Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0T
C
Comments and Guidelines
Aneurysm
Clip
Yes
C*
Yes
C* • Aneurysm clips made from ferromagnetic materials are contraindicated
for MR procedure, since excessive magnetically induced forces may displace
these clips, causing serious injury or death.
• By comparison aneurysm clips classified as non-ferromagnetic
(eg. Titanium alloy) have been tested and shown to be safe for patients
undergoing MR procedures at 1.5T or lower.
• All aneurysm clips must be checked and documented for MRI compatibility.
• Every aneurysm clip placed here at UCSF since 1985, is safe at both
1.5T and 3T. ( Sugita T2 aneurysm Clip, Yasargil Phynox aneurysm clip (FE),
Yasargil Titanium aneurysm clip (FT))
• Aneurysm clips places at an outside hospital, must have written documentation
stating the make, model, and date of insertion. This information must be reviewed
and confirmed for MRI compatibility before the patient is allowed in the scan room.
• Due to the increased artifacts generated by the 3 Tesla, the 1.5Tesla should be the
scanner of choice, unless specific high resolution imaging is needed and has been
approved by the radiologist.
http://www.mrisafety.com/safety_article.asp?subject=146
http://www.radiology.ucsf.edu/patient-care/patient-safety/mri
C*=Conditional, see comments
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0T
C
Comments and Guidelines
Breast
Tissue
Expander
No No • Adjustable breast tissue expanders and mammary implants are utilized for breast reconstruction following mastectomy, for the correction of breast and chest-wall deformities
and underdevelopment, for tissue defect procedures, and for cosmetic augmentation. These
devices are typically equipped with either an integral injection site or a remote injection
dome • it is recommended that a patient with a breast tissue expander that has a metallic
component be identified prior to MRI so that the radiologist is aware of the potential
problems related to the generation of artifacts as well as a possible injury.
• There are various breast tissue expanders that have magnetic ports to allow for a more
accurate detection of the injection site. These devices are substantially attracted to the
static magnetic fields of MR systems and, therefore, may be uncomfortable, injurious, or
contraindicated for patients undergoing MR procedures.
• Expanders which are unsafe for MRI include the following expanders:
Contour Profile Tissue Expander
McGhan Medical Breast Tissue Expander
Magna-Site Tissue Expander
http://www.mrisafety.com/safety_article.asp?subject=16
MRI IMPLANT TABLE EDITED MARCH 2013
Cardiac
Loop
Recorder
Yes
C*
Yes
C*
Reveal Plus (Model # 9526) Insert able Loop Recorder (ILR)
Magnetic and Radio Frequency (RF) fields produced by MRI may adversely affect the data being
stored by the Reveal Plus Insert able Loop Recorder (ILR).
Since the ILR contains ferromagnetic components, the strong magnetic field of the MRI system
may apply a mechanical force on the ILR. The patient may be able to feel this magnetic force on
the ILR. While this does not represent a safety hazard, the patient should be made aware of this
possibility to avoid undue patient concern.
Reveal DX (Model # 9528) and Reveal XT (Model # 9529) Insert able Cardiac Monitor
Non-clinical testing demonstrates that Reveal DX and Reveal XT devices are safe for use in the
MRI environment when used according to these instructions.
The Reveal DX and Reveal XT can be safely scanned in patients under the following conditions:
MRI IMPLANT TABLE EDITED MARCH 2013
MRI Equipment Requirements
• The MRI equipment must be a closed bore, cylindrical magnet with a static magnetic field
of 1.5 Tesla (T) or 3.0 T.
• MRI equipment must be used in normal operating mode (based on standards defined in
IEC 60601-2-33). Refer to Reveal DX or Reveal XT Clinician Manual for additional
information.
• If local or surface coils are needed, please refer to the Reveal DX or Reveal XT Clinician
Manual for additional information.
MRI Procedural Requirements
• Verify that the Reveal DX or the Reveal XT is at least 6 weeks post-implant. This
waiting period allows sufficient time for implant pocket and wound healing and minimizes
the effects of device “tugging” caused by the magnetic fields.
• Verify that the Reveal DX or Reveal XT implant location is in the subcutaneous tissue of
the chest region.
• Uninterrupted duration of active scanning (when RF and gradients are on) over the chest
during MRI must not exceed 30 minutes. A waiting period of at least 10 minutes is required
if additional chest scans are necessary.
• Check that additional implantable devices are not present, including abandoned pacing leads.
• Medtronic has not tested interactions with all other implanted devices or abandoned pacing leads.
• The MRI procedure may overwrite the recorded data in the Reveal DX or the Reveal XT. Print
or save the data stored in the Reveal DX or Reveal XT to diskette before the MRI.
MRI IMPLANT TABLE EDITED MARCH 2013
• The Patient Assistant and the Medtronic CareLink Model 2090 Programmer are not MRI safe
and should not be brought into MRI controlled room (magnet room). Continued below…….
Post-MRI operation
• Check the programmed parameters of the Reveal DX or the Reveal XT after the MRI
procedure.
• Print or save the data collected during the MRI procedure to diskette because the MRI procedure
may temporarily affect the Reveal DX or the Reveal XT’s event detection and recording. The date
and time of the MRI procedure should be recorded under the Patient icon in the notes section for
future reference. Note: The Cardiac Compass trend data cannot be cleared and might show
irregularities at the time of the MRI operation.
www.mrisafety.com/safety_info.asp
Device 1.5
T C
3.0T
C
Comments and Guidelines
Cardiac
Pacemaker
Yes
C*
No The following guidelines have been set here at UCSF to scan a patient with a pacemaker or ICD:
• A risk- benefit ratio must be established and benefit from the MRI examination
MRI IMPLANT TABLE EDITED MARCH 2013
should clearly outweigh the risks of the procedure.
• MR imaging in the thoracic region is subject to additional risk and this
should be considered when evaluating the benefit to risk for a specific patient.
• Multiple MR exams in a single patient should also be avoided, as there is potential
for increased risk and this should needs to be considered when determining if the
benefits clearly outweigh the risks.
• The study should not be performed if similar clinical information could be
obtained by a different image modality
• The MRI scan must be approved by an attending radiologist and an
electro-physiologist.
• For Pacemaker patients: The patient must be non-pacemaker dependent.
• For ICD patients: Risks associated with MR examination in patients with ICD
are expected to be higher than in patients with pacemakers. Therefore,
MR examination of patients with ICDs should not be performed unless there
are highly compelling circumstances and when the benefits clearly outweigh
the risks. ICD patients must be non-pacemaker dependent.
• The patient must have the pacemaker/ ICD in place for more than 6 weeks.
• The patient must be not at or near Elective Replacement Indicator (ERI).
• Abandoned temporary epicardial (cut at skin) leads are not a contraindication
for MRI (please see images a. and b. bellow). Abandoned/capped trans venous
leads are a contraindication to MRI (please see image c. bellow).
MRI IMPLANT TABLE EDITED MARCH 2013
c.
MRI IMPLANT TABLE EDITED MARCH 2013
• Prior to the scan date, a cardiology consult must be scheduled, where it will be determined if the
pacemaker/ICD patient is safe to have an MRI. At this time, a written informed consent will be
established. If a patient is referred from and outside physician, a visit to the UCSF device clinic
should be scheduled before the MRI is acquired.
• An electro-physiologist, or cardiologist or device nurse (with pacemaker/ICD monitoring
experience) must be present before, during, and after the MRI exam to validate the performance
integrity of the pacemaker/ICD prior to the MRI exam, monitor the patient during the exam, and
check the pacemaker/ICD performance after the exam. An electro-physiologist should be
carrying a pager and be within quick reach of the MRI scanner during the entire examination. For
patients with ICDs, a physician with electrophysiological expertise should perform post scan
device reprogramming and defibrillation threshold testing. Arrangements should be made for the
patient to follow-up with electro physiologist at 3 months after the MRI.
• The patient should be monitored continuously during the MR procedure (e.g., blood pressure,
pulse rate, oxygen saturation, and ECG).
• Appropriate personnel, a crash cart, and defibrillator must be available throughout the procedure
to address an adverse event.
• Maintain visual and voice contact throughout the procedure with the patient.
• The patients must be able to communicate in case any complications arise.
• Instruct the patient to alert the MR system operator of any unusual sensations or problems so
that, if necessary, the MR system operator can immediately terminate the procedure.
• The patients will only be scanned during normal workday hours (8A-5P) on 1.5T scanners, only
in hospital setting with appropriate trained personnel. For urgent, on-call cases, scans may need
MRI IMPLANT TABLE EDITED MARCH 2013
to be done after 5P or on weekends, although it is preferable to wait until normal workday hours
if possible.
• The MRI will only be scheduled during the day and appointments will be coordinated to ensure
that an MRI physicist and an attending radiologist are involved to tailor the study for minimum
SAR while obtaining the essential clinical information.
• Patients with MR conditional Medtronic pacemaker (Revo MRI SureScan pacing system) can
undergo an MRI study only if the MR isocenter is superior to the C1 vertebra or inferior to the
T12 vertebra. In that case, Medtronic specific guidelines should be followed. In case MR
isocenter is contained between C1 and T12, specific MR safety protocols described above should
be followed.
• The current protocol is effective now for cardiac imaging and neuroimaging studies. Additional
clinical indications will be considered in the future.
http://www.mrisafety.com/safety_article.asp?subject=18
http://www.radiology.ucsf.edu/patient-care/patient-safety/mri
http://www.medtronic.com/for-healthcare-professionals/products-therapies/cardiac-rhythm/pacemakers/revo-
mri-pacing-system/index.htm
C*=Conditional, see comments
MRI IMPLANT TABLE EDITED MARCH 2013
Device
1.5
T C
3.0T
C
Comments and Guidelines
Cochlear
Implants
No No Some types of cochlear implants employ a relatively strong cobalt samarium magnet used in conjunction
with an external magnet to align and retain a radio frequency transmitter coil. Other types of cochlear
implants are electronically activated. Consequently, MR procedures are strictly contraindicated in
patients with these types of implants because of the possibility of injuring the patient and/or damaging or
altering the function of the cochlear implants.
Device 1.5
T C
3.0T
C
Comments and Guidelines
Deep Brain
Stimulator
(DBS)
Yes
C*
No • Deep brain stimulation (DBS) is a surgical procedure used to treat a variety
of disabling neurological symptoms—most commonly the debilitating symptoms
of Parkinson’s disease.
• The DBS system consists of three components: the lead, the extension, and the
neurostimulator. The neurostimulator (the "battery pack") is usually implanted
under the skin near the collarbone. In some cases it may be implanted lower in
the chest or under the skin over the abdomen.
• The greatest concern for electronically activated or electrically conductive implants
in the brain is excessive MR imaging-related heating, which can cause irreversible
tissue damage.
• This may lead serious injury to the patient, including the possibility of transient
dystonia, paralysis, coma, or even death.
• The Deep brain stimulator must turn off and checked before the exam, and turned
MRI IMPLANT TABLE EDITED MARCH 2013
on and checked after the exam. The primary contact is Monica Volz, Pg 443-5493.
• Patients with DBS implants can have MRI imaging of the brain with a head coil
(transmit-receive head coil, where the RF is not transmitted by the body coil
• Imaging of parts of the body other than the head is prohibited.
http://www.medtronic.com/your-health/parkinsons-disease/therapy/our-dbs-therapy-products/activa-
rc/index.htm
MRI IMPLANT TABLE EDITED MARCH 2013
Device
1.5
T C
3.0T
C
C* = Conditional, see comments
Comments and Guidelines
Drug
Infusion
Pump
Yes
C* No A drug infusion pump is used for automatic delivery of antineoplastic agents, morphine, or narcotics.
The infusion pump has a motor that may have ferromagnetic properties, a magnetic switch, and is
programmed by telemetry. Potential issues pertaining to MRI imaging of patients with a drug infusion
pump include the following:
1) Stopping of needed medication
2) Temporary or permanent stall of the motor
3) Local tissue heating adjacent to implant during imaging
4) Increase potential for Peripheral nerve stimulation (PNS)
Although many drug infusion pumps are contraindicated for MRI, there are currently some pumps that
can be exposed to the MRI environment during imaging under strict conditional requirements. Always
review the manufacturer guidelines for the specific pump before introducing the patient into the
magnet room and proceeding with the exam. Strictly adhere to the manufacturer’s conditional
requirements for managing the pump prior to, during, and post imaging. The following are
examples of conditional drug infusion pumps:
AccuRx Constant Flow Implantable Pump (St. Jude Medical, Plano, TX)www.sjm.com
SynchroMed (II,EL) , implantable drug infusion device (Medtronic Inc.)http://www.medtronic.com
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0T
C
C* = Conditional, see comments
Comments and Guidelines
External
Ventricular
Drain
Yes Yes • An external ventricular drain (EVD) is a device used in neurosurgery that relieves raised
intracranial pressure and monitors CSF fluid levels.
• The EVDs utilized here at UCSF is the Integra Limitorr, and the Integra MoniTorr.
• Both are safe to use on either the 1.5 or 3.0T. This device does not cause artifacts.
http://www.integra-ls.com//PDFs/NeuroCriticalCare/Limitorr%20Sales%20Brochure.pdf
Device 1.5
T C
3.0T
C
Comments and Guidelines
Integra
Camino
Bolt
No No • Camino Bolt catheter is an intracranial pressure monitoring device. It consists of a cranial bolt
surgically implanted into the skull in which a pressure transducer is placed. At the end of the
pressure transducer is a transducer connector, which is attached to a cable that is attached to a
monitor.
• The Camino Bolts utilized here at UCSF is made by Integra Neurosciences. Currently, the FDA
MRI IMPLANT TABLE EDITED MARCH 2013
has requested that Integra revalidate their results concerning the Camino Bolt and its MR
compatibility. Until such time that the manufacturer has completed their revalidation process and
can produce a document that states that Camino Bolt and the various models are MRI safe in a
1.5T and or a 3.0T MR environment, Camino Bolts will be contraindicated for both the 1.5T and
3.0T.
www.integra-ls.com/PDFs/Camino/Camino%20110-4HM%20PI.PDF
Device 1.5
T C
3.0T
C
Comments and Guidelines
InterStim
Therapy
Yes
C*
No • InterStim Therapy is similar to a pacemaker for the bladder. The implantable device uses mild
electrical pulses to target the communication problem that exists between the sacral nerves and
the brain to tell the bladder to work correctly.
• The therapy treats urinary retention and the symptoms of overactive bladder, including urinary
urge incontinence and significant symptoms of urgency-frequency.
• As with any neurostimulator, only the brain can be imaged using a transmit-receive head coil.
http://www.mrisafety.com/safety_article.asp?subject=158
http://professional.medtronic.com/pt/uro/snm/prod/interstim/manuals-technical-
resources/index.htm
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0T
C
Comments and Guidelines
Linux
Reflux
Manageme
nt
System
No No • Consists of a series of titanium beads, each with a magnetic core, connected together with
titanium wires to form a ring shape. It is implanted around the lower end of the
esophagus to help control gastro esophageal reflux (GERD).
• Patients who have the LINX device should not be exposed to , or undergo, an MRI.
Exposure to MRI could cause serious injury to the patient and the device may damaged.
http://www.fda.gov/MedicalDevices/ProductsandMedicalProcedures/DeviceApprovalsandClearances/Re
cently-ApprovedDevices/ucm300790.htm
MRI IMPLANT TABLE EDITED MARCH 2013
Device
1.5
T C
3.0T
C
C* = Conditional, see comments
Comments and Guidelines
Magnimplant
Magnatract
Sternum
Implant
No No Magnimplant” and “Magnatract”in a combined system to correct for pectus excavatum or sunken chest
deformity, in pediatric patients. As part of the system “Magnimplant” is an internal magnet that is
surgically placed behind the sternum, and “Magnatract” is an external brace with a magnet implanted.
The devices combine to exert a defined magnetic force to the chest wall.
This device system is totally contraindicated for MRI at any field strength.
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0
T C
Comments and Guidelines
Programmable
Ventricular
Shunt
Yes
C*
Yes
C*
Programmable Valves are implantable devices that provide constant
Intra-ventricular pressure and drainage of CSF for the management of hydrocephalus.
• Subjecting the valve to strong magnetic fields may change the setting of the valve(shunt)
The use of Magnetic Resonance (MR) systems up to 3-Tesla will
not damage the valve mechanism, but may change the setting of the valve. Confirm the
valve setting after an MRI procedure.
• The performance level setting should always be checked before and after MRI exposure. Contacts to check shunt: Pedi – Carolyn 443-4726, Adult – Gwen 443-
4769
• The following programmable shunts have been approved for the 3.0 Tesla MRI: Codman Hakim (Johnson and Johnson), Delta Shunt (Medtronics),proGAV
programmable valve (Aesculap), Pulsar Valve (Sophysa USA), Strata
(Medtronics), Sophy Adjustable pressure valve (Sophysa USA)
• Due to the increased artifacts generated by the 3 Tesla, the 1.5Tesla should
be the scanner of choice, unless specific high resolution imaging is needed and has been
approved by the radiologist.
•
http://www.mrisafety.com/safety_article.asp?subject=175
MRI IMPLANT TABLE EDITED MARCH 2013
Device 1.5
T C
3.0
T C
Comments and Guidelines
Vagus
Nerve
Stimulator
Yes
C*
Yes
C* • The VNS Therapy generator is an implantable, programmable pulse generator that delivers a
precise pattern of stimulation to the left vagus nerve.
• MRI procedures should not be performed on patients
with VNS Therapy Systems who have a Lead break. A broken Lead
should be removed prior to MRI. Suspected Lead breaks should be
confirmed by performing appropriate diagnostic procedures and
consultation with Cyberonics. Broken Lead wires present increased
risk of thermal injury to patients during MRI procedures
• Magnetic resonance imaging (MRI) should not be performed with a magnetic resonance body
coil in the transmit mode. The heat induced in the Lead by an MRI body scan can cause injury.
• Imaging restricted to brain or head imaging utilizing a transmit-receive head coil.
• The device must be turned off (Pulse generator output programmed to 0.0 mA) prior to the exam
and the device checked and reset after the exam. See below.
MRI IMPLANT TABLE EDITED MARCH 2013
http://us.cyberonics.com/en/vns-therapy/healthcare-professionals/manuals/
http://www.mrisafety.com/safety_article.asp?subject=41
•