  HealthGuard-15 Owners Manual -Version 1.4 1 HealthGuard-15 Owners Manual Appearance and Part Names Integrated Cuff Arm Rest Power Supply Optical Standard Software CD USB Cable Scale Serial Cable for Scale Hardware (in bag) The HealthGuard-15 should include all of the co mponents identifi ed above. The HealthGuard-15 measures the blood pressure and estim ates the body fat of adults aged 18 and over.

Owners Manual-hg15 v1_4

Embed Size (px)

DESCRIPTION

This is the owners manual

Citation preview

  • HealthGuard-15 Owners Manual -Version 1.4 1

    HealthGuard-15 Owners Manual Appearance and Part Names

    Integrated Cuff Arm Rest Power Supply

    Optical Standard Software CD USB Cable

    Scale Serial Cable for Scale Hardware (in bag) The HealthGuard-15 should include all of the components identified above. The HealthGuard-15 measures the blood pressure and estimates the body fat of adults aged 18 and over.

  • HealthGuard-15 Owners Manual -Version 1.4 2

    Getting Started Assembling & Setting Up Your HG-15 First you'll need enough room to place the scale, the integrated cuff, and your computer next to each other as shown in the illustration below.

    Next, open and remove the contents of the hardware bag (2 phillips head screws, 2 washers, and 2 thumb screws). Attach the are rest to the cuff as shown below:

  • HealthGuard-15 Owners Manual -Version 1.4 3

    Once assembled, attach the cuff to the table top using the supplied thumbscrews:

    Attach the Cords You can now connect all of the components to complete the assembly of your HG-15: Key: A- Communication Cable B - Power Cord C - USB Cable

    You can now proceed with installation of the HG-15 software.

  • HealthGuard-15 Owners Manual -Version 1.4 4

    Installing the HG-15 Software Minimum PC Requirements PC Running Windows XP or Vista 32 mb RAM Software Installation Insert the Software CD into your CD drive. The CD will start automatically, and you will see the following screen:

    Follow the on screen prompts to complete installation of the software. Note: If the CD does not start automatically, view the contents of the CD and double click the setup.exe file. The HG-15 can now be connected to your PC.

  • HealthGuard-15 Owners Manual -Version 1.4 5

    Make the following connections in the following order: 1. Power Connection Connect the power supply to the Power Connection located at the end of the Interface Box. Then plug the power supply into a standard wall socket to provide power to the HG-15. 2. Serial Cable Connect the serial cable to the scale and then to the cuff. 3. PC Connection The HealthGuard-15 communicates with your PC via a USB Cable. Simply connect the USB cable to any available USB port on your PC. Then plug the connector on the opposite end of the USB Cable into the HealthGuard-15.

    Note: The HG-15 may take 2-5 minutes to warm up before it is recognized by your PC.

  • HealthGuard-15 Owners Manual -Version 1.4 6

    Using the HealthGuard-15 Start the Software Start the software by double clicking the HG icon on your desktop. You will see the following display:

  • HealthGuard-15 Owners Manual -Version 1.4 7

    Zero Adjusting the HG-15 Before using the HG-15, the body fat measurement probe should be zero adjusted. To start the Zero Adjust procedure simultaneously press the "Ctrl" key and the "Z" key (Ctrl Z). You will see the following dialogue box:

    Place the optical standard in the cuff as shown and click "OK"

  • HealthGuard-15 Owners Manual -Version 1.4 8

    When the Zero Adjust is complete, you will see the following dialogue box:

    Click "OK" to acknowledge the completed Zero Adjust. Futrex recommends that a Zero Adjustment be performed each time the unit is powered up. Selecting Metric or English (or Imperial) Measurement Units. The default measurement for the HG-15 is English (or Imperial) units. To put the HG-15 into Metric Mode, simultaneously press the "Ctrl" key and the "M" key (Ctrl M). To return to English (or Imperial) units simultaneously press the "Ctrl" key and the "E" key (Ctrl E).

    Click OK to acknowledge the change.

  • HealthGuard-15 Owners Manual -Version 1.4 9

    Performing a Measurement To start a measurement click the "Begin" Button on the HG screen:

    You will then be prompted to enter the following data:

    Use the mouse to select Male or Female

  • HealthGuard-15 Owners Manual -Version 1.4 10

    Click the "Age in Years" Box Then the keypad shown below will appear.

    Enter your age, and click DONE.

  • HealthGuard-15 Owners Manual -Version 1.4 11

    You will then be prompted for your height. Click the "Height in Inches" box (it will indicate "Height in cm" if the software is

    configured for Metric measurement units), and the keypad will appear once again.

    Enter your height, and click DONE

  • HealthGuard-15 Owners Manual -Version 1.4 12

    You will then be prompted for your weight. Click the "Use Scale" box, or "Enter Manually" box.

    If you selected "Enter Manually," use the on screen keypad to enter your

    height, and click DONE. If you selected "Use Scale," the following screen will appear.

    Step on the scale. When weight has been successfully measured, the following dialogue box will appear:

  • HealthGuard-15 Owners Manual -Version 1.4 13

    Click OK, and the screen will look like this:

    Place the left arm into the cuff, and press Start. The following screen will appear:

  • HealthGuard-15 Owners Manual -Version 1.4 14

    The measurement may be stopped at any time by clicking the Abort button shown above, OR by pressing the red Emergency Stop Button on the side of the cuff housing (see photo).

    It is very important that the test subject be still and quiet during the evaluation. Talking or movement can cause errors in the measurements.

  • HealthGuard-15 Owners Manual -Version 1.4 15

    As soon as the test is complete, a results screen will appear hat provides an evaluation of the information that was gathered.

    The results will remain visible for 60 seconds. The user has the option of simply writing them down, or clicking the "Print Results" button for a 3 page report. To obtain the printed report, the "Print Results" button must be clicked within 60 seconds of completing the test.

  • HealthGuard-15 Owners Manual -Version 1.4 16

    Entering Blood Pressure, Weight or Body Fat Bias Values Your HealthGuard-15 has been custom calibrated to ensure accurate measurement of blood pressure and body composition; however, over time it may be necessary to adjust the readings by entering bias values in the software

    1. Using Windows Explorer, view the folders on your C Drive, and locate the folder named "Futrex."

    2. Open this folder and you will find a sub folder named "Kiosk."

    3. This folder will contain a file named bias.txt. Open this file. You will see the following window.

  • HealthGuard-15 Owners Manual -Version 1.4 17

    4. Enter the appropriate numbers. The amended file may look something like this:

  • HealthGuard-15 Owners Manual -Version 1.4 18

    Troubleshooting Problem Probable Cause(s) Corrective Measures Display Shows HG-15 NOT CONNECTED

    a) Bad Connection between the HG-15 and the USB Port b) No Power to the HG-15 c.) HG-15 is not warmed up

    a) Ensure that the USB Cable is connected to both the PC and the HG-15 b) Ensure that the power supply is plugged into a working outlet, and that it is connected to the HG-15 cuff. c.) The HG-15 takes 1-2 minutes to warm up (sometimes a bit longer). Wait for a minute or two, and the try again.

    Poor Repeatability on Blood Pressure Measurements

    Subject moved during the evaluation.

    Repeat the evaluation, and be sure to remain still and quiet during the evaluation.

    Unusually high blood pressure readings

    Subject has not relaxed enough to be at normal resting blood pressure.

    Have the subject sit quietly for 2-3 minutes, and then repeat the test.

    Error measuring Body Fat. Display is ---

    Sleeve was in the way of the body fat sensor.

    Ensure that the subject has his or her sleeve rolled up. The skin must be bare for a body fat reading.

    Poor Repeatability on Body Fat Measurements

    Incorrect Zero Adjustment Procedure Subject moved during the evaluation Sleeve was in the way of the body fat sensor.

    See the Zero Adjust Procedure on page 5 Repeat the evaluation, and be sure to remain still and quiet during the evaluation. Ensure that the subject has his or her sleeve rolled up. The skin must be bare for a body fat reading.

    Error taking Measurement when stepping on scale

    Bad connection between the scale and the HG-15

    Ensure that the Serial Cable is connected to both the scale and the HG-15