µox Compact troubleshooting guide

Start the slide show then click Next

Page 2: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Compact

Home

Symptoms of the problem (click on the relevant line)

No display on the unitThe machine has frozen up

The oxygen consumption is too high.

The machine switches on but nothing happens even if you override manually.The right hand pressure gauge is below 2.5 bars.

The right hand pressure gauge is above 2.5 bars

The pressure on the right hand pressure gauge does not reach 2.5 bars

The pressure on the left hand pressure gauge remains at 0.

The pressure on the left hand pressure gauge is above 2 bars.

The Norton tube (noodle) on the equipment outlet disconnects constantly.

The equipment will not run in automatic mode.

The equipment does not work and the machine is displaying an error code.

Page 3: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the equipment connected to a power supply?

Page 4: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Disconnect and reconnect the equipment to reinitialise it.

Has the machine restarted?

Page 5: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the equipment connected to the regulator properly?

And is the pressure at 6 bars?

(sometimes the connection is poor due to oxygenation)

Page 6: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the pressure on the right hand pressure gauge at 2.5 bars?

Page 7: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the equipment connected to the regulator properly?

And is the pressure at 6 bars?

(sometimes the connection is poor due to oxygenation)

Page 8: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Manually override in small increments by turning the regulator clockwise to 2.5 bars.

Did this help to solve your problem?

Yes No

Home

Page 9: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the pressure on the left hand pressure gauge over 2.5 bars?

Page 10: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Completely unscrew the regulator, carry out a manual override in order to remove any impurities, then reset the pressure to 2.5 bars.

If the problem persists, the doser module is defective. Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 11: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Carry out a manual override. Are the solenoid valves working?

Page 12: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Are the Norton tubes (noodles) at the doser outlet and the tank inlet disconnected?

Page 13: Next Home µox Compact troubleshooting guide Start the slide show then click Next

If yes If no

Home

Check the ceramic element; it may be broken or the pipe connecting to the diffuser may be disconnected.

Page 14: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the device's outlet coupling connected properly?

(risk of a poor connection due to oxygenation).

Page 15: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser, carry out a manual override and the gas

should flow.

Does the left hand pressure gauge fall again immediately?

Page 16: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the diffuser clogged?

Page 17: Next Home µox Compact troubleshooting guide Start the slide show then click Next

If yes If no

Home

Check that the pipe between the device and the diffuser is not squashed or blocked, disconnect the Norton tube from the diffuser and carry out a manual override. The gas

should flow and the left hand pressure gauge should fall again immediately.

Page 18: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Have the tank and dose volumes been programmed?

Page 19: Next Home µox Compact troubleshooting guide Start the slide show then click Next

YesHome

Is the programmed dose compatible with the volume programmed? (see table below)

Tank vol

10 100 250 500 1000 2000 5000Cali-bratio

n value

min - max

50 0.4 - 200

0.1 - 90

0.1 - 40 0.1 - 2 0.1 - 9 0.1 - 4,5

0.1 - 1.8

100 0.7 - 200

0.1 - 180

0.1 - 72

0.1 - 36

0.1 - 18 0.1 - 9

0.1 - 3.6

250 1.6 - 200

0.2 - 200

0.1 - 180

0.1 - 90

0.1 - 45

0.1 - 22.5 0.1 - 9

500 3.2 - 200

0.3 - 200

0.2 - 200

0.1 - 180

0.1 - 90 0.1 - 45

0.1 - 18

1000 6.3 - 200

0.7 - 200

0.3 - 200

0.2 - 200

0.1 - 180 0.1 - 90

0.1 - 36

2000 12.5 - 200

1.3 - 200

0.5 - 200

0.3 - 200

1.3 - 200

0.1 - 180

0.1 - 72

5000 31.3 - 200

3.2 – 200

1.3 - 200

0.7 - 200

0.4 - 200

0.2 - 200

0.1 - 180

No

Page 20: Next Home µox Compact troubleshooting guide Start the slide show then click Next

If yes If no

Home

Check that there are no leaks in the circuit; see attached procedure.

Page 21: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Defective machine or supply

Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 22: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The equipment is ready for use.

Did this help to solve your problem?

Yes No

Home

Page 23: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The O2 supply circuit is blocked.

Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 24: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The equipment’s outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation).

Replace the coupling.

Doser outlet 2/4 (ref: 1.6SortieDoseur2/4).

Did this help to solve your problem?

Yes No

Home

Page 25: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank.

Did this help to solve your problem?

Yes No

Home

Page 26: Next Home µox Compact troubleshooting guide Start the slide show then click Next

You can repair it with a stainless steel olive or replace the entire tube.

ref: 215TA10MO

Did this help to solve your problem?

Yes No

Home

Page 27: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Replace the coupling.

Doser outlet 2/4 (ref: 1.6SortieDoseur2/4).

Did this help to solve your problem?

Yes No

Home

Page 28: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank.

Did this help to solve your problem?

Yes No

Home

Page 29: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Pipe blocked - replace it.

Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it.

If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in

it needing to be replaced.

It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use.

Did this help to solve your problem?

Yes No

Home

Page 30: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Clean the ceramic element (see attached procedure); check that it is working properly in a bucket before putting it back in the tank.

Did this help to solve your problem?

Yes No

Home

Page 31: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Pipe blocked - replace it.

Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it.

If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in

it needing to be replaced.

It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use.

Did this help to solve your problem?

Yes No

Home

Page 32: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The machine has lost its calibration value; to check it disconnect and reconnect the device. The first screen that appears will indicate the calibration value. If this value is 0, you will need to reprogramme it.

Programming the calibration value requires you to open the device.

Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 33: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Repair following the attached instructions or contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 34: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Check the power supply and connections.

Did this help to solve your problem?

Yes No

Home

Page 35: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The machine is defective. Contact our after-sales service for a replacement.

Did this help to solve your problem?

Yes No

Home

Page 36: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Set the O2 bottle outlet pressure to 6 bars.

Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to

operate.

Did this help to solve your problem?

Yes No

Home

Page 37: Next Home µox Compact troubleshooting guide Start the slide show then click Next

While carrying out a manual override, gently tap on the black screw next to the right hand pressure gauge; a solenoid valve may be stuck.

Otherwise, the solenoid valve or the solenoid valve coil is defective.

Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 38: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The equipment’s outlet connection is not connected or is not connected securely (risk of poor connection due to oxygenation).

Possibly replace the regulator outlet coupling.

Ref: 3.1.SD1

Did this help to solve your problem?

Yes No

Home

Page 39: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Set the O2 bottle outlet pressure to 6 bars.

Reminder: it is the difference in pressure between the supply at 6 bars and the back pressure at 2.5 bars that acts as a reference for the device to

operate.

Did this help to solve your problem?

Yes No

Home

Page 40: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Dosing module defective; contact our after-sales service to replace it.

Did this help to solve your problem?

Yes No

Home

Page 41: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Check the regulator setting by turning it anti-clockwise; if the pressure does not drop, the dosing module is defective. Contact our after-sales service for

a replacement.

Did this help to solve your problem?

Yes No

Home

Page 42: Next Home µox Compact troubleshooting guide Start the slide show then click Next
Page 43: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Defective solenoid valve coil.

Contact our after-sales service for a replacement.

Did this help to solve your problem?

Yes No

Home

Page 44: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Leak in the pipe between the device and the tank. Repair it with a joint or replace it.

Ref: 1.6Manchon

Did this help to solve your problem?

Yes No

Home

Page 45: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Pipe blocked - replace it.

Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it.

If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in

it needing to be replaced.

It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use.

Did this help to solve your problem?

Yes No

Home

Page 46: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Yes No

Home

Is the ceramic element in the diffuser clogged?

Page 47: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Internal circuit in device blocked.

Contact after-sales service.

Note: This happens if you switch off the device without taking the diffuser out of the wine or if you leave it to soak in cleaning solution without disconnecting it.

If effect, the wine or cleaning solution will consume the oxygen in the tubes and rise up to the dosing device, crystallise, block all the tubes and may even damage the dosing module, resulting in

it needing to be replaced.

It is very important to remove the diffuser from the tank and to disconnect it as soon as it is no longer in use.

Did this help to solve your problem?

Yes No

Home

Page 48: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Defective dosing module.

Contact our after-sales service for a replacement.

Did this help to solve your problem?

Yes No

Home

Page 49: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Enter a tank volume in hL and a dose in mL/L/month.

Did this help to solve your problem?

Yes No

Home

Page 50: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Often it is an inputting error. Enter the desired values again.

Did this help to solve your problem?

Yes No

Home

Page 51: Next Home µox Compact troubleshooting guide Start the slide show then click Next

The machine has lost its programme, contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 52: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Contact our after-sales service.

Did this help to solve your problem?

Yes No

Home

Page 53: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Thank you for using Vivelys' online help.We hope that we have met your expectations, and are available to help if you need

any further information.

Please do not hesitate to contact us by email or telephone:

+33 (0)4 67 85 68 [email protected]

Home

Page 54: Next Home µox Compact troubleshooting guide Start the slide show then click Next

Do you need more help?Were you unable to solve the problem using the online troubleshooting guide?

Please contact our technical support team by email or telephone:

+33 (0)4 67 85 68 [email protected]

Home