Numerical Assessment of SpecificAbsorption Rate in the Human BodyCaused by NFC DevicesS. Cecil1, G. Schmid1, K. Lamedschwandner1, A. Oberleitner2, M. Bammer2, J. Morak2, G. Schreier2

1Seibersdorf Laboratories, Austria2Austrian Institute of Technology AIT

Page 2: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 2

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Introduction

Page 3: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 3

NFC (Near Field Communication) technologyuses inductive coupling (= magnetic field coupling)

between transmitter and receiver

at 13.56 MHz.

Introduction

The relatively high magnetic field strengthin the immediate surrounding of NFC devices

gives rise to the question about local personal exposure of users.

Page 4: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 4

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 5: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 5

Regulatory framework

The EU R&TTE Directive 1999/5/EC coversall equipment, which uses the radiofrequency spectrum.

e.g. mobile phones, car door openers, …. and of course NFC devices!

The applicable Essential Requirements of the Directive includethe protection of health and the safety of the user and any other person.

Page 6: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 6

Harmonised standards

EU Directives refer toso called Harmonized Standards which specify

exposure limits and compliance evaluation methods.

Harmonized Standards give a presumption of conformity.

For NFC devices the product standard EN 50364 has to be applied.

Page 7: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 7

Exposure assessment

The EN 50364 refersfor general public exposure

to the Basic Restrictions and Reference Levels from the European Council Recommendation 1999/519/EC.

The limit levels of the EC council recommendation are equal to the general public ICNIRP limits.

Page 8: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 8

Limit levels

Basic restrictions* for general public exposure:Local SAR limits are understood as averaged over 10g of contiguoustissue and over a 6 minute time interval

*) Basic restrictions for the general public according to EC council recommendation and ICNIRP recommendation.

For SAR values below the basic restrictions it can be assumed that there are no adverse health effects.

Page 9: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 9

incident EM-wavetransmitted fraction

absorbed fraction(dissipated power)

What is SAR?

temperature elevationdue to absorption

absorbing medium (body)

Physical quantity for tissue heating due to absorption of electromagneticpower:

SAR … Specific Absorption Rate [W/kg]

reflected fraction

ρσ

2ESAR =

σ...conductivity of the lossy medium [S/m]

ρ...mass density [kg/m³]

E...electric field strength inside the tissue [V/m]

Page 10: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 10

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 11: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 11

SAR measurementprocedure uses

a real device (DUTs)operated at its maximum output power.Simplified, homogeneous body models(phantoms)filled with tissue simulating liquid.Isotropic miniature probesmounted to a 3-axes positioning system.

Measurement results are conservativeestimates for SAR values in the real body!SAR cannot be measured in vivo!

SAR-measurement setup, accreditedEMC test laboratory Seibersdorf

Page 12: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 12

Spatial SAR distributions

inside the tissue simulating liquid were measured for

0-0,001 0,001-0,002 0,002-0,003 0,003-0,004 0,004-0,005

0,04-0,05

0,03-0,04

0,02-0,03

0,01-0,02

0-0,01

SAR [W/kg]

SAR [W/kg]

For each device the worst case was considered: device touching the phantom bottomBased on this spatial SAR distribution the 10g-averaged SAR can be computed.

Nokia 6212c: Cell phone with integratedNFC interface

ACR* 122: Stationary NFC smart cardreader

*) Advanced Card Systems Ltd., Hong Kong

Page 13: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 13

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 14: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 14

SAR simulation

Numerical computations use3D full wave simulation platform SEMCAD X based on the Finite Differences in Time Domain (FDTD) method.

MRI-based anatomically correct body models

with approx. 85 different tissues characterised

by their dielectric properties at 13.56 MHz.

Numerical models of the NFC devices.

Human body model „Duke“Simulation model of the NFC mobile

Page 15: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 15

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 16: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 16

Investigated scenarios

4 different scenarios were simulated:NFC mobile in the hand of „Louis“ while connecting a NFC readerNFC mobile on the ear of „Louis“ while phoning and NFC activeNFC reader on the chest of „Duke“ with 20 mm distance to the coilNFC reader on the hip of „Duke“ with 5 mm distance to the coil

Body model „Duke“: male, 34 years, 70 kg, 174 cmBody model „Louis“: male, 14 years, 50 kg, 165 cm

The Virtual Family

Duke

The Virtual Classroom

Louis

Page 17: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 17

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 18: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 18

Results (1)

Scenario „NFC mobile in the hand while connecting a NFC reader“Maximum 10g-averaged SAR: 6.52 mW/kg

Nokia 6212c in the hand of „Louis“ SAR distribution in a cross section through the hand

Page 19: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 19

Results (2)

Scenario: „NFC reader on the chest“Maximum 10g-averaged SAR: 1.52 mW/kg

ACR 122 on the chest of „Duke“ SAR distribution in a cross section of „Duke“

Page 20: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 20

Results (3)

The 13.56 MHz magnetic fieldscause induction of RF currents inside body parts close to a NFC device=> dissipation of electric power in the tissue

The simulation results showedmaximum 10g-averaged SAR-values more than 2 orders of magnitudebelow the limit levels (2 W/kg for head and trunk, 4 W/kg for limbs).

d=20mm

d=5mm

The given SAR values correspond to a continuous field generation. Averaging over a 6 minute time interval will lead to significant lower SAR values because of duty cycles <1 in reality!

Page 21: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 21

Comparison with measurements

The measurement resultsconfirm the conservativity of the SAR measurements because the 10g-averaged SAR values are higher compared to the simulations.

Measurement results are conservative estimates for SAR values in the real body!

Page 22: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 22

IntroductionLimit levels for NFC devicesSAR measurement methodSAR simulation methodInvestigated scenariosResultsConclusion

Content

Page 23: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 23

Experimental and numerical results showed thatmaximum 10g-averaged SAR valuescaused by the considered NFC devices and exposure scenarioswere more than 2 orders of magnitude below the SAR limit levels.

=> Personal exposure due to NFC devices cause SAR levels far below the limits.

=> No evidence that the fields of NFC devices can cause adverse health effects.

Important: Manufacturers have to demonstrate the compliance with personal exposure limits in their conformity assessment procedure for NFC devices(CE marking of compliant equipment)!

Conclusion

Page 24: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 24

Thank you for your attention!

This project was carried out in cooperation with AIT business unit Biomedical Systems.

Page 25: Numerical Assessment of Specific Absorption Rate in the ......Numerical Assessment of Specific Absorption Rate in the Human Body Caused by NFC Devices S. Cecil 1, G. Schmid , K. Lamedschwandner1,

Kurt Lamedschwandner 25

DI Dr. Kurt Lamedschwandner, MBA SEIBERSDORF LABORATORIES

Seibersdorf Labor GmbH, 2444 Seibersdorf, AustriaT +43 (0) 50 550-2805, F +43 (0) 50 550-2813kurt.lamedschwandner@seibersdorf-laboratories.atwww.seibersdorf-laboratories.at