49
AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses: www.analytical-controls.com www.paclp.com AC Analytical Controls BV P.O Box 10054, 3004 AB Rotterdam Innsbruckweg 35, 3047 AG Rotterdam, the Netherlands Phone : +31-10-462 4811 Fax : +31-10-462 6330 E-mail : [email protected] AC Analytical Controls Inc 8824 Fallbrook Dr., Houston, Texas 77064, USA Phone : +1-2815800339 Fax : +1-2815800719 E-mail : [email protected] AC Analytical Controls Asia Pacific Ltd 30 Robinson Road, Robinson Towers #03-04 048546 Singapore Phone : +65-324-9017 Fax : +65-324-9019 E-mail : [email protected] Operating Manual

Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses: AC Analytical Controls BV

  • Upload
    ledien

  • View
    241

  • Download
    7

Embed Size (px)

Citation preview

Page 1: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

AC OxyTracer Analyzer on 7890 GC

Manual part number 23070.020 version 2011/1.0

Contact addresses:

www.analytical-controls.com www.paclp.com

AC Analytical Controls BV P.O Box 10054, 3004 AB Rotterdam Innsbruckweg 35, 3047 AG Rotterdam, the Netherlands Phone : +31-10-462 4811 Fax : +31-10-462 6330 E-mail : [email protected]

AC Analytical Controls Inc 8824 Fallbrook Dr., Houston, Texas 77064, USA Phone : +1-2815800339 Fax : +1-2815800719 E-mail : [email protected]

AC Analytical Controls Asia Pacific Ltd 30 Robinson Road, Robinson Towers #03-04 048546 Singapore Phone : +65-324-9017 Fax : +65-324-9019 E-mail : [email protected]

Operating Manual

Page 2: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Copyright Analytical Controls 2011

All Rights Reserved. Reproduction, adaptation, or translations without prior written permission is prohibited, except as allowed under the copyright laws.

First edition, December 1995

Printed in the Netherlands

MS-DOS and Microsoft are registered trademarks and Windows is a trademark of Microsoft Corporation.

Warranty

The information contained in this document is subject to change without notice.

Analytical Controls makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose. Analytical Controls shall not be liable for errors contained herein or for incidental or consequential damage in connection with the furnishing, performance, or use of this material.

Safety Symbols

Warnings in the manual or on the instrument must be observed during all phases of operation, service, and repair of this instrument. Failure to comply with these precautions violates safety standards of design and the intended use of the instrument. Analytical Controls assumes no liability for the customer's failure to comply with these requirements.

WARNING

A warning calls attention to a condition or possible situation that could cause injury to the user.

CAUTION

A caution calls attention to a condition or possible situation that could damage or destroy that product or the user's work.

Safety Information

Throughout this manual, the reader and/or system operator will be required

to utilize hazardous chemicals. During the performance or routine maintenance, the reader and/or system operator may be exposed to potentially dangerous electrical voltages and/or other hazards. To reduce the personal risk involved, the following guidelines are established. These requirements are within accepted standards for general analytical laboratory operation. Please refer to established safety requirements and government safety and disposal for additional information.

Chemical Hazards

WARNING HYDROCARBONS ARE TOXIC. UTILIZE THE CONTENTS OF ALL AC STANDARDS ONLY IN A FUME HOOD. AVOID INHALING FUMES AND CON-TACT WITH ANY PART OF THE BODY.

WARNING LIGHT HYDROCAR-BONS ARE EXTREME-LY FLAMMABLE. THE FUMES AND VAPORS FROM LIGHT HYDRO-CARBONS ARE EASILY IGNITED. KEEP THE CONTENTS OF ALL AC VIALS AWAY FROM FLAMES, SPARKS, OR SOURCES OF HEAT.

WARNING HYDROGEN IS A FLAMMABLE GAS. IF HYDROGEN OR ANY OTHER FLAMMABLE GAS IS USED, PERIODIC LEAK TESTS SHOULD BE PERFORMED. BE SURE THAT THE HYDROGEN SUPPLY IS OFF UNTIL ALL CON-NECTIONS ARE MADE AND ENSURE THAT THE INLET FITTINGS ARE EITHER CONNECTED TO A COLUMN OR CAPPED AT ALL TIMES HYDROGEN GAS IS PRESENT IN THE INSTRUMENT. DO NOT USE PLASTIC OR RUBBER GAS LINES FOR HYDROGEN.

CAUTION All gases are dangerous when compressed. Do not store cylinders where they might be dropped or exposed to excess

heat. Always direct the flow of compressed gases away from people.

Electrical Hazards

WARNING DANGEROUS ELECTRICAL VOLTAGES ARE PRESENT IN MANY PARTS OF THE SYSTEM, EVEN WITH THE POWER SWITCHES TURNED OFF. UNPLUG ALL THE MAIN POWER CORDS FROM THE POWER SOURCE FOR ALL THE SYSTEM COMPONENTS PRIOR TO SERVICING ANY OF THE SYSTEM COMPONENTS.

WARNING METAL COLUMNS ARE EXCELLENT CONDUCTORS OF ELECTRICITY. THE OVEN HEATER COILS INSIDE THE OVEN IN THE GAS CHROMA-TOGRAPH ARE NOT ELECTRICALLY INSULATED AND CARRY LETHAL VOL-TAGE. TO AVOID CONTACT WITH LETHAL ELECTRICAL VOLTAGE WHILE CHANGING METAL COLUMNS, TURN OFF THE MAIN POWER TO THE HP GAS CHROMATOGRAPH BEFORE REMOVING OR INSTALLING THE COLUMN. KEEP THIS COLUMN AWAY FROM POTENTIAL LIVE ELECTRICAL SOURCES AT ALL TIMES WHILE HANDLING.

Page 3: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Contents

Page 4: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Contents

Version 2011/1.0 Contents-2

TABLE OF CONTENTS PAGE

Pre-installation ....................................................................... 1

Pre-installation Requirements ...................................................................................... 1-2

Gas Requirements .......................................................................................................... 1-3 Gas Purity ...................................................................................................................... 1-3

Physical Requirements .................................................................................................. 1-4

Electrical Requirements ................................................................................................. 1-5

Pre installation Checklist ............................................................................................... 1-6 System Configuration .................................................................................................... 1-6 Software......................................................................................................................... 1-6

Recording Important Numbers ..................................................................................... 1-7

Installation .............................................................................. 2

Installing the Hardware .................................................................................................. 2-2 Introduction .................................................................................................................... 2-2 Cabling (General) .......................................................................................................... 2-2 Analyzer configuration ................................................................................................... 2-2 Installing columns and jumpers on the Capillary Deans bracket............................... 2-5

Installation of the liner ................................................................................................... 2-8

Installation of capillary column to the TPI. .................................................................. 2-9

Column Installation to the micro fluidic switch ........................................................ 2-10

Installing the AC OxyTracer methods ........................................................................ 2-11 Introduction .................................................................................................................. 2-11 Installing the AC methods ............................................................................................ 2-11 Microsoft Windows Instructions .............................................................................. 2-11

Operation ................................................................................ 3

Introduction..................................................................................................................... 3-2

Summary of Method ....................................................................................................... 3-3

Configuration .................................................................................................................. 3-4

Principle of Analysis ...................................................................................................... 3-5

Instrument setup ............................................................................................................ 3-7 Adjusting flows and pressures ....................................................................................... 3-7 PCM pressure ................................................................................................................ 3-7 Flows ............................................................................................................................. 3-8 Swivel test ...................................................................................................................... 3-8 Pre column analysis..................................................................................................... 3-10 Real pre column analysis............................................................................................. 3-10 Iterative method ........................................................................................................... 3-12 Calibration .................................................................................................................... 3-15 Sample handling .......................................................................................................... 3-16 Calculation ................................................................................................................... 3-16 Detectability ................................................................................................................. 3-17 Minimum Detection limit .......................................................................................... 3-17 Linearity: .................................................................................................................. 3-17

Repeatability ................................................................................................................ 3-17

Samples ......................................................................................................................... 3-18

Method description ...................................................................................................... 3-20

Page 5: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Contents

Version 2011/1.0 Contents-3

Spare parts kit and consumables kit OxyTracer ....................................................... 3-24

Page 6: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

1

Pre-installation

Page 7: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Pre-installation Requirements

Version 2011/1.0 1-2 of 7

Pre-installation Requirements

This chapter contains pre-installation procedures for the AC Analytical Controls OxyTracer Analyzer. Because the OxyTracer analyzer is based upon the Agilent 7890A Series gas chromatograph (GC), make sure your laboratory meets the environmental, weight, power, and gas requirements as described in the Site Preparation Checklist for the Agilent GC. This section contains gas requirements, physical requirements, electrical requirements, a pre-installation checklist for the AC OxyTracer Analyzer, and a record form.

Page 8: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Gas Requirements

Version 2011/1.0 1-3 of 7

Gas Requirements

Gas Purity

Some gas suppliers furnish “instrument” or “chromatographic” purity grades of gas that are intended specifically for chromatographic use. We recommend only these grades for use with Agilent GC detectors.

Only very low ppm levels (≤0.5 ppm) of oxygen and total hydrocarbons should be present in the gas supplies. Oil-pumped air supplies are not recommended because they may contain large amounts of hydrocarbons. The addition of high-quality moisture, hydrocarbon, and oxygen traps immediately after the main tank pressure is highly recommended. Refer to the section "Gas Supply" of the Agilent GC Site Preparation Checklist for more information on using traps.

Table 1-1. Gas Purity Recommendations

Carrier Gases

Gas type Helium

Pre-

pressure

6 bar 90 psi

Percent

purity

99.995%

Flame Ionization Detector Support Gases

Gas type Hydrogen Makeup gas (Nitrogen)

Air

Percent

purity

99.995% 99.995% (water- and oil- free) from a compressor

Page 9: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Physical Requirements

Version 2011/1.0 1-4 of 7

Physical Requirements

• Width - Approximately 2.1 meters (7 feet) for the complete system.

• Depth - Approximately 0.6 meters (2 feet) for the complete system.

• Height - Approximately 0.8 meters (2 feet 8 inches) for the complete system.

• Weight - 70 KG (155 lb.)

The analyzer may be installed in any reasonably accessible area.

THE GAS CHROMATOGRAPH WILL VENT HOT OVEN AIR FROM THE REAR OF

THE GAS CHROMATOGRAPH WHEN THE OVEN IS COOLING DOWN TO

AMBIENT TEMPERATURE. ENSURE NOTHING IS PLACED BEHIND THE GAS

CHROMATOGRAPH THAT MAY CONTACT THE HOT EXHAUST AIR. AC

MANUFACTURES AN ALUMINUM "OVEN EXHAUST DEFLECTOR" (AC PART NO.

59.40.200) THAT DEFLECTS THE EXHAUST AIR UP, AWAY FROM THE REAR OF

THE GAS CHROMATOGRAPH. THIS DEVICE EASILY AND QUICKLY ATTACHES

TO THE REAR OF THE GAS CHROMATOGRAPH. PLEASE CONTACT YOUR

LOCAL AC OFFICE SHOULD YOU DESIRE TO INSTALL THIS DEVICE.

WARNING

Page 10: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Electrical Requirements

Version 2011/1.0 1-5 of 7

Electrical Requirements

A proper earth ground is required for GC operations.

For North America and geographical areas that use 110 - 120 VAC :

• Gas chromatograph: 120 VAC. Please refer to the Agilent Site Preparation Checklist for additional information if required.

• All accessories: 110 - 120 VAC, 4 standard 3 prong grounded outlets.

For Europe and geographical areas that use 220/230/240 VAC

• Gas chromatograph: 220/230/240 VAC, Please refer to the Agilent Site Preparation Checklist for additional information if required.

• All accessories: 220 - 240 VAC, 4 grounded outlets

Maximum power consumption of the 7890A gas chromatograph is 2,250 VA for a regular oven and 2,950 for a fast-heating oven.

Do not turn on the oven until all electrical and gas connections have been completed and checked for correctness and leaks.

NOTE

NOTE

Page 11: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Pre installation Checklist

Version 2011/1.0 1-6 of 7

Pre installation Checklist

System Configuration

The Agilent system bundle configuration changes on a regular basis (typically every 6 months or so) to be faster, have a faster computer chip, and have a larger hard disk drive (HDD). The following are system guidelines:

Mininum PC Configuration

• Processor: 1.5 Ghz Pentium 4

• RAM: 512 MB

• Hard disc:40 GB

• Video: 1280 x 1024 resolution (SXGA)

• Removable Media: ATAPI CD, CD-RW or DVD drive

• Mouse: MS windows compatible pointing device

• LAN: 10/100baseT

Software • Microsoft Windows Windows XP Professional, Service pack 3

• Agilent G2070BA ChemStation32bit software for 1 GC

• Agilent G2071BA ChemStation32bit software for additional GC

The Agilent G2070BA ChemStation software consists of "core" software plus one Agilent GC instrument driver. Each additional Agilent GC requires the Agilent G2071BA additional instrument software, up to a maximum of three add-ons for four instruments per system.

NOTE

Page 12: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Pre-installation

Recording Important Numbers

Version 2011/1.0 1-7 of 7

Recording Important Numbers

In any verbal or written correspondence with Analytical Controls or Agilent concerning your ChemStation software, you will need to know the following information:

Agilent G2070BA Revision Code Example B.04.0X

Agilent G2070BA License No.

Microsoft Windows Type and Revision Code

Microsoft Windows License No.

Agilent Gas Chromatograph Serial No.

Agilent Computer Serial No.

Take a few minutes now to fill in the empty boxes above. The revision code is found on the diskettes or CD-ROM. The license numbers are found on the documentation shipped with the software.

Page 13: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

2

Installation

Page 14: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-2 of 11

Installing the Hardware

Introduction

This chapter contains installation procedures for the AC OxyTracer Analyzer for the Agilent 7890A Series gas chromatograph (GC). Refer to the 7890A GC installation Poster and to the 7890A GC User information for installation information. If the 7890A Automatic Liquid Sampler (ALS) is included, refer to the appropriate sections in the 7890A documentation.

Cabling (General)

The Agilent GC communicates electronically with other instruments through cables. Again refer to the 7890A GC installation Poster

• The Agilent 7890A Series GC is connected to the computer via LAN.

Automatic

liquid

sampler

7890Computer

1 LAN interface PC - GC

1

Mouse

Printer

Figure 2-1 Cabling configuration for the Agilent 7890A Series GC.

Analyzer configuration

The OxyTracer analyzer based on the Agilent 7890A gas chromatograph is configured with a TPI inlet, a capillary Deans bracket, a capillary pre column, a capillary analysis column, an auxiliary pressure control (PCM), a three way valve with a restrictor and a Flame Ionization Detector.

Page 15: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-3 of 11

FID

H2

FID

Air

Make-u

p N

2

Figure 2-2 Rear side connections 7890A GC

TEMPERATURE-SENSITIVE ITEMS (GAS CYLINDERS, CHEMICALS,

REGULATORS, ETC) SHOULD NOT BE LOCATED IN THE PATH OF THE

HEATED EXHAUST. ALSO, EXERCISE CARE WORKING AT THE REAR OF

THE INSTRUMENT DURING COOL-DOWN CYCLES TO PREVENT POSSIBLE

BURNS.

The instrument is factory-tested and is shipped with all columns pre installed. Following description should be followed in case capillary columns are broken or when new columns are installed.

WARNING

Page 16: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-4 of 11

Table 2-1. Columns for the AC OxyTracer Analyzer

Column Description Part No. for

Replacement

Maximum

temp °C

1 Pre column (capillary methyl

silicone, 0.8 mm OD)

10.74.011 325

2 Analysis column (capillary Lowox,

0.8 mm OD)

10.73.030 300

Monitor (0.5 mm OD) 21052.020 350

Table 2-2. Valves Used in the Instrument

Valve Assy 80102.300

Table 2-3. Ferrules / liners

Graphite ferrule 0.8 mm (10 pcs) 21040.007

Graphite ferrule Monitor 0.5 mm (10 pcs) 21041.001

Siltite ferrule 0.8 mm (10 pcs) 21040.026

Siltite ferrule 0.4 mm (10 pcs) 21040.025

Liner Siltek (10 pcs) 21032.007

Page 17: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-5 of 11

Installing columns and jumpers on the Capillary Deans bracket

The capillary Deans bracket (Figure 2-3) consists of a metal assembly and springs that are fixed to the oven shroud. A capillary pre column, a capillary analysis column and a capillary monitor are mounted on this assembly. The columns are connected to the micro fluidic switch which is attached to the oven left side

Before continuing, turn the oven and any heated zones off. Turn off all gases and turn off main power switch of the GC to avoid any potential hazard

1. Open the oven

2. Place the pre column (10.74.011) on the bracket

3. Attach one of the springs to the pre-column.

4. Place the analysis column (10.73.030) on the bracket

5. Attach the second spring to the analysis column

6. Connect the inlet of the analysis column to bottom connector of the micro fluidic switch.

7. Connect the outlet of the analysis column to the FID

8. Connect the pre column inlet to the TPI inlet

9. Connect the outlet of the pre column to middle connector of the micro fluidic switch

10. Connect the monitor column to top connector of the micro fluidic and fix the monitor to the pre column hangar, to prevent breaking it. (0.5 ferrule)

11. Connect the other end of the monitor to the vent connector located in front of the FID.

Do not use Liquid leak detector on the capillary connections

WARNING

WARNING

Page 18: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-6 of 11

Figure 2-3 Capillary Deans bracket

It is important to mount the columns in such a way as to minimize the tension in the columns.

Always wear protective glasses when working with fused silica capillary columns.

The AC OxyTracer Analyzer comes with pre installed and checked-out columns. The capillary analysis column, the capillary pre column and the capillary monitor column are all connected on the bracket to the various connectors and to the Inlet and the Detector. Special heat resistant wire is used to fix the columns to the bracket.

NOTE

WARNING

NOTE

Page 19: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the Hardware

Version 2011/1.0 2-7 of 11

Figure 2-4 Cross sectional overview of TPI

Table 2-4. General assay part numbers

Drawin

g

Part number Description

1 30.30.530 TPI Septum Inlet head

2 30.30.520 TPI Inlet head base

3 30.30.500 TPI Inlet housing assy.

4 30.30.510 TPI Inlet heating block

40.01.049 TPI Heater/sensor assy (70W40V)

70.40.200 Nut holder

30.20.064 Septum retainer nut

30.23.005 Capillary Column nut (2pcs)

30.30.550 TPI Isolation

Page 20: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installation of the liner

Version 2011/1.0 2-8 of 11

Installation of the liner

The TPI in the OxyTracer analyzer uses a special deactivated glass liner to prevent the loss of oxygenated compounds by adsorption. This liner can be installed at the topside of the injector.

The top of the TPI is connected to the base using a nickel seal. This seal must be replaced each time the top is removed from the base in order to avoid leaks.

Before installation of the liner turn oven temperature and Inlet to off, to work in a safe environment.

1 Turn off the inlet temperature regulation and the inlet EPC module.

2 Remove the TPI inlet head base (use wrench 15.10.403).

3 Disconnect the column.

4 Use TPI Liner remover (15.10.404) to remove old liner and TPI inlet seal.

Use a clean tissue to handle the liner and seal in order to prevent contamination of the liner.

4 Control liner before installation. Tapered side must be located at the bottom end of the inlet.

5 Place liner (part 10.32.007) into the injector by pushing it slowly into the inlet.

6 Place a new seal (10.10.194) on the TPI body. The seal is self-aligning.

Damage to the sealing beads on TPI body and TPI Septum cap holder will affect the fitting performance and cause system leakage. Handle with care.

7 Check if the liner top is not above the nickel seal. Loosen your column nut to lower the liner. Install the column again after action 8.

8 Reconnect the TPI inlet head base on inlet finger tight and turn max. 1/8 “ using the wrench.

9 Install the column (see page 9).

Over tightening will damage the sealing beads and possibly cause system leakage.

Note

Note

Warning

Warning

Page 21: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installation of capillary column to the TPI.

Version 2011/1.0 2-9 of 11

Installation of capillary column to the TPI.

The TPI uses a special deactivated glass liner to make a direct injection technique possible. The column is attached to the liner by pushing it to the tapered internal side of the liner and tightens using a graphite ferrule.

The inside diameter of the column must be 0.53 mm.

1 Turn off the GC oven.

2 Turn off the inlet temperature regulation and the inlet EPC module and let the system cool down.

3 Insert the column through the injector nut and graphite ferrule.

4 Cut a small piece of the column using a column cutter to make a clean and nice formed column end.

5 Lead the column end into the column connector and push upwards until the column stops on the liner.

Do not apply excessive force.

6 Tighten the column nut.

7 Install other end of column to the middle connector of the micro fluidic switch.

Note

Warning

Page 22: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Column Installation to the micro fluidic switch

Version 2011/1.0 2-10 of 11

Column Installation to the micro fluidic switch

The capillary columns and capillary jumpers are connected to the micro fluidic switch. The micro fluidic switch is made in such a way that there is hardly any dead volume resulting in good chromatography. All connections are made with 0.8 SilTite ferrules, except for the monitor column where a 0.4 SilTite ferrule is used.

The SilTite ferrules must be pre-swaged.

For instruction on pre-swaging SilTite Ferrules, see Swaging SilTite Ferrules.

Page 23: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Installation

Installing the AC OxyTracer methods

Version 2011/1.0 2-11 of 11

Installing the AC OxyTracer methods

Introduction

The AC OxyTracer analyzer does not need extra software besides ChemStation. The methods used for the oxygenates analyses are supplied on a CD.

A working knowledge of Microsoft Windows, and Agilent G2070BA ChemStation is required to operate the AC OxyTracer Analyzer successfully.

Installing the AC methods

These installation steps assume that the following software is already installed on the computer's hard disk:

• Microsoft Windows

• Agilent G2070BA ChemStation revision B.03.0x or higher

Microsoft Windows Instructions

1 Turn the computer and monitor on. The computer will boot up and start Microsoft Windows.

2 Insert the CD with the methods and copy the methods to the appropriate directory.

NOTE

Page 24: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

3

Operation

Page 25: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-2 of 24

Introduction

This section describes the operation, configuration, and analysis of the AC Analytical Controls OxyTracer Analyzer. The AC system is based upon the Agilent 7890A Series gas chromatograph (GC). Refer to the appropriate Agilent manuals for detailed information on the Agilent GC.

This system is dedicated to the analysis of trace oxygenated compounds in hydrocarbon liquids with a maximum boiling point of 250 °C. Oxygenates that can be determined are ethers, alcohols and ketones.

The analyzers is designed to the analysis of methanol, ethanol, n-Propanol*, i-Propanol, 2-butanol, i-Butanol**, t-butanol, methyl tert-butyl ether (MTBE), ethyl tert-butyl ether, (ETBE), tert-amyl methyl ether (TAME), diisopropyl ether (DIPE), Acetone and methyl ethyl ketone (MEK) in hydrocarbon streams (in particular naphthas) with a final boiling point below 250°C. The individual oxygenates can be determined from 0.1 to 500 ppm.

The system allows methods to be developed for the analysis of other oxygenated components.

*n-Propanol and I-Propanol are not separated and elute as one peak

** I-butanol, t-butanol en 2-butanol elute as one peak

Page 26: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-3 of 24

Summary of Method

The OxyTracer analyser is based upon the principle described by D.R. Deans

(Chromatographia, 1 (1968), 18). The principle of this technique is to control the

pressure between two columns and to direct the effluent of a (pre) column to

several, usually two, columns. By using columns of a very different nature (e.g. a

polar and a non-polar column) difficult separations can be achieved.

The sample is injected onto a methyl silicone pre column which elutes lighter

hydrocarbons to a monitor column and retains the oxygenated and heavier

hydrocarbons. Just before the oxygenated component elutes from this column,

the valve and thus a PCM pressure is switched and the components coming from

the pre column are directed to the analysis column. After the oxygenated

components have eluted from the pre column to the analysis column, the valve is

switched back to its original position, switching the pressure again. All the other

components remaining on the pre column are now directed to the monitor column

and the vent. The oxygenated components are now separated from the

remaining hydrocarbons on the analysis column and are detected by the Flame

Ionization Detector. The detector response, proportional to the component

concentration is recorded. The peak areas are measured and the concentration

of each component is calculated with reference to the external standard.

FID

TPI

PCM

pre column

analysis column

monitorVent

Figure 3-1. The AC OxyTracer flow diagram

Page 27: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-4 of 24

Configuration

The AC OxyTracer Analyzer uses an Agilent 7890A Series GC with electronic

pneumatics control (EPC) configured with a TPI inlet, a capillary Column bracket

a three-way valve with a capillary restrictor, a micro fluidic switch, a PCM (EPC)

pressure regulator, a vent and a Flame Ionization Detector. A second Flame

Ionization Detector is optional for monitoring the vent signal.

The capillary Column bracket is configured with a methyl silicone pre column, a

capillary analysis column (Lowox) and a monitor column. The micro fluidic switch

is located in the GC oven. The three-way valve with the capillary restrictor, as

well as the monitor vent are located on top of the gas chromatograph. A second

FID is optional to monitor the vent.

Figure 3-2. The AC OxyTracer – capillary Deans bracket

Page 28: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-5 of 24

Principle of Analysis

In standby, the Valve is in the off position (Fig. 3-3). The PCM pressure is

supplied at point 1.This means the pressure at point 1 is higher than the midpoint

pressure (2) So the pre column effluent is directed to the monitor column and the

vent. After injection of the sample, through the TPI a pre separation on boiling

point takes place on the pre column. The light hydrocarbons quickly pass both

the pre column and the monitor column to the vent. The heavier hydrocarbons

and the oxygenated compounds are retained. The monitor column does not

retard any components; its main function is to have an equal pressure as the

analysis column making pressure switching between the two channels possible.

FID

TPI

Detector

PCM

pre column

analysis column

monitor

1

2

3Vent

Figure 3-3. Pre column analysis - Valve in Off position

Just before the oxygenated components elute from the pre column, the valve is

switched (Fig 3.4). The PCM pressure is now supplied at point 3, making the

pressure at point 3 higher then the midpoint pressure, so the effluent from the

pre column is directed towards the analysis column. The oxygenated

components together with the hydrocarbons with similar boiling points elute onto

the analytical column. After the elution of the last oxygenated component, the

valve is switched again. The remaining hydrocarbons are thus directed towards

the monitor and the vent.

The fraction that was cut from the pre column contains oxygenated compounds

and some hydrocarbons with similar boiling points. This fraction is now separated

on the analysis column. Separation is done in a two stage oven ramped

temperature program. The oxygenates are measured on the Flame Ionization

Detector.

The capillary restrictor, which is connected to the three-way valve creates a little

‘leak flow’ thus preventing the diffusion of sample or components in the stainless

steel tubing leading to the valve.

Page 29: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-6 of 24

FID

TPI

Detector

PCM

pre column

analysis column

monitorVent

1

2

3

Figure 3-4. Cutting components - Valve in On position

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

0

20

40

60

80

100

120

140

FID1 A, (C:\DATAOX~1\D030715B\001F0107.D)

2.9

88

9.8

83 -

E

TB

E 1

0.1

31 -

M

TB

E 1

0.3

32 -

D

IPE

12.0

08 -

T

AM

E

14.2

38 -

M

eth

anol

14.6

11 -

A

ceto

ne

16.0

44 -

M

EK

16.2

35 -

E

thanol

17.6

94 -

Is

o &

n-P

ropanol 18.7

53 -

T

BA

& Iso &

2-B

uta

nol

19.2

77 -

n-B

uta

nol

Figure 3-5. Chromatogram 2.00 to 2.65 min cut from pre column

separated on analysis column (standard containing 14 oxygenates)

Page 30: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-7 of 24

Instrument setup

The instrument is factory-tested and is shipped with all columns pre installed. If

during transportation a capillary jumper or a column breaks, or if later on for

some reason a column is changed or any other chromatographic condition

changes, the flows and pressures will have to be checked. Also the monitor

column length needs to be readjusted or replaced when the analysis column is

replaced.

Adjusting flows and pressures

The Deans principle of switching with pressure will only function properly if there

is a pressure balance, meaning identical flows on both the monitor as the

analysis channel. Another condition is a suitable pressure difference between the

head pressure and the PCM pressure.

The actual Front inlet and the PCM pressures are instrument dependant. The

pressures are increased / decreased till an optimal separation of the Ether group

(ETBE, MTBE and DIPE) is achieved. See for the actual values the Test

document provided with the instrument.

PCM pressure

The PCM pressure provides the head pressure for both the analysis column as

well as the monitor column. Also it provides the ‘switching pressure’ of the

effluent from the pre column. A pressure difference of 2 kPa was found

satisfactory for this instrument.

If the PCM pressure is too low this may result in no switching at all, causing all

components to be directed to the monitor and vent. Another consequence of a

low switching pressure may be bad switching, resulting in tailing peaks, or no

peaks at all.

If the PCM pressure is too high, components are trapped on the pre column and

do not elute at all.

NOTE

Page 31: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-8 of 24

Flows

The following procedure describes how to setup the flows and pressures in the

system.

When all the columns have been installed (see installation chapter) the

flows/pressures have to be set. Load the pre column method (precol.M).

When the GC is ready, switch off the detector flows (air, hydrogen and make-up

flow) and measure the column flow on the detector and on the monitor vent.

These flows should not differ more than 5 %. To balance the flows, the monitor

length has to be tuned, the column flow on the analysis column is fixed and

cannot be varied.

The default monitor column length is one meter, if the flow through the monitor

column is lower than the flow through the analysis column this means the monitor

column has too much restriction, so it should be shortened.

Undo the monitor column from the vent, cut five centimeters from the column

end, remount the column and measure the flows again. Repeat this procedure

until the column flows on detector and vent are within 5 % of each other.

When the monitor column length has been adjusted, a test has to be performed

to establish whether the columns are properly connected.

If the flow through the monitor is higher than the flow through the analysis

column, the restriction of the monitor is too low. In this case a new length of

monitor column has to be mounted and tuned.

Swivel test

A method to verify that all columns have been correctly mounted is the swivel

test. The method to perform this test is Swivel.M. In this method the valve is

switched every 0.10 min whilst a pure component (e.g. Cyclohexane) elutes from

the pre column. Switching the valve off in this peak must give an immediate drop

in signal on the detector. Also it must give an immediate rise, when the valve is

switched back again.

Load the method SWIVEL.M, adjust the PCM pressure at 36 kPa and inject

Cyclohexane.

Page 32: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-9 of 24

min2 2.5 3 3.5 4 4.5 5 5.5 6 6.5

pA

0

10

20

30

40

50

60

70

80

90

FID1 A, (A:\D040209\SIG10098.D)

Figure 3-6. Swivel test on Cyclohexane

Visually check the chromatogram. An example of a good swivel test is given in

figures 6.

In case of a bad swivel test, this means there is a dead volume somewhere in the

system. Most likely one of the capillary connections (column or jumper) is not

okay and must be checked.

Page 33: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-10 of 24

Pre column analysis

With the introduction of electronic pressure control, the retention time stability in

capillary gas chromatography increased enormously. Therefore also the cutting

times for the OxyTracer analyzer are very stable and hardly need readjusting.

However when a new column is installed or when some other chromatographic

condition changes or when one wants to setup a method for a specific oxygenate

it may be necessary to re-determine the cutting times.

There are two ways of determining the correct cutting times. The real pre column

analysis of a qualitative pre column standard which allows one to record the cut

times from the chromatogram. Or an iterative method whereby a sequence is run

with different methods each with varying cut times.

Real pre column analysis

To establish the correct valve cutting times, a pre column analysis of a cutting

time standard has to be run. This qualitative standard has to contain the

oxygenates of interest in n-Dodecane. In the pre column method the pre column

effluent is directed to the monitor column and the FID. So the pre column

chromatogram reflects the exact times at which the oxygenates elute from the

pre column.

To make a pre column chromatogram, the monitor column has to be connected

to the FID instead of the vent. The following procedure has to be followed:

• Cool down the GC oven to ambient

• Undo the analysis column from the Flame ionization detector

• Leave the column end of the analysis column loose in the oven

• Undo the monitor column from the vent

• Connect the monitor column to the Flame Ionization detector

• Heat the oven to 100 °C

When the GC is ready, the pre column analysis can be performed.

Page 34: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-11 of 24

min0.5 1 1.5 2 2.5 3 3.5

pA

0

100

200

300

400

500

600

FID2 B, (A:\007B0703.D)

2.398

2.686

Methanol

n-Butanol

T1 T2

Figure 3-7. Pre column chromatogram

Load the pre column method (PRECOL.M), set the pcm pressure to 36 kPa and

inject the pre column standard. From the obtained chromatogram, determine the

two cut times. (Figure 8) T1 and T2.

Run the standard a second time to make sure the correct cut times are obtained.

Now before samples can be analyzed, first the column and monitor have to be

reconnected:

• Cool down the GC oven to ambient

• Undo the monitor column from the Flame ionization detector

• Connect the analysis column to the Flame ionization detector

• Connect the monitor column to the vent

• Heat the oven to 100 °C

The earlier established cut times must then be entered in the analysis method.

(OXY.M). Load the method OXY.M,

Select: View, Instrument, Edit Parameters…,

Runtime and edit (Replace) the

Valve 1 On and Off time. (see figure 8)

Valve 1 On time = T1

Valve 1 Off time = T2

Finally Save the method.

Page 35: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-12 of 24

Figure 3-8. Edit runtime table

Iterative method

In this method the cut times are determined by iteration. A set of cut times is

tried, and depending on the result a new adapted set of cut times is tried, this

step is repeated until the final cut times are found.

This method of cut times determination can be automated. Different methods are

made, each only differing in cut times. For each method T1 and T2 should be

varied with steps of 0.05 min. Starting for instance with a T1 of 2.50 minutes and

a T2 of 3.00 minutes. These methods are all saved with unique names. (e.g.

Cut_1 etc.)

Then make a sequence and analyze the pre column standard with the different

methods. Afterwards review the chromatograms that were obtained and select

the one with the proper cut times. An example can be found in figure 5.

If n-Butanol is missing, T2 should be increased. If Methanol is missing T1 should

be decreased.

Page 36: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-13 of 24

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

5

10

15

20

25

30

35

40

45

FID1 A, (C:\DATAOX~1\D030716\SIG10036.D)

14.1

77 -

M

eth

anol

19.2

67 -

n-B

uta

nol

Figure 3-9. Analysis of pre column standard

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

5

10

15

20

25

30

FID1 A, (C:\AASEQU~1\013F2201.D)

10.0

31 -

M

TB

E

14.1

93 -

M

eth

anol 1

4.5

07 -

A

ceto

ne

16.2

05 -

E

thanol

Figure 3-10. Calibration chromatogram – containing MTBE, Methanol and

Ethanol (acetone contamination)

The Lowox column can trap impurities in the carrier gas relatively easy.

Therefore a standby temperature of 200 °C is recommended.

NOTE

Page 37: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-14 of 24

Table 3-1. Elution order of some oxygenated compounds

Component

Trivial name IUPAC name

Ether Diethyl ether

Etbe 2-Ethoxy-2-methyl propane

Mtbe 2-Methoxy-2-methyl propane

Dipe 2-Isopropoxy propane

Tame 2-Methyl-2-methoxy butane

Dipropyl ether n-Propyl ether

Butyl ethyl ether

Butyraldehyde Butanal

Methanol

Acetone 2-Propanone

Iso valeraldehyde 3-Methyl butanal

Valeraldehyde n-Pentanal

MEK 2-Butanone

Ethanol

Diethyl ketone 3-Pentanone

Methyl propyl ketone 2-Pentanone

i-Propanol

propanol

2-Popanol

t-butanol

i-butanol

2-butanol

t-Butyl alcohol

2-Methyl-1-propanol

sec-Butyl alcohol

n-Butanol

Page 38: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-15 of 24

Calibration

After the correct cut times have been set and the method has been saved, a

calibration should be performed.

To calibrate the method, a calibration standard has to be analyzed. This

calibration standard should contain the oxygenates of interest at low level.

An external calibration is performed. Prepare a calibration standard as described

in ASTM D-4307. Inject the calibration standard using method Oxy.m. Make sure

the proper concentrations are filled out in the calibration table, check the

chromatography (integration) and calibrate

Data Analysis – Calibration – Recalibrate

The response factors are now calculated and stored in the calibration file. The

method is now ready for sample analysis.

Used components must be of known purity, hydrocarbons used must also be free

of the components to be analyzed.

NOTE

Page 39: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-16 of 24

Sample handling

Effort must be taken to assure representative samples and standards are

obtained. Some oxygenated compounds are very reactive. Methanol is a very

good example, at very low concentrations it can readily be lost in the vials it is put

into.

This means that sampling becomes very critical. Also calibration samples can

change in time. So it is necessary to make fresh standards on a regular basis.

Store calibration standards capped below 5 °C when not in use.

Another point of attention therefore is the injection. It is necessary to use the

washing options and vials of the auto sampler to flush the syringe after an

injection and prior to an injection. Fill the washing vials with an oxygenate-free

hydrocarbon (e.g. n-Nonane, or Cyclohexane)

Calculation

Quantitative results are based on calibration analyses obtained by injecting a

fixed volume of a reference blend. With the reference blend, the relation between

peak area and component concentration is determined. This response factor is

calculated by ChemStation as follows:

Ri= Ci/Ai

Where: Ri = Response factor (ppm Wt/µV)

Ci = concentration compound (ppm Wt)

Ai = peak are of the compound (µV)

The unknown concentration of a component x in a sample is calculated as

follows:

Cx = Ax * Rx

Where: Cx = concentration component (ppm Wt)

Rx = Response factor component (ppm Wt/µV)

Ax= peak area of the component (µV)

NOTE

Page 40: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-17 of 24

Detectability

Minimum Detection limit

This is defined as the smallest concentration that can be distinguished from a

blank analysis. (With a 95 % confidence level)

LDL is 0.1 ppm Wt

Linearity:

Linear response from 0.1 to 500 ppm Wt

Repeatability

The repeatability at a 10 ppm concentration level for Methanol and Ethanol is

better than 5 % (RSD), for MTBE the repeatability at the same level is better than

1 % (RSD).

Page 41: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-18 of 24

Samples

Following are some example chromatograms of various samples.

Figure 3-11. Reformate containing 0.7 ppm MTBE & 2.6 ppm MEK.

Figure 3-12. Virgin Naphtha contains 13.0 ppm MTBE, 19 ppm MEK and

various alcohols

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

0

10

20

30

40

50

FID1 A, (C:\DATAOX~1\SIG10041.D)

2.2

51

2.3

49

2.5

23

2.6

67

2.9

04

3.4

96

4.7

07 5

.012

5.1

76

5.2

59

6.9

52

7.7

40

10.1

79 -

M

TB

E

13.0

52

13.6

51

15.6

20

15.8

27

16.0

91 -

M

EK

16.8

16

17.0

38

17.3

26

17.7

30 -

Is

o &

n-P

ropanol

18.8

47 -

T

BA

& I

so &

2-B

uta

nol

19.3

25 -

n-B

uta

nol

19.8

67

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

0

10

20

30

40

50

FID1 A, (C:\AASEQU~1\025F0701.D) 2

.265

2.2

88

2.3

47

2.3

75

2.3

96

2.4

19

2.4

77

2.5

03

2.5

59

2.6

33

9.3

07

9.9

90 -

M

TB

E

14.6

06 -

A

ceto

ne

16.0

65 -

M

EK

17.7

17

Page 42: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-19 of 24

Figure 3-13. Raffinate contains 35 ppm MTBE and 0.9 ppm Acetone

min0 2.5 5 7.5 10 12.5 15 17.5 20 22.5

pA

0

10

20

30

40

50

FID1 A, (C:\AASEQU~1\024F0403.D)

2.2

63

2.3

36

2.3

61

2.3

91

2.4

82

2.5

10

2.5

96

2.6

64

2.8

33

2.9

28

3.0

52

3.2

34

3.5

44

3.7

96

3.8

54

3.9

76

4.0

57

4.1

64

10.0

32 -

M

TB

E

14.6

28 -

A

ceto

ne

Page 43: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-20 of 24

Method description

Table 3-2 Oxy Method

Oven

Initial temperature:

100 °C Initial time: 5 min

Rate: 5 °C/min Final temp 130 °C

Final time 0 min

Rate 10 °C/min Final temp 225 °C

Final time 9.5 min

Total run time 30 min

Inlet (He) PCM B-1 (He)

Temperature: 200 °C Initial pressure 36 kPa*

Front inlet (He): 38 kPa*

FID

Temperature: 250 °C

Flow (H2): 35 ml/min

Flow (Air): 350 ml/min

Makeup (N2): 20 ml/min

Data rate: 50 Hz

Run Time Table

Specifier Typical Time** Setting Description

Valve 1 initial OFF Sample to monitor

Valve 1 2.00 ON Sample to analysis column

Valve 1 3.50 OFF Sample to monitor

GC Injector

Sample washes 3 Injection volume:

1.0 µL

Sample pumps 6

PostInj Solvent A washes

4

*See the Test document provided with the instrument for the actual pressure settings for the instrument

**Times will be fine-tuned and included with checkout documentation for each analyzer

Page 44: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-21 of 24

Table 3-3 Precol Method

Oven

Initial temperature:

100 °C Initial time: 5 min

Rate: 5 °C/min Final temp 130 °C

Final time 0 min

Rate 10 °C/min Final temp 225 °C

Final time 9.5 min

Inlet (He) PCM B-1 (He)

Temperature: 200 °C Initial pressure 36 kPa*

Front inlet (He): 38 kPa*

FID

Temperature: 250 °C

Flow (H2): 35 ml/min

Flow (Air): 350 ml/min

Makeup (N2): 20 ml/min

Data rate: 50 Hz

Run Time Table

Specifier Typical Time** Setting Description

Valve 1 initial OFF Sample to monitor

GC Injector

Sample washes 3 Injection volume:

0.1 µL

Sample pumps 6

PostInj Solvent A washes

4

*See the Test document provided with the instrument for the actual pressure settings for the instrument

**Times will be fine-tuned and included with checkout documentation for each analyzer

Page 45: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-22 of 24

Table 3-4. Swivel method

Oven

Initial temperature:

100 °C Initial time: 5 min

Rate: 5 °C/min Final temp 130 °C

Final time 0 min

Rate 10 °C/min Final temp 225 °C

Final time 9.5 min

Inlet (He) PCM B-1 (He)

Temperature: 200 °C Initial pressure 36 kPa*

Front inlet (He): 38 kPa*

FID

Temperature: 250 °C

Flow (H2): 35 ml/min

Flow (Air): 350 ml/min

Makeup (N2): 20 ml/min

Data rate: 50 Hz

Run Time Table

Specifier Typical Time* Setting Description

Valve 1 initial OFF Sample to monitor

Valve 1 2.00 ON Sample to analysis column

Valve 1 2.10 OFF Sample to monitor

Valve 1 2.20 ON Sample to analysis column

Valve 1 2.30 OFF Sample to monitor

Valve 1 2.40 ON Sample to analysis column

Valve 1 2.50** OFF Sample to monitor

Page 46: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-23 of 24

GC Injector

Sample washes 3 Injection volume:

1.0 µL

Sample pumps 6

PostInj Solvent A washes

4

*See the Test document provided with the instrument for the actual pressure settings for the instrument

** To be repeated until the run table is full (maximum is 75 settings)

Page 47: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Operation

Samples

Version 2011/1.1 3-24 of 24

Spare parts kit and consumables kit OxyTracer

Spares parts kit CCG4000.100

Valve Assy 80102.300

Aluminum washer (10 pcs) 21011.040

FTA restrictor 30.40.015

Consumables kit CCG4000.200

Capillary Lowox (analysis column) 10.73.030

Capillary methyl silicone (pre column) 10.74.011

Monitor 21052.020

SilTite ferrule 0.8 (10 pcs) 21040.026

SilTite ferrule Monitor 0.4 (10 pcs) 21040.025

Liner Siltek (10 pcs) 21032.007

Septa, 9.5 mm (50 pcs) 21040.004

Syringe (10 µL) 30.25.001

Sample box 14 oxygenates 20001.520

Page 48: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

4

Index

Page 49: Operating Manual - ptplab.net · AC OxyTracer Analyzer on 7890 GC Manual part number 23070.020 version 2011/1.0 Contact addresses:   AC Analytical Controls BV

Index

Version 2011/1.0 4-2 of 5

Index

A AC OxyTracer flow diagram ........ 3-3

C cabling ........................................ 2-2

cabling configuration ............... 2-2 calibration

AC Oxytracer analyzer .........3-15 Column configuration .................. 2-4 configuration

AC OxyTracer Analyzer ........ 3-4 configuration, system .................. 1-6 consumables kit .........................3-24

G gas purity .................................... 1-3 gas requirements ........................ 1-3 GC method ................................3-21

Oxy ......................................3-20 GC method: ...............................3-21

H Helium ........................................ 1-3 hydrocarbon

trap ....................................... 1-3 Hydrogen .................................... 1-3

I installation

AC OxyTracer methods .......2-11 hardware ............................... 2-2

Installation AC methods .........................2-11

Introduction ................................. 3-2

L LAN ............................................. 2-2

O Oven Exhaust Deflector .............. 1-4

P physical requirements ................. 1-4 pre installation checklist .............. 1-6 preinstallation requirements ........ 1-2 principle of analysis

AC OxyTracer Analyzer ......... 3-5

R rear side connections .................. 2-3 Requirements

electrical ................................ 1-5

S software ...................................... 1-6

installation ........................... 2-11 Spare parts kit ........................... 3-24 Spare parts kit and consumables kit OxyTracer ................................. 3-24 system configuration ................... 1-6

T trap

hydrocarbon .......................... 1-3 moisture ................................ 1-3 oxygen ................................... 1-3

V valves .......................................... 2-4