OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station. Make sure it is safe & orderly. Always keep unqualified/unauthorized personnel away from the test area. Familiarize yourself with safety protocols in the event of a problem. Exercise caution and never touch products or connections during a test. T rain operators. Never touch clips directly and always connect the return lead first. Y ou should always know when a test is being performed. WARNING: THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH ELECTRICAL SAFETY TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING AND USING A TEST INSTRUMENT. CONTACT US AT [email protected] FOR MORE INFO ON HOW TO GET TRAINED ON ELECTRICAL SAFETY TESTING. Quick Start Guide

OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

  • Upload
    others

  • View
    3

  • Download
    0

Embed Size (px)

Citation preview

Page 1: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

OMNIA® IIModels 8204/8254, 8206/8256, 8207/8257

SAFETY CHECKLISTSurvey the test station. Make sure it is safe & orderly.

Always keep unqualifi ed/unauthorized personnel away from the test area.

Familiarize yourself with safety protocols in the event of a problem.

Exercise caution and never touch products or connections during a test.

Train operators. Never touch clips directly and always connect the return lead fi rst.

You should always know when a test is being performed.

WARNING: THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH ELECTRICAL SAFETY TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING AND USING A TEST INSTRUMENT. CONTACT US AT [email protected] FOR MORE INFO ON HOW TO GET TRAINED ON ELECTRICAL SAFETY TESTING.

Quick Start Guide

Page 2: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

INSTRUMENT SETUP

Connect the female end of the standard NEMA style power cord into the input power receptacle on the rear panel of the instrument. Plug the male end of the cord into a grounded power source.

Connect the interlock disable key to the signal input connector on the rear panel of the instrument. This is required to run a test.

Turn the POWER switch to ON.

WARNING: LOCATE A SUITABLE TESTING AREA WITH A CLASS I GROUNDED OUTLET. BE SURE THAT YOUR CLASS I OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST.

SENSE+H.V.

SENSE-

CURRENT RETURN

SIGNAL INPUTSIGNAL OUTPUT

REMOTE OUTPUT

277V~MAX. 16A MAX.

EXTERNAL CASE CND PROBE HI PROBE LOLL NN

SCANNER 1 SCANNER 2

RS232

T10A 250V

GND

CAT. II

115V~ /230V~10A MAX., 50Hz/60HZ

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV16A AX. MAX.

OUTPUT

REMOTE INTERFACE

INPUT

DUT OUTPUTSDUT POWER I/P MD CIRCUITS

SENSE+H.V.

SENSE-

CURRENT RETURN

SIGNAL INPUTSIGNAL OUTPUT

REMOTE OUTPUT

277V~MAX. 16A MAX.

EXTERNAL CASE CND PROBE HI PROBE LOLL NN

SCANNER 1 SCANNER 2

RS232

T10A 250V

GND

CAT. II

115V~ /230V~10A MAX., 50Hz/60HZ

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV16A AX. MAX.

OUTPUT

REMOTE INTERFACE

INPUT

DUT OUTPUTSDUT POWER I/P MD CIRCUITS

TESTRESET

POWER

ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES

MyMenu

1 2 ABC DEF

GHI JKL MNO

PQRS TUV WXYZ

3

4 5 6

7

EXIT ENTER

SCANNER STATUS

H.V.

GND

HI

LO

CH 1 2 3 4 5 6 7 8

ON8 9

• 0 < - -

TESTRESET

POWER

CAUTIONHIGH VOLTAGE

5KV MAX.

H.V.

SENSE+ SENSE-

CURRENT RETURN

MENUSetup System

Setup TestsPerform Tests

VERI-CKEKInformation

OMNIA II 8257

SIGNAL INPUTSIGNAL OUTPUTSCANNER 1 SCANNER 2

REMOTE INTERFACE

8206/8256 8207/8257

8206/8256 8207/8257

To Grounded Power Source

2

1

2

3

Page 3: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

After a test has been set, you may edit the test parameters at any time.

ADD NEW TESTS

EDIT TEST PARAMETERS

TEST SETUP

3

START

Select the Add soft key.From the Main Menu selectthe Setup Tests soft key.

Select the a Test Type.

MyMenu

0001 Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

MyMenu

MyFile Name: 0001 TEST

0001 Dielectric Withstand

Insulation Resistance

AC Ground Bond

DC Continuity

Run Test

Line Leakage

SETUP TESTS

MyMenu

AC - WITHSTANDVoltage

HI-LimitTotalLO-LimitTotal

Ramp Up

Dwell Time

Ramp Down

Range :Voltage 0V-5000V

Frequency

Arc Detect

1240 V

10.00mA

1.0s

5

0.00mA

10.00mA

0.00mA

0.00mA0.1s

0.0s

60Hz

OFF

Arc Sense

HI-LimitRealLO-LimitReal

Offset

Step Name

Defaults

More

Use the arrow keys to choose aparameter to edit.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

EXIT ENTER

MyMenu

AC - WITHSTANDVoltage

HI-LimitTotalLO-LimitTotal

Ramp Up

Dwell Time

Ramp Down

Frequency

Arc Detect

1240 V

10.00mA

1.0s

5

0.00mA

10.00mA

0.00mA

0.00mA0.1s

0.0s

60Hz

OFF

Arc Sense

HI-LimitRealLO-LimitReal

Offset

Step Name

Defaults

More

Enter the value for each parameter by using the alpha/numeric key pad. Select ENTER to save and go to the next parameter.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

AC - WITHSTANDVoltage

HI-LimitTotalLO-LimitTotal

Ramp Up

Dwell Time

Ramp Down

Frequency

Arc Detect

1240 V

10.00mA

1.0s

5

0.00mA

10.00mA

0.00mA

0.00mA0.1s

0.0s

60Hz

OFF

Arc Sense

HI-LimitRealLO-LimitReal

Offset

Step Name

Defaults

More

Select EXIT key followed by the ENTER key to save your selection and return to the Setup Tests screen.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

MENUSetup System

Setup TestsPerform Tests

VERI-CKEKInformation

OMNIA II 8257

Setup System

Perform Tests

Edit

START

Use the arrow keys to choose a test to edit within the Setup Tests screen.

MyMenu

0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ

File Name: 0001

Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

Use the arrow keys to choose a parameter to edit.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

EXIT ENTER

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

EXIT ENTER

Select EXIT key. You will return to the Setup Tests screen.

0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA

Select the Edit soft key.

MyMenu

0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ

File Name: 0001

Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA

MyMenu

0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ

File Name: 0001

Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA

MyMenu

INSULATION RESISTANCEVoltage

HI-Limit

LO-Limit

Ramp Up

Dwell Time

Delay Time

Range :Voltage 0V-6000V

500 V

0.00MΩ

0.5s

0.0s

0.05MΩ

0.000uA

0.1s

1.0s

Ramp Down

Charge-Lo

OFFDUT Output

Step Name

Defaults

Enter the value for each parameter by using the alpha/numeric key pad. Select ENTER to save and go to the next parameter.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

INSULATION RESISTANCEVoltage

HI-Limit

LO-Limit

Ramp Up

Dwell Time

Delay Time

Range :Hi-Limit 0.05MΩ - 50000MΩ 0=OFF

500 V

0.00MΩ

0.5s

0.0s

0.05MΩ

0.000uA

0.1s

1.0s

Ramp Down

Charge-Lo

Step Name

Defaults

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

INSULATION RESISTANCEVoltage

HI-Limit

LO-Limit

Ramp Up

Dwell Time

Delay Time

Range :Hi-Limit 0.05MΩ - 50000MΩ 0=OFF

500 V

0.00MΩ

0.5s

0.0s

0.05MΩ

0.000uA

0.1s

1.0s

Ramp Down

Charge-Lo

Step Name

Defaults

OFFDUT Output

OFFDUT Output

Add

Delete

Page 4: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

Adding multiple tests to the same screen automatically creates a multiple step test sequence.

Note: The test will run in a sequential order unless Single Step control is set to ON.

Test programs can be saved and loaded as fi les for easy recall of test conditions.

4

TEST SETUPSET TEST SEQUENCES

SET TEST SEQUENCES

START

Select the Single Step soft key to toggle ON or OFF.

To perform a Single Step test, select the Perform Tests soft key.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

MENU

Setup System

Setup Tests

Perform Tests

VERI-CKEK

Information

OMNIA II 8257

MyMenu

MENU

Setup System

Setup Tests

Perform Tests

VERI-CKEK

Information

OMNIA II 8257

MyMenu

0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ

File Name: 0001

PERFORM TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA Single Step

Fail Stop

Load Files

Results

OFF

ON

Select the EXIT key. You will return to the Main Menu.

MyMenu

0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ

File Name: 0001

PERFORM TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA Single Step

Fail Stop

Load Files

Results

ON

ON

Setup Tests

VERI-CKEK

PERFORM TESTS

Fail Stop ON

START

From the Main Menu selectthe Setup Tests soft key.

If a test file is not already loaded, load a file by selecting the File soft key.

Use the arrow keys to choose the file.Select the Load soft key to load the file.

MyMenu

MENU

Setup System

Setup Tests

Perform Tests

VERI-CKEK

Information

OMNIA II 8257 MyMenu

MyMenu

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

File Name: 0001 TEST

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

EXIT ENTER

0001 0002 TEST0003 SEAN0004

File Name:New File

Save

Save As

Delete

Load

FILE SETUP

The test file is loaded.

Setup System

Perform Tests

Prompt

Fail Stop ON

Delete

MyMenu

File Name: 0002 TEST

Single Step

Fail Stop

Load Files

Results

OFF

ON

PERFORM TESTS

0002 LLT Settings 10000uA 0.5s 0.0uA

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA

Page 5: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

5

CREATE A NEW TEST FILE

DELETE A TEST STEP

START

From the Main Menu selectthe Setup Tests soft key.

Select the File soft key.

Select the New File soft key.

MyMenu

MENUSetup System

Setup TestsPerform Tests

VERI-CKEKInformation

OMNIA II 8257 MyMenu

MyMenu

0001 Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

This returns you to the Setup Testsmenu with a blank sequence. Thefile name is displayed at the bottom.

You may begin setting up a full test sequence by selecting the Add soft key.

MyMenu

File Name: 0002 TEST

Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

0002 File Name:

New File

Save

Save As

Delete

Load

FILE SETUP

Use the alpha/numeric key pad to give the file a name. Select ENTER to save.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

Use the alpha/numeric key pad to give

MyMenu

CREATE FILECaps Lock

Symbol Key

ON

File Name:

MyMenu

File Name: 0002 TEST

Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

Setup System

Perform TestsPrompt

Fail Stop ON

Edit

Save

0002 TEST

0001

0002 0002

0001

START

From the Main Menu select the Setup Tests soft key.

If a test file is not already loaded, load a file by selecting the File soft key.

Select the Load soft key to load the file. Use the arrows to select the test you would like to delete. Select the delete soft key to delete the Test Step.

MyMenu

MENUSetup System

Setup TestsPerform Tests

VERI-CKEKInformation

OMNIA II 8257 MyMenu

MyMenu

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

File Name: 0001 TEST

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA Add

Edit

Delete

Prompt

File

Fail Stop ON

SETUP TESTS

EXIT ENTER

0001 0002 TEST0003 SEAN0004

File Name:New File

Save

Save As

Delete

Load

FILE SETUP

MyMenu

0001 0002 TEST0003 SEAN0004

File Name:New File

Save

Save As

Delete

Load

FILE SETUP

Setup System

Perform TestsPrompt

Fail Stop ON

Delete Delete

Page 6: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

6

ADAPTER BOX CONNECTIONS (8204/8254)The adapter box (P/N 38482) provides an easy way to connect the OMNIA II to a device under test (DUT). Note: If you are testing a Class I product, safety agencies may require you to perform a Ground Bond test on the ground circuit of the DUT along with the Hipot test.

Using the output ports on the rear panel, plug the black high voltage lead in the High Voltage terminal.

Plug the sense lead into the Sense (+) terminal. Plug the high current test lead into the Current terminal. These two red leads act as the ground connection.

Connect the ground return clip (P/N 38490) to the dead metal on the chassis of the DUT. Plug the sense lead into the Sense (-) terminal. Plug the return lead into the Return terminal. Always check to make sure a good connection is made between the DUT and the return clip.

Plug the power cord of the DUT into the adapter box receptacle.

SENSE+H.V.

SENSE-

CURRENT RETURN

SIGNAL INPUTSIGNAL OUTPUTSCANNER 1 SCANNER 2

RS232

T10A 250V

GND

CAT. II

115V~ /230V~10A MAX., 50Hz/60HZ

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV MAX.

GROUNDTEST

OUTPUT

CH

H.V.TEST

OUTPUT

OUTPUT

REMOTE INTERFACE

INPUT

G-COM. CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8

CA

UTI

ON

HIG

H V

OLT

AG

E5K

VAC

MA

X.6K

VDC

MA

X.

NL

CA

UTI

ON

HIG

H V

OLT

AG

E5K

VAC

MA

X.6K

VDC

MA

X.

ADAPTER BOX CONNECTIONS (8206/8256 & 8207/8257)Note: Use the rear output terminal for connecting the adapter box (P/N 38578) on the 8206/8256 and 8207/8257 models.

Using the DUT output ports on the rear panel, plug the line lead from the adapter box into the L terminal and the neutral lead into the N terminal.

Attach the red ground lead from the adapter box to the GND terminal.

Attach the black case return (P/N CBLHC40-10TL) lead into the case terminal.

Connect the ground return clip to dead metal on the chassis of the DUT.

Plug the line power cord from the DUT into the adapter box receptacle.

REMOTE OUTPUT

277V~MAX. 16A MAX.

EXTERNAL CASE CND PROBE HI PROBE LOLL NN

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV16A AX. MAX.

DUT OUTPUTSDUT POWER I/P MD CIRCUITS

SENSE+H.V.

SENSE-

CURRENT RETURN

SIGNAL INPUTSIGNAL OUTPUT

REMOTE OUTPUT

277V~MAX. 16A MAX.

EXTERNAL CASE CND PROBE HI PROBE LOLL NN

SCANNER 1 SCANNER 2

RS232

T10A 250V

GND

CAT. II

115V~ /230V~10A MAX., 50Hz/60HZ

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV MAX.

CAUTIONHIGH VOLTAGE

5KV16A AX. MAX.

OUTPUT

REMOTE INTERFACE

INPUT

DUT OUTPUTSDUT POWER I/P MD CIRCUITS

CA

UTI

ON

HIG

H V

OLT

AG

E5K

VAC

MA

X.6K

VDC

MA

X.

NL

CA

UTI

ON

HIG

H V

OLT

AG

E5K

VAC

MA

X.6K

VDC

MA

X.

TEST CONNECTIONS

1

2

3

1

2345

Page 7: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

TEST LEAD CONNECTIONSIf you want you can use test leads to connect the OMNIA II to the DUT.

Plug the black return lead (P/N 38490) into the Return terminal. Plug black sense lead into the Sense (-) terminal.

Attach the clip-terminated end of the return lead to the dead metal on the chassis of the DUT.

Connect the high voltage lead (P/N 4040A-08) into the High Voltage terminal on the front panel of the instrument.

Attach the high voltage lead to the current-carrying conductors of the DUT.

Did you know? When performing a Ground Bond test on a Class I product connect the red high current lead (P/N 38489) into the current terminal and red sense lead into the sense (+) terminal. The clip at the other end will connect to the ground pin of the DUT plug.

Did you know? If necessary, a 10 x 20 cm piece of foil can be attached to the enclosure to simulate a full hand contact to the DUT. This foil acts as the return point. The return and sense (-) terminal are attached to the foil. Always check to make sure you made a good connection between the DUT and the return clip.

Class I Product

Class II Product

TESTRESET

POWER

ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES

MyMenu

1 2 ABC DEF

GHI JKL MNO

PQRS TUV WXYZ

3

4 5 6

7

EXIT ENTER

SCANNER STATUS

H.V.

GND

HI

LO

CH 1 2 3 4 5 6 7 8

ON8 9

• 0 < - -

TESTRESET

POWER

CAUTIONHIGH VOLTAGE

5KV MAX.

H.V.

SENSE+ SENSE-

CURRENT RETURN

TESTRESET

POWER

ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES

MyMenu

1 2 ABC DEF

GHI JKL MNO

PQRS TUV WXYZ

3

4 5 6

7

EXIT ENTER

SCANNER STATUS

H.V.

GND

HI

LO

CH 1 2 3 4 5 6 7 8

ON8 9

• 0 < - -

TESTRESET

POWER

CAUTIONHIGH VOLTAGE

5KV MAX.

H.V.

SENSE+ SENSE-

CURRENT RETURN

FOIL

7

TEST CONNECTIONSWARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.

1

2

3

4

Page 8: OMNIA II QSG - Associated Research · OMNIA ® II Models 8204/8254, 8206/8256, 8207/8257 SAFETY CHECKLIST Survey the test station.Make sure it is safe & orderly. Always keep unqualifi

View the results from performed tests. Results include test type, reason for failure, and test parameters.

RESULTS

START

From the Perform Tests screen, use the arrow keys to choose the file to see the test results.

Select the Results soft key.

Results include test type, reason for failure and test parameters.

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

EXIT ENTER

Select the EXIT key to return to thePerform Tests screen

MyMenu

File Name: 0002 TEST

Single Step

Fail Stop

Load Files

Results

OFF

ON

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA

MyMenu

File Name: 0002 TEST

Single Step

Fail Stop

Load Files

Results

OFF

ON

PERFORM TESTS

0002 LLT Settings 10000uA 0.5s 0.0uA

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA

Load Files

MyMenu

File Name: 0002 TEST

RESULTS

0002 LLT Settings 10000uA 0.5s 0.0uA

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA

0002 LLT Settings 10000uA 0.5s 0.0uA

PERFORM TESTS

1 2 ABC 3 DEF

4 5 JKLGHI 6 MNO

7 8 TUVPQRS 9

EXIT ENTER

WXYZ

• 0 — < - -

MyMenu

File Name: 0002 TEST

Single Step

Fail Stop

Load Files

Results

OFF

ON

0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA 0002 LLT Settings 10000uA 0.5s 0.0uA

PERFORM TESTSLLTPASS

264.0179.9

Voltage

Leakage RMSV

uA

179.9I-Maximum

uA

179.6MC Volt RMS

NeutralReverseGround

Probe Ph-Pl

1.0Time

s

mV

key to return to the

For additional information about these and other key features of the OMNIA® II, please consult the full Operation and Service Manual or call us toll-free 1-800-858-TEST (8378) or +1-847-367-4077