Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2 , Ingrid Lai 2 , Andrew Kerr 1,2 and L. John Schreiner 1,2 New Technologies in Radiation Oncology 101 1 Cancer Centre of Southeastern Ontario, Kingston Health Sciences Centre and 2 Departments of Oncology and Physics, Queen’s University, Kingston, ON, Canada Acknowledgements: Funding support from Canadian Institute of Health Research (CIHR) and Ontario Consortium for Adaptive Interventions in Radiation Oncology (OCAIRO) and industry partner Best Theratronics, Canada is much appreciated.

New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Digital Portal Imaging in Cobalt-60 Radiation Therapy

Matthew Marsh1,2, Ingrid Lai2, Andrew Kerr1,2 and L. John Schreiner1,2

New Technologies in Radiation Oncology 101

1 Cancer Centre of Southeastern Ontario, Kingston Health Sciences Centre and 2 Departments of Oncology and Physics, Queen’s University, Kingston, ON, Canada

Acknowledgements: Funding support from Canadian Institute of Health Research (CIHR) and Ontario Consortium for Adaptive Interventions in Radiation Oncology (OCAIRO) and industry partner Best Theratronics, Canada is much appreciated.

Page 2: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Introduction 1

• Image guidance for patient set-up verification is a standard of care in modern radiation therapy.

• Cobalt-60 (Co-60) units are still the only treatment options in many clinical settings, but most units do not have imaging capabilities.

• The addition of even basic imaging could be hugely beneficial.

• Co-60 imaging has some challenges: • Lower contrast images at 1.25 MeV photon energy

(Compton density contrast dominates) • Large source size diminishes resolution

Page 3: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Introduction 2

• In this work we investigate if these issues can be overcome by Co-60 digital imaging with an electronic portal imaging device (EPID)

• Rationale: Prior to the introduction of on-board kV X-ray imaging, portal imaging was the most effective way to verify the patient’s position for external beam radiation therapy.

Page 4: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Methods I: Apparatus

• Treatment units: • Th-780C Co-60 unit (Best Theratronics, Kanata, Canada) • Varian Clinac 6EX Linac (Varian,Palo Alto, CA, USA)

• Imaging panels: • XRD1640 amorphous Si EPID (PerkinElmer, Germany) • aSi500 EPID on 6EX (Varian, USA)

Imaging Setup

SAD

SDD

Page 5: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Methods II: Image Acquisition / Processing

• Image Acquisition: • Frame integration times: 133 ms for XRD1640,

100 ms for aSi500. • 4 frames were averaged for each image

• Image Processing • MATLAB (The Math Works, Natick, MA, USA) • Iterative deconvolution, contrast-limited adaptive

histogram equalization and Gaussian smoothing.

• Phantoms: • CIRS 801-P pelvis

(CIRS, Norfolk, VA) • SBU-4 “Rando”

(Kyoto Scientific)

Page 6: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

• Co-60 EPID images of SBU-4 “Rando” with Perkin Elmer XRD1640

• Image quality depends on setup geometry and improves when the panel is close to the patient

Results I: Co-60 EPID IMAGES (H&N)

SAD=80cm SDD=100cm SAD=100cm SDD=125cm SAD=80cm SDD=120cm SAD=100cm, SDD=140cm

Page 7: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

SAD=100cm, SDD=140cm

SAD=80cm SDD=120cm

• Usable images are obtained with 100/140 cm configuration, would be suitable for 360° rotation of a gantry mounted panel.

Results I b: Co-60 EPID IMAGES (H&N)

Page 8: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Results II: IMAGES (Pelvis)

• Co-60 EPID images with XRD1640 (Perkin Elmer) and aSi500 (Varian) portal imagers at SAD=100cm , SDD=140cm distances

• AP pelvic images are superb after image processing (contrast-limited adaptive histogram equalization, CLAHE). Key landmarks on all major bones in the region can be identified in the sagittal view.

Co-60 , XRD1640 Co-60 , aSi500

Page 9: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Results II b: IMAGES (Pelvis)

• In fact Co-60 images compare favorably with 6MV EPID images aSi500 (Varian) portal imagers at SAD=100cm, SDD=140cm.

• Varian imaging improved recently with move to aSi1000.

Co-60 , XRD1640 Co-60 , aSi500 6MV, aSi500

Page 10: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Results III: Image Quality Metrics

Measured point spread functions. (Left) Original. (Right) After deconvolution sharpening (deconv. kernel radius 14px at 80/100 & 100/125, 17px at 100/140, 18px at 80/120).

Point Spread Function (PSF) • Does indicate Co-60 penumbra problem: PSF is 2-4 mm in

radius. • Deconvolution sharpening can reduce PSF width, at the

cost of some ringing.

0 5 10 15 200

0.02

0 .04

0 .06

0 .08

0 .1

Pos tion (px)

Me

asu

rem

en

t (a

rb.)

PS F

80/10080/120100/125100/140

0 5 10 15 200

0.02

0 .04

0 .06

0 .08

0 .1

Pos tion (px)

Me

asu

rem

en

t (a

rb.)

PS F Afte r De convolution

80/10080/120100/125100/140

Page 11: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Results III: Image Quality Metrics

Raw square and sine wave MTFs determined using PipsPro (Standard Imaging, Inc.) technique for different SAD, SDD (cm).

Modulation Transfer Functions

• MTF measurements via PipsPro QC3 phantom show a penumbra overlap effect for SAD/SDD 80/120 cm.

• Best results are obtained with the detector panel as close as possible to the object being imaged.

Page 12: New Digital Portal Imaging in Cobalt-60 Radiation Therapy · 2018. 5. 7. · Digital Portal Imaging in Cobalt-60 Radiation Therapy Matthew Marsh 1,2, Ingrid Lai , Andrew Kerr 1,2

Discussion / Conclusions

• The image quality achievable with the Co-60 system approaches that of a 6 MV linac’s portal imager, and is adequate to identify bony anatomy with respect to the treatment field. • Sharpness in Co-60 images is somewhat lacking in

comparison to 6 MV images, but this can be improved using image processing.

• Resource-limited cancer centres with Co-60

units can increase treatment accuracy and efficiency by adding a gantry-mounted or independent EPID combined with image processing software.