines change

Personal conflicts: time management, overloads schedule, illness

3 2 6

Communication among members to know each other’s schedules, understand critical path, seek help when needed

Team Manager

7

Material acquisition delay Prototype cannot be built and tested

Long lead times, parts not ordered on time

1 2 2

Contact with vendors, determine parts with long lead times, order by week 7

Team Manager

8

Incorrect material handling Overload system capabilities Misuse of supplies

2 2 4

Responsible team members in charge of their components. Understand system capabilities and specifications

Engineering Lead

9Unable to meet customer specifications

Project failure, unhappy customer

Weight of device too heavy2 3 6

Be cautious of component weights when creating detailed design

Engineering Lead

10Device contains sharp edges, harms the users 1 3 3

Ensure all points of contact will not harm user, no pressure points

Mechanical Engineers

11

Memory overflow Device would go into an error state

Not enough memory on the micro controller and associated memory systems 2 2 4

Ensure enough memory is available on micro controller

Computer Engineers

Page 28: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Project PlanID Task Name Complete (%) Completetion Goal When

Completed Issues/Comments1 Define Project 2 Review customer needs 90% 10/2/2012 Review after concepts selected3 Review customer specifcations 90% 10/2/2012 Review after concepts selected4 Finalized functional decomposition 100% 9/21/2012 9/21/2012 5 Determine work breakdown structure 100% 10/5/2012 9/28/2012changes after systems review6 Observe walking patterns at naz clinic 100% 9/24/2012 9/24/2012 7 Concept Generation 8 Create benchmark matrix 90% 9/28/2012 9 Define components 100% 9/14/2012 9/28/2012

10 Establish system possibilities 100% 9/14/2012 9/21/2012 11 Create system comparions (Pros and Cons) 100% 9/28/2012 10/2/2012 12 Develop Proposed Design 13 Assign team member specific jobs 100% 9/21/2012 9/21/2012Proposed design approval14 Develop psedeocode 100% 9/28/2012 9/28/2012 15 Review previously developed sensor code 100% 9/28/2012 9/28/2012 16 Feasibility analysis 80% 10/2/2012 17 Compile Systems Design Review 18 Schedule review 100% 9/25/2012 9/28/2012 19 Create risk assessment 100% 10/2/2012 10/2/2012 20 Create review report out 95% 9/28/2012 21 Develop project schedule 90% 10/2/2012 22 Part Selection 23 Select components 40% 10/12/2012 24 Create budget breakdown 0% 10/12/2012 Need component specifications 25 Detail component information (specs, vendor) 5% 10/19/2012 26 Purchase parts 0% 10/26/2012 relient on specs 24-2627 Detailed Design 28 Create BOM 0% 10/19/2012 29 Update risk assessment on going 30 Verify design output 0% 10/23/2012 31 Finalize system architecture 0% 10/16/2012 32 Prepare drawings, schematics, and flow charts 0% 10/26/2012 33 Identify critical design path 0% 10/12/2012 34 Document design changes (if any) 0% 11/10/2012 35 Test Plan 36 Determine component testing 0% 10/26/2012 37 Create testing guide/SOP 0% 11/3/2012 38 Estimate resource requirements 0% 11/10/2012 39 Create data collection sheets 0% 11/10/2012 40 Additional Tasks 41 Update EDGE on going

Page 29: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

1. Scope of project is to design a modified AFO that includes: Energy storage medium Foot rotation device Terrain sensing system Microcontroller

2. We will focus on a detailed design following the “Mechanical Locking Mechanism” concept.

Closing Remarks/Comments

Page 30: P13002: Ankle-Foot Orthotic Un-Tethered, Mechanical

Questions?