Upload
osman
View
66
Download
0
Tags:
Embed Size (px)
DESCRIPTION
MQXC Nb -Ti 120mm 120T/m 2m models. G.A.Kirby MQXC has been running for 4 to 5 years and has had many contributors to the design, construction ,testing From CERN, CEA, US labs and industry. Over view. MQXC cross section reminder. Aperture Zero assembly over view - PowerPoint PPT Presentation
Citation preview
1
MQXCNb-Ti 120mm 120T/m 2m models
G.A.Kirby
MQXC has been running for 4 to 5 years and has had many contributors to the design, construction ,testing
From CERN, CEA, US labs and industry
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
2
Over view
• MQXC cross section reminder.• Aperture Zero assembly over view• Heat extraction features (that can be used on all Hilumi magnets)
• collared aperture Warm Field measurements• yokes magnet Warm Fields measurements.• Aperture Zero vertical test plan • Project plan for magnets Zero, One ,& Two• Teat Extraction Horizontal test plan.
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
3
MQXC Magnet Cross Section
Collaring Keys
Yoke
Yoke key
Pressuer vessel shellTwo layer coil
Axial assembly Rod
Holes for heat exchangers
Magnetic shims
Main Collar Insert Collar
Quench heater
Bean screen
Yoke fixing pin
Bus bar, assembly, lifting, feathers.
Main Magnet Parameters
Gradient 118 T/mAperture 120 mmWorking point
80% SS
Current 12’850 A
Open shoe
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
4
Heat extraction features
Collaring Model with open ground insulation The yellow open structure provides helium path from coils to the heat exchangers
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
5
Thanks to Pier Paolo Granieri for cable images
Heat extraction features
Collars spacing give 6% open
Open collaring shoes
Open quench heater
We see the open helium paths through the “fish-bones’ and the open quench
heater between layers.
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
6
MQXC project over view-Coil manufacture for 3 apertures completed.-99% of components delivered for all 3 magnets.(We are only missing the half shells , which have been replaced by 4 rods for vertical test)
-Magnet zero Warm field measurements very good result! Waiting for 1.9K test in SM18. (end May/ June). Cold test will start to measure heat extraction. (new expected heat load ~860 W @ 1.9K !)
-Magnet 1 Aperture assembly started. (cold test June /July)
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
7
Pole assembly
Quench heater mounted on top of layer one coil
Soldering the inter layer joint
Fish-bone on top of layer 2
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
8
Magnet assembly over view
Vertical assembly 9m magnet will need horizontal assembly
Large movements during compressing coil, insulation and collars 2 to 3 cm on dia.
Full length collaring keys
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
9
End flange and joint box
End flanges welded to
collaring keys
Multi layer joint box
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
10
4 – 80mm diameter rods temporary replacement for outer shell
Yoke packing factor too high at 99.6%
we need to reduce to 98% for the next
magnet .
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
11
Aperture Warm Field Measurementswithout Yoke
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
12
Yoked magnet Warm field measurement with large coil set.
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
13
MQXC Magnetic shims. 8 rods in the collars, and 8 plates between collar’s and yoke.
We can adjust final field quality + or – several units
Magnetic shims can me mounted or removed easily
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
14
Results from Warm Field measurements in yoked magnet zero
b3 a3 b4 a4 b5 a5 b6 a6 b7 a7 b8 a8 b9 a9 b10 a10 b11 a11 b12 a12 b13 a13 b14 a14 b15 a15
-1.0
-0.6
-0.2
0.2
0.6
1.0Integral Harmonics @ 27 mm
10A 15A 20A
Units
Magnetic measurements of the MQXC-1 show all together a very slightly improved field quality with respect to the collared coil. The mulitpole components at 27 mm reference radius are all below 1 unit in 10-4. The effect of the iron yoke on the allowed multipoles is slightly lower than expected (-0.1 instead of -0.48 units for the b6 and -0.005 instead of -0.02 for the b10). It will have to be seen if this is confirmed with the new and more appropriate search coil. The b6 component in the straight section is about 1 unit higher than expected, but in the right direction. The b6 could be further reduced by installing all the tuning rods. This would be useful also for the field quality at nominal excitation, because the saturation effect on the iron yoke is only of the order of 0.4 units.
Thanks to Prof. Stephan Russenschuck and his team
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
15
Vertical 1.9 K test plan. • protection test set up dump, QH testing "• training to high field• magnetic field measurement • joint resistance check• beam heating tests, with inter-pole heaters and QH • measurement of Ramp rate loss • thermal cycle to room temp• Change magnetic shims ?• cool to 1.9K • Repeat high field training and field measurement
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
16
MQXC (0,1,&2 Magnet Planning ) for 2012 plan
Magnet Assy “0”SM 18 Cold testing
Magnet “0”
Assy Magnet “1”
JuneEaster August
Assy Magnet “2”
17
Heat Extraction Cold Horizontal Testing
• We plan to use a std. MQ cryostat• The horizontal test is to see if we can
extract the equivalent large heat load (860W over the triplet) (see Rob’s talk later )
• Heaters on the cold bore and in the magnet simulate the beam beating ~ 20W on the 2m magnet
2M- MQXCSM18 Test stand
MQ vacuum vessel
Valve box to feed two heat exchanger tubes
through magnet
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
Glyn A Kirby, LARP CM18/HiLumi LHC Meeting May 2012
18
We are Pushing on with MQXC hope to finish vertical testing this year
The End