1/9/2018 1 Multimodality Imaging in Aortic Diseases: Federico M Asch MD, FASE, FACC Chair, ASE Guidelines and Standards Committee MedStar Washington Hospital Center MedStar Health Research Institute Georgetown University Washington, DC Federico M Asch MD, FASE, FACC Chair, ASE Guidelines and Standards Committee MedStar Washington Hospital Center MedStar Health Research Institute Georgetown University Washington, DC January 2018 I have no financial disclosures related to this presentation I have no financial disclosures related to this presentation

Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

1

Multimodality Imaging in Aortic Diseases:

Multimodality Imaging in Aortic Diseases:

Federico M Asch MD, FASE, FACCChair, ASE Guidelines and Standards Committee

MedStar Washington Hospital CenterMedStar Health Research Institute

Georgetown UniversityWashington, DC

Federico M Asch MD, FASE, FACCChair, ASE Guidelines and Standards Committee

MedStar Washington Hospital CenterMedStar Health Research Institute

Georgetown UniversityWashington, DC

January 2018

I have no financial disclosures related to this presentationI have no financial disclosures related to this presentation

Page 2: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

2

J Am Soc Echocardiogr 2015;28:119-82

GUIDELINES AND STANDARDS

asecho.org Guidelineswww.

Survey for Imaging of the Aorta

Page 3: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

3

Survey for Imaging of the AortaSurvey for Imaging of the Aorta

1. Uniform protocol for measuring aorta1. Uniform protocol for measuring aorta

Inner-to-inner

Outer-to-outer

58%

42%

Survey for Imaging of the AortaSurvey for Imaging of the Aorta

2. Which of the following is your recommended/preferred time to measure the aortic root ?

2. Which of the following is your recommended/preferred time to measure the aortic root ?

End-systole

End-diastole

44%

56%

Page 4: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

4

Survey for Imaging of the AortaSurvey for Imaging of the Aorta

3. Which of the following do you consider to be the reliable level of resolution of current imaging modalities?

3. Which of the following do you consider to be the reliable level of resolution of current imaging modalities?

1 mm

2 mm

3 mm

4 mm

5 mm

41%

47%

6%

6%

0

Survey for Imaging of the AortaSurvey for Imaging of the Aorta

4. Which of the following degrees of enlargement do you feel is significant during annual/serial follow-up of aortic size ?

4. Which of the following degrees of enlargement do you feel is significant during annual/serial follow-up of aortic size ?

>2 mm

>3 mm

>4 mm

>5 mm

18%

23%

24%

35%

Page 5: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

5

Imaging TechniquesImaging Techniques• Chest X‐ray

• Echo  (TTE, TEE, 3D‐echo, epiaortic)

• Intravascular echo (IVUS)

• Intracardiac echo (ICE)

• CT/MDCT

• Magnetic resonance imaging

• Aortography

• Chest X‐ray

• Echo  (TTE, TEE, 3D‐echo, epiaortic)

• Intravascular echo (IVUS)

• Intracardiac echo (ICE)

• CT/MDCT

• Magnetic resonance imaging

• Aortography

Page 6: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

6

Diseases of the Thoracic AortaDiseases of the Thoracic Aorta

• Acute aortic syndromes

• Thoracic aortic aneurysms

• Acute aortic syndromes

• Thoracic aortic aneurysms

continued . . .

- Aortic dissection- Intramural hematoma- Penetrating aortic ulcer- Ruptured aortic aneurysm

- Bicuspid aortic valve-related aortopathy- Marfan syndrome- Other genetic diseases

(Ehlers-Danlos; Loeys-Dietz, Turner syndrome,etc)

Diseases of the Thoracic AortaDiseases of the Thoracic Aorta

• Traumatic injury of thoracic aorta

• Aortic coarctation

• Atherosclerosis

• Aortitis

• Traumatic injury of thoracic aorta

• Aortic coarctation

• Atherosclerosis

• Aortitis- Noninfectious

- Infectious

Page 7: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

7

Measuring the Aorta

What to look for?What to look for?

• Aortic Valve morphology

• Normal Aortic size in the adult:

– Ao root < 40 mm

– Ascending Ao < 37 mm

– Descending Aorta < 28 mm

• These values are, however, very variable.

• Aortic Valve morphology

• Normal Aortic size in the adult:

– Ao root < 40 mm

– Ascending Ao < 37 mm

– Descending Aorta < 28 mm

• These values are, however, very variable.

Page 8: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

8

Aortic size to be adjusted by body size and age

Aortic size to be adjusted by body size and age

< 20 yo 20‐40 yo >40 yo

Roman M et al. Am J Cardiol 1989;64:507

Complications and Asc Ao sizeImportance of accurate measurements

Complications and Asc Ao sizeImportance of accurate measurements

Elefteriades, JACC 2010;55:841 

Page 9: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

9

Complications and Asc Ao sizeImportance of accurate measurements

Complications and Asc Ao sizeImportance of accurate measurements

Elefteriades, JACC 2010;55:841 

Asc Aortic size at time of Type A Dissection

Asc Aortic size at time of Type A Dissection

Pape et al for IRAD. Circulation 2007;116:1120

60% < 5.5 cm

40% <5 cm

Page 10: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

10

PRE POST PRE POST

B.

*

Weinsaft J, Asch F, GenTAC. JACC 2016.

Aortic size predicts dissection even after prophylactic aortic graft surgery

Aortic size predicts dissection even after prophylactic aortic graft surgery

• Reproducibility is similar in all modalities. A variation error in aortic measurements of ≅ 3mm should be assumed

• Consider:– 2D vs 3D methods– Gated imaging to select timing of the cardiac cycle– Blind spots / area of interest– Frequency of required follow‐up– Additional benefit of each technique– Contraindications for each modality– Availability and expertise at each center

• Reproducibility is similar in all modalities. A variation error in aortic measurements of ≅ 3mm should be assumed

• Consider:– 2D vs 3D methods– Gated imaging to select timing of the cardiac cycle– Blind spots / area of interest– Frequency of required follow‐up– Additional benefit of each technique– Contraindications for each modality– Availability and expertise at each center

Which test to order?TTE vs TEE vs CT vs MRIWhich test to order?

TTE vs TEE vs CT vs MRI

Page 11: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

11

• Most importantly, measurements are very variable among different imaging modalities and even within each modality.

• For proper follow‐up, it is critical to use the same modality and compare side to side images.

• Most importantly, measurements are very variable among different imaging modalities and even within each modality.

• For proper follow‐up, it is critical to use the same modality and compare side to side images.

MDCTMDCT

Strength:‐ 3D – Multiplanar‐ Entire aorta and branches‐ Landmarks/site

Weaknesses:‐ Need for contrast‐ Radiation‐ Ao Root (non‐gated)‐ Cross‐sections (axial CT)

Page 12: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

12

TTETTE

Strengths:‐ Aortic Root, AI‐ Standardized measurements‐ Availability‐ Safety‐ Great Screening tool for

‐ Ascending aorta‐ Arch ‐ Abdominal aorta

Weaknesses:‐ All other segments

TEETEE

Strength:‐ Accurate measurements‐ Other cardiac structures

Weaknesses‐ Blind spot: Arch and vessels‐ No landmarks‐ Tortuous Aorta‐ Frequent follow‐ups‐ Semi‐ invasive

Page 13: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

13

MRIMRI

Strength:‐ 3D – Multiplanar‐ Entire aorta and branches‐ Aortic walls‐ Landmarks/site‐ No contrast‐ No Radiation

Weaknesses:‐ Contrast (MRA)‐Ao Root (non‐gated)

Elefteriades, JACC 2010;55:841

Elongation and tortuosity can induce measurement errors

Elongation and tortuosity can induce measurement errors

Page 14: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

14

Importance of Imaging techniqueImportance of Imaging technique

Mendoza D, Weinsaft J et al. Ann Thorac Surg, 2011;92:904

Importance of Imaging techniqueImportance of Imaging technique

Mendoza D, Weinsaft J et al. Ann Thorac Surg, 2011;92:904

Page 15: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

15

Mendoza D, Weinsaft J et al. Ann Thorac Surg, 2011;92:904

Echo vs CT measurements‐ GenTACEcho vs CT measurements‐ GenTAC

• 189 cases without grafts were identified with echo and CT 

performed within 30 days, and no events in‐between

• Inner edge to inner edge, systolic measurements

Page 16: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

16

Sinus of Valsalva = good correlationSinus of Valsalva = good correlation

N= 109ICC=0.82

Difference

-3

-2

-1

0

1

2

3

Average

2 3 4 5 6 7 8

sinus_valsalva_max Echo vs. CT

Bland-Altman Plot

SinoTubular Junct = good correlationSinoTubular Junct = good correlation

N= 94ICC=0.75

Difference

-2

-1

0

1

2

Average

1 2 3 4 5 6 7 8

stj_max Echo vs. CT

Bland-Altman Plot

Page 17: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

17

Asc Aorta = good correlationAsc Aorta = good correlation

N=87ICC=0.77

Difference

-3

-2

-1

0

1

2

3

Average

1 2 3 4 5 6

asc_max Echo vs. CT

Bland-Altman Plot

AV Annulus – poor correlationAV Annulus – poor correlation

N=94ICC=0.49

1 2 3 4 5

12

34

5

AV_Annulus_1_max_CT

AV_A

nnul

us_1

_max_

Ech

o

y0.401x1.077

Difference

-3

-2

-1

0

1

2

3

Average

1 2 3 4 5

av_annulus_max Echo vs. CT

Bland-Altman Plot

For cases with Gated CT:N=23, ICC=0.74

Page 18: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

18

Asch FM, GenTAC invest. JACC img 2016;9:219‐26

Standardization of measurements is critical for reproducibility

Variation between clinical centers and Core lab was lower for Echo than CT / MRI

Measurement technique should be 

standardized for:

‐ All imaging modalities

‐ All Age groups

Page 19: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

19

J Am Soc Echocardiogr 2015;28:119-82

GUIDELINES AND STANDARDS

asecho.org Guidelineswww.

Page 20: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

20

Measure perpendicular to the long-axis of the aorta

Measurement OptionsMeasurement Options

• Inner edge-inner edge

• Outer-outer

• Leading edge-leading edge

• Inner edge-inner edge

• Outer-outer

• Leading edge-leading edge

Page 21: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

21

Recommended time to measure

the aortic root

A. End‐systole

B. End‐diastole

- Greater reproducibility (Ao pressure more stable in late diastole)

- End-diastole easy to ID by QRS

Bossone, Yuriditsky, Asch et al. J Am Soc Echocardiogr 2016;29:166‐72

Differences in Measurement timing and technique are small

Page 22: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

22

TTE TEE CT and MRI

Page 23: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

23

Thoracic AorticAneurysms

Page 24: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

24

You want to explore the ENTIRE Aorta, AT LEAST ONCE

Acute AorticSyndromes

Page 25: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

25

TEE CT-scan

Dissection:Descending Thoracic Aorta

Mechanisms of Aortic RegurgitationA B

C

Page 26: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

26

Intramural Hematoma

Ulcerated Aortic Plaque

Page 27: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

27

Summary

Regardless of the Imaging Modality,methods should be unified

Leading edge, End diastole

Summary

Regardless of the Imaging Modality,methods should be unified

Leading edge, End diastole

Page 28: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

28

SummarySummary

• Indications for specific modality depends on:

• TTE used most often for aortic root assessment

• Indications for specific modality depends on:

• TTE used most often for aortic root assessment

- Accuracy for specific diseases- Availability- Cost/benefit ratio

SummarySummary

• CT-scan high resolution of entire aorta

• MRI greatest morphologic and dynamic

• TEE optimal procedure for guidance in OR

• CT-scan high resolution of entire aorta

• MRI greatest morphologic and dynamic

• TEE optimal procedure for guidance in OR

information without radiation, but lesswidely available

including arch, mesenteric, and renal vessels

safely performed in critically ill patients, even those on ventilators

Page 29: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

29

Diseases of the AortaDiseases of the Aorta

QUIZQUIZ

Which factor is most important to optimize Aortic diameter 

measurements reproducibility? 

Which factor is most important to optimize Aortic diameter 

measurements reproducibility? 

• A ‐ Measure always in end diastole

• B ‐ Use Leading edge to leading edge convention

• C ‐ Use same imaging modality and methods with side by side comparison

• D ‐ Only use CTA as it is more reliable and accurate

• A ‐ Measure always in end diastole

• B ‐ Use Leading edge to leading edge convention

• C ‐ Use same imaging modality and methods with side by side comparison

• D ‐ Only use CTA as it is more reliable and accurate

Page 30: Multimodality Imaging in Aortic DiseasesElongation and tortuosity can induce ... • CT-scan high resolution of entire aorta •MRI greatest morphologic and dynamic • TEE optimal

1/9/2018

30

Which factor is most important to optimize Aortic diameter 

measurements reproducibility? 

Which factor is most important to optimize Aortic diameter 

measurements reproducibility? 

• A ‐ Measure always in end diastole

• B ‐ Use Leading edge to leading edge convention

• C ‐ Use same imaging modality and methods with side by side comparison

• D ‐ Only use CTA as it is more reliable and accurate

• A ‐ Measure always in end diastole

• B ‐ Use Leading edge to leading edge convention

• C ‐ Use same imaging modality and methods with side by side comparison

• D ‐ Only use CTA as it is more reliable and accurate

Thank youThank you