Upload
sasha-kozmonaut
View
124
Download
0
Embed Size (px)
Citation preview
The Optimal Production of Methyl Acetate Using a Microreactor
Summer Research Program 2007
Elizabeth LeitnerValerie PiazzaMaria del Mar RojasAlex KitayevPedro CuacuasGrace BrockCorinna Wang
Table of Contents
Page Section
3 Abstract
4 Introduction
6 Experimental
9 Results & Discussion
19 Conclusion
20 Recommendations
21 Works Cited
22 Acknowledgements
23 Appendix I – Reaction Diagram
24 Appendix II - Calculations
2
Abstract
The purpose of this experiment was to find the optimal condition for producing
methyl acetate. Methyl acetate is made by the esterification of methanol and acetic acid
with an addition of a catalyst, sulfuric acid. Several runs were carried out under different
conditions. Different catalyst amounts were tried with different temperatures and
different pump settings to find the ideal conditions in which the yield of methyl acetate
would be the highest. Using the conversion rate of methanol, the production rate of
methyl acetate was determined. The optimal settings for producing the maximum
conversion of methanol to methyl acetate was 1.0mL catalyst at 40˚C at a pump setting of
15%/35%, which produced a yield of 87.6%.
3
Introduction
Methyl acetate is a commonly used chemical in industry around the world.
Methyl acetate is a component in such items as nail polish removers, glues and used as a
substitute for acetone [1-2]. Methyl acetate is a product of an esterification reaction
involving methanol and acetic acid. Sulfuric acid must be added as a catalyst in order for
the reaction to significantly take place. This is because the catalyst lowers the activation
energy, allowing the sulfuric acid to increase the rate of the reaction without being
consumed. Various conditions can be changed to produce different yields of methyl
acetate. By changing the temperature, flow rate of reactants and amounts of catalyst
used, the amount of methyl acetate produced will vary. To analyze the final product, a
gas chromatograph was used. A sample of the final product was injected into the gas
chromatograph and the results were displayed on the integrator, which included
information such as the retention times for each component as well as the area.
There are two methods that are used to produce methyl acetate. One method uses
reactive distillation and the other method requires the use of a microreactor. Liquid
phase reactions commonly use reactive distillation. Methanol is placed in a container
with acetic acid and sulfuric acid where they are heated. Since methyl acetate (the
product) has the lowest boiling point (56.9˚C) compared to methanol (64.7˚C), acetic acid
(118.7˚C) and the sulfuric acid catalyst (338˚C), which have higher boiling points, it will
be the first compound to vaporize [3-6]. The methyl acetate vapors flow through the
distillation column into a separate container where the methyl acetate is then condensed
back into its liquid phase. According to Le Chatelier’s Principle, if the product is
4
removed from the reaction, the reaction shifts to the product side to compensate for the
lost products [7].
The other method involves the use of a microreactor. For the experiment, the
microreactor was used to mix methanol with acetic acid within narrow channels to
produce methyl acetate. There were inlet tubes for both methanol and acetic acid and a
tube dedicated for the heat exchange fluid to alter the temperature if necessary, as well as
an outlet tube to collect the effluent. These narrow conditions better helped the
substances to mix for a more optimal product.
Reactive distillation and the microreactor both provide a way to produce methyl
acetate. The benefit of using the microreactor in comparison with reactive distillation is
that the microreactor is far safer than distillation. Since the reaction process is confined
into a smaller space, it is easier to both operate and service. Due to its small size, the
microreactor is more easily detected for malfunctions. Also once the malfunctions are
detected they are easier to fix. Since reactive distillation is a large process, it would be far
more difficult to repair. Part of the research was to determine if an array of
simultaneously working microreactors would create the same yield of methyl acetate as a
reactive distillation column under the most optimal conditions.
5
Experimental
The goal of the project was to find the optimal settings for the best production of
methyl acetate using methanol, acetic acid and a catalyst, sulfuric acid. In order to
accomplish this goal various experiments were run. The whole project was separated into
three stages. The first area of experimentation involved the gas chromatograph (GC).
The GC separates and identifies the substances involved in the reaction. Each substance
injected into the GC passes through a column and, in the process, was slowed down
depending on its volatility. As a result, the various substances leave the GC at distinct
rates based on their retention times. A connected Hewlett-Packard 3395 Integrator
served to express the information recognized by the GC through a unique curve for each
substance injected. In addition, it states the retention time of each substance next to its
corresponding curve as well as the areas of each peak. Although it is not immediately
apparent which curve corresponds to which substance, repeated experimentation served
to identify the specific peak and retention time of each compound injected into the GC.
Initially, each substance was run separately through a Gow-Mac Series 580 Gas
Chromatograph in order to ascertain the characteristics: total area, percent area and
retention time. Next, a mixture of the three substances was run through the GC to see if
three different peaks would appear. However, it was imminently apparent that the
methanol and methyl acetate had similar peaks and therefore it was necessary to adjust
the settings. Originally the column temperature was at 110˚C, the detector temperature
was at 185˚C and the injection temperature was at 170˚C. This was problematic due to
the fact that the substances were being processed too quickly, primarily caused by high
temperatures. As a result, the machine could not distinguish between the methanol and
6
methyl acetate. The temperatures needed to be reduced in an effort to slow down the
process. After thorough experimentation it was revealed that the most effective
temperatures were a column temperature of 41˚C, a detector temperature of 128˚C and an
injection port temperature of 127˚C. Furthermore, it was established that the methanol
peak was detected first by the integrator, followed by the methyl acetate and the acetic
acid.
After the GC settings were modified for their ideal use, the pump settings for each
substance needed to be calibrated. The pumps used were Index Corporation Model
#701C Micropumps. A series of processes of methanol and acetic acid was completed in
order to establish the proper calibration settings for each substance. First, an empty
beaker was weighed and recorded on a Denver Instrument APX-2001 balance. Once the
mass was observed, the settings of the methanol pump were tested. Beginning with zero
percent, methanol was run for fifteen seconds into a waste beaker to establish a constant
flow. Then it was moved into the empty beaker for the next fifteen to twenty seconds.
After the pumps were shut the beaker was then weighed again to determine how much
methanol had been collected. This same process was repeated with pump settings
ranging from five percent to forty percent in increments of five percent. The same
calibration process was used for the acetic acid.
In order to analyze samples of methyl acetate and its reactants, methanol and
acetic acid were processed through a microreactor at room temperature without the
presence of a catalyst. A steady state of two seconds was established during which the
sample was not collected. This was done in order to allow time for the reaction to take
place. Then the sample was collected in a beaker for two seconds. This was the sample
7
used to analyze the yield of methyl acetate. The beaker that contained the effluent was
placed in a beaker of ice water. The procedure was repeated at various pump settings,
starting at 20%/20% (methanol pump setting/acetic acid pump setting) at increasing
intervals of five percent until 40%/40%. Then, the procedure was repeated at 35%/15%
and 15%/35%. This was done in an effort to see the impact using different amount of
methanol and acetic acid had on the final results.
In order to determine whether a catalyst would affect the outcome, the reaction
was repeated using 1mL of sulfuric acid. Then the amount of sulfuric acid used was
increased by half a milliliter after each set of runs up to three milliliters. The pump
setting did not differ between the reaction conducted without a catalyst and the reaction
conducted with a catalyst. These reactions were conducted at room temperature.
Another factor involved in the experiment was temperature. The reaction was
conducted at temperatures of 28˚C (room temperature), 40˚C and 50˚C using the same
pump settings. The temperatures were altered using a Corning Stirrer/Hot Plate and
monitored using an Extech RTD thermometer. After the reaction was completed for the
specific pump settings and temperatures, the product of the reaction was collected and
transported to the GC in an ice bath. This was done in an effort to stop the reaction
altogether. This specific reaction is a reversible reaction, so icing the product will
prevent the reverse reaction. Using a syringe, a sample of 0.5μL was extracted and
injected into the GC. Then the attached integrator identified the peaks and retention
times for each substance.
8
Results & Discussion
Initially, the calibration curves for methanol (Figure 1) and acetic acid (Figure 2)
were computed. The results are shown below.
Calibration Curve of Methanol
y = 0.1008x - 0.5423R2 = 0.9873
-0.5
0
0.5
1
1.5
2
2.5
3
3.5
4
0 5 10 15 20 25 30 35 40 45
Pump Setting (%)
Flow
Rat
e (g
/sec
)
Figure 1 - Calibration curve of methanol
Calibration Curve of Acetic Acid
y = 0.1319x - 0.7931R2 = 0.9508
0
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
0 5 10 15 20 25 30 35 40 45
Pump Setting (%)
Flow
rate
(g/s
ec)
Figure 2 - Calibration curve of acetic acid
It was necessary to calibrate both the pumps and the GC in order to compute the
flow rate and determine how much of each component was being released from the
microreactor at different time settings. Also, the factor of human error had to be taken
9
into consideration because the amount of the mixture being inserted could have been off
by a fraction of a microliter, which would change the results entirely.
The GC was calibrated using different amounts of pure methanol to find an
average area and number of moles calibrated, which can be observed in Figure 3. The
area computed by the integrator was not able to be used mainly because of human error.
The areas printed out were scattered due to the fact that 0.5μL was not always injected,
sometimes it was a little more and at other times it was a little less. The percentages were
always accurate because no matter how much was injected the same percentage would
always be calculated. Therefore, the percentages were more accurate and were better to
use. Additionally, it was necessary to find the average area for a 0.5μL injection because
the average area would be used as a constant once the calculations were completed. As a
result, all the computation would be accurate. The pumps were calibrated so that the
flow rate could be calculated, and so that the amount of each substance released from the
microreactor could be determined at particular time settings.
10
GC Calibration of Methanol
y = 8E+09x + 23175R2 = 0.9869
0
100000
200000
300000
400000
500000
600000
0 0.00001 0.00002 0.00003 0.00004 0.00005 0.00006 0.00007
Number of Moles
Aver
age
Area
Figure 3 - GC calibration of methanol
The conversion of methanol is directly related to the setting in which it is being
pumped. The higher the pump setting, a higher percent of conversion yielded. As shown
in the “calibration calculations,” the pump setting was increased from 20% to 25%,
which raised the average methanol conversion from 24.43845% to 25.91316%. At room
temperature, the optimal condition found was at 35% acetic acid and 15 % methanol,
which yielded an average of 87.6% and was the highest conversion of methanol recorded.
The same correlation was true at 40˚C and 50˚C, as demonstrated in figure 5 and figure 6,
respectively. This result is due to the fact methanol is consumed more rapidly when there
is a smaller amount present in comparison with that of acetic acid.
11
Average Methanol Conversion at Room Temperature
0
5
10
15
20
25
30
0 5 10 15 20 25 30 35 40
Pump Setting (%)
Ave
rage
Con
vers
ion
(%)
Figure 4 - Average methanol conversion with varying pump setting (Room Temp)
Average Methanol Conversion at 40˚C
0
5
10
15
20
25
30
0 10 20 30 40 50
Pump Setting (%)
Ave
rage
Con
vers
ion
(%)
Figure 5 - Average methanol conversion with varying pump setting (40˚C)
12
Average Methanol Conversion at 50˚C
0
5
10
15
20
25
30
35
40
45
0 10 20 30 40 50
Pump Setting (%)
Ave
rage
Con
vers
ion
(%)
Figure 6 - Average methanol conversion with varying pump setting (50˚C)
In addition to the role of the pump settings, temperature was also a large factor in
the rate of collision of the particles. The higher the temperature was, the higher the rate of
collision. In this experiment the temperature never exceeded 50ºC because the boiling
point of methyl acetate was 56.9ºC and this could potentially have an impact on the
outcome of the experiment.
13
Average Conversion of Methanol (Temperature)
28
29
30
31
32
33
34
35
0 10 20 30 40 50 60
Temperature (˚C)
Aver
age
Conv
ersi
on (%
)
Figure 7 - Average conversion of methanol with varying temperature
Another factor that was thought to have an affect on the reaction, along with
pump settings and heat, was sulfuric acid, the catalyst. A catalyst was necessary for the
reaction to occur because without it the acetic acid and methanol cannot produce methyl
acetate. This can be observed in Table1.
Pump Setting
Conversion of Methanol
20%/20% 1.60593595125%/25% 2.42050928730%/30% 7.22344031135%/35% 4.12228632135%/15% 8.57467769
Table 1 – Conversion of methanol without a catalyst
In theory, a catalyst is supposed to speed up a reaction [8]. Therefore; increasing
the amount of catalyst should also make the reaction occur more rapidly.
When all the trials were completed, the results were recorded and the methanol
conversions were calculated, as seen in Figure 8. The 0mL catalyst runs yielded the least
14
conversion of methanol while the 3.0mL trials yielded the greatest conversion of
methanol. Most of the results correspond with the fact that the greater the concentration
of the catalyst the greater the conversion of methanol. However, there was a break in this
trend between 1.0mL of catalyst and 1.5mL of catalyst. Instead of an increase in
conversion of methanol there was a decrease between these two points. While the source
of the error is unknown, it is likely that there was a problem with the pumps or the
general procedure when conducting the experiment with 1.5mL and 2.0mL of catalyst.
Although the conversions of methanol began to increase after 1.5mL catalyst, the
conversion at 2.0mL and 2.5mL catalyst was relatively the same as the conversion at
1.0mL. This could be due to the fact that the change in the amount of catalyst was
recognizable, but not as significant until a more drastic change occurred, from 1.0mL to
3.0mL for instance.
When analyzing Figure 8, it became apparent that, while the presence of a catalyst
is essential for the reaction to occur, the amount of catalyst present appears to be
insignificant. This is so because the average conversion for each amount of catalyst used,
disregarding the conversion value for 1.5mL catalyst, only varied by less than 2%. This
percentage is within the experimental error. With each step taken in this experiment there
were many opportunities for error, including measuring the wrong amount of substance
into the beaker. Also, while transporting the mixture to the lab to be analyzed by the GC,
the ice water bath could possibly not have been cold enough, which would not prevent
the product from continuing to react.
15
Average Methanol Conversions (Catalyst Experiment)
0
5
10
15
20
25
30
35
0 0.5 1 1.5 2 2.5 3 3.5
Catalyst (mL)
Ave
rage
Con
vers
ion
(%)
Figure 8 - Average methanol conversion with varying amounts of catalyst
To see the effect that the amount of methanol and acetic acid had in relation to
each other, different pump settings of 35%/15% and 15%/35% were tested. According to
Table 2, the average conversion of methanol was lower at 35%/15% than at 15%/35%.
This is because there was a lower amount of acetic acid to react with a greater amount of
methanol. Consequently, the conversion of methanol was much higher at 15%/35%
because there was a greater quantity of acetic acid reacting with a smaller quantity of
methanol.
Pump Settings
Average Conversion of Methanol
35%/15% 15.1386252315%/35% 73.952639
Table 2 – Average conversion of methanol for different pump settings ratios
16
When conducting any lab experiment, results may sometimes be skewed due to
human and inherent errors. Such human errors in this particular lab include the different
mixtures of sulfuric acid that were produced. Graduated cylinders were used to make sure
the right amount of chemicals was mixed together. However, eyes can only see to a
certain degree of accuracy. There were human errors with the pump settings for the
methanol and acetic acid. Sometimes the pumps were set to the incorrect settings, which
produced odd results. After evaluating the results produced by the integrator was it
apparent that the pump settings were not set properly. Those runs had to be done again.
There was also the issue of how the syringe was inserted into the gas chromatograph. At
times the syringe was not inserted properly and got stuck halfway through the injection
port. The syringe then had to be pulled out and inserted once again. The proper injection
technique was to stand and inject the syringe directly. If errors still occurred, the septum
was checked and the injection port was tightened. As a result of improper injections,
some of the mixture may have evaporated, which would produce smaller peaks and areas
than expected. Additionally, the gas chromatograph was very hot, so if the syringe was
not inserted correctly, the mixture would heat up and would therefore react. This would
completely defeat the purpose of keeping the mixture in an ice bath.
There were also inherent errors such as the precision of the graduated cylinder
that was used to create various mixtures of sulfuric acid. That graduated cylinder was
only so accurate. As with the gas chromatograph, the machine needed to be set to certain
conditions in order for there to be a clear separation between all the peaks. However, like
the graduated cylinders, the machine was only so accurate because after multiple
injections of the mixture, the percent area of each peak varied slightly.
17
Additionally, residue contained in the columns of the GC could have led to
inaccurate areas. This could be due to residual water or acetic acid that did not
completely exit the column. Also, the pumps of the microreactor could not have been set
exactly at the desired settings because they did not have many tick marks for them to be
set precisely.
18
Conclusion
Several trends were observed to occur as different variables were tested. As the
pump settings increased, so did the average conversion of methanol. Additionally, when
there was no presence of a catalyst the average conversion was very low and when the
catalyst was present, the conversion was much higher. However, as long as there was the
presence of catalyst, the average conversion did not change much between varying
amounts of catalyst. Also, as the temperature increased the average conversion appeared
to increase exponentially. The average conversion of methanol was greatest when there
was excess amount of acetic acid to react with a limited amount of methanol.
The optimal conditions for producing the greatest conversion of methanol to
methyl acetate appeared to be at a high temperature, with more acetic acid than methanol,
and with the presence of a catalyst. From the experiments conducted, the optimal settings
for producing the maximum conversion of methanol to methyl acetate was 1.0mL
catalyst at 40˚C at a pump setting of 15%/35%, which produced a yield of 87.6%.
19
Recommendations
After completing the experiment, it was recognized that there should be more
experimentation with a higher range of catalyst, possibly starting at 0.1mL and going up
to 10mL. Also, while this experiment concluded at 50ºC, a higher range of temperature
settings could be more effective. In addition, one could experiment with different
combinations of pump settings. Since the 15%/35% and 35%/15% settings were so
significant in the experimental trials, one could try using pump settings in which the ratio
of methanol and acetic acid are in different proportions. Furthermore, additional
investigation of GC settings could provide for a more distinct separation in the peaks of
methanol and methyl acetate. Lastly, more runs could be conducted in order to establish a
more precise average for the collected data.
20
Works Cited
[1] “Methyl Acetate.” Chemicalland21. Arokor Holdings, Inc. 2000 (http://www.chemicalland21.com/industrialchem/solalc/METHYL%20ACETATE.htm).
[2] “Methyl Acetate is offered as a replacement for acetone.” ThomasNet. ThomasNet. 2006 (http://news.thomasnet.com/fullstory/459580).
[3] “Methyl Acetate.” W ikipedia . Wikimedia Foundation, Inc. 2 July 2007 (http://en.wikipedia.org/wiki/Methyl_acetate).
[4] “Material Safety Data Sheet.” Iowa State University. Iowa State University. 14 March 2001 (http://avogadro.chem.iastate.edu/MSDS/methanol.htm).
[5] “Safety Data for Acetic Acid.” University of Oxford. University of Oxford. 6 January 2006 (http://physchem.ox.ac.uk/MSDS/AC/acetic_acid.html).
[6] “Safety Data for Sulfuric Acid.” University of Oxford. University of Oxford. 31 August 2005 (http://physchem.ox.ac.uk/MSDS/SU/sulfuric_acid_concentrated.html).
[7] Clark, Jim. “Le Chatelier’s Principle.” Chemguide. Chemguide. 2002 (http://www.chemguide.co.uk/physical/equilibria/lechatelier.html).
[8] Clark, Jim. “The Effect of Catalysts on Reaction Rates.” Chemguide. Chemguide. 2002 (http://www.chemguide.co.uk/physical/basicrates/catalyst.html).
21
Acknowledgements
A special thanks to:
Con Edison, Sponsor
Professor Okorafor, Chemical Engineering Advisor
Kemi Adeyeba & Albert Wang, Teaching Assistants
Mike Westbrook, Lab Technician
Susan Dorsey, Program Director
The Cooper Union, Program Site
22
Appendix I - Reaction Diagram
Reaction of Methanol and Acetic Acid
ON
OFF
50
0
25ON
OFF
50
0
25ON
OFF
50
0
25
AceticAcid
Methanol
Heater and Stirrer
23
Appendix II – Calculations
Acetic Acidm b
0.1319 -0.7931
Pump Setting (%)
Flow Rate (g/sec)
Flow Rate (mL/sec)
# of moles after 2 sec
Concentration (mol/mL)
20 1.8449 1.758722593 0.061445462 0.00848711825 2.5044 2.387416587 0.156329588 0.01600051230 3.1639 3.016110582 0.197496879 0.01605571335 3.8234 3.644804576 0.23866417 0.01609207740 4.4829 4.27349857 0.279831461 0.01611784115 1.1854 1.130028599 0.073995006 0.00754927235 3.8234 3.644804576 0.23866417 0.024506112
Methanolm b
0.1008 -0.5423
Pump Setting (%)
Flow Rate (g/sec)
Flow Rate (mL/sec)
# of moles after 2 sec
Concentration (mol/mL)
20 1.4737 1.861202324 0.091991261 0.02541247825 1.9777 2.497726699 0.123451935 0.02527089330 2.4817 3.134251074 0.154912609 0.02518756135 2.9857 3.770775448 0.186373283 0.02513266440 3.4897 4.407299823 0.217833958 0.0250937735 2.9857 3.770775448 0.186373283 0.03802912415 0.9697 1.224677949 0.060530587 0.012430599
CombinedPump
Setting (%)Flow Rate (mL/sec)
Amount collected in 2 sec (mL)
20 3.619924917 7.23984983425 4.885143286 9.77028657230 6.150361655 12.3007233135 7.415580024 14.8311600540 8.680798393 17.36159679
35/15 4.900804047 9.80160809415/35 4.869482525 9.73896505
24