Upload
ewald
View
213
Download
0
Embed Size (px)
Citation preview
Noise properties of grating-based x-ray phase contrast computedtomography
Thomas Kohlera)
Philips Technologie GmbH Forschungslaboratorien Rontgenstraße, 24–26 D–22335 Hamburg, Germany
Klaus Jurgen EngelPhilips Technologie GmbH Forschungslaboratorien Weißhausstraße, 2 D–52066 Aachen, Germany
Ewald RoesslPhilips Technologie GmbH Forschungslaboratorien Rontgenstraße, 24–26 D–22335 Hamburg, Germany
(Received 1 September 2010; revised 29 November 2010; accepted for publication 2 December
2010; published 20 July 2011)
Purpose: To investigate the properties of tomographic grating-based phase contrast imaging with
respect to its noise power spectrum and the energy dependence of the achievable contrast to noise
ratio.
Methods: Tomographic simulations of an object with 11 cm diameter constituted of materials of
biological interest were conducted at different energies ranging from 25 to 85 keV by using a wave
propagation approach. Using a Monte Carlo simulation of the x-ray attenuation within the object, it
is verified that the simulated measurement deposits the same dose within the object at each energy.
Results: The noise in reconstructed phase contrast computed tomography images shows a maxi-
mum at low spatial frequencies. The contrast to noise ratio reaches a maximum around 45 keV for
the simulated object. The general dependence of the contrast to noise on the energy appears to be
independent of the material. Compared with reconstructed absorption contrast images, the recon-
structed phase contrast images show sometimes better, sometimes worse, and sometimes similar
contrast to noise, depending on the material and the energy.
Conclusions: Phase contrast images provide additional information to the conventional absorption
contrast images and might thus be useful for medical applications. However, the observed noise
power spectrum in reconstructed phase contrast images implies that the usual trade-off between noise
and resolution is less efficient for phase contrast imaging compared with absorption contrast imag-
ing. Therefore, high-resolution imaging is a strength of phase contrast imaging, but low-resolution
imaging is not. This might hamper the clinical application of the method, in cases where a low spa-
tial resolution is sufficient for diagnosis. VC 2011 American Association of Physicists in Medicine.
[DOI: 10.1118/1.3532396]
Key words: phase contrast tomography, grating-based phase contrast imaging, Talbot interferometer
I. INTRODUCTION
In conventional x-ray imaging, the contrast is generated by
the attenuation of the x-ray beam by the photoelectric effect
and the Compton effect. An alternative way to generate con-
trast is to measure the phase shift that an object imposes on
the “wave front” of an x-ray beam.
Although a number of phase contrast imaging (PCI) tech-
niques exist,1 PCI plays only a minor role in the domain of
medical imaging, mainly because of the stringent require-
ments of known techniques on coherence and monochroma-
ticity of the x-ray beams. These are met by synchrotron
x-ray sources but unmet by conventional x-ray tubes as
almost exclusively used in the medical domain. Recently, a
technique well known in the visible range of wavelengths
has been adopted to the x-ray domain, namely so-called Tal-
bot-Lau interferometry.2–5 This technique permits the use of
conventional tubes and functions with partially coherent
beams generated by gold gratings placed close to the x-ray
tube’s exit window. Among all known techniques of phase-
contrast imaging in the hard (medical) x-ray domain, Talbot-
Lau interferometry requires the smallest coherence volume
for phase-sensitive imaging.4,41
The grating-based setup allows tomographic reconstruc-
tion of the three-dimensional distribution of the refractive
index.6–8 The resulting images of biological samples like a
rat brain6 or a rat heart9 show excellent soft tissue contrast
that would likely add clinical value in diagnostic imaging if
it is applicable to human imaging.
Currently, experimental setups for the grating-based PCI
use x-ray energies in the range from 20 to 60 keV (Ref. 10)
meaning that a wide range of medical applications are
already in reach from an energy perspective.
Noise in phase contrast x-ray imaging is different to that
of absorption contrast imaging as shown, e.g., for the case of
propagation-based PCI.11,12 In fact the noise shows a broad
spatial auto-covariance profile, or, in other words, there is a
significant amount of low-frequency noise.
The purpose of the present work is to investigate the noise
features of tomographic grating-based PCI, viz., the noise
S106 Med. Phys. 38 (7), July 2011 0094-2405/2011/38(7)/S106/11/$30.00 VC 2011 Am. Assoc. Phys. Med. S106
power spectrum and the energy dependence of the contrast
to noise ratio (CNR) that can be achieved.
II. METHOD
This section starts with a brief summary of the main
aspects of grating-based differential phase contrast imaging
(DPCI). An in-depth description can be found in Ref. 13.
The setup for grating-based DPCI is shown in Fig. 1, where
only the most important optical elements required for phase
contrast imaging are shown.3,5 A plane wave of coherent
x-rays hits the sample. Due to refraction, the phase front
behind the object is distorted. The distorted wavefront passes
a beam-splitter grating G1, which creates a characteristic inter-
ference pattern at the location of the analyzer grating G2. The
interference pattern is imaged by measuring the x-ray inten-
sity using a detector D immediately behind G2 at several rela-
tive transverse positions of the gratings G1 and G2.
The pitch g1 of the beam-splitter grating G1 is typically in
the order of a few micrometers, and the interference pattern
at the position of G2 changes on the same length scale. By
analyzing the interference pattern using an absorber grating
G2, the fine changes which the object imposes on the inter-
ference pattern can still be resolved, despite the fact that the
pixel size of the detector D is much larger than the character-
istic length scale of the interference pattern. This is true,
however, only as long as the interference pattern is still peri-
odic with the spatial frequency matching the pitch of G2,
which is the case as long as the gradient of the phase front
just before G1 is constant over the pixel area.
II.A. Simulation
In phase contrast imaging, the interaction of the object
with the x-ray beam is described by wave optics.14 The com-
plex refractive index of the material is
n ¼ 1� dþ ib; (1)
where d describes a phase shift that the object imposes onto
the wavefront and b is related to the linear attenuation coeffi-
cient l by
l ¼ 4pbk
(2)
with k being the wavelength of the x-rays, which is related to
the energy E of the photons by the relation E¼ hc=k where his the Planck constant and c is the speed of light. The propaga-
tion of a monochromatic x-ray beam with wave vector~k ¼ ðkx; ky; kzÞ represented by a complex wave field wx is
mathematically described by the Helmholtz equation
ðr2 þ k2Þwxðx; y; zÞ ¼ 0; k ¼ j~kj ¼ x=c ¼ 2p=k: (3)
If we assume that the wave field is known at the plane y¼ 0
before the object and that the wave propagates along the yaxis, i.e., jkxj; jkzj � jkyj, then the wave field behind the
object, i.e., at y¼D, can be calculated according to the pro-
jection approximation as
wxðx;D; zÞ � e�ikDe�ikÐ D
0ðdðx;y;zÞ�ibðx;y;zÞÞdywxðx; 0; zÞ: (4)
For later reference, it is convenient to introduce here the two
projections of interest, namely
aðx; zÞ ¼ðD0
2kbðx; y; zÞdy (5)
Uðx; zÞ ¼ðD0
kdðx; y; zÞdy: (6)
The projection a(x,z) is used to reconstruct an attenuation
contrast image, whereas the projection U(x,z) is used to
reconstruct the phase contrast image.
In our simulations, we applied the projection approxima-
tion to calculate the propagation of the initial wave front
through the object to the position of G1, i.e., the grating G1
is located at y¼D. For the further propagation from G1 to
G2, separated by a distance d, we used the Fresnel propaga-
tion operator
DFd ¼ expðikdÞF�1 exp � id
2kðk2
x þ k2z Þ
� �F ; (7)
where F and F�1 denotes the two-dimensional Fourier
transform with respect to the coordinates x and z and its
inverse, respectively.
In summary, the simulation of the measurement process
consists of the steps
(1) Setup of a plane wave wx(x, 0, z) at the plane y¼ 0 as
wx(x,0,z) ¼ 1.
(2) Application of the projection approximation, Eq. (4),
using analytically calculated projections a(x,z) and
U(x,z) to obtain the wave field wx(x,D,z) at the location
of the grating G1.
(3) Simulating the effect of the phase grating G1 by modulat-
ing the phase of wx(x,D,z) with a rectangular function with
pitch g1, resulting in a modulated wave field ~wxðx;D; zÞ.(4) Forward propagation of ~wxðx;D; zÞ to the position of
grating G2 by application of the Fresnel operator result-
ing in ~wxðx;Dþ d; zÞ.FIG. 1. Grating-based phase contrast CT setup.
S107 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S107
Medical Physics, Vol. 38, No. 7, July 2011
(5) Simulating for each relative position of the analyzer gra-
ting G2 with respect to the phase grating G1 a detector
reading by:
(a) Simulating the effect of the analyzer grating G2 by
modulating the amplitude of ~wxðx;Dþ d; zÞ with a
rectangular function with pitch g2 resulting in
wxðx;Dþ d; zÞ.(b) Simulating an integrating detector by integrating
jwxðx;Dþ d; zÞj2 over the detector pixel area.
II.B. System setup
Figure 2 shows the mathematical phantom used in the
study. The material compositions were taken from,15 and the
optical properties of these compositions were calculated
based on the formulas described in Ref. 16 using atomic
form factors as tabled in Ref. 17.
Most system parameters were taken from the experimen-
tal setup used by Pfeiffer et al.:6 The total of 721 parallel
beam projections over 180 with 1024 detector pixels cover-
ing 13 cm field of view were simulated (128 lm pixel size).
Since we simulated only a single line system, the coordinate
z was dropped in Eq. (4), which is equivalent to simulating
an object that is homogeneous in z. For each projection
angle, eight different relative positions of G1 with respect to
G2 were simulated in order to retrieve the absorption contrast
and phase contrast images as described in Sec. II C. The
scans were simulated for energies from 25 to 85 keV in steps
of 5 keV assuming monochromatic, coherent x-rays in order
to evaluate the energy dependence of the CNR. The pitches
of the gratings G1 and G2 were g1 = 4 lm and g2¼ 2 lm and
kept constant. G1 was simulated as a pure phase grid impos-
ing 0 or p phase shift to the wave field and G2 was a perfect
analyzer grid, i.e., the amplitude was either not or com-
pletely damped. The distance from G1 to G2 was adjusted for
each energy to the first fractional Talbot distance given by
the relation
d ¼ g21
8k; (8)
resulting in distances ranging from d¼ 40.3 mm for 25 keV
to d¼ 137.1 mm for 85 keV.
For some evaluations, the results are also compared with
a conventional CT system having the same pixel size.
Results from this setup are labeled as obtained by incoherent
absorption contrast imaging (IACI) whereas absorption con-
trast results from the grating-based setup are labeled as
obtained by coherent absorption contrast imaging (CACI).
II.C. Data processing and reconstruction
Absorption and differential phase contrast projections
were obtained from the eight detector readings at different
relative positions of G1 with respect to G2 by the standard
processing as described in Ref. 3: The intensity variation of
the measured intensity at detector pixel j as a function of the
displacement n of the grating G2 in x-direction is modeled as
fjðnÞ ¼ Ijð1þ Vj cosð2pn=g2 þ ujÞÞ; (9)
where Ij is the mean transmitted intensity, Vj is the visibility,
and uj is a phase parameter. All three parameters or deriva-
tives of them can be used for tomographic imaging,3,18 but
here, we restrict ourselves to absorption contrast imaging
(ACI, related to the parameter Ij) and to differential phase
contrast imaging (related to the parameter uj).The two quantities needed for DPCI and CACI were
derived from the l¼ 1, …, N samples acquired at relative dis-
placements nl¼ lg2=N of the transmitted intensity fjl¼ fj(nl)
(which are obtained by the step 5(b) as described in Sec. II A)
for each individual detector pixel j according to:
Ij ¼1
N
XN
l¼1
fjl (10)
uj ¼ argXN
l¼1
fjl cosð2pl=NÞþ iXN
l¼1
fjl sinð2pl=NÞ !
; (11)
where arg(c) denotes the main branch of the argument of a
complex number c, which is sometimes also written as
arg(c)¼ atan2(=(c), <(c)). It is important to note that the
phase uj cannot be calculated unambiguously from the
measurements since the codomain of arg is (�p, p]. This is
reflected in Eq. (11) by the use of the symbol uj rather than
uj in Eq. (9). Phases uj outside the interval (�p, p] are
wrapped into this interval.42
From the first quantity, the conventional absorption pro-
jection is computed by calculating pixel by pixel
aj ¼ � logIj
�I
� �; (12)
FIG. 2. Phantom used for simulations. The diameter of the object is 110
mm, the background material is breast mammary gland adult 1. The objects
have a diameter of 5.1 mm. All objects are invariant in z direction.
S108 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S108
Medical Physics, Vol. 38, No. 7, July 2011
where �I is the intensity of the blank scan, which was constant
for the entire detector in our simulation. The set of all aj
derived from one projection is a discrete version of the con-
tinuous projection as introduced in Eq. (5).
The values uj are proportional to the derivative of the pro-
jection of the real part of the complex refractive index with
respect to x, in the following denoted as the phase gradient,
with the exception that phase gradient wrapping can occur.
Specifically, if we start from the relation given in Ref. 3
(dropping the index j for the sake of compact notation and
stressing that the left hand side is u and not u)
u ¼ kd
g2
@U@x
; (13)
with k being the wavelength of the x-rays, and insert for the
distance d the mth fractional Talbot distance
dm ¼ mg2
1
8k; (14)
where m is an odd positive integer as well as the relation g1
¼ 2g2 for the parallel beam geometry used in this study, the
phase is
u ¼ mg2
2
@U@x) @U
@x¼ 2
mg2
u: (15)
Now we define the dynamic range of the measurement
setup as the range of @U=@x that can be correctly recovered
from a measurement of u, i.e., as the range of @U=@x,
where u ¼ u. For a setup using the first fractional Talbot
distance (m¼ 1) we infer from Eq. (15) that the dynamic
range of the estimated phase gradient @U=@x is (�p, p)
phase difference per one half of the grating pitch of G2. For
setups using a higher order fractional Talbot distance, the
dynamic range of @U=@x is even smaller. Gradients outside
this interval are wrapped into this interval because the arg
in Eq. (11) wraps phases outside the interval (�p, p) into
this interval.
In a previous work,19 we observed strong cupping or
capping artifacts in tomographic phase contrast CT recon-
structions, which we attributed to phase wrapping and to
the fact that near the object border, the approximation that
the gradient of the phase front over the physical detector
width is constant becomes poor since the differential pro-
jection is singular here. We first assumed that this effect
has not been observed previously since in former experi-
ments the object was either thin4,7,13,20,21 or embedded in a
cup with fluid and the blank scan was taken with the cup
entirely filled with fluid,6,8,9 whereas in the simulations
conducted in Ref. 19 and here, the object is rather large.
However, this effect was not observed in a recent study by
Bartl et al.22 although a large object (17 cm diameter) was
embedded in air. After a careful reinvestigation, our hy-
pothesis turned out to be wrong: The phase wrapping was
not the cause of the artifacts. In order to discuss the possi-
ble appearance of phase wrapping, let us consider a cylin-
drical object consisting of a material that creates a phase
shift of a per unit length. The phase front U behind the
object in the projection approximation is
UðxÞ ¼ 0 for jxj > R
2affiffiffiffiffiffiffiffiffiffiffiffiffiffiffiR2 � x2p
for jxj � R
�(16)
resulting in a phase gradient for jxj<R
@U@x¼ � 2xaffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
R2 � x2p : (17)
Now we can calculate from Eqs. (15) and (17) the positions
xw, where the gradient leaves the dynamic range of the sys-
tem, which happens when the phase u hits the border of the
dynamic range of the arg function, i.e., if the true phase ubecomes �p or þp. Thus we need to solve the equation
6p¼! u ¼ mg2
2
�2xaffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiR2 � x2p (18)
for x, which turns out to be
xw ¼ �Rffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffiffi
1þ ðmg2a=pÞ2q : (19)
For the parameters in our setup (m¼ 1, R¼ 55 mm,
g2¼ 2 lm, and a � 46 rad=mm at 25 keV), we obtain
xw ¼ �54:976 mm, i.e., only the outermost 24 lm of the
object’s projection on each side are affected. In other words,
the problem of phase wrapping will not be visible for the
system with a simulated pixel size of 128 lm. However, it is
still true that the phase gradient @U=@x in this outermost
area varies strongly since it is approaching a singularity
and thus, an accurate estimate of the phase gradient is
impossible.
Finally we found that the cupping and capping artifacts
disappeared almost completely after decreasing the sample
spacing used when the Helmholtz equation was applied [cf.,
Eq. (4)] to 8 nm, meaning that wx(x, 0, 0) was discretized on
a grid in x with spacing 8 nm and this spacing was kept until
the final step of integrating the square of the wave function
over the pixel area. Additionally, we placed the object
slightly off-center to create a variablity in the sampling of
the edges.
Reconstruction was performed using filtered back-projec-
tion using a conventional ramp-filter for the absorption con-
trast image23 and a Hilbert filter for the phase contrast
image24 with the additional trick of using a half-pixel-shift as
described in Ref. 25 for antialiasing purposes. In order to sup-
press aliasing in both absorption and phase contrast imaging,
cubic spline interpolation was used during back-projection.
II.D. CNR measurement
The CNR for each object o was calculated as
CNRo ¼So � SBG
ro(20)
where So is the reconstructed image value of the object o,SBG is the reconstructed image value of the background ma-
terial (i.e., breast mammary gland adult 1) and ro is the esti-
mated noise level in the object o. Note that in our definition,
the CNR may be positive or negative, which is unusual but it
S109 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S109
Medical Physics, Vol. 38, No. 7, July 2011
is useful for the discussion of the results to be able to distin-
guish between negative and positive contrasts.
For each object o, the signal level So was determined as
the average reconstructed valued within a square of 2.56 mm
edge size centered within the object, using a pixel size of
20 lm. The background signal was determined as the aver-
age reconstructed valued within a square of 2.56 mm edge
size located at the center of the object.
As will be seen in the results section, the phase contrast
images contain a considerable amount of low-frequency
noise, which makes a contrast measurement difficult. There-
fore, the contrast was measured in noise-free reconstructions.
Furthermore, since we still observe minor cupping artifacts
in the reconstructed phase contrast images, in particular for
low energies, we determined the noise level in the difference
images between the noisy and the noiseless case.
II.E. Relative contrast measurement
The relative contrast for each object o was calculated as
RCo ¼So � SBG
SBG
(21)
where So and SBG are the same quantities as defined in
Sec. II D. For the phase contrast images, where the contrast
is exclusively caused by differences in the electron density,
this quantity should equal for each object the relative differ-
ence in electron density
Dno
nBG
¼ no � nBG
nBG
; (22)
where no is the electron density of the object and nBG is the
electron density of the background.
II.F. Modulation transfer function measurements
One aspect of this study is to compare the CNR of DPCI
and ACI. Of course, a fair comparison is only possible if the
CNR is measured in images with the same resolution. In
order to measure the spatial resolution, the modulation trans-
fer function (MTF) was determined.
Typically, the MTF is measured using a bead or a wire
with a size much smaller than the detector pixel size.26,27
This approach is, however, impractical for DPCI since the
system is actually insensitive to objects much smaller than
the pixel size. This is due to the fact that the phase gradient
is averaged over the pixel area. For objects being projected
completely onto a single pixel, the resulting mean gradient is
zero and thus, the object is invisible in DPCI.43 A second
aspect that needs to be kept in mind is that we performed
simulations with coherent x-rays which may result in edge-
enhancement effects in the reconstructed absorption contrast
images. This effect is well visible in the absorption images
obtained by the grating-based measurement setup, e.g., in
Ref. 3 which were obtained using synchrotron radiation.
Thus, this effect needs to be considered.
Therefore, the MTF was determined from the edge
response function measured directly in the phantom where
the CNR evaluation is done with the exception that the
noise-free data were used: For each object, 360 radial lines
starting from the center of the object were reconstructed on a
5 lm grid and averaged. Averaging was required to become
insensitive to remaining aliasing artifacts in the recon-
structed images. The point spread function was estimated by
numerical differentiation of the resulting edge-response fol-
lowed by normalization with the step height. Finally, the
MTF was calculated as the absolute value of the one-dimen-
sional Fourier-transformation of the point spread function.
This procedure of calculating the MTF neglects the influence
of the discrete calculation of the derivative, which is a minor
effect since the sampling density of 5 lm is much smaller
than the pixel size (128 lm). Second, the finite curvature of
the edge is neglected. However, since all lines are perpendic-
ular to the edge, its curvature has only a second order effect.
In any case, the main purpose of the MTF calculation is to
compare the different imaging methods which are all influ-
enced equally.
II.G. Noise evaluation
For each calculated scenario, we first adapted the number
of source photons such that the total energy deposited into
the object was constant, thus allowing for a meaningful com-
parison of the CNR in the different reconstructed images.
The dose calculation was performed using a Monte Carlo
simulation considering photoelectric absorption and Comp-
ton scattering. For this dose calculation, the 3D structure of
the object was assumed to be a cylinder of infinite length in zdirection and the energy deposited within the cylinder was
taken as a dose measure. Thereafter, we deduced analytically
the expected mean number of detected photons for each de-
tector pixel, and created white noise according to the Poisson
distribution.
For the calculation of the noise power spectrum (NPS), a
central square of 50 mm edge size was reconstructed using a
10242 image matrix. Object structures were removed by sub-
tracting the noise-free reconstruction. The NPS is deter-
mined by a 2D FFT followed by a binning into several
frequency bands of 3 lp/cm width and a calculation of the
total signal intensity in each frequency bin. The resulting
spectra are further normalized to the total noise power in the
image.
For the calculation of the noise ro required for CNR
measurements, a similar procedure was performed: The
noise-free reconstruction was subtracted from the noisy
reconstruction and ro was calculated as the square root of
the variance of all image values in the object. A single noise
realization was not sufficient for a stable noise estimation
since the objects are relatively small and the NPS shows sub-
stantial low-frequency noise for the DPCI case. Therefore,
the noise estimate of ten independent noise realizations was
averaged.
III. RESULTS
Figure 3 shows an overview of the reconstruction results
for the absorption and phase contrast images for selected
energies. The detailed evaluation of MTF and CNR are
S110 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S110
Medical Physics, Vol. 38, No. 7, July 2011
performed on the central part containing four objects of in-
terest. This region of interest (ROI) is shown in Fig. 4.
III.A. Modulation transfer function
Results for the MTF measured at the object edges of the
four objects within the ROI are shown in Fig. 5 for DPCI,
CACI, and IACI. Exemplarily, we selected the results at
25 keV.
The following observations can be made: First of all, the
MTF of DPCI matches well the MTF of IACI for all cases.
Second, the MTF of CACI shows some variability. We
attribute this variation to so-called edge-enhancement
effects, which are well known in the field of x-ray imaging
using coherent sources.
In the following, we will give a qualitative reasoning for
the observed MTFs in the CACI case. To begin with, we
recall a basic formula for the edge-enhancement phenom-
enon in propagation-based x-ray imaging. Let I0(x) denote
the intensity directly behind the object, i.e., in the so-call
contact plane. For a weakly absorbing object, the intensity
Id(x) at a distance d behind the object can be calculated as28
IdðxÞ ¼ I0ðxÞ 1� kd
2pr2UðxÞ
� �; (23)
where U is the phase of the wavefront as introduced in
Sec. II A.
For a qualitative discussion of the effect, we further
assume that kd2pr2UðxÞ � 1, so that the estimation a(x) of
the projection of the linear attenuation coefficient (which is
used for CACI) is approximately
~aðxÞ ¼ � logIdðxÞ
�I¼ � log
I0ðxÞ�Iþ log 1� kd
2pr2UðxÞ
� �
� aðxÞ � kd
2pr2UðxÞ; (24)
meaning that the estimated projection is perturbed by an
additive term � kd2pr2UðxÞ.
The effect of edge-enhancement is often seen in
propagation-based phase contrast imaging at the object
FIG. 3. Sample reconstruction results of the phantom
simulated for selected energies. The top row shows the
absorption contrast images, the bottom row the recon-
structed phase contrast images. Level and window are
set as in Fig. 4.
FIG. 4. Central region of interest of the images shown
in Fig. 3. The top row shows the absorption contrast
images and the bottom row the reconstructed phase
contrast images. For each individual image, the level is
set to the background material and window is selected
such that the full dynamic range of the region of inter-
est is shown.
S111 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S111
Medical Physics, Vol. 38, No. 7, July 2011
boundary, i.e., at the air–tissue interface. At this interface,
the object has a positive contrast in both b and d compared
to the background.
Now we turn again to our simulation results. It is impor-
tant to keep the contrasts of the different objects as shown in
Fig. 4 (leftmost column) in mind: Water and adipose tissue
lipoma create the least contrast in the phase contrast image
among the four objects and the coherent absorption contrast
image is only to a very small extent affected by edge-
enhancement and the MTF of CACI matches the other two
MTFs. The material breast mammary gland adult 2 shows
decent positive contrast in phase contrast and the absorption
contrast images. The projection of this object is subject to
edge-enhancement resulting in an improvement of the MTF
at higher frequencies. A very special situation is present for
the material breast (whole) 50/50, which shows positive con-
trast in the absorption image but negative contrast in
the phase contrast image. Furthermore, the contrast in the
absorption image is rather low whereas the contrast in the
phase contrast image is relatively strong, which leads to a
strong influence of the phase contrast on the MTF of CACI.
The fact that the material has opposite contrasts in the
absorption contrast and the phase contrast image is key to
understand the observed effect of a degradation of the MTF:
As argued in the discussion of Eq. (24), the additive perturb-
ance � kd2pr2UðxÞ leads to the edge-enhancement. If the con-
trast of d changes its sign, it also changes the sign of the
disturbing term and the edge-enhancement turns into a
smoothing.
We need to stress here that the change of the MTF for
CACI due to edge-enhancement can give only a qualitative
indication about how the MTF is affected by the propagation
of the coherent wave field in a real experiment. For a correct
quantitative prediction of this effect it would be required to
take into account the finite size of the object along the x-ray
beam, since the object size (110 mm) is comparable to the
distance d between the two gratings (40.3 mm for the shown
case of 25 keV). Therefore, edge-enhancement of the same
magnitude will already take place while the beam propagates
through the object to the grating G1. Furthermore, the
approximation that the object is only weakly absorbing is
not very accurate for the object studied here.
III.B. Noise power spectrum
The visual appearance of noise can be seen already in
Fig. 4, showing that phase contrast images show predomi-
nantly low-frequency noise. The noise power spectra are
shown in Fig. 6. For comparison, the NPS graph in Fig. 6
shows additionally the NPS of IACI. Evidently and not sur-
prising, the NPS for CACI and IACI are identical.
In order to illustrate one important aspect of the different
noise power spectra, the ROI as shown in Fig. 4 was once
reconstructed as described in Sec. II C and for comparison,
this image was further smoothed with a Gaussian kernel of
full-width-half-maximum of 0.3 mm. The resulting images
are shown in Fig. 7. The noise level in the absorption con-
trast images decreases by the smoothing by a factor of 3.0,
FIG. 5. MTFs obtained at 25 keV for the objects within the ROI. The MTF graphs are ordered according to the position of the objects in the ROI shown in the
center.
S112 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S112
Medical Physics, Vol. 38, No. 7, July 2011
whereas the noise reduction for the phase contrast image is
only a factor of 1.4.
III.C. Energy dependence of CNR
A detailed analysis of the CNR as a function of x-ray
energy was performed on the four objects shown in the ROI,
see Fig. 4. The analysis is shown in Fig. 8. The curves of
CACI and IACI differ in their scale by a factor offfiffiffi2p
reflect-
ing the absorption of half the photons in the analyzer grating
G2 in CACI.
As stated before, a fair comparison of CNR between dif-
ferent imaging methods requires that the methods provide
the same resolution. The results of MTF measurements as
shown in Fig. 5, however, indicate that CACI and DPCI do
not provide the same resolution. Strictly speaking, only the
CNR curves of IACI and DPCI shown in Fig. 8 should be
compared, but not the CACI curve.
III.D. Contrast measurements
Table I sumarizes for the material within the ROI the rela-
tive difference in electron density as reported in Ref. 15 and
the measured relative contrasts in the phase contrast images.
Since the electron densities are given with three significant
digits in Ref. 15, the uncertainty in the relative differences in
the electron density is in the order of 0.1% and the measured
relative contrasts match well the expected values.
IV. DISCUSSION
The noise power spectrum of the phase contrast image
shows a maximum at low frequencies, see Fig. 6. This can
be easily understood from the reconstruction kernel used in
the reconstruction: The ramp filter used in absorption con-
trast imaging damps low frequencies and boosts the higher
frequencies leading to the well-known peak in the noise
power spectrum near the Nyquist frequency of the detection
system (actually, due to the smoothing effect of interpolation
during back-projection, the peak is well below the Nyquist
frequency of 39 lp=cm). On the other hand, the Hilbert filter
used in phase contrast imaging treats all frequencies equally,
leading to a higher noise power at lower frequencies.44 One
of the important implications of this feature for tomographic
medical imaging is that the superior CNR of phase contrast
images compared with absorption contrast as demonstrated
in Refs. 6 and 9 will diminish for applications, where a lower
spatial resolution is required: In absorption contrast imaging,
noise and resolution can be traded off very efficiently since
smoothing also removes noise in the frequency range, where
most of the noise is. However, for phase contrast imaging,
smoothing only suppresses the tail of the noise power spec-
trum and does not lead to a substantial increase in CNR. For
the example shown in Fig. 7, the CNR of the absorption con-
trast image after smoothing improves by a factor of 3.0 in
CACI whereas it improves only by a factor of 1.4 in the
DPCI. This effect should always be considered if the two
modalities are compared with respect to the achieved CNR.
Regarding CNR, it should also be noted that the CNR in the
phase contrast image can be increased by increasing the dis-
tance between G1 and G2 to a higher fractional Talbot order
because the sensitivity to the gradient of the phase front is
increased, see Eq. (15). This was for example used to obtain
the brain images in Ref. 6, where the 9th fractional Talbot
order was used, resulting in a sensitivity increase by a factor
of nine. The price to pay for this increase in CNR is that the
phase wrapping as discussed in Sec. II C will appear and the
problem of unwrapping the phase gradient will become a
demanding task. Finally, we need to stress that we simulated
a perfect DPCI system, i.e., perfect gratings have been used
and a perfectly coherent and monochromatic x-ray beam was
simulated. In a real clinical system, the CNR of the DPCI
would be reduced due to a decrease in the visibility caused
FIG. 6. Noise power spectra of absorption and phase contrast imaging.
FIG. 7. ROI reconstruction at 85 keV at system resolution, i.e., reconstructed with a pure ramp and Hilbert filter, respectively (a and c) and the same images
smoothed using a Gaussian kernel of 0.3 mm full-width-half-maximum (b and d) for absorption contrast (a and b) and phase contrast (c and d).
S113 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S113
Medical Physics, Vol. 38, No. 7, July 2011
by imperfections of the grating, the limited coherence of the
x-ray source, and the finite bandwidth of the x-rays.
The MTF measurements show that a direct and fair com-
parison of the CNR obtained by DPCI and CACI is difficult
since the MTF for CACI shows a dependency on the object’s
material. In our simulations, we found that the MTF in
CACI is improved whenever the contrasts of the object have
the same sign in d and b. Only for a single object (and even
for this only for energies below 35 keV), the contrasts have
opposite signs leading to a degradation in the MTF. Thus,
we may state that the loss of CNR in CACI due to the ana-
lyzer grating G2 is typically compensated partially by a
slight improvement of the MTF.
With respect to the noise properties of reconstructed
phase contrast images, it is also worth comparing the fea-
tures observed here with previous investigations for other ac-
quisition methods for phase contrast x-ray imaging: For the
in-line phase contrast imaging method, the auto-covariance
of the noise was studied by Chou and Anastasio.12 Similar to
our results for the grating-based method, they found that the
auto-covariance function is rather broad, which is equivalent
to predominantly low frequency noise. While in Ref. 12, the
reconstruction involved an explicit retrieval of the projection
of the real part of the refractive index, there are also other
reconstruction algorithms, which do not need this intermedi-
ate step. Interestingly, these algorithm are found to be stable
at high frequencies and unstable at low frequencies,28–30
where the instability originates from a singularity, viz., a first
order pole, in the reconstruction kernel at zero frequency.
This feature implies at least qualitatively the same noise
behavior.
The CNR of phase contrast images measured at constant
dose for the sample phantom are qualitatively very similar:
For all objects, the magnitude of the CNR in DPCI shows a
maximum around 45 keV. The existence of a maximum at a
distinct energy independent of the object is caused by several
effects: First, the phase shift introduced by the objects is
exclusively created by the electron density, see Sec. III D,
with an energy dependence of E�1: the real part of the re-
fractive index d has an energy dependence of E�2, and one
factor of E is included in the wave vector k in the projection
approximation, see Eqs. (4) and (6). Thus, it is expected
that the energy dependence of the CNR of all objects in the
phantom is qualitatively the same. The second cause is the
different contributions of the photoelectric effect and
FIG. 8. CNR measurements for the objects within the ROI as a function of energy.
TABLE I. Relative difference in electron density of the materials in the ROI
as derived from (Ref. 15) and measured relative contrast as defined in
Sec. II E. The measurements for RCo are independent of the energy.
Material o Dno=nBG RCo
Adipose tissue lipoma �0.9% �1.0%
Breast mammary gland adult 2 þ2.7% þ2.7%
Breast (whole) 50=50 �2.4% �2.5%
Water þ1.2% þ1.1%
S114 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S114
Medical Physics, Vol. 38, No. 7, July 2011
Compton scatter to the total attenuation. For lower energies,
the photoelectric effect dominates. However, it decreases
approximately as E�3, whereas the phase contrast decreases
only as E�1. Thus, going to higher energies decreases the
contrast in the phase image, but this is more than compen-
sated by the fact that at constant dose more photons can be
used for imaging since fewer of them are absorbed. How-
ever, as soon as the Compton effect contributes significantly
to the attenuation at higher energies, the total transmission
decreases again and the decreasing phase contrast is—in the
constant dose scenario—no longer compensated by an even
faster decreasing absorption.
Herzen et al.31 showed in experiments using several dif-
ferent liquids that phase and absorption contrast images
show complementary information in the sense that the CNR
is sometimes larger in the phase contrast image and some-
times in the absorption contrast image. Our simulation using
material compositions of biological tissues confirms that this
feature is relevant for medical imaging.
Conceptually, the grating-based DPCI setup invests half
the dose, namely the dose that is absorbed in the analyzer
grating G2, to obtain additional information about the object.
Qi et al.,21 for instance, derived quantitative images of the
electron density and the effective atomic number. It should
be acknowledged that there are also other setups which can
provide the same information without investing dose in the
first place: These methods are based on dual energy acquisi-
tions or spectrally resolved detection followed by a separa-
tion of the total attenuation into the contributions of the
photoelectric effect and Compton scattering.32–34 In these al-
ternative methods, there is no dose penalty to pay in the first
place. Furthermore, it has been demonstrated that spectral
CT can also detect specifically materials which have a
K-edge in the used spectral range, providing even further
useful information.35,36 Although it should be theoretically
possible to detect these materials also in the phase contrast
signal (this is due to the relationship between d and b given
by the Kramers–Kronig relation), it has not been shown
experimentally yet. From a system point of view, this alter-
native option to obtain the additional information might be
preferred since the required system modifications are moder-
ate compared with the system modifications required for a
DPCI system. On the other hand, it is unclear currently,
whether these methods can provide the additional informa-
tion at a CNR-level that is comparable to that obtained by a
DPCI setup, which is not a-priori clear since the material
decomposition from dual energy acquisitions suffers from
noise correlation.37 Finally, as mentioned before, the gra-
ting-based DPCI setup can provide a third reconstructed
image, the so-called decoherence image, based on the visi-
bility function, see Eq. (9). Such an image cannot be
obtained by dual energy or spectral CT. Thus, the potential
clinical benefit of the decoherence image might be finally a
key argument whether a grating-based DPCI setup is worth
being used for clinical applications.
a)Electronic mail: [email protected]. A. Lewis, “Medical phase contrast x-ray imaging: current status and
future prospects,” Phys. Med. Biol. 49, 3573–3583 (2004).
2A. Momose, S. Kawamoto, I. Koyama, Y. Hamaishi, K. Takai, and Y.
Suzuki, “Demonstration of x-ray talbot interferometry,” Jpn. J. Appl.
Phys. 42, L866–L868 (2003).3T. Weitkamp, A. Diaz, C. David, F. Pfeiffer, M. Stampanoni, P. Cloetens,
and E. Ziegler, “X-ray phase imaging with a grating interferometer,” Opt.
Express 13, 6296–6304 (2005).4F. Pfeiffer, C. Kottler, O. Bunk, and C. David, “Hard x-ray phase tomography
with low-brilliance sources,” Phys. Rev. Lett. 98, 108105-1–108105-4
(2007).5A. Momose, W. Yashiro, H. Maikusa, and Y. Takeda, “High-speed x-ray
phase imaging and x-ray phase tomography with Talbot interferometer
and white synchrotron radiation,” Opt. Express 17, 12540–12545 (2009).6F. Pfeiffer, O. Bunk, C. David, M. Bech, G. L. Duc, A. Bravin, and P.
Cloetens, “High-resolution brain tumor visualization using three-dimen-
sional x-ray phase contrast tomography,” Phys. Med. Biol. 52, 6923–6930
(2007).7F. Pfeiffer, O. Bunk, C. Kottler, and C. David, “Tomographic reconstruc-
tion of three-dimensional objects from hard x-ray differential phase con-
trast projection images,” Nucl. Instrum. Methods Phys. Res. A 580, 925–
928 (2007).8M. Bech, T. H. Jensen, R. Feidenhans’l, O. Bunk, C. David, and F. Pfeiffer,
“Soft-tissue phase-contrast tomography with an x-ray tube source,” Phys.
Med. Biol. 54, 2747–2753 (2009).9T. Weitkamp, C. David, O. Bunk, J. Bruder, P. Cloetens, and F. Pfeiffer,
“X-ray phase radiography and tomography of soft tissue using grating
interferometry,” Eur. J. Radiol. 68, S13–S17 (2008).10T. Donath, F. Pfeiffer, O. Bunk, W. Groot, M. Bednarzik, C. Grunzweig,
E. Hempel, Popescu, M. S., Hoheisel, and C. David, “Phase-contrast
imaging and tomography at 60 keV using a conventional x-ray tube
source,” Rev. Sci. Instrum. 80, 053701-1–053701-4 (2009).11C.-Y. Chou and M. A. Anastasio, “Influence of imaging geometry on noise
texture in quantitative in-line x-ray phase-contrast imaging,” Opt. Express
17, 14466–14480 (2009).12C.-Y. Chou and M. A. Anastasio, “Noise texture and signal detectability
in propagation-based x-ray phase-contrast tomography,” Med. Phys. 37,
270 – 281 (2010).13T. Weitkamp, C. David, C. Kottler, O. Bunk, and F. Pfeiffer, “Tomography
with grating interferometers at low-brilliance sources,” Proc. SPIE 6318,
63180S (2006).14D. M. Paganin, Coherent X-Ray Optics (Oxford University Press, Oxford,
2006).15“Photon, electron, proton, and neutron interaction data for body tissues,”
ICRU Report No. 46” (International Commission on Radiation Units and
Measurements, Bethesda, MD, 1992).16B. L. Henke, E. M. Gullikson, and J. C. Davis, “X-ray interactions: photo-
absorption, scattering, transmission, and reflection at E = 50–30000 eV, Z
= 1–92,” At. Data Nucl. Data Tables 54, 181–342 (1993).17C. T. Chantler, K. Olsen, R. A. Dragoset, A. R. Kishore, S. A. Kotochigova,
and D. S. Zucker, “X-ray form factor, attenuation, and scattering tables (ver-
sion 2.1),” National Institute of Standards and Technology (2005).18M. Bech, O. Bunk, T. Donath, R. Feidenhans’l, C. David, and F. Pfeiffer,
“Quantitative x-ray dark-field computed tomography,” Phys. Med. Biol.
55, 5529–5539 (2010).19T. Koehler, K. J. Engel, and E. Roessl, “Noise properties of grating based
x-ray phase contrast computed tomography,” in Proceedings of the firstinternational conference on image formation in X-ray computed tomogra-phy (Salt Lake City, Utah, USA, 2010), pp. 181–184.
20J. Zambelli, N. Bevins, Z. Qi, and G.-H. Chen, “Radiation dose efficiency
comparison between differential phase contrast CT and conventional
absorption CT,” Med. Phys. 37, 2473–2479 (2010).21Z. Qi, J. Zambelli, N. Bevins, and G.-H. Chen, “Quantitative imaging of
electron density and effective atomic number using phase contrast CT,”
Phys. Med. Biol. 55, 2669–2677 (2010).22P. Bartl, J. Durst, W. Haas, E. Hempel, T. Michel, A. Ritter, T. Weber,
and G. Anton, “Simulation of x-ray phase-contrast computed tomography
of a medical phantom comprising particle and wave contributions,” Proc.
SPIE 7622, 76220Q (2010).23A. C. Kak and M. Slaney, Principles of Computerized Tomographic Imag-
ing (IEEE Press, New York, 1987).24G. W. Faris and R. L. Byer, “Three-dimensional beam-deflection optical
tomography of a supersonic jet,” Appl. Opt. 27, 5202–5212 (1988).25F. Noo, J. Pack, and D. Heuscher, “Exact helical reconstruction using
native cone-beam geometries,” Phys. Med. Biol. 48, 3787–3818 (2003).
S115 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S115
Medical Physics, Vol. 38, No. 7, July 2011
26“IEC 61223–3–5, Acceptance tests - imaging performance of computed
tomography X-ray equipment, draft” (2003).27D. Schafer, J. Wiegert, and M. Bertram, “Method for the determination of
the modulation tranfer function (MTF) in 3D X-ray imaging systems with
focus on correction for finite extent of test objects,” Proc. SPIE 6510,
65104Q (2007).28A. V. Bronnikov, “Theory of quantitative phase-contrast computed
tomography,” J. Opt. Soc. Am. A 19, 472–480 (2002).29A. Groso, M. Stampanoni, R. Abela, P. Schneider, S. Linga, and R. Muller,
“Phase contrast tomography: An alternative approach,” Appl. Phys. Lett. 88,
214104 (2006).30A. V. Bronnikov, “Phase-contrast CT: Fundamental theorem and recon-
struction algorithms,” in Proceedings of the Fully3D (2007), pp. 429–432.31J. Herzen, T. Donath, F. Pfeiffer, O. Bunk, C. Padeste, F. Beckmann, A.
Schreyer, and C. David, “Quantitative phase-contrast tomography of a liq-
uid phantom using a conventional x-ray tube source,” Opt. Express 17,
10010–10018 (2009).32R. E. Alvarez and A. Macovski, “Energy-selective reconstructions in x-ray
computerized tomography,” Phys. Med. Biol. 21, 733–744 (1976).33B. J. Heismann, J. Leppert, and K. Stierstorfer, “Density and atomic number
measurements with spectral x-ray attenuation method,” J. Appl. Phys. 94,
2073 (2003).34W. Zou, T. Nakashima, Y. Onishi, A. Koike, B. Shinomiya, H. Morii, Y.
Neo, H. Mimura, and T. Aoki, “Atomic number and electron density mea-
surement using a conventional x-ray tube and a CdTe detector,” Jpn. J.
Appl. Phys. 47, 7317–7323 (2003).35E. Roessl and R. Proksa, “K-edge imaging in x-ray computed tomography
using multi-bin photon counting detectors,” Phys. Med. Biol. 52, 4679–
4696 (2007).
36E. Roessl, B. Brendel, J. P. Schlomka, A. Thran, and R. Proksa, “Sensitivity
of photon-counting K-edge imaging: Dependence on atomic number and
object size,” in Proceedings IEEE Nuclear Science Symposium ConferenceRecord NSS ‘08” (2008), pp. 4016–4021.
37W. A. Kalender, E. Klotz, and L. Kostaridou, “An algorithm for noise sup-
pression in dual energy CT material density images,” IEEE Trans. Med.
Imaging 7, 218–224 (1988).38A. Olivo and R. Speller, “Image formation principles in coded-aperture based
x-ray phase contrast imaging,” Phys. Med. Biol. 53, 6461–6474 (2008).39Z.-F. Huang, K.-J. Kang, L. Zhang, Z. Chen, F. Ding, Z. Wang, and Q.-G.
Fang, “Alternative method for differential phase-contrast imaging with
weakly coherent hard x rays,” Phys. Rev. A 79, 013815-1–013815-6
(2009).40D. C. Ghiglia and M. D. Pritt, Two-Dimensional Phase Unwrapping -
Theory, Algorithms, and Software (John Wiley & Sons, Inc., 1998).41We do not consider here the alternative grating-based setups proposed in
Refs. 38 and 39.42Note that it is in the general case impossible to detect, whether phase
wrapping occurred or not. Only if it is ensured that the phase gradi-
ent is sampled properly by the system, i.e., if the true phase u does
not vary by more than p between adjacent pixel in a row, then one-
dimensional phase unwrapping by using Itoh’s method can be
employed.43However, such a small object would appear in the dark field image since
the variation of the phase gradient leads to a loss of the visibility, see
Eq. (9) and (Ref. 18).44Due to the inhomogeneous sampling of the Radon space and thus also the
Fourier space with a series of parallel projections, this equal treatment of
all frequencies does not lead to a flat noise power spectrum.
S116 Kohler, Engel, and Roessl: Noise properties of phase contrast CT S116
Medical Physics, Vol. 38, No. 7, July 2011