Upload
others
View
5
Download
0
Embed Size (px)
Citation preview
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Operator’s Manual
ST-120S Automatic Torque Tester
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 2 of 82 Rev A 28-Aug-17
Table of Contents
For convenience, this manual is divided into several chapters. A brief description of each chapter follows. Chapter 1: Overview ………………........................................................ pg 6 This chapter provides a brief introduction, common uses, and brief description on content to follow. Chapter 2: Specifications and Description .......................................... pg 8 This chapter contains information about the ST-120S and a detailed picture describing parts to the torque tester. Chapter 3: Unpacking and Inspection ............................................... pg 12 This chapter is a guide to ensure that the ST-120S is received in proper condition and that all necessary components have been located. Chapter 4: System Power Up ............................................................. pg 15 This chapter contains the required information for electrically and pneumatically connecting the system components and then powering up the system. Chapter 5: Tooling Set Up …………..................................................... pg 17 This chapter contains the required information for installing and adjusting the tooling that will be used for testing. Utilizes he Manual Mode Menu Chapter 6: Quick Start ....................................................................... pg 23 This chapter contains a step-by-step procedure to help the user become familiar with the basic operation of the ST-120S Chapter 7: Tester Settings ................................................................. pg 27 This chapter contains instructions on how to navigate and edit the ST-120S settings including Data output, screen format and more Chapter 8: Test Types and Their Menus ............................................. pg 30 This chapter contains a description of the ST-120S test types and the menus that are used to select and program them.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 3 of 82 Rev A 28-Aug-17
Chapter 9: Product Management ……............................................... pg 37 This chapter contains instructions on how to navigate and edit the Product
Management Menu so that the operator can enter/modify product specific variables.
Chapter 10: Example Product Management Setups .......................... pg 39 This chapter contains several examples of common product set-ups including cap tightening and release testing with CR, TEB, and CT caps. Chapter 11: Password Management ………........................................ pg 47 This chapter contains instructions on how to navigate and edit the administrative
passwords for setup and recalibration actions.
Chapter 12: Timers ………………………...…......................................... pg 48 This chapter contains instructions on how to navigate and edit the delay timers
during testing for all test modes.
Chapter 13: Data and Communication ..…......................................... pg 50 This chapter contains information stored and recalled in the Data Log Menu as well
as the data management and communication types of the ST-120S
Chapter 14: System Calibration......................................................... pg 52
This chapter is intended to acquaint the user with the required Calibration
Equipment, the Calibration Verification Process, and the process for performing a
User Calibration.
Chapter 15: Gold Bottle Verification .................................................. pg 63
This chapter describes the Gold Bottle and the procedure for using it to verify the calibration of the ST-120S.
Chapter 16: Troubleshooting ............................................................ pg 66
This chapter is a guide to assist in the recognition of any possible malfunctions, identification of their probable causes, and corrective actions to solve the problem.
Chapter 17: Preventative Maintenance ............................................. pg 68
This chapter is a guide to assist in the proper care and maintenance schedule to ensure long lasting quality and use.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 4 of 82 Rev A 28-Aug-17
Appendix 1: ST-120S Spare Parts Kit ................................................. pg 72
This addition is to show the operator common spare parts associated with the
ST-120S
Appendix 2: Torque Transducer Specifications ................................. pg 74
This addition contains information and specifications about the Torque Transducer
Appendix 3: Variables in Torque Testing ........................................... pg 76
This addition is to inform about the many variables that can affect torque readings
Appendix 4: Torque Testing Standards ….......................................... pg 81
This addition is a quick reference for the various standards including the ones
developed by ASTM and ISBT
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 5 of 82 Rev A 28-Aug-17
Conventions The following conventions are used in this manual.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 6 of 82 Rev A 28-Aug-17
Chapter 1: Overview
What’s Inside This manual is designed to provide the user with a step-by-step procedure from receipt of the system through operation, maintenance and calibration.
Introduction
Mesa Laboratories, Inc. is an industry leader in the manufacturing of precision torque measuring equipment. Over forty years of experience in the torque-testing field is reflected in every automated system. The SureTorque ST-120S Automated Torque Tester by Mesa Labs is a test instrument intended to be incorporated into a quality control program of bottle packaging companies. The ST-120S is a motorized testing equipment with a strain gage based torque transducer used to measure the removal/applied torque of threaded closures. The equipment was developed in accordance with various ASTM D10.31 and ISBT voluntary bottle testing standards. Mesa Labs recommends the user to obtain copies of the standards developed by the ASTM D10.31 subcommittee and ISBT (International Society of Beverage Technologists) for reference.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 7 of 82 Rev A 28-Aug-17
Common Uses The automated test system can be used for the following purposes: Cap tightening: The unit may be used as a portable precision cap tightener in low volume production facilities. Closure integrity testing: The torque required to remove the cap from the container threads is an indicator of packaging quality. Low removal torque values may indicate a poor cap product design or bad capping process which may result in product contamination or leakage. Testing the thread break torque within a quality control program (combined with accelerated aging and leak testing) will ensure that your packaged products are consistently and reliably closed and product leakage/contamination is minimized or eliminated. Package Development: Thread break/tamper evident band break/strip and applied torque testing can be utilized for package development and marketing. New materials and designs can be evaluated by comparing the value and repeatability of test results. Validation: The test results from the SureTorque tester can be used to validate and set the limits of the packaging process. Packaging Machinery Troubleshooting: The torque testing results can aid in troubleshooting a packaging machine configuration or operation.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 8 of 82 Rev A 28-Aug-17
Chapter 2: Specifications and
Description
What’s Inside This chapter contains information about the ST-120S and a detailed picture describing parts to the torque tester.
Technical Specifications
Testing
Torque Tests Cap release test, Final application, Fatigue, Tamper evident break
Torque Range 0-88 lbf•in (0-10N•m)
Torque Transducer Strain gauge
Accuracy +/- 0.5% for static loads over full scale
Repeatability +/- 0.2 lbf•in for one set up
Resolution Min 0.1 lbf•in
Maximum Overload 150%
Rotation 0-60 RPM, CW and CCW, Infinite
Modes Position, Speed or Torque control
Torque Units d•Nm, N•m, kg•cm, kg•m, oz•in, lbf•in
Calibration Certifiable (NIST)
Standards Meets or exceeds NIST, ASTM, FDA, GMP
Electrical
Control PLC-based system
Analog Signal 16bit/12.5ms conversion
Response Time (avg) <12.5ms
HMI Display 3.5” 256-Color Touch panel
Communication RS-232/RS-485 (USB, Ethernet optional)
Real-Time Clock 7-year typical battery backup
Utilities
Air Source 80-125 PSI, 5 SCFM
Power Source 100-240VAC, ~2.5A, 50-60 Hz
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 9 of 82 Rev A 28-Aug-17
Environment
Temperature Range 32-122°F (0-50°C)
Relative Humidity 10% to 95% (non-condensing)
Mounting Flat surface or desktop
Container Sizes
Height Range 1.0” (with base adapter) to 16.75”
Diameter Range 0.25” to 7.50” (with 10” platform)
Dimensions
Dimensions 31” x 12" x 16” (HxWxD)
Weight 65 lbs.
Materials of Construction
Exterior Anodized Aluminum and Stainless Steel
Hardware Stainless Steel
Tooling Anodized Aluminum, Nickle Plated Steel, Polyurethane, Acrylic, HDPE, UHMW
Tooling and Options
Tooling
Collet CT Standard threaded cap collet
Collet CR Child resistant cap collet
Angular Gripper Spray type or odd shaped cap gripper
Spring Loaded Serrated or non-threaded closures (beer cap)
Standard Clamp Set Round container clamps (left and right)
Clamp Adapter Set Adapts legacy tooling to new design
“V” Clamp Set Clamps for square or glass containers
Top load Chuck Chuck housing with adjustable spring top load fixture (5- 25 lbf). Use for child-resistant and push-to-turn closures.
Axial Float Chuck Chuck housing with floating mechanism. Use to minimize top load during testing for standard thread caps.
Height Base Adapter Platform base for short containers
Locating Base Adapter Locating base for square or round bottom containers
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 10 of 82 Rev A 28-Aug-17
Options
DAQ Option 1 Captures peak torque, time and date stamp. Windows based software, version 1.10 – 1.12.
DAQ Option 2 Captures Realtime Torque vs. Time or Angle vs. Torque. Windows based software, version 3.1.
DAQ Custom Consult factory
TorqueIQ SPC Software
FDA 21 CFR Part 11 compliant record keeping and Statistical Process Control Software
Serial Printer For all torque tester models. Interface: RS-232. Paper dimensions: 57.5, 69.5, or 76.0 mm (W) x 83.0 mm dia. Reliability: MTBF-180,000 hour
Laboratory Kit ASTM Calibration kit in protective case with 3 weights and 4 pulleys for 25 points of calibration.
Portable Kit ASTM Calibration kit in protective case with 1 weight and 2 pulleys for 3 points of calibration
Gold Bottle For daily “Quick Check” of calibration, 0-20 lbf•in range.
Safety Enclosure 1” frame with ¼” thick clear abrasion resistant polycarbonate for protection
Top load Monitoring Electronically controlled and monitored top load system
Automatic Product Detection
Automatic storage and recall of test formulas for fast switchover
Indicator Light System Red, Amber, and Green Indicator lights for easy pass/fail indication
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 11 of 82 Rev A 28-Aug-17
Description The ST-120S is a precision test instrument utilizing a unique torque and speed
controlled servo motor, a patented pressure controlled closure gripping
mechanism and a strain gage based torque transducer to measure the opening
and closing torque of threaded closure systems. Test methods were developed
based on various ASTM, ISBT standards and current testing practices at the
world’s largest bottle packaging companies. For general arrangement and change
part descriptions please refer to the pictures below.
Base
Dual Activation Start Buttons
Collet
Chuck
Self-Centering
Clamp Assembly
Clamps
NEMA power enclosure
(Optional)
Column
Back panel with pressure
regulators and air filters E-Stop
Touch Screen HMI+PLC
Head Assembly
Column height
Adjustment handles
Locating Base (Optional)
Air Filters
Quick Release Pin
Platform
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 12 of 82 Rev A 28-Aug-17
Chapter 3: Unpacking and
Inspection What’s Inside This chapter is a guide to ensure that the ST-120S is received in proper condition and that all necessary components have been located.
Packaging Your Torque meter was carefully inspected, both electrically and mechanically before shipment. Follow the unpacking procedure to carefully remove the tester and check for any signs of damage that may have occurred during shipment.
Immediately report any shipping damage to the shipping agent.
Unpackaging Procedure Step 1: Remove outer shipping shell. Retain and use the original packing materials in case reshipment is necessary. Step 2: Move inside box and place to side for later use.
Contains all change
parts purchased
along with standard
equipment
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 13 of 82 Rev A 28-Aug-17
Step 3: Unscrew 4 wing nuts securing machine to shipping frame.
Step 4: Carefully move the ST-120S to table top or testing platform.
Replace foam frame parts and wing nuts on bottom shipping frame.
Step 5: Remove lock screw for the Column Cable. Located on the
bottom of the Base plate of the ST-120S. (3/16” Hex)
IMPORTANT!! Lock screw for column cable must be
removed before any testing can be done.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 14 of 82 Rev A 28-Aug-17
Check List The following items are shipped with all ST-120S testers:
Input power cable
Input power supply
USB cord
Air hose with fitting
Operator’s manual
A Certificate of Calibration for your new ST-120S is included behind the front cover of the manual.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 15 of 82 Rev A 28-Aug-17
Chapter 4: System Power Up
What’s Inside This chapter contains the required information for electrically and pneumatically connecting the system components and then powering up the system.
Powering the ST-120S The system is energized by first inserting the external power supply into the receptacle on the back panel of the head (see Below) and then plugging the input power cable into an AC OUTLET.
Now by actuating the red rocker power switch located on the back panel, the system will be energized and ready for use.
The ST-120S is supplied for use with 100 - 240 VAC 50 - 60 HZ. Consult factory or your local sales representative for operation with other voltages. The use of a surge protector or power conditioner with the ST-120S is recommended as a protective measure against electrical noise.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 16 of 82 Rev A 28-Aug-17
Pneumatic Set Up The system is properly connected to an air supply by first connecting the air fitting supplied to a clean, dry, air supply of 80 psi at 4 cfm. Next connect the airline to the 1/8” NPT fitting at air filter on the back panel.
Initial Display After Power-up The following Main Menu will appear on the touch screen display whenever the system is turned on. The manufacturer and software version information is shown on the first “splash” screen during the boot-up procedure. Startup/Support Display Main Menu
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 17 of 82 Rev A 28-Aug-17
Chapter 5: Tooling Set Up
What’s Inside This chapter contains the required information for installing and adjusting the
tooling that will be used for holding the bottle, gripping the cap and positioning
the platform.
The tooling pictured here may not necessarily be the tooling shipped with your system however the descriptions for installing this tooling is still applicable for most other tooling packages.
Installing Tooling for Holding the Bottle
SureTorque testers require the use of product specific tooling, it is important to
use the correct change parts for the testing. Usually change parts are labeled for
easy identification.
To allow safe pneumatic actuations, lower the platform to the bottom of the
vertical support then release the locking handle. Lower the platform assembly to
the bottom of the column. Secure the platform assembly with the locking handle.
Now slide the clamps into the clamp bracket and secure the clamps by tightening
the thumb screw.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 18 of 82 Rev A 28-Aug-17
Custom Printed locating bases, including
those used for Automatic Product detection,
should be mounted on the platform before
adjusting the height. This is to incorporate the
added height of the base and ensuring
accurate placement to secure the container.
Installing Tooling for Gripping the Cap
All Mesa supplied chucks are mounted on the drive shaft and secured by tightening 2 hex set screws
If utilizing a standard, axial/floating, or top load chuck, a
pneumatic tube is needed and attached to the drive shaft. A
Collet is also needed to grip the closure. The collet is
inserted into the chuck, and secured with the quick release
pin.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 19 of 82 Rev A 28-Aug-17
Adjusting the Platform
The Platform was designed to position the container at a convenient height for loading and unloading test samples while providing for the minimum amount of vertical motion.
Step 1: Once the clamps, chuck, and collet are installed, adjust the platform
height by lowering/raising the platform until the clamp profile matches the
container profile. When the correct clamping position is found, secure the platform
by tightening the adjustment knobs.
Step 2: On the PLC, navigate to the Manual Mode by selecting Setup on the
Main Menu. The tester will prompt the operator for a password. (default
password is 1111). The operator will then select Manual Mode
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 20 of 82 Rev A 28-Aug-17
Step 3: Activate the platform cylinder by selecting the PLATFORM U/D key on
the HMI.
Step 4: Release the platform assembly locking handle and, using the crank
handle, raise the assembly with one hand while centering the bottle with the
other ensuring that the cap slides inside the collet.
Step 5: Make sure the collet does not interfere with any part of the container or
the tamper evident band. Generally, it is the best to insert approximately 3/5 of
the cap (excluding the TEB) in the collet, then secure the platform assembly with
the locking handle.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 21 of 82 Rev A 28-Aug-17
Step 6: Press the CHUCK key to squeeze the cap. This will also center the
container. The squeezing pressure may be adjusted with the mini regulator on the
rear panel. Try to manually rotate the container body to confirm that the collet is
holding the cap securely.
If the cap is slipping in the collet, make sure that:
A. the chuck cylinder or airline is not leaking (listen for the sound of
leaking air, or remove the collet and look at the actuating plastic ring
lowering and raising inside the chuck assembly).
B. the correct collet is used and is not excessively worn.
C. the air pressure controlling the squeezing force/pressure of the collet
is not too low.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 22 of 82 Rev A 28-Aug-17
Step 6: Press the CLAMP key to activate the self-centering clamp cylinder.
Inspect the setup for possible alignment issues. To make sure the clamps are
holding the container securely, deactivate the platform and chuck cylinders by
pressing the appropriate keys on the screen, then manually apply torque on the
cap while the container is clamped. If the container is slipping, make sure that:
A. the clamp cylinder or airline is not leaking (listen for the sound of
leaking air).
B. the correct clamps are used and they are not excessively worn.
C. the air pressure controlling the clamping force/pressure is not too low.
The clamp pressure can be adjusted with the “clamp” mini regulator
on the rear panel.
Step 7: The final step, check the overall setup, deactivate the clamp, the chuck
and the platform cylinders by pressing the corresponding keys and then activate
them again one by one in the correct order: first the clamps, second the platform,
third the chuck. If clearances, positions and pressures look good, the setup is
done. Press the BACK key to navigate to the Main Menu
It is always a good idea to document platform and column
heights for future reference. This will allow for quick and easy
adjustment for multiple container heights.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 23 of 82 Rev A 28-Aug-17
Chapter 6: Quick Start
What’s Inside This chapter contains a step-by-step procedure to help the user become familiar with the basic operation of the ST-120S.
Main Menu
The Main Menu allows the user to navigate to the main areas of the control system. This screen may be accessed from any of the displays by pressing the BACK key a couple of times.) Press the appropriate icon to test caps, calibrate the machine, explore historical measurement data or setup the machine.
SETUP: Manual operation and general electronic setup CALIBRATE: Calibration, linearity check and recalibration TEST: Torque Test, up to 20 pre-configured test profiles DATA LOG: Historical measurement data
Testing Navigate to the Test screen from the Main Menu by selecting the Test icon. Applied (clockwise) and Release (counterclockwise) testing may be initiated from here to determine the torque required to rotate a cap both in CW and CCW directions. By adjusting the rotational speed, torque ramp, maximum rotation and other variables, various torque tests may be performed. The sequence of operation and the operating principles are detailed below.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 24 of 82 Rev A 28-Aug-17
Test Settings:
BACK: Return to “Main Menu” PREV RECORD: Select previous test recipe NEXT RECORD: Select next test recipe SETTINGS: Displays test recipe settings screen COUNTER RESET: Resets batch counter COUNTER DETAILS: Displays counter results
Step 1: Place a container on the platform then press the dual activation safety start buttons. During this process, and anytime when pneumatic cylinders activate, keep yourself and others clear of the moving parts. Step 2: Once the clamps are closed and the chuck is activated, pinch points are eliminated and the product is firmly supported from all sides. You may release the start buttons now. The chuck starts rotating at the pre- defined velocity, torque ramp and acceleration. Check the display during the cycle to analyze the torque/ angle trend real-time. Step 3: During the test cycle, observe the container/collet for indication of concentricity, excessive side load or downward pressure, and fine-tune the mechanical setup if required.
Anytime during the operation, you may press the BACK key to stop the automatic test if the unit is not able to measure the torque/angle properly, mechanical or electrical adjustments are required, or if there is an emergency.
Step 4: Depending on the selected testing method, either when the maximum rotation is exceeded, or when the peak torque is measured, the chuck stops and the screen displays the peak value. The actuators return to home position and the product is ready for removal.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 25 of 82 Rev A 28-Aug-17
Default Settings SETUP: Default Password:1111
Product Management: Name: 1. Product
Cap ID: NA
Collet ID: NA
Container ID: NA
Clamp ID:: NA
Head height: 0.00
Left clamp: 0.00
Clamp height: 0.00
Right clamp: 0.00
Product Management 2/2: Cap type: CT
Cap material: #1: PET
Cap color: NA
Container type: ROUND
Container material: #1: PET
Container color: NA
TEB: NO
Liner: NO
Cap OD: 0.00
Cap ID: 0.00
Cap height: 0.00
Container OD: 0.00
Container ID: 0.00
Container height: 0.00
Container length: 0.00
Container width: 0.00
Passwords: Setup: 1111
Calibration: 1113
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 26 of 82 Rev A 28-Aug-17
Miscellaneous: Screen format: Data and Graph
Serial data: Result only - PC
Data entry: Enabled
Date/time: mm/dd/yy hh:mm:ss
Timers: Clamp on: 00:00:40
Chuck on: 00:00:50
Test start: 00:01:00
New cycle: 00:01:00
CALIBRATION: Default Password: 1113
Torque: Preferred Unit: 1 lbfin
Pulley Radius: 02.00 in
Weight: 010.000 lb
Pressure: Preferred Unit: PSIG
Reference Low: 0
Reference High: 80.0
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 27 of 82 Rev A 28-Aug-17
Chapter 7: Tester Settings
What’s Inside This chapter contains instructions on how to navigate and edit the ST-120S
settings including Data output, screen format and more
Navigation to Menu To view/edit the tester settings, first navigate to the Other Settings menu by
selecting Setup from the Main Menu. The tester will prompt the operator for a
password. (default password is 1111). Next select Other Settings
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 28 of 82 Rev A 28-Aug-17
Other Settings Menu More Settings Menu
From the Other Settings menu’s, the operator can specify:
Screen Format: Select from 3 display representations
Data Only: Displays only the numeric angle and torque
values.
Graph and Data: Provides not only numeric values on
the screen but also the angle vs. torque graph of the
measurement.
Pass/Fail: The most basic of the result representations, it
only displays whether the test passed or failed the
specifications.
Serial Data: The data output may be set to the following:
Result only – PC: Data output is formatted for
compatibility with the ST-DAQ software.
Result only – Printer: Data is formatted for printout on
a 3” wide paper roll.
Real-time – Angle vs. Torque: The data output is a
continuous stream containing the angle and corresponding
torque values.
Real-time – Angle vs. Time: Data output is a
continuous stream of the time and the corresponding
torque values.
No Output: When this option is selected, there is no
automatic data output from the torque tester.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 29 of 82 Rev A 28-Aug-17
Test Settings: The editing of test variables may be enabled/disabled by
changing this variable. See Chapter 8
Date and Time: Change the system date and time.
Start Switches: Determines what procedure happens when start buttons
are pressed.
Press Only: Machine begins testing when both switches
are pressed simultaneously
Press and Hold: Machine begins testing when both
switches are pressed simultaneously. Both start buttons
must be continuously held throughout testing otherwise
tester immediately stops running and realigns to starting
position.
Number of Test Controls Test recipes being displayed when Test Settings
are
Recipes to Display: disabled. Up to 49 test recipes. Test Recipes can be
navigated through on the Test Screen through the Next
Record and Previous Record buttons.
Run Offset CCW: Controls Dead band cutoff for calibration in CCW direction
Run Offset CW: Controls Dead band cutoff for calibration in CW direction
Pressure Alarm: Enable/Disable audible alarm due to Low Air Pressure
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 30 of 82 Rev A 28-Aug-17
Chapter 8: Test Types and Their
Menus
What’s Inside This chapter contains a description of the ST-120S test types and the menus that are used to select and program them.
Test Types The St-120S was designed to perform 5 test modes and can be selected in the Test Settings screen. To get to this screen from the Main Menu, Select the Test Option. The test mode can be selected and changed by pressing the TEST MODE: on the Navigation Screen
The 5 test modes include:
A. Release (counter-clockwise) B. Applied (clockwise) C. Release then Applied D. Applied then Release E. Strip
(other test modes are available upon customer customization)
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 31 of 82 Rev A 28-Aug-17
Enabling Test Settings In order to edit test settings for each test mode, the operator must first enable
the option by navigating to the Other Settings menu. The test settings for each
test mode are intentionally disabled due to minimizing operator error without first
applying the administrative password. To enable the test settings, first navigate to
the Other Settings menu by selecting Setup from the Main Menu. The tester
will prompt the operator for a password. (default password is 1111). Next select
Other Settings
In the Other Settings Menu, press the toggle switch for TEST SETTINGS to
enable editing of the test setting for each test mode.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 32 of 82 Rev A 28-Aug-17
Release and Applied Settings
After enabling the test settings, the operator can now edit product specific test
settings in the Test Settings. Navigate to the Test Settings by selecting Test from
the Main Menu. Next, toggle the Test Mode selection to display the product
specific Test Mode.
Product Name: Select from the list of available products.
Test Mode: Select from Applied, Release, Release then Applied or
Applied then Release test modes
Unit: Select from ozfin, lbfin, lbfft, gfcm, kgfm, or Nm
Platform U/D: Enable/Disable platform actuation. When doing a quick
accuracy verification with the Gold Bottle verification tool
or when testing a pump dispenser head, it is necessary to
disable the platform actuation.
NO of Cycles: For fatigue testing enter a number between 1 and 32,000.
This is the number of tests the machine will automatically
conduct on the supplied container/cap. For example, in an
applied then release test the number of cycles may be set
to 1000 in order to evaluate the torque degradation over
several repetitive opening/closing cycles.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 33 of 82 Rev A 28-Aug-17
To edit the test mode settings, press the Settings button to enter the Settings
Menu.
Release Settings Menu Applied Settings Menu
The operator can toggle between the Release Settings menu and the Applied
Settings Menu when the test mode selected is either Release then Applied or
Applied then Release.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 34 of 82 Rev A 28-Aug-17
From the Settings menu, the operator can specify:
Applied Set-point: The torque applied in a cap tightening cycle.
Torque Limit: The maximum permissible torque in a removal cycle.
Applied Fallback: Only applicable in Strip Test mode when the torque based
validation method is selected. This variable defines the
amount of drop in the torque that validates a peak as a
strip torque. One may also consider this variable as the
sensitivity of the strip test.
Release Fallback: When sensitivity based testing is enabled this variable
allows the automatic determination of thread break torque
(or other characteristic drops in torque). In other words,
the release fallback defines how sensitive the control is to
the torque variations/transients during a cycle.
Please note that the fallback values have no effect on the
operation when rotation based test methods are active.
Max rotation: The maximum allowed rotation during a test cycle.
(Degrees)
Ramp time (s): This is a timer defining the speed of the torque ramp up.
60s means the maximum torque is reached in 60s. When
the timer is set to 0, the torque applied during a cycle is
constant and defined by the starting torque value.
Starting torque: The torque applied at the beginning of the test. Default
value is 2.5%-5%, approximately 0 lbfin. 100% is
approximately 80 lbfin. (Units are in %)
Speed (rpm): The angular speed of the rotation during a test cycle.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 35 of 82 Rev A 28-Aug-17
Method: Torque: Recommended in Applied mode to immediately
stop the test upon reaching the set-point.
Rotation: Recommended in removal cycle where rotation
based validation is required. CR and TEB closures
Top load: Optional: It is possible to adjust the amount of force the
platform cylinder applies vertically while the product is
being tested.
High pressure: Recommended for most CT (continuous
thread) closures to determine the thread break torque or
apply torque to a set-point.
Low pressure: Recommended to measure the thread
break torque on CR (child resistant) caps. By adjusting the
platform height and the pressure in the platform actuator
(with the mini regulator on the rear panel) it is possible to
optimize the top load, making CR cap engagement
repeatable and minimizing the introduced friction.
“Zero”: Recommended when both the thread break and
the tamper evident band break torque must be measured
in one test cycle. This method allows a low adjustable
pressure to compensate for the weight of the platform
assembly and “float” the platform as the cap is being
unthreaded. The “floating platform” effect minimizes top
load and allows the accurate measurement of both thread
break and tamper evident bridge break in one test cycle.
Warning Limits: Configure these variables to display warning and error
& Error Limits: messages on the screen. (low error < low warning < pass
< high warning < high error). For example: if the
permissible removal torque range on a cap is 14-18lbfin,
set the low error and warning to 14lbfin, the high error
and warning to 18lbfin. When required, allow a separate
range for the warnings.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 36 of 82 Rev A 28-Aug-17
Operational Modes
Many different test modes may be created by selecting Rotation-based or
Torque based measurements and/or changing the adjustable platform
pressure/top load.
In Rotation based measurement, the chuck moves until the angle exceeds the
predefined limit, as opposed to the torque based measurement where the break
torque is automatically determined because the torque dropped after the thread
break or other distinctive break point (for example the engagement torque of a
child-resistant cap’s outer shell) is reached.
When the Torque based measurement is enabled the unit can automatically
determine the break torque with the help of the Release Fallback parameter. The
Fallback is the amount of drop in the torque the machine must measure in order
to qualify the peak torque as the thread break torque. The Fallback may be also
considered as the sensitivity of the automatic torque test.
When testing products with multiple “break” points, increasing the Fallback value
will eliminate the possibility of measuring a local force drop as the absolute thread
peak torque. On the other hand, when the Fallback is increased too much, the
tester may lose its ability to consistently recognize the real thread break torque.
Under these circumstances it may be preferable changing to the rotation based
testing method. This will allow the tester to rotate the chuck to the angle defined
by the maximum rotation variable and then display the peak.
To achieve repeatable and accurate results, we recommend that
the customers analyze every cap/container system and configure
the runtime variables accordingly.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 37 of 82 Rev A 28-Aug-17
Chapter 9: Product Management
What’s Inside This chapter contains instructions on how to navigate and edit the Product
Management Menu so that the operator can enter/modify product specific
variables.
Navigation to Menu These variables are for documentation purposes only; their use is optional and the
settings/values stored in these fields do not have any effect on the operation of
the machine. To navigate to the Product Management menu, first select Setup
from the Main Menu. The tester will prompt the operator for a password.
(default password is 1111). Next select Product Management
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 38 of 82 Rev A 28-Aug-17
Product Management Product Management Menu Product Management More Menu
Description of the product management data fields:
Name: Customer specified product name
Cap ID, Container ID: Customer specified cap and container IDs.
Collet ID, Clamp ID: Identification entry for each product specific tooling.
Platform height: The position of the top of the platform
Clamp height: The position of the top of the clamp slide
Left clamp, right clamp: When applicable, these fields can be used to
document the right and left clamp positions.
Cap, Container type, The type, shape, material and color of the
shape, material, color: cap/container system.
TEB, Liner: These fields may be used to document the presence
of a liner or tamper evident band.
Cap, Container Document product dimensions in these fields.
Dimensions:
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 39 of 82 Rev A 28-Aug-17
Chapter 10: Example Product
Management Setups
What’s Inside This chapter contains several examples of common product set-ups including cap tightening and release testing with CR, TEB, and CT caps.
Example 1: Cap Tightening Cycle This example is to demonstrate the process of setting up a new product in the product management menu and then assigning that product to a specific test profile. For this example, we will test a generic closure and assign it to a cap tightening recipe. Step 1: Navigate to the Product Management menu, first select Setup from the Main Menu. The tester will prompt the operator for a password. (default password is 1111). Next select Product Management. Step 2: Rename the Product to: Cap Test 1 Step 3: Return to Setup Menu and select Other Settings Step 4: In the Other Settings Menu, press the toggle switch for TEST
SETTINGS to enable editing of the test setting for each test mode. Set test
recipes to 1 then return to Main Menu by pressing the Back button twice.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 40 of 82 Rev A 28-Aug-17
Step 5: Navigate to the Test Menu
Step 6: Edit the Product selection by toggling the menu and selecting Cap Test
1. Toggle the Test Mode to Applied and Ensure Platform U/D is Enabled and
NO of Cycles is set to 1. Toggle the Unit to lbfin.
Step 7: Navigate to the Settings Menu and apply the following settings:
Applied Set-point: Based on closure specs (Usually ½ Diameter in mm)
Max rotation: 180
Torque Ramp: 30s
Start torque: 5%
Speed (rpm): 5 rpm
Method: Torque
Top load: High pressure
Low Warning: 0.000
High Warning: 80.000
Low Error: 0.000
High Error: 80.000
Step 8: Press the start switches to perform the Applied test
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 41 of 82 Rev A 28-Aug-17
Example 2: Release torque testing of a CT closure This example is to demonstrate the process of setting up another product in the product management menu and then assigning that product to a different test profile. For this example, we will test a CT (Continuous Thread) closure and assign it to a Release test recipe. A standard chuck is used in this example. Step 1: Navigate to the Product Management menu, first select Setup from the Main Menu. The tester will prompt the operator for a password. (default password is 1111). Next select Product Management. Step 2: Rename the Product to: Cap Test CT Step 3: Return to Setup Menu and select Other Settings Step 4: In the Other Settings Menu, press the toggle switch for TEST
SETTINGS to enable editing of the test setting for each test mode. Set test
recipes to 2 then return to Main Menu by pressing the Back button twice.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 42 of 82 Rev A 28-Aug-17
Step 5: Navigate to the Test Menu. Cap Test 1 is still associated to Test 1. Press
the Next Record to edit Test 2.
Step 6: Edit the Product selection by toggling the menu and selecting Cap Test
CT. Toggle the Test Mode to Released and Ensure Platform U/D is Enabled
and NO of Cycles is set to 1. Toggle the Unit to lbfin.
Step 7: Navigate to the Settings Menu and apply the following settings:
R Limit: Based on closure specs (Usually ½ Diameter in mm)
Fallback: .5
Speed (rpm): 5 rpm
Torque Ramp: 30s
Method: Torque
Start torque: 5%
Max Rotation: 360
Top load: High pressure
Low Warning: 0.000
High Warning: 80.000
Low Error: 0.000
High Error: 80.000
Step 8: Press the start switches to perform the Release test
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 43 of 82 Rev A 28-Aug-17
Example 3: Release torque testing of a CR closure This example is to demonstrate the process of setting up another product in the product management menu and then assigning that product to a different test profile. For this example, we will test a CR (Child-Resistant) closure and assign it to a Release test recipe. A top load chuck is used in this example. Step 1: Adjust the platform to engage the CR Cap. This will make the test repeatable and minimize the introduced friction. To Adjust the Platform, please refer to Chapter 5. Step 2: Navigate to the Product Management menu, first select Setup from the Main Menu. The tester will prompt the operator for a password. (default password is 1111). Next select Product Management. Step 3: Rename the Product to: Cap Test CR Step 4: Return to Setup Menu and select Other Settings Step 5: In the Other Settings Menu, press the toggle switch for TEST
SETTINGS to enable editing of the test setting for each test mode. Set test
recipes to 3 then return to Main Menu by pressing the Back button twice.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 44 of 82 Rev A 28-Aug-17
Step 6: Navigate to the Test Menu. Cap Test 1 and Cap Test CT are still
associated to Test 1 and 2 respectively. Press the Next Record to edit Test 3.
Step 7: Edit the Product selection by toggling the menu and selecting Cap Test
CR. Toggle the Test Mode to Released and Ensure Platform U/D is Enabled
and NO of Cycles is set to 1. Toggle the Unit to lbfin.
Step 8: Navigate to the Settings Menu and apply the following settings:
R Limit: Based on closure specs (Usually ½ Diameter in mm)
Fallback: .5
Speed (rpm): 5 rpm
Torque Ramp: 30s
Method: Rotation
Start torque: 5%
Max Rotation: 360
Top load: Low pressure
Low Warning: 0.000
High Warning: 80.000
Low Error: 0.000
High Error: 80.000
Step 9: Press the start switches to perform the Release test on a CR cap
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 45 of 82 Rev A 28-Aug-17
Example 4: Release torque testing of a TEB closure This example is to demonstrate the process of setting up another product in the product management menu and then assigning that product to a different test profile. For this example, we will test a closure with a TEB (Tamper Evident Band) and assign it to a Release test recipe. An axial floating chuck is used in this example. Step 1: Adjust the platform to engage the cap and ensure the platform is at a height that both the TEB and release torque can be measured (Float). The “floating platform” effect minimizes top load and allows the accurate measurement of both thread break and tamper evident bridge break in one test cycle. To Adjust the Platform, please refer to Chapter 5. Step 2: Navigate to the Product Management menu, first select Setup from the Main Menu. The tester will prompt the operator for a password. (default password is 1111). Next select Product Management. Step 3: Rename the Product to: Cap Test TEB Step 4: Return to Setup Menu and select Other Settings Step 5: In the Other Settings Menu, press the toggle switch for TEST
SETTINGS to enable editing of the test setting for each test mode. Set test
recipes to 4 then return to Main Menu by pressing the Back button twice.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 46 of 82 Rev A 28-Aug-17
Step 6: Navigate to the Test Menu. Cap Test 1, Cap Test CT and Cap Test CR
are still associated to Test 1, 2, and 3 respectively. Press the Next Record to edit
Test 4.
Step 7: Edit the Product selection by toggling the menu and selecting Cap Test
CR. Toggle the Test Mode to Released and Ensure Platform U/D is Enabled
and NO of Cycles is set to 1. Toggle the Unit to lbfin.
Step 8: Navigate to the Settings Menu and apply the following settings:
R Limit: Based on closure specs (Usually ½ Diameter in mm)
Fallback: .5
Speed (rpm): 5 rpm
Torque Ramp: 30s
Method: Rotation
Start torque: 5%
Max Rotation: 360
Top load: Zero pressure
Low Warning: 0.000
High Warning: 80.000
Low Error: 0.000
High Error: 80.000
Step 9: Press the start switches to perform the Release test on a TEB closure
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 47 of 82 Rev A 28-Aug-17
Chapter 11: Password Management
What’s Inside This chapter contains instructions on how to navigate and edit the administrative
passwords for setup and recalibration actions
Navigation to Menu Passwords are implemented to safe guard against accidental changes in the setup
as well as protecting the calibration from inadvertently being overwritten with
invalid measurements. To navigate to the Passwords menu, first select Setup
from the Main Menu. The tester will prompt the operator for a password.
(default password is 1111). Next select Password. Passwords can be edited here
and up to 4 numeric digits.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 48 of 82 Rev A 28-Aug-17
Chapter 12: Timers
What’s Inside This chapter contains instructions on how to navigate and edit the delay timers
during testing for all test modes.
Navigation to Menu To navigate to the Passwords menu, first select Setup from the Main Menu. The
tester will prompt the operator for a password. (default password is 1111). Next
select Timers.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 49 of 82 Rev A 28-Aug-17
Timers Menu Timers are in the format (mm:ss:hh) and can range from .5 sec up to 99 minutes.
Each timer can be selected to change the actuation delays.
Timers / Delays
Timers:
Clamp On: This timer is used to delay the clamp activation.
Chuck On: This timer delays the chuck cylinder after starting a test.
Test Start: Delays the rotation at the beginning of the test.
New cycle: Delay after a test is finished.
Suggested Delay:
Clamp On: 00:00.40
Chuck On: 00:00.50
Test Start: 00:01.00
New cycle: 00:01.00
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 50 of 82 Rev A 28-Aug-17
Chapter 13: Data and
Communication
What’s Inside This chapter contains information stored and recalled in the Data Log Menu as
well as the data management and communication types of the ST-120S
Data Log Menu The Data Log Menu can be accessed by pressing the Data Log icon from the
Main Menu.
The Data Log section contains both the Historical Data log as well as the Statistics
Log. Data can be sent to a PC for either a single record or for all stored data.
Historical Data Statistics
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 51 of 82 Rev A 28-Aug-17
Data Management and Communication Several communication options may be selected in the Setup/Other Settings
submenu. For more info on the selection please see Chapter 7: Tester Settings.
The serial communication is set to RS232, configured for 9600/8/n/1. Available
output options:
• The data output may be formatted to ASCII characters to be printed on 3”
roll paper.
• It may be formatted to ASCII strings, including peak torque value, date and
time stamps, etc. This is compatible with the previous SureTorque data
acquisition software.
• Data may be configured for real-time data acquisition to analyze torque
transients (analog data is sampled in approximately 10ms intervals at min.
16-bit resolution).
• Data may be saved in PLC memory without sending any data out on the
com port.
For more information on communication, contact the manufacturer.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 52 of 82 Rev A 28-Aug-17
Chapter 14: System Calibration
What’s Inside This chapter is intended to acquaint the user with the required Calibration
Equipment, the Calibration Verification Process, and the process for performing a
User Calibration.
General Calibration Information
The ST-120S is supplied with a factory calibration that is stored on the PLC.
Checking the calibration of the ST-120S at least once a year
is recommended when the system is used frequently.
Required Calibration Equipment
The following list of equipment is required to perform a Calibration. When ordering this equipment please provide the Model Number and Serial Number of the machine that the equipment will be used on. Minimum Required Equipment:
1. 2” Calibration Pully with wire 2. Chuck to house calibration pully 3. Calibration Frame 4. 10 lb. Calibration Weight (NIST Traceable)
All equipment can be found in the ST-120S calibration kit, 000-1413: Kit Includes:
1. Calibration Pullies: 0.5”, 1”, 1.5”, 2” Radius 2. Calibration Frame 3. Calibration Weights: 2 lb., 5 lb., 10 lb. (All weights NIST Traceable)
4. Protective Carrying Case
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 53 of 82 Rev A 28-Aug-17
Calibration Pully Set
The calibration pully set include a 0.5”, 1”,
1.5”, and 2” Radius pullies. The different sizes
are designed to allow for multiple calibration
points with the weight set. (see chart below)
Calibration Frame
The calibration frame is designed to be fixed to
the ST-120S column during calibration. It
includes a frictionless bearing that allows for
the most accurate calibration.
Calibration Weight Set
The calibration weight set consists of three weights that including 2 lb., 5 lb., and 10 lb. All weights are NIST traceable.
Protective Carrying Case
The Protective carrying case is designed to hold all calibration equipment in a foam designed insert.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 54 of 82 Rev A 28-Aug-17
Torque Calibration Weight Chart
The required calibration point in lbfin is produced by selecting the appropriate
weight to pully ration. This can be computed by multiplying the weight by the
pully radius to acquire a desired torque value.
This chart shows the weights and pullies that Mesa Supplies when a calibration kit
is purchased. The resulting field represents the lbfin the ST-120S should read.
Calibration Weights (lbs.)
Pully
Radius (in) 2 5 2 + 5 10 2 +10 5 +10
2 +
5 +10
0.5 1
lbfin
2.5
lbfin
3.5
lbfin
5
lbfin
6
lbfin
7.5
lbfin
8.5
lbfin
1 2
lbfin
5
lbfin
7
lbfin
10
lbfin
12
lbfin
15
lbfin
17
lbfin
1.5 3
lbfin
7.5
lbfin
10.5
lbfin
15
lbfin
18
lbfin
22.5
lbfin
25.5
lbfin
2 4
lbfin
10
lbfin
14
lbfin
20
lbfin
24
lbfin
30
lbfin
34
lbfin
Calibration Menu The Calibration Menu can be accessed by pressing the Calibrate icon from the
Main Menu.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 55 of 82 Rev A 28-Aug-17
Torque Calibration Equipment Set Up Procedure This procedure is used to detail how the calibration equipment is connected to the
system to perform a calibration verification as well as a user calibration.
Step 1: Release the platform assembly locking handle and, using the crank
handle, drop the assembly to its lowest point.
Step 2: Remove clamps, and the collet from the chuck. Insert desired calibration
pully and wire into chuck.
Quick Release Pin Location
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 56 of 82 Rev A 28-Aug-17
Step 3: Remove the platform by loosening the adjustment knobs and replacing
the platform with the calibration frame.
Step 4: Adjust the height of the column and calibration frame until top of bearing
is aligned with the center of the calibration pully wire. This is to ensure no
additional friction is added to the system.
Wire is parallel with benchtop
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 57 of 82 Rev A 28-Aug-17
Torque Calibration Verification Procedure This procedure is used to detail how to perform a calibration verification. To start
a verification, ensure the calibration equipment has been set up properly and the
calibration menu is shown on the PLC.
Step 1: Rotate the chuck so that the pulley is in a position that allows the wire to
wrap around the pulley a min. of 90 degrees. The wire must run parallel to the
base between the two pulleys).
Step 2: Lock the shaft by pressing the Lock or Unlock Shaft button. If it’s not
locked the chuck is free to rotate and the weight may fall or drive the chuck until
a hard stop.
Wire Connection Location in Back
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 58 of 82 Rev A 28-Aug-17
Step 3: With no weight added to the pully, press the Zero Offset and the Reset
Peak buttons to zero out the system for an accurate verification.
Step 4: Hang a weight from the pully wire and verify that the torque readout
matches the Calibration chart for the pully and weight used. The torque readout is
considered within range for a +/- 0.5%.
If the torque readout is out of tolerance, follow the User Calibration Procedure
below to recalibrate the display.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 59 of 82 Rev A 28-Aug-17
Torque User Calibration Procedure
This procedure is used to detail how to perform a Torque User Calibration to
recalibrate the tester. To start a torque calibration, ensure the calibration
equipment has been set up properly and the calibration menu is shown on the
PLC.
Step 1: Rotate the chuck so that the pulley is in a position that allows the wire to
wrap around the pulley a min. of 90 degrees. The wire must run parallel to the
base between the two pulleys).
Step 2: Lock the shaft by pressing the Lock or Unlock Shaft button. If it’s not
locked the chuck is free to rotate and the weight may fall or drive the chuck until
a hard stop.
Step 3: With no weight added to the pully, press the Zero Offset and the Reset
Peak buttons to zero out the system for an accurate verification.
Step 4: Press the RECAL button, the tester may prompt the operator for a
password. (default password is 1113).
Step 5: Accept or change the calibration variables (unit system [0-ozfin, 1-lbfin, 3-
kgfcm, 6dNm], pulley radius and weight). Use the numeric keypad to enter a
variable then press the ENTER key to accept the change. When all variables are
entered, follow the instructions on the bottom of the screen.
Step 3
Step 4 Step 2
Step 3
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 60 of 82 Rev A 28-Aug-17
Step 5: Unload the pulley and press the NEXT key. This registers the zero offset
of the calibration.
Step 6: Hang a weight from the pully wire on the right side of the pulley, let it
settle, then press the NEXT key. This registers the gain of the calibration for
counterclockwise (CCW) direction.
Step 7: Hang the weight on the left side of the pulley, let it settle, then press the
Next key. This registers the gain of the calibration for clockwise (CW) direction.
Step 8: The display is now calibrated both CW and CCW directions, press the
NEXT key again to return to the calibration check screen and test linearity by
hanging different weights on the pulley, or just press the BACK key to return to
the Main Menu.
The SureTorque tester is a precision torque measurement
device. The torque is transformed to an electrical signal with a
strain gage based force transducer. As in the case of any other
strain gage based measurement device, it is recommended to
check the calibration and/or re-calibrate the device as often as
possible. Usually the recalibration timeframe is not more than 6
months, but proper determination depends on many factors,
such as the duty cycle and the average load on the transducer,
also the possibility of overloading the strain gage. On-site
calibration services, calibration kits and certified weights are
optionally available from Mesa Labs.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 61 of 82 Rev A 28-Aug-17
Pressure User Calibration Procedure This procedure is used to detail how to perform a Pressure User Calibration to
recalibrate the tester. To start a pressure calibration, ensure the calibration menu
is shown on the PLC.
Pay special attention to pneumatic actuators operating at 80 PSI,
as pressure may cause serious damage to operator, equipment
and product.
Step 1: Press the Calibrate Pressure button.
Step 2: Check the Pressure Readout if readings are not in tolerance, recalibrate by
pressing the RECAL button and follow the on-screen instructions.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 62 of 82 Rev A 28-Aug-17
Step 3: Adjust the Mini Regulator for the chuck on the back panel all the way out
to produce a Low reference point. Select Next.
Step 4: Adjust the Mini Regulator for the chuck on the back panel all the way in to
produce a Low reference point. Select Next.
Step 5: Adjust the Mini Regulator for the clamp on the back panel all the way out
to produce a Low reference point. Select Next.
Step 6: Adjust the Mini Regulator for the clamp on the back panel all the way in to
produce a Low reference point. Select Next.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 63 of 82 Rev A 28-Aug-17
Chapter 15: Gold Bottle Verification
What’s Inside This chapter describes the Gold Bottle and the procedure for using it to verify the calibration of the ST-120S.
Gold Bottle Description
The Model 1612 Gold Bottle is used to quickly verify the
calibration and operational repeatability of the Model ST-
120S Torque Tester. This bottle is pre-set at the factory
to a customer specified torque value.
Each Gold Bottle is labeled with its exact torque rating.
Since the cap torque of this bottle is the torque
produced
by a permanent magnet assembly, the value will always
be
the same. Therefore, when tested on the ST-120S, the
measured value should always repeat unless something
has been damaged.
The bottle is normally supplied with the customer’s cap,
or with a custom printed cap (optional), so that the
customer’s Collet can be used in the testing process.
These Gold Bottles are readily available and their torque
can be certified on an annual basis
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 64 of 82 Rev A 28-Aug-17
Gold Bottle Verification
This procedure is used to detail how to perform a Gold Bottle verification. To start
a verification, Remove the clamps and ensure the calibration menu is shown on
the PLC.
Step 1: Lock the shaft by pressing the Lock or Unlock Shaft button. If it’s not
locked the chuck is free to rotate and the weight may fall or drive the chuck until
a hard stop.
Step 2: Insert the cap of the Gold Bottle into the chuck. A min of 60% of the
closure should be inserted. Press the Chuck On/Off to close the chuck around
the cap
Step 3: While the Gold Bottle is stationary, press the Reset Peak and Zero
Offset buttons. The Peak Torque reading should display 0.000
Step 4: Slowly turn the Gold Bottle clockwise until the peak reading is displayed.
Record the peak reading for validation
Step 5: While the Gold Bottle is stationary, press the Reset Peak and Zero
Offset buttons. The Peak Torque reading should display 0.000
Step 6: Slowly turn the Gold Bottle counter-clockwise until the peak reading is
displayed. Record the peak reading for validation.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 65 of 82 Rev A 28-Aug-17
Step 7: Check the verification rating on the Gold Bottle and compare with the
peak readings recorded and verify they are within specified tolerance. To check
the specified tolerance, please refer to the specifications table. If peak readings
are not in specified tolerance, Repeat Test. If verification fails a second test, a
calibration should be done.
Step 8: While holding the Gold Bottle, release the chuck by pressing Chuck
On/Off
Step 9: Unlock the shaft by pressing Lock or Unlock shaft key. Ensure Shaft
Status displays “Unlocked”.
Gold Bottle Specifications
Small
Torque Range: .5 – 4 in-lbs
Gold Bottle Tolerance: +/- 0.2 in-lbs
ST-120S Machine Tolerance: +/- 0.5 in-lbs
Standard
Torque Range: 3 – 16 in-lbs
Gold Bottle Tolerance: +/- 0.2 in-lbs
ST-120S Machine Tolerance: +/- 0.5 in-lbs
Large
Torque Range: 8 – 43 in-lbs
Gold Bottle Tolerance: +/- 0.2 in-lbs
ST-120S Machine Tolerance: +/- 0.5 in-lbs
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 66 of 82 Rev A 28-Aug-17
Chapter 16: Troubleshooting
What’s Inside This chapter is a guide to assist in the recognition of any possible malfunctions, identification of their probable causes, and corrective actions to solve the problem.
Troubleshooting
Refer to the Machine Tuning Sheet in the product manuals when making any
adjustments to the machine. This is a general troubleshooting guide; therefore,
some malfunction conditions and/or corrective applications may not apply to your
model Electric Torque Tester.
Only qualified personnel should troubleshoot this machine
All personnel should stay clear of moving parts
All guards and safety features must be replaced before this
machine is returned to service. Failure to comply with these
warnings may cause personal injury!
Mechanical Troubleshooting Guide
Malfunction Probable Cause
Unusually High
Torque Reading:
Check clearance between the top, inside surface of the Collet’s
relief cut counter bore, and the top of the container/closure.
Assure that this clearance is from .06” to .012”
Unusually Low
Torque Reading:
Check the rigidity of the container clamps
Container Must Not Rotate!!
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 67 of 82 Rev A 28-Aug-17
Electrical Troubleshooting Guide
Malfunction Probable Cause Corrective Action
ST-120S will not
Power-Up:
a. No AC Power to main Panel
b. Main AC fuse missing/blown
c. Main disconnect not in “on”
position
d. “Power On” switch faulty
a. Check Power Connections
b. Replace fuses
c. Turn disconnect to “on”
position
d. Replace rocker-switch
ST-120S will not
Start:
a. No AC Power
b. Control or main fuse
missing/blown
c. Line voltage not within +/- 10%
a. Check connections and
disconnect. Reconnect
power supply
b. Replace fuse
c. Install isolation X-former
Pneumatic Troubleshooting Guide
Malfunction Probable Cause Corrective Action
Cannot Get Proper
Air Supply:
a. Air regulator defective
b. Air gauge defective
c. Leak in air system
d. Insufficient air supply
a. Replace air regulator
b. Replace air gauge
c. Locate leak and correct
d. Check for restrictive kinks or
leaks in air hoses and
connections
Discrete Air
Components Not
Reading/
Functioning:
a. Excessive moisture in system
b. Component defective
c. Defective rear panel fuse
d. Low air Pressure
a. Check System filter
b. Replace component
c. Replace fuse
d. Check air supply and insure
air pressure at main
regulator at 80 psi
Water in Air
Supply:
a. Filter defective
b. Filter dirty
a. Replace filter
b. Clean or Replace filter
Discrete Air
Components
Malfunctioning:
a. Excessive moisture in system
b. Air supply dirty
c. Leak in component or hose
connections
a. Check filtering system
b. Check filtering system
c. Locate leak and correct
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 68 of 82 Rev A 28-Aug-17
Chapter 17: Preventative
Maintenance
What’s Inside This chapter is a guide to assist in the proper care and maintenance schedule to ensure long lasting quality and use.
Maintenance Schedule At every startup: Check for leaks
Optional: Verify the accuracy of the measurement system
using the Gold Bottle Verification, see chapter 15
Weekly: Check condition of change parts (asymmetric wear may
indicate setup/concentricity issues). If the polyurethane
surfaces in the change parts are hardened and/or dirty,
wipe them with a clean, soft cloth/paper towel and
isopropyl alcohol.
Monthly: Clean stainless-steel enclosures.
Clean all other parts with clean, soft cloth and Formula 409
Glass and surface cleaner or similar.
Check the air filter assembly for fluid/debris trapped inside
the bowl, replace cartridge and clean bowl if necessary.
Actuate the pneumatic cylinders in manual mode and
check for leaks (a drop in air pressure or hissing sound).
Check for worn/chafed/broken air hoses and electrical
power supply cords and replace if necessary.
Annual: Check calibration, see chapter 14. Recalibrate if necessary.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 69 of 82 Rev A 28-Aug-17
Cleaning Tips Stainless Steel Surfaces: To remove dust, use a clean, soft cloth.
To remove light soil, use a clean, soft cloth/sponge and
Formula 409 Glass and surface cleaner or similar.
To remove moderate and heavy soil, use a damp cloth
or sponge and clean with a non-abrasive cleaner such
as a baking soda paste then carefully rinse and dry.
To polish and prevent fingerprints, follow with Sheila
Shine or similar.
If the stainless steel enclosures show signs of rust, this
is typically only a surface discoloration and a mild,
nonabrasive stainless steel cleaner and a damp cloth
can be used to remove surface buildup and restore
shine.
Anodized Aluminum: Because anodizing is so hard, you want to use an
abrasive cleaning technique with a gentle soap.
Do not use harsh acidic or alkaline cleaners because
they may destroy the finish. Use solvents with care as
they may stain the finish. Regardless of the technique,
be sure to try a test area first.
One recommended technique is to use an abrasive
cleaning sponge with mild dish washing liquid. Always
try a test small area first to prevent a widespread
problem.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 70 of 82 Rev A 28-Aug-17
Acrylic Surfaces: Do not use cleaners containing ammonia. Cleaners such
as Windex or 409 will harm your new piece of acrylic.
Use only products specifically recommended for
cleaning acrylic such as Novus #1, or Brillianize, and a
soft cloth.
Never use a dry cloth or your hand to clean your acrylic!
This rubs the dirt and dust INTO the acrylic as much as
it rubs it OFF. First, blow the dust or dirt off, or use
water and a soft cloth to float the dirt off. Then use a
recommended cleaner to complete the job.
A mild solution of dish detergent and warm water takes
off stubborn dirt easily without harming the acrylic if
you've run out of Novus #1 or Brillianize.
Fine scratches can be removed with a mild abrasive
polish such as Novus #2, or Novus #3. Heavier
scratches, such that you can feel with your fingernail,
will require some elbow effort to remove with a bit of
sanding and buffing. A series of grits (150,400,600)
followed by a buffing wheel and available buffing
compound, will restore the acrylic surface.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 71 of 82 Rev A 28-Aug-17
Lubrication Tips Cylinders in SureTorque models can operate for years without lubrication. Generally all SMC cylinders in the machine (including CXSM and NCDG models) can operate without lubrication. The BIMBA “Pneu-turn” rotary actuator can operate without lubrication. Chuck cylinder: can operate without lubrication for years Should you decide to lubricate pneumatic components, please read the recommendations below. Lubrication of non-lube type SMC cylinders: These cylinders are lubricated at the factory and can be used without any further lubrication. Should you decide to lubricate, use class 1 turbine oil (without additives) ISO VG32. Stopping lubrication later may lead to malfunction due to the loss of the original lubricant. Therefore, lubrication must be continued once it has been started. Lubrication of a lube style SMC cylinders: Install a lubricator in the pneumatic circuit and use Class 1 turbine oil (with no additives) ISO VG32. Do not use machine oil or spindle oil. The Pneu-Turn Rotary Actuator is pre-lubricated at the factory for extensive, maintenance-free operation. However, the life of the rotary actuator can be lengthened by providing additional lubrication with an air-line mist lubricator or direct introduction of oil to the actuator every 500 hours of operation. Recommended oils for Buna N seals are medium to heavy inhibited hydraulic and general-purpose oil. If High Temperature seals, use Dow Corning #710. Other types of prelube are available upon request. The rack and pinion gear and ball bearings are pre-lubricated at the factory for extensive, maintenance-free operation. If additional lubrication should be required, use a high-grade bearing grease. Chuck cylinder: every 3 yrs. replace the rubber O-rings and apply a light coating of silicone based lubricant on the new O-rings. Only a thin film of lubricant should be used to prevent seal damage.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 72 of 82 Rev A 28-Aug-17
Appendix 1: ST-120S Spare Parts
Kit
What’s Inside This addition is to show the operator common spare parts associated with the ST-
120S
Spare Parts Kit
000-1022 Assy, Cable, Counter Balance, 15.56"
066-1637 Torque Transducer, 0-10Nm 066-1047 Motion
controller
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 73 of 82 Rev A 28-Aug-17
066-0974 Expansion module 066-0975 Adaptor I/O
066-1046 Stepper motor 000-1446 PCA, Encoder Interface Circuit
066-0945 Solenoid Valve, 4 Port 066-1197 Filter, Air
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 74 of 82 Rev A 28-Aug-17
Appendix 2: Torque Transducer
Specifications
What’s Inside This addition contains information and specifications about the Torque Transducer
Specifications This device utilizes an electronic strain gauge to measure very small amounts of
changes as it is flexed or strained in either CW or CCW directions. The mechanical
changes are measurable increases or decreases in electrical resistance. Our
transducer has the ability to compensate over a wide range of temperatures and
hold its accuracy for a cycle of over 1,000,000 measurements (under the specified
operating conditions).
Reaction type torque transducer
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 75 of 82 Rev A 28-Aug-17
Transducer Specifications
Accuracy: 0.1% FS
Available Ranges: 5, 10, 20 Nm
Output: 2.0 mV/V ± 0.1%
Zero Balance: ±1% max.
Temperature Effect Zero: ±0.05% / 10K max.
Span: ±0.05% / 10K max.
Non-linearity: ±0.1% max.
Hysteresis: ±0.1% max.
Non-repeatability: ±0.03% max.
Creep over 30 minutes ±0.04% max.
Input Resistance: 350 ohm ± 2
Output Resistance: 350 ohm ± 2
Excitation (recommended): 6-10 V
Excitation (Max): 12 V
Insulation Resistance: 5 Gohm min.
Temperature Compensated: -10°C to +40°C
Operating: -20°C to +65°C
Storage: -50°C to +85°C
Overload, Safe: 150%
Breaking Load: 300% max
Protection: IP 65
Corrosion protection: Chem Nickel
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 76 of 82 Rev A 28-Aug-17
Appendix 3: Variables in Torque
Testing
What’s Inside This addition is to inform about the many variables that can affect torque readings
Applied Torque In most cases, the higher the applied torque, the higher the thread break torque.
This is true up to the strip torque where the threads break/deform irreversibly in a
cap tightening cycle. Without the cap/container manufacturer’s guidance, it is
recommended to start with an application torque that equals to the cap outer
diameter in mm divided by two (in inch-pounds [in-lbs., lbf-in]). The release
torque / applied torque quotient depends on the specific cap design. It is usually
in the 0.6-0.9 range and higher for glass and lower for plastic bottles. It is not
uncommon to see values out of this range and dwell time usually decreases the
quotient over time.
Sensitivity of Automatic Thread Break torque measurement In the automatic detection of the thread break torque, two methods may be used
to validate the result.
A. Fallback based peak torque validation: this is the fastest and most cost-
effective way to measure thread break torque on a cap. In CR cap
applications, special attention must be made when fine tuning the fallback
value to avoid validating the shell engagement as the thread break torque.
B. Rotation based peak torque validation: to overcome the problem introduced
by the torque drop during the CR engagement an additional encoder can be
used to validate the thread break torque. The rotation limit must be set
according to the worst-case scenario of the shell engagement.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 77 of 82 Rev A 28-Aug-17
C. If fallback based validation is used during a measurement and the fallback is
set at <1.5lbfin, the tester will stop in <25 degrees and display the
engagement torque. To avoid the false readout, either the fallback must be
increased above 1.5lbfin (the recommended fallback for CR caps is 2.5lbfin),
or rotation based validation must be used and the rotation limit set at ~40
degrees.
D. See the top load vs. rotation and torque vs. rotation trends below to
understand the variations during a CR cap removal cycle. The vertical axis
represents both toque (lbfin) and top load (lbf) while the horizontal axis is
the rotation in degrees.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 78 of 82 Rev A 28-Aug-17
Cap and container squeezing pressure, top load The variation in cap/container squeezing pressure and the top load on the cap
may also affect the torque reading. The pressure variation is caused by either
force or contact area variation. The larger the contact area and/or the higher the
force compressing the cap and the container threads, the higher the torque
readout will be. Thus, in some applications it is important to monitor the container
and/or cap squeezing pressures and the top load force.
Speed of torque ramp, update rate of torque digitization There are two phenomenon a package engineer must be aware of when setting a
torque ramp setting:
A. When the torque ramp up is fast compared to the conversion time of the
digitizing device, the removal torque readout on the digital machine can be
considerably lower than the real peak due to the slow analog to digital
conversion speed. The error originating from the low sampling speed is not
to be confused with the quantization error. To understand the error
originating from inappropriate digitization, look at the graphs below and/or
find more information on the internet about the Nyquist-Shannon sampling
theorem and resolution/quantization noise.
Optimal chuck pressure:
The contact area is “minimal” and
the torque readout is not affected.
High chuck pressure and/or excessive
top load: The contact area is increased
resulting in higher torque readouts.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 79 of 82 Rev A 28-Aug-17
B. When the torque ramp up time is slower, the removal torque tends to be
lower because the gradually increasing fatigue lowers the peak force
required to finally break the threads. If the torque ramp is faster, the thread
break torque is usually higher.
Even if a manual torque tester had fast digitizer circuit, the lack of torque
ramp control may still cause considerable variation from one operator to
another, and even for one operator depending on how fast he/she manually
applied torque on a specific package.
Analog torque signal (red) sampled with a 12 bit, 100ms AD converter. The peak readout with this digitizer is ~4.9 lbfin at
200ms.
The same analog signal (red) sampled (green) with a 12 bit, 10ms AD converter. The peak readout with this
system is ~5.5 lbfin at 160ms.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 80 of 82 Rev A 28-Aug-17
Dwell Time In various experiments, it has been established that release torque levels are
highest immediately after application and then gradually decrease to a stable level
over a period of time (days/weeks). The rate of the release torque decay is
greatest in the first couple of hours/days and then reduces at a decreasing rate
before reaching its stable level. Production processes such as hot filling or those
using heat activated glue cap systems can produce a big difference in the release
torque readout when compared with results measured in a laboratory
environment.
Product Variations Minor changes in mold, material and liner can be also major contributing factors in
torque variations. See an example of different liner alignments and how it relates
to the contact surface area and the release torque.
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 81 of 82 Rev A 28-Aug-17
Appendix 4: Torque Testing
Standards
What’s Inside This addition is a quick reference for the various standards including the ones
developed by ASTM and ISBT
ASTM Developed by D10.31:
• D2063-91(2002) Standard Test Methods for Measurement of Torque
Retention for Packages with Continuous Thread Closures
• D3198-97(2002) Standard Test Method for Application and Removal Torque
of Threaded or Lug- Style Closures
• D3469-97(2002) Standard Test Methods for Measurement of Vertical
Downward Forces to Disengage Type IIA Lug-Style Child-Resistant Closures
• D3470-97(2002) Standard Test Method for Measurement of Removal Lug
Strippage of Type IIA Child-Resistant Closures
• D3472-97(2002) Standard Test Method for Reverse-Ratchet Torque of Type
IA Child-Resistant Closures
• D3474-90(2002) Standard Practice for The Calibration and Use of Torque
Meters Used in Packaging Applications
• D3475-07 Standard Classification of Child-Resistant Packages
• D3481-06 Standard Test Method for Manual Shelling Two-Piece Child-
Resistant Closures That Are Activated by Two Simultaneous Dissimilar
Motions
• D3810-97(2002) Standard Test Method for Minimum Application Torque of
Type IA Child- Resistant Closures
• D3968-97(2002) Standard Test Method for Monitoring of Rotational Torque
of Type IIIA Child- Resistant Closures
• D7257-06 Standard Test Method for Automated Shelling Two-Piece Child-
Resistant Closures That Are Activated by Two Simultaneous Dissimilar
Motions
Mesa Laboratories • 12100 6th Awe. • Lakewood, CO 80228
Phone: (303) 987-8000 • Fax: (303) 987-8989 • E-mail: [email protected] • Web: torque.mesalabs.com
Page 82 of 82 Rev A 28-Aug-17
• D5094-90 Standard Test Methods for Gross Leakage of Liquids from
Containers with Threaded or Lug-Style Closures
• D5419-95 Standard Test Method for Environmental Stress Crack Resistance
(ESCR) of Threaded Plastic Closures
ISBT Voluntary Standard Test Methods for Plastic Flat Top Closures
• Top Closures
• Back-off
• Ball Impact
• Drop Impact
• Elevated Temperature Cycle
• Opening Performance
• Pressure Retention: Zahm Nagel
• Pull Up
• Removal Torque
• Secure Seal Test
• Security
• Strip Torque
• Top Load Vent
• Vacuum Retention
International Test Standards • DIN EN 12377:1998 Packaging - Flexible tubes - Test method for the
tightness of closures
• DIN EN 14401:2004 Rigid plastics containers - Methods to test the
effectiveness of closures
• SS-EN 12377 Packaging - Flexible tubes - Test method for the air tightness
of closures
USPC Test Methods • CHAPTER 671: Containers – Permeation