GeneralSpecifications
Yokogawa Electric Corporation2-9-32, Nakacho, Musashino-shi, Tokyo, 180-8750 JapanTel.: 81-422-52-5617 Fax.: 81-422-52-6792
Model FLXA212-Wire Analyzer
GS 12A01A02-01E
n GeneralThe model FLXA21® 2-Wire Analyzer, one model of FLEXA® series, offers single or dual sensor measurement. The modular-designed analyzer offers 4 kinds of measurements – pH/ORP (oxidation-reduction potential), contacting conductivity (SC), inductive conductivity (ISC) or dissolved oxygen (DO) – with the respective sensor module.For dual sensor measurement, the combination of two same type sensor inputs – pH/ORP and pH/ORP (analog sensor only), SC and SC, and DO and DO – are available with two sensor modules. Dual sensor measurement offers additional functionalities; calculated data function and redundant system.Variety of calculated data from two measuring parameters is selectable for each measurement. On the redundant system built on two measuring parameters of two sensor inputs, main output parameter is automatically switched over to the second sensor output in case of the main sensor’s failure condition.Addition to conventional analog pH/ORP sensors, the analyzer FLXA21 can be connected to Yokogawa’s digital sensor, FU20F pH/ORP SENCOM Sensor.
In the FLXA21 Human Machine Interface (HMI), 2-wire type analyzer FLXA21 offers easy touch screen operation and simple menu structure in 12 languages. Menus of display, execution and setting are displayed in a selected language.The analyzer FLXA21 automatically recognizes the installed sensor module and prepares the necessary menus for right configuration, even for dual sensor measurement.For immediate measurement, the FLXA21 offers quick setup functionality. The quick setup screen appears when the analyzer is powered. Only a few setups – date/time, language, basic sensor configurations and output – will start the measurement.
The FLXA21 offers the best accuracy in measurement with temperature compensation functionality and calibration functionality. Sensor diagnostics and sensor wellness indication make measurement reliable. Logbook of events and diagnostic data is a useful information source for maintenance.
For the wide range of industrial environment, the FLXA21 is designed with the enclosure of plastic, stainless steel or stainless steel with corrosion-resistant coating. And, for hazardous location, the FLXA21 has approvals of ATEX, IECEx, FM, CSA and NEPSI.
nFeatures•4kindsofmeasurements;pH/ORP,SC,ISCandDO•Dualsensormeasurementon2-wiretypeanalyzer;pH/ORPandpH/ORP,SCandSC,andDOandDO
•Calculateddatafromdualsensormeasurement•Redundantsystemondualsensormeasurement•ConnectionofdigitalFU20FpH/ORPSENCOMSensor
•Easy touch screen operation on 2-wire type analyzer•SimpleHMImenustructurein12languages•Quicksetupmenuforimmediatemeasurement•Indicationofsensorwellness•Enclosure – plastic, stainless steel or stainless steel
with corrosion-resistant coating•Hazardouslocationapprovals–ATEX,IECEx,FM,CSAandNEPSI
GS12A01A02-01E©Copyright Mar. 2010
8th Edition Jan. 06, 2014
2
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
nGeneralSpecifications1. Basic MeasurementObject/SensorType•pH/Oxidation-reductionPotential(pH/ORP)(analog
sensor)•Conductivity(SC)•InductiveConductivity(ISC)•DissolvedOxygen(DO)•pH/Oxidation-reductionPotential(pH/ORP)(digital
sensor)Note: Theavailablemeasurementobjectdependsona
sensormoduleinstalledontheanalyzer. AnalyzerStructure
Module structureCompositionofAnalyzerOne(1)HousingassemblyOne(1)ortwo(2)Sensormodules
CombinationofSensorModulewhentwomodulesareinstalledCombinationsoftwosamesensormodulesareavailable; pH/ORPandpH/ORP(analogsensor) SCandSC DOandDO
2. Measurement2-1. pH/Oxidation-reductionPotential(pH/ORP)
withanalogsensors InputSpecificationDualhighimpedanceinput(≥1012Ω)
InputRangepH: -2to16pH(withoption/K:0to14pH)ORP: -1500to1500mVrH: 0to100rHTemperature:
Pt1000: -30to140ºCPt100: -30to140ºC6k8: -30to140ºCPTC10k: -30to140ºCNTC8k55: -10to120ºC3kBalco: -30to140ºCPTC500: -30to140ºC
OutputRangepH: min.span1pH max.span20pHORP: min.span100mV max.span3000mVrH: min.span2rH max.span100rHTemperature:min.span25ºC max.span170ºC
Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)pH
Linearity: ±0.01pHRepeatability:±0.01pHAccuracy:±0.01pH
ORPLinearity: ±1mVRepeatability:±1mVAccuracy:±1mV
TemperaturewithPt1000,6k8,PTC10k,NTC8k55,3kBalco,
PTC500Repeatability:±0.1ºCAccuracy:±0.3ºC
withPt100Linearity: ±0.4ºCRepeatability:±0.1ºCAccuracy:±0.4ºC
2-2. Conductivity(SC) InputSpecificationTwoorfourelectrodesmeasurementwithsquarewaveexcitation,usingmax60m(200ft)cable(WU40/WF10)andcellconstantsfrom0.005to50.0cm-1
InputRangeConductivity:
min.: 0µS/cmmax.: 200mSx(Cellconstant) (overrange2000mS/cm)
Resistivity:min.: 0.005kΩ/(Cellconstant)max.: 1000MΩxcm
Temperature:Pt1000: -20to250ºCPt100: -20to200ºCNi100: -20to200ºCNTC8k55: -10to120ºCPb36(JISNTC6k): -20to120ºC
OutputRangeConductivity: min.0.01µS/cm max.2000mS/cm(max90%zero
suppression)Resistivity: min.0.001kΩxcm max.1000MΩxcm(max90%zero
suppression)Temperature: min.span25ºC max.span270ºC
Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)Conductivity
2µSxKcm-1to200mSxKcm-1
Accuracy:±0.5%F.S.1µSxKcm-1to2µSxKcm-1
Accuracy:±1%F.S.Resistivity
0.005kΩ/Kcm-1to0.5MΩ/Kcm-1
Accuracy:±0.5%F.S.0.5MΩ/Kcm-1to1MΩ/Kcm-1
Accuracy:±1%F.S.Temperature
withPt1000,Pb36,Ni100Accuracy:±0.3ºC
withPt100,NTC8k55Accuracy:±0.4ºC
TemperaturecompensationNaCltable: ±1%Matrix: ±3%
Stepresponse: 90%(<2decades)in7secondsNote: "F.S."meansmaximumsettingvalueofanalyzer
output. "K"meanscellconstant. YOKOGAWAprovidesconductivitysensorsofwhich
cellconstantsare0.1to10cm-1.
Jan. 06, 2014-00
3
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
2-3. InductiveConductivity(ISC) InputSpecificationCompatiblewiththeYokogawainductiveconductivityISC40serieswithintegratedtemperaturesensor:NTC30korPt1000.
InputRangeConductivity: 0to2000mS/cmat25ºCreference
temperature.Temperature: -20to140ºCCablelength: max.60meterstotallengthoffixedsensor
cable+WF10(J)extensioncable. Influenceofcablecanbeadjusted
bydoinganAIRCALwiththecableconnected to a dry cell.
OutputRangeConductivity:
min.span:100µS/cmmax.span: 2000mS/cm(max90%zero
suppression)Temperature: min.span25ºC max.span160ºC
Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)(Outputspanis0-100µS/cmormore)Conductivity:
Linearity: ±(0.4%F.S.+0.3µS/cm)Repeatability:±(0.4%F.S.+0.3µS/cm)
Temperature:±0.3ºCStepresponse: 90%(<2decades)in8secondsNote: "F.S."meansmaximumsettingvalueofanalyzer
output.2-4. DissolvedOxygen(DO) InputSpecificationTheFLXA21acceptsoutputfrommembranecoveredDissolvedOxygensensors.ThesesensorscanbeGalvanictype,wherethesensorgeneratesitsowndrivingvoltageorPolarographictype,wherethesensorusesexternaldrivingvoltagefromtheconverter.Theinputrangeis0to50µAforGalvanicsensorsand0to1microAforPolarographicsensors.Fortemperaturecompensation,theFLXA21acceptsPt1000(DO30sensor)andNTC22kelements(OXYFERMandOXYGOLDsensors).
InputRangeDO30sensor: DissolvedOxygen:0to50mg/l(ppm) Temperature: -20to150ºC Note:ProcesstemperatureforDO30is0to40ºC
Hamiltonsensors:Oxyferm:
Measurementrange:10ppbto40ppmTemperaturerange: 0to130ºC
OxygoldG:Measurementrange:2ppbto40ppmTemperaturerange: 0to130ºC
OxygoldB:Measurementrange:8ppbto40ppmTemperaturerange: 0to100ºC
OutputRangeDOconcentration:
mg/l(ppm):min.: 1mg/l(ppm)max.: 50mg/l(ppm)
ppb:min.: 1ppbmax.: 9999ppb
%saturation:min.: 10%max.: 600%
Temperature:min.span25ºCmax.span170ºC
Performance(Accuracy)(Thespecificationsareexpressedwithsimulatedinputs.)Performanceinppmmode:
Linearity: ±0.05ppmor±0.8%F.S.,whicheverisgreater
Repeatability:±0.05ppmor±0.8%F.S.,whicheveris greater
Accuracy:±0.05ppmor±0.8%F.S.,whicheverisgreater
Performanceinppbmode:Linearity: ±1ppbor±0.8%F.S.,whicheveris
greaterRepeatability:±1ppbor±0.8%F.S.,whicheveris
greaterAccuracy:±1ppbor±0.8%F.S.,whicheveris
greaterTemperature
Linearity: ±0.3ºCRepeatability:±0.1ºCAccuracy:±0.3ºC
Note: "F.S."meansmaximumsettingvalueofanalyzeroutput.
2-5. pH/Oxidation-reductionPotential(pH/ORP)withdigitalsensor,FU20FpH/ORPSENCOMSensor
InputSpecificationBi-directionaldigitalcommunication(RS-485)
betweenFU20FandFLXA21 InputRange(dependingonFU20F)pH: 0to14pHORP: -1500to1500mVrH: 0to100rHTemperature:-10to105ºC
OutputRangepH: min.span1pH max.span20pHORP: min.span100mV max.span3000mVrH: min.span2rH max.span100rHTemperature:min.span25ºC max.span170ºC
Jan. 06, 2014-00
4
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
3. Electrical OutputSignalGeneral: Oneoutputof4-20mADC Note:Tolerance:±0.02mA
Bi-directionalHARTdigitalcommunication,superimposedonmA(4-20mA)signal
Outputfunction: LinearorNon-linear(21-steptable)Burnoutfunction:(NAMUR43exceptISC)
WithoutHART/PH201G: Down:3.6mA (signal:3.8to20.5mAforpH/ORP,SC
andDO) (signal:3.9to20.5mAforISC) Up:22mAWithHART/PH201G: Down:3.6mAforpH/ORP,SCandDO Down:3.9mAforISC (signal:3.8to20.5mAforpH/ORP,SC
andDO) (signal:3.9to20.5mAforISC) Up:22mA
PowerSupplyNominal24VDClooppoweredsystem
One(1)Sensormodule(1input): 16to40VDC(forpH/ORP(analog
sensor),SCandDO) 17to40VDC(forISC) 21to40VDC(forpH/ORPSENCOM
sensor)Two(2)Sensormodules(2inputs): 22.8to40VDC(forpH/ORP(analogsensor),
SCandDO)Note: WhentheFLXA21isusedinthemulti-dropmodeof
HARTcommunication,theoutputsignalischangedfrom12.5mADCto4mADCjustafterthepoweristurnedon.Enoughpowersupplyfortheinstrumentsistobeprovided.
MaximumLoadResistancepH/ORP(analogsensor),SCandDO: RefertotheFigure1.ISCandpH/ORPSENCOMsensor: RefertotheFigure2.
Message language: 12(English, Chinese, Czech, French,
German,Italian,Japanese,Korean,Polish,Portuguese,RussianandSpanish)
Oneanalyzerhasall12languages.Note: Descriptionforaselectionoflanguageand
languagenamesarewritteninEnglish.Note: OnlyEnglishalphabetandnumericareavailable
foratagnumber,anadditionaldescriptionforeachvalueonthedisplayscreenandpasswords.
Note: Onlyformessagelanguageonthescreen,12languagesareprovided.
4. Mechanicalandothers HousingCase: •Plastic(Polycarbonate) •Stainlesssteelwithoutpainting •Stainlesssteelwithepoxycoating •StainlesssteelwithurethanecoatingCasecolorandfinish:
Color: Silvergray(equivalenttoMunsell3.2PB7.4/1.2)
(forplasticcase,stainlesssteelcaseswithcoating)
Finish: Electropolishing(forstainlesssteelcasewithoutpainting)
Window: Polycarbonate(flexible)Windowframeforstainlesssteelcases: Polycarbonate,color:silvergray
(equivalenttoMunsell3.2PB7.4/1.2)Protection:IP66(exceptCanada),Type4X(except
Canada),Type3S/4X(Canada) PlateMainnameplate:insidecasecoverRegulationplate: on the case outside
CableandTerminalCablesize:
Outerdiameter:6to12mm(suitableforM20cablegland)3.4to7mm(groundingcableforplasticcase)
Terminalscrewsize:M4 torqueofscrewup:1.2N•mWireterminal: Pinterminal,ringterminalandspade
terminalcanbeusedforanalyzer’spowersupplyterminalsandsensorterminals.
Forthegroundingterminalonthestainlesssteelcase,ringterminalshouldbeused.
Pinterminal:pindiameter:max.1.9mm Ringandspadeterminal:width:max.7.8mm
Jan. 06, 2014-00
250
600
1000
0 24.718
V - 11.50.022
R =
1295
1617 22.8 4040Voltage (V)
2-sensor measurement
Load
resi
stan
ce (Ω
)
Digital Communication Range (HART)
Figure1 SupplyVoltageandLoadResistance forpH/ORP(analogsensor),SCandDO
250304
600516
1000
0 24.718
V - 11.50.022
R =
1295
17 4022.86 40
Except SENCOM
Voltage (V)
Load
resi
stan
ce (Ω
)
Digital Communication Range (HART)
18.2 21
Figure2 SupplyVoltageandLoadResistance forISCandpH/ORPSENCOMsensor
DisplayLCDwithatouchscreen: Black/White:213x160pixelsContrastadjustmentavailableonthetouchscreen
5
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
CableEntryPlasticcase:
1-Sensormeasurement: 3 holes, M20cableglandx3pcs, Sleevex1pc(forgroundingcableline)2-Sensormeasurement: 4 holes, M20cableglandx4pcs, Sleevex1pc(forgroundingcableline)
Stainlesssteelcase: 3 holes, M20cableglandx3pcs Close up plug x 1 pcNote: Cableglandandplugaredeliveredwithananalyzer,
butnotassembledintotheanalyzer. MountingMountinghardware(option): •Universalmountingkit(Note) •Pipeandwallmountinghardware •PanelmountinghardwareNote: Thiskitcontainsthepipeandwallmounting
hardwareandthepanelmountinghardware.Hood(option): •Stainlesssteel •Stainlesssteelwithurethanecoating •Stainlesssteelwithepoxycoating
StainlessSteelTagPlateWhentheadditionalcode“/SCT”withatagnumberisspecified,thetagplateonwhichthetagnumberisinscribedisdeliveredwiththeanalyzer.Tag plate is hanging type.
ConduitAdapterUsingoptionaladapter •G1/2(quantity:4) •1/2NPT (quantity:4) •M20x1.5 (quantity:4)Theseconduitadaptersaredeliveredwithananalyzer,butnotassembledintotheanalyzer.
SizeofHousingCasePlastic: 144x144x151mm(LxWxD)(without
cablegland)Stainlesssteelcase: 165x165x160mm(LxWxD)(without
cablegland) WeightApprox.1kg(Plastichousing)Approx.2kg(Stainlesssteelhousing)
ShippingDetailsPackagesize: Approx.340x340x370mm(LxWxH)
AmbientOperatingTemperature-20to+55ºC
StorageTemperature-30to+70ºC
Humidity10to95%RH(Non-condensing)
DocumentFollowingdocumentsaredeliveredwithananalyzer;
Papercopy:Start-upManual written in EnglishSafetyPrecautions written in English
CD-ROM:Start-upManual written in English
User'sManual written in EnglishSafetyRegulationsManual forEuropeanregion written in 25 languagesGeneralSpecifications written in EnglishTechnicalInformation forHARTCommunication written in EnglishUserSettingTable of5kindsofmeasurement/sensortype written in English
RegulatoryComplianceSafety: EN61010-1 UL61010-1 CAN/CSAC22.2No.61010-1EMC: EN61326-1ClassA,Table2(Forusein
industriallocations) EN61326-2-3 AS/NZSCISPR11 KoreaElectromagneticConformity
StandardInstallationaltitude:2000morlessCategorybasedonIEC61010:I(Note1)PollutiondegreebasedonIEC61010:2(Note2)Note1:Installationcategory,calledover-voltagecategory,
specifiesimpulsewithstandvoltage. Equipmentwith"CategoryI"(ex.twowire
transmitter)isusedforconnectiontocircuitsinwhichmeasuresaretakentolimittransientover-voltagestoanappropriatelylowlevel.
Note2:Pollutiondegreeindicatesthedegreeofexistenceofsolid,liquid,gasorotherinclusionswhichmayreducedielectricstrength.Degree2isthenormalindoorenvironment.
Explosion-proof(Intrinsicallysafetypeandnon-incendive)(forsuffixcode:-EA):
ATEXIntrinsicallysafeapprovalApplicablestandard
ExplosiveAtmospheresEN60079-0:2009GeneralrequirementsEN60079-11:2007 Equipmentprotection
byintrinsicsafety“i”EN60079-26:2007 Equipmentwith
equipmentprotectionlevel(EPL)Ga
EN60529:1992 Degreesofprotectionprovidedbyenclosures(IPCode)
TypeofprotectionII1GExiaIICGa Group:II Category: 1G T4:forambienttemperature:–20to55ºC T6:forambienttemperature:–20to40ºC Atmospherepressure:80kPa
(0.8bar)to110kPa(1.1bar) DegreeofProtectionofthe
Enclosure:IP66IECExIntrinsicallysafe
ApplicablestandardIEC60079-0:2007 Part0:General
requirementsIEC60079-11:2006 Part11:Equipment
protectionbyintrinsicsafety“i”IEC60079-26:2006 Part26:Construction,
testandmarkingofGroupIIZone0 electrical apparatus
Jan. 06, 2014-00
6
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
IEC60529:2001 Degreesofprotectionprovidedbyenclosures(IPCode)
TypeofprotectionExiaIICGa T4:forambienttemperature:–20to55ºC T6:forambienttemperature:–20to40ºC Atmospherepressure:80kPa
(0.8bar)to110kPa(1.1bar) DegreeofProtectionofthe
Enclosure:IP66FMIntrinsicallysafeandnonincendiveapproval
ApplicablestandardFM-3600:2011 ApprovalStandardfor
ElectricEquipmentforuseinHazardous(Classified)LocationsGeneralRequirement
FM-3610:2010 ApprovalStandardforIntrinsicallySafeApparatusandAssociatedApparatusforUseinClassI,II,andIII,Division1,Hazardous(Classified)Locations
FM-3611:2004 NonincendiveElectricalEquipmentforUseinClassIandII,Division2andClassIII,Divisions1and2,Hazardous(Classified)Locations
FM-3810:2005 ElectricalEquipmentforMeasurement,ControlandLaboratoryUse
NEMA250:1991 EnclosuresforElectricalEquipment(1000VoltsMaximum)
ANSI/IEC60529:2004 Degreesofprotectionprovidedbyenclosures(IPCode)
ANSI/ISA60079-02009 Part0:GeneralRequirements
ANSI/ISA60079-112011Part11:Equipmentprotectionbyintrinsicsafety“i”
TypeofprotectionClassI,Division1,GroupsA,B,CandD
(IntrinsicallySafe)ClassI,Division2,GroupsA,B,CandD
(Nonincendive)ClassI,Zone0,inHazardous(Classified)
Locations(IntrinsicallySafe)ClassI,Zone2,GroupIIC,inHazardous
(Classified)Locations(Nonincendive)
AExiaIICFor all protection type, T4:forambienttemperature:-20to55°C T6:forambienttemperature:-20to40°CAtmospherepressure:80kPa(0.8bar)to
110kPa(1.1bar)DegreeofProtectionoftheEnclosure:
NEMAType4XandIP66CSAIntrinsicallysafeandnonincendiveapproval
ApplicablestandardCAN/CSAC22.2No.94-M1991 Special
PurposeEnclosuresCAN/CSAC22.2No.157-92
IntrinsicallySafeEquipmentforUseinHazardousLocations
C22.2No213-M1987 Non-incendiveElectricalEquipmentforUseinClassI,Division2HazardousLocations
CAN/CSA-E60079-0-07 Electricalapparatusforexplosivegasatmospheres-Part0:Generalrequirements
CAN/CSA-E60079-11-02Electricalapparatusforexplosivegasatmospheres-Part11:Intrinsicsafety“i”
IEC60529:2001 Degreesofprotectionprovidedbyenclosures(IPCode)
Typeofprotection(C22.2)ClassI,Division1,GroupsA,B,CandD
(IntrinsicallySafe)ClassI,Division2,GroupsA,B,CandD
(Nonincendive)For all protection type, T4:forambienttemperature:-20to55°C T6:forambienttemperature:-20to40°CAtmospherepressure:80kPa(0.8bar)to
110kPa(1.1bar)AmbientHumidity:0to100%(Nocondensation)DegreeofProtectionoftheEnclosure:Type4X
Typeofprotection(E60079)ExiaIIC T4:forambienttemperature:-20to55°C T6:forambienttemperature:-20to40°CAtmospherepressure:80kPa(0.8bar)to
110kPa(1.1bar)AmbientHumidity:0to100%(Nocondensation)DegreeofProtectionoftheEnclosure:IP66
NEPSIIntrinsicallysafeapprovalApplicableStandard
GB3836.1-2010 Explosiveatmospheres-Part1:Equipment-Generalrequirements
GB3836.4-2010 Explosiveatmospheres-Part4:Equipmentprotectionbyintrinsicsafety“i”
GB3836.20-2010Explosiveatmospheres-Part20:Equipmentwithequipmentprotectionlevel(EPL)Ga
TypeofprotectionExiaIICGaT4:forambienttemperature:-20°Cto55°CT6:forambienttemperature:-20°Cto40°CAtmospherepressure:80kPa(0.8bar)to
110kPa(1.1bar)DegreeofProtectionoftheEnclosure:IP66
ElectricalParameters(Exia)Eachhousingassembly(basemodule)andeachsensormodulearerespectivelycertificated.Inputparametersofsensormodulemeetoutputparametersofhousingassembly.
Housingassembly
Inputparameters
Supplyandoutputcircuit(terminals+and-): Ui,Vmax =30V Ii,Imax =100mA Pi,Pmax =0.75W Ci =13nF Li =0mH (Linearsource)
Outputparameters
Measuringmoduleinputcircuit(CN2orCN3onBackboard) UoVt,Voc =13.65V Io,It,Isc =50mA Po =0.372W Co,Ca =80nF Lo, La =7.7mH
Jan. 06, 2014-00
7
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
pH/ORPmodule,SCmodule,andDOmodule
Inputparameters
Ui,Vmax =13.92V Ii,Imax =50mA Pi,Pmax =0.374W Ci =40nF Li =2.9mH
Outputparameters
Sensorinputcircuit(pH:terminals11through19,SC:terminals11through16,DO:terminals11through18) UoVt,Voc =11.76V Io,It,Isc =116.5mA Po =0.3424W Co,Ca =100nF Lo,La =1.7mH
ISCmodule
Inputparameters
Ui,Vmax =13.92V Ii,Imax =50mA Pi,Pmax =0.374W Ci =40nF Li =7.7mH
Outputparameters
Sensorinputcircuit(terminals11through17) UoVt,Voc =11.76V Io,It,Isc =60.6mA Po =0.178W Co,Ca =100nF Lo,La =8mH
ControlDrawing(ATEXandIECExtypes)
Measuring module 1
Note: The measuring module on this drawing means the sensor module on this General Specifications.
Measuring module 2
Sensor 1
Sensor 2
(*)
(Note 1)Safety BarrierHousing Assembly
Supply -
Supply +
→ Non Hazardous LocationHazardous Location ←
-
+
Electricaldataareasfollows; MaximumVoltage(Ui)=30V MaximumCurrent(Ii)=100mA MaximumPower(Pi)=0.75W InternalCapacitance(Ci)=13nF InternalInductance(Li)=0mHNote1:Theoutputcurrentmustbelimitedbyaresistor“R”
suchthatImaxout=Uz/R(linearsource).Note2:SafetybarriercertifiedbyanotifybodyinEUas
ATEXshouldbeused.Note3:Whenusingnonisolationbarrier,connect(*)toIS
earthingsystem.Note4:Measuringmodule2isinstalledwhenrequired.
Whenmeasuringinductiveconductivity,onlyonemodulecanbeinstalled.
ControlDrawing(FMtype)Followingcontentsrefer“DOC.No.IKE039-A12”
Measuring module 1
Measuring module 2
(Refer to Note 7)
Sensor 1
Sensor 2(Note 1)
Safety BarrierHousing AssemblySupply -
Supply +
→ Non Hazardous LocationHazardous Location ←
-
+
(*)
Class I, Division 1, Groups A, B, C, and DClass I, Zone 0 and 1, Group IICT4 for Ta = 55°C, T6 for Ta = 40°C
2 wire analyzer
Note: The measuring module on this drawing means the sensor module on this General Specifications.
Electricaldataareasfollows;Input MaximumInputVoltage(Ui)=30V MaximumCurrent(Ii)=100mA MaximumPower(Pi)=0.75W InternalCapacitance(Ci)=13nF InternalInductance(Li)=0mHSensorInputCircuit
TypeofMeasuringModule pH,SCandDO ISCMaximumVoltage(Uo) 11.76V 11.76VMaximumCurrent(Io) 116.5mA 60.6mAMaximumPower(Po) 0.3424W 0.178WExternalCapacitance(Ca,Co) 100nF 100nFExternalInductance(La,Lo) 1.7mH 8mH
Note1:Inanysafetybarrierused,theoutputcurrentmustbelimitedbyaresistor“R”suchthatImaxout=Uz/R.
Note2:ThesafetybarriershallbeFMEntity-Approvedassociatedapparatus/barrierwhere:
BarrierVoc,Uo≤30V; BarrierIsc,Io≤100mA; BarrierPo≤0.75W; BarrierCa,Co≥13nF+Ccable; BarrierLa,Lo≥LcableNote3:Whenusingnonisolationbarrierconnect(*)toIS
earthingsystem.Note4:pHandSCSensor(s)areofapassivetypeto
beregardedas‘simpleapparatus’sameas06ATEX0218X,06ATEX0219,IECExKEM06.0052X, FM3028779, 06ATEX0220X, 06ATEX0221,IECExKEM06.0053Xortheoneindividuallycertifiedwithrelevantparameters.
Note5:ISCSensor(s)areISC40Sof00ATEX1067Xortheoneindividuallycertifiedwithrelevantparameters.
Note6:DOSensor(s)areofapassivetypetoberegardedas‘simpleapparatus’ortheoneindividuallycertifiedwithrelevantparameters.
Note7:Measuringmodule2maynotmounted.AsforISCmodule,onlyonecanbemounted.
Note8:InstallpertheNationalElectricalCode(NFPA70)Note9:WARNING-Potentialelectrostaticcharging
hazard Electrostaticchargemaycauseanexplosion
hazard.Avoidanyactionsthatcausethegenerationofelectrostaticcharge,e.g.,rubbingwith a dry cloth.
Note 10:AsanalternativetoinstallingtheFLXA21inDivision2usingClassI,Division2wiringmethods,theFLXA21maybeinstalledinDivision2usingnonincendivefieldwiringinaccordancewiththeNationalElectricalCode(NFPA70)usingthesameparametersidentifiedforintrinsicallysafeentityinstallations.TheAssociatedNonincendiveApparatusshallhavenonincendivefieldwiringconnectionswhichareFMApprovedforuseintheClassI,Division2location.
ControlDrawing(CSAtype)
Measuring module 1
Measuring module 2
(Refer to Note 7)
Sensor 1
Sensor 2
Housing AssemblySupply -
Supply +
→ Non Hazardous LocationHazardous Location ←
(*)
Refer to NoteSafety Barrier
-
+
Intrinsically SafeGroup IIC, Zone 0Class I, Division 1
Non-incendiveClass I, Division 2,Groups A, B, C, D
2 wire analyzer
Not use safety barrier but CSA certified equipment use in Non-incendive
Note: The measuring module on this drawing means the sensor module on this General Specifications.
Jan. 06, 2014-00
8
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
Electricalparameters(Intrinsicallysafe)HousingAssembly
Supplyandoutputcircuit(terminals+and-) Ui(Vmax)=30V,Ii(Imax)=100mA,
Pi(Pmax)=0.75W,Ci=13nF,Li=0mHMeasuringmoduleinputcircuit(CN2orCN3on
Backboard) Uo(Vt,Voc)=13.65V,Io(It,Isc)=50mA,
Po=0.372W,Co(Ca)=80nF,Lo(La)=7.7mH
pHmodule,SCmoduleandDOmodule Ui(Vmax)=13.92V,Ii(Imax)=50mA,
Pi(Pmax)=0.374W,Ci=40nF,Li=2.9mHSensorinputcircuit(terminals11through19) Uo(Vt,Voc)=11.76V,Io(It,Isc)=116.5mA,
Po=0.3424W,Co(Ca)=100nF,Lo(La)=1.7mH
ISCmodule Ui(Vmax)=13.92V,Ii(Imax)=50mA,
Pi(Pmax)=0.374W,Ci=40nF,Li=7.7mHSensorinputcircuit(terminals11through17) Uo(Vt,Voc)=11.76V,Io(It,Isc)=60.6mA,
Po=0.178W,Co(Ca)=100nF,Lo(La)=8mH
Installationrequirementsbetweenhousingassemblyandsafetybarrier
Uo≤UiIo≤IiPo≤PiCo≥Ci+CcableLo≥Li+LcableVoc≤VmaxIsc≤ImaxCa≥Ci+CcableLa≥Li+LcableUo,Io,Po,Co,Lo,Voc,Isc,CaandLaareparametersofbarrier.
ElectricalParameters(Nonincendive)HousingAssembly
Supplyandoutputcircuit(terminals+and-) Ui(Vmax)=30V,Ci=13nF,Li=0mHMeasuringmoduleinputcircuit(CN2orCN3on
Backboard) Uo(Vt,Voc)=13.65V,Io(It,Isc)=50mA,
Co(Ca)=80nF,Lo(La)=7.7mHpHmodule,SCmoduleandDOmodule Ui(Vmax)=13.92V,Ci=40nF,Li=2.9mHSensorinputcircuit(terminals11through19) Uo(Vt,Voc)=11.76V,Io(It,Isc)=116.5mA,
Co(Ca)=4uF,Lo(La)=4.5mHISCmodule Ui(Vmax)=13.92V,Ci=40nF,Li=7.7mHSensorinputcircuit(terminals11through17) Uo(Vt,Voc)=11.76V,Io(It,Isc)=60.6mA,
Co(Ca)=4uF,Lo(La)=19mHNoteforIntrinsicallySafeInstallation:
1: Inanysafetybarrierused,theoutputcurrentmustbelimitedbyaresistor“R”suchthatIo=Uo/RorIsc=Voc/R.
2: ThesafetybarriermustbeCSAcertified.3: Inputvoltageofthesafetybarriermustbelessthan
250Vrms/Vdc4: Whenusingnonisolationbarrierconnect(*)toIS
earthingsystem.5: pHandSCSensor(s)areofapassivetypeto
beregardedas‘simpleapparatus’sameas06ATEX0218X,06ATEX0219,IECExKEM06.0052X, FM3028779, 06ATEX0220X, 06ATEX0221,IECExKEM06.0053Xortheoneindividuallycertifiedwithrelevantparameters.
6: ISCSensor(s)areISC40Sof00ATEX1067Xortheoneindividuallycertifiedwithrelevantparameters.
7: DOSensor(s)areofapassivetypetoberegardedas‘simpleapparatus’ortheoneindividuallycertifiedwithrelevantparameters.
8: Measuringmodule2maynotmounted.AsforISCmodule,onlyonecanbemounted.
9: InstallationshouldbeinaccordancewithCanadianElectricalCodePartIandLocalElectricalCode.
10:DonotalterdrawingwithoutauthorizationfromCSA.NoteforNonincendiveInstallation:
1: Theparameterforsensorinputcircuitmustbetakeninto account when installed.
2: InstallationshouldbeinaccordancewithCanadianElectricalCodePartIandLocalElectricalCode.
3: DonotalterdrawingwithoutauthorizationfromCSA.ControlDrawing(NEPSItypes)
Measuring module 1
Note: The measuring module on this drawing means the sensor module on this General Specifications.
Measuring module 2
Sensor 1
Sensor 2
(*)
(Note 1)Safety BarrierHousing Assembly
Supply -
Supply +
→ Non Hazardous LocationHazardous Location ←
-
+
Electricaldataareasfollows; MaximumVoltage(Ui)=30V MaximumCurrent(Ii)=100mA MaximumPower(Pi)=0.75W InternalCapacitance(Ci)=13nF InternalInductance(Li)=0mHNote1:Theoutputcurrentmustbelimitedbyaresistor“R”
suchthatImaxout=Uz/R(linearsource).Note2:SafetybarriercertifiedbyNEPSIshouldbeused.Note3:Whenusingnonisolationbarrier,connect(*)toIS
earthingsystem.Note4:Measuringmodule2isinstalledwhenrequired.
Whenmeasuringinductiveconductivity,onlyonemodulecanbeinstalled.
5. DigitalCommunication KindofDigitalCommunication•HARTorPH201GdedicateddistributorNote: Onlyonekindofdigitalcommunicationisavailable
foroneanalyzer. OutputValueParameter(HART)Fourvalueparameters(measuredvalues)areavailableforonedigitalcommunication.•For1-sensormeasurement,theseparametersare
measuredvalues.•For2-sensormeasurement,refertothenextitem.
DigitalCommunicationof2-SensorMeasurement(HART)Evenwhentwosensormodulesareinstalled,onlyonedigitalcommunicationisavailablefor2-sensormeasurement.Fourvalueparameterscanbeselectedfromthefollowings; Measuredvaluesoftwosensors Calculateddataof2-sensormeasurement Redundantsystemoutput
SpecificContactOutputwithdedicateddistributor,modelPH201G(StyleB)Thedistributor,modelPH201G,isdesignedtoconnect with the 2-Wire Analyzer.Thisdistributorsuppliesdrivepowertotheanalyzerandreceivessimultaneously4-20mADCsignalfromthe analyzer.Thissignalisconvertedto1-5VDCsignalinthedistributor.Thisdistributoralsoreceivesdigitalsignalssuperimposedonthe4-20mADCsignal,andprovidescontactoutputs
Jan. 06, 2014-00
9
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
6. Model&SuffixCodes
Model Suffixcode Optioncode Description
FLXA21 ······················································································· ···················· 2-Wire Analyzer
Powersupply -D ···················· Always-D
Housing -P-S-U-E
················································································
PlasticStainlesssteelStainlesssteel+urethanecoatingStainlesssteel+epoxycoating
Display -D ···················· Anti-glareLCD
Type -AA-EA
········································
General purposeATEX,IECEx,FM,CSA,NEPSI(Note5)
1st input -P1-C1-C5-D1-S1
····································································································
pH/ORP(Note7)Conductivity(SC)Inductiveconductivity(ISC)Dissolvedoxygen(DO)pH/ORP(SENCOMsensor)(Note8)
2ndinput(Note1) -NN-P1-C1-D1
················································································
Without inputpH/ORP(Note7)Conductivity(SC)Dissolvedoxygen(DO)
Output -A ···················· 4-20mA+HART
— -N ···················· Always -N
Languageset(Note2) -LA ···················· English and 11 languages
Country(Note3) -N-J
········································
GlobalexceptJapanJapan
— -NN ···················· Always -NN
Option Mounting hardware
Hood
Tag plateConduit adapter
Measurementlaw
/UM/U/PM/H6/H7/H8/SCT/CB4/CD4/CF4/K
Universalmountingkit(Note4)PipeandwallmountinghardwarePanelmountinghardwareHood,stainlesssteelHood,stainlesssteel+urethanecoatingHood,stainlesssteel+epoxycoatingStainlesssteeltagplateConduitadapter(G1/2x4pcs)Conduitadapter(1/2NPTx4pcs)Conduitadapter(M20x1.5x4pcs)WithMeasurementLawcertificate(Note6)
Notes:1: Whena2ndinputisselected,onlythesamekindofthe1stinputisavailable. Forexample,whena1stinputis“-P1”,the2ndinputmustbethesame“-P1”. ThecombinationofISCandISCisnotavailable.And,thecombinationofSENCOMsensorandSENCOMsensorisnot
available,either.2: Theselanguagesaremessagelanguagesontheanalyzer’sdisplay. OneanalyzerhasEnglishand11languages. Alllanguagesareasfollows;English,Chinese,Czech,French,German,Italian,Japanese,Korean,Polish,Portuguese,
RussianandSpanish.3: WhenananalyzerisusedinJapan,itmustmeettheJapaneseMeasurementLaw. OnlySIunitsmustbeusedontheanalyzeranditsdocumentsinJapan.4: Theuniversalmountingkitcontainsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).5: Thetype“-EA”isintrinsicallysafetypeofATEX,IECEx,FM,CSAandNEPSI,andnon-incendiveofFMandCSA.
Jan. 06, 2014-00
Input/Outputsignal:Numberofavailabledrive/signalpoint: 1Outputsignal:1-5VDC(2points)(Note)Loadresistance: 2kΩorless(1-5VDCoutput)Isolationsystem: Loopisolationtype
Note: Twooutputsignalsforoneanalyzer'sanalogoutputareprovided.Two1-5VDCoutputsignalsaresame.
Contact output:Contact rating: 250VAC,maximum100VA 220VDC,maximum50VA
Holdcontactoutput: NCcontact,normallyenergized Contactcloseswhenpowerisoffor
duringHoldsituation.Fail contact output: NCcontact,normallyenergized Contactcloseswhenpowerisoffor
duringFail/Warningconditions.Wash contact output: NOcontact Contact closes during wash cycles.
10
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
6: TheanalyzerwithJapaneseMeasurementLawcertificateisavailableonlyforthefollowingmodel; FLXA21-D-[Housingcode]-D-AA-P1-NN-A-N-LA-J-NN/[optioncodeexcept/K]/K OnlyonepHmeasurementwithananalogsensoriscertified.Theoutputsignalof4-20mAiscertified.HARTcommunication
isnotcertified.7: ThisinputistobecomefromananalogpH/ORPsensor.8: Whentheanalyzerisconnectedwiththedigitalsensor,FU20FpH/ORPSENCOMSensor,onlythefollowingmodelis
available; Type: Generalpurpose(-AA) 2ndinput:Withoutinput(-NN) Option: except“/K”
nDimensionsandMountingPlasticHousing
44.9137
120.2
151
22.5 80
4-M6 depth 5(For sensor 2 cable)
For sensor 1 cable
For power supply
For grounding cable
137
80144
24.7
26.5
26.5
141144
FB4_01.ai
Unit: mm
ConduitAdapter(Optioncode:/CB4,/CD4,/CF4)
Adapter
49(1.93")
G1/2 screw (/CB4), 1/2 NPT screw (/CD4)M20x1.5 screw (/CF4)
Approx.55(2.2")
Packing
Unit: mm(inch)Nut
F0204.ai
Jan. 06, 2014-00
11
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
(Note)Theuniversalmountingkit(/UM)containsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).
Panelmountinghardware(Optioncode:/PM,/UM)
2-M5 length 35
185
178
100
195
135
Pan
el th
ickn
ess
1 to
12
Spacing panel cutout
4-M6 *
138+10
138
+1 0
FB4_03.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Wallmountinghardware(Optioncode:/U,/UM)
144 164
13
10070
144
224
200
50
15
For wall mounting3-ø10 holes
4-M6 *
FB4_02.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Pipemountinghardware(Optioncode:/U,/UM)
151
100
200
5017
4
144(205)
Pipe mounting (Horizontal) Pipe mounting (Vertical)
Pipe 50A(ø60.5)
M8 U-bolt
4-M6 *
FB4_04.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Jan. 06, 2014-00
12
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
Stainlesssteelhood(Optioncode:/H6,/H7,/H8)184 220
(70)(100)
9144
144
7257
(50)
199
FB4_05.ai
Unit: mm
StainlessSteelHousing
80165
13722.244
150160
32.5
4-M6 depth 5(For sensor 2 cable)For sensor 1 cable
For power supply
Plate(Only with coating)
80137
165
25.3
26.5
FB4_06.ai
Unit: mm
ConduitAdapter(Optioncode:/CB4,/CD4,/CF4)
Adapter
49(1.93")
G1/2 screw (/CB4), 1/2 NPT screw (/CD4)M20x1.5 screw (/CF4)
Approx.55(2.2")
Packing
Unit: mm(inch)Nut
F0204.ai
Jan. 06, 2014-00
13
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
(Note)Theuniversalmountingkit(/UM)containsthepipeandwallmountinghardware(/U)andthepanelmountinghardware(/PM).
Panelmountinghardware(Optioncode:/PM,/UM)
2-M5 length 35
185
178
100
195
134
Pan
el th
ickn
ess
1 to
12
Spacing panel cutout
4-M6 *Plate(Only with coating)
138+10
138
+1 0FB4_08.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Wallmountinghardware(Optioncode:/U,/UM)
165 17313
70100
4-M6 *
165
200
5015
234.
5
For wall mounting3-ø10 holes
FB4_07.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Jan. 06, 2014-00
14
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
Pipemountinghardware(Optioncode:/U,/UM)
160
100
200
5018
4.5
165(214)
Pipe mounting (Horizontal) Pipe mounting (Vertical)
Pipe 50A(ø60.5)
M8 U-bolt
4-M6 *
FB4_09.ai
Unit: mm
*: Tighten the four screws to a torque of 2 N•m.
Stainlesssteelhood(Optioncode:/H6,/H7,/H8)184
220
(70)(100)
9165
165
7257
(50)
199
FB4_10.ai
Unit: mm
Jan. 06, 2014-00
15
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
nWiringDiagrams
(Plastic housing)(Stainless steel housing)
*2 *2
*4
*5
*6
*6
n n n n n n n n n
n n n n n n n n n
+–
*1: Use a 2-wire shielded cable with an outside diameter of 6 to 12 mm.*2: Connect the analyzer to gland. (Class D ground: 100 ohm or less) The way of connecting the grounding cable varies depending on the plastic housing and stainless steel
housing. In the case of the plastic housing, connect the grounding cable to the terminal of the power module inside,
and in the case of the stainless steel housing, connect the grounding cable to the terminal of the housing. Use a cable with an outside diameter of 3.4 to 7 mm for the grounding line of the plastic housing. The minimum cross sectional area of the protective grounding cable should be 0.75 mm2.*3: This line is connected to a distributor or 24V DC power supply.*4: Terminal numbers for each sensor module are shown below.*5: Two modules of the same kind of measurement/sensor type can be installed. When measuring inductive
conductivity or pH/ORP with the SENCOM sensor, only one module can be installed.*6: The terminal box may be necessary depending on the sensor cable length and the distance between the
analyzer and the sensor. The SENCOM sensor is to be connected directly to the analyzer without a terminal box.*7: Two outputs, output 1 and output2, of PH201G or SDBT are same signals.
*1
WTB10 or BA10Terminal box FLXA21 2-Wire Analyzer
Sen
sor
WTB10 or BA10Terminal box
Sen
sor
PH Module
SC Module
ISC Module
DO Module
SENCOM Module
161519171318141211
PH
NC1817141315161211
DO
15161417131211
ISC
NC161514131211
SC
8786NC848382NC
SENCOM
Power supply
Output 1 (1-5V DC)
20 to 130V DCor80 to 138V AC, 47 to 63Hz
Power supply20 to 130V DCor80 to 138V AC, 47 to 63Hz
HOLD WASHFAIL
Output 2 (1-5V DC)
Output 1 (1-5V DC)
Output 2 (1-5V DC)
Case of DistributorPH201G (Style B)
Case of DistributorSDBT
(*3)
*3
*7
*7
(*3)
A+
– CMN CMNB
CD
FH
L N
A+
–+
–B
1+
–2
FH
L N
ba
dc
fe
Jan. 06, 2014-00
16
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
nInquirySpecificationsSheetforFLXA212-WireAnalyzer
Makeinquiriesbyplacingcheckmarks()inthepertinentboxesandfillingintheblanks.
1. GeneralInformationCompanyname ContactPerson; Department; Plantname; Measurementlocation; Purposeofuse; Indication,Recording,Alarm, Control
2. MeasurementConditions(1)Processtemperature; to Normally [°C](2)Processpressure; to Normally [kPa](3)Flowrate; to Normally [l/min](4)Flowspeed; to Normally [m/s](5)Slurryorcontaminants; No, Yes(6)Nameofprocessfluid; (7)Componentsofprocessfluid; (8)Others;
3. InstallationSite(1)Ambienttemperature; to [°C](2)Location;Outdoors,Indoors(3)Others;
4. Requirements1stInput; pH/ORP(analogsensor)Conductivity(SC)Inductiveconductivity(ISC) Dissolvedoxygen(DO)pH/ORP(digitalsensor,FU20F)2ndInput; With(sameas1stInput) Without
4.1 pH/ORP(analogsensor)
1stInput(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection; Electrode, Holder,pHConverter,Cleaningsystem,Terminalbox, Accessories(4)Electrodecablelength; 3m,5m,7m,10m,15m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guide pipe, Submersion, Flow-through, Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod; No cleaning, Ultrasoniccleaning, Jet cleaning, Brush cleaning (8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;
2ndInput(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection; Electrode, Holder,pHConverter,Cleaningsystem,Terminalbox, Accessories(4)Electrodecablelength; 3m,5m,7m,10m,15m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guide pipe, Submersion, Flow-through, Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod; No cleaning, Ultrasoniccleaning, Jet cleaning, Brush cleaning (8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;
Jan. 06, 2014-00
17
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
4.2 Conductivity
1stInput(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Detector/sensor; SC4AJ Twoelectrodesystem(0.02cm-1)Twoelectrodesystem(0.1cm-1) SC8SG Twoelectrodesystem(0.01cm-1)Twoelectrodesystem(10cm-1), Fourelectrodesystem(10cm-1) SC210G Twoelectrodesystem(0.05cm-1)Twoelectrodesystem(5cm-1)(4)Detector/sensormountingmethod; SC4AJ Adaptermounting, Welding socket, Weldingclamp SC8SG Screw-in, Flow-through SC210G Screw-in, Flange, Flow-through, Screw-inwithgatevalve(5)Electrodecablelength; SC4AJ3m,5m,10m,20m SC8SG 5.5m,10m,20m SC210G 3m,5m,10m,15m,20m(6)Others;
2ndInput(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Detector/sensor; SC4AJ Twoelectrodesystem(0.02cm-1)Twoelectrodesystem(0.1cm-1) SC8SG Twoelectrodesystem(0.01cm-1)Twoelectrodesystem(10cm-1), Fourelectrodesystem(10cm-1) SC210G Twoelectrodesystem(0.05cm-1)Twoelectrodesystem(5cm-1)(4)Detector/sensormountingmethod; SC4AJ Adaptermounting, Welding socket, Weldingclamp SC8SG Screw-in, Flow-through SC210G Screw-in, Flange, Flow-through, Screw-inwithgatevalve(5)Electrodecablelength; SC4AJ3m,5m,10m,20m SC8SG 5.5m,10m,20m SC210G 3m,5m,10m,15m,20m(6)Others;
4.3 Inductiveconductivity
(1)Measuringrange; (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection;ISC40GJSensor,Holder,Converter,BA20Terminalbox, WF10JExtensioncable(4)Sensormountingmethod; ISC40FDJImmersionholder,ISC40FFJFlow-throughholder, ISC40FSJDirectinsertionadapter(5)ISC40GJSensorcablelength; 5m,10m,15m,20m(6)WF10JExtensioncablelength; 5m,10m,20m,30m,40m(7)Others;
Jan. 06, 2014-00
18
AllRightsReserved.Copyright©2010,YokogawaElectricCorporation GS12A01A02-01E
4.4 Dissolvedoxygen
1stInput(1)Measuringrange; 0to50mg/L (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection; Electrode, Holder,Converter,Cleaningsystem, Terminalbox, Maintenance parts set, Calibrationset(4)Electrodecablelength; 3m,5m,10m,15m,20m(5)Typeofholder; Guide pipe, Submersion, Flow-through, Suspension, Angledfloatingball,Verticalfloatingball(6)Cleaningmethod; No cleaning, Jet cleaning(7)Others;
2ndInput(1)Measuringrange; 0to50mg/L (2)Transmissionoutput;4to20mADC(3)Systemconfigurationselection; Electrode, Holder,Converter,Cleaningsystem, Terminalbox, Maintenance parts set, Calibrationset(4)Electrodecablelength; 3m,5m,10m,15m,20m(5)Typeofholder; Guide pipe, Submersion, Flow-through, Suspension, Angledfloatingball,Verticalfloatingball(6)Cleaningmethod; No cleaning, Jet cleaning(7)Others;
4.5 pH/ORP(digitalsensor,FU20F)(1)Measuringrange;pH0to14ORP to mV (2)Transmissionoutput;4to20mADCpHORPTemperature(3)Systemconfigurationselection; Electrode, Holder,pHConverter,Cleaningsystem, Accessories(4)Electrodecablelength; 3m,5m,10m,20m, m(5)Electrodeoperatingpressure; 10kPaorless,Morethan10kPa(6)Typeofholder; Guide pipe, Submersion, Flow-through, Suspension,Angledfloatingball, Verticalfloatingball(7)Cleaningmethod; No cleaning, Jet cleaning(8)Sampletemperature;-5to105°C,-5to100°C,-5to80°C(9)Others;
Subjecttochangewithoutnotice.Jan. 06, 2014-00