OMNIA® IIModels 8204/8254, 8206/8256, 8207/8257
SAFETY CHECKLISTSurvey the test station. Make sure it is safe & orderly.
Always keep unqualifi ed/unauthorized personnel away from the test area.
Familiarize yourself with safety protocols in the event of a problem.
Exercise caution and never touch products or connections during a test.
Train operators. Never touch clips directly and always connect the return lead fi rst.
You should always know when a test is being performed.
WARNING: THIS GUIDE WAS CREATED FOR OPERATORS HAVING SOME FAMILIARITY WITH ELECTRICAL SAFETY TESTING. AN ELECTRICAL SAFETY TESTER PRODUCES VOLTAGES AND CURRENTS THAT CAN CAUSE HARMFUL OR FATAL ELECTRIC SHOCK. TO PREVENT ACCIDENTAL INJURY OR DEATH, THESE SAFETY PROCEDURES MUST BE STRICTLY OBSERVED WHEN HANDLING AND USING A TEST INSTRUMENT. CONTACT US AT [email protected] FOR MORE INFO ON HOW TO GET TRAINED ON ELECTRICAL SAFETY TESTING.
Quick Start Guide
INSTRUMENT SETUP
Connect the female end of the standard NEMA style power cord into the input power receptacle on the rear panel of the instrument. Plug the male end of the cord into a grounded power source.
Connect the interlock disable key to the signal input connector on the rear panel of the instrument. This is required to run a test.
Turn the POWER switch to ON.
WARNING: LOCATE A SUITABLE TESTING AREA WITH A CLASS I GROUNDED OUTLET. BE SURE THAT YOUR CLASS I OUTLET HAS BEEN TESTED FOR PROPER WIRING. READ THE SAFETY CHECKLIST OF THIS GUIDE BEFORE STARTING TO TEST.
SENSE+H.V.
SENSE-
CURRENT RETURN
SIGNAL INPUTSIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL CASE CND PROBE HI PROBE LOLL NN
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~10A MAX., 50Hz/60HZ
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
SENSE+H.V.
SENSE-
CURRENT RETURN
SIGNAL INPUTSIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL CASE CND PROBE HI PROBE LOLL NN
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~10A MAX., 50Hz/60HZ
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES
MyMenu
1 2 ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH 1 2 3 4 5 6 7 8
ON8 9
• 0 < - -
TESTRESET
POWER
CAUTIONHIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE-
CURRENT RETURN
MENUSetup System
Setup TestsPerform Tests
VERI-CKEKInformation
OMNIA II 8257
SIGNAL INPUTSIGNAL OUTPUTSCANNER 1 SCANNER 2
REMOTE INTERFACE
8206/8256 8207/8257
8206/8256 8207/8257
To Grounded Power Source
2
1
2
3
After a test has been set, you may edit the test parameters at any time.
ADD NEW TESTS
EDIT TEST PARAMETERS
TEST SETUP
3
START
Select the Add soft key.From the Main Menu selectthe Setup Tests soft key.
Select the a Test Type.
MyMenu
0001 Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
MyMenu
MyFile Name: 0001 TEST
0001 Dielectric Withstand
Insulation Resistance
AC Ground Bond
DC Continuity
Run Test
Line Leakage
SETUP TESTS
MyMenu
AC - WITHSTANDVoltage
HI-LimitTotalLO-LimitTotal
Ramp Up
Dwell Time
Ramp Down
Range :Voltage 0V-5000V
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA0.1s
0.0s
60Hz
OFF
Arc Sense
HI-LimitRealLO-LimitReal
Offset
Step Name
Defaults
More
Use the arrow keys to choose aparameter to edit.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
EXIT ENTER
MyMenu
AC - WITHSTANDVoltage
HI-LimitTotalLO-LimitTotal
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA0.1s
0.0s
60Hz
OFF
Arc Sense
HI-LimitRealLO-LimitReal
Offset
Step Name
Defaults
More
Enter the value for each parameter by using the alpha/numeric key pad. Select ENTER to save and go to the next parameter.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
AC - WITHSTANDVoltage
HI-LimitTotalLO-LimitTotal
Ramp Up
Dwell Time
Ramp Down
Frequency
Arc Detect
1240 V
10.00mA
1.0s
5
0.00mA
10.00mA
0.00mA
0.00mA0.1s
0.0s
60Hz
OFF
Arc Sense
HI-LimitRealLO-LimitReal
Offset
Step Name
Defaults
More
Select EXIT key followed by the ENTER key to save your selection and return to the Setup Tests screen.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
MENUSetup System
Setup TestsPerform Tests
VERI-CKEKInformation
OMNIA II 8257
Setup System
Perform Tests
Edit
START
Use the arrow keys to choose a test to edit within the Setup Tests screen.
MyMenu
0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
Use the arrow keys to choose a parameter to edit.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
EXIT ENTER
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
EXIT ENTER
Select EXIT key. You will return to the Setup Tests screen.
0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA
Select the Edit soft key.
MyMenu
0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA
MyMenu
0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ
File Name: 0001
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA
MyMenu
INSULATION RESISTANCEVoltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :Voltage 0V-6000V
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
OFFDUT Output
Step Name
Defaults
Enter the value for each parameter by using the alpha/numeric key pad. Select ENTER to save and go to the next parameter.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
INSULATION RESISTANCEVoltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :Hi-Limit 0.05MΩ - 50000MΩ 0=OFF
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
INSULATION RESISTANCEVoltage
HI-Limit
LO-Limit
Ramp Up
Dwell Time
Delay Time
Range :Hi-Limit 0.05MΩ - 50000MΩ 0=OFF
500 V
0.00MΩ
0.5s
0.0s
0.05MΩ
0.000uA
0.1s
1.0s
Ramp Down
Charge-Lo
Step Name
Defaults
OFFDUT Output
OFFDUT Output
Add
Delete
Adding multiple tests to the same screen automatically creates a multiple step test sequence.
Note: The test will run in a sequential order unless Single Step control is set to ON.
Test programs can be saved and loaded as fi les for easy recall of test conditions.
4
TEST SETUPSET TEST SEQUENCES
SET TEST SEQUENCES
START
Select the Single Step soft key to toggle ON or OFF.
To perform a Single Step test, select the Perform Tests soft key.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
MyMenu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257
MyMenu
0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ
File Name: 0001
PERFORM TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA Single Step
Fail Stop
Load Files
Results
OFF
ON
Select the EXIT key. You will return to the Main Menu.
MyMenu
0002 IR Settings 500V 0.00MΩ 0.5s 0.05MΩ
File Name: 0001
PERFORM TESTS0001 ACW Settings 1240V 10.00mA 1.0s 0.000mA Single Step
Fail Stop
Load Files
Results
ON
ON
Setup Tests
VERI-CKEK
PERFORM TESTS
Fail Stop ON
START
From the Main Menu selectthe Setup Tests soft key.
If a test file is not already loaded, load a file by selecting the File soft key.
Use the arrow keys to choose the file.Select the Load soft key to load the file.
MyMenu
MENU
Setup System
Setup Tests
Perform Tests
VERI-CKEK
Information
OMNIA II 8257 MyMenu
MyMenu
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
File Name: 0001 TEST
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
EXIT ENTER
0001 0002 TEST0003 SEAN0004
File Name:New File
Save
Save As
Delete
Load
FILE SETUP
The test file is loaded.
Setup System
Perform Tests
Prompt
Fail Stop ON
Delete
MyMenu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
PERFORM TESTS
0002 LLT Settings 10000uA 0.5s 0.0uA
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA
5
CREATE A NEW TEST FILE
DELETE A TEST STEP
START
From the Main Menu selectthe Setup Tests soft key.
Select the File soft key.
Select the New File soft key.
MyMenu
MENUSetup System
Setup TestsPerform Tests
VERI-CKEKInformation
OMNIA II 8257 MyMenu
MyMenu
0001 Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
This returns you to the Setup Testsmenu with a blank sequence. Thefile name is displayed at the bottom.
You may begin setting up a full test sequence by selecting the Add soft key.
MyMenu
File Name: 0002 TEST
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
0002 File Name:
New File
Save
Save As
Delete
Load
FILE SETUP
Use the alpha/numeric key pad to give the file a name. Select ENTER to save.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
Use the alpha/numeric key pad to give
MyMenu
CREATE FILECaps Lock
Symbol Key
ON
File Name:
MyMenu
File Name: 0002 TEST
Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
Setup System
Perform TestsPrompt
Fail Stop ON
Edit
Save
0002 TEST
0001
0002 0002
0001
START
From the Main Menu select the Setup Tests soft key.
If a test file is not already loaded, load a file by selecting the File soft key.
Select the Load soft key to load the file. Use the arrows to select the test you would like to delete. Select the delete soft key to delete the Test Step.
MyMenu
MENUSetup System
Setup TestsPerform Tests
VERI-CKEKInformation
OMNIA II 8257 MyMenu
MyMenu
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
File Name: 0001 TEST
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA Add
Edit
Delete
Prompt
File
Fail Stop ON
SETUP TESTS
EXIT ENTER
0001 0002 TEST0003 SEAN0004
File Name:New File
Save
Save As
Delete
Load
FILE SETUP
MyMenu
0001 0002 TEST0003 SEAN0004
File Name:New File
Save
Save As
Delete
Load
FILE SETUP
Setup System
Perform TestsPrompt
Fail Stop ON
Delete Delete
6
ADAPTER BOX CONNECTIONS (8204/8254)The adapter box (P/N 38482) provides an easy way to connect the OMNIA II to a device under test (DUT). Note: If you are testing a Class I product, safety agencies may require you to perform a Ground Bond test on the ground circuit of the DUT along with the Hipot test.
Using the output ports on the rear panel, plug the black high voltage lead in the High Voltage terminal.
Plug the sense lead into the Sense (+) terminal. Plug the high current test lead into the Current terminal. These two red leads act as the ground connection.
Connect the ground return clip (P/N 38490) to the dead metal on the chassis of the DUT. Plug the sense lead into the Sense (-) terminal. Plug the return lead into the Return terminal. Always check to make sure a good connection is made between the DUT and the return clip.
Plug the power cord of the DUT into the adapter box receptacle.
SENSE+H.V.
SENSE-
CURRENT RETURN
SIGNAL INPUTSIGNAL OUTPUTSCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~10A MAX., 50Hz/60HZ
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV MAX.
GROUNDTEST
OUTPUT
CH
H.V.TEST
OUTPUT
OUTPUT
REMOTE INTERFACE
INPUT
G-COM. CH1 CH2 CH3 CH4 CH5 CH6 CH7 CH8
CA
UTI
ON
HIG
H V
OLT
AG
E5K
VAC
MA
X.6K
VDC
MA
X.
NL
CA
UTI
ON
HIG
H V
OLT
AG
E5K
VAC
MA
X.6K
VDC
MA
X.
ADAPTER BOX CONNECTIONS (8206/8256 & 8207/8257)Note: Use the rear output terminal for connecting the adapter box (P/N 38578) on the 8206/8256 and 8207/8257 models.
Using the DUT output ports on the rear panel, plug the line lead from the adapter box into the L terminal and the neutral lead into the N terminal.
Attach the red ground lead from the adapter box to the GND terminal.
Attach the black case return (P/N CBLHC40-10TL) lead into the case terminal.
Connect the ground return clip to dead metal on the chassis of the DUT.
Plug the line power cord from the DUT into the adapter box receptacle.
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL CASE CND PROBE HI PROBE LOLL NN
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV16A AX. MAX.
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
SENSE+H.V.
SENSE-
CURRENT RETURN
SIGNAL INPUTSIGNAL OUTPUT
REMOTE OUTPUT
277V~MAX. 16A MAX.
EXTERNAL CASE CND PROBE HI PROBE LOLL NN
SCANNER 1 SCANNER 2
RS232
T10A 250V
GND
CAT. II
115V~ /230V~10A MAX., 50Hz/60HZ
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV MAX.
CAUTIONHIGH VOLTAGE
5KV16A AX. MAX.
OUTPUT
REMOTE INTERFACE
INPUT
DUT OUTPUTSDUT POWER I/P MD CIRCUITS
CA
UTI
ON
HIG
H V
OLT
AG
E5K
VAC
MA
X.6K
VDC
MA
X.
NL
CA
UTI
ON
HIG
H V
OLT
AG
E5K
VAC
MA
X.6K
VDC
MA
X.
TEST CONNECTIONS
1
2
3
1
2345
TEST LEAD CONNECTIONSIf you want you can use test leads to connect the OMNIA II to the DUT.
Plug the black return lead (P/N 38490) into the Return terminal. Plug black sense lead into the Sense (-) terminal.
Attach the clip-terminated end of the return lead to the dead metal on the chassis of the DUT.
Connect the high voltage lead (P/N 4040A-08) into the High Voltage terminal on the front panel of the instrument.
Attach the high voltage lead to the current-carrying conductors of the DUT.
Did you know? When performing a Ground Bond test on a Class I product connect the red high current lead (P/N 38489) into the current terminal and red sense lead into the sense (+) terminal. The clip at the other end will connect to the ground pin of the DUT plug.
Did you know? If necessary, a 10 x 20 cm piece of foil can be attached to the enclosure to simulate a full hand contact to the DUT. This foil acts as the return point. The return and sense (-) terminal are attached to the foil. Always check to make sure you made a good connection between the DUT and the return clip.
Class I Product
Class II Product
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES
MyMenu
1 2 ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH 1 2 3 4 5 6 7 8
ON8 9
• 0 < - -
TESTRESET
POWER
CAUTIONHIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE-
CURRENT RETURN
TESTRESET
POWER
ELECTRICAL SAFETY COMPLIANCE ANALYZEROMNIA® SERIES
MyMenu
1 2 ABC DEF
GHI JKL MNO
PQRS TUV WXYZ
3
4 5 6
7
EXIT ENTER
SCANNER STATUS
H.V.
GND
HI
LO
CH 1 2 3 4 5 6 7 8
ON8 9
• 0 < - -
TESTRESET
POWER
CAUTIONHIGH VOLTAGE
5KV MAX.
H.V.
SENSE+ SENSE-
CURRENT RETURN
FOIL
7
TEST CONNECTIONSWARNING: DO NOT TOUCH THE DEVICE UNDER TEST ONCE YOU START THE TEST.
1
2
3
4
View the results from performed tests. Results include test type, reason for failure, and test parameters.
RESULTS
START
From the Perform Tests screen, use the arrow keys to choose the file to see the test results.
Select the Results soft key.
Results include test type, reason for failure and test parameters.
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
EXIT ENTER
Select the EXIT key to return to thePerform Tests screen
MyMenu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA
MyMenu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
PERFORM TESTS
0002 LLT Settings 10000uA 0.5s 0.0uA
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA
Load Files
MyMenu
File Name: 0002 TEST
RESULTS
0002 LLT Settings 10000uA 0.5s 0.0uA
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA
0002 LLT Settings 10000uA 0.5s 0.0uA
PERFORM TESTS
1 2 ABC 3 DEF
4 5 JKLGHI 6 MNO
7 8 TUVPQRS 9
EXIT ENTER
WXYZ
• 0 — < - -
MyMenu
File Name: 0002 TEST
Single Step
Fail Stop
Load Files
Results
OFF
ON
0001 ACW Settings 1240 V 10.00mA 1.0s 0.000mA 0002 LLT Settings 10000uA 0.5s 0.0uA
PERFORM TESTSLLTPASS
264.0179.9
Voltage
Leakage RMSV
uA
179.9I-Maximum
uA
179.6MC Volt RMS
NeutralReverseGround
Probe Ph-Pl
1.0Time
s
mV
key to return to the
For additional information about these and other key features of the OMNIA® II, please consult the full Operation and Service Manual or call us toll-free 1-800-858-TEST (8378) or +1-847-367-4077
©2019 Associated Research • arisafety.com