Operate Model 503DR Ml HP-2 or Ml HP-3 Hydroprobe Neutron Moisture Gauge Type REFERENCE Document No. TO-320-022 Rev/Mod G-1 Release Date 12/13/2016 Page 1 of 34 Tank Farm Plant Operating Procedure Surveillance GCX-2 CHANGE HISTORY ( LAST 5 REV-MODS ) Rev-Mod Release Date Justification: Summary of Changes G-1 12/13/2016 Document title change Update title for TFC-ESHQ-ENV-STD-06. G-0 11/10/2016 Procedure Update to Current Field Work Conditions Clarification of shipping/storage container. Update Personnel Safety and RadCon sections. Clarify example of probe readout and probe depth F-1 07/11/2016 Inconsequential Change Updated Records Section. F-0 10/08/2014 Periodic Review Pages 4, 14-16, & 25 Changed warning into step. Removed the “critical step” icons. E-1 05/06/2013 Operations Request. Changed title of procedure and changed from CONTINUOUS to REFERENCE use. Added, deleted and revised steps to provide a more efficient performance of task(s). Table of Contents Page 1.0 PURPOSE AND SCOPE ................................................................................................................ 3 1.1 Purpose................................................................................................................................ 3 1.2 Scope ................................................................................................................................... 3 2.0 INFORMATION............................................................................................................................. 3 2.1 Terms and Definitions......................................................................................................... 3 2.2 General Information ............................................................................................................ 3 3.0 PRECAUTIONS AND LIMITATIONS......................................................................................... 4 3.1 Personnel Safety.................................................................................................................. 4 3.2 Equipment Safety ................................................................................................................ 4 3.3 Radiation and Contamination Control ................................................................................ 5 3.4 Environmental Compliance ................................................................................................ 5 4.0 PREREQUISITES .......................................................................................................................... 6 4.1 Special Tools, Equipment, and Supplies............................................................................. 6 4.2 Performance Documents ..................................................................................................... 6 4.3 Field Preparation ................................................................................................................. 7

Operate Model 503DR Ml HP-2 or Ml HP-3 Hydroprobe Neutron

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

1 of 34

Tank Farm Plant Operating Procedure Surveillance

GCX-2

CHANGE HISTORY ( LAST 5 REV-MODS )

Rev-Mod Release Date Justification: Summary of Changes

G-1 12/13/2016 Document title change Update title for TFC-ESHQ-ENV-STD-06.

G-0 11/10/2016 Procedure Update to Current

Field Work Conditions

Clarification of shipping/storage container. Update Personnel

Safety and RadCon sections. Clarify example of probe readout

and probe depth

F-1 07/11/2016 Inconsequential Change Updated Records Section.

F-0 10/08/2014 Periodic Review Pages 4, 14-16, & 25 Changed warning into step. Removed the

“critical step” icons.

E-1 05/06/2013 Operations Request.

Changed title of procedure and changed from CONTINUOUS to

REFERENCE use. Added, deleted and revised steps to provide

a more efficient performance of task(s).

Table of Contents Page

1.0 PURPOSE AND SCOPE ................................................................................................................ 3

1.1 Purpose ................................................................................................................................ 3

1.2 Scope ................................................................................................................................... 3

2.0 INFORMATION............................................................................................................................. 3

2.1 Terms and Definitions......................................................................................................... 3

2.2 General Information ............................................................................................................ 3

3.0 PRECAUTIONS AND LIMITATIONS......................................................................................... 4

3.1 Personnel Safety.................................................................................................................. 4

3.2 Equipment Safety ................................................................................................................ 4

3.3 Radiation and Contamination Control ................................................................................ 5

3.4 Environmental Compliance ................................................................................................ 5

4.0 PREREQUISITES .......................................................................................................................... 6

4.1 Special Tools, Equipment, and Supplies............................................................................. 6

4.2 Performance Documents ..................................................................................................... 6

4.3 Field Preparation ................................................................................................................. 7

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

2 of 34

5.0 PROCEDURE ................................................................................................................................. 8

5.1 Borehole/Drywell Cleanliness Check ................................................................................. 8

5.2 Prepare the MI HP-2 or MI HP-3 Hydroprobe for Operation .......................................... 11

5.3 Obtain Soil Moisture Readings with MI HP-2 or MI HP-3 Hydroprobe ......................... 14

5.4 Confirm Data in MI HP-2 or MI HP-3 Hydroprobe ......................................................... 18

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification with MI HP-2 or

MI HP-3 Hydroprobe ........................................................................................................ 19

5.6 Disposition ........................................................................................................................ 30

5.7 Records ............................................................................................................................. 30

Data Sheet 1 - Borehole/Drywell Data ..................................................................................................... 31

Data Sheet 2 - DR503 Soil Moisture Gauge Performance Verification Data........................................... 33

Attachment 1 - Display Error Messages ................................................................................................... 34

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

3 of 34

1.0 PURPOSE AND SCOPE

1.1 Purpose

This procedure provides instructions for using CPN Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe at an assigned tank farm vadose zone borehole/drywell or for obtaining data

for Hydroprobe-Neutron Source performance verification.

This procedure can be performed in multiple locations. A work area and/or location

specific hazard analysis must be performed prior to starting the activity per

TFC-ESHQ-S_SAF-C-02.

1.2 Scope

This procedure applies to the setup of the Hydroprobe for borehole/drywell logging,

operating, and obtaining and transferring borehole/drywell data from the Hydroprobe to a

computer.

1.2.1 This procedure also applies to obtaining data for Hydroprobe-Neutron Source

performance verification.

2.0 INFORMATION

2.1 Terms and Definitions

ORRSR - Onsite Routine Radioactive Shipment Record

2.2 General Information

2.2.1 See Attachment 1 for a listing of the display and meaning of error messages.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

4 of 34

3.0 PRECAUTIONS AND LIMITATIONS

3.1 Personnel Safety

WARNING - Attempting to catch and regain control of the probe and cable if it slips

free during removal or insertion into the borehole/drywell could result in

personnel injury.

WARNING - Failure to handle the probe and source carefully or keep it at a maximum

distance could result in increased Neutron exposure.

WARNING - Failure to store the neutron source in its shipping/storage container when

not in use, could result in increased Neutron exposure.

WARNING - Removal of radioactive liquid could result in the spread of contamination.

3.1.1 Use of a neutron source requires users to be qualified as Source User (Course

356424). Issue and return of this source must be conducted by a Source

Custodian or Source Custodian delegate.

3.1.2 Poor lighting conditions and snow cover may hide uneven ground, survey

stakes, digging, etc. from the view of personnel performing this procedure.

3.2 Equipment Safety

CAUTION - Performing steps in this section (5.5) while precipitation is occurring or

when considerable water or snow is present near the probe, could bias the

data obtained.

CAUTION - To obtain the most accurate readings the moisture detection system should

have a Performance Verification performed. The Performance

Verification needs to be performed with the 503DR Hydroprobe meter and

source/detector together.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

5 of 34

3.3 Radiation and Contamination Control

3.3.1 The neutron source is to be stored in its shipping/storage container when not

in use.

3.3.2 Cleanliness check swabs will be controlled as radioactive material after use

and will be surveyed by a Health Physics Technician (HPT) prior to exiting

the farm.

3.3.3 Removal of contaminated liquid requires HPT coverage and cannot be

performed using the long term Radiological Work Permit that permits routine

drywell logging/monitoring activities.

3.3.4 Failure to handle the probe and source carefully or keep it at a maximum

distance could result in increased Neutron exposure.

3.3.5 When this procedure is worked in radiological areas, an approved

radiological work permit (RWP) is required. If radiological conditions or

work performed falls outside the scope of the RWP, all work activities must

be discontinued until a new or revised RWP has been issued in accordance

with TFC-ESHQ-RP_RWP-C-03.

3.4 Environmental Compliance

3.4.1 If borehole/drywell is found to be contaminated during use of this procedure,

Environmental must be notified per the Environmental On-Call List.

3.4.2 The following controls specified in the appropriate section of

TFC-ESHQ-ENV-STD-06, As Low As Reasonably Achievable Control

Technology (ALARACT) Requirements Standard, are applicable:

At a minimum, pre- and post-job surveys shall be taken and

documented on a Radiological Survey Report.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

6 of 34

4.0 PREREQUISITES

4.1 Special Tools, Equipment, and Supplies

The following equipment and supplies may be needed to perform this procedure:

Clean rags

Swabbing equipment

Rubber gloves

Face shield

BLUE copy of Onsite Routine Radioactive Shipment Record

Magnetic caution signs

Measuring tape with swab

CPN International Inc. Model 503DR Ml HP-2 or Ml HP-3 Hydroprobe

Moisture Models A, B, C, E, F, and G located at site adjacent to weather station.

4.2 Performance Documents

The following procedure may be needed to perform this procedure:

TO-100-052, Perform Waste Generation, Segregation, Accumulation and Clean-

up.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

7 of 34

4.3 Field Preparation

4.3.1 CHECK that there is a calibration certificate on file for the hydroprobe-

neutron source combination used, or that engineering personnel have verified

hydroprobe-neutron source combination is acceptable for use.

4.3.2 IF obtaining data for Hydroprobe-Neutron Source performance verification

AND

IF directed by Shift Manager/OE, GO TO Section 5.5.

NOTE - The ORRSR provides the Emergency Response Information required per HNF-

5183 Tank Farms Radiological Control Manual, Article 423.13.

4.3.3 WHEN transporting the neutron source, ENSURE Blue copy of the Onsite

Routine Radioactive Shipment Record (ORRSR) is in possession.

4.3.4 PERFORM Neutron and Gamma surveys at shielded storage cask to ensure

levels are as specified on the ORRSR prior to transporting.

4.3.5 PERFORM Neutron and Gamma surveys on the source: Contact, 30cm to

validate previous levels.

4.3.5.1 INFORM the work team of these levels.

4.3.5.2 IF significant changes from previous readings are identified,

PERFORM the following in the most logical order:

a. PLACE work in a safe condition.

b. NOTIFY Surveillance Manager and RadCon

Management.

4.3.6 ENSURE Process Engineer has provided a copy of Data Sheet 1 for each of

the following:

Borehole/drywell to be sampled with borehole/drywell number

Well ID number

Depths of readings to be taken.

4.3.7 PERFORM a work area and/or a location specific hazards analysis per

TFC-ESHQ-S-SAF-C-02.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

8 of 34

5.0 PROCEDURE

WARNING

Failure to store the Neutron source in its shipping/storage container

when not in use, could result in increased Neutron exposure.

WARNING

Failure to handle the probe and source carefully or keep it at a

maximum distance could result in increased Neutron exposure.

NOTE - The above WARNINGs apply to all steps related to the use of the source within

Section 5.0.

- Sections 5.1 and 5.2 can be performed in any logical order.

- Sections 5.1 through 5.3 are performed for each borehole/drywell identified by Process

Engineer.

5.1 Borehole/Drywell Cleanliness Check

5.1.1 IF cleanliness checks have been performed within 30 days of running the

scan AND

IF cleanliness checks have been documented on a Radiological Survey

Report, SKIP Section 5.1 as it is not required.

WARNING Removal of radioactive liquid could result in the spread of contamination.

5.1.2 REMOVE borehole/drywell plug to gain access to borehole/drywell.

5.1.3 LOWER swab into the borehole/drywell to check for contamination.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

9 of 34

5.1 Borehole/Drywell Cleanliness Check (Cont.)

NOTE - The Health Physics Technician (HPT) is responsible for monitoring the tape.

5.1.4 CONTINUOUSLY MONITOR tape with contamination survey instrument

for any indication of increasing meter response AND

REEL UP swab.

5.1.4.1 IF background is greater than 150 counts per minute, CHECK in

a low background area per RWP requirements.

5.1.5 IF the radiation/contamination survey meter indicates contamination, STOP

removing the swab.

5.1.5.1 NOTIFY the following:

Surveillance Manager

Shift Manager

RadCon Manager.

5.1.6 IF the radiation/contamination survey meter response does not increase and

swab is wet or dry (Normal Condition), PERFORM the following:

5.1.6.1 WHEN the swab is in proper position, STOP raising swab.

5.1.6.2 PLACE bucket near well opening.

5.1.6.3 PLACE swab in bucket.

5.1.7 MONITOR swab for contamination.

5.1.8 IF background radiation levels are too high to perform a contamination

survey of the swab, PERFORM the following:

5.1.8.1 CONTAIN the swab.

5.1.8.2 MOVE the swab to an area of lower background radiation.

5.1.9 IF swab is wet, PERFORM the following:

5.1.9.1 PERFORM smear of swab.

5.1.9.2 IF smear is not saturated, PERFORM field count.

5.1.9.3 IF unable to field count, HAVE smear monitored at the count lab

facility.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

10 of 34

5.1 Borehole/Drywell Cleanliness Check (Cont.)

5.1.10 IF contamination is detected at greater than 5,000 dpm/Probe Area, Beta

Gamma or 100 dpm/Probe Area Alpha, PERFORM the following:

5.1.10.1 RE-INSTALL borehole/drywell cap

5.1.10.2 NOTIFY the following:

Surveillance Manager

Shift Manager

RadCon Manager.

5.1.11 DISPOSE of contaminated items (i.e., double bag, place in satellite

accumulation area and fill out waste inventory sheet per TO-100-052).

5.1.12 IF contamination is not detected, PLACE reel and swab into proper

container.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

11 of 34

5.2 Prepare the MI HP-2 or MI HP-3 Hydroprobe for Operation

CAUTION

To obtain the most accurate readings the moisture detection system

should have a Performance Verification performed. The Performance

Verification needs to be performed with the 503DR Hydroprobe meter

and source/detector together.

5.2.1 CONFIRM screen displays “READY”.

5.2.1.1 IF a decimal point is displayed in lower left or a bar is displayed

in the left character, ENSURE the battery is charged.

5.2.2 PRESS TIME AND

CONFIRM screen displays indicates “16”.

5.2.2.1 IF display does not indicate 16, PUSH STEP until display reads

16 AND

PUSH ENTER.

5.2.3 PRESS UNITS AND

CONFIRM screen displays indicates “MUNT CNT”.

5.2.3.1 IF display does NOT indicate “MUNT CNT”, PUSH STEP until

display reads MUNT CNT AND

PUSH ENTER.

5.2.4 ATTACH the MI-HP-2 or MI-HP-3 meter to the source/detector AND

DOCUMENT on Data Sheet 1 the source/detector, meter identification

numbers and depth or number of readings.

5.2.5 REVIEW the copy of Data Sheet 2 that is kept with the source/detector

shipping/storage container AND

CONFIRM that the source/detector and meter have had a Performance

Verification performed within the last year.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

12 of 34

5.2 Prepare the MI HP-2 or MI HP-3 Hydroprobe for Operation (Cont.)

5.2.6 SET format as follows:

5.2.6.1 PUSH FMT.

5.2.6.2 PUSH STEP twice.

5.2.6.3 OBTAIN number of depths or measurements to be entered from

Data Sheet 1.

5.2.6.4 WHEN “DEPTHS” is displayed on screen, ENTER identified

number of depths from engineering AND

PUSH STEP.

5.2.6.5 WHEN screen displays “SET FMT?” PUSH ENTER twice.

5.2.6.6 CONFIRM screen displays “READY”.

5.2.7 OBTAIN Chi-square value as follows:

5.2.7.1 PLACE the case away from other radioactive sources.

5.2.7.2 ENSURE the following:

Neutron Probe shipping/storage container on stable surface

No other radioactive sources are within 30 feet of gauge

No significant source of hydrogen is within 10 feet

All personnel are at least 10 feet away after starting the

reading.

5.2.7.3 POSITION neutron probe on yellow square on top of neutron

probe shipping/storage container.

5.2.7.4 PUSH STD.

5.2.7.5 PUSH STEP three times and “NEW STD” will be displayed on

screen.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

13 of 34

5.2 Prepare the MI HP-2 or MI HP-3 Hydroprobe for Operation (Cont.)

5.2.8 OBTAIN a standard count as follows:

5.2.8.1 WHEN “NEW STD?” is displayed, PUSH ENTER.

NOTE - Read-out unit will beep every few seconds during standard count.

- Displayed numbers are not entered on final Data Sheet until tasks

are within range.

5.2.8.2 WAIT 4 to 5 minutes for completion of standard counts.

5.2.8.3 WHEN beeping stops, CONFIRM screen on read-out unit

displays a standard count (e.g. S 7145 or S 11034).

5.2.8.4 PUSH STEP AND

CONFIRM previous standard is displayed (e.g. P XXXX).

5.2.8.5 PUSH STEP AND

CONFIRM Chi number is displayed.

5.2.8.6 CONFIRM Chi indicates 0.75 to 1.25.

5.2.8.7 RECORD Chi number on Data Sheet 1.

5.2.8.8 IF Chi is lower than 0.75 or higher than 1.25, RETRY by

performing Steps 5.2.8.1 through 5.2.8.6.

5.2.8.9 IF Chi fails again, RECHARGE the battery AND

NOTIFY technical POC or Engineering.

5.2.8.10 PUSH STEP AND

RECORD standard count on Data Sheet 1.

5.2.8.11 PUSH STEP AND

RECORD previous standard count on Data Sheet 1.

5.2.8.12 PUSH STEP AND

PUSH ENTER until “READY” is displayed on screen.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

14 of 34

5.3 Obtain Soil Moisture Readings with MI HP-2 or MI HP-3 Hydroprobe

WARNING Attempting to catch and regain control of the probe and cable if it

slips free during removal or insertion into the borehole/drywell could

result in personnel injury.

5.3.1 ENSURE neutron survey instrument is positioned within 3 feet of well to

monitor for changes in neutron levels.

5.3.1.1 HPT MONITOR for changes in radiological conditions.

NOTE - Readings are generally taken from engineering established depth up to required

depth of well casing.

5.3.2 POSITION Hydroprobe unit over well in ground AND

SLOWLY LOWER cable until probe is at lowest depth indicated on

Data Sheet 1 not getting any closer than 15 cm (6in) above the soil surface.

5.3.3 PUSH LOG.

5.3.4 WHEN screen displays a number, ENTER borehole/drywell ID number you

are logging AND

PUSH ENTER twice.

5.3.5 IF incorrect number is entered for borehole/drywell being logged,

PERFORM the following:

5.3.5.1 PUSH CLEAR until screen displays “READY”.

5.3.5.2 PUSH LOG.

5.3.5.3 ENTER correct well number AND

PUSH ENTER.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

15 of 34

5.3 Obtain Soil Moisture Readings with MI HP-2 or MI HP-3 Hydroprobe

(Cont.)

NOTE - XX = the deepest depth you entered in Step 5.2.6.

5.3.6 WHEN screen displays “TAKE XX”, PUSH START.

NOTE - Logging takes 16 seconds for each depth interval. The read-out unit will beep

once when logging starts and twice when the logging is complete and a number

will be displayed on screen (e.g. M10 3456, where M10 = depth, 3456 =

measurement).

5.3.7 WHEN number is displayed, RECORD CNT value on Data Sheet 1.

5.3.8 IF measurement is last measurement to be taken, GO TO Step 5.3.14.

5.3.9 PUSH ENTER AND

CONFIRM next depth measurement identified on Data Sheet 1 is displayed.

WARNING Attempting to catch and regain control of the probe and cable if it

slips free during removal or insertion into the borehole/drywell could

result in personnel injury.

5.3.10 MOVE cable up to next depth interval identified on Data Sheet 1.

5.3.11 PUSH START.

5.3.12 REPEAT Steps 5.3.7 through 5.3.11 until data is obtained for all depths

listed on Data Sheet 1.

5.3.13 IF a mistake is made during logging, PERFORM the following:

5.3.13.1 PUSH CLEAR.

5.3.13.2 WHEN screen displays “DATA OK?”, PUSH CLEAR to go

back to the READY screen.

5.3.13.3 GO TO Step 5.3.3.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

16 of 34

5.3 Obtain Soil Moisture Readings with MI HP-2 or MI HP-3 Hydroprobe

(Cont.)

5.3.14 WHEN the last measurement is completed, PRESS ENTER AND

PERFORM the following:

5.3.14.1 PUSH CLEAR AND

CONFIRM “DATA OK?” is displayed.

5.3.14.2 PRESS ENTER to go back to READY.

5.3.15 IF a depth is missed while logging,

OR

IF there is doubt all depths have been logged, CHECK data as follows:

5.3.15.1 PUSH STEP to scroll through all depths.

5.3.15.2 IF screen displays “TAKE XX” for any depth, return probe to

appropriate depth in well and PUSH ENTER to log missed

depth.

5.3.15.3 WHEN satisfied all depths are logged, CONFIRM “DATA

OK?” is displayed.

5.3.15.4 GO TO Step 5.3.14.1.

WARNING Attempting to catch and regain control of the probe and cable if it

slips free during removal or insertion into the borehole/drywell could

result in personnel injury.

5.3.16 PUSH ENTER to save data.

5.3.17 RAISE cable until probe is in the shielded position.

5.3.18 INSTALL borehole/drywell cap.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

17 of 34

5.3 Obtain Soil Moisture Readings with MI HP-2 or MI HP-3 Hydroprobe

(Cont.)

5.3.19 IF last borehole/drywell reading has been obtained,

AND/OR

IF end of work day has been reached, PERFORM the following:

5.3.19.1 RETURN to location where Chi was performed.

5.3.19.2 PERFORM Steps 5.2.7.3 through 5.2.8.5.

5.3.19.3 RECORD Chi in Comments Section of Data Sheet 1.

5.3.19.4 PUSH STEP AND

RECORD Standard Count in Comments Section of Data Sheet 1.

5.3.20 REPEAT Section 5.1 through Step 5.3.19 for each borehole/drywell

identified by Process Engineer.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

18 of 34

5.4 Confirm Data in MI HP-2 or MI HP-3 Hydroprobe

5.4.1 RECALL data as follows:

5.4.1.1 PUSH RCL.

5.4.1.2 PUSH STEP until desired borehole/drywell ID is displayed AND

PUSH ENTER.

5.4.1.3 PUSH STEP to review data.

5.4.1.4 PUSH CLEAR to return to borehole/drywell ID.

5.4.1.5 REPEAT Steps 5.4.1.2 through 5.4.1.4 until all desired data has

been reviewed.

5.4.1.6 PUSH CLEAR AND

CONFIRM “READY” is displayed.

5.4.2 GO TO Section 5.6.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

19 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe

CAUTION

Performing steps in this section (5.5) while precipitation is occurring or

when considerable water or snow is present near the probe, could bias

the data obtained.

NOTE- The steps in this section are performed upon request.

Prepare to Obtain Performance Data

5.5.1 ENTER Hydroprobe meter number and neutron source/detector serial

number on Data Sheet 2.

5.5.2 INSTALL a stop on neutron source/detector cable at a distance of between

37.25 inches and 38.25 inches from the top of the amphenol connector on

neutron source/detector.

5.5.3 REMOVE covering over 8 in. ID Soil Standards A, B and C and 6 in. ID

Soil Standards E, F, and G to expose the central hole.

Pre- Performance Verification Check Data

5.5.4 ENSURE meter is attached to neutron probe AND

CONNECT neutron probe cable.

5.5.5 CONFIRM screen displays “READY”.

5.5.5.1 IF a decimal point is displayed in lower left or a bar is displayed

in the left character, ENSURE battery is charged.

5.5.6 PRESS TIME AND

ENSURE screen display indicates “16”.

5.5.6.1 IF display does not indicate “16”, PRESS STEP until display

reads “16” AND

PRESS ENTER.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

20 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.7 PRESS UNITS AND

CONFIRM screen display indicates “MUNT CNT”.

5.5.7.1 IF display does not indicate “MUNT CNT”, PRESS STEP until

display reads “MUNT CNT” AND

PRESS ENTER.

5.5.8 SET format.

5.5.8.1 PRESS FMT.

5.5.8.2 PRESS STEP twice.

5.5.8.3 WHEN “DEPTHS” is displayed on screen, ENTER“6”,

OR

IF unit will not accept “6”, ENTER “06”.

5.5.8.4 PRESS STEP.

5.5.8.5 WHEN screen displays “SET FMT?”, PRESS ENTER twice.

5.5.8.6 CONFIRM screen displays “READY”.

5.5.9 OBTAIN Chi-square value.

5.5.9.1 ENSURE the following:

Neutron Probe shipping/storage container on stable surface

No other radioactive sources are within 30 feet of gauge

No significant source of hydrogen is within 10 feet

All personnel are at least 10 feet away after the readings are

started.

5.5.9.2 POSITION the neutron probe on yellow square on top of

neutron probe shipping/storage container.

5.5.9.3 PRESS STD.

5.5.9.4 PRESS STEP twice and “Chi” will be displayed on the screen.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

21 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.9.5 CONFIRM Chi number is displayed.

5.5.10 OBTAIN a standard count.

5.5.10.1 PRESS STEP.

5.5.10.2 WHEN “NEW STD?” is displayed, PRESS ENTER.

NOTE - Read-out unit will beep every few seconds during standard count.

5.5.10.3 WAIT approximately 4 to 5 minutes for completion of standard

counts.

5.5.10.4 WHEN beeping stops, CONFIRM screen on read-out unit

displays a standard count (e.g., S 7145 or S 11034).

5.5.10.5 PRESS STEP AND

CONFIRM previous standard is displayed (e.g., P XXXX).

5.5.10.6 PRESS STEP AND

CONFIRM Chi number is displayed.

5.5.10.7 CONFIRM Chi number indicates 0.75 to 1.25.

5.5.10.8 IF Chi number is lower than 0.75 or higher than 1.25,

PERFORM Steps 5.5.9.3 through 5.5.10.7 to obtain Chi again.

5.5.10.9 IF Chi number is lower than 0.75 or higher than 1.25 on recount,

CONTACT Process Engineer.

5.5.10.10 RECORD Chi number on Data Sheet 2.

5.5.10.11 WHEN Chi value is obtained and recorded, PRESS STEP AND

RECORD standard count on Data Sheet 2.

5.5.10.12 PUSH STEP AND

RECORD previous standard count on Data Sheet 2.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

22 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.10.13 PUSH STEP AND

THEN PRESS ENTER.

Source in Shield Count Data

5.5.11 POSITION neutron source/detection on the yellow square on top of neutron

probe shipping/storage container.

5.5.12 PRESS LOG.

5.5.13 WHEN screen displays a number, ENTER “1” for the Source in Shield

Count data AND

PRESS ENTER twice.

5.5.14 IF an incorrect number is entered, PERFORM the following:

5.5.14.1 PRESS CLEAR until screen displays “READY”.

5.5.14.2 PRESS LOG.

5.5.14.3 ENTER correct verification data series number AND

PRESS ENTER.

NOTE - XX equals the depth entered in Step 5.5.7.

5.5.15 WHEN screen displays “TAKE XX”, PRESS START.

NOTE - A total of six readings will be taken, Logging takes 16 seconds for each

reading. The read-out unit will beep once when logging starts and twice when

the logging is complete. A number will be displayed on screen (e.g., M3

3456). M3 = depth, 3456 = measurement.

- All sourced of hydrogen should be at least 10 feet away when taking

measurements, including personnel, moisture models, bottled water, truck tires,

etc.

5.5.16 WHEN number is displayed, RECORD CNT value for Log 1on

Data Sheet 2.

5.5.17 IF measurement is the last measurement to be taken, GO TO Step 5.5.22.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

23 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.18 PRESS ENTER AND

CONFIRM next reading identified on Data Sheet 2 is displayed.

5.5.19 PRESS START.

5.5.20 REPEAT Steps 5.5.16 through 5.5.19 until data are obtained for all six

readings AND

RECORD data on Data Sheet 2.

5.5.21 IF a mistake is made during logging, PERFORM the following:

5.5.21.1 PRESS CLEAR.

5.5.21.2 WHEN screen displays “DATA OK?”, PRESS CLEAR to go

back to READY screen.

5.5.21.3 GO TO Step 5.5.14.2.

5.5.22 WHEN last measurement is completed, PRESS ENTER AND

PERFORM the following:

5.5.22.1 PRESS CLEAR AND

CONFIRM “DATA OK?” is displayed.

5.5.22.2 PRESS ENTER to go back to READY.

Source in Soil Calibration Standards Count Data

NOTE - Data Sheet 2 is setup to obtain data in Logs 2, 3, 4, 5, 6 and 7 from soil

standards G, B, E, C, F and A respectively since soil standards are located from

South to North as G, B, E, C, F, and A.

5.5.23 POSITION neutron source/detector unit over center of Soil Standard G, B,

E, C, F, or A AND

SLOWLY LOWER cable until probe is at stop installed in Step 5.5.2.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

24 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

NOTE - Readings must be obtained with the bottom of the detector approximately 36

inches into the Soil Standard, and with the detector centered in the hole.

5.5.24 PRESS LOG.

5.5.25 WHEN screen displays a number, ENTER “2” for the first set of Soil

Standard data AND

PRESS ENTER twice.

5.5.26 IF an incorrect number is entered, PERFORM the following:

5.5.26.1 PRESS CLEAR until screen displays “READY”.

5.5.26.2 PRESS LOG.

5.5.26.3 ENTER correct calibration data series number AND

PRESS ENTER.

NOTE - XX equals the depth entered in Step 5.5.7.

5.5.27 WHEN screen displays “TAKE XX”, PRESS START.

NOTE - Logging takes 16 seconds for each reading. The read-out unit will beep once

when logging starts and twice when the logging is complete. A number will be

displayed on screen (e.g., M3 3456, where M3 = depth and, 3456 =

measurement).

5.5.28 WHEN number is displayed, RECORD CNT value for appropriate Log

number on Data Sheet 2.

5.5.29 IF measurement is the last measurement to be taken, GO TO Step 5.5.34.

5.5.30 PRESS ENTER AND

CONFIRM next reading identified on Data Sheet 2 is displayed.

5.5.31 PRESS START.

5.5.32 REPEAT Steps 5.5.28 through 5.5.31 until data are obtained for all six

readings AND

RECORD Soil Standard data on Data Sheet 2.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

25 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.33 IF a mistake is made during logging, PERFORM the following:

5.5.33.1 PRESS CLEAR.

5.5.33.2 WHEN screen displays “DATA OK?”, PRESS CLEAR to go

back to the READY screen.

5.5.33.3 GO TO Step 5.5.26.2.

5.5.34 WHEN the last measurement is completed, PRESS ENTER AND

PERFORM the following:

5.5.34.1 PRESS CLEAR AND

CONFIRM “DATA OK?” is displayed.

5.5.34.2 PRESS ENTER to go back to READY.

WARNING

Attempting to catch and regain control of the probe and cable if it slips

free during removal or insertion into the borehole/drywell could result

in personnel injury.

5.5.35 RAISE cable until probe is in the shielded position.

5.5.36 PRESS ENTER to save data.

5.5.37 REPEAT Steps 5.5.23 through 5.5.36 for remaining Soil Standards AND

ENTER “3” in Step 5.5.25 for Log number 3, “4” in Step 5.5.25 for Log

number 4, “5” in Step 5.5.25 for Log number 5, “6” in Step 5.5.25 for Log

number 6, and “7” in Step 5.5.25 for Log number 7.

5.5.38 REPLACE covers over Soil Standards.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

26 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

Source in Air Count Data

5.5.39 POSITION neutron source/detector in the air at least three feet off the

ground and away from any significant hydrogen source, such as water, fuel,

oil, personnel, etc.

5.5.40 PRESS LOG.

5.5.41 WHEN screen displays a number, ENTER “8” for the source in air data

AND

PRESS ENTER twice.

5.5.42 IF an incorrect number is entered for the verification data series, PERFORM

the following:

5.5.42.1 PRESS CLEAR until screen displays “READY”.

5.5.42.2 PRESS LOG.

5.5.42.3 ENTER correct verification data series number AND

PRESS ENTER.

NOTE - XX equals the depth entered in Step 5.5.8.3.

5.5.43 WHEN screen displays “TAKE XX”, PRESS START.

NOTE - Logging takes 16 seconds for each reading. The read-out unit will beep once

when logging starts and twice when the logging is complete. A number will be

displayed on screen (e.g., M3 3456, where M3 = depth and 3456 =

measurement).

5.5.44 WHEN number is displayed, RECORD the CNT value for Log number “8”

on Data Sheet 2.

5.5.45 IF measurement is the last measurement to be taken, GO TO Step 5.5.50.

5.5.46 PRESS ENTER AND

CONFIRM next reading identified on Data Sheet 2 is displayed.

5.5.47 PRESS START.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

27 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.48 REPEAT Step 5.5.44 until data are obtained for all six readings AND

RECORD each on Data Sheet 2 for Source in Air data.

5.5.49 IF a mistake is made during logging, PERFORM the following:

5.5.49.1 PRESS CLEAR.

5.5.49.2 WHEN screen displays “DATA OK?”, PRESS CLEAR to go

back to the READY screen.

5.5.49.3 GO TO Step 5.5.42.

5.5.50 WHEN last measurement is completed, PRESS ENTER AND

PERFORM the following:

5.5.50.1 PRESS CLEAR AND

CONFIRM “DATA OK?” is displayed.

5.5.50.2 PRESS ENTER to go back to READY.

Post- Performance Verification Check Data

5.5.51 OBTAIN Chi-square value.

5.5.51.1 ENSURE the following:

Neutron Probe shipping/storage container on stable surface

No other radioactive sources are within 30 feet of gauge

No significant source of hydrogen is within 10 feet

All personnel are at least 10 feet away after the readings are

started.

5.5.51.2 POSITION the neutron probe on yellow square on top of

neutron probe shipping/storage container.

5.5.51.3 PRESS STD.

5.5.51.4 PRESS STEP twice and “Chi” will be displayed on the screen.

5.5.51.5 CONFIRM Chi number is displayed.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

28 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.52 OBTAIN a standard count.

5.5.52.1 PRESS STEP.

5.5.52.2 WHEN “NEW STD?” is displayed, PRESS ENTER.

NOTE - Read-out unit will beep every few seconds during standard count.

5.5.52.3 WAIT approximately 4 to 5 minutes for completion of standard

counts.

5.5.52.4 WHEN beeping stops, CONFIRM screen on read-out unit

displays a standard count (e.g., S 7145 or S 11034).

5.5.52.5 PRESS STEP AND

CONFIRM previous standard is displayed (e.g., P XXXX).

5.5.52.6 PRESS STEP AND

CONFIRM Chi number is displayed.

5.5.52.7 CONFIRM Chi number indicates 0.75 to 1.25.

5.5.52.8 IF Chi number is lower than 0.75 or higher than 1.25,

PERFORM Steps 5.5.51.3 through 5.5.52.7 to obtain Chi again.

5.5.52.9 IF Chi number is lower than 0.75 or higher than 1.25 on recount,

CONTACT Process Engineer.

5.5.52.10 RECORD Chi number on Data Sheet 2.

5.5.52.11 WHEN Chi value is obtained and recorded, PRESS STEP AND

ENTER standard count on Data Sheet 2.

5.5.52.12 PUSH STEP AND

PUSH STEP AND

RECORD Post- Performance Verification Check Standard

Count, P Value, and Chi Value on Data Sheet 2.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

29 of 34

5.5 Obtain Data for Hydroprobe-Neutron Source Performance Verification

with MI HP-2 or MI HP-3 Hydroprobe (Cont.)

5.5.53 SIGN/DATE Data Sheet 2 AND

GO TO Section 5.6.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

30 of 34

5.6 Disposition

5.6.1 ROUTE applicable Data Sheet(s) to Process Engineer.

5.7 Records

5.7.1 PERFORM the following for records identified within this procedure.

5.7.1.1 RECORD the number of times the record was generated in

applicable column

OR

PLACE a check mark () in the N/A column.

5.7.1.2 SUBMIT the package to SST Leak Detection Engineer or

alternate designated engineer.

Records Submittal Checklist

Number

of times

completed

N/A

()

Data Sheets

Data Sheet 1 - Borehole/Drywell Data

Data Sheet 2 - DR503 Soil Moisture Gauge Performance Verification

Data

SST Leak Detection Engineer or alternate designated engineer SEND the completed records

to the Central Shift Office for records retention.

/ /

Signature Print (First and Last) Date

SST LD Engineer/Alternate

The record custodian identified in the Company-Level Record Inventory and Disposition

Schedule (RIDS) is responsible for record retention in accordance with TFC-BSM-

IRM_DC-C-02.

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

31 of 34

Data Sheet 1 - Borehole/Drywell Data

NOTE - This data sheet may be copied as necessary for each borehole/drywell to be sampled.

This data sheet may be expanded if additional depth is required.

Source/Detector

Number: Meter Number: Tank Number:

Borehole/Drywell # ________

ID__________________

Standard

Count Value _________

Chi Value ______ P Value _________

(Previous Standard Count

Value)

60 ft/#__________ 45 ft/#__________ 30 ft/#__________ 15 ft/#__________

59 ft/#__________ 44 ft/#__________ 29 ft/#__________ 14 ft/#__________

58 ft/#__________ 43 ft/#__________ 28 ft/#__________ 13 ft/#__________

57 ft/#__________ 42 ft/#__________ 27 ft/#__________ 12 ft/#__________

56 ft/#__________ 41 ft/#__________ 26 ft/#__________ 11 ft/#__________

55 ft/#__________ 40 ft/#__________ 25 ft/#__________ 10 ft/#__________

54 ft/#__________ 39 ft/#__________ 24 ft/#__________ 9 ft/#__________

53 ft/#__________ 38 ft/#__________ 23 ft/#__________ 8 ft/#__________

52 ft/#__________ 37 ft/#__________ 22 ft/#__________ 7 ft/#__________

51 ft/#__________ 36 ft/#__________ 21 ft/#__________ 6 ft/#__________

50 ft/#__________ 35 ft/#__________ 20 ft/#__________ 5 ft/#__________

49 ft/#__________ 34 ft/#__________ 19 ft/#__________ 4 ft/#__________

48 ft/#__________ 33 ft/#__________ 18 ft/#__________ 3 ft/#__________

47 ft/#__________ 32 ft/#__________ 17 ft/#__________ 2 ft/#__________

46 ft/#__________ 31 ft/#__________ 16 ft/#__________ 1 ft/#__________

/ /

Signature Print (First and Last) Date

Operator

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

32 of 34

Data Sheet 1 - Borehole/Drywell Data (Cont.)

Source/Detector

Number: Meter Number: Tank Number:

Borehole/Drywell # ________

ID__________________

Standard

Count Value _________

Chi Value ______ P Value _________

(Previous Standard Count

Value)

120 ft/#__________ 105 ft/#__________ 90 ft/#__________ 75 ft/#__________

119 ft/#__________ 104 ft/#__________ 89 ft/#__________ 74 ft/#__________

118 ft/#__________ 103 ft/#__________ 88 ft/#__________ 73 ft/#__________

117 ft/#__________ 102 ft/#__________ 87 ft/#__________ 72 ft/#__________

116 ft/#__________ 101 ft/#__________ 86 ft/#__________ 71 ft/#__________

115 ft/#__________ 100 ft/#__________ 85 ft/#__________ 70 ft/#__________

114 ft/#__________ 99 ft/#__________ 84 ft/#__________ 69 ft/#__________

113 ft/#__________ 98 ft/#__________ 83 ft/#__________ 68 ft/#__________

112 ft/#__________ 97 ft/#__________ 82 ft/#__________ 67 ft/#__________

111 ft/#__________ 96 ft/#__________ 81 ft/#__________ 66 ft/#__________

110 ft/#__________ 95 ft/#__________ 80 ft/#__________ 65 ft/#__________

109 ft/#__________ 94 ft/#__________ 79 ft/#__________ 64 ft/#__________

108 ft/#__________ 93 ft/#__________ 78 ft/#__________ 63 ft/#__________

107 ft/#__________ 92 ft/#__________ 77 ft/#__________ 62 ft/#__________

106 ft/#__________ 91 ft/#__________ 76 ft/#__________ 61 ft/#__________

/ /

Signature Print (First and Last) Date

Operator

Comments:

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

33 of 34

Data Sheet 2 - DR503 Soil Moisture Gauge Performance Verification Data

Hydroprobe Meter Number

Neutron Source Serial Number

Pre-Data Verification Checks

Standard Count Value

P Value (Previous Standard Count Value)

Chi Value

Calibration Data

Log 1

With Source in

Source Shield

Log 2

With Source in 6 in.

Model G Standard

Log 3

With Source in 8 in.

Model B Standard

Log 4

With Source in 6 in.

Model E Standard

Log 5

With Source in 8 in.

Model C Standard

Log 6

With Source in 6 in.

Model F Standard

Log 7

With Source in 8 in.

Model A Standard

Log 8

With Source in

Air

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Time

(sec)

16

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

Count

Number

Total

Counts

6 6 6 6 6 6 6 6

5 5 5 5 5 5 5 5

4 4 4 4 4 4 4 4

3 3 3 3 3 3 3 3

2 2 2 2 2 2 2 2

1 1 1 1 1 1 1 1

Post- Data Verification Checks

Standard Count Value

P Value (Previous Standard Count Value)

Chi Value

Comments/Observations:

Sign/Print (First and Last)/Date:

Operate Model 503DR Ml HP-2 or Ml HP-3

Hydroprobe Neutron Moisture Gauge

Type

REFERENCE Document No.

TO-320-022 Rev/Mod

G-1 Release Date

12/13/2016 Page

34 of 34

Attachment 1 - Display Error Messages

Operating Error Messages

Message Meaning Action

NO STDCT No moisture standard count. Perform Steps 5.2.6 thru 5.2.8.10.

FULL Record log full. Obtain different instrument.

NOT HERE ID entered into RECALL does not exist. Enter correct ID.

EMPTY Record log empty when RECALL pressed. Load record(s) via LOG.

NO DATA Record log empty when PRINT pressed. Obtain data.

SET UNIT Calibration coefficients undefined for CT or RD

unit.

Select CT per Step 5.2.3.

Internal Check (Hardware) Errors

Message Meaning Action

RAM XXXX Non-destructive ram test failure at address xxx.

Contact Process

Engineer. ROM XXXX Program memory checksum error. XXXX is the non-zero sum.

EEP XXXX EEPROM read after write failure at address XXXX.

CK X . XXXX Timer clock test period error 1-X.XXX >0.01 seconds.

MCTR ERR Moisture counter failure.

Program Traps

Message Meaning Action

I DIV C Integer divide by zero.

Contact Process

Engineer DIV BY 0 Real divide by zero.

NEG ROOT Square root of negative number.

PBUS ERR Pcode buss error.

EEP XXXX EEPROM read after write failure at address XXX.

ERR 100 Computer bus error, undefined restart.

ERR 102 Message number too big.

ERR 108 Real display roundup exponent underflow.

ERR 109 Onclear pstack pointer greater than current pstack pointer.

ERR 110 Onclear machine stack pointer greater than current machine stack pointer.

ERR 113 Undefined pcode.

ERR 116 Spurious timeout.

ERR 117 Pcode pointer underflow (machine stack running pcode).

ERR 120 Onclear stack underflow.

ERR 201 Depth index too big.

ERR 203 Record index too big.

ERR 204 Count too big to store in record.

ERR 217 Spurious initialization