NUI, Galway X-Ray Diffraction Best Practice P. McArdle Feb 2010
1. Saving the X-ray Tube
2. Getting the most from your data
3. Large Unit Cell Parameter Problem (>45Å)
4. Using CheckCIF
1. Saving the X-Ray Tube
Use 50kV/40mA and automatic turn down to 20kV/5mA at the end of data collection
Right click X-Ray Button and select Generator Options
2. Getting the most out of your data.Start a second instance of CrysAlis (An instance with no Diffo. Control Buttons)Click START STOP. Click Load New Experiment.Open your Data Set PAR file.
Click the button beside the Peak Hunting Iconand select Peak hunting with user settings.This will overwrite any peak list
Click Next twice.
Select Follow model changes, Moderate wobbling
Click Finish
Check the Rmerge in Shelxoutput.
Typical SHELXL RUNFloating origin restraints generated
7056 Reflections read, of which 995 rejected
-10 =< h =< 10, -6 =< k =< 10, -13 =< l =< 13, Max. 2-theta = 50.70
0 Systematic absence violations
Inconsistent equivalents etc.
h k l Fo^2 Sigma(Fo^2) N Esd of mean(Fo^2)
6 0 1 51141.65 212.72 3 1812.89 5 0 5 32120.26 232.67 3 1444.75
2 Inconsistent equivalents
2683 Unique reflections, of which 0 suppressed
R(int) = 0.0166 R(sigma) = 0.0250 Friedel opposites not merged
Maximum memory for data reduction = 2400 / 26284
3. Large Unit Cell Parameter Problem (>45Å)
This looks OK spots seem resolved but Detector Dist = 45mm and the c = 45Å and 49% indexation.
Further on in the data set the spots are not resolved.It is necessary to increase the resolution in 3 Dimensions x, y and z.If x and y are the coordinates on the detector then z is normal to the detector.
Solution: - Use a longer Detector Distance to increase xy resolution and a smaller oscillation angle to increase z resolution.At the start of the pre-experiment set Detector dist. to 70mm and click dc edit runs
4. Using CheckCIF
Oscail can combine the CIF from ShelxL with the CIF from the Diffo.
When you have refined a structure using ShelxL you should have the following Commands in the INS fileACTA, SIZE, TEMP (ºC).Use Oscail to Add/Replace the Space Group.
Oscail gives the following Message.
Click the Windows Explorer Button and you can copy the files to your memory stick. The files you need are:Job.INXJob.HKLJob_od.CIFJob.p4p (if you have indexed the faces)
Find the CIF file written by the Diffo. And click OpenThe number of ?s left in the CIF is written to the screen this is usually 6.
You can use the Oscail file checker the ? Button to examine the CIF file.If you follow the defaults a typical output is: ################################################## The SHEL command has been used and the resolution has been limited to 0.8300 Angstroms sin(THETA/LAMBDA) = 0.6024 Acta.Cryst. require at least 0.6 THETA LIMIT = 25.3503 There are AFIX instructions in the INS file ################################################## A CIF file exists _diffrn_measured_fraction_theta_max 0.994 _diffrn_reflns_theta_full 25.35 _diffrn_measured_fraction_theta_full 0.994 >95% expected CIF contains _diffrn_reflns_theta_max 29.0786 This is greater than the SHEL limit################################################## Comments and/or advice SHEL 999 0.83 leaves SIN(theta)/lamda > 0.6 Your Flack is_refine_ls_abs_structure_Flack 0.02(4) If this is OK add the following after Flack in cif_chemical_absolute_configuration; absolute configuration established by anomalous dispersion effects in diffraction measurements on the crystal.;