f Monitor of Beam in the Abort Gap Randy Thurman-Keup DOE Review of Tevatron Operations at FNAL March 29-31, 2005 a.k.a. Abort Gap Integrator (AGI)

Monitor of Beam in the Abort Gap

Embed Size (px)

DESCRIPTION

Monitor of Beam in the Abort Gap. a.k.a. Abort Gap Integrator (AGI). Randy Thurman-Keup DOE Review of Tevatron Operations at FNAL March 29-31, 2005. AGI – TeV Bunch Structure. Bunch. 21 buckets. 1113 RF buckets total. The Tevatron operates with 36 bunches in 3 groups called trains - PowerPoint PPT Presentation

Citation preview

Page 1: Monitor of Beam in the Abort Gap

fMonitor of Beam in the

Abort Gap

Randy Thurman-Keup

DOE Review of Tevatron Operations at FNAL

March 29-31, 2005

a.k.a. Abort Gap Integrator (AGI)

Page 2: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 2

AGI – TeV Bunch Structure

The Tevatron operates with 36 bunches in 3 groups called trains

Between each train there is an abort gap that is 139 RF buckets long RF bucket is 18.8 ns Abort gap is 2.6 s

139 buckets

21 buckets 1113 RF buckets totalTrain

Bunch

Abort Gap

Page 3: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 3

AGI - Motivation During an abort

Abort kicker magnet ramps up during abort gap Beam in the abort gap is directed towards magnets, CDF,

etc… Quenches (in the past, 10 x 109 caused quenches) CDF silicon detector damage

Previous monitors relied on counters external to the beampipe that were timed with abort gap Measured stuff leaving the abort gap, not stuff still in it

Use synchrotron radiation to directly measure abort gap beam Want to be sensitive to a DC beam that is 1 part in 104 of

the total beam

Page 4: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 4

AGI - Apparatus Made use of existing synchrotron light

system Measures beam profile

Added beam splitter and gated photomultiplier tube Photomultiplier had to be gateable and

insensitive to light present just before the gate (bunch intensity is several thousand times brighter than DC beam) Custom gating circuit for generic photomultiplier

Pulse two dynodes Hamamatsu gated MCP style PMT on loan from LBNL

Page 5: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 5

AGI - Apparatus

Al pickoff mirror

Quartz Window

X,Y mirrorSplitter

Optional 100:1 Filter

Beampipe

Synch Light Box

Optional 25:1 Filter

Image intensifier,CID etc…

SyncLite System

X,Y mirror

Narrowband blue filter

Lens

Synchrotron Light

Page 6: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 6

AGI – Data Acquisition

PMT signal is integrated and digitized using locally built integrator and commercial VME ADC

Readout is through a VxWorks front end which also controls the timing of the PMT and integrator gates

Average 1000 samples in each of the 3 abort gaps

Page 7: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 7

AGI – Commissioning

Installed ~June 2004 LBNL gated MCP-PMT installed ~July 2004 and

in use since then Calibrations

Bunch based (using an optical attenuator) TeV Electron Lens (TEL) based

TEL normally cleans the abort gaps Turn TEL off and compare growth of abort gap beam

with total beam from DCCT measurement Looked at abort gap structure using

photon counting

Page 8: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 8

AGI – TEL Study

TEL

TotalBeam

AGI’s

• When TEL is turned off, DC beam stops being ejected and accumulates around the ring including the abort gap

The beam loss rate decreases

• Determine excess beam by measuring deviation of total beam from projected value with TEL on

December 2004

Page 9: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 9

AGI – TEL Study

TEL

Abort gap beamper total beam deviation from TEL on

Abort gap beamper AGI (3 gaps)

Abort gap beamper Synclite

June 2004

Projected total beam from TEL on subtracted from total beam

Page 10: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 10

AGI - Status

TEL

TeV Energy

Beginning of storeMarch 2005

AGI1, AGI2, andAGI3 are measurementsof the abort gap beamat three differentpositions withinthe abort gap

Their behavior dependson their position relativeto the action of the TEL

AGI1

AGI2

AGI3

AGI2 AGI1 AGI3

Abort Gap

TEL On

Page 11: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 11

AGI – Status• Longitudinal Damper problem• Caused beam to be shaken out

of the bunches

Feb 21, 2005

Page 12: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 12

AGI – Longitudinal Structure

0 50 100 150 200 250 300 350 4000

0.01

0.02

0.03

0.04

0.05

0.06

0.07

0.08

0.09

0.1

Time (2ns)

Bunch 1 Bunch 36

End of Abort Gap 3

Bunch 36 is 2 s to the left of this plot

The red crosses show the location of the center of the RF buckets.

These data were collected using a PMT pulse-over-threshold counting technique These small

bunches are ~105 smaller than a normal bunch

Page 13: Monitor of Beam in the Abort Gap

f

3/30/2005 R. Thurman-Keup - DOE Review 13

AGI – Conclusion

Integral part of TeV operations Checked by the sequencer before normal beam

aborts Watched by CDF and MCR

i.e. I get paged when it hiccups

Next Steps Measure pbars in abort gaps (TeV request) Consider separating the abort gap monitor from

the normal synchrotron light system