OPERATION/MAINTENANCE MANUAL

PORTABLE TRITIUM MONITOR

MODEL 400AC

OVERHOFF TECHNOLOGY CORPORATION 1160 US ROUTE 50, MILFORD, OHIO, USA

Page 2: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

~~OVERHOFF TECH OLOGY CORPORaTIO

CE DECLARATION OF CONFORMITY

We, The Manufacturer: OVERHOFF TECHNOLOGY CORPORATION 1160 U.S. Route 50, Milford, Ohio, 45150

Declare under our sole responsibility that the following equipment:

Description: Tritium in Air Monitor, Portable Model Series 400

Model Designation: 400AC

Application of Council Directive (s):

Electromagnetic Compatibility Directive (EMC) 89/336/EEC and amending directives 91/263/EEC, 92/31/EEC, 93/68/EEC

Low Voltage Directive 73/23EEC, and amending directive 93/68/EEC

to which this declaration relates is in conformity with the relevant provisions of the following standard(s) when installed in conformance with the installation instructions contained in the product documentation:

EN61326 EN61 000-3-2 EN61 000-3-3 EN61 000-4-2 Level 2 EN61 000-4-3 EN61 000-4-4 EN61000-4-5 EN61 000-4-6 EN61000-4-11

Pertinent LVD section of :

BS EN 61010-1

Class A, Radiated/Conducted Emissions Limits Conducted Emissions, Equipment Class A Conducted Emissions (Voltage Fluctuations) Electrostatic Discharge Immunity Electromagnetic Field Immunity Electrical Fast Transient Immunity Surge Immunity RF Conducted Immunity Voltage Dips Immunity

Safety requirements for electrical equipment for measurement, control and laboratory use, Part 1: General requirements: 2001

Technical information is maintained at: Overhoff Technology Corporation 1160 U.S. Route 50, Milford, Ohio 45150

We, the undersigned, her by declare at the product(s) specified above conforms to the listed directive(s) and standard(s)

Signature:

Full Name:

Position:

PO ox 82. 1160 U.S Route SO. i forcf, Ohio. 45 0-9705 usa Tel phone 3 248 240 Fac imil 5 3 248 24 2

e-m iI [email protected] we 5 e h p www.overo.com

Page 3: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

SAFETY NOTICE

This Tritium Monitor has been designed and tested in accordance with EN 61010-1. To ensure that the monitor is used safely, follow all safety and operating instructions in this manual. If the monitor is not used as described in this manual, the safety features of the monitor might be impaired.

� Do not use the monitor unless it is fully assembled, housed inside the case secured with the screw fasteners it was supplied with.

� Turn off the power and remove the plug connection from the AC power converter before removing batteries.

� Make sure the battery covers are properly closed and secured.

� Remove the battery from the monitor if the monitor is to be stored for long periods.

� This instrument has not been designed for indiscriminate opening or disassembly of the internal parts. The bias voltage batteries are always connected to the ionization chamber cans, even when the instrument is switched OFF. The electronics contains highly sensitive semiconductors which are destroyed by even the slightest electrostatic discharge

� Not suitable for use in wet locations

� Not suitable for use in explosion hazard environments

� Refer to instruction label when connecting to the external jack for supplementary power input. DO NOT EXCEED 3.5VDC! Use only the AC power converter that was supplied with the monitor.

SYMBOLS

The following international symbols are used with this manual and equipment:

Important Safety Information in Manual

DC (direct current)

Page 4: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

TABLE OF CONTENTS SECTION PAGE 1.0. INTRODUCTION 1 1.1. PHYSICAL DESCRIPTION 2.0. TECHNICAL SPECIFICATIONS 2 3.0. CIRCUIT DESCRIPTION 4 3.1. IONIZATION CHAMBERS 3.2. ELECTROMETER 3.3. SIGNAL PROCESSING AMPLIFIER 5 4.0. CONFIGURATION 6 4.1. EXTERNAL FEATURES 4.2. HOSE CONNECTIONS 5.0. OPERATION 7 5.1. RS232 OUTPUT 8 6.0. CALIBRATION METHODS 9 6.1. CALIBRATION SOFTWARE INSTALLATION 10 6.2. GAMMA CALIBRATION SETUP 11 FIGURE 1: GAMMA CALIBRATION SETUP 6.3. GAMMA CALIBRATION PROCEDURE 12 6.4. CALIBRATION SOFTWARE FRONT PANEL 14 6.5. CALIBRATION SOFTWARE MENUS 15 6.6. CALIBRATION REPORTS 16 6.7. HOW THE GAMMA CALIBRATION SOFTWARE WORKS 17 6.8. TRITIUM GAS CALIBRATION 18 6.9. TRITIUM GAS CALIBRATION PROCEDURE 6.10. A FINAL NOTE ON CALIBRATION 7.0. MAINTENANCE 19 7.1. D-CELL BATTERY REPLACEMENT 20 8.0. SERVICE AND SUPPORT 21 9.0. WARRANTY 10.0. REPLACEABLE PARTS

FIGURE 2: FRONT PANEL CONTROLS 22

FIGURE 3: HOSE BARB CONNECTION 23

FIGURE 4: GAMMA CHECK LOCATION 24

FIGURE 5: OUTPUT CONNECTIONS 25

FIGURE 6: J2 OUTPUT CABLE DRAWING 26

DRAWING, FUNCTIONAL BLOCK DIAGRAM 27

DRAWING, GENERAL ARRANGEMENT SH 1 OF 2

DRAWING, GENERAL ARRANGEMENT SH 2 OF 2

Page 5: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 1

1.0. INTRODUCTION

The Model 400AC portable tritium monitor is a small, high sensitivity, hand held, rechargeable battery operated, and fully gamma-compensated survey meter with RS232 serial data output and user-recalibration features. Note: Do not use non-rechargeable batteries in the 400AC.

1.1. PHYSICAL DESCRIPTION

The Model 400AC uses four ionization chambers arranged in a cruciform pattern. Two ionization chambers, with a total volume of 400 cc, are used for measurement; the other two chambers, with an identical total volume of 400 cc, serve for gamma compensation. The sample stream is drawn through the ionization chambers by means of a small rotary vane pump, which is plumbed to the outlet of the furthest down stream ionization chamber. It is recommended to always use a good quality particulate filter at the instrument sampling inlet to prevent entry of dust particulates.

A large easy to read liquid crystal digital panel meter with a range from 0.1 to 1999.9 MBq/m3 (3 to 54000 μCi/m3) is used for measurement display. Other units of measurement, such as μCi/m3, DAC and μSv/hr, or others may be specified by the user at time of requesting a unit. The instrument exhibits a basic sensitivity of the order of 0.1 MBq/m3, which it is able to attain due to the fact that it is immune to response to both radon, gamma, and cosmic ray noise.

A pair of “D” size NiMH batteries supplies power. NiCd batteries are not recommended, because of undesirable “memory” characteristics. Onset of battery depletion is signaled by illumination of an red LED located next to the meter face. An external power supply can be used by attaching to a small receptacle located on the side of the instrument case. When new, the batteries will fully charge in 10 hours time, and can power the 400AC for over 16 hours (with the pump on). The same LED also signals red when the high voltage power supply (HVPS) used for ionization chamber bias has a malfunction.

A ten-position alarm selector switch, adjustable over partial scale 0.2 – 100 MBq/m3 is located on the front panel. An OFF position is included. An acoustic signaler emits a steady tone if the measurement exceeds the set point. An intermittent tone is heard if the sample air flow has been interrupted. The acoustic alarm is silenced by a “MUTE” push button. The alarms are non-latching. A toggle switch on the front panel switches the gamma compensation chambers ON/OFF. For normal operation, the GAMMA COMPENSATION switch should be in the ON position.

An RS232 serial data output has been provided. The instrument case is constructed of light weight aluminum. A handle is attached for hand held survey use.

Gas flow connections are made externally to the instrument by appropriate attachment of flexible plastic hose. See Figure 3. An external disposable in-line filter has been provided.

Page 6: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 2

SENSITIVITY 0.1 MBq/m3 ACCURACY ±10% GAMMA COMPENSATION four chambers in a cruciform pattern to completely eliminate

errors due to external gamma radiation Gamma response less than 0.2 μCi/m3 per mR/hr.

GAMMA COMP. ON/OFF a front panel toggle switch allows the user to manually

switch the gamma compensation ON or OFF. RESPONSE RATE 30 seconds to reach 90% of final reading, NOISE LEVEL ±0.1 MBq/m3, 1 S.D. (10 second electronic time constant) ZERO STABILITY after 30 seconds (or less) warm up, zero drift less than ±0.1 MBq/m3) ALARM (ACOUSTIC) 1. ten position selector switch set point for signal alarm 0.2 – 100 MBq/m3, steady tone. An OFF position is included. 2. low flow produces an intermittent tone 3. Mute switch silences audible tone ALARM (VISUAL) LEVEL: red LED; when tritium level exceeds set point FLOW: yellow flashing LED; low pump flow LOW BAT: red LED; D-cell batteries need to be replace

-AND- HVPS: red LED illuminates to indicate a malfunction

with the high voltage power supply (HVPS) used to bias the ionization chambers

EXTERNAL CONNECTIONS RS232 serial data output for tritium measurement, level alarm status, and calibration DUST FILTER in line disposable cartridge, Pall P/N 12082 SAMPLING SYSTEM 2 hose barb ports are located on the front panel PUMP miniature internal pump for 3 to 5 volume changes per minute, typically 1.5 to 2 LPM.

2.0. TECHNICAL SPECIFICATIONS, MODEL 400AC MEASUREMENT DISPLAY 4 1/2 digit LCD 0.1 to 1999.9 MBq/m3 (3 – 54000 μCi/m3)

Page 7: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 3

TECHNICAL SPECIFICATIONS, MODEL 400AC, continued IONIZATION effective volume: 400 cm3 CHAMBER VOLUME port to port volume: 440 cm3 POWER two “D” size NiMH batteries, external jack for charging Note: do not use non-rechargeable batteries! ENVIRONMENTAL 0° C to +40° C 20 to 90 % R.H. Non-Condensing CASE light weight aluminum SIZE AND WEIGHT 7.6" L, 5.2" W, 6.9" H excluding handle, 6.5 lbs (3 kg) ACCESSORIES

• 2 “D” size NiMH batteries (installed) • Sniffer hose • Dust filter • 2.3 meter long cable for RS232 serial data. Mini-

DIN plug for J2 output connector at one end with DB9 plug for serial connection on other end

• Power converter 100-240 VAC, 50/60 Hz, .25 A to 3.3 Vdc @ 1.2A-1.5A 5.5 mm O.D. x 2.1 mm I.D. Plug Center pin is positive DO NOT SUBSTITUTE, use only the recommended power supply

CALIBRATION choice of two modes: Automatic Calibration ( with included PC-based software ) and Manual Calibration ( by adjusting a calibration potentiometer ).

Page 8: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 4

3.0. CIRCUIT DESCRIPTION

3.1. IONIZATION CHAMBERS

In its simplest form, an ionization chamber is an enclosed volume with two electrodes. Voltage is applied between the electrodes, generating an electric field, which will segregate and collect electric charges, which are created by nuclear events occurring inside the chambers. Nuclear events may consist of ionization of air molecules by external or internal alpha, beta or gamma radiation.

The 400AC monitors are designed to measure tritium. Activity of tritium decay is such that a concentration of 1 μCi/m3 in a volume of one liter will generate an ionization current of about 0.95 x 10-15 amperes. This is a very weak current.

Alpha pulses from naturally occurring radon, are much more energetic, they can produce short current bursts of up to 10-13 amperes during decay, and therefore appear as large noise "spikes" which can seriously impair tritium measurement.

Gamma radiation also has a strong effect. In practice, a gamma radiation field of 1 mR/hr will create approximately the same amount of ionization as 90 μCi/m3 of tritium.

A tritium monitor, in order to measure to low concentrations, must be able to respond only to tritium and be immune to alpha or gamma radiation. For this purpose, a second ionization chamber system has been included to balance out any ionization current contribution from external gamma radiation.

In the 400AC instruments, the four ionization chambers are arrayed in a cruciform pattern, ensuring almost perfect gamma compensation in all directions.

The ionization chamber polarizing voltage is supplied by an electronic high voltage power supply. The surfaces of the ionization chambers have a thin coat of paint for insulation, but it is best to avoid touching them.

3.2. ELECTROMETER

Also known as a transimpedance amplifier, it serves the purpose of converting the extremely feeble ionization current into a voltage suitable for further signal processing and measurement display.

The heart of the electrometer consists of a specially selected dual ultra high impedance semiconductor device which has been chosen both for ultra low internal current leakage as well as long term d.c. stability.

In the 400AC instruments, the electrometer is directly attached to the ionization chamber cluster and is protected by a rectangular metal cover.

CAUTION: This instrument has not been designed for indiscriminate opening or disassembly of the internal parts. The bias voltage batteries are always connected to the ionization chamber cans, even when the instrument is switched OFF. It contains highly sensitive semiconductors which are destroyed by even the slightest electrostatic discharge

Page 9: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 5

3.3. SIGNAL PROCESSING AMPLIFIER

A single printed circuit board attached directly to the front face of the instrument contains all power supply and signal processing electronics.

Proprietary circuitry is used for the recognition and elimination of transient signals due to radon or high-energy cosmic ray pulses. The 400AC instruments, with digital display, use a dedicated internal circuit to disable the pulse rejection circuit when the measured signal reaches approximately 3.0 MBq/m3 (60 μCi/m3).

An OFFSET control is furnished in order to adjust the reading to zero in case of offsets caused by tritium contamination of the chambers or otherwise.

A front panel control has been provided for adjustment of the set point (level) at which the acoustic alarm is desired to sound. The acoustic signaler has the second function of alerting the user that sample gas flow is impeded.

All power supplies are regulated. A small warning LED on the front panel will illuminate when the battery terminal voltage has dropped to about 2.2 V. This signals that the batteries should be recharged – WITHIN 1 HOUR AFTER THE WARNING LED ILLUMINATES. It will also illuminate when the high voltage power supply (HVPS) operates outside of specified voltage range.

Two plug and jack connections are found on the sides of the instrument. One is for an external charger for the NiMH batteries, and the other is the RS-232 cable connection to a personal computer.

Page 10: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 6

4.0. CONFIGURATION 4.1. EXTERNAL FEATURES The front panel features include: 1. the digital panel meter, 0.1- 1999.9 MBq/m3 (3 – 54000 μCi/m3) 2. function control knob, OFF, MEASURE, SAMPLE 3. alarm level control knob, 0.2- 100.0 MBq/m3

4. low battery LED, with dual function, indicates red when HiMH batteries need to be re-charged and indicates red to notify the user if the high voltage power supply (HVPS) is operating out of specification

5. signal level alarm LED 6. low flow alarm LED 7. acoustic signaler 8. mute push-button 9. calibration potentiometer (under phillips head screw) 10. offset potentiometer (small knob) 11. SAMPLE IN/OUT hose barbs 12. toggle switch for GAMMA COMPENSATION ON/OFF control 13. two compartments for one D-cell NiMH battery each, covers for these are secured

from the inside of the instrument to deter replacement with non-rechargeable type. Side features include: 14. jack for external charger, 3.3Vdc, 1.2-1.5A, DO NOT SUBSTITUTE, use only the

recommended power supply 15. 8-pin mini-din receptacle for RS232

16. GAMMA CHECK labels for position of check source, refer to section 6.0. 17. snap holder for dust filter

4.2. HOSE CONNECTIONS

A sniffer hose can be attached to the hose barb on filter housing for the SAMPLE inlet NOTE 1: NEVER OPERATE THE INSTRUMENT WITHOUT A DUST FILTER IN THE SAMPLE STREAM NOTE 2: THE INSTRUMENT MUST BE IN THERMAL EQUILIBRIUM WITH ITS SURROUNDINGS GENERAL OPERATION NOTES:

The following information is provided to the user to ensure stable and accurate performance.

The monitor can be hand carried or located on any flat surface, such as a table top, or, it can be mounted to a wall bracket or shelf, or on a small moveable cart. In all cases, the instrument must be protected against vibration, shock, moisture and dirt

NOT SUITABLE FOR USE IN WET LOCATIONS NOT SUITABLE FOR USE IN EXPLOSION HAZARD ENVIRONMENTS

Page 11: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 7

5.0. OPERATION

Ensure that a dust filter is connected at the SAMPLE inlet.

The following steps are necessary and sufficient to operate the instrument:

1. Set measurement alarm level to desired value.

2. Set the GAMMA COMPENSATION switch to the desired position. Normally the switch should be ON. The OFF position is provided for functional checks or calibration using a gamma source.

3. Rotate mode switch to “MEASURE”. The “low flow” LED will flash, since the pump is inactive. The “mute” switch will silence the intermittent tone if desired.

4. Allow two to three minutes for the instrument to stabilize, then (if necessary) rotate the OFFSET control to achieve a zero reading on the meter. Allow some time for the reading to stabilize, rotate the OFFSET control in small increments as required.

The instrument is now ready for use. In this mode the ionization chambers are active, but the pump is not. The instrument is in a so-called “standby mode” ready to sample the instant the mode switch is advanced to the next position.

5. Rotate the mode switch to the SAMPLE position. Now, the pump is operating and the low flow indication will be eliminated.

If it is desired to operate the instrument continuously on an external power supply, only a source of 3.3 V d.c. with 1.5 A current capacity should be used. Attaching the external power plug will automatically charge the batteries.

DO NOT PLUG IN THE EXTERNAL POWER SUPPLY WITHOUT D-CELL BATTERIES INSTALLED IN THE MONITOR

NOTE: If the audible alarm is an intermittent tone, sample flow through the chambers is below specification. This could be due to an obstructed sampling hose or other reason.

It is IMPERATIVE that the sample stream be free from dust, dirt or moisture. Not only will the instrument show erratic behavior, but also it may cease to function entirely. If moisture is ingested, then continued pumping to evaporate and expel the moisture can be attempted. If this fails, the instrument must be returned to the factory for service.

Condensation can occur if an instrument is brought from a cold environment into warmer surroundings.

Furthermore, temperature changes to the instrument, both lower to higher as well as higher to lower will create transient currents in the electrometer which can appear as large phantom measurement signals.

The instrument must be allowed to thermally equilibrate to its surroundings prior to use. When there is an OFFSET due to thermal disequilibrium, use the following procedure:

Page 12: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 8

OFFSET COMPENSATION:

1. Switch the instrument into the measure mode 2. After approximately three minutes. The instrument should indicate 0000 on the digital

panel meter. An offset of 0.2 – 0.5 MBq/m3 (5 - 15 �Ci/m3) is typical for situations due to temperature changes. This offset should disappear as thermal equilibrium is attained.

3. Adjust the “offset” compensation potentiometer as required. The location is shown in Figure 2.

NOTE: The rotation direction for the adjustment is clockwise for change in a positive direction. Use very small amounts of rotation, the display will not

react immediately because there is a 10 second time constant.

5.1 RS232 OUTPUT

The RS232 interface serves the purpose of transmitting data to any PC via a standard RS232 link, and also allows user recalibration of the unit.

Use the supplied Data Viewer / Calibrator software to view the data ( you could also use Hyper Terminal, or other suitable applications - the COM Port setting should be 9600 baud, 1 start bit, 8 data bits, parity = none, 1 stop bit, and flow control = none ). The COM port settings are already set correctly in 400 Data Viewer / Calibrator.

Data is sent every 1 second. All data is in ASCII so the position of each parameter not be constant. Here is the format of the data communication:

":3879;181;010000;000002;uCi/m^3;0;0;518;0;1.05;1;......\r\n" [0] [1] [2] [3] [4] [5;6; 7; 8; 9;10] [11]

[0] :Serial Number [1] Sequence Number [2] Flags: High Tritium Alarm ( 1 = Alarm )

Flow Alarm ( 1 = Alarm ) Calibration in progress ( 1 = in progress ) Power Supply Fault ( 1 = Fault ) Communication Error ( 1 = Error ) Calibration Error ( 1 = Error )

[3] Tritium Value [4] Units [5] Volatile Cal pot value ( 0 - 255 ) [6] Non-Volatile Cal pot value ( 0 - 255 ) [7] Battery Value ( 0 - 1023 ) [8] Calibration State: 0 not calibrating

1 low rail 2 high rail 3 final

[9] Microcode version # [10] High Voltage Power Supply Flag( 1 = OK, 0 = Fault) [11...] Message 1

Page 13: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 9

6.0 CALIBRATION METHODS

Tritium monitors employing ionization chambers, such as the 400AC, may be calibrated with either of two methods. The first method consists of injecting a known activity of tritium gas; the second method uses external gamma radiation of known field strength. The tritium gas method is time consuming, and is quite difficult to perform with precision without the use of an apparatus called a gas calibrator. This method serves as baseline calibration from which the gamma response (the second method) can be cross-correlated. The second method uses an external gamma field. In this instance, the compensation ionization chambers are disconnected (gamma compensation is OFF). Overhoff Technology has provided calibration software so that the user can perform a gamma only calibration.

PERIODIC GAMMA CHECK If a tritium monitor has previously been calibrated by any other method, gas or gamma, a low intensity gamma radiation source check can be used as a quick verification of monitor performance. Labels on the sides of the instrument case is the defined location for “GAMMA CHECK”. When using the identical gamma check source, at the defined spot, it should always produce the same instrument response, provided, of course, temperature and pressure variations are taken into account. This source check may be performed at a frequency of your choice, it could be daily, weekly or monthly. We recommend a small laboratory gamma check source of the type which is commonly intended for G-M counters or other survey instruments. For example; a 10 micro Curie, Cesium-137 check source should be sufficient for a monitor reading of 4.0 – 8.0 MBq/m3 (100-200 μCi/m3). IMPORTANT: Do not adjust the calibration when performing a gamma check.

Page 14: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 10

6.1 CALIBRATION SOFTWARE INSTALLATION AND CONFIGURATION 1. Load the calibration software onto a PC running Windows XP, Vista, or Win 7: Insert the Overhoff CD in your CD drive, and view the contents. Go to the folder “400AC”, then to the folder “Data Viewer and Calibrator”. Run Setup.exe and hit “Next” 3 times.

2. Connect the 400AC to a serial port on the PC, using the supplied RS232 cable. If your computer does not have an RS232 port ( 9-pin DIN ) connector, you will need to use a USB-to-RS232 adapter. These adapters usually come with a CD that contains drivers - these drivers must be loaded also.

3. Set the Mode Switch on the 400AC to the “Measure” position.

4. If you are using a USB to RS232 adapter,

you must determine which COM port number points to the USB adapter, using the following menu:

Start -> Control Panel -> System -> “Hardware” tab -> “Device Manager” button -> Ports (COM and LPT) Expand this box: The COM port number in this example is 8.

Page 15: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 16: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 12

6.3 GAMMA CALIBRATION PROCEDURE

1. Start the software: Start -> All Programs -> Overhoff Technology -> 400 Viewer and Calibrator

2. ( This step should not be necessary, unless you are using a USB to RS232 adapter ) Set

the COM port number in software using this menu:

Settings -> Change COM Port

Enter the port number ( see section 6.1.4 ) and click “OK”.

(The software will remember to use this COM port the next time it is run.)

Page 17: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 13

3. Calibrate the 400AC: a. Start Calibration:

Calibration -> Start Calibration

b. Set the Calibration Source to “Gamma”. Enter the Reference Value that you were asked to calculate in section 6.2. Note: for gamma calibration, the

reference value is multiplied by a conversion factor of 76 uCi/m3 per mRem/h. To use a different conversion factor, click this button:

You are permitted to enter a new

conversion factor between 70 and 90 uCi/m3 per mRem/h ( or between 7 and 9

uCi/m3 per uSv/h.). c. Click “OK”. Wait approximately 6 minutes while the software displays a 1, 2, and 3, then hit “Accept” or “Reject” to accept or reject the results of the calibration.

Page 18: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 14

6.4 CALIBRATION SOFTWARE FRONT PANEL Concentration (large field): Displays the one-second average value of tritium concentration. Average Concentration: Displays the 60-second average value of tritium concentration. Reference Value: Displays the user-entered radiation value of the test source. Alarms: Alarms: High Tritium Alarm: Will be red when the measured tritium value exceeds the 400AC alarm set point. Low Flow Alarm: Will be red when the pump is not working correctly. Power Supply Fault Alarm: Will be red upon a failure of any of the three internal power supplies. High Voltage Battery Alarm: Will be red if the high voltage battery needs to be replaced. The high

voltage battery check is done once when the 400AC is powered ON. The results of this check are transmitted in each packet thereafter, but the high voltage battery is not checked again until the 400AC is powered OFF and then ON again.

Battery Charge Remaining: approximate time. When new, the batteries will fully charge in 10 hours time, and can power the 400AC for over 16 hours (with the pump on). Other Fields: Sequence Number: the 400AC tags each packet it sends with number, starting at 0 on powerup. Packets Received: the number of packets received by the calibration program. 400 Serial Number: the 400AC sends its serial number in the packet. Volatile DCP: The “temporary” test value of the DCP, which is used during calibration. Non-Volatile DCP: the “permanent” value of the DCP, which is set when calibration completes. Microcode Version: version of code contained in the 400AC’s microprocessor.

History: a buffer of the past 50 tritium values. The newest value is at the top. You can change the size of this buffer in the text box.

Calibration Panel: shows the stage of calibration, and the number of seconds remaining in that stage.

Page 19: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 15

6.5 CALIBRATION SOFTWARE MENUS

Starts a calibration. Restores the DCP to its midpoint ( 127 ), which is the factory default setting. Prints a calibration report that has already been created and saved.

Allows entry of source value, conversion factor, and preferred units. See the next section for an explanation Changes the COM port number. Digitally Controlled Pot value ( 0 – 255). Lists any communication or program errors. Not of interest to users. Switches it off or on.

As of Version 1.5, an additional “Version” menu choice next to “Settings” displays the software version number.

Page 20: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 16

6.6 CALIBRATION REPORTS NOTE: Microsoft Word 2003 or later is a requirement for generating calibration reports. A sample calibration report appears at the end of this manual on page 26. Calibration Reports are initially disabled. You can enable calibration reports to be generated, automatically, by displaying the Calibration Report Information Form in the menus: Settings - > Calibration Reports. There are several fields into which you can enter information to be included in the report.

A check here enables report generation. Source Isotope Source ID number A check here means yes. Your name. Automatically print a copy; “ask” will give you a choice.

Finished calibration reports are stored in subdirectories under My Documents -> 400 Calibration Reports. These subdirectories are named by year: 2008, 2009, etc. The Calibration Reports are Microsoft Word files. By default, reports are given filenames which are the serial number of each 400AC. Example: 3547.doc. A form will be displayed after each successful calibration, allowing you to specify a different report filename ( you can hit <return> to use the default name ). Calibration files are not overwritten – duplicate filenames will be prevented by using version numbers, such as 3547_1.doc, 3547_2.doc, etc.

Page 21: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 17

6.7 HOW THE GAMMA CALIBRATION SOFTWARE WORKS There are 3 stages of calibration. Each stage takes 2 minutes, for a total of 6 minutes. When calibration is complete, a “Digitally Controlled Potentiometer” (DCP) is set within the 400AC.

1) In Stage 1, the DCP is set to its maximum value. The instrument reading is allowed to settle. A measurement is then taken.

2) In Stage 2, the DCP is set to its minimum value. The instrument reading is allowed to settle. A measurement is then taken.

3) Using the results of Stage 1 and Stage 2, a value is calculated so that the instrument will read the expected value of the reference source. The DCP is set to this value for Stage 3, and the instrument is allowed to settle one last time

After stage 3 finishes, you are asked to accept or reject the results of the calibration. Compare the Average Concentration field with the Reference Value field - the final readings should be very close to each other. If you accept the results, the DCP will be set to the new value. If you reject the results, the DCP will be reset to the value it contained before the calibration attempt. After you accept or reject, the 400AC’s compensation chambers are switched back online, and the readout value will then decrease to nearly zero. Note: Do not change the DCP value after the calibration is finished! Note: When calibration fails, it is usually because the DCP could not be set because the reference value was out of range. An error form will display the range of reference values over which the 400AC can be calibrated.

Page 22: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 18

6.8 TRITIUM GAS CALIBRATION

The user is not only limited to calibration with a gamma source. The tritium gas calibration method requires an apparatus called a gas calibrator. The manufacturer of the gas calibrator generally provides instructions for the use, and these should be followed.

Some general hints can be given;

It is important that the calibration sample be well circulated through the entire calibration system loop.

Adequate time should be allowed for the system pressure and temperature to come to equilibrium, and that no excess pressure is built up.

The inclusion of a previously calibrated "master" or "reference" tritium monitor in the sampling loop is highly recommended.

The calibration can actually be repeated for several levels of tritium activity. This is not done to verify the linearity of the tritium monitor (which is highly linear) but to ensure that the calibration process itself is free from subtle errors.

6.9 TRITIUM GAS CALIBRATION PROCEDURE

The procedure is the same as for Gamma Calibration, with one difference: the Calibration Source should be changed to “Tritium Gas” in the Reference Source Information form in step b.

6.10 A FINAL NOTE ON CALIBRATION Normally calibration is adjusted with a digital control potentiometer and the calibration software that has been provided with the instrument. There is a hardware calibration potentiometer, however it should not be adjusted. It is located behind the Phillips screw adjacent to the letters CAL that are silkscreened on the front panel. Large changes in calibration are evidence of malfunction. The instrument may be returned to the factory for service. Contact us by Telephone (513) 248-2400, by Fax (513) 248-2402 or Email [email protected]

Page 23: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 19

7.0. MAINTENANCE

Overhoff 400 series portable instruments have been designed for many years of trouble free service. OPERATOR MAINTENANCE The following operational checks may be performed at daily, weekly or monthly intervals to suit. Inspect dust filter for excessive dust build up. Check the flow rate. Does the pump have sufficient flow such that the Low Flow Alarm is not indicated when 10ft of the sniffer hose is connected to the inlet of the dust filter? GAMMA CHECK, If a tritium monitor has previously been calibrated by any other method, gas or gamma, a low intensity gamma radiation source check can be used as a quick verification of monitor performance. On the side of the instrument case towards the front which is the defined location for “GAMMA CHECK”. When using the identical gamma check source, at the defined spot, it should always produce the same instrument response, provided, of course, temperature and pressure variations are taken into account. This source check may be performed at a frequency of your choice, it could be daily, weekly or monthly. We recommend a low intensity gamma check source of the type which is commonly intended for G-M counters or other survey instruments. For example; a 10 micro Curie, Cesium-137 check source should be sufficient for a monitor reading of 100-200 μCi/m3. IMPORTANT: Do not adjust the calibration when performing a gamma check. Manipulate the alarm set point potentiometer to verify correct functioning of the alarm. If the instrument is suspected of DRIFT, the zero reading may be verified. This should be done by an instrument engineer or technician. SUPERVISORY MAINTENANCE The following tasks are the responsibility of the supervisory engineering staff. 1. Calibration verification is to be performed at yearly intervals, or as otherwise specified. 2. Response checks (in case of need for cursory verification of the operational status of the

ionization chambers and of the whole system), of the system may be tested by using a low strength gamma radiation check source. This must be done under the strict supervision of a health physicist. The gamma source is brought into proximity of each ionization chamber and the response is observed.

FACTORY MAINTENANCE A determination that the system appears to have suffered a functional failure should require that the factory be notified (telephone (513) 248-2400, facsimile (513) 248-2402). Engineering assistance via telephone or facsimile, will be supplied by the manufacturer OVERHOFF TECHNOLOGY CORPORATION. Should it appear to be necessary to return the instrument to our factory, authorization for the return must be obtained from Overhoff Technology Corporation prior to shipping. In-freight charges will be borne by the customer.

Page 24: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 20

Very little maintenance is required for the 400AC tritium monitors, but some periodic attention may be necessary, especially if the instrument is to be used in adverse environments.

Pump life is in excess of 1000 hours of actual use; ensuring that the instrument is operated only with dust filters in line preserves its life.

When not in use, the monitor should be stored in a cool dry environment. 7.1. D-CELL BATTERY REPLACEMENT

The Model 400AC uses (2) “D” size NiMH batteries. The batteries will need to be recharged within an hour after the low battery light illuminates. A cover has been installed over each battery holder in the attempts to minimize the chance of accidental replacement using non-rechargeable batteries. Do not use alkaline or carbon-zinc cells.

NOTE: USE ONLY NiMH BATTERIES IN THIS INSTRUMENT.

Step 1. Remove Case Remove the three phillips head screws on the bottom of the Model 400AC. The case can now be removed allowing access to the battery cover mounting screws. It may be necessary to disconnect the sample tube attached to the filter outlet.

This instrument contains components that are easily destroyed if the case is opened and handled without proper precaution. If damage occurs during battery replacement the repair will not be covered under warranty. Avoid touching the ionization chambers and the PCB to reduce the risk of damage.

Step 2. Remove Battery Covers

The battery covers are fastened in place by two phillips head screws. Remove these two screws and the cover to gain access to the battery tube cap. Do not remove the two hex nuts.

Step 3. Replace Batteries Access to the batteries is made by twisting off the caps located on the front panel of the instrument.

Step 4. Installation Installation of the battery caps, battery covers and case is the reverse of Steps 1 and 2.

Page 25: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …

Page 21

8.0. SERVICE AND SUPPORT

This instrument contains highly sensitive semiconductors which are destroyed by even the slightest electrostatic discharge if the case is opened and the instrument is handled without proper precaution.

Special training can be given to qualified technical personnel who are entrusted with

instrument service and repair responsibility. Warranty is void if maintenance or repair (other than that which is listed in this manual) is performed by an unauthorized repair facility.

OVERHOFF TECHNOLOGY CORPORATION Telephone (513) 248-2400 Facsimile (513) 248-2402

Email: [email protected] 9.0. WARRANTY

All instruments built by Overhoff Technology Corporation are warranted to perform as claimed.

Defective components or workmanship of the instrument will be corrected free of charge for parts or labor within a period of one year from delivery. Nonperformance of the instrument as a result of negligence on behalf of the customer is not covered by this warranty.

Should it appear to be necessary to return the instrument to our factory, authorization for the return must be obtained from Overhoff Technology Corporation prior to shipping. In-freight charges will be borne by the customer. 10.0. REPLACEABLE PARTS

The following parts and components are disposable items and may be obtained from Overhoff Technology Corporation or from any original supplier:

Battery, primary power "D” size, NiMH, 10,000mAh, (qty 2 req’d) Dust Filter Pall No. 12082 Ionization Chamber Can P/N 1020686 Pump P/N 50084 Sample Inlet Brass, P/N 22BH-4-2 Hose Barb Sample Outlet Brass, P/N 230-4-2 Hose Barb Panel Meter P/N DMO-742W AC power converter KPTS05-03315U RS232 Cable J2-400AC-1M (2.3 meter length cable is standard, a

maximum length of 15 meters is available)

Page 26: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 27: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 28: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 29: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 30: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …
Page 31: OPERATION/MAINTENANCE MANUAL PORTABLE TRITIUM …