OPTICAL SENSOR

BLOOD LEAK DETECTORNON-INVASIVE BLOOD LEAK DETECTION

Page 2: OPTICAL SENSOR BLOOD LEAK DETECTOR › fileadmin › media › ENG › ... · leaks. The optical sensor uses a light source with an opti-mum wavelength in order to provide the best

FOR SMALLEST TRACES OF BLOOD

TECHNICAL DATA

SONOTEC US Inc.

190 Blydenburgh RdSuite 8, 2nd FloorIslandia, New York 11749USA

phone +1 631 / 415 4758e-mail [email protected] www.sonotecusa.com

SONOTEC Ultraschallsensorik Halle GmbH

Nauendorfer Str. 206112 Halle (Saale)Germany

phone +49 (0)345 / 133 17-0e-mail [email protected] www.sonotec.eu

SALES & SUPPORT

SONOTEC preserves the right to change technical specifi cations without further notice. (Rev. 3.1 / 2018-11-02)

Certifi ed according to ISO 9001 & EN ISO 13485

The blood leak detector BLD is a non-invasive optical sensor, developed to detect smallest amounts of blood in clear fl uids on transparent plastic tubing. It was designed for medical devices such as dialysis machines.

As blood should not cross the blood / dialysate membrane, each dialysis machine needs a detector to monitor blood leaks. The optical sensor uses a light source with an opti-mum wavelength in order to provide the best sensitivity regarding the detection of the red blood cells. A threshold

Highly resistant to ambient light

IEC 60601-2-16 compliant

Customized designs available

Wavelength and sensitivity adjustable to different applications

Easy parameterization and testing

Made in Germany

triggers an alarm if the amount of blood in the fl uid exceeds the limits. The required sensitivity is defi ned by the inter-national standard IEC 60602-2-16. Accordingly, the sensor detects 0.35 ml / min of blood at a hematocrit level of 0.32 at the maximum specifi ed fl ow rate. Sophisticated algorithms in the sensor guarantee an excellent protection against ambient light. The sensor can be easily adjusted to similar applicati-ons by selecting a light source with a different wavelength. Modifi cations of the design or interfaces are available as customer-specifi c solutions.

MEASURING PRINCIPLE Optical, in transmission

TUBING Transparent plastic tubing

OUTER DIAMETER 5.5 to 7.0 mm

POWER SUPPLY +3.3 to +5.5 VDC, ≤ 50 mA

OPERATING TEMPERATURE 5 to 50 °C

RESPONSE TIME 100 ms

BLD01 BLD02

INTERFACE UART Digital

STANDARDSSafety requirements: IEC 60601-2-16:2008,

EMC: IEC 60601-1-2:2007

SOFTWARE

Development according to IEC 62304:2006, Safety class C

UNIQUE FEATURES