One room - all applications Philips EasyDiagnost Eleva Specification Digital radiography and fluoroscopy room with the DRF room solution The EasyDiagnost Eleva is a R/F system designed to perform a wide range of applications from gastro-intestinal, various iodine and vascular procedures to standard radiography. By combining the EasyDiagnost Eleva with a choice of a digital vertical stand and/or a digital detector in the table you can create your own 2-in-1 DRF room solution. Set up a radiography and fluoroscopy room that perfectly fits your needs – from a single detector version to a high-end full digital room solution with all advantages of filmless and even cassetteless workflow.

One room - all applications - Philips

  • Upload
    others

  • View
    12

  • Download
    1

Embed Size (px)

Citation preview

Page 1: One room - all applications - Philips

One room - all applicationsPhilips EasyDiagnost Eleva Specification Digital radiography and fluoroscopy room with the DRF room solution

The EasyDiagnost Eleva is a R/F system designed to perform a wide range of

applications from gastro-intestinal, various iodine and vascular procedures to

standard radiography. By combining the EasyDiagnost Eleva with a choice of a

digital vertical stand and/or a digital detector in the table you can create your

own 2-in-1 DRF room solution. Set up a radiography and fluoroscopy room

that perfectly fits your needs – from a single detector version to a high-end

full digital room solution with all advantages of filmless and even cassetteless

workflow.

Page 2: One room - all applications - Philips
Page 3: One room - all applications - Philips

Content

04 Main components and options

06 Fluoroscopy unit EasyDiagnost Eleva

10 DoseWise

11 X-ray generation

13 Image detection

14 Image display

15 Extended Digital Imaging

19 Ceiling suspension CS

20 Over-table technique

22 Digital vertical stand VS

24 Vertical stand VS advanced

26 Eleva workspot

27 Height adjustable trolley TA-M

29 Image plate reader PCR Eleva S Plus

30 DRF room solution – Configuration proposals

32 Room planning

35 General

Page 4: One room - all applications - Philips

Main components and options*

EasyDiagnost Eleva ✓

Spot-film device (digital camera) ✓

Under-table X-ray tube assembly ✓

Tube lift ❍

Image detection, incl. image intensifier and TV camera ✓

Image display, incl. console Eleva Examination Control ✓

LCD monitor in examination room ✓

Monitor trolley for LCD monitor ✓

Ceiling suspension for one or two LCD monitors ❍

Extended digital imaging, incl. image processing and storage with ViewForum ✓

Angiography (with Bolus Chase) ❍

Footswitch for fluoroscopy and exposure ✓

Accessories ✓/❍

DoseWise ✓/❍

High voltage generator Velara RF ✓

DICOM Worklist Management ❍

DICOM MPPS ❍

DICOM image export ✓

DICOM print ✓

DICOM Media on CD-R ❍

mShield ✓

Bucky tray ❍

EasyLat ❍

Digital in-table detector ❍

� ✓ = standard ❍ = optional

Main components and options

* Please contact your local sales specialist for additional system extensions and accessories

Page 5: One room - all applications - Philips

Main components and options

Ceiling suspension with X-ray tube assembly with control grip and collimator (CS2/4)

Digital vertical stand VSDigital flat detectorMotorized tilting – 20º to +90ºRemovable gridAutomatic exposure control (AEC)� AEC measuring fields

Vertical stand VS advanced tiltCassette trayMotorized tilting – 20º to +90ºRemovable gridAutomatic exposure control (AEC)� AEC measuring fields

Vertical stand VS advancedCassette trayRemovable gridAutomatic exposure control (AEC)� AEC measuring fields

Height adjustable trolley TA-M ❍

Eleva workspot for displaying and processing of digital flat detector images ❍

UNIQUE image processing ✓

Reject analysis ❍

PCR Eleva S Plus ❍

Integration in existing PCR Eleva system ❍

iSite integration ❍

Philips CAD Chest solution xLNA ❍

✓ = standard❍ = optional

Page 6: One room - all applications - Philips

The EasyDiagnost Eleva stand

consists of:

• Table

• Cassette tray, digital detector unit or

EasyLat (option)

• Spot-film device (Digital Camera)

• Compressor

• Tube assembly

• Tube lift (option)

Table

DimensionsTable heightTable top

8�0 mm2,000 mm x �90 mm

(�2.7")(78.7" x 27.2")

Table tiltTilt angle

Speed

90° / –20°90° / –�0°90° / –��°90° / –8�°1°/s to �°/s variable

Table top

MovementsLateralLongitudinal

–100 mm to +90 mm8�0 mm

(–�.9" to +�.�")(��")

Movement speedLateralLongitudinalwith bolus chase acquisition

�2 mm/s�0 mm/s0–1�0 mm/s

(1.7"/s)(2.�"/s)(0–�.9"/s)

X-ray attenuation aquivalent0.7 mm Al typical(at 100 kV, 2.7 mm Al HVL)

Maximum patient weightfor all movementsstatic, in horizontal or verticaltable position (incl. footrest)

180 kg2�0 kg

(�00 lbs)(��0 lbs)

✓✓ = standard❍ = optional

Fluoroscopy unitEasyDiagnost Eleva

Page 7: One room - all applications - Philips

7

2115

(6'

11.

3")

2950

(9'

8.1

")

1558 (5' 1.3")

1775

(5'

9.9

")

1500 (4' 11.1")1075 (3' 6.3")

554

(21.

8")

2399

for

23 c

m I.

I. (7

' 10.

5" fo

r 9"

I.I.)

2390

for

38 c

m I.

I.

(7' 1

0.1"

for

15"

I.I.)

2950

(9'

8.1

")

All dimensions in mm (feet/inches)

750 (29.5")

340 (13.4") 520

(20.5")

1340 (4' 4.8")

620 (24.4")

film

38 cm I.I. (15")

975 (38.4") 1

2000 (6' 6.7")

835 (32.9")

835 (32.9")

690

(27.

2")

100

(3.9

")85

(3

.3")

554

(21.

8")

833

(32.

8")

max

. 210

4 (m

ax. 6

' 10.

8")

626

(24.

6")

155

(6.1

")49

0(1

9.3"

)

512

190

750

(29.

5")

140

1075 (3' 6.3") 885 mm (34.8") – with EasyLat

108.5 108.5 (4.3")

1000 (3' 3.4")

1

2115

(6'

11.

3")

2950

(9'

8.1

")

1558 (5' 1.3")

1775

(5'

9.9

")

1500 (4' 11.1")1075 (3' 6.3")

554

(21.

8")

2399

for

23 c

m I.

I. (7

' 10.

5" fo

r 9"

I.I.)

2390

for

38 c

m I.

I.

(7' 1

0.1"

for

15"

I.I.)

2950

(9'

8.1

")

2115

(6'

11.

3")

2950

(9'

8.1

")

1558 (5' 1.3")

1775

(5'

9.9

")

1500 (4' 11.1")1075 (3' 6.3")

554

(21.

8")

2399

for

23 c

m I.

I. (7

' 10.

5" fo

r 9"

I.I.)

2390

for

38 c

m I.

I.

(7' 1

0.1"

for

15"

I.I.)

2950

(9'

8.1

")

Page 8: One room - all applications - Philips

8

Spot-film device (Digital Camera)

MovementsLateralParkingLongitudinal

220 mm�20 mm7�0 mm

(8.7")(20.�")(29.�")

Gridwithout tube lift option

with tube lift option

�0 l/cmr = 10:1, fo = 800 mm�0 l/cmr = 10:1, fo = 900 mm

Grid parking Automatic movement to parking position ✓

Compressor

Compression cone Integrated in spot-film device ✓

Compression movement �0� mm (12") ✓

Compression stop 10 steps, manually selectable ✓

Compression pressure servo supported ✓

Tube lift (option)

Speed �0 mm/s (2"/s) ✓

Focus – Table top distance �10 mm to ��0 mm (20" to 2�.�") ✓

✓ = standard❍ = optional

Fluoroscopy unitEasyDiagnost Eleva (contd.)

Page 9: One room - all applications - Philips

9

750 (29.5")

340 (13.4") 520

(20.5")

1340 (4' 4.8")

620 (24.4")

film

38 cm I.I. (15")

975 (38.4") 1

2000 (6' 6.7")

835 (32.9")

835 (32.9")

690

(27.

2")

100

(3.9

")85

(3

.3")

554

(21.

8")

833

(32.

8")

max

. 210

4 (m

ax. 6

' 10.

8")

626

(24.

6")

155

(6.1

")49

0(1

9.3"

)

512

190

750

(29.

5")

140

1075 (3' 6.3") 885 mm (34.8") – with EasyLat

108.5 108.5 (4.3")

1000 (3' 3.4")

1

All dimensions in mm (feet/inches)

750 (29.5")

340 (13.4") 520

(20.5")

1340 (4' 4.8")

620 (24.4")

film

38 cm I.I. (15")

975 (38.4") 1

2000 (6' 6.7")

835 (32.9")

835 (32.9")

690

(27.

2")

100

(3.9

")85

(3

.3")

554

(21.

8")

833

(32.

8")

max

. 210

4 (m

ax. 6

' 10.

8")

626

(24.

6")

155

(6.1

")49

0(1

9.3"

)

512

190

750

(29.

5")

140

1075 (3' 6.3") 885 mm (34.8") – with EasyLat

108.5 108.5 (4.3")

1000 (3' 3.4")

1

Page 10: One room - all applications - Philips

10

DoseWise

✓ = standard❍ = optional

Less radiation time

FluoroGrab ✓

Dynamic FluoroGrab ✓

Preset collimation ✓

Collimation on LIH ❍

Automatic parameter setting ✓

High resolution LCD monitor (flicker-free) ✓

More awareness

Dose calculation ✓

Dose measurement ❍

Dose display ✓

Dose reporting ✓

X-ray disable button ✓

X-ray shieldingfixed and flexible lead curtains againstscattered radiation

Smart beam

Pulsed Fluoroscopy (PF), In-pulse controlled ❍

Grid Controlled Fluoroscopy (GCF), In-pulse controlled

Intelligent Exposure (IQX), In-pulse controlled ✓

SpectraBeam RF ✓

Adaptive Measuring Field ✓

Tube lift ❍

Iris shutter and two pairs of parallel shutters ✓

Parkable Grid (in spot-film device) ✓

✓ = standard❍ = optional

Page 11: One room - all applications - Philips

11

X-ray generation

Generator Velara RF

Power pack �0 kW�� kW80 kW

Tube support SRO 2��0SRM 22�0 GSSRO ��100 (for CS)

Exposure techniques • Manual: kV-mAs or kV-mA-s• Automatic Exposure Control (AEC), based on

falling load and fixed current techniques• Intelligent Exposure (IQX), In-Pulse controlled• Automatic kV reduction techniques• Tomo Density Control (TDC)

Fluoroscopy techniques • Automatic kV/mA control during fluorscopy• Pulsed Fluoroscopy (PF), In-pulse controlled• Grid Controlled Fluoroscopy (GCF),

In-pulse controlled

Tube voltage exposureTube voltage fluoroscopy

�0 to 1�0 kV�0 to 110 kV

Tube current exposureTube current fluoroscopy

1 to 1100 mA0.2 to � mA

depending on power pack

mAs product 0.� to �00 mAs0.� to 8�0 mAs

with AECwithout AEC

Exposure times 1 ms to � s with AEC

Access times (tube dependent)Standby to fluoroscopyFluoroscopy to radiographyRadiography to fluoroscopy

< 0.� s< 0.8 s< 0.8 s

Dose managementDose calculationDose measurement

Entrance dose, entrance dose rate, dose-area productDose-area product

Safety Tube overload protectionAutomatic mains voltage compensation

✓ = standard❍ = optional

Page 12: One room - all applications - Philips

12

Tube SRO 2550 SRM 2250 GS SRO 33100 (for CS)

Focal spot 0.� / 1.0 0.� / 1.0 0.� / 1.0

Anode diameter 90 mm 100 mm 90 mm

Anode angle 1�º 1�º 1�º

Anode heat storage capacity �00 kHU (220 kJ) �80 kHU (280 kJ) �80 kHU (280 kJ)

Maximum voltage 1�0 kV 12� kV 1�0 kV

Max. continuous heat dissipation 2�0 W �00 W 2�0 W

Max. heat content of assembly 1700 kHU (12�0 kJ) 20�0 kHU (1�10 kJ) 1700 kHU (12�0 kJ)

Grid switch No Yes No

Collimator for Under-table technique Over-table technique

Collimator controlBuilt-in -controller and motor controllersPower-up self testAuto calibration at power-upError logging, reporting

✓ ✓

SpectraBeam RFProgrammable filter pre-setsMax. number of pre-filters

�✓

Rectangular collimationMotor drivenTwo pairs of parallel shuttersMax. opening �1 cm (20")Positioning accuracy 0.�% of SID

✓ ✓

Circular collimationOctagonal ShapeMotor drivenPositioning accuracy 0.�% of SID

✓ –

Collimation on LIHFor regular shuttersOn last fluoroscopy imageOn last exposure image

–––

X-ray generation (contd.)

Page 13: One room - all applications - Philips

1�

Image Intensifier (I.I.)

Mode2�0 mm, 170 mm, 1�0 mm �10 mm, 2�0 mm, 200 mm, 170 mm�80 mm, �10 mm, 2�0 mm, 170 mm

(9", 7", �.�") or(12", 10", 8", 7") or(1�", 12", 10", 7")

Input window Titanium

Detective Quantum Efficiency ��%

Modulation Transfer Factor ��% at 1 lp/mm in �8 cm (1�'') mode

Central spatial resolution �80 mm (1�'') mode:170 mm (�.7'') mode:

� lp/mm (1�.7 lp/ '')� lp/mm (2�.� lp/ '')

Conversion factor in �8 cm (1�'') mode,typical

�00 cd/m2/(mR/s); ��,8 cd/m2/(µGy/s)

Structure noise Negligible

Contrast ratio170 mm (�.7'') (10% disc): �80 mm (1�'') (10% disc):

��:122:1

Image detection

Camera

Type CCD

Matrix 102� x 102� and �12 x �12

Dynamic range Between �� and 70 dB

Digital output 12 bit

Horizontal / Vertical scan reverse

Yes

Dose control, automatic In-pulse controlled

Automatic Gain Control In-frame gain control

Measurement fields 2 fields configurable

Application range (dose) � to 10 000 nGy/image

Page 14: One room - all applications - Philips

Image display

Monitor

Type LCD (Liquid Crystal Display) ✓

Monochrome display ��0 mm (18'') ✓

Progressive display (flicker free) ✓

Native format 1280 x 102� SXGA ✓

Nominal brightness �00 Cd/m2 ✓

Monitor carriers

Trolley for one LCD monitorfor two LCD monitors

Ceiling suspension for one LCD monitorfor two LCD monitors

✓ = standard❍ = optional

1�

Page 15: One room - all applications - Philips

1�

Extended Digital Imaging

Fluoroscopy

Digital high resolution continuous fluoroscopy ✓

Grid Controlled Fluoroscopy: 0,�-�0 pulses/sec (�0/�0Hz) ❍

Pulsed fluoroscopy, 0,�-� pulses/sec (�0/�0Hz) ❍

FluoroGrab ✓

Dynamic FluoroGrab (up to �0 images/sec) ✓

Fluoro images freeze (last image hold) ✓

Trace subtract fluoroscopy ❍

Dynamic noise reduction ✓

Acquisition

First time right IQ with harmonization and auto windowing ✓

Matrix size 102�2 and �122 ✓

Acquisition speed: up to 8 images/sec (102�2 and �122) ✓

Acquisition speed: up to �0 images/sec (�122) ❍

Adjustable acquisition speed (during run) ✓

Live subtraction ❍

Standard acquisition RAM capacity (0.� GB) ✓

Max. acquisition RAM capacity (1 GB) ❍

External video display ✓

Bolus chase (non subtracted) ❍

✓ = standard❍ = optional

Page 16: One room - all applications - Philips

1�

Image processing

Harmonization ✓

Processing depth (bit) 1�

Video invert ✓

Zoom variable up to x �.7 ✓

Magnifying glass and panning ✓

Manual adjustable contrast/brightness/edge enhancement ✓

Electronic shutters (automatic & manual) ✓

Flip/rotate ✓

Annotation 'free text and advanced' ✓

Measurements: length, angle and area ✓

Vascular image processing

Subtract (single mask) ❍

Run subtract (multiple mask) ❍

Measurements extended: area, histogram, vessel analysis (automatic) ❍

Re-masking ❍

Pixel shift ❍

Split screen pixel shift ❍

Auto warp (automatic pixel shift) ❍

Land marking/road mapping: acquisition ❍

View trace (max. opacification) ❍

CO2 imaging (min. opacification) ❍

✓ = standard❍ = optional

Extended Digital Imaging (contd.)

Page 17: One room - all applications - Philips

17

Viewing

View switch time between patients stored in RAM: 0.1 sec ✓

View switch time between runs stored in RAM (within one patient): 0.1 sec ✓

Dynamic view & run cycle: upt to �0 fps ✓

Overview (user selectable number of images) ✓

Pictorial index ✓

Parallel working ❍

Multimodality viewing (CT/MR/…) in control- and exam room ❍

Direct Mouse Manipulation (DMM) ✓

Default display protocols ✓

User log-on ✓

Image storage

Hard disk ✓

CD-recordable (background storage and retrieval) ❍

Flagging for selective storing, printing and export of images ✓

(Automatic) export of image to network servers (e. g. PACS) ❍

Image reconstruction

ColonMap (complete abdominal overview) ❍

Bolus Chase Reconstruction (complete overview of peripheral vasculature) ❍

✓ = standard❍ = optional

Page 18: One room - all applications - Philips

18

Printing

Direct HCU printing (point to point in background) ❍

DICOM Print SCU ✓

Paper printing ✓

Manual compose print (user dependent layout) ✓

Protocol print (user selects layout from a list) ✓

Auto protocol print ✓

Print protocol editor ❍

Automatic annotation of hardcopies ✓

Networking

DICOM store SCP ✓

DICOM store SCU/SCP ❍

DICOM storage commit SCU ❍

DICOM query/retrieve SCU/ SCP ❍

DICOM worklist management SCU ❍

DICOM MPPS SCU ❍

DICOM presentation state ✓

DICOM grey scale display standard ✓

Multitasking (background communication) ✓

Serviceability❍

Remote service support ✓

Digital image quality measurement tool ✓

✓ = standard❍ = optional

Extended Digital Imaging (contd.)

Page 19: One room - all applications - Philips

19

The ceiling suspension CS2 / CS4

for over-table technique consists of:

• Column

• Rail system

• X-ray tube assembly with collimator

and control grip

Ceiling suspension CS2 / CS4 option

Column �-part Al telescopic column;adaptable to individual room heights

Weight-balanced vertical movement 1,�80 mm (�8.�") ✓

Format-sensing to control the automatic collimator

Tube support SRO 2��0 (see page 1�)SRO ��100 (see page 1�)

Longitudinal travelwith extension rails

�,��0 mm�,0�0 mm

(1�9")(2��")

Transverse movementCS2CS�

1,�7� mm�,19� mm

(�0")(118")

Longitudinal/transverse centering up to 7 lock-in positions can be set

Tube assembly rotationaround horizontal axisaround vertical axis

12�º+1��º / –180º

Laser slit light ✓

Ceiling suspension CS

✓ = standard❍ = optional

brake rail

max

. cei

ling

heig

ht 3

110

(10'

2.4

") * min

. 125

0 (4

' 1.2

")m

in. 1

235

(4' 0

.6")

(w

ithou

t ß-

stop

)

mov

emen

t ra

nge

0...1

480

(4' 1

0.3"

)m

ovem

ent

rang

e 0.

..149

5 (4

' 10.

9")

(with

out

ß-st

op)

max

. 273

0 (8

' 11.

5")

380

(15"

)

* for max. maneuverability

All dimensions

in mm

(feet/inches)

Page 20: One room - all applications - Philips

20

Over-table technique

EasyLat

Type Fold-out cassette holder forhigh quality lateral exposures

Amplimat measuring chamber ✓

Cassette-size sensing ✓

Grid 70 l/cm, r = 1�:1,f0 = 110 cm, stationary

Cassette formatsfold-in

fold-out

1� cm x 18 cm to �� cm x �� cm(�" x 7" to 1�" x 17")�� cm x �� cm to �� cm x �� cm(1�" x 1�" to 1�" x 17")

✓ = standard❍ = optional

900 (35.4")

min

. cei

ling

heig

ht 2

720

(8' 1

1.1"

) *

min

. 860

(33

.9")

mov

emen

t ra

nge

0...1

480

(4' 1

0.3"

)

mov

emen

t ra

nge

0...1

495

(4' 1

0.9"

) (w

ithou

t ß-

stop

)m

in. 8

45 (

33.3

")

(with

out

ß-st

op)

max

. 234

0 (7

' 8.1

")38

0(1

5")

1200 (3' 11.2")

510

(20.

1")

1500 (4' 11.1")2653 (8' 8.5")

3000 (9' 10.1")

2000 (6' 6.7")1322 (4' 4.1") 1678 (5' 6.1")

1875 (6' 1.8")1090 (3' 6.9")

588(23.1")

375(14.8")

72(2.8")

* for max. maneuverability

All dimensions in mm (feet/inches)

Page 21: One room - all applications - Philips

Bucky unit in table

Amplimat measuring chamber ✓

Automatic cassette size sensing ✓

Removable grid �� l/cmr = 12:1 or 8:1,f0 = 100 cm, oscillating(additional grid types are optional available)

Cassette formats 1� cm x 18 cm to �� cm x �� cm(�'' x 7'' to 1�'' x 17'')

✓ = standard❍ = optional

Digital detector unit

Amplimat measuring chamber ✓

Preset collimation ✓

Exam-type dependent collimation presets ✓

Type Flat detector made of amorphous silicon

Detector size �� cm x �� cm (17" x 17")

Image matrix size �,000 pixels x �,000 pixels

Pixel size 1�� m

Image resolution up to �.� Lp/mm

Dimension �9� mm x �7� mm x 18� mm

(2�.�" x 22.�" x 7.2")

Removable grid �� lines/cm, r = 8, f0 = 1�00 mm (�'7.1")

(additional grid types are optional available)

✓ = standard❍ = optional

21

Page 22: One room - all applications - Philips

Digital vertical stand VS

Column

Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit

Vertical movement range �00 mm to 1,800 mm (center beam) (11.8" to �’11")

Installation Floor and wall attachment or floor only (option)

Grips

Two patient grips On left and right of the detector unit

Patient stretch grip (option)

• Arranged on left or right of the detector unit• Rotatable

Brakes

Brakes Electromagnetic, self-locking

22

Detector unit

Type Flat detector made of amorphous silicon

Detector size �� cm x �� cm (17" x 17")

Image matrix size �,000 pixels x �,000 pixels

Pixel size 1�� m

Image resolution up to �.� Lp/mm

Dimension �9� mm x �7� mm x 18� mm (2�.�" x 22.�" x 7.2")

Angle of tilt – 20º to +90º, motorized tilting

Weight approx. 2�.� kg (�8.� lbs)

Removable grid �� l/cm, r = 8, f0 = 1�00 mm (�'7.1")(additional grid types are optionally available)

Grid storage For up to 2 grids within the detector unit

Page 23: One room - all applications - Philips

2�

58 (2.3")

102 (4")

575

(22.

6")

406 (16")710 (28")

825

(32.

5")

1173

(3'

10.

2")

1819

(5'

11.

6")

596 (23.5") 227(8.9")

1990

(6'

6.4

")20

80 (

6' 9

.9")

≤150

0 (4

' 11.

1")

≥300

(11.

8")

≤210

5 (6

' 10.

9")

417 (16.4")

dete

ctor

pla

ne

Buck

y un

it fr

ont

side

596

54 (2.1")

902

(35.

5")

1286 (4' 2.6")

dete

ctor

pla

ne

Buck

y un

itfr

ont

side

596

(23.

5")

54 (2.1")

300(11.8")

902

(35.

5")

with spacer (option)

spacer

64 (2.5")107 (4.2")

1990

(6'

6.4

")20

80 (

6' 9

.9")

986 (38.8")

596

(23.

5")

902

(35.

5")

713 (28.07")

20°

≥300

(11.

8")

≤150

0 (4

' 11.

1")

≤210

5 (6

' 10.

9")

≥550

(21

.7")

dete

ctor

pla

ne

Buck

y un

itfr

ont

side

54 (2.1")

with tilting

with tilting

58 (2.3")

102 (4")

575

(22.

6")

406 (16")710 (28")

825

(32.

5")

1173

(3'

10.

2")

1819

(5'

11.

6")

596 (23.5") 227(8.9")

1990

(6'

6.4

")20

80 (

6' 9

.9")

≤150

0 (4

' 11.

1")

≥300

(11.

8")

≤210

5 (6

' 10.

9")

417 (16.4")

dete

ctor

pla

ne

Buck

y un

it fr

ont

side

596

54 (2.1")

902

(35.

5")

1286 (4' 2.6")

dete

ctor

pla

ne

Buck

y un

itfr

ont

side

596

(23.

5")

54 (2.1")

300(11.8")

902

(35.

5")

with spacer (option)

spacer

The digital vertical stand VS

consists of:

• Column

• Detector unit

• Grid

• 2 user interfaces (left & right)

• 2 patient grips (left & right)

All dimensions

in mm (feet/inches)

Page 24: One room - all applications - Philips

2�

Vertical stand VS advanced

Column

Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit

Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")

Installation Floor and wall attachment or floor only (option)

Grips

Two patient grips On left and right of the detector unit

Patient stretch grip(option)

• Arranged on left or right of the detector unit• Rotatable

Brakes

Brakes Electromagnetic, self-locking

Vertical stand VS advanced

with/without tilt

Detector unit (cassette tray)

Dimensions (w x h) �9� mm to �7� mm (2�.�" to 22.�")

Angle of tilt (option) – 20º to +90º, motorized tilting

Automatic exposure control (AEC)

� AEC measuring fields

Operating 2 control units (left & right) and wireless remote control

Removable grid �� l/cm, r = 8, f0 = 1�00 mm(additional grid types are optionally available)

Grid storage For up to 2 grids within the detector unit

Cassette formats 18 cm x 2� cm to �� cm x �� cm(8" x 10" to 1�" x 17")

Page 25: One room - all applications - Philips

2�

Column

Hardware Counterbalanced rugged column for motorized and manual vertical movement of the detector unit

Vertical movement range �00 mm to 1,800 mm (center beam) (11.82" to �'10.87")

Installation Floor and wall attachment or floor only (option)

Grips

Two patient grips On left and right of the detector unit

Patient stretch grip(option)

• Arranged on left or right of the detector unit• Rotatable

Brakes

Brakes Electromagnetic, self-locking

All dimensions in mm (feet/inches)

58 (2.3") 102 (4")

575

(22.

6")

406 (16")710 (28")

825

(32.

5")

1173

(3'

10.

2")

1819

(5'

11.

6")

596 (23.5") 227 (8.9")

160(6.3")

1990

(6'

6.3

")20

80 (

6' 9

.9")

≤150

0 (4

' 11.

1")

≥300

(11.

8")

≤210

5 (6

' 10.

9")

537 (21.1")

dete

ctor

pla

ne

Buck

y un

it fr

ont

side

596

(23.

5")

54 (2.1") 54 (2.1")

902

(35.

5")

1281 (4' 2.4")

dete

ctor

pla

ne

Buck

y un

itfr

ont

side

596

(23.

5")

300(11.8")

902

(35.

5")

with spacer (option)

spacer

58 (2.3")102 (4")

1990

(6'

6.3

")20

80 (

6' 9

.9")

981 (38.6")

596

(23.

5")

120

(4.7

")

902

(35.

5")

708 (27.9")

20°

≥300

≤150

0 (4

' 11.

1")

≤210

5 (6

' 10.

9")

≥549

(21.

6")

dete

ctor

pla

ne

Buck

y un

itfr

ont

side

54 (2.1")

with tilting (option)

with tilting (option)

The vertical stand VS advanced

consists of:

• Column

• Detector unit (cassette tray)

• Grid

• 2 user interfaces (left & right)

• 2 patient grips (left & right)

Page 26: One room - all applications - Philips

2�

Eleva workspot

Computer

Hard disk 1�0 GB (equivalent to more than �,000 images)

RAM storage capacity 2 GB

Operating frequency 2 GHz

Interfaces • Gigabit Ethernet / DICOM interface• Detector interface

Image data

Data volume Up to 18 MB/image

Matrix depth 1� bit/pixel

Console

Monitor • 17"-TFT display with resolution 1,280 x 1,02�• 19"-TFT display with resolution 1,280 x 1,02� (optional)• Touch screen operation

Keyboard with mouse and function buttons

For entering administrative patient data and for operating the screen menus

Dimensions A B C

Touch monitor 17" ��� mm ��0 mm 2�� mm

(17.1") (1�.9") (9.�")

Touch monitor 19" (optional)

��� mm ��7 mm 2�1 mm

(17.2") (18.0") (9.�")

PC 21� mm ��8 mm 27� mm

(8.�") (17.�") (10.8")

C

C

B

B

A

A

• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •• • • • • •

Page 27: One room - all applications - Philips

27

Height adjustable trolley TA-M

Trolley

Height of table adjustment �00 mm to 870 mm (2�.�" to ��.�")

Dimensions with accessories, max. 2,270 mm x 900 mm (7' �.�" x ��.�")

Total weight 1�0 kg (28� lbs)

Accessories All accessories for the table TH-S can be used with this trolley as well

Brakes At each side

Table top

Type Floating table top of sandwich design with carbon fiber overlay

Dimension table top (l x w)Standard versionBariatric version

2,200 mm x �20 mm2,200 mm x �70 mm

(7' 2.�" x 2�.�")(7' 2.�" x 2�.�")

X-ray transparent areaStandard versionBariatric version

1,7�0 mm x �20 mm1,7�0 mm x �70 mm

(�' 8.1" x 2�.�")(�' 8.1" x 2�.�")

Thickness of table top �� mm (2.1")

Range of horizontal movementstransverselongitudinal

100 mm (�.�")floating by fixing the vertical axes of the wheels

Attenuation equivalent 1.2 mm Al equivalent at 100 kV

Patient load 22� kg (�9� lbs)

Page 28: One room - all applications - Philips

28

2270 (7' 5.4")

620

(24.

4")1)

100

(3.9

")10

0(3

.9")

600

(23.

6")

870

(34.

3")

890

(35"

)

54 (

2.1"

)

1) Bariatric version 670 mm (26.4")

All dimensions in mm (feet/inches)

Height adjustable trolley TA-M (contd.)

Page 29: One room - all applications - Philips

Image plate reader PCR Eleva S Plus

590 (23.3")

810

(31.

9")

43 (

1.7"

)

380 (15.0")

Image plate reader PCR Eleva S Plus

Type The image plate reader PCR Eleva S Plus consists of• Cassette feeding unit for the cassette,• Reader unit.

Plate througput Up to 97 plates/h (18 cm x 2� cm), see table on page 1� (high-speed mode allowing up to 9� plates/h for �� cm x �� cm)

Cassette release �7 s to �� s, depending on cassette size and type

Cassette feeding unit• Removal of exposed image plate from the cassette• Reading of the image plate (cassette) barcode• Transport of the exposed image plate to and from the

reader unit• Loading the cassette with the erased plate• Release of newly loaded cassette

Reader unit:Scanning resolution

Read functionErase function

100 m or 200 m(200 m through either software re-sizing or high-speed mode)�0 m (Option. Not available in the US)12 bit analog/digital conversion of informationExposure-dependent erasure of residual information on the image plate using white light

All dimensions in mm (inches)

29

Page 30: One room - all applications - Philips

DRF room solution – Configuration proposals

Configuration 1

EasyDiagnost Eleva with Bucky tray or EasyLat• Ceiling suspended tube carrier (CS2/�)• Tiltable digital vertical stand VS

Configuration 2

EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)

Configuration 3

EasyDiagnost Eleva with digital detector in table• Ceiling suspended tube carrier (CS2/�)• Tiltable digital vertical stand VS

�0

Page 31: One room - all applications - Philips

Optional available*:

• CR-integration

• Height adjustable trolley TA-M

• Vertical stand VS advanced

with/without tilt

* Please contact your local sales specialist for additional configurations and accessories

�1

Page 32: One room - all applications - Philips

�2

Room planning

12

1150

280 91

31 28

10502040

31005500

240

824

24

14

22

2321

32

29

600

1200

1000

30

5050

9

10

8

34

34

6

7

1

33

25

13

2726

16 3

2

18

20

11

17

19

54

35 EasyDiagnost Eleva

Key

1 DI cabinet

2 Image Receiver Module

� Eleva examination control (EEC)

� Keyboard

� Mouse

� Generator cabinet (Velara)

7 EasyDiagnost Eleva main cabinet

8 DI cabinet rear cover

9 Generator cabinet rear cover

10 Main cabinet rear cover

11 System controller (Sysca) PC

12 Indication box (IBO)

Key (contd.)

1� Acquisition work spot

1� Infra red receiver

1� Slave monitor (18")

1� Monitor Acquisition work spot (18")

17 ViewForum computer

18 Mouse

19 Keyboard

20 Viewing monitor (18") (maybe combined with �WM [shared made])

21 Footswitch

22 Monitor carriage

2� Tilting axis

Key (contd.)

2� Center of table top

2� Speed control

2� Handswitch colon map

27 Exposure handswitch

28 Area for service purposes

29 Space required for installing the X-ray module

�0 Space required for inserting the table top (alternatively: 2.0� m at the head end in 0° position)

�1 If necessary, order an insulating kit for ceiling rails locally

Key (contd.)

�2 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision

�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)

�� Min. space �–7 mm between 2 Cabinet Rear Cover

�� EasyDiagnost Eleva

Layout proposal for EasyDiagnost Eleva with under-table technique (single plane)

All dimensions in mm (feet/inches)

Page 33: One room - all applications - Philips

��

Room planning

�3220�6360

3140*

32

15

54SID max. 1800

835

16

985.5 � 7.5

29

1000 1000

28

35

6

835

31

1935

1500

*

5800 17

18

19

479

11

730

27

2

23

24

22

5

20

21

25

26

14

4

3

13

10

12

33

8

9

1

34

34

36 EasyDiagnost Eleva

Layout proposal for EasyDiagnost Eleva with over-table technique (second plane)

Key

1 DI cabinet

2 Image Receiver Module

� Eleva examination control (EEC)

� Keyboard

� Mouse

� Ceiling rails ceiling suspension

7 Ceiling rails MCS

8 Generator cabinet (Velara)

9 EasyDiagnost Eleva main cabinet

10 DI cabinet rear cover

11 Monitor ceiling suspension

Key (contd.)

12 Generator cabinet rear cover

1� Main cabinet rear cover

1� System controller (Sysco) PC

1� Indication box (IBO)

1� BuckyDiagnost VS advanced tilt

17 Infra red receiver

18 Examination monitor (18")

19 Reference monitor (18")

20 Slave monitor (17")

21 ViewForum computer

22 Mouse

2� Keyboard

Key (contd.)

2� Viewing monitor (18") (maybe combined with �WM [shared made])

2� Speed control

2� Handswitch colom map

27 Exposure handswitch

28 Footswitch

29 Ref. cross point

�0 If necessary, order an insulating kit for ceiling rails locally

�1 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision

Key (contd.)

�2 SID max. 1800 mm with table in upright position

�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)

�� Min. space �–7 mm between 2 Cabinet Rear Cover

�� Tube in position “free exposure”

�� EasyDiagnost Eleva

All dimensions in mm (feet/inches)

Page 34: One room - all applications - Philips

��

Key

1 DI cabinet

2 Image Receiver Module

� Eleva examination control (EEC)

� Keyboard

� Mouse

� Ceiling rails ceiling suspension

7 Ceiling rails MCS

8 CS

9 Generator cabinet (Velara)

10 EasyDiagnost Eleva main cabinet

11 DI cabinet rear cover

Room planning

Layout proposal for DRF room solution with digital vertical stand (minimum room layout without applicational restrictions)

Key (contd.)

12 Monitor ceiling suspension

1� Generator cabinet rear cover

1� Main cabinet rear cover

1� System controller (Sysco) PC

1� Indication box (IBO)

17 Acquisition work spot

18 BuckyDiagnost VS digital with motorized tilting

19 Infra red receiver

20 Examination monitor (18")

21 Reference monitor (18")

22 Monitor Acquisition work spot (18")

Key (contd.)

2� ViewForum computer

2� Mouse

2� Keyboard

2�Viewing monitor (18")(maybe combined with �WM) [shared model])

27 Footswitch

28 Speed control

29 Handswitch colon map

�0 Exposure handswitch

�1 If necessary, order an insulating kit for ceiling rails locally

Key (contd.)

�2 Important!In case the CS rail runs above the EasyDiagnost image intensifier, enlarge the minimum ceiling height by 1�0 mm to prevent collision

�� External rails (optional)

�� Scopomat in work position

�� For cooling requirements it is absolutely necessary to separate E- and B-cabinet from each other!The minimum space between E- and B-cabinet have to be at least 70 mm! The M-cabinet can positioned side by side to the B-cabinet, the E-cabinet or the wall (see drawing)

�� Min. space �–7 mm between 2 Cabinet Rear Cover

�7 EasyDiagnost Eleva

All dimensions in mm (feet/inches)

15

18

8

19 7

12

21

20

29

28

17 22 3

4 5

225

26

24

15

23

31

30

35

36

1 10 9

36

27

32

336

11 1314

� 6100

5800

37 EasyDiagnost Eleva

Page 35: One room - all applications - Philips

��

General

Power supply Digital verticalstand VS

Eleva workspot EasyDiagnost Eleva Eleva Examination Control

Generator 50 kW

Mains voltage 2�0 V ±10% 100 V to 2�0 V ±10% 2�0 V ±10% 2�0 V ±10%

�00 V ±10%; � phase

Mains frequency �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz �0 Hz / �0 Hz

Current input max. 1.7 A max. � A max. 7.� A max. � A max. 1�� A

Fuse �A 10 A 10 A 10 A � x �� A

Heat dissipation (average)

100 W 12� W �00 W - max. 1�00 W �0 W �00 W

Safety Tube overload protection

Automatic mains voltage compensation

Ambient conditions

Working temperature +18˚C to +�0˚C

Storage temperature –10˚C to +��˚C

Rel. humidity (in operation) 20% to 7�% (without condensation)

Rel. humidity (storage) 10% to 9�%

Air pressure 700 hPa to 10�0 hPA

Conformity

The DRF system complies with the provisions of the Medical Device Direktive 9�/�2 EEC (9�) and with IEC, FDA, PAL and CSA standards.

Page 36: One room - all applications - Philips

© 2007 Koninklijke Philips Electronics N.V.All rights are reserved.

Philips Medical Systems DMC GmbH reserves the right to make changes in specifications and/or to discontinue any product at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.

Printed in The Netherlands.��12 9�2 291�1/70� * NOV 2007

Philips Medical Systems is part of

Royal Philips Electronics

Interested?

Would you like to know more about our imaginative

products? Please do not hesitate to contact us.

We would be glad to hear from you.

On the web

www.medical.philips.com

Via email

[email protected]

By fax

+�1 �0 27 �� 887

By mail

Philips Medical Systems

Global Information Center

P.O. Box 128�

��02 BG Eindhoven

The Netherlands

By phone

Asia

Tel: +8�2 2821 �888

Europe, Middle East, Africa

Tel: +�9 70�1 ��� 22��

Latin America

Tel: +�� 11 212� 07��

North America

Tel: +1 800 229 ��17