Upload
allen-lawrence
View
221
Download
0
Tags:
Embed Size (px)
Citation preview
Novel Instrumentation for future TeV e+e- colliders
Detectors for bunch length measurement
and Beam loss monitoringMayda Velasco, Anne Dabrowski,
Alex CarterNorthwestern University
DOE site visitSeptember 2006
A. Dabrowski, September 05 20061/33
Overview CTF3 Facility
A. Dabrowski, September 05 2006
Drive Beam Injector
Drive Beam Accelerator
PETS Line
30 GHz source Stretcher
Delay Loop TL1 (2006)
CR (2006)
TL2
(2007)
CLEX
(2007)
30 GHz tests
RF photo-injector test
(2006-2007)
Electron beam facility• 1.5microsecond pulse• 150MeV electron LINAC• 3.5 Amp current • Newly commissioned delay loop• Very rich environment for beam instrumentation
development
2/33
Overview Northwestern CTF3 Activities
A. Dabrowski, September 05 2006
Drive Beam Injector
Drive Beam Accelerator
PETS Line
30 GHz source Stretcher
Delay Loop TL1 (2006)
CR (2006)
TL2
(2007)
CLEX
(2007)
30 GHz tests
RF photo-injector test
(2006-2007)
‘Gallery’
Timing & 100MHz ADC’s
0-2kVPower supply
Cables Patch Panel
Beam position monitor
Acceleratingstructure
Quadrupoles
e-
‘Tunnel’
Beam Loss Monitoring
Pickup for bunch length measurement
3/33
Outline
• RF pickup for bunch length measurement
– Principle of the measurement – Advantages of this type of bunch length measurement?– Status of the hardware designed, installed and tested in
2005/2006– Hardware and design improvements still to be installed
before the Fall data taking.– Work planned for the Fall:
• Final setup installed, tested• Analysis performed in normal beam operation
• Beam loss Monitoring
– Reminder devices installed and fully instrumented since 2003
– Ongoing work in optimization of setup
A. Dabrowski, September 05 20064/33
A. Dabrowski, September 05 2006
Principle of the measurement I
The frequency spectrum of the bunch contains all the information about the bunch shape and length.
The field induced in the waveguide has a temporal evolution directly related to that of the electron bunch
Hence it is possible to derive the spatial charge distribution from the frequency spectrum of the field excited in the waveguide
In this measurement, one takes advantage of the power spectrum in the frequency domain. For a given beam current, the larger the power spectrum amplitude, the shorter the bunch length.
The RF-pickup detector measures the spectrum of the electromagnetic field of the bunch collected by a rectangular RF-pickup connected to a waveguide.
5/33
Principle of the measurement II
22
)( e
The smaller the bunch length the larger the signal at a given frequency
The smaller the bunch length the larger measured power
To reach the < ps sensitivity, need to have reach of high frequencies
Solid: σt = 1 ps
Dash: σt = 2 ps
Dash-dot: σt = 3 ps
Pow
er
Sp
ectr
um
[a
.u.]
Pow
er
Sp
ectr
um
[a
.u.]
• Consider an electron bunch of N particles with longitudinal charge distribution Λ(z)• The frequency spectrum is found by Fourier transformation the fields, e.g.
A. Dabrowski, September 05 20066/33
A. Dabrowski, September 05 2006
• Advantages– Non-intercepting / Non destructive– Easy to implement in the beam line– Relatively low cost (compared to streak camera and RF
deflector)– Relatively good time resolution (ns) follow bunch length
within the pulse duration– Measure a single bunch or a train of bunches– Relative calibration within measurements
• Short comings in the calibration– Beam position sensitive– Sensitive to changes in beam current
• If used with the RF deflector and/or a streak camera (method of measuring bunch length at CTF3)– Can provide an excellent cross calibration of device
Advantages of the RF-Pickup
7/33
A. Dabrowski, September 05 2006
• An RF pickup was installed in CTF2– Rectangular waveguide coupled to a rectangular
hole made on the beam pipe surface– Using the mixing technique it measured bunches as
short as 0.7ps. It was limited by a maximum mixing frequency of 90 GHz.
• This device was dismantled in 2002, and is no longer being used at CTF3.
• Goal is to re-install the device with an improved design– Increase maximum frequency reach to max mixing
at 170 GHz, to reach bunch length measurements of 0.3ps
– Design a ceramic/diamond RF window for good vacuum and transmission at high frequency
– Spectral analysis by single shot FFT analysis from a large bandwidth waveform digitizer
RF-pickup device installed in CTF2
C. Martinez et al, CLIC note 2000-020
8/33
A. Dabrowski, September 05 2006
mixing technique in CTF3
Install one WR-28 pickup, with WR-28 waveguide to transport the signal to the gallery
Use a series of filters, and waveguide pieces on the same mounting board in the gallery to separate the signal from the beam in the following frequency ranges:
WR-28 Waveguide components, and mixer at 26.5 GHz
WR-12 Waveguide components and mixer at 56.5 GHz
and at 74.5 GHz
(27 – 39) GHz
(46 - 54) GHz
(75 – 90) GHz
(60 – 72) GHz
WR-6 Waveguide components with mixer at 157 GHz
(143 – 157) GHz
(157 – 170) GHz
Recuperated from CTF2
NEW
NWU9/33
Investment in new hardware
A. Dabrowski, September 05 2006
Components purchased by Northwestern– Local oscillator (Down
converter) • LO 157 GHz • RF 142-177 GHz
– D-band waveguide components (waveguide WR-6 1.65 mm x 0.83 mm, cutoff 110 GHz)
• D-band Horn (gain 20dB)• D-band fixed attenuator (10
dB)• D-band waveguide 5cm Status:
Installed and tested - OK
10/33
Investment in new hardware – Filters
A. Dabrowski, September 05 2006
• Fabrication of Brass filter at 157 GHz– Conical edge at angle of 20 degrees,
thickness 3 x diameter of hole• Used as a high/low pass filter to study
frequency band:– 157 GHz-14 GHz – 157 GHz+14 GHz
• Other filters to be used from CTF2 setup: 26.5 GHz, 40 GHz, 56.5 GHz, 74.5 GHz
Scale (mm)
Status:Installed and tested - OK
11/33
Investment in new hardware – Digitizer
A. Dabrowski, September 05 2006
Acqiris DC282High-Speed 10-bit PXI/CompactPCI® Digitizer•
4 channel
3 GHz bandwidth with 50 Ω high-frequency front end
2 GHz bandwidth with 50 Ω standard front end
2-8 GS/s sampling rate
Auto-install software with application code examples for C/C++, MATLAB, National Instruments LabVIEW etc
Provide a single shot FTT in real timeStatus:Currently being installed
12/33
Installation of Pickup and waveguidesInstallation of Pickup and waveguides
• October 2005 Shutdown
• Pickup with Al203 window (3.35 mm thickness) installed in the CT line
• Signal from BPR0523 split, and ½ taken to the gallery along ~ 15m of WR-28 waveguide
A. Dabrowski, September 05 2006
CT
.BH
D 0
510
CT
.BP
M 0
515
CT
. WC
M 0
525
CT
.VV
S 0
512
CT
.QD
D 0
520
CT
.QF
D 0
530
CT
. PH
M 0
560
CT
.MT
V 0
550
CT
S.M
TV
060
5
CT
.BH
E 0
540
Dum
p
Dump
CT
.DH
D05
05
CD
.DH
L 01
01
CD
.BP
M 0
105
CD
.VP
I 010
4
CT
.BP
R 0
532
13/33
Installed analysis station in GalleryInstalled analysis station in Gallery
A. Dabrowski, September 05 2006
1st mixing @ 26.5 GHz
2nd mixing @ 56.5 GHz
3rd mixing @ 74.5 GHz
Analysis board for
(28 – 90) GHz
14/33
A. Dabrowski, September 05 2006
Examples of Signals from May 2006Examples of Signals from May 2006
Mixer: 56.5 GHzLO: 9.6 GHz
Peaks: 0.1 GHz Beam signal 66 GHz
Equipment
recuperated from
CTF2 installed
correctly &
functions well
Mixer: 26.5 GHzLO: 9.6 GHz
Peaks: 0.1 GHzBeam signal 36 GHz
15/33
Plots Alex Carter
A. Dabrowski, September 05 2006
Signal for analysis to the Oscilloscope
Input signal from the Local Oscillator for the second stage of mixing
Beam signal from WR-28 waveguide
Output of 1st mixing stage at 157 GHz
157 GHz brass filter.
Analysis of
(143 – 170) GHz
(setup in transmission or reflection)
Installed analysis station in GalleryInstalled analysis station in Gallery
16/33
A. Dabrowski, September 05 2006
Analysis of Signals from May 2006Analysis of Signals from May 2006
Mixer: 157 GHzLO: 11.4 GHz
Peaks: 0.4 GHzBeam signal 168 GHz !
New 157 GHz mixer works well!
Showed proof of principle that power measurements at high frequency can be done at > 150 GHz!
Reach sub-pico second bunch length measurement
17/33
Optimization of Window
rf
c
Material Thickness Epsilon
Al203 window 3.35 ± 0.07 mm 9.8
CVD diamond window 0.500 ± 0.005 mm (~6 at 30 GHz measured @ CERN
A. Dabrowski, September 05 2006
@ 90 GHz through Al203 λ is effectively ~ 1 mm
Al203 window not optimized for good transmission designed a thin (0.5mm) diamond window with lower εr. (~6 at 30 GHz measured @ CERN)
•Design prototype completed.
•Machining and brazing test happening now.
plan install window in machine before Fall data taking.
0.5mm
rf
c
18/33
Work needed before Fall data taking
• Complete machining and brazing of new diamond Window.
• Redesign analysis board in gallery:– All 4 mixing stations read out at once with newly
installed large bandwidth waveform digitizer.– Careful attention must be paid to alignment
• Write software:– For new Acquiris data acquisition.– To remotely control the local oscillator
• Write analysis code to use the new system for data taking during the run.
A. Dabrowski, September 05 200619/33
Calibration of device - Fall data taking
A. Dabrowski, September 05 2006
Chicane optics & bunch length measurements - 2004
Magnetic chicane (4 dipoles)
RF Deflector Screen
Betatron phase advance(cavity-profile monitor)
Beta function at cavity and profile monitor
Beam energy
RF deflector phase
RF deflectorwavelength
Deflecting Voltage
Bunch length
y0
y
Deflecting mode TM11
RF deflector off RF deflector on
Slide from Thibaut Lefevre (CERN)
20/33
Calibration of device - Fall data taking
A. Dabrowski, September 05 2006
Bunch Length Measurement with the Streak Camera
OTR@ MTV0550SR@ MTV0361
Two different bunch compression settings in the Frascati Chicane
= 8.9ps = 4.5ps
Maximum Sweep speed @ 10ps/mm
21/33
Calibration of device - Fall data taking
A. Dabrowski, September 05 2006
OTR@ MTV0550SR@ MTV0361
= 8.9ps = 4.5ps
Betatron phase advance(cavity-profile monitor)
Beta function at cavity and profile monitor
Beam energy
RF deflector phase
RF deflectorwavelength
Deflecting Voltage
Bunch length
y0
y
Deflecting mode TM11
Calibration Strategy:
For various settings on the chicane, take bunch length measurements using both the RF deflector, and the RF-pickup. Calibrate the response function of the pickup.
Once calibrated, the pickup can be installed anywhere else in the machine where a bunch length measurement is needed.
The RF-pickup is a much less expensive device than the RF deflector & Streak camera.
Should have better resolution than the Streak camera.22/33
Summary of work on Bunch Length detector
A. Dabrowski, September 05 2006
• An RF-pickup has been installed in the CT line in CTF3• The setup should provide a bunch length measurement
up to 0.3ps• The RF-pickup is installed close to the RF deflector for a
good cross calibration of bunch length.• Proof of principle during May 2006:
– Measuring high frequencies with a WR-28 waveguide pickup was possible!
– The Mixer at 157 GHz was tested and works well.– The filter machined at 157 GHz works well.
• Improved RF diamond window for high frequencies is being machined and brazed, and on schedule to be installed in machine before Fall data taking.
• The new acquiris data digitizing scope is ready for installation, and programming of the DAQ is schedules for the fall.
• The RF-pickup device is flexible, and can be installed in various locations once calibrated
23/33
BLM s installed along the LinacBLM s installed along the Linac
Beam position monitor
Acceleratingstructure
Quadrupoles
e-
Yz
x
Typical Linac section
Design beam optics
Goal : To install the BLM at a position where the beam could be lost
Install 3 chambers and 1 Faraday Cup per Girder which have an accelerating structure
A. Dabrowski, November 24 2004
Chamber installed
24/33
Beam Loss monitoring
A. Dabrowski, September 05 2006
‘Gallery’
Timing & 100MHz ADC’s
0-2kVPower supply
Cables Patch Panel
Beam position monitor
Acceleratingstructure
Quadrupoles
e-
‘Tunnel’
•Full setup since 2003
•3 detectors installed per girder.
•Cross calibration with Faraday cup
•Electronics with 2 gain ranges (x10 difference)
25/33
Beam Loss monitoring
A. Dabrowski, September 05 2006
Goal is to measure secondary shower, and from there determine if possible where the original source of the shower was:
From Geant3.21 simulation [1]
Secondary e+/e- flux distribution at 100cm from the point of the beam loss in the X/Y plane transverse to the beam line
Secondary e+/e- flux versus the longitudinal position. Each curve corresponds to the output signal amplitude of each of the 4 detectors located at different z positions along a LINAC module.
[1] M. Wood, CTF3 note 064, (2004)
26/33
Beam Loss monitoring
A. Dabrowski, September 05 2006
Energy distribution of photons and electrons produced by 0.1% loss of 25MeV electrons in the LINAC coming out at an angle of 30mrad
Signals in the detector will have dominant contributions from both photons and electrons.
Cross sections, at low photon energy < 1 MeV photoelectric effect; > 10 MeV pair production.
Signals has contributions from Photoelectric effect, secondary emission and ionisation27/33
Calibration PlateauCalibration Plateau for chambers filled with He or Ar gas taken in the CTF3 machine, normalized to the Faraday cup
Linear response between Chamber and Faraday cup
Calibration Factor ~ 6 depending on beam loss shower shape
A. Dabrowski, September 05 2006
0 100 200 300 400 5000
2
4
6
8
10
12
He Ar
Am
plitu
de (
a.u.
)
Bias Voltage (V)
4000 5000 6000 7000 8000 9000 10000-24
-16
-8
0
V= 450V V= 400 V V= 300 V V= 200 V V= 100 V V= 0 V
Out
put
Vol
atg
e (m
V)
Time (ns)
Length of Plateau greater for Ar as expected. He breaks down at > 450 V
28/35
Girder 6
Girder 5
Beam loss with nominal beam opticsChamber signals for nominal beam optics have good signal and time response with only low x20 gain low x20 gain amplifieramplifier (and 50 Ω resistor)
Linear response between Chamber and Faraday cup
Calibration Factor ~ 6 depending on beam loss shower shape
System Provides a Mapping of beam loss in Z along girder !
A. Dabrowski, September 05 200629/35
Girder 6
High gain (x200) for the electronics
Girder 5
Very Low lossesVery Low losses on Girder 5 /6
Losses at the limit of the BPM resolution … signals on Girder 6 visible with high gain electronics, but on Girder 5 we are not sensitive to losses BUT perhaps there were no losses?
High gain (x200) for the electronics
A. Dabrowski, September 05 200630/35
Very Low Losses on Girder 6
gain (x200(x201K)) for the electronics
Data taken with very low loses in the beam …
Modified the electronics to increase the sensitivity… 50Ω input impedance to the amplifier replaced by 1kΩ (x20 more)
• still with the amplifier gain of x20 or x200
• Drawback : we loose the time resolution (x20 slower)
Since the Calibration Factor between SEM and Faraday cup X5, and input impedance factor is x20, here we are in the less than ‰ region of beam loss A. Dabrowski, November 24 2004
In this condition… SEM sees the beam loss, but the Faraday Cup in the noise.
31/35
Alternative detector for region of low beam loss
1000 1500 2000 2500 3000
-300
-200
-100
0
100
PMT as a Cherenkov monitorA
mpl
itude
(m
V)
t (ns)
Signal from PMT
Relative calibration with Faraday Cup x200 at 300 V
A. Dabrowski, November 24 2004
In the 1st cavity
Girders 11/12 …
Energy ~ 60-100 MeV•From Simulation, beam loss is more collimated at higher energy …
•A PMT as Cherenkov detector was installed in the 1st Cavity, together with a Faraday Cup.
•Change PMT voltage, depending of sensitivity required
•PMT have relatively low dynamic range (103) within a given voltage gain•At 300V bias voltage, the PMT is x200 more sensitive than the Faraday cup
•Here we estimated the loss to be lower than 10-4 of the beam current
32/33
Summary … Beam Loss on Linac
A. Dabrowski, November 24 2004
Chambers are sensitive to beam loss along the girder …. Additional monitoring complementary to BPM has been used since 2003 In regions of very low loss, Chambers sensitive if one sacrifices resolution, and integrates over large impedance.
Alternate electronics are being considered for the Fall, for a test in Low loss environment. Amplifier installed inside chamber box (discussions with Jim Crisp at Fermilab Beams division lot of experience building electronics for beam current monitors). Benefit from reducing the RC (capacitance dominated by cable increase input impedance and maintain time resolution)
33/33
A. Dabrowski, September 05 2006
Backup Slides
A. Dabrowski, September 05 2006
Why is this measurement needed?
• Optical radiation• Streak camera -------------------- xxxxxxx xxxxxxx > 200fs• Non linear mixing ----------------- xxxxxxx xxxxxxx Laser to RF jitter : 500fs• Shot noise frequency spectrum -- xxxxxxx xxxxxxx Single bunch detector
• Coherent radiation• Interferometry ------------------- xxxxxxx xxxxxxx• Polychromator --------------------- xxxxxxx xxxxxxx
• RF Pick-Up -------------------------------- xxxxxxx xxxxxxx xxxxxxx > 500fs
• RF Deflector ----------------------------- xxxxxxx xxxxxxx xxxxxxx
• RF accelerating phase scan -------------- xxxxxxx xxxxxxx xxxxxxx High charge beam
• Electro Optic Method• Short laser pulse ------------------ xxxxxxx xxxxxxx xxxxxxx Laser to RF jitter : 500fs• Chirped pulse ---------------------- xxxxxxx xxxxxxx xxxxxxx > 70fs
• Laser Wire Scanner ---------------------- xxxxxxx xxxxxxx xxxxxxx Laser to RF jitter : 500fs
1 n! Limitations
Performances of Bunch Length detectors Performances of Bunch Length detectors (table thanks to Thibaut Lefevre, CERN)(table thanks to Thibaut Lefevre, CERN)