GeniusTM 2
Thermometer Checker/Calibrator
Operating Manual
F/C˚
F/C˚
Checker / Calibrator
For use with Model 303000 Tympanic Thermometers
Table of Contents
Page
Section I - Functional Description................................................................................................................................................... 1
Section II - Precautions and Warnings........................................................................................................................................... 2
Section III - Icon Identification........................................................................................................................................................ 3
Section IV - Required Equipment.................................................................................................................................................... 4
Section V - System Setup.................................................................................................................................................................. 4
Section VI - Operational Procedure................................................................................................................................................. 5
ApplyPower.................................................................................................................................................................................. 5
Start-upScreen............................................................................................................................................................................. 5
ReminderScreen........................................................................................................................................................................... 5
MainMenu.................................................................................................................................................................................... 5
Preferences.................................................................................................................................................................................... 5
LanguageSelection....................................................................................................................................................................... 5
SetUser/SiteInformation.............................................................................................................................................................. 5
SetDateandTime......................................................................................................................................................................... 6
CheckCalibration........................................................................................................................................................................... 6
Section VII - Error Messages............................................................................................................................................................. 8
Section VIII - Checker/Calibrator Specifications............................................................................................................................ 9
Section IX - Cleaning Instructions................................................................................................................................................... 9
Section X - Troubleshooting........................................................................................................................................................... 10
Section XI - Customer Service........................................................................................................................................................ 11
Section XII - Warranty..................................................................................................................................................................... 11
Section XIII - Electromagnetic Conformity Declaration............................................................................................................. 12
ThisproductcontainssoftwaresolelyownedbyCovidien.Covidiengrantstheuseranon-exclusive,limitedlicensetousethesoftwareaccordingtotheoperatinginstructions.AcopyofthelicensecanbeobtainedfromCovidien.
en
GeniusTM2Checker/Calibrator Table of Contents1
Section I - Functional Description
ThismanualdescribestheoperationoftheGenius™2Checker/Calibrator.ThisdevicehasbeendevelopedtochecktheaccuracyoftheGenius™2TympanicThermometerandautomaticallyrecalibratethethermometer,ifnecessary.Inaddition,theGenius2Checker/CalibratorhasthecapabilitytogenerateatestreportforeachthermometertestedandsaveittoaUSBflashdrive.TheGenius2Checker/CalibratorwillworkwithallGenius™2Thermometersoftwarerevisions.
TheGenius2Checker/Calibratorcontainstwoindependentlycontrolledinfraredcalibrationtargetsthataresimilartofactorycalibrationtargets.These“blackbody”targetsaredesignedtohaveefficientradiativeheattransferandproduceinfraredradiationthatapproximatesthehumanearattemperaturesof32.2°C(90.0°F)and40.6°C(105.0°F).
F/C˚
F/C˚
1 2 3 4
8 910
5
6
7
11
F 3.15A 250V
Checker / Calibrator
Figure 1 - Front Panel Layout
Legend
1 PowerConnector 7 HighTemperatureTarget2 PowerSwitch 8 HighTemperatureTargetLED3 FuseHolder 9 LCDDisplay4 LowTemperatureTargetLED 10 EncoderKnob5 LowTemperatureTarget 11 USBport6 Genius2ThermometerConnector
GeniusTM2Checker/Calibrator Table of Contents2
Section II - Precautions and Warnings
• TheGenius2Checker/Calibratorisdesignedtobeoperatedasspecifiedinthismanual.ProtectionprovidedbythesafetydevicesintheGenius2Checker/Calibratormaybeimpairedifthedeviceisnotusedinaccordancewiththedirectionscontainedinthismanual.
• TheGenius2Checker/Calibratorenclosureisnotwaterproof.Avoidspillingliquidofanykindontothefrontpanelofthedevice.Likewise,theGenius2Checker/CalibratorispoweredfromanACpowersource.Avoidthepotentialforcontactbetweenliquidandanyexternalsurfaceofthedevicewhileconnectedtothemainssource.
• Thesurfacequalityoftheinteriorofthereferenceblackbodytargetsiscriticaltotheiroperationasefficientinfraredradiators.Careshouldbeexercisedwhenworkingneartheentrancetotheblackbodytargetstoensurethatnocontactismadewiththesurfaceinsidetherubberflaps.
• Intheeventofablownfuse,replacethefusewithafusethatexactlymatchestheratedvoltage,current,andformfactorasspecifiedintheGenius2Checker/CalibratorSpecificationssectionofthismanual.MakesurethedeviceisdisconnectedfromtheACpowersourcebeforechangingablownfuse.
• Thisdeviceshouldnotbeusedinthepresenceofflammableanesthetics.Thereisriskofanexplosioninthepresenceofthese anesthetics.
• Therearenouser-serviceablecomponentsinthisdevice.Donotremovethefrontpanel.Referservicingtoqualifiedservice personnel.
• Useonlythesuppliedpowermodulewiththisdevice.• WhenpromptedtoinserttheGenius2Thermometerintooneoftheblackbodytargets,thereisa60secondwindowoftimeto
accomplishtheinsertion.Ifthetimedelayexceedsthisperiod,theprocedurewillbeterminated.• Priortochecking/calibratingaGenius2Thermometer,ensurethatthethermometerlensisclean.Effectivecleaningwill
ensurethatthecalibrationtolerancesarecorrectlyverified.RefertothecleaninginstructionsintheGenius2Thermometeroperating manual.
• WheninstallingaGenius2probecoverontothetipoftheGenius2Thermometer,useonlycoversthatareinacassetteandnotreleased.Trytoavoidallhandcontactwiththeprobecoversinordertopreventinaccuratereadings.Donotreuseprobecovers.UseonlyGenius2probecovers.
• LettheGenius2Checker/Calibratorwarmupafterpower-upforatleast15minutesbeforeuse.MakesurethattheGenius2Thermometers,probecoversandtheGenius2Checker/Calibratorhavehadenoughtimetoequilibratetoroomtemperaturebeforeproceeding.Allowalongerwarmupperiodifthedeviceorthermometerswereattheextremelimitsofallowedambienttemperature.TheportabilityoftheGenius2Checker/Calibratorwillallowforrelocationofthedevicetoareasthatmeettheambienttemperaturerequirements.
• TomaintainElectromagneticCompatibility(EMC)conformance,useaUSBflashdrivethatisCEmarkedandmeetsEN55022 standard.
• DonotputorresttheGenius2ThermometeronthesurfaceoftheGenius2Checker/Calibratoratanytime.Thethermalsensorsinsidethethermometerprobetipwilldetecttheheattransferredfromtheblackbodytargetsthroughthefrontpanelsurface.Thiswillcauseinstabilityofthethermometersensors,whichwillproduceerroneoustemperaturereadings.
• TheGenius2Checker/Calibratormayonlybeusedindoors,inanareafreefromdraftsandwideambienttemperatureswings.• TheUSBportonthefrontpanelisintendedtobeusedwithonlyaUSBFlashmemorydrive.NootherUSBdevicesshouldbe
insertedintotheUSBport.TheUSBflashmemorydriveshouldbeinstalledpriortodevicepower-up.• IntheeventthattheGenius2Checker/Calibratorisdropped,returntheunittothefactoryforre-calibration.
GeniusTM2Checker/Calibrator Table of Contents3
Section III - Icon Identification
SN SerialNumber
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
CorrugatedRecycles
STERILE
Keep AwayFrom Sunlight Keep Dry
Single Use
LOT
0123
FRAGILETHIS SIDE UP
DEHP – FREE
0°
40°
Avoid TemperatureExtremes
Do not use if package is opened or damaged.
STERILE R
ASTERILE
EOSTERILE
STERILE
STERILE R
ASTERILE
EOSTERILE
Humidity LimitationAtmospheric pressure limitation
Use By
C US
LATEX – FREE
Caution, consultaccompanying
documents
ConsultInstructions
For Use
MRMR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT ANDRADIOACTIVE SOURCES
NONSTERILE
2STERILIZE
DO NOT RESTERILIZE
POWDER – FREE
POWDER
Manufacturer
ConsultAccompanyingDocuments
CE MARK SYMBOLS
NON-CE MARK SYMBOLS
ELECTRONIC SYMBOLS AND CERTIFIED TESTING MARKS
CorrugatedRecycles
STERILE
Keep AwayFrom Sunlight Keep Dry
Single Use
LOT
0123
FRAGILETHIS SIDE UP
DEHP – FREE
0°
40°
Avoid TemperatureExtremes
Do not use if package is opened or damaged.
STERILE R
ASTERILE
EOSTERILE
STERILE
STERILE R
ASTERILE
EOSTERILE
Humidity LimitationAtmospheric pressure limitation
Use By
C US
LATEX – FREE
Caution, consultaccompanying
documents
ConsultInstructions
For Use
MRMR UNSAFE NON-PYROGENIC
PROTECT FROM HEAT ANDRADIOACTIVE SOURCES
NONSTERILE
2STERILIZE
DO NOT RESTERILIZE
POWDER – FREE
POWDER
DateofManufacture
StandbyPower DoubleorReinforcedInsulationProtection
ReplacewithSameRatedFuse CEMark
USBPort DisposeofasElectricalandElectronicWaste
F/C˚
HighTemperatureTarget
LISTED(LABORATORY EQUIPMENT)
46EA
ULListingInformation
F/C˚
LowTemperatureTarget
Caution:IndoorUseOnly
Genius2ThermometerConnection
GeniusTM2Checker/Calibrator Table of Contents4
Section IV - Required Equipment
1. Genius2Checker/Calibrator 2. PowermodulewithACadapter 3. Probecable 4. Operator’sManual 5. Genius2Thermometerstobetested 6. Genius2disposableprobecovers 7. SmallPhillipsscrewdriver
Section V - System Setup
The following actions should precede use of the Genius 2 Checker/Calibrator: 1.ConnectthecorrectACplugadapterforyourcountrytothepowermodule. 2.Connectthepowermodulecabletothepowerjackonthefrontpanel. 3.ConnectthepowermoduletoanACpoweroutlet. 4.Turntheunitonviatheswitchonthefrontpanel. 5.LettheGenius2Checker/Calibratorunitwarmupforatleast15minutes. 6.SetasideacassettewithnewGenius2probecovers. 7.LettheGenius2Thermometersequilibratetotheroomtemperature. 8.PreparetheGenius2Thermometersbythoroughlycleaningthelensespriortouse.RefertotheGenius2Thermometeroperating
manualforcleaninginstructions. 9.RemovethebatterycompartmentaccesspanelbyusingasmallPhillipsscrewdriver.RemoveallthreeAAAbatteries. 10.ConnecttheprobecableroundconnectortothefrontpaneloftheGenius2Checker/Calibrator. 11. Connecttheotherendoftheprobeconnectorintothebatterycompartment.Theconnectorwillsnapintoplacewherethe batterieswere.RefertoFigure2.
Figure 2 - Insertion of Probe Connector into Genius 2 Battery Compartment
GeniusTM2Checker/Calibrator Table of Contents5
Section VI - Operational Procedure
ThefollowingprocedurewilldescribethestepsrequiredtooperatetheGenius2Checker/Calibrator.Inordertofacilitatefamiliaritywiththeproduct,itwouldbeadvisabletousethissectionofthemanualduringthecalibrationcheckofthefirstfewthermometers.Installation of USB Flash Drive (Optional) InstallaUSBflashdriveintotheUSBportlocatedonthefrontpanelifareportfilewrittentotheUSBflashdriveisdesired.IftheUSBflashdriveisinstalledafterperformingacalibrationcheck,anerrormessagemayoccurwhenattemptingtowritethereportfileiftheoperatordoesnotwaitfortheUSBflashdrivetocompleteinitialization.IftheGenius2Checker/Calibratorispoweredoff,allinternalmemory(notyetsavedtotheUSBflashdrive)willbelost.
Apply Power MakesuretheAC-adapterendofthepowermoduleispluggedintoagroundedoutlet.Connectthepowermoduletothefrontpanelpowerconnector.ApplypowertotheGenius2Checker/Calibratorbyusingthepowerswitchonthefrontpanel.
Start-up Screen ThefirstscreenontheLCDdisplayistheInitializationScreen,whichdisplaysthesoftwarerevisionanddateoftheGenius2Checker/Calibrator.Ifthereisaninternalproblemduringinitialization,anErrorscreenwillbedisplayed(SeeSectionVII).
Reminder Screen ThenextscreenontheLCDdisplayisaremindertolettheGenius2Checker/Calibratorwarm-upforatleast15minutesafterapplyingpower.Thiswilllettheinfraredblackbodytargetswarmuptothepropertemperatures.Toexitthisscreen,presstheencoderknob.
Main Menu TheMainMenuscreenisdisplayednext.Thisscreenhasfourmenuoptions:
CheckCalibration Preferences SetUser/SiteInformation SetTime/DateToenteranymenuoption,usetheencoderknobtoscrolloverandhighlightthedesiredmenuoption,thenpresstheencoderknob.
Preferences ThePreferencesmenuisusedtosetupthelanguageoption.PresstheencoderknobtoentertheLanguageSelectionscreen.
Language Selection TheLanguageSelectionscreenliststheavailablelanguagesfortheLCDscreentext.Toselectalanguage,scrollovertheselectionanditwillbehighlighted.PresstheencoderknobtocompletetheselectionandreturntotheMainMenuscreen.Theselectedlanguagewillbestoredinnon-volatilememoryandwillbeusednexttimetheunitispoweredon.
Set User/Site Information TheSetUser/SiteInformationscreenisusedtoentertheoperator’snameandtheorganizationand/orsiteinformation.Thisinformationwillbewrittentothetestreportfile.
Toentertheinformation,usetheencoderknobtoscrollthroughthealphabetforagivenletterpositionintheoperatornameorsiteinformation.Whenthecorrectletterisshown,presstheencoderknobtoselecttheletterandadvancetothenextletterposition.Toenteraspace,justpresstheencoderknobwhenaspaceorblankletterisshowninthatletterposition.Theusernamecanbeupto20letterslong.Usetheencoderknobtoscrolltotheorganizationfieldandenteranameforthesiteinasimilarmanner.Toexitthemenu,scrolltotheDONEtextandpresstheencoderknob.
GeniusTM2Checker/Calibrator Table of Contents6
Set Date and Time TheSetDateandTimescreenmenuoptionisusedtosetupthedateandtimethatisusedonthetestreport.Thetimeanddateareinternallystoredandautomaticallyupdatedwhentheunitispoweredup.
Toentertime,usetheencoderknobtoselectthehoursandminutes.Note:thehoursissetupusinga24hourclockformat,where1PMisenteredas13:00etc.Whenthevalueiscorrect,pushtheencoderknobtomovetothenextfieldofthetimeandthentothedatefield.ThedatefieldissetupusingaMMDDYYYYformat.Afterthelastfieldofthedateisentered,thenextencoderpushwillexitthismenuandreturntotheMainMenu.Note:thetimeanddateareshownatthebottomoftheMainMenu.
Check Calibration ThismenuoptionperformsthecalibrationcheckofaGenius2Thermometer,andifrequired,automaticallyrecalibratesthethermometerusingaseriesofstep-by-stepscreens.Beforeenteringthismenuoption,makesurethattherequiredequipmentlistedinSectionIVisavailable.Thisincludesacassettewithunused(new)probecovers.
Step 1: Warmupofinfraredblackbodytargets.Thisscreenisdisplayedwhilethetargetsarecheckedforpropertemperature.BoththeLowTargetandtheHighTargetarechecked,andwhentheyareatthecorrecttemperature,thescreenwillchangeto“OK”foreachtarget.Thereisaprompttopresstheencoderknobtocontinuetothenextscreen.Ifeithertargetcannotreachthecorrecttemperature,anErrormessagescreenwillbedisplayed.Ifthishappens,checktobesuretheGenius2Checker/Calibratoriswithinthecorrectambient(room)temperaturerange,andhasbeengivenatleast15minutesofwarmuptimesincepowerwasswitchedon.
Step 2: InspecttheGenius2Thermometerthatwillbetested.Ensurethatthethermometerlensisnotscratchedandisfreefromearwaxand/orothercontaminants.Ifthethermometerlensis“dirty”,cleanthelensaccordingtotheinstructionsintheGenius2Thermometeroperatingmanual.Oncethethermometerlensisclean,connecttheGenius2ThermometertotheChecker/CalibratorbypluggingtheroundendofthethermometerinterfacecableintotheroundconnectorsocketontheGenius2Checker/Calibratorfrontpanel(refertoFigure3).OpenthebatterydoorofthethermometerbyusingasmallPhillipsscrewdriver.Removethebatteriesandsetaside.Connecttheotherendofthethermometerprobecableintothebatterycompartment,makingsuretheconnector“snaps”intoplacesecurely.RefertoFigure2inSectionV.
Note: Anewprobecovermustbeusedpriortoeverytargetinsertion.Fingeroilsonpreviouslyusedprobecoversorre-usedprobecoversthathavebeenheatedbyGenius2Checker/Calibratortargetcancauseerroneousreadingswhichmayresultinafailedcalibrationcheckorrecalibration.Therefore,itisextremelyimportantthatanewprobecoverisusedeachtime.
Step 3: Presstheencoderknobtocontinue,theninstallanewGenius2probecoverontotheGenius2Thermometer.Makesuretouseacoverthatisheldinthecassette.Oncethecoverisonthethermometer,donotpointtheprobetipatanyobject,includinghands,fingersorLCDdisplay,asthiswillcauseaninaccuratetemperaturereading.Inspecttheprobecovertomakesureitisfullyseated(nospacebetweenthecoverandthetipbase)andnoholes,tears,orwrinklesarepresentintheplasticfilm.Whenthecoverison,inserttheprobetipintotheLowtemperaturetarget,whichwillhaveablinkingLEDindicatornexttoit.Makesurethattheprobetipisfullyinsertedintothetargetwell.RefertoFigure3forproperinsertion.Inordertoavoidatime-outerror,insertthecoveredprobetipintothetargetwithin60secondsfromthetimetheencoderknobispressed.
GeniusTM2Checker/Calibrator Table of Contents7
F/C˚
F/C˚
Checker / Calibrator
Figure 3 - Proper Insertion of Genius 2 Probe Tip into Target
Step 4: Theinsertionoftheprobetipintothetargetwillautomaticallybedetectedandthescreendisplaywillchangeto“PressGenius2Thermometerscanbutton.”PressthescanbuttonontheGenius2ThermometerwhileitisinsertedintotheLowtemperaturetarget.TheGenius2Thermometerwilloutput3shortbeepsasittakesthetemperatureoftheblackbodytarget.Afterthetemperatureistaken,thenextscreenwilldisplay.
Step 5: WithdrawthethermometerfromtheLowtemperaturetargetandejecttheprobecover.Thescreendisplaywillchangeto“InstallanewprobecoverandinsertprobeintotheHightarget.”Again,useanewGenius2probecoverfromthecassetteandbecarefulnottopointtheprobetipatanyobject,includinghands,fingersorLCDdisplay.Inspecttheprobecoverstomakesureitisfullyseated(nospacebetweenthecoverandthetipbase)andnoholes,tears,orwrinklesarepresentintheplasticfilm.Whenthecoverison,inserttheprobetipintotheHightemperaturetarget,whichwillhaveablinkingLEDindicatornexttoit.Makesurethattheprobetipisfullyinsertedintothetargetwell.
Step 6: TheinsertionoftheprobetipintotheHightargetwillautomaticallybedetectedandthescreendisplaywillchangeto“PressGenius2Thermometerscanbutton.”PressthescanbuttonontheGenius2ThermometerwhileitisinsertedintotheHightemperaturetarget.TheGenius2Thermometerwilloutput3shortbeepsasittakesthetemperatureoftheblackbodytarget.Afterthetemperatureistaken,thenextscreenwilldisplay.Step 7: IftheGenius2thermometeriswithinspecifiedaccuracylimits,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerpassedcalibrationcheck.IftheGenius2thermometerisfoundtobeoutsidethespecifiedaccuracylimits,thedisplaywillindicatetotheuserthattheGenius2Checker/CalibratorisenteringintoacalibrationsequencethatrequirestheusertorepeatSteps3through6threemoretimesinordertogatherrecalibrationdata.FollowingthelastpassthroughSteps3through6,theGenius2Checker/CalibratorwillattempttorecalibratetheGenius2thermometer.Iftherecalibrationissuccessful,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerpassedcalibration.Iftherecalibrationisunsuccessfulorthethermometerisunabletobecalibrated,aResultsscreenwillbedisplayedshowingthethermometerserialnumberalongwithanindicationthatthethermometerfailedcalibration.
GeniusTM2Checker/Calibrator Table of Contents8
Step 8: Afterthepass/failResultsscreenhasbeendisplayed,presstheencoderknobtodisplayasecondResultsscreencontainingthetargettemperatures,theGenius2readings,andthetargetversusthermometervariances.Step 9: Thenextscreenwilldisplay“WritereporttoUSBflashdrive?”anddisplayYesandNobelow.Ifareportisdesired,insertaflashdriveintothefrontpanelUSBport.Note:allowabout30secondsafterinsertionfortheUSBflashdrivetoinitialize.UsetheencoderknobtoscrolltoeitherYesorNoandpresstheencoderknobtomakeaselection.IfNoisselected,thescreenreturnstotheMainMenu.IfYesisselected,areportiswrittentotheUSBflashdrive,usingthethermometerserialnumberforfilenameand.TXTasthefileextension.Thefilecontainsthetestresultsaswellasotherinformationsuchasserialnumber,date,time,username,andorganization.ThisfileisasimpleASCIItextfileandisonlyavailableinEnglish.ThisfilecanbereadbyMicrosoftWindows®operatingsystemandMicrosoftWindows®notepadandsenttoaprinter.Afterthefileiswritten,thedisplaywillshow“Pushknobtocontinue.”PresstheencoderknobtoreturntotheMainMenu.Step 10: Aftercompletingacalibrationcheck/recalibration,removetheprobeconnectorfromthebatterycompartment,usingarockingmotiontofreetheconnector.Thenreinstallthe3AAAbatteries,observingtheproperpolarityorientation.Finally,reinstallthebatterycompartmentcoverandtightenthescrew.Step 11: RemoveUSBflashdrivebeforeclosingthelidofthecase.ThelidshouldbeclosedwhentheGenius2Checker/Calibratorisnotinuse.
Section VII - Error Messages
Theerrormessageswillhaveanumberassociatedwiththem,suchasError1,etc.Thefollowingisanexplanationoftheerrors:
Error 1: Internal Error ThiserrorisanindicationofaproblemwithintheGenius2Checker/Calibrator.Ifthiserrorpersistsaftercyclingpoweroffandbackon,returntheunitforfactoryservicing.
Error 2: Unable to communicate with Genius 2 Thermometer. ThiserrorisanindicationofeitheralackofcommunicationoramiscommunicationbetweentheGenius2ThermometerandtheGenius2Checker/Calibrator.Checkbothendsofthethermometerprobecable,especiallytheendthatconnectsinsidethebatterycompartment.Thepinsinthethermometerprobecablecanlosespringactioniftheyarebentordamaged.Afterinspectingtheprobecable,re-attemptthecalibrationcheck.Iftheproblempersists,tryadifferentGenius2Thermometer,ifavailable.Ifthisdoesnotresolvetheproblem,returntheunitforfactoryservicing,alongwiththethermometercableandtheGenius2Thermometerthatindicatedthiserror.
Error 3: TimeoutThiserrorisanindicationthattheoperatortooktoolongtoperformanoperation.ThiserrormightbeseenwhenconnectingtotheGenius2Thermometer,orwhileinstallingaprobecoveronthethermometer.Thetimeoutissetforoneminute.Tryrepeatingtheoperationthatgavethiserror.
Error 4: Target Out of Range Thiserrorisanindicationthatoneoftheblackbodytargetsisnotwithintherangeoftemperatureallowed.Checkthattheunitisbeingoperatedwithintheallowedambienttemperatureoperatingrange.RefertoSectionVIII.Ifoperatingwithinthespecifiedambientconditionsdoesnoteliminatetheerror,theblackbodytarget(s)maybedamaged.Returntheunitforfactoryservicing.
Error 5: USB Flash Drive Write Error ThiserrorisanindicationofaproblemwhenwritingthereportfiletoaUSBflashdrive.CheckthatthewriteprotectswitchontheUSBdriveisinthe“off”position.Makesurethatthereissufficientfreespaceonthedrive,andthatsecurityorencryptionfeaturesarenotinuse.
Error 6: Ambient Temperature Out of RangeThiserrorisanindicationthattheambienttemperatureisnotwithinthespecifiedrangeforperforminganaccuraterecalibration.Checkthattheunitisbeingoperatedwithintheallowedambienttemperatureoperatingrange.RefertoSectionVIII.
GeniusTM2Checker/Calibrator Table of Contents9
Error 7: Wrong TargetThiserrorisanindicationthattheGenius2Thermometerhasbeeninsertedintowrongtargetorwasinsertedintothetargetprematurely.Pleasetakecaretofollowtheinstructionsonthescreenandnotperformstepspriortotheinstructionsbeingdisplayed.
TheErrorscreenwillbedisplayedfor60secondsbeforereturningtotheMainMenu.ToexittheErrorscreensooner,presstheencoderknob.
Section VIII - Checker/Calibrator Specifications
Temperature Set PointsLow Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.2°C+/-0.3°C(90.0°F+/-0.5°F)High Target . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40.6°C+/-0.3°C(105.0°F+/-0.5°F)
Target Temperature Accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .InitialLimit+/-0.03°C(0.05°F) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Postyear1+/-0.06°C(0.10°F)
Genius 2 Thermometer Accuracy After Recalibration*Readings between 36. 0 °C and 39.0°C(96.8 °F and 102.2 °F). . . . . . . . . . . . . . . . . . . . . . . . +/-0.2°C(+/-0.4°F)Readings less than 36. 0 °C (96.8 °F) or greater than 39.0°C(102.2 °F). . . . . . . . . . . . . . . . . . . . +/-0.3°C(+/-0.5°F)
Warm-up Time Before Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15minutes(fromnon-extremeambient)Ambient Temperature Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21.1-26.7°C(70-80°F)Relative Humidity Operating Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50+/-20%,non-condensingAltitude Range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . Upto2000meters(6563feet)Storage Temperature Range . . . . . . . . . . . . . . . . . . . . . . . . . -25°C-55°C(-13°F-131°F)85%RHnon-condensing
Approximate SizeLength . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31.1cm(12.25”)Depth . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15.9cm(6.25”)Width . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26.0cm(10.25”)Weight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3.3kg(7.2lbs)Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100-240V0.8-0.4A47-63HzFuse Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . SchurterFSF5x20mmFastBlow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .LowBreakingCapacity250V,3.15ARecommended Checker/Calibrator Calibration Interval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .OneYearESD Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .MeetsIEC61000-4-2Non-continuousCriteriaCPollution Degree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2Installation Category. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IndoorUseOnly*PostrecalibrationaccuracyusingtheGenius2Checker/Calibratormaynotnecessarilybeequivalenttofactorycalibration
Section IX - Cleaning Instructions
Caution: Do not immerse the checker/calibrator, cables, or power cord in water or other cleaning solution; clean using a damp (not wet) cloth. Failure to follow cleaning procedures described herein could result in hazards to users.
AswithanyACpoweredelectricaldevice,caremustbetakentopreventliquidfromenteringtheChecker/Calibratortoavoidelectricalshockhazard,firehazard,ordamagetoelectricalcomponents.
GeniusTM2Checker/Calibrator Table of Contents10
General Cleaning InstructionsCleaningoftheGenius2Checker/Calibratormaybeperformedasfollows:
• First, unplug theGenius2Checker/Calibratorfrom power source. Never clean the device while connected to the mains supply.
• Amilddetergentshouldbeusedforgeneralcleaningofoutsidesurfaces.Also,aclothdampedwithisopropylalcoholcanbeusedtocleantheunit.Do not useabrasivesorsolvents.Avoidexcessmoisturearoundthetargetassembliesandtheelectricalconnectors.
• Donotattempttocleanthetargetbelowtherubberflaps.• Donotcleanthetargetnestsorflapswithalcohol.
IfthereisanydoubtaboutthecompatibilityofacleaningagentwithpartofthisequipmentoritsmaterialscontactCovidiencustomerservice(SeeSectionXI).
Section X - Troubleshooting
Some of the common problems are as follows:
Problem Probable Cause Corrections
Genius2Checker/Calibratorwillnotpowerup
Genius2Checker/CalibratornotreceivingACpower CheckACpowerconnections
Blownfuse Replacefuse
TestreportdoesnotwritetoUSBflashdrive
USBflashdrivenotformatted UsePCorlaptoptoformatflashdrive
USBflashdrivewriteprotected UsePCorlaptoptoturnoffwriteprotection
USBflashdrivedoesnotcontainenoughfreememoryspace UsePCorlaptoptodeleteunnecessaryitemsorusedifferentUSBflashdrive
USBflashdriveincompatiblewithGenius2Checker/Calibrator UsedifferentUSBflashdrive
Genius2Checker/CalibratorhasnothadenoughtimetodetectthenewlyinsertedUSBflashdrive
Retrywriteprocess
MultipleGenius2thermometersfailrecalibration
Probecoverstuckinoneoftheblackbodytargets Removeprobecoverfromblackbodytargetusingsmallneedle-nosepliersorhemostats.Donotscratchtheinteriorsurfaceofthetarget.
Dirtyprobecover Useanewprobecoverforeachreading.Donotre-useaprobecover.Probecoversshouldneverbetouchedbyhand.
Dirtythermometerlens CleanthethermometerlenspertheinstructionsintheGenius2operatingmanual.
Genius2thermometernotseatedcorrectly EnsurethattheGenius2thermometeriscompletelyseatedinthetargetandattheproperangle.
Genius2thermometers,probecovers,and/orchecker/calibratornotatroomtemperature.
EnsurethattheGenius2thermometers,probecovers,and/orchecker/calibratorhavehadenoughtimetoequilibratetoroomtemperaturebeforeproceeding.
Error6repeatedlyappearswhiletestingthesameGenius2thermometer
PotentialpowerconsumptionissuewithGenius2thermometer Repeatcalibrationcheck/recalibrationwithadifferentGenius2thermometerandreturnproblemthermometerforfactoryservicing.
Genius2thermometerseemstoreadhighorlowfollowingasuccessfulrecalibration
Probecoverreusedduringcalibrationcheck/recalibration Repeatcalibrationcheck/recalibrationusinganewprobecoverpriortoeverytargetinsertionduringtheprocess.
GeniusTM2Checker/Calibrator Table of Contents11
Section XI - Customer Service
TheGenius2Checker/Calibratorrequiresfactorycalibrationandinspectiononceayear.NotifyandreturnthedevicetoCovidienonorbeforethecalibrationdatefortheChecker/Calibrator.TheChecker/Calibratorwilldisplayamessageonemonthpriortothecalibrationdateasareminder.
Caution:There are no serviceable parts inside, return to factory for service.
Intheeventthatitisnecessarytoreturnaunitforrepair,pleaseobservethefollowing:
1. ContactCovidienasshownbelow,oryourlocalcustomerservicerepresentativeforanAuthorizedReturnNumberandshipping instructions.
2. Shipinsuredparceltoyourlocalservicecontactortheappropriatelocationbelow.
Parts ListingPleasecontactyourlocalcustomerservicecenterorsalesrepresentativeforthepartslistedbelow.
DescriptionGenius2Checker/Calibrator. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303097RefurbishedGenius2Checker/Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .303096
Section XII - Warranty
CovidienwarrantstheGenius2Checker/Calibratortobefreeofdefectsinmaterialsandworkmanshipundernormaluseandserviceforaperiodof1yearfromthedateofdeliverybyCovidientothefirstpurchaser.Liabilityhereunderislimitedtorepairorreplacement,atCovidien’soption,atCovidienservicefacilityforanyproduct,whichshallundernormaluseandservice,appeartoCovidientohavebeendefectiveinmaterialorworkmanship.Thiswarrantyshallnotapplyto,andCovidienshallnotberesponsiblefor,anylossarisinginconnectionwiththepurchaseofanyproductthathasbeenrepairedbyanyoneotherthanCovidienoritsauthorizedrepresentative,orwhichhasbeensubjecttomisuse,neglect,oraccident,orwhichhasbeenusedotherwisethaninaccordancewiththeinstructionsfurnishedbyCovidien.Covidienneitherassumesnorauthorizesanyrepresentativeorotherpersontoassumeforitanyliabilityotherthanasexpresslysetforthherein.
Covidien disclaims all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular purpose or application other than as expressly set forth in the product labeling. In no event will Covidien be liable for any incidental, indirect, or consequential damages in connection with the purchase or use of this product.
UnitedStatesCovidien5920LongbowDriveBoulder,CO803011-800-448-0190
CanadaCovidien7300TransCanadaHighwayPointe-Claire,QCH9R1C7,Canada1-877-664-8926
OutsideUS&CanadaCovidienUnit2TalismanBusinessCenter,LondonRoadBicester,OX266HR,UK1+44-1869-32806
GeniusTM2Checker/Calibrator Table of Contents12
Section XIII - Electromagnetic Conformity Declaration
TheGenius2Checker/CalibratorhasbeenbuiltandtestedaccordingtoIEC/EN61010-1andEN61326-1standards
TheChecker/Calibratorisintendedforuseintheelectromagneticenvironmentspecifiedbelow.Theuseroftheequipmentshouldassurethatitisusedinsuchanenvironment.
Emissions Test Compliance Electromagnetic Environment -Guidance
ConductedandRadiatedEmissions(EN61326-1)
Group1,ClassB TheChecker/Calibratoremissionsareverylowandarenotlikelytocauseinterferenceinnearbyelectronicequipment.
HarmonicCurrent(EN61000-3-2)
ClassA
VoltageFluctuationandFlicker(EN61000-3-3)
Complies
RadiatedDisturbanceImmunity(EN61000-4-3)
Complies
ConductedDisturbanceImmunity(EN61000-4-6)
Complies
PowerFrequencyMagneticFieldImmunity(EN61000-4-8)
Complies
VoltageDipsandInterruptsImmunity(EN61000-4-11)
Complies
ElectricalFastTransient/BurstsImmunity(EN61000-4-4)
Complies
SurgeImmunity(EN61000-4-5) Complies
ElectrostaticDischarge(EN61000-4-2)
MeetsCriteriaC,Non-Continuous
Manual No. 1038307
COVIDIEN, COVIDIEN with logo and Covidien logo are U.S. and internationally registered trademarks of Covidien AG.
Microsoft and Windows are either registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.
Other brands are trademarks of a Covidien company.
© 2011 Covidien.
Made in USA. Covidien llc, 15 Hampshire Street, Mansfield, MA 02048 USA.
Covidien Ireland Limited, IDA Business & Technology Park, Tullamore.
www.covidien.com 1-800-448-0190 [T]
REV 12/2011
Identification of a substance that is not contained or present within the product or packaging.Identification d’une substance non présente dans le produit ou l’emballage.Identifizierung einer Substanz, die weder im Produkt noch in der Verpackung enthalten bzw. vorhanden ist.Identificazione di una sostanza che non è contenuta né presente nel prodotto o nella confezione.Identificación de una sustancia no contenida o que no esté presente dentro del producto o embalaje.Identifiering av ett ämne som inte ingår eller förekommer i produkten eller förpackningen.Identificatie van een stof die niet in het product of de verpakking is vervat of aanwezig is.Identificação de uma substância não contida ou não existente no produto ou embalagem.Tuotteessa tai pakkauksessa olemattoman aineen tunnistus.Identifikation af et stof, der ikke er indeholdt eller til stede i produktet eller emballagen.Αναγνώριση ουσίας που δεν περιέχεται ή δεν υπάρχει στο προϊόν ή στη συσκευασία.Identifikace látky, která není obsažena nebo přítomna v produktu nebo obalu.Olyan anyag azonosítása, amelyet sem a termék, sem a csomagolás nem tartalmaz, vagy amely azokban nincs jelen.Определение вещества, не содержащегося или не присутствующего в продукте и упаковке.Identyfikacja substancji nie zawartej i nieobecnej w produkcie lub jego opakowaniu.Ürün veya ambalajında bulunmayan veya var olmayan bir madde tanımlaması.Identifikasjon av et stoff som ikke er til stede i produktet eller emballasjen.