MiniGRAIL“Extreme Make-over”
Giorgio FrossatiKamerlingh Onnes LaboratoryLeiden University
In collaboration with ROG group AURIGA-group
The spherical gravitational wave antenna MiniGRAIL is being prepared for the next cool-down, using the experience acquired with the last cool-down of 2005. The problems found were:
1- Lack of sensitivity of the transducers when cooled to 60mK. The thermally excited resonance were barely above the thermal noise. The best spectral strain sensitivity was ~8x10-21 /√Hz , still more than two orders of magnitude lower than the quantum limit of 4x10-23/√Hz
2- Instability of the SQUIDs due to RF noise and too high an intrinsic SQUID noise
3- Relatively low cooling power of the dilution refrigerator limiting the minimum sphere temperature to 60 mK.
6 new transducers
Improve sensitivity new transducers:
• Larger surface area
• Smaller mass
6 new transducers
-Super-conducting shielding-Magnetic (Glasmet) shielding
-Low frequency vibrations damping
Decrease the external EM perturbations:
6 new transducers
6 new transformers
super-conducting and magnetic
shielding
Improve transducer read out:
6 new transducers
6 new transformersNew double
SQUIDs
super-conducting and Mu-metal shielding
Improve SQUID sensitivity:
2-stage Twente
2-stage Trento
6 new transducers
6 new transformers new dilution
refrigerator
Super-conducting and mu-metal shielding
new dilution refrigerator unit:
• large surface mixing chamber and heat exchangers
• expected cooling power of at least 50 microW at 30 mK