ISL Temperature Measurement and Control Page - 1
ANSI CALIBRATIONType E, CHROMEL® (+) Constantan (-), has the highest EMFoutput of any standardized metallic thermocouple. If used unprotected,Type E wires are not subject to corrosion at subzero temperatures.They can be used in inert, oxidizing or reducing atmospheres. Becausethey cover a wide range with a single calibration curve, Type Ethermocouples are preferred for computer applications.
Type J. Iron (+) Constantan (-), is the most commonly usedcalibration. It is suitable for use in a vacuum, inert, oxidizing or reducingatmosphere. If unprotected the iron wire may be attacked by ammonia,nitrogen and hydrogen atmospheres. In sub-zero temperatures the ironwire may rust or become brittle.
Type T. Copper (+) Constantan (-), is commonly used for subzero to700°F temperature. Preferred to Type J for sub-zero applicationsbecause of copper’s higher moisture resistance, as compared to iron. Ifunprotected, it will still function in a vacuum, inert, oxidizing or reducingatmosphere.
Type K, CHROMEL® (+) ALUMEL® (-), is generally used tomeasure high temperature to 2300°F. It should not be used for accuratetemperature measurements below 900°F after prolonged exposureabove 1400°F. If unprotected it can be used only in inert or oxidizingatmospheres. It has a short life in alternately oxidizing and reducingatmospheres and in reducing atmospheres, particularly in the 1500 to1850°F range.
Type S. Platinum-10% Rhodium (+) vs. Platinum (-).Type R. Platinum-13% Rhodium (+) vs. Platinum (-). Thesethermoelements should always be protected from contamination byreduced oxides, metallic vapors or other oxides at high temperatures.Platinum protective sheaths or Alumina are used at temperatureswhich preclude the use of base metal sheaths. Insulation should besilica free to prevent contamination. Type S is frequently used forcalibration and checking. Type R has a slightly greater sensitivity andconsequently is used more frequently in industrial applications.
Type B. Platinum-30% Rhodium (+) vs. Platinum-6% Rhodium (-).For use between 1000 and 3175°F. Intended to prevent the problemsexperienced with types S and R such as (1) weakening of the pureplatinum leg due to excessive grain growth and (2) calibration shift dueto pure platinum wire picking up rhodium volatilized from the alloy wireat 1500°C. The flatness of the temperature-millivolt curve at normalreference junction ambient temperature permits the use of copperextension wire.
Type W. Tungsten-5% Rhenium (+) vs. Tungsten-26% Rhenium (-).These thermoelements possess excellent stability at temperatures inthe 3000°F to 4000°F range. For use at high temperatures a protectiveatmosphere must be provided, such as hydrogen, inert gas or vacuum.They are extremely sensitive to mechanical damage and should behandled carefully to prevent breakage.
JUNCTION TYPE
GroundedJunction (G)
The junction is welded to the sheath. Junction is completely sealedfrom any contamination and response time is typically 1/3 quickerthan ungrounded junction.
UngroundedJunction (U)
The junction is physically and electrically insulated from the sheath.Used in cycling or high electrical noise applications.
ExposedJunction (E)
The junction is extended one diameter beyond sheath for fastresponse time. Use where mechanical damage will not occur.
Th
erm
al E
MF
(m
V)
Temperature (°C)
ThermocouplesTECHNICAL DATAT.D.
Thermal EMF (mV)
Page - 2 ISL Temperature Measurent and Control
TECHNICAL DATARTD’S (Resistance Temperature Detectors)
An RTD sensing element consists of a wire coil or deposited film ofpure metal. The element’s resistance increases with temperature in aknown and repeatable manner. RTD’s exhibit excellent accuracy over awide temperature range and represent the fastest growing segmentamong industrial temperature sensors. Their advantages include:
· Temperature range: Models in this catalog cover temperatures from -436 to 1582°F (-260 to 850°C).
· Repeatability and stability.· Sensitivity: The voltage drop across an RTD provides a much larger output than a thermocouple.· Linearity: Platinum and copper RTD’s produce a more linear re- sponse than thermocouples or thermistors. RTD non-linearities can be corrected through proper design of resistive bridge networks.·Low system cost: RTD’s use ordinary copper extension leads and require no cold junction compensation.· Standardization: Manufacturers offer RTD’s to industry standard curves, most commonly 100Ω platinum with a Temperature Coefficient of Resistance of 0.00385 Ω/Ω/°C in two tolerance classes (Class A: +0.06% @ 0°C, Class B: +0.12% @ 0°C).
Wire Wound ElementThe standard RTD element used in ISL’s probe assemblies are madeof 99.99% pure platinum wire wound about a ceramic or glass coreand hermetically sealed within a ceramic or glass capsule.Platinum wire was chosen as it best meets the needs of precisionthermometry. It resists contamination, can be highly refined and ismechanically and electrically stable. This provides for closeinterchangeability between elements with negligible drift and error withage. On special request, ISL can make available RTD elements madewith other wire materials.
Thin Film ElementMade by platinum being deposited as a film on a substrate and thenencapsulated. This method allows for the production of small fastresponse, accurate sensors.
RTD lead configurationsBecause an RTD is a resistance type sensor, resistance introduced byconnecting extension wires between the RTD and control instrumentwill add to readings. Furthermore, this additional resistance is notconstant but increases with ambient temperature.You can reduce leadwire error by:
· Using larger gauge extension wires.· Specifying an RTD with greater sensitivity.· Employing a 3 or 4-wire resistance cancelling circuit.· Using a 2-wire current transmitter.
Calculating leadwire error “E”Leadwire Ω (from table below) x Wire Length(ft)
Element Sensitivity (Ω/Ω/°C)
Lead configuration 1 provides one connection to each end of thesensor. This construction is suitable where the resistance of the run ofthe lead wire may be considered as an additive constant in the circuit,and particularly where the changes in lead resistance due to ambienttemperature changes may be ignored.
Lead configuration 2 provides one connection to one end and two to theother end of the sensor. Connected to an instrument designed toaccept three wire input, compensation is achieved for lead resistanceand temperature change in lead resistance. This is the most commonlyused configuration.
Lead configuration 3 provides two connections to each end of the sensor.This construction is used for measurements of the highest precision.
Lead configuration 4 is similar to Lead configuration 3 except that aseparate pair of wires is provided as a loop to provide compensationfor lead resistance and ambient temperature changes in leadresistance.
Style 1
Style 2(standard)
Style 3
Style 4
Comparison of Thermocouples and Pt RTD’s
CriterionStandard-Grade Thermocuple ISA J ISA K
StandardDIN Pt RTD
Accuracy@ 0°C100°C500°C
Time ConstantTip SensitiveUpper Temp.
+2.2°C+2.2°C+3.9°C
1.7 sec*Yes
870°C
+2.2°C+2.2°C+3.9°C
1.7 sec*Yes
1260°C
+0.3°C+0.5°C+3.0°C
5.0 sec**Yes***
850°C***
*1/4” OD probe, grounded junction **1/4” OD probe*** Contact Factory
T.D.
E=
Leadwire AWG Ohms/ft. at 25°C
12141618202224262830
0.00160.00260.00410.00650.01030.01650.02620.04180.06660.1058
=(error in °C)
ISL Temperature Measurement and Control Page - 3
Sensors for the plastics industry
Thermocouples and RTD’s used in the plastic industry are usually spring loaded against the mea-suring surface using a bayonet cap, bayonet adapter together and fixed fittings. Size and configu-rations vary according to application but serve the same purpose.
If what you use is not shown, call our sales and engineering staff.
Catalog # Description Bayonet lock cap
Compression nut andsleeve
1/8” bore
Bayonet Adapters and Fixed Fittings1
How to orderOn the fixed bayonet style (H44) it is veryimportant that the immersion depth “U”includes the hole depth plus the immersiondepth. Also, specify Thread size.Example: H44 - 1 1/4”- 1/8 NPT
1/8” ADJUSTABLE W/ BRASS FERRULE1/8” ADJUSTABLE W/ NYLON FERRULE1/8” ADJUSTABLE W/ STAINESS FERRULE
Adjustable Bayonet Fitting for 1/8” dia. probe
U
7/16”
H439 H439-NY H439-SS
ISL Temperature Measurement and ControlPage - 4
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
WT Style Flexible Wire
♦ ♦ ♦ ♦ ♦ Fast response time♦ ♦ ♦ ♦ ♦ Low cost♦ ♦ ♦ ♦ ♦ Multiple uses♦ ♦ ♦ ♦ ♦ Exposed or enclosed junction
♦ ♦ ♦ ♦ ♦ Wire: Thermocouple grade, special limits♦ ♦ ♦ ♦ ♦ Insulation: Teflon® , Fiberglass or PVC♦ ♦ ♦ ♦ ♦ Optional stainless steel overbraid available♦ ♦ ♦ ♦ ♦ Wide range of insulation materials to meet any temperature range, high humidity and corrosive environments
U 2”
1- 20 AWG 2- 24 AWG 3- 30 AWG 4- 36 AWG X- Other ( Specify)
WIRE SIZE
JUNCTION E - Exposed I - Isolated
LEADWIRE Standard Overbraid
Fiberglass G O Teflon® T B PVC P V
W T 0 0
LEADWIRE LENGTH ( inches )
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
Specify “U” inches (001-999)
Page - 5
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ISL Temperature Measurement and Control
RT Style Ring Terminal
♦ ♦ ♦ ♦ ♦ Surface temperature measurement up to 800° F♦ ♦ ♦ ♦ ♦ Easily attached to any flat or curved surface with a screw♦ ♦ ♦ ♦ ♦ Large variety of ring sizes
♦ ♦ ♦ ♦ ♦ High temperature “SS” ring terminals♦ ♦ ♦ ♦ ♦ Standard 20 AWG, Teflon® or Fiberglass T/C wire with optional SS overbraid♦ ♦ ♦ ♦ ♦ Specify screw size and temperature
U 2”
brazed
U - Ungrounded G - Grounded
JUNCTION
SCREW SIZE- Hole Dia. ( In )
Standard Overbraid Fiberglass G O Teflon® T B
LEADWIRE LENGTH ( inches)
LEADWIRE
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
A - No.6 0.144 B - No.8 0.169 C - No.10 0.196 D - 1/4 0.266 E - 3/8 0.390 F - 1/2 0.525
Specify “U” inches (001-999)
R T 0 0
ISL Temperature Measurement and ControlPage - 6
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ST Style Brass Shim
♦ ♦ ♦ ♦ ♦ Low profile; typically - .012” thick♦ ♦ ♦ ♦ ♦ Easily cemented to any surface
♦ ♦ ♦ ♦ ♦ Standard shim material is brass or stainless; other materials are available to meet any specific requirement♦ ♦ ♦ ♦ ♦ Thermocouple wire used: PVC, Teflon and Fiberglas 20 - 36 AWG♦ ♦ ♦ ♦ ♦ Specify: shim size, calibration and length
S T 0 0
A - 20 AWG B - 24 AWG C - 30 AWG X - Other ( Specify )
WIRE SIZE
LEADWIRE
LEADWIRE LENGTH ( inches )
.12”Thick
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
Specify “U” inches (001-999)
U 2” .75”
.75
SHIM MATERIAL
S - StainlessB - Brass
Standard Overbraid Flex Armor Fiberglass G O 1 Teflon® T B 3 PVC P V 2
Page - 7
10
ISL Temperature Measurement and Control
EX Style Flexible Extension
Features
U
U
CALIBRATION
TERMINATION “A”
LEAD WIRE
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
S - Standard 1 1/2” split leadsL - 1 1/2” split leads with spade lugsP - Standard plugJ - Standard jackM- Miniature plugF - Miniature jackX - Other ( Specify )
Standard Overbraid Flex Armor PVC P V 2 Fiberglass G O 1 Teflon® T B 3
LEADWIRE LENGTH ( Inches )
Specify “U” inches (006-999)
TERMINATION “B”
SPECIAL If none , Enter “0” If required (Specify)
E X 0 0 0
Termination “A” Termination “B”
Stainless Steel Armor
Stainless Steel Overbraidor withouth Overbraid
♦ ♦ ♦ ♦ ♦ Quick disconnect♦ ♦ ♦ ♦ ♦ Full protection of extension wire against harsh industrial environments♦ ♦ ♦ ♦ ♦ Fully compensated thermocouple materials to eliminate reading errors
♦ ♦ ♦ ♦ ♦ Extension wire: Special limits, 20 AWG, stranded, PVC, Fiberglass and Teflon insulation♦ ♦ ♦ ♦ ♦ SS flexible armor, SS overbraid or standard♦ ♦ ♦ ♦ ♦ Custom design to any specification
1
S -L -P -J -
M -F -X -
Standard 1 1/2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
General Specifications
ISL Temperature Measurement and ControlPage - 8
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
PT Style Tube and Wire
♦ ♦ ♦ ♦ ♦ Low cost♦ ♦ ♦ ♦ ♦ Quick delivery♦ ♦ ♦ ♦ ♦ Wide temperature range
♦ ♦ ♦ ♦ ♦ Sheath material is 316 SS♦ ♦ ♦ ♦ ♦ Cold - End epoxy sealed or crimped♦ ♦ ♦ ♦ ♦ Select from variety of T/C wire to meet the operating temperature and environment specifications
P T
A
1 3/8”
A 2”
2” U
U
TUBE DIA. 4 - 1/8” 6 - 3/16” 8 - 1/4”
JUNCTION
0 - Straight 4 - 45 deg. Bend 9 - 90 deg. Bend
BEND ANGLE
IMMERSION LENGTH “A” ( inches ) A- 2 D- 3.5 G- 5 J- 6.5 L- 8 O- 9.5 R- 12
B- 2.5 E- 4 H- 5.5 J- 7 M- 8.5 P- 10 x- Other (Specify) C- 3 F- 4.5 I- 6 K- 7.5 N- 9 Q- 11
LEADWIRE
LEADWIRE LENGTH ( inches )
Standard Overbraid Flex Armor Fiberglass G O 1 Teflon® T B 3
J - Type J K - Type K T - Type T E - Type E P - 100 Ω Thin film RTD*
Specify U inches (001-000)
Grounded Ungrounded Single G U Duplex D C
Page - 9
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ISL Temperature Measurement and Control
♦ ♦ ♦ ♦ ♦ Spring loaded to assure contact with measured surface♦ ♦ ♦ ♦ ♦ Fixed bayonet fitting for quick twist-lock installation♦ ♦ ♦ ♦ ♦ Low cost & quick delivery
♦ ♦ ♦ ♦ ♦ Tube: .188 O.D., 316 SS (std)♦ ♦ ♦ ♦ ♦ Steel spring, nickel plated brass cap♦ ♦ ♦ ♦ ♦ Wide temperature range♦ ♦ ♦ ♦ ♦ Wide variety of wire insulation materials♦ ♦ ♦ ♦ ♦ Fixed immersion length
FB Style Fixed Bayonet
F B 0
BEND ANGLE 0 - Straight 1 - 45 degree bend 9 - 90 degree bend X - Other (Specify)
IMMERSION LENGTH “A” A - 2 D - 3.5 G - 5 J - 6.5 B - 2.5 E - 4 H - 5.5 K - 7 C - 3 F - 4.5 I - 6 X - Other (Specify)
LEADWIRE LENGTH ( inches )
U 1 5/8”
3/4”
“A”
“A”
“A”1”
U
U 2”
Grounded Ungrounded Single G U Duplex D C
JUNCTION
Standard Overbraid Flex ArmorFiberglass G O 1Teflon® T B 3
LEADWIRE
J - Type J K - Type K T - Type T E - Type E P - 100 Ω Thin film RTD*
Specify “U” inches (001-999)
ISL Temperature Measurement and ControlPage - 10
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
J - Type J K - Type K T - Type T E - Type E P - 100 Ω Thin film RTD*
♦♦♦♦♦ Bayonet cap threads length of flex armor♦♦♦♦♦ Compression of flex armor loads tip♦♦♦♦♦ Fits standard bayonet adaptors
♦♦♦♦♦ .188” tube diameter- 316SS (std)♦♦♦♦♦ Leadwire available with Fiberglass or Teflon®
insulation♦♦♦♦♦ Standard terminations available
AA Style Adjustable Armor
A A 0
LEADWIRE LENGTH ( inches )
U
Grounded Ungrounded Single G U Duplex D C
JUNCTION
SHEATH DIAMETER
4 - 1/8” : 6 - 3/16” : 8 - 1/4”
TIP LENGTH
1 - 3/8” : 2- 1/2” :X= Specify
Stranded OverbraidFiberglass G OTeflon T B
LEADWIRE
Specify “U” inches (001-999)
3”
Page - 11
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ISL Temperature Measurement and Control
AS Style Adjustable Spring
♦♦♦♦♦ Variable adjustable sensor fits a largerange of hole depths
♦♦♦♦♦ Bends to any angle♦♦♦♦♦ Eliminates immersion depth variations
♦♦♦♦♦ .188” O.D. 316SS tube (std)♦♦♦♦♦ Stainless steel spring O.D. .250”♦♦♦♦♦ Standard terminations available
U
A S
J - Type J K - Type K T - Type T E - Type E P - 100 Ω Thin film RTD*
JUNCTION Grounded Ungrounded Single G U
SHEATH DIAMETER
4 -1/8” : 6 - 3/16” : 8 - 1/4”
TIP LENGTH
1 - 3/8” : 2- 1/2” :X= Specify
SPRING LENGTH1 -2 -3 -
6 inch spring8 inch spring12 inch spring
LEADWIRE LENGTH (inches)
Standard OverbraidFiberglass G OTeflon® T B
LEADWIRE
Spring length
Specify “U” inches (001-999)
2”
ISL Temperature Measurement and ControlPage - 12
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
MB Style Melt Bolt
E MBLength A
M B 1 0
J - Type JK - Type K
SHEATH DIAMETER4 - .125”6 - .188”
IMMERSION LENGTH “A”123
- Flush- 1/4”- 1/2”
45X
- 3/4”- 1”- Other (Specify)
Grounded Ungrounded Single G U Duplex D C
JUNCTION
MELT BOLT LENGTH
3 - 3 inch6 - 6 inch9 - 9 inch
SHEATH MATERIAL
1 - 304SS
♦♦♦♦♦ Designed for insertion directly intomolten plastic
♦♦♦♦♦ Available with various lead extensions,Flex Armor and SS Overbraid.
Shown with sheath extension.
♦♦♦♦♦ 304SS sheath and melt bolt♦♦♦♦♦ Melt bolt lengths are 3 inch, 6 inch and 9
inch
LEAD LENGTH “E”
Specify whole inches:(001- 999)
Page - 13
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ISL Temperature Measurement and Control
BT Style Nozzle
♦ ♦ ♦ ♦ ♦ Low profile and short depth♦ ♦ ♦ ♦ ♦ Rotating threaded screw♦ ♦ ♦ ♦ ♦ Wide temperature range♦ ♦ ♦ ♦ ♦ Available in standard and metric thread
♦ ♦ ♦ ♦ ♦ Threaded screw, all stainless steel♦ ♦ ♦ ♦ ♦ Steel spring for wire strain relief♦ ♦ ♦ ♦ ♦ Standard 24 AWG Wire
2”U
U
LEADWIRE LENGTH ( inches )
Specify “U” inches (001-999)
0
A - 1/4 - 20 UNF, B - 8/32 thread C - 10/32 thread D - 1/4 - 28 UNF X - Other (Specify)
BOLT SIZE
LEADWIRE
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
B T 0 0 0
Standard OverbraidFiberglass G OTeflon® T B
ISL Temperature Measurement and ControlPage - 14
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
BT Style Beaded Base Metal
♦ ♦ ♦ ♦ ♦ Low cost♦ ♦ ♦ ♦ ♦ High temperature up to 2300° F for type “K” thermocouple♦ ♦ ♦ ♦ ♦ Long life if properly protected with a thermowell or protection tube
♦ ♦ ♦ ♦ ♦ Standard thermocouple calibrations are; J & K♦♦♦♦♦ Standard wire: 8 and 14 AWG♦ ♦ ♦ ♦ ♦ Color coded for proper instrument hook-up♦ ♦ ♦ ♦ ♦ Select from oval or round insulators♦ Custom designs to meet any requirement,
contact factory
B - Bare (no insulators) O - Oval Ceramic R - Round Ceramic
JUNCTION
LENGTH ( Inches )
****
** Not Apply
2”
2 or 4 Hole Ceramic InsulatorsButt Weld
U
J - Type J K - Type K T - Type T E - Type E X - Other ( Specify )
Specify whole inches: 01 - 999
B T 0 0 0 0
1 - 08 AWG 2 - 14 AWG 3 - 20 AWG X - Other ( Specify )
WIRE SIZE
ELEMENT INSULATION
Style Twisted Butt Welded Welded Single T B Duplex D W
Color Coaded
Page - 15
Features
CALIBRATION
TERMINATION
SPECIAL If none , Enter “0” If required (Specify)
General Specifications
1
S -L -P -J -
M -F -X -
Standard 2” split leadsSplit leads with spade lugsStandard plugStandard jackMiniature plugMiniature jackOther (Specify)
ISL Temperature Measurement and Control
NB Style Noble Metal
♦ ♦ ♦ ♦ ♦ High temperature up to 3000° F♦ ♦ ♦ ♦ ♦ Most accurate of all thermocouples standard grade R & S: +/- 0.25 % reference grade R & S: +/- 0.1 % within range of 400° to 1400°C♦ ♦ ♦ ♦ ♦ Fast response time; typically under .3 Sec
♦ ♦ ♦ ♦ ♦ Standard grade platinum wire is provided unless reference grade is requested♦ ♦ ♦ ♦ ♦ High purity alumina 2 or 4 hole insulators♦ ♦ ♦ ♦ ♦ Standard termination is 1 1/2” leads with ceramic fish spine insulators and copper sleeves
1 1/2”
1/2”
CopperSleeve
Ceramic FishSpine Insulators
ExposedJunction
S - Type S ( Pt vs Pt/Rh ) R - Type R ( Pt vs Pt/Rh ) B - Type B ( Pt vs Pt/Rh ) W - Type W ( Tg vs Tg/Rh )
A - 24 AWG B - 30 AWG
ELEMENT SIZE
CONSTRUCTION S - Single D - Duplex LENGTH “U” ( Inches )
U
High Purity Alumina Insulator
U
Specify whole inches: 01 - 999
N B E 0 0 0
P - Standard PlugJ - Standard Jack
ISL Temperature Measurement and ControlPage - 16
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
SHEATH MATERIAL
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
JUNCTION
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
M A
MA Style Cut and Strip
U “L”
SPECIAL REQUIREMENT
“L” STRIP LENGTH
COMPRESSION FITTING
SHEATH LENGTH “U”
If required, Enter “X”and Specify “0” if not
X - 0 D - 2” A - 1/2” E - 2 1/2” B - 1” F - 3” C - 1 1/2”
Please Specify Thread Size
Specify in Whole Inches01 - 999
!!!!!Fast delivery!!!!!Standard sheaths!!!!!T/C material protected by sheath
!!!!!Available in standard or special limits (99.6%)MgO.
!!!!!Can be supplied with 0.020” to .375” dia.sheath.
*For noble metal price contact factory.
Page - 17ISL Temperature Measurement and Control
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
SHEATH MATERIAL
JUNCTION
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
ML Style Spring loaded
M L 0
SPECIAL REQUIREMENT
LEAD LENGTH “L”
“U” SHEATH LENGTH
Enter “U” Lengthin Whole Inches (0-999)
Enter Lead Lengthin Whole Inches(0-99)
If Required, Enter “Xand Specify. “0” If Not
U
!!!!!Spring loaded assures contact withbottom of well
!!!!!“L” length is flexible lead to allow forexpansion of well
!!!!!Spring is adjustable
!!!!!Standard or special limits of error - Hardpack MgO insulation
!!!!!Available in 1/8”, 3/16”, or 1/4” O.D.!!!!!High temp spring
Flexible lead “L”
ISL Temperature Measurement and ControlPage - 18
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
SHEATH MATERIAL
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
JUNCTION
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
M C 0 0
MC Style Plug Termination
U
P - Standard PlugJ - Standard JackM - Miniature PlugF - Miniature JackHP- High Temp. Standard PlugHJ- High Temp. Standard Jack
COLD END TERMINATION
SPECIAL REQUIREMENT
“U” SHEATH LENGTH
If required, Enter “X” and Specify. “0” if Not
Specify in Whole Inches (0-999)
!!!!!Quick disconnect thermocouple assemblies!!!!!Standard or miniature connectors!!!!!Standard connector rated to 425° F. Also, avillable with high temperature
connnector to 900° F.
!!!!!Available in standard or special limits oferror
!!!!!Hard packed MgO.!!!!!Sheath diameter available from .010” to .375”!!!!!Available sheath temperatures to 2880° F with Pyrosyl “B”
Mini Plug & Jack dual elemt available only formax. diameter 0.125”.
Notes:
*For noble metal price contact factory.
Page - 19ISL Temperature Measurement and Control
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
SHEATH MATERIAL
JUNCTION
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
M F
MF Style Metal Transition
L 2” 1.8” U
COLD END TERMINATION
If Required, Enter “X”and Specify. “0” If Not
“U” Sheath Length
SPECIAL REQUIREMENT
S - Standard 1 1/2” LeadsP - Standard PlugJ - Standard JackM - Miniature PlugF - Miniature JackHP- High Temp. Standard PlugHJ- High Temp. Standard Plug
LEAD LENGTH “L”Specify Length in Whole Feet (01-99
LEADWIRE CONSTRUCTION
Specify Length in Whole Feet (0-99)
!!!!!Rugged stainless steel transition!!!!!Flexible leadwire, Stranded or SolidAlso, availlable with Plug & Jack cold endterminetion!!!!!Diameters available from .010” to .375”
!!!!!Hard packed MgO insulation!!!!!1/4” O.D. transition for smaller diameter!!!!!3/8 O.D. transition for larger dia. sheath!!!!!400° F epoxy
Solid: Standard OverbraidFiberglass F BTeflon T EPVC P VStranded:Fiberglass S OTeflon® H UPVC M R
*For noble metal (R & S) price contact factory.
ISL Temperature Measurement and ControlPage - 20
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
SHEATH MATERIAL
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
JUNCTION
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
M E
ME Style Large Transition
2” L 1.8” U
SPECIAL REQUIREMENT
S - Standard 1 1/2” LeadsP - Standard PlugJ - Standard Jack
LEAD LENGTH “L”
Enter Length in Whole Feet01 to 99
LEADWIRE CONSTRUCTION
Solid Flex ArmorFiberglass 1Teflon® 3PVC 2
Enter Length in Whole Feet
SHEATH “U” LENGTH
COLD END TERMINATION
If Required, Enter “X”and Specify. “0” If Not
!!!!!Large transition for flex armor!!!!!PVC or Teflon coated for added protection!!!!!Crush resistant
!!!!!5/16” O.D. x 0.44” I.D. transitionaccomadates standard 3/16” I.D. x 9/32” O.D.SS flex or 1/4” I.D. x 11/32” O.D. flex!!!!!Maximum continuous operating temperatureof 400° F for large transition. Higher
ranges available
5/16” O.D.
*For noble metal (R & S) price contact factory.
Page - 21ISL Temperature Measurement and Control
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
SHEATH MATERIAL
JUNCTION
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
2” U
Double threadedHex FittingDF4 shown
2”1 1/4”
U
Epoxy Seal Process ConnectionSF6 shown
MP Style Fixed Fitting
M P
2”
SHEATH LENGTH “U”
Enter Length in Feet 01 to 99
LEAD TYPE
F - FiberglassT - Teflon®
P - PVC
If Required, Enter “X”and Specify. “0” If Not
SPECIAL REQUIREMENT
!!!!!Process connection and transition all in one!!!!!Available in 1/4”, 1/2” and 3/4” NPT connection!!!!!Epoxy seal, 400°F
!!!!!Standard or special limits of error hardpacked MgO insulation
!!!!!2” leadwire is stranded for flexibility!!!!!Stainless steel process fittings
Note:
Lead lengths over 2” contact factory
!!!!!Use in thermowells or directly in process
PROCESS CONNECTION
SF4 - 1/4” NPTSF6 - 1/2” NPTSF8 - 3/4” NPT
DF2 - 1/4” NPTDF4 - 1/2” NPTDF6 - 3/4” NPT
*For noble metal (R & S) price contact factory.
ISL Temperature Measurement and ControlPage - 22
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
SHEATH MATERIAL
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
JUNCTION
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
MS Style Spring Loaded
M S
U
Spring Loaded Adjustable1/2” x 1/2” NPT304 SS Hex Fitting
PROCESS CONNECTION
SHEATH LENGTH “U”
Enter Length in WholeFeet 01 to 99
SL4 - .188”O.D. 1/2” NPTSL6 - .250”O.D. 1/2” NPT
LEAD TYPE
Fiberglass - FTeflon® - TPVC - P
If Required, Enter “X”and Specify. “0” If Not
SPECIAL REQUIREMENT
!!!!!Adjustable spring loaded fitting ensures directsurface contact
!!!!!Fast response, great as bearing sensor!!!!!Use with well and protection head
!!!!!1/2” x 1/2” NPT hex fitting, 304SS!!!!!Fitting adjustable spring travel of 1/2”
3”
*For noble metal (R & S) price contact factory.
Page - 23ISL Temperature Measurement and Control
Features General Specifications
Std. Limits J K T E N (R S)* Spl. Limits 1 2 3 4 5 6 7
CALIBRATION
SHEATH DIAMETER
1 - 304 SS 3 - 316 SS 2 - 310 SS 6 - INC. 600
SHEATH MATERIAL
JUNCTION
2
4 - .063”5 - .125”6 - .188”
7 - .250”8 - .313”9 - .375”
Grounded Ungrounded Exposed Single G U E Dual Common D C P Dual Isolated - S T
!!!!!Connection heads provide superior dust andmoisture resistance
!!!!!Heads are available in Aluminum, Cast Iron,Explosion proof and Polypropylene
!!!!!Optional temperature transmitters canmount inside connection head
M H
Enter “U” Lengthin Whole Feet
U
MH Style Head Termination
SPECIAL REQUIREMENT
CONNECTION HEAD TYPE
1 - Standard Cast Aluminum2 - Standard Cast Iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
If Required, Enter “X”and Specify. “0” If Not
“U” SHEATH LENGTH
!!!!!Sheath diameter available from .063” to.375”
PROCESS CONNECTION
SF4 - 1/4” NPTSF6 - 1/2” NPTSF8 - 3/4” NPT
DF2 - 1/4” NPTDF4 - 1/2” NPTDF6 - 3/4” NPT
!!!!!Hex fittings are made of stainless steel
*For noble metal (R & S) price contact factory.
U
ISL Temperature Measurement and ControlPage - 24
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
5 - .125” 6 - .188” 7 - .250”
SHEATH O.D.
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
3 RE Style Tube & Wire
R E
UL 1 1/2”
!!!!!Standard industrial leads available infibergass, teflon and PVC insulation
!!!!!Accurate and dependable reading!!!!!Economical and quick delivery
!!!!!Available in diameters .125” to .250”!!!!!Tube and wire max temperatures 300° C!!!!!Epoxy sealed to resist moisture
SPECIAL REQUIREMENT
If Required Enter “X”and Specify. “0” If Not
COLD END TERMINATION
C - Three Pin PlugE - Three Pin JackS - Standard 1 1/2” Leads
LEADWIRE LENGTH “L”
Enter “L” Length inWhole Feet 01 to 99
LEADWIRE CONSTRUCTION*
Standard OverbraidFiberglass F BTeflon T EPVC M R
* *
Page - 25ISL Temperature Measurement and Control
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
5 - .125” 6 - .188” 7 - .250”
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
3
!!!!!Replacement spring loaded element forwell assemblies
!!!!!Epoxy sealed against moisture
!!!!!Sheath diameters .188” and .250”!!!!!Mineral insulated available for applications
above 300°C!!!!!Teflon® or fiberglass insulation
SPECIAL REQUIREMENT
If Required Enter “X”and Specify. “0” If Not
LEADWIRE LENGTH “L”
LEADWIRE CONSTRUCTION
T - Teflon ( 400 F )F - Fiberglass ( 800 F )
Enter “L” Length inWhole Feet 01 to 99
U
Epoxy sealed
RT Style Spring loaded, replacement
*
** Contact factory for mineral insulated 650°C
3” flexable leads
R T
ISL Temperature Measurement and ControlPage - 26
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
5 - .125” 6 - .188” 7 - .250”
SHEATH O.D.
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
3
R C 0 0 0
U
SPECIAL REQUIREMENT
If Required, Enter “X”and Specify. “0” If Not
COLD END TERMINATION
C - Three Pin PlugE - Three Pin JackW - Plug & Jack
RC Style Plug Termination
Optional 3 Pin Jackor Code “W”
!!!!!Connector available in standard or mini sizes!!!!!Available with mating Jack to allow for quick
connect / disconnect
!!!!!Available in diameters of .125” to .250”!!!!!Dual element available!!!!!Connector Temperature to 400°F
**Contact factory for mineral insulated (650°C).
N/A
Page - 27ISL Temperature Measurement and Control
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
5 - .125” 6 - .188” 7 - .250”
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
3
R L
SPECIAL REQUIREMENT
If Required Enter “X”and Specify. “0” If Not
COLD END TERMINATION
C - Three Pin PlugE - Three Pin JackW - Plug and JackS - Standard 1 1/2” Leads
LEADWIRE LENGTH “L”
Enter “L” Length inWhole Feet 01 to 99
LEADWIRE CONSTRUCTION
L 1.8”” U
RL Style Metal Transition
!!!!!Spring strain relief protects lead wire againstsharp bends in the transition area
!!!!!Stainless steel transitions are crimped tosheath and epoxy filled for temperatures upto 400°F
!!!!!Available in .125” to .250” sheath O.D.
Standard OverbraidFiberglass F BTeflon T EPVC M V
* *
ISL Temperature Measurement and ControlPage - 28
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
5 - .125” 6 - .188” 7 - .250”
SHEATH O.D.
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
3 RD Style Fixed Fitting
3” 2” U
If Required Enter “X”and specify. “0” If Not
LEADWIRE CONSTRUCTION
SPECIAL REQUIRMENTS
R D
!!!!!Process connection designed for use withconnection head
!!!!!Mineral insulation available for tempera-tures above 300°C
!!!!! .125” to .250 O.D. sheaths available!!!!!Stainless steel process fittings available in
sizes 1/4” to 3/4” NPT
PROCESS FITTINGPART NO. MALE N.P.T.
DF2 1/4” DF4 1/2” DF6 3/4”
Fiberglass FTeflon TPVC P
**Contact factory for mineral insulated (650°C).
* *
N/A
Page - 29ISL Temperature Measurement and Control
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
5 - .125” 6 - .188” 7 - .250”
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
3 RS Style Spring Loaded
3” U
R S
LEADWIRE CONSTRUCTION
If Required Enter “X”and specify. “0” If Not
SPECIAL REQUIRMENTS
SPRING LOADED PROCESS BUSHING
PART NO. TUBE O.D. MALE NPT
SL4 .188” 1/2” SL6 .250” 1/2”
!!!!!Spring loaded element ensures contact withwell bottom
!!!!! .188” & .250” sheath diameters!!!!! 1/2”x 1/2” process threads
StandardFiberglass FTeflon T
**Contact factory for mineral insulated (650°C).
N/A
ISL Temperature Measurement and ControlPage - 30
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
5 - .125” 6 - .188” 7 - .250”
SHEATH O.D.
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
3 RH Style Connection Head
U
R H D 4
SPECIAL REQUIREMENT
CONNECTION HEAD TYPE
1 - Standard Cast Aluminum2 - Standard Cast Iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
If Required, Enter “X”and Specify. “0” If Not
!!!!!Connection heads provide superior dustand moisture resistance
!!!!!Screw top and flip top covers available
!!!!!Sheath sizes of .125” to .250” O.D.!!!!!1/2”x 1/2” NPT stainless hex fitting!!!!!Connection heads available in aluminum, cast
iron and polypropylene
“L” LENGTH FRACTIONAL
0 - 0 4 - 1/2”1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 7/8”3 - 3/8”
**Contact factory for mineral insulated (650°C).
* *
N/A
Head mounted
temperature tran
smitters
also available
Page - 31ISL Temperature Measurement and Control
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
5 - .125” 6 - .188” 7 - .250”
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
3 RH Style Connection Head
L
R H
CONNECTION HEAD TYPE
1 - Standard Cast Aluminum2 - Standard Cast Iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
S 6
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If Not
!!!!!Connection head provides superior dust andmoisture resistance
!!!!!Designed for use with compression fittings
!!!!!Sheath construction available in low temp.max 300°C and mineral insulated 650°C max
!!!!!Connection heads available in Aluminum,cast iron and polypropylene
Head mounted
temperature tran
smitters
also available**Contact factory for mineral insulated (650°C).
* *
N/A
0
ISL Temperature Measurement and ControlPage - 32
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “U”
Specify Whole Feet
5 - .125” 6 - .188” 7 - .250”
SHEATH O.D.
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 A 316 SS 2 B Inconel 600 3 C
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
3
CONNECTION HEAD TYPE
1 - Standard Cast Aluminum2 - Standard Cast Iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If Not
RB Style Spring Loaded
U
R B 0 0
1/2” NPT
!!!!!Spring loaded fitting ensures contact withbottom of well
!!!!!Connection head provides superior dustand moisture resistance
!!!!! .188” & .250” sheath diameters available!!!!!1/2” of spring loaded travel!!!!!1/2” NPT stainless steel process connection
**Contact factory for mineral insulated (650°C).
N/A
0
Page - 33ISL Temperature Measurement and Control
Features
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
SHEATH LENGTH “L”
Specify Whole Inches
3- 316ss
SHEATH MATERIAL MAX 260° C (500° F)
General Specifications
4RN Style Connection Head
3” U
R N
7 - 250”
7 3
CONNECTION HEAD TYPE
1- Polypropylene2- Epoxy coated Aluminum
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If Not
T - TRI-CLAMP®
SANITARY CAP STYLE
A -1 1/2” D -3”B -2” X -Other (Specify)C -2 1/2”
SANITARY CAP SIZE/ TUBE O.D.
!!!!!3-A certified for sanitary clean-in-placeapplications
!!!!!Epoxy coated or polypropylene head resistsharsh washdown chemicals
!!!!! .250” O.D. 316 ss sheath
®Tri-Clamp is a registeredtrademark of Tri-Clover, Inc.
0
ISL Temperature Measurement and ControlPage - 34
Features General Specifications
STANDARD PT-100 RTD .00385CLASS WIRE WOUND THIN FILM
CLASS A A FCLASS B B M
SHEATH O.D.
100 Ohm Single 2 3 4 100 Ohm Duplex D L F
ELEMENT 2 Wire 3 Wire 4 Wire
SHEATH LENGTH “L”
Specify Whole Inches
3- 316ss
SHEATH MATERIAL MAX 260° C (500° F)
4
7- .250”
7 3
LEADWIRE LENGTH “E”
Enter “E” Length inWhole Feet 01 to 99
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If Not
RY Style Teflon Lead
A -1 1/2” D -3”B -2” X -Other (Specify)C -2 1/2”
SANITARY CAP SIZE/ TUBE O.D.
R Y
!!!!!3-A certified for sanitary clean-in-placeapplications
!!!!!FEP Teflon® flexible extension
!!!!! .250” O.D. sheath, polished 316ss sheath andsanitary cap
!!!!!100Ω Ω Ω Ω Ω 2 , 3 and 4 wire circuits available
UL
T - TRI-CLAMP®
SANITARY CAP STYLE
ISL Temperature Measurement and Control Page - 35
Section 4 Thermistors
What’s a Thermistor ?The Word Thermistor is derived from THERMally sensitive resISTOR .there are two major types: NTC and PTC.NTC Thermistors have a Negative Temperature Coefficient of resistance.PTC Thermistors have a Positive Temperature Coefficient of resistanceNTC Thermistors exhibit a decrease in electrical resistance with increasing temperature.PTC Thermistors exhibit an increase in electrical resistance with increasing temperature.Thermistors are actually classified as Ceramic Semiconductors. Manufacturing precisionThermistors involves advanced ceramics technology, solid state chemistry and electronics.Depending on the materials and methods of fabrication, they are generally used inthe temperature range of - 50°to 150°C and up to 300°C for some glass encapsulated units,however their Temperature vs Resistance curve is Non-Linear.The resistance value of Thermistors are typically referenced at 25°C ( R25 ). For most applica-tions R25 values are between 100 ohms and 100 Kohms.
Fig. 1 Thermistor Probe Typical Construction
At Eustis Co. Thermistor probes can be custom built to meet most demandingapplications, whether your concern is: Size, Interchangeability, Response Time Ruggedness, Response Time or Temperature, we can build It.
Applications:
* Automotive* HVAC* Biomedical* Military / Aerospace* Food Processing* Electronics
Sensor Manufacturing Capabilities
O.D.
1”MIN. IMMERS.
OVERALL SHEATH LENGTHCONNECTING WIREEXTENSION LENGTH
CABLE END CLOSURE HOUSINGTHERMISTOR ELEMENT(BEAD TYPE SHOWN)
4
Page - 36 ISL Temperature Measurent and Control
Portable Probes
!!!!!Probes available in thermocoupleRTD and thermistor elements
!!!!!Plastic, Teflon or SS handle available!!!!!A variety of probe tips available,
standard immersion tip shown below
U 4”Please Specify
!!!!!Probe diameter available in standard.125” O.D. to .250” O.D.
!!!!!PVC cord with molded mini plug.
UPlease Specify
4”
P H
CALIBRATION
J K T (x for other)
SHEATH MATERIAL1-3-6-
304SS316SSINC. 600
SHEATH O.D.
5-6-7-
.125”
.188”
.250”
Piercing tip - Used formeasuring temperatureswhere penetration intofrozen or bulk foods isrequired.
Air tip - For fast accuratemeasuring air or gas.
Sheath length “U”
Specifiy Whole inches
BEND ANGLE0-4-5-
Straight45°90°
TIP STYLE
A-B-C-
FlatAirPiercing
T-C-
Straight cordCoil cord
LEAD
(bends are made at 6” “L”)
0 0 0
If required, enter “X” andspecify. “0” if not.
SPECIAL REQUIREMENT
4
ISL Temperature Measurement and Control Page - 37
Section 5 Industrial Assemblies
INDUSTRIAL THERMOCOUPLE & RTD ASSEMBLIES
To order a complete assembly from thissection:
Step 1: Determine the part number for sensingelement.
Step 2: Determine Part number of Thermowell fromsection .
Example: Sensor, Extension & Connection head.
♦♦♦♦♦ Part Number:♦♦♦♦♦ Mineral insulated♦♦♦♦♦ Type J♦♦♦♦♦ 1/4” O.D.♦♦♦♦♦ 304 stainless steel♦ Ungrounded junction♦ As required element length♦ Spring loaded♦ Nipple-Union-Nipple extension♦ General purpose Aluminum head♦ No special requirements
Example: Thermowell
Part Number: TT300A045030
♦♦♦♦♦ Threaded, tapered thermowell♦ 3/4” NPT Process connection♦ .260” bore♦ 4 1/2” immersion length “U”♦ No lag♦ 316 stainless steel♦ No special requirements
ConnectionHead
ExtensionAssembly(type 1)
Sensorelement
Thermowell
5
Page - 38 ISL Temperature Measurent and Control
Industrial MW Style - T/C
M W
!!!!! Industrial, mineral insulated assemblies foruse with thermowells
!!!!!Seven varieties of connection heads!!!!!Spring loading available
Nom. 6” Nom. 2”
U
U
U
U
Type 1
Type 2
Type 3
Type 4
Nom. 4”
Std. Limits J K T E N Spl. Limits 1 2 3 4 8
CALIBRATION
5 - .125” X- Other 6 - .188” 7 - .250”
SHEATH DIAMETER
1 - 304 SS 4 - 446 SS 2 - 310 SS 5 - 347 SS 3 - 316 SS 6 - INC. 600
SHEATH MATERIAL
Grounded Ungrounded Single G U Dual Common D C Dual Isolated - S
JUNCTION
ELEMENT LENGTH “U”When ordering a complete assemblyincluding Thermowell, specify “TR”otherwise, specify “U” in whole inches
1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 3/4”3 - 3/8” 7 - 7/8”4 - 1/2” O - N/R (complete assy)
“U” IN FRACTIONAL INCHES
1 - Nipple-Union-Nipple 2 - Nipple-Union 3 - Nipple 4 - Head Only
EXTENSION
1- General Purpose Aluminum2 - General Purpose Cast Iron3 - Heavy Duty Aluminum4 - Heavy Duty Cast iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
Special Requirementsenter “X” and Specify
Y - YESN - NO
SPRING - LOADED
Thermowell P/NSee Section 6
5
ISL Temperature Measurement and Control Page - 39
B W
Industrial BW Style - T/C
!!!!!Beaded elements for use with thermowells!!!!!Available in four extension configurations
!!!!!Seven conneciton head varieties to choosefrom
!!!!!Element size available 14 AWG and 8 AWG!!!!!Noble metal thermocouples are supplied with
Alumina insulators, 24 AWG wire is standard
CONNECTION HEAD
UU4” NOM.
UU
6” NOM. 2” NOM.
Type 1
Type 2
Type 3
Type 4
Std. Limits J K T E N (S R B)* Spl. Limits 1 2 3 4 8 3 5 6
CALIBRATION
ELEMENT LENGTH “U”When ordering a complete assemblyincluding Thermowell, specify “TR”otherwise, specify “U” in whole inches
1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 3/4”3 - 3/8” 7 - 7/8”4 - 1/2” O - N/R (complete assy)
“U” IN FRACTIONAL INCHES
1 - Nipple-Union-Nipple 2 - Nipple-Union 3 - Nipple 4 - Head Only
EXTENSION
1- General Purpose Aluminum2 - General Purpose Cast Iron3 - Heavy Duty Aluminum4 - Heavy Duty Cast iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
Thermowell P/NSee Section 6O
Special Requirementsenter “X” and Specify
*For type R,S and B pricing contact the factory.
1 - 08 AWG 2 - 14 AWG 3 - 24 AWG (S R B)* X - Other ( Specify )
WIRE SIZE
ELEMENT INSULATION
Style Twisted Butt Welded Welded Single T B Duplex D W
B - Bare (no insulators) O - Oval Ceramic R - Round Ceramic
O
5
JUNCTION
Page - 40 ISL Temperature Measurent and Control
Industrial RW Style - RTD
Features General Specifications
!!!!!RTD Elements for use with Thermowells!!!!!Seven connection head varieties to choose
from
R W
SHEATH DIAMETER
ELEMENT LENGTH “U”When ordering a complete assemblyincluding Thermowell, specify “TR”otherwise, specify “U” in whole inches
1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 3/4”3 - 3/8” 7 - 7/8”4 - 1/2” O - N/R (complete assy)
“U” IN FRACTIONAL INCHES
1 - Nipple-Union-Nipple 2 - Nipple-Union 3 - Nipple 4 - Head Only
EXTENSION
1- General Purpose Aluminum2 - General Purpose Cast Iron3 - Heavy Duty Aluminum4 - Heavy Duty Cast iron5 - Explosion Proof Aluminum6 - Explosion Proof Cast Iron7 - Polypropylene
Special Requirementsenter “X” and Specify
100 Ohm PLATINUM RTD .00385CLASS WIRE WOUND THIN FILM
AB
AB
FM
6 - .188”7 - .250”
ELEMENT
100 Ohm Single100 Ohm Dual
2 Wire 3 Wire 4 Wire2 3 4D L -
Y - YESN - NO
SPRING - LOADED
CONNECTION HEAD
!!!!!Four standard extension configurations, type 1,2,3and 4
!!!!!Available up to 800°C temperature
Nom. 6” Nom. 2”U
U
U
U
Type 1
Type 2
Type 3
Type 4
Nom. 4”
NOTE:**For M.I. cable construction and 800 deg/C contact factory.
Thermowell P/NSee Section 6
Tube & Wire M.I. Cable**SHEATH MATERIAL Max 300° C Max 650° C
304 SS 1 N/A 316 SS 2 B Inconel 600 3 C
5
ISL Temperature Measurement and Control Page - 41
Section Thermowells and Protection tubes
Section 6 Thermowells
Threaded ThermowellSocket-weldWeld inFlangedVan StoneMetal TubesCeramic Tubes
6
Page - 42 ISL Temperature Measurent and Control
THERMOWELL MATERIALS GUIDE
Oxygen 75 ALL SteelOleic Acid See Fatty AcidsOxalic Acid 212 ALL MonelPhotographic Bleaching 100 ALL 304SSPalmitc Acid See Fatty AcidsPhosphoric Acid 212 ALL 316SSPhenol 212 ALL 316SSPotassium Compounds See Sodium CompoundsPropane 300 SteelRosin 700 100% 316SSSea Water 75 MonelSoap & Detergents 212 Monel 304SSSodium Bicarbonate 212 20% 316SSSodium Bisulphite 212 20% 304SSSodium Bisulphate 212 20% 304SSSodium Carbinate 212 40% 316SSSodium Chloride 300 30% MonelSodium Chromate 212 ALL 316SSSalt or Brine See Sodium ChlorideSodium Cyanide 212 ALL 304SSSodium Hydroxide 212 30% 316SSSodium Hypochlorite 75 10% Hast. CSodium Nitrate 212 40% 304SSSodium Nitrate 75 20% 316SSSodium Phosphate 212 10% SteelSodium Silicate 212 10% SteelSodium Sulfide 212 30% 316SSSodium Sulfite 212 10% 316SSSodium Sulfate 212 30% 304SSSodium Thiosulfate 212 ALL 304SSSteam 304SSStearic Acid See Fatty AcidsSugar Solution See GlucoseSulfur 500 304SSSulfur Chloride 75 DRY 316SSSulfur Dioxide 500 DRY 316SSSulfur Trioxide 500 DRY 316SSSulfuric Acid 212 10% 316SSSulfuric Acid 212 10-90% Hast. BSulfuric Acid 212 90-100% 316SSSulfuric Acid Fuming 175 Hast. CSulfurous Acid 75 20% 316SSTitanium Tetrachlonde 75 ALL 316SSTannic Acid 75 40% Hast. BToluene 75 SteelTrichloracetice Acid 75 ALL Hast. BTrichlorethylene 300 DRY MonelTurpentine 75 316SSVarnish 150 SteelZinc Chloride 212 ALL Hast. BZinc Sulfate 212 ALL 316SS
Corrodent Temp Conc. Recom. °F % Material
Copper (10) SulfateCopper Plating Solution(Cyanide)Copper Plating Solution(Acid)Corn OilCreosoteCrude OilEthyl AcetateEthyl Chloride, DRYEthanolEthyleneGlycol(Uninhibited)Ethylene OxideFatty AcidsFernc ChlorideFerric SulfateFormaldehydeFormic AcidFreonFlourine, AnhydrousFurturalGasolineGlucoseGlue ph 6-8GlycerineHydrobromic AcidHydrochloric Acid(37 -38%)Hydrogen Chloride, DryHydrocyanic AcidHydrofluoric AcidHydrogen Flouride, DryHydrofluogilicic AcidHydrogen PeroxideKeroseneLacquers & ThinnersLatic AcidLimeLinseed OilMagnesum ChlorideMagnesium Hydroxide(or Oxide)Magnesum SulfateMercuric ChlorideMercuryMethylene ChlorideMethyl Chloride, DryMilk fresh or sourMolassesNatural GasNitric AcidNitric Acid
300 ALL 316SS180 304SS
75 304SS
200 304SS200 ALL 304SS300 MonelSee Lacquer Thinner500 SteelSee Alchohols212 ALL 304SS
75 Steel500 ALL Hast. C75 ALL Hast. C300 ALL 304SS212 40% 316SS300 ALL 316SS300 Steel100 304SS450 316SS300 Steel300 304SS300 ALL 304SS212 ALL Brass212 ALL Hast. C225 ALL Hast. B
500 304SS212 ALL 304SS212 60% Monel175 Steel212 40% Monel125 10-100% 304SS300 ALL Steel300 ALL 304SS300 ALL 316SS212 ALL 316SS75 Steel212 50% Nickel75 ALL 304SS
212 40% 304SS75 10% Hast. C700 Steel212 ALL 304SS75 Steel180 304SSSee Glucose70 304SS75 ALL 304SS300 ALL 316SS
Acetic AcidAcetic AnhydrideAcetoneAcetyleneAlcoholsAlum Potassium(orSodium)Aluminum ChlorideAluminum SulfateAmmonia, DryAmmonium Hydroxide(Ammonia, Aqua)Ammonium ChlorideAmmonium NitrateAmmonium SulfateAmyl AcetateAnilineAsphaltAtmosphere (industrialand Marine)Barium CompoundsBeerBenzene (Benzol)Benzoic AcidBleaching PowderBoraxBordeaux MixtureBoric AcidBromineButaneButyl AlcoholButyric AcidCalcium BisulphiteCalcium ChlorideCalcium HydroxideCalcium HypochloriteCarbolic AcidCarbon Dioxide, DryCarbonated WaterCarbonated BeveragesCarbon DisulphideCarbon TetrachlorideChlorine, DryChlorine, MoistChloracetic AcidChloroform, DryChromic AcidCiderCitric AcidCopper (10) ChlorideCopper (10) Nitrate
212 ALL Monel300 Nickel212 ALL 304 SS400 304 SS212 ALL 304 SS300 ALL Hast. C
212 ALL Hast B212 ALL 316 SS212 ALL 304,316SS212 ALL 304,316SS
300 50% Monel300 ALL 304SS212 ALL 316SS300 ALL 304SS25 Monel250 304SS
304SS
See Calcium70 304SS212 Steel212 ALL 316SS70 15% Monel212 ALL Brass200 304SS400 ALL 316SS125 DRY Monel400 ALL SteelSee Alcohols212 Hast. C75 ALL Hast. C212 ALL Hast. C300 20% Hast. CSee Bleaching PowderSee Phenol800 ALL Brass212 ALL 304SS212 304SS200 304SS125 ALL Monel100 Monel100 ALL Monel212 ALL Monel212 Monel300 ALL Hast. C300 ALL 304SS212 ALL Hast. C212 ALL Hast. C300 ALL 316SS
Corrodent Temp Conc. Recom. °F % Material
Corrodent Temp Conc. Recom. °F % Material
THERMOWELLSManufactured from drilled bar stock, Pyrocoms thermowells provide protection from pressure, gas and liquid elements. Thick
walls provide sturdy protection for the sensor against high velocity and corrosive environments. Below is a helpfull guide of recomendedmaterials for specific corrodents.
Monel® is a registerd trade markof the Inco family of companiesHasteloy B® and Hasteloy C® areregistered trade marks of HaynesInternational
6
Page - 43ISL Temperature Measurement and Control
PROCESS CONNECTION “P”
3 - 3/4” NPT1 - 1” NPT2 - 1/2” NPT
A - .260” BoreB - .385” Bore
IMMERSION “U” LENGTH (In)
BORE DIAMETER “M”
Enter “U” Dimensionfrom table below
LAG “T” (In)If Required Enter inWhole Inches, “0” If Not
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™7 - Brass8 - Hastelloy™9 - Monel™
MATERIAL*
If Required, Enter “X”and Specify
SPECIAL REQUIREMENT
6TS Style Threaded Straight
T S 0 0
Note: *For other lengths and materials not listed above contactfactory.**For thermowells with lag contact factory.
ISL Temperature Mesurement and ControlPage - 44
PROCESS CONNECTION “P”
3 - 3/4” NPT1 - 1” NPT2 - 1/2” NPT
A - .260” BoreB - .385” Bore
IMMERSION “U” LENGTH (In)
BORE DIAMETER “M”
Enter “U” Dimensionfrom table below
LAG “T” (In)If Required Enter inWhole Inches, “0” If Not
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™7 - Brass8 - Hastelloy™9 - Monel™
MATERIAL*
If Required, Enter “X”and Specify
SPECIAL REQUIREMENT
6 TR Style Threaded Reduced
T R 0 0
Note: *For other lengths and materials not listedabove contact factory.**For thermowells with lag contact factory.
Page - 45ISL Temperature Measurement and Control
PROCESS CONNECTION “P”
3 - 3/4” NPT1 - 1” NPT2 - 1/2” NPT
A - .260” BoreB - .385” Bore
IMMERSION “U” LENGTH (In)
BORE DIAMETER “M”
Enter “U” Dimensionfrom table below
LAG “T” (In)If Required Enter inWhole Inches, “0” If Not
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™7 - Brass8 - Hastelloy™9 - Monel™
MATERIAL*
If Required, Enter “X”and Specify
SPECIAL REQUIREMENT
6TT Style Threaded Tapered
T T 0 0
Note: *For other lengths and materials not listed above contactfactory.**For thermowells with lag contact factory.
ISL Temperature Measurement and ControlPage - 46
6 W Style Socket Weld Thermowell
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If Not
MATERIAL*
0 - Carbon Steel1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™7 - Brass8 - Hastelloy™9 - Monel™
W 0 0
PIPE SIZE “P”
3 - 3/4” ( Actual 1.050” )1 - 1” ( Actual 1.315” )
BORE DIAMETER “M”
A - .260”B - .385”
IMMERSION “U” LENGTH (In)Enter “U” Dimmensionfrom table below
S - StraightR - ReducedT - Tapered
0
Note: *For other lengths and materials not listed above contactfactory.
STYLE
Style “S” Style “R”
Style “T”
Page - 47ISL Temperarure Measurement and Control
6
PIPE SIZE “P”
3 - 3/4” ( Actual 1.050” )1 - 1” ( Actual 1.315” )
BORE DIAMETER “M”
A - .260”B - .385”
IMMERSION “U” LENGTH (In)Enter “U” Dimmensionfrom table below
SPECIAL REQUIREMENTIf Required, Enter “x”and Specify. “0” If Not
MATERIAL*
0 - Carbon Steel1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™7 - Brass8 - Hastelloy™9 - Monel™
WI Style Weld-In
W I 0 0
Note: *For other lengths and materials not listed above contactfactory.
0
Full length taper provided when “U”
ISL Temperature Measurement and ControlPage - 48
6 F Style Flanged Thermowells
F
FLANGE SIZE (In)
1 - 3/4” 4 - 2” 2 - 1” 5 - 3” 3 - 1-1/2”FLANGE RATING (Lbs)
A - 150 D - 900 B - 300 E - 1500 C - 600 F - 3000
FLANGE MATERIAL
1 - 304 SS 6 - Inconel 600™2 - 310 SS 8 - Hastelloy™3 - 316 SS 9 - Monel™
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If not
THERMOWELL MATERIAL*
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 600™8 - Hastelloy™9 - Monel™
IMMERSION “U” LENGTH (In)
Enter “U” Dimmensionfrom table below
STYLE
Note: *For other lengths and materials not listed above contactfactory.
S - StraightR - ReducedT - Tapered
BORE DIAMETER “M”A - .260”B - .385”
0
Style “S”
Style “R”
Style “T”
Page - 49ISL Temperarure Measurement and Control
6
V
BORE DIAMETER “M”A - .260”B - .385”
SPECIAL REQUIREMENTIf Required, Enter “X”and Specify. “0” If not
THERMOWELL MATERIAL*
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS5 - 347 SS6 - Inconel 6008 - Hastelloy9 - Monel
IMMERSION “U” LENGTH (In)
Enter “U” Dimmensionfrom table below
PIPE SIZE ( NPS)
1 - 1” Nominal ( 1.315” )2 - 1-1/2” Nominal ( 1.90” )3 - 2” Nominal ( 2.375” )
Van Stone Thermowells
0
Note: *For other lengths and materials not listed above contactfactory.
STYLE
S - StraightR - ReducedT - Tapered
0 0
Style “R”
Style “S”
Style “T”
ISL Temperature Measurement and ControlPage - 50
6
PIPE SIZE ( NPT )
3 - 1/2” ( .840” )4 - 3/4” ( 1.05” )5 - 1” ( 1.315 )
PROCESS CONNECTION
Bushing FlangeA - 1/2” NPT 1 - 3/4B - 3/4” NPT 2 - 1C - 1” NPT 3 - 1 1/2D - 1 1/4” NPT 4 - 2E - 1 1/2” NPT 5 - 30 - No Bushing or Flange
FLANGE RATING ( Lbs )
0 - No FlangeA - 150 D - 900B - 300 E - 1500C - 600 F - 2500
MATERIAL Flange or Bushing
0 - None 3 - 316 SS1 - 304 SS 8 - Hastelloy2 - 310 SS 9 - Monel
SPECIAL REQUIREMENT
1 - 304 SS2 - 310 SS3 - 316 SS4 - 446 SS6 - Inconel 6008 - HastelloyH - HR-160
PIPE MATERIAL*
If Required, Enter “X” andSpecify. “0” If Not
“T” LENGTH (In) Enter in Whole Inches
DIMENSION “U” FRACTIONAL0 - 0 4 - 1/2”1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 7/8”3 - 3/8”
IMMERSION “U” LENGTH
Enter “U” Dimension inWhole Inches
PIPE SCHEDULE**1 - 40 3 - 1602 - 80 4 - XX
M Style Metal Protection Tube
M
UT
T U
Process Connection
RF Flange Shown
Page - 51ISL Temperarure Measurement and Control
6
TUBE DIAMENTER*
O.D. I.D.1 - 3/8” 1/4”2 - 7/16” 5/16”3 - 1/2” 3/8”4 - 11/16” 7/16”5 - 3/4” 1/2”6 - 1” 3/4”
HEAD CONNECTION
0 - None2 - 1/2” NPT3 - 3/4” NPT
PROCESS CONNECTION**
0 - None2 - 1/2” NPT3 - 3/4” NPT (Standard)
SPECIAL REQUIREMENT
If Required, Enter “X”and Specify. “0” If Not
TUBE MATERIALA - Alumina (3400°F)M - Mullite (2750°F)C - Cer-Met (2200°F)H - Halsic I (3000°F)S - Silicon Carbide (3000°F)
IMMERSION “U” FRACTIONAL
0 - 0 4 - 1/2”1 - 1/8” 5 - 5/8”2 - 1/4” 6 - 7/8”3 - 3/8”
IMMERSION “U” LENGTH
Enter “U” Dimensionin Whole Inches
C 0 0
C Style Ceramic Protection Tube
U
U
Process ConnectionHead Connection
Plain Tube No Fitting
0
Note: *For other lengths and materials not listed abovecontact factory.