LEARNING OBJECTIVES
UL 1365 listing
Additional Installations/Applications
NFPA 99 HISTORY
2018 brings the requirement of a third party inspection prior to concealing the medical gas piping. • Must be performed by an independent ASSE 6020 or 6030
• Labels on pipe and valves to be inspected
• Must witness the initial pressure test (150psi or 1.5X operating)
• Documented and given to ASSE 6030 verifier
CHAPTER 5: INSPECTION
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
5.1.10.1.4 Tubes shall be one of the following: 1. Copper remains unchanged from previous code editions 2. Listed corrugated medical tubing (CMT) fabricated from copper alloy No.
51000 strip, meeting ASTM B103/ B103M, Standard Specification for Phosphor Bronze Plate, Sheet, Strip, and Rolled Bar, with a design margin of 3.5, externally coated with a nonmetallic sheath marked with the manufacturer’s marking. The listing shall include testing to demonstrate that CMT systems can be consistently gas-purged with results equivalent to comparable medical gas copper tubing
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
5.1.10.1.5 CMT shall have a flame spread index of 25 or less and a smoke developed index of 50 or less as determined by ASTM E84 or S102, Standard Test Method for Surface Burning Characteristics of Building Materials.
5.1.10.1.6 CMT shall be identified by the manufacturer as suitable for oxygen service at a minimum of every 0.92 m (3 ft).
5.1.10.2.1(3) CMT is allowed for Vacuum/WAGD piping
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
5.1.10.3.2 Positive pressure patient gas systems, medical support gas systems, vacuum systems, and WAGD systems constructed of CMT shall have turns, offsets, and other changes in direction made by bending the tubing up to the minimum bend radius or by fittings in accordance with 5.1.10.3.1
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
5.1.10.11.4.4 Supports for CMT shall be in accordance with the CMT manufacturer’s installation instructions.
5.1.10.11.10.3 CMT systems shall be installed by ASSE 6010 qualified installers using the CMT manufacturer’s instructions.
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
UL 1365 Corrugated Medical Tubing • Developed in accordance with NFPA 99 standards • Limited to aboveground or underground inside a conduit • Limited to 2” and smaller • Series of 13 different testing procedures
CHAPTER 5: CMT CORRUGATED MEDICAL TUBING
UL 1365 PERFORMANCE TESTING
UL 1365 PERFORMANCE TESTING
UL 1365 REQUIREMENTS
• Tube shall carry manufactures marks every two feet with: o Manufactures trademark or identifying symbol o Manufactures part number o Maximum operating pressure o Factory designation if applicable
• Fitting shall be permanently marked with: o Manufactures trademark or identifying symbol o Manufactures part number o Factory designation if applicable
CMT FEATURES
• Long continuous lengths/ 80% reduction in brazed joints
• Minimal brazing/ with the use of axially swaged fittings it’s possible to eliminate brazing
• Reduced risk of contamination
• Inherently resistant to seismic events
• Return the system to service substantially quicker/ up to a 5X faster installation time
COPPER VS CMT TYPICAL 8 HOUR DAY
Typical 8 hour day / 1 installer Copper (hrs)
CMT (hrs)
Arrive on site, set up equipment, move material into location
2 1
Obtain hot work permit and prepare fire system 1 0
Install piping / Braze piping 3 6
Allow piping to cool to touch/clean up work area 1 1
Fire watch 1 0
Total Time 8 8
PICKUP OF AN ADDITIONAL 4 HRS WHERE INSTALLER CAN BE ACTUALLY INSTALLING PIPE
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Project scope: Relocate existing zone valve from a patient care area to a nurse station • New location was 65’ down an existing corridor • Vacuum run in 1.25” copper • Oxygen run in 1” CMT • Tie in to existing O2 system with Axial Swaged fittings, eliminating all hot
work in the facility for the Oxygen system • Hangers and supports pre-installed and not accounted for in comparison
of systems
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Medical Vacuum System: (Traditional Copper) • 150’ of 1.25” copper • 28 fittings (90’s 45’s, couplings) • 64 man hours for completion
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Oxygen System: (CMT) • 150’ of 1” MediTrac CMT • 4 Meditrac end fittings • 2 axial swaged fittings for tie in • 12 man hours for completion
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Installer CMT benefits compared to copper: • Reduced man hours (premium time) by 82% • Reduced fittings (possible failures) by 79% • Eliminated exposure and risk of hot work
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Medical Gas Verifier CMT benefits compared to copper: • Eliminate the need to remove nitrogen left in system from
the purge • Reduced shut-down and verification time • Reduced risk of contaminants in system
(Nitrogen or particulates)
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
Facility CMT benefits compared to copper: • Work done on a single Saturday vs multiple nights • Less work in patient care areas = Lower patient impact • Reduced risk of contaminants in system
(Nitrogen or particulates) • Less facility staff involvement allows for staff to focus on other
projects • 55% reduction in total installed cost
MELROSE WAKEFIELD CASE STUDY COPPER/CMT RUN IN PARALLEL
ADDITIONAL INSTALLATIONS
ADDITIONAL INSTALLATIONS
ADDITIONAL INSTALLATIONS
Omega Flex, Inc. • 451 Creamery Way • Exton, PA 19341 1-800-671-8622 ISO 9001 Registered Company MTUS-510 09/18 © Omega Flex, Inc. 2018
New zone valve box, with Meditrac installed for oxygen.
(continued on back)
Location: Melrose Wakefield Hospital, Melrose, MA
• Saves Time • Saves Money • Increases Installation
Options
CASE STUDY MediTrac® in Hospital Renovation
Until the innovative new product, Corrugated Medical Tubing (CMT), was adopted into the Health Care Facilities Code NFPA 99 – 2018; installation, retrofits and repairs of medical gas piping systems were confined to the rigid installation practices of hard-drawn seamless copper pipe.
MediTrac, the world’s first flexible medical gas piping system, bends to a facility’s needs, resulting in shorter downtimes. MediTrac decreases installation time by greatly reducing the number of fittings, potential leak points and chance of contamination during installation.
Project Overview: OmegaFlex® the makers of MediTrac® partnered with Acute Medical Gas Services from Goffstown, NH on the first US based BETA site installation at Melrose-Wakefield Hospital in Melrose, Massachusetts. The project included the relocation of an existing zone valve box from the patient care area to a designated area outside of the patient care area as required per NFPA 99.
Installation Overview The new zone valve box, containing shutoffs for both a vacuum and oxygen line, was located on the other side of a corridor beyond the fire stop roughly 65 feet away. For comparison purposes the vacuum line was done in 1-1/4 hard-drawn copper pipe and the oxygen lines were done in 1 MediTrac CMT. All piping runs were done
in parallel taking the same route between connection points. In total approximately 75 of piping was used for each run (150 total per system).
The piping run included a vertical rise out of the new zone valve box followed by a series of 90° and 45° bends to the 65 straight run down the corridor. The piping then turned 90 degrees across the corridor and was tied into the existing supply and distribution lines. Both piping systems utilized the same existing hanger supports down the 65’ corridor run.
Summary of Piping System Connections Medical Vacuum Brazed Hard-drawn copper pipe details: 2 runs total • 150 of hard - drawn K copper pipe • 28 fittings with a total of 56 brazed joints
- 12 – 90° elbows - 4 – 45° elbows - 12 couplings
Medical Oxygen MediTrac® CMT details: 2 runs total
• 150 of MediTrac® CMT • 8 fittings total with 4 brazed joints
- 4 MediTrac® CMT fittings - 2 couplings - 2 axial swaged hard-drawn copper pipe connections
(continued from front)
meditrac.us
Termination of MediTrac CMT piping is easy with the patent pending MediTrac fitting.
MediTrac cuts the same as hard-drawn copper pipe but navigates confined spaces and corners without the hot-work.
CASE STUDY MediTrac® in Hospital Renovation
As with all rigid piping installations elbows are required when turning corners and changing direction to route around, above, or below obstacles. MediTrac CMT is able to bend around corners and flex to avoid potential obstacles. Resulting in no intermediate connections.
In this case two 1 MediTrac CMT lines were pulled together, much like pulling wire or cabling through a hole in the fire wall down the entire length of the corridor in roughly 15 minutes. An additional 30 minutes later the piping rough in was complete and ready for fitting connections.
Labor Man Hours per System Overview of man hours and comments received from Acute Medical Gas Services
Medical Vacuum Brazed Hard-drawn copper pipe details
• 64 total man hours to tie in the new valve box and prepare for the final tie in to existing line.
Medical Oxygen MediTrac® CMT details • 12 total man hours to pull and connect,
- Note: Prefab of a T for an alarm sensor and reducing couplings to make the final tie-in required. So, in theory, you could subtract several hours from this time.
Overall the use of MediTrac® resulted in a 79% reduction in joints for the piping system and a minimum of an 82% reduction in labor man hours. Additional benefits of the MediTrac® system as noted by Acute Medical include:
Ease of transport to the job site
Reduced cost in support items such as brazing rods and purge gas
Contamination possibility is greatly reduced
Lack of intermediate joints and the systems flexibility resolve fire prevention during difficult brazing situations unlike hard drawn copper systems (wire, ducts with dust inside, walls, gas piping, etc.)
Omega Flex, Inc. • 451 Creamery Way • Exton, PA 19341 1-800-671-8622 ISO 9001 Registered Company MTUS-511 03/19 © Omega Flex, Inc. 2019
Product: MediTrac® Corrugated Flexible Medical Tubing
Location: Henrico Doctors’ Hospital, Henrico, VA
• Saves Time • Saves Money • Increases Installation
Options
CASE STUDY MediTrac® in Hospital Renovation
MediTrac® by OmegaFlex® continues its introduction into the market by providing innovation from end to end with another installation.
Project Overview OmegaFlex® the makers of MediTrac® partnered with Pure-Line Plumbing from Durham, NC on a medical vacuum system retrofit at Henrico Doctors’ Hospital in Henrico, Virginia. The project included the removal of an existing oxygen only zone valve box and installation of a new O2 and vacuum zone valve box. This newly added box is used to service 9 patient care rooms with updated oxygen outlets and newly added vacuum outlets. Existing oxygen piping was reused and a hybrid system utilizing copper pipe and MediTrac® corrugated medical tubing was installed for the vacuum lines. This hybrid system installation highlights the easy and efficiency of using MediTrac® CMT seamlessly with traditional rigid copper tubing materials.
Installation Overview The tie in to the closest vacuum line was approximately 110 away from the updated
zone valve box. 1 MediTrac® was used to navigate multiple changes in direction in a tightly packed space above the ceiling with 2 through wall penetrations. Routing of rigid copper piping in this location would have been difficult due to the limited space available for routing of the piping system and the brazing that would have been required for the traditional piping system.
The outlet from the zone valve box was connected to roughly 28 of 3/4 MediTrac® CMT. Piping was installed vertically up the wall cavity to the space above the drop ceiling and routed into the patient care area via a 3rd through penetration. At this point the piping system was transitioned from MediTrac® to rigid copper pipe that was routed in an “L” configuration in the ceiling of the corridor outside of the individual patient care rooms. A total of six (6) brazed copper tees were installed to provide seven (7) drops off the copper piping to service all nine (9) patient care rooms.
In total there was roughly 110 of 1 and 185 of ¾ MediTrac® CMT installed in this project in conjunction with approximately 68 of ¾ rigid brazed copper. (continued
OmegaFlex® the makers of MediTrac® partnered with Pure-Line Plumbing from Durham, NC on a medical vacuum system retrofit at Henrico Doctors’ Hospital in Henrico, Virginia.
Typical Installation Benefits • 55% Reductions in Total Installed Cost • 80% Reductions in Brazed Joints (Fire Hazard) • 70% Fewer Fittings (Less Potential Leak Points) • 5 Times Faster to Install (Minimal Downtime and
Maximum Patient Care)
(continued from front)
1" MediTrac ® CMT connection to zone valve box for vacuum.
3/4" MediTrac ® CMT drop to patient care room from rigid copper main.
CASE STUDY MediTrac® in Hospital Renovation
Installation Highlights The initial 110 run from the existing vacuum line to the new zone valve box was pulled and roughed in in 40 minutes. That time included the drilling of 2 through penetrations. Final tie in took the total install time of this section to roughly 1 hour. It was estimated that to complete this same run with rigid copper piping would have taken at least 4.5 hours and required a large number of braze joints to navigate the tight installation location.
While most of the tie-ins from the copper line to the patient care rooms were not very long, ranging from 15 to 35, the reduction of overhead work to fit and braze copper ells and couplings saved on installation time. That savings on installation time allowed the job to be completed in the short window available for the installation.
• Overall, this project went from days to hours for completion
• Contamination possibility was greatly reduced
• There was a significant reduction in intermediate joints
• The hospital was able to return to service and back to providing patient care in a short amount of time