MODIS and VIIRS TEB Performance

Chris Moeller, Greg Quinn

University of Wisconsin

MODIS/VIIRS Calibration Workshop

November 18, 2019

Page 2: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

20 22 23 24 25

27 28 29 30 31

32 33 34 35 36

Terra MODIS (C6.1) SNOs with MetOp-A IASI

Page 3: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

20 22 23 24 25

27 28 29 30 31

32 33 34 35 36

Aqua MODIS (C6.1) SNOs with MetOp-A IASI

Page 4: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

Switch to VIIRS V3 L1B

VIIRS – CrIS Comparisons

SNPP N-20

Page 5: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

VIIRS – CrIS Comparisons

SNPP N-20

A LUT correction is planned for N-20 L1B processing that will reduce scene temperature dependence.

Page 6: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

VIIRS – IASI Comparisons

SNPP N-20

A LUT correction is planned for N-20 L1B processing that will reduce scene temperature dependence.

Page 7: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

SNPP VIIRS SNOs with MetOp-A IASI

M12 M13I4

M14 M15 I5 M16

Page 8: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

N-20 VIIRS SNOs with MetOp-A IASI

M12 M13I4

M14 M15 I5 M16

Page 9: MODIS and VIIRS TEB Performance · Ja124_2018-Nov05 2019 SNOs: N20 Avg VilRS - IASI A; All Bands M12 M13 M14 M15 M16 0.6 -0.6 200 220 VIIRS 240 260 280 (K) 300 Brightness Temp Mae,2012

Summary

• Terra MODIS shows trending in PVLWIR (B27-30) that could be problematical for climate studies. Electronic Xtalk likely remains root cause.

• Aqua MODIS trends more stable but watching PVLWIR.

• SNPP VIIRS trending well.

• Some differences in biases between SNPP and N-20 VIIRS as a function of scene temperature.