Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 34 [email protected] www.teseq.com 1 NSG 4070 APPLICATION FOR MIL-STD 461F CS 114 Standards: MIL-STD 461F CS 114 Frequency range: 10 kHz to 200 MHz Curve 1 to 5: see diagrams Modulation: 1 kHz pulse modulation, 50% duty cycle Test method: Substitution method with monitoring probe Monitoring probe: only for information Test parameter Signal generation: NSG 4070-0 Modulator: included in NSG 4070-0 Power meter: 3x included in NSG 4070-0 Power amplifier: max. 30 W required RF Switch: SW 4070 Directional coupler: DCP 0100 Current injection probe: CIP 9136A Monitoring probe: MD 4070 Calibration jig: PCJ 9201B Termination: 50 10 W Attenuation: 1x 20 dB 30 W, 2x 10 dB 30 W, 1x 6 dB 30 W Software: incl. in NSG 4070 or optional C3I or WIN 6000 Equipment RF out Amp in < +10 dBm Amp out ch.1 < +27 dBm ch.2 < +20 dBm ch.3 < +20 dBm Power meter NSG 4070 Power USB Tuning StSize StSize StSize Local Back Stop Run Hold 0 . 1 4 7 8 5 2 3 6 9 Enter Step STO FRQ LVL RCL MOD 2nd Help RF NSG 4070 Power meter ch. 1 stress level ch. 2 forward power ch. 3 reverse power Amp in Amp out 50 Termination NSG 4070 RF output NSG 4070 User port Amp in Amp out Power amplifier inputs Calibration jig MD 4070 Ground plane Interlock Switch (part of SW 4070) Switch (part of SW 4070) 2 A A 1 B B DIRECTIONAL COUPLER Calibration set-up for monitoring probe Remarks: The monitoring probe MD 4070 needs to be calibrated in the way of its use (active, passive or with switching at a specific frequency from active to passive). WARNING: The power meter inputs are very sensitive. It is the user’s responsibility to ensure that the selected test levels does not damage the equipment. Any hardware/setup changes should be calculated before starting the test.

NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 [email protected] 1 NSG 4070

  • Upload
    others

  • View
    4

  • Download
    0

Embed Size (px)

Citation preview

Page 1: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

1

NSG 4070 ApplicAtioN for Mil-StD 461f cS 114

Standards: MIL-STD 461F CS 114Frequency range: 10 kHz to 200 MHzCurve 1 to 5: see diagramsModulation: 1 kHz pulse modulation, 50% duty cycleTest method: Substitution method with monitoring probeMonitoring probe: only for information

Test parameter

Signal generation: NSG 4070-0Modulator: included in NSG 4070-0Power meter: 3x included in NSG 4070-0Power amplifier: max. 30 W requiredRF Switch: SW 4070Directional coupler: DCP 0100Current injection probe: CIP 9136AMonitoring probe: MD 4070 Calibration jig: PCJ 9201BTermination: 50 Ω 10 WAttenuation: 1x 20 dB 30 W, 2x 10 dB 30 W, 1x 6 dB 30 W Software: incl. in NSG 4070 or optional C3I or WIN 6000

Equipment

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig MD 4070 Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Calibration set-up for monitoring probe

Remarks: The monitoring probe MD 4070 needs to be calibrated in the way of its use (active, passive or with switching at a specific frequency from active to passive).

WARNING: The power meter inputs are very sensitive. It is the user’s responsibility to ensure that the selected test levels does not damage the equipment. Any hardware/setup changes should be calculated before starting the test.

Page 2: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #3

Auxiliaryequipment

AEEUT

Equipmentunder test

Insulating

MD 4070

Test level for MIL STD 461F CS114 curve #1

Calibration set-up (for curve #1)

Test set-up with monitoring probe (for curve #1)

37

77 7769

30

40

50

60

70

80

90

100

110

0,01 1 30 200Frequency in MHz

Cur

ren

t in

dB

uA

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: Calibration with a 20 dB increased level.

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: 20 dB, 30 WTermination: 50 Ω 10 W

Remarks: Test with additional 20 dB attenu-ator (attenuator #3) between directional coupler and BCI probe. Use of MD 4070 on PM ch.1: The monitoring probe is in the linear measuring range above 54 dBμA stress level (above 70 kHz). The MD 4070 needs to be used in the active mode.

Page 3: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

3

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #3

Auxiliaryequipment

AEEUT

Equipmentunder test

Insulating

MD 4070

Test level for MIL STD 461F CS114 curve #2

Calibration set-up (for curve #2)

Test set-up with monitoring probe (for curve #2)

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: Calibration with a 16 dB increased level.

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: 16 dB, 30 WTermination: 50 Ω 10 W

Remarks: Test with additional 16 dB attenu-ator (attenuator #3) between directional coupler and BCI probe. Use of MD 4070 on PM ch.1: The monitoring probe is in the linear measuring range above 57 dBμA stress level (above 50 kHz). The MD 4070 needs to be used in the active mode.

43

8375

83

30

40

50

60

70

80

90

100

110

0,01 1 30 200Frequency in MHz

Cur

ren

t in

dB

uA

Page 4: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

4

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #3

Auxiliaryequipment

AEEUT

Equipmentunder test

Insulating

MD 4070

Test level for MIL STD 461F CS114 curve #3

Calibration set-up (for curve #3)

Test set-up with monitoring probe (for curve #3)

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: Calibration with a 10 dB increased level.

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: 10 dB, 30 WTermination: 50 Ω 10 W

Remarks: Test with additional 10 dB attenu-ator (attenuator #3) between directional coupler and BCI probe. Use of MD 4070 on PM ch.1: The monitoring probe is in the linear measuring range above 57 dBμA stress level (above 30 kHz). The MD 4070 needs to be used in the active mode.

49

8981

89

30

40

50

60

70

80

90

100

110

0,01 1 30 200Frequency in MHz

Cur

ren

t in

dB

uA

Page 5: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

5

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #3

Auxiliaryequipment

AEEUT

Equipmentunder test

Insulating

MD 4070

Test level for MIL STD 461F CS114 curve #4

Calibration set-up (for curve #4)

Test set-up with monitoring probe (for curve #4)

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: No additional attenuator required.

Attenuator #1: 10 dB, 30 W Attenuator #2: not in useAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: No additional attenuator required. Use of MD 4070 on PM ch.1: The monitoring probe is in the linear measuring range above 66 dBμA stress level (above 30 kHz). The MD 4070 needs to be used in the active mode from 10 to 300 kHz and in the passive mode above 300 kHz.

57

9789

97

30

40

50

60

70

80

90

100

110

0,01 1 30 200Frequency in MHz

Cur

ren

t in

dB

uA

Page 6: NSG 4070 ApplicAtioN for Mil-StD 461f cS 114 - teseq.de · Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T +49 30 56 59 88 35 F +49 30 56 59 88 34 desales@teseq.com 1 NSG 4070

Teseq GmbH Landsberger Str. 255 12623 Berlin Germany T + 49 30 56 59 88 35 F + 49 30 56 59 88 [email protected] www.teseq.com

6

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

50 Ω Termination

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

Calibration jig BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #2

RF out

Amp in

< +10 dBm

Amp outch.1 < +27 dBm

ch.2 < +20 dBm

ch.3 < +20 dBmPower meter

NSG 4070

PowerUSB

Tuning

StSizeStSize StSize

Local

Back StopRun

Hold

0 .

1

4

7 8

5

2 3

kHzdBm6

9 MHzdBµV

HzV

Enter

Step1

STO

FRQ LVL

RCL

Step2

Step3

MOD

2nd

Help

RFON/OFF

NSG 4070Power meter ch. 1 stress levelch. 2 forward powerch. 3 reverse power

Amp in Amp out

NSG 4070RF output

NSG 4070User port

Amp in Amp out

Power amplifierinputs

BCI probe Ground plane

Interlock

Switch (part of SW 4070)

Switch (part of SW 4070)

2

A

A

1

B

B

DIRECTIONAL COUPLER

Attenuator #1 Attenuator #3

Auxiliaryequipment

AEEUT

Equipmentunder test

Insulating

MD 4070

Test level for MIL STD 461F CS114 curve #5

Calibration set-up (for curve #5)

Test set-up with monitoring probe (for curve #5)

69

109101

109

30

40

50

60

70

80

90

100

110

0,01 1 30 200Frequency in MHz

Cur

ren

t in

dB

uA

Attenuator #1: not in use Attenuator #2: 10 dB, 30 WAttenuator #3: not in useTermination: 50 Ω 10 W

Remarks: Power meter channel 1 needs to be protected with a 10 dB attenu-ator.

Attenuator #1: not in use Attenuator #2: not in useAttenuator #3: not in use Remarks: No additional attenuator required. Use of MD 4070 on PM ch.1: The MD 4070 needs to be used in the active mode in the range 10 kHz to 70 kHz and in the passive mode above 70 kHz.

83-253100 E02 MIL-STD 461F Sep. 2010