37
Kwon, Ji-Young Application Chief Manager Ultrasound Division Canon Medical Systems Korea Associates Course (10:40 – 11:20) June 6 th , 2019

초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Kwon, Ji-YoungApplication Chief Manager

Ultrasound Division

Canon Medical Systems Korea

Associates Course(10:40 – 11:20)

June 6th, 2019

Page 2: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

1. Ultrasound Instruments

2. Image Mode

3. Quality Assurance

Contents

2

Page 3: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

1. Ultrasound Instruments

(1) Main System

(2) Transducers

2. Image Mode

3. Quality Assurance

Contents

3

Page 4: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Advantages & Disadvantage of Ultrasound

4

Main advantages

Noninvasive examination

Possible to observe cross-sectional images in real-time

System itself is compact (Easy to be installed and moved to wards)

Main disadvantages

Need experience for proper operation, and lack reproducibility a little

Page 5: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound Main Body System

5

1) Pulser : produces electrical voltage bursts.

2) Transducer : vibrates to convert electrical voltage into pulse of

ultrasound, and again convert echo into tiny electrical voltage called

“echo signal”.

3) Receiver : amplifies received signal.

4) Memory : stores in memory until complete image is created.

5) Display : displays on monitor in real-time with frame averaging

15~30 fps.

6) Recording Device : film cameras, laser cameras, printer, video

recorders, PACS, etc.

Page 6: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Principle of Real-time Scanning

6

1. Electronic Scanning

- Broad bandwidth electronic arrays

- Electronic focusing

- Conventional transducers

2. Mechanical Scanning

- Contains moving part (crystal or mirror) within transducer

- Fixed focus

- 3D/4D transducers

Page 7: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound Transducers

7

1. Linear Array Transducers

- Linear switched arrays : subgroup firing of elements in sequential

fashion

- Rectangular FOV : all scan lines are parallel to each other

- Mainly used for superficial region : breast, thyroid, MSK, vessel, etc.

Elements

Scanning lines

Page 8: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound Transducers

8

2. Convex Array Transducers

- Curved switched arrays : subgroup firing of elements in sequential

fashion with wide arc

- Fan-shaped FOV : scan lines are not parallel to each other

- Mainly used for abdomen, pelvis, OB, etc.

Elements

Scanning lines

Page 9: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound Transducers

9

3. Phased Array Transducers

- Generally called sector transducer : phasing delays to steer with

firing all elements together

- Pie-shaped FOV : fewer elements and much smaller foot print

- Mainly used for cardiac, TCS, etc.

Elements

Scanning lines

Page 10: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound Transducers

10

4. Matrix Transducers

- Generally called 1.5D arrays transducer

- Electronic focusing in slice thickness plane

- More uniform image than conventional array transducers

Matrix

1.5D Array

Conventional

1D Array

Scan plane

Slice thickness

plane

Page 11: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

1. Ultrasound Instruments

2. Image Mode

(1) B Mode

(2) Color Mode

(3) Doppler Mode

3. Quality Assurance

Contents

11

Page 12: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Principle of Display Modes

12

1. A-mode (Amplitude mode)

- First form of echo display

- Presented as vertical spikes along horizontal baseline

2. B-mode (Brightness mode)

- Generally called 2D mode presented by brightness of dot

3. M-mode (Motion or Time-motion mode)

- Produced by tracking single B-mode scan line orienting it vertically

4. Doppler mode

- Color flow display with mapping or spectral display with waveform

using Doppler shift calculated by Doppler equation

5. 3D/4D mode

- Obtained by series of 2D slices of volume of tissue and

constructing by computer

- Static or real-time displays

Page 13: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

B Mode Optimization

13

1. Overall Gain

- Amplification function of RF signal

- Increases brightness of all echoes in image

Brighter

Darker

Proper gain setting

-> uniform quality

Under gain setting

-> too dark and not clear

Over gain setting

-> too bright on the higher

echogenic part

Page 14: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

B Mode Optimization

14

2. TGC or STC

- Compensation function in depth direction

- Adjusts with same brightness regardless of their location in anatomy

Proper TGC setting Improper TGC setting

Page 15: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

B Mode Optimization

15

3. Depth Range

- Determines PRF and affects frame rate

- Effect of magnification

Shallower display

Deeper display

Normal depth setting Proper depth setting

for GB polyp observation

Depth and FOV adjustment

-> higher frame rate

Page 16: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

B Mode Optimization

16

4. Focus

- A process of reducing beam width

- Improves lateral resolution and increases spatial intensity and sensitivity

Shallow position

-> poor quality except near field

Medium position

-> uniform quality

Deep position

-> improved quality in deep part

Page 17: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

B Mode Optimization

17

5. Frequency

- Trade-off between resolution and penetration

- Improves axial resolution in higher frequency

- Improves penetration in lower frequency

Deep penetration mode

High resolution mode

High frequency

-> good resolution, but less power

Standard frequency

-> medium quality

Low frequency

-> good penetration in far field,

but less resolution in near field

Page 18: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Doppler Mode

18

1. Doppler Effect

- Wave phenomenon in which there is change in frequency and wavelength of wave due to

motion between source of wave and receiver.

2. Doppler Shift Frequency

- Difference(Fd) between received frequency(Fr) and transmitted frequency(Ft)

3. Doppler Equation

Page 19: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Color Doppler Mode

19

1. Color Flow Imaging

- Doppler shifts are detected within “color box”.

- It is possible to know the blood flow direction.

- In general, the blood flow coming close to the probe is displayed in red

color, and the flow going away from the probe is displayed in blue color.

2. Power Doppler Imaging

- Signals colored according to the intensity of Doppler signal are displayed.

- High flow volume is displayed bright and low flow volume dark.

3. High Resolution Doppler Imaging

- Increased sensitivity to flow, so useful for slow flow or small vessels.

- Improves color spatial resolution, so less blooming artifact.

4. Micro Flow Imaging

- Latest color Doppler technique for micro-vascular flow.

- Excellent for low velocity flow with less clutter artifact.

Superb Micro-vascular Imaging Monochrome SMI

Page 20: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Spectral Doppler Mode

20

1. Pulsed Wave Doppler

- Provides most detail about blood flow : not only presence and direction, but additional

information such as blood velocity and detailed characteristics of flow.

- Only Doppler shift from within sample volume is evaluated, giving good range resolution.

2. Continuous Wave Doppler

- Most use of echocardiography.

- Not transmitting pulses, but continuous stream, and no range resolution.

Continuous Wave

Pulsed Wave

Page 21: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Color Mode Optimization

21

1. Color Gain

- Start with higher setting, then reduce gain as appropriate without noise

More color

Less color

Page 22: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Color Mode Optimization

22

2. Scale

- Controls PRF and determines velocity range

- Appropriate without aliasing

Lower velocity range

Higher velocity range

< CCA distal > < ICA proximal > < ICA proximal >

Velocity range : 28.8 cm/s Velocity range : 28.8 cm/s Velocity range : 38.4 cm/s

Page 23: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Color Mode Optimization

23

3. Color Box Steering

- Angle between color box and blood flow

- 10~30 degrees to flow direction

Left side steering Right side steering

CCA distal - Neutral CCA distal - 20° steering to Rt.

ICA prox. - 20° steering to Rt. ICA prox. - 10° steering to Lt.

Page 24: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

PW Mode Optimization

24

1. Doppler Gain

- Same procedure as used for color Doppler

Stronger signal

Weaker signal

Proper gain setting

-> peak velocity is seen well

Under gain setting

-> weak signal of spectrum

Over gain setting

-> too strong signal with

background noise

Page 25: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

PW Mode Optimization

25

2. Baseline

- Optimize display scale for particular flow without aliasing

- Typically baseline is adjusted first, then Doppler scale

Page 26: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

PW Mode Optimization

26

3. Scale

- Set to display blood velocity and accurately determine velocity

Lower velocity range

Higher velocity range

< Vert. A > < Vet. A > < ICA proximal >

Velocity range : 100 cm/s Velocity range : 70 cm/s Velocity range : 120 cm/s

Page 27: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

PW Mode Optimization

27

4. Doppler Cursor Steering

- Angle between pulsed Doppler cursor line and blood flow

- 10~30 degrees to flow direction

- Mostly activates with color box steering, but can control independently

Left side steering Right side steering

10° steering - underestimation 20° steering 30° steering - over estimation

< Same Scale Condition >

Page 28: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

PW Mode Optimization

28

5. Doppler Angle Correction

- To calculate blood velocity accurately

- Try to correct at 30~60 degrees, but not at greater than 60 degree

- No need to correct angle if you want to measure RI value

60° correction 30° correction - underestimation 0° - no correction

< Same Steer Angle >

Page 29: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Pitfalls in Color & PW Doppler Interpretation

29

Common mistake is to assume that lighter shades of color close to Nyquist limit

indicate faster moving blood in Color Doppler.

Lighter shades of color indicate higher Doppler shifts

Color represents Doppler shifts, not blood velocity

= Positive channel

= Negative channel

Page 30: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Pitfalls in Color & PW Doppler Interpretation

30

Second common error is to assume that shifts above baseline are positive in PW

Doppler.

Positive channel Negative channel

Negative channel Positive channel

Page 31: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

1. Ultrasound Instruments

2. Image Mode

3. Quality Assurance

(1) QA Test

(2) Cleaning, Disinfecting, and Sterilization

Contents

31

Page 32: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Quality Assurance

32

What is quality assurance program?

- All procedures necessary to maintain proper and consistent equipment operation

- To keep diagnostic quality of US image

- Good quality assurance will ┌ ensure proper equipment operation

├ detect gradual changes in performance

├ minimize equipment down time

├ reduce number of repeat scans

└ optimize operator and patient safety

Page 33: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound QA Device

33

Tissue-mimicking phantom

ATS-539 (USA)

Page 34: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Ultrasound QA Test

34

Dead Zone : ring-down zone or ring-down space

Axial Resolution : along length of beam

Lateral Resolution : across beam

Gray Scale and dynamic range

Vertical measurement : range accuracy (depth calibration accuracy)

Horizontal measurement : horizontal accuracy (lateral accuracy)

Sensitivity : maximum depth of penetration

Focal Zone : beam Profile (shape of beam)

Functional Resolution : uniformity (definition and fill-in)

Conformance Criterion

Conformance Reference

Reference

Page 35: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Equipment Cleaning, Disinfecting, and Sterilization

35

Cleaning (Cleansing)

- Removal of soil, hence reduction in numbers of microorganisms from surface

- By washing with Soap and water, or wiping with 70% Alcohol (Ethyl alcohol, Enzyme)

- Essential first step prior to either disinfection or sterilization

- Appropriate for transducer used for normal abdominal scanning

Disinfecting

- Inactivation of vegetative bacteria, viruses and fungi, but not bacterial spores

- By Boiling method, Glutaraldehyde (Metricide, Cidex, Widex, Procide)

- Used in contact with intact mucous membranes

- Essential for vaginal transducer

Sterilization

- Complete destruction of microorganisms, including bacterial spores

- By Autoclaving(steam under pressure), Ethylene Oxide(EO gas), Plasma

- Most common approach when used in operating room

- Simple method of maintaining sterility : to cover transducer with sterile sheath

Page 36: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

36

Page 37: 초음파기기의 특성과 관리imedimp4.castcity.net/kacu/20190609/ba3/01.pdf · 2019. 7. 18. · Ultrasound QA Test 34 Dead Zone : ring-down zone or ring-down space Axial Resolution

Thank You for Your Attention