Upload
santoshnbsm
View
225
Download
0
Embed Size (px)
Citation preview
7/29/2019 nbsm metrology
1/3
Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70
Prepared on March 4, 2013
1/3
Item I, Laptops
Hard Drive : >500GB SATA 5400rpm
Display: > 15.6 inch, Led
Processor: > Intel 2nd Gen Core i5 , Speed - >2.3GhZ, Bus Speed : >1333 MHz
Graphics Memory: 1GB
Memory: > 4GB DDR3 Speed: >1333 MHz
Hard Drive Capacity: > 500GB Serial ATA - 5400.0 rpm
Operating System: Microsoft Windows 8 64 bit , Microsoft Windows 7 Professional 64 bit
Network & Communication : WAN, Bluetooth, Ethernet, USB, HDMI
Power: >6 hours
Item 2, Low temperature liquid baths
Temperature ranges from30C to 150C
Stability to 0.02 C
Uniformity: 0.05 CAccuracy: 0.3C
Calibration Depth: not less than 300 mm
Computer communications: USB /RS-232/LAN/WAN
Item 3, Thermometric Bridge (For calibration of SPRTs, Thermistors etc)
Measuring Range
Temperature: -50 to 600 C ;
Resistance: 0 to 400
Resolution
Temperature: 0.1 mK (Pt 100)Resistance: 50 m
Measurement Uncertainty: Within the values of its resolution (95% Confidence Level & at (202)C
Sensor Connection: 4 wire
Number of sensor inputs: 2
Computer communications: USB/RS-232/LAN/WAN
(Calibration Certificates from laboratories accredited as per ISO/IEC 17025 required)
Assumed Cost: $8000
Item 4: Laboratory Ice Cube makers
Required No. : 1
Production Capacity >30 Kg/24 HoursStorage Capacity > 1 Year
Accessories Ice Cube Crusher (Mechanical)
Assumed Cost: $1000
Item 5: Data Logger with humidity and temperature sensor : 1 set
7/29/2019 nbsm metrology
2/3
Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70
Prepared on March 4, 2013
2/3
Temperature:
Thermistor Probe Measurement Range:40 C to 125 C (40 F to 257 F)
Internal Resolution: 0.1 C typ
Thermistor Probe Temperature Accuracy (Overall Error): 0.5 C typ, 2.0 C max
Alarm Threshold Range (Software-Selectable):40 C to 125 C (40 F to 257 F) range for both high
alarms and low alarms
Data Sampling
Sampling Rate (Software-Selectable): 10 s, 30 s, 1 min, 5 min, 30 min, 1 hr, 6 hr, 12 hr
Temperature Samples: More than 1,000,000 max
Temperature Units: C or F
Relative Humidity
Measurement Range: 0% RH to 100% RH
Accuracy (Overall Error Between 20% RH and 80 %RH): 3.0% RH typ
Alarm Threshold Range: 0% RH to 100% RH (high and low alarms)
Data Sampling
Sampling Rate (Software-Selectable): 10 s, 30 s, 1 min, 5 min, 30 min, 1 hr, 6 hr, 12 hr
Combined Temperature and Humidity Samples: More than 500,000 maxTemperature Units: C or F
Other requirements
Rechargeable battery-powered:
Internal lithium polymer battery (rechargeable using included microUSB cable)
Internal Storage:
Logs more than 1 million temperature samples and more than 500,000 temperature/humidity samples
LCD Display:
Integrated LCD screen displays current readings, min/max readings, high and low alarm occurrences,
and battery charge status
Connectivity: USB, WiFi
Software
Item 6: Force Transducer for Calibration of Universal testing machine/ Compressive testing
machine (1 Set)
Capacity: 2 MN
Class : 0.5 (as per ISO 376)
Direction of force: Tensile and Compression
Protection: IP67
Normal Sensitivity: 2mV/V
Calibration Certificate: from ISO 17025 accredited Laboratory
Other requirements: Mounting for tensile and compressive mode, display unit
Item 6: Force Transducer for Calibration of Universal testing machine/ Compressive testing
machine (1 Set)
Capacity: 500kN
Class : 0.5 (as per ISO 376)
Direction of force: Tensile and Compression
Protection: IP67
7/29/2019 nbsm metrology
3/3
Items to be purchased in Scientific Metrology Division, NBSM for FY-2069/70
Prepared on March 4, 2013
3/3
Normal Sensitivity: 2mV/V
Calibration Certificate: from ISO 17025 accredited Laboratory
Other requirements: Mounting for tensile and compressive mode, display unit