MICROSCOPENAVIGATIONMICROSCOPE INTERFACE CABLES
Hardware User GuideRevision 1.0Date of issue: 2020-05-26 (ISO 8601)Copyright 2020, Brainlab AG Germany. All rights reserved.
TABLE OF CONTENTS1 GENERAL INFORMATION.............................................................................................5
1.1 Contact Data ........................................................................................................................................5
1.2 Legal Information ...............................................................................................................................6
1.3 Symbols................................................................................................................................................7
1.4 Using the System ...............................................................................................................................8
1.5 Compatibility with Medical Devices and Software ....................................................................10
1.6 Training and Documentation..........................................................................................................11
2 Microscope Interface Cables...................................................................................13
2.1 Microscope Navigation ...................................................................................................................13
2.2 Cables Overview ...............................................................................................................................14
2.3 Strain Relief .......................................................................................................................................16
2.4 Cleaning..............................................................................................................................................17
3 LEICA MICROSCOPES ..................................................................................................19
3.1 Connection to Curve 1.0 .................................................................................................................19
3.2 Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)..........................20
3.3 Connection to Curve Ceiling-Mounted ........................................................................................21
3.4 Connection to Curve Navigation ...................................................................................................22
3.5 Connection to Kick 1.x and Kick 2.x ............................................................................................23
4 HAAG-STREIT MICROSCOPES.............................................................................25
4.1 Connection to Curve 1.0 .................................................................................................................25
4.2 Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)..........................26
4.3 Connection to Curve Ceiling-Mounted ........................................................................................27
4.4 Connection to Kick 1.x and Kick 2.x ............................................................................................28
5 ZEISS MICROSCOPES ...................................................................................................29
TABLE OF CONTENTS
Hardware User Guide Rev. 1.0 Microscope Navigation 3
5.1 Connection to Curve 1.0 .................................................................................................................29
5.2 Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)..........................30
5.3 Connection to Curve Ceiling-Mounted ........................................................................................31
5.4 Connection to Curve Navigation ...................................................................................................32
5.5 Connection to Kick 1.x and Kick 2.x ............................................................................................33
TABLE OF CONTENTS
4 Hardware User Guide Rev. 1.0 Microscope Navigation
1 GENERAL INFORMATION1.1 Contact Data
Support
If you cannot find information you need in this guide, or if you have questions or problems, contactBrainlab support:
Region Telephone and Fax Email
United States, Canada, Centraland South America
Tel: +1 800 597 5911Fax: +1 708 409 1619
Brazil Tel: (0800) 892 1217 [email protected]
UK Tel: +44 1223 755 333
Spain Tel: +34 900 649 115
France and French-speakingregions Tel: +33 800 676 030
Africa, Asia, Australia, EuropeTel: +49 89 991568 1044Fax: +49 89 991568 5811
JapanTel: +81 3 3769 6900Fax: +81 3 3769 6901
Expected Product Lifetime
Brainlab provides five years of service for this product.
Feedback
Despite careful review, this user guide may contain errors. Please contact us [email protected] if you have improvement suggestions.
Manufacturer
Brainlab AGOlof-Palme-Str. 981829 MunichGermany
GENERAL INFORMATION
Hardware User Guide Rev. 1.0 Microscope Navigation 5
1.2 Legal Information
Copyright
This guide contains proprietary information protected by copyright. No part of this guide may bereproduced or translated without express written permission of Brainlab.
Brainlab Trademarks
• Brainlab® is a registered trademark of Brainlab AG.• Curve® is a registered trademark of Brainlab AG.• Kick® is a registered trademark of Brainlab AG.
Non-Brainlab Trademarks
• Leica, ARVEO® is a registered trademark of Leica Microsystems CMS GmbH.• HAAG-STREIT® is a registered trademark of Haag-Streit Holding AG.• KINEVO®, OPMI®, Pentero®, Pentero C®, PENTERO® and TIVATO® are registered
trademarks of Carl Zeiss Meditec AG.
Patent Information
This product may be covered by one or more patents or pending patent applications. For details,see: www.brainlab.com/patent.
CE Label
The CE label indicates that the Brainlab product complies with the essential re-quirements of the Council Directive 93/42/EEC ("MDD").Microscope Navigation is a Class IIb product according to the rules establishedby the MDD.
Disposal Instructions
When a medical device reaches the end of its functional life, clean the device of all biomaterial/biohazards and safely dispose of the device in accordance with applicable laws and regulations.
Only dispose of electrical and electronic equipment in accordance with statutory regu-lations. For information regarding the WEEE (Waste Electrical and Electronic Equip-ment) directive or relevant substances that could be present in the medical equipment,visit:www.brainlab.com/sustainability
Report Incidents Related to This Product
You are required to report any serious incident that may have occurred related to this product toBrainlab, and if within Europe, to your corresponding national competent authority for medicaldevices.
Sales in US
US federal law restricts this device to sale by or on the order of a physician.
Legal Information
6 Hardware User Guide Rev. 1.0 Microscope Navigation
1.3 Symbols
Warnings
WarningWarnings are indicated by triangular warning symbols. They contain safety-criticalinformation regarding possible injury, death or other serious consequences associatedwith device use or misuse.
Cautions
Cautions are indicated by circular caution symbols. They contain important informationregarding potential device malfunctions, device failure, damage to device or damage toproperty.
Notes
NOTE: Notes are formatted in italic type and indicate additional useful hints.
Product Symbols
Symbol Explanation
Keep dry
Manufacturer's batch code
Reference (article) numberNOTE: Indicates the Brainlab product number.
Manufacturer
Consult instructions for use
U.S. federal law restricts this device to sale by or on order of a physician
Unique Device Identifier
Medical Device
GENERAL INFORMATION
Hardware User Guide Rev. 1.0 Microscope Navigation 7
1.4 Using the System
Abbreviated Device Description
Microscope Navigation is software that runs on a Brainlab IGS platform or Brainlab navigationsystem, which consists of a computer, a display and an infra-red (IR) tracking camera. It can beused in conjunction with other Brainlab Image Guided Surgery (IGS) software. The connectionbetween the surgical microscope and the Brainlab navigation system is established by adedicated microscope integration cable. Designed to assist during surgeries that utilize a surgicalmicroscope, the software links the microscope view to the preoperative patient image data andprovides information based on:• The field of view through the microscope.• The microscope position relative to the patient.• The patient's medical imaging data.The microscope position is tracked during surgery so that patient image data can be displayed/superimposed over the corresponding patient anatomy, either on the monitor or through themicroscope's Head-Up Display (HUD).For orientation, the optical axis and focus point are displayed on the corresponding location on thepatient's image set. The focus point can also be displayed in the connected navigation application.The software provides functionality to verify and correct a patient registration provided by the IGSsoftware.Furthermore, it offers motorized movement functionality of supported surgical microscopes.
Intended Use and Indications for Use
The Microscope Navigation is a device, that when used with a Brainlab navigation system andcompatible instrument accessories, is intended as image guided planning and navigation systemto enable open and minimally invasive surgery.It links an instrument and the view of the surgical field (e.g., video, view through surgicalmicroscope) to a virtual computer image space on patient image data being processed by thenavigation workstation.The device is indicated for any medical condition in which a reference to a rigid anatomicalstructure can be identified relative to images (CT, CTA, X-Ray, MR, MRA and ultrasound) of theanatomy.
Place of Use
The application is developed to run on Brainlab IGS Platform or navigation system, which shall beused in hospital environments, specifically in rooms which are appropriate for surgicalinterventions (e.g. operation rooms).
User Profiles
The intended user profile is defined as follows:• Brainlab software, hardware and platforms are used by surgeons (neuro, ortho, spinal, trauma),
surgeon’s assistants and OR-nurses.• Brainlab hardware and platforms are handled by trained OR staff and maintained by trained
Brainlab personnel.
Patient Population
There are no demographic or regional limitations for patients. It is up to the surgeon to decide ifthe system shall be used to assist a certain treatment.
Using the System
8 Hardware User Guide Rev. 1.0 Microscope Navigation
Careful Handling of Hardware
System components and accessory instrumentation are comprised of precise mechanicalparts. Handle them carefully.
Plausibility Review
WarningBefore patient treatment, review the plausibility of all information input to and output fromthe system.
GENERAL INFORMATION
Hardware User Guide Rev. 1.0 Microscope Navigation 9
1.5 Compatibility with Medical Devices and Software
Compatible Non-Brainlab Surgical Microscopes
Model Manufacturer
• HS 5-1000 (Hi-R 1000 + FS 5-33)• HS 3-1000 (Hi-R 1000 + FS 3-43)• HS Hi-R 700 (FS 2-23, FS 3-43, FS 5-33)• MÖLLER 20-1000 (Hi-R 1000 + FS 4-20)
HAAG-STREIT SURGICAL/ MÖLLER-WEDEL GmbHRosengarten 1022880 WedelGermany
• M530 OHX• M530 OH6• ARVEO• M720 OH5/OHC5• M525 OH4/OHC4, F50/F40, C50/C40
Leica Microsystems (Schweiz) AGMax Schmidheiny-Strasse 2019435 HeerbruggSwitzerland
• KINEVO 900• TIVATO 700• OPMI PENTERO 900• OPMI PENTERO 800• OPMI Pentero/Pentero C
Carl Zeiss Meditec AGSite OberkochenRudolf-Eber-Straße 1173447 OberkochenGermany
Additional surgical microscopes may become available after release of this manual. ContactBrainlab support if you have any questions regarding compatibility with Brainlab software.
Non-Brainlab Devices
Using medical device combinations which have not been authorized by Brainlab may adverselyaffect safety and/or effectiveness of the devices, and endanger safety of patient, user, and/orenvironment.
Compatibility with Medical Devices and Software
10 Hardware User Guide Rev. 1.0 Microscope Navigation
1.6 Training and Documentation
Brainlab Training
Before using the system, all users must participate in a mandatory training program held by aBrainlab authorized representative to ensure safe and appropriate use.
Reading User Guides
This guide describes complex medical software or medical devices that must be used with care.It is therefore important that all users of the system, instrument or software:• Read this guide carefully before handling the equipment• Have access to this guide at all times
GENERAL INFORMATION
Hardware User Guide Rev. 1.0 Microscope Navigation 11
Training and Documentation
12 Hardware User Guide Rev. 1.0 Microscope Navigation
2 MICROSCOPE INTERFACECABLES
2.1 Microscope Navigation
General Information
Microscopes from various manufacturers can be set up for use with Brainlab navigation systems.The connection between the surgical microscope and the Brainlab navigation system isestablished by a dedicated microscope integration cable.The navigation system can determine the viewing direction, the focus point, and the field of viewdiameter of the microscope and display the information on the navigation screen.To set up a microscope:• A Microscope Adapter Set must be attached to the microscope.• The microscope and the navigation system must be connected using the correct microscope
integration cable.For more information on the Microscope Adapter Set, refer to the Cranial/ENT Optical TrackingInstrument User Guide.
Before Using
Read the relevant Software User Guide before performing microscope navigation.
MICROSCOPE INTERFACE CABLES
Hardware User Guide Rev. 1.0 Microscope Navigation 13
2.2 Cables Overview
Cables List
①
② ③ ④
⑤
⑥⑦⑧
Figure 1
No. Cable Name Article No.
① BREAKOUT INTERFACE CABLE FOR CURVE 1.0 15234B
② BNC CABLE 10 M 18562-08
③ MICROSCOPE INTERFACE CABLE 2.0 (HAAG-STREIT/MOELLER/OLYMPUS) 15204
④ SVGA MONITOR CABLE MALE-MALE 10 M 15229A
⑤ (CAN BUS) MICROSCOPE INTERFACE CABLE 2.0 (LEICA) 15222C
⑥ MICROSCOPE INTERFACE CABLE 2.0 (ZEISS) 15217C
⑦
MICROSCOPE INTERFACE CABLE 3.0 (ZEISS) (shown) 15235-01
MICROSCOPE INTERFACE CABLE 3.0 (LEICA) 15236-01
MICROSCOPE INTERFACE CABLE 3.0 (HAAG-STREIT/MOELLER) 15237-01
⑧
MICROSCOPE INTERFACE CABLE (ZEISS KINEVO 900) 15240
MICROSCOPE INTERFACE CABLE 4.0 (ZEISS KINEVO 900STEREO) 15241-01
MICROSCOPE INTERFACE CABLE 4.0 (ZEISS KINEVO900/TIVATO 700) (shown) 15241-02
MICROSCOPE INTERFACE CABLE 4.0 (ZEISS) 15241-03
MICROSCOPE INTERFACE CABLE 4.0 (LEICA) 15241-04
Cables Overview
14 Hardware User Guide Rev. 1.0 Microscope Navigation
Cabling
• Do not connect additional hardware to the original setting after calibration.• Do not connect devices other than a valid calibrated video camera to the video port of the
navigation system.• Make sure that there is enough space for the movement range of the microscope head when
connecting it.• Do not expose the system to direct UV light, as it may damage equipment.
NOTE: For further information about electrical safety, see the relevant System and TechnicalUser Guide and/or the manufacturer's instructions for use for your microscope.
MICROSCOPE INTERFACE CABLES
Hardware User Guide Rev. 1.0 Microscope Navigation 15
2.3 Strain Relief
About Strain Relief
Ensure safe usage by using strain relief loops to keep the cabling free from snagging hazards.
How to Attach the Strain Relief
①
②
③ ④
⑤Figure 2
Step
1. Put the hook and loop strap through the strain relief loops of the microscope cable, wrap-ping the strap once around the cable ①.
2. Attach the microscope cable with the hook and loop strap:• To the rail of the ceiling unit, when using the Curve Ceiling-Mounted and Buzz Navi-
gation (Ceiling-Mounted) systems ②.• To the strain relief adapter, when using the Curve Navigation 17700 ③, Curve ④ and
Kick systems ⑤.NOTE: The Curve Navigation 17700 system will be referred to from this point forward asCurve Navigation.
Strain Relief
16 Hardware User Guide Rev. 1.0 Microscope Navigation
2.4 Cleaning
Before You Begin
Ensure that the system is fully shut down and disconnected from mains power before beginningcleaning.
No Automatic Disinfection
Do not use automatic cleaning and disinfection procedures for the system components.
No Sterilization
Do not sterilize any part of the system. High temperatures from sterilization may damagecomponents.
No Liquids
Ensure that liquids do not enter the system, as this could damage the component and/or theelectronics.
Disinfectant Types
The cabling shall only be cleaned using the following disinfectant types:• alcohol-based (e.g., Meliseptol, Mikrozid AF Liquid)• alkylamine-based (e.g., Incidin Plus 2%)• active oxygen-based (e.g., Perform)• aldehyde/chloride-based (e.g., Antiseptica Kombi - Flächendesinfektion)
NOTE: Use only surface disinfectants released in your specific market.
Cleaning the Cables
To clean the microscope interface cables:
Step
1. Shut down the connected systems.
2. Disconnect the cable from the connected systems.
3. Close the connectors using cable caps if provided.
4. Clean the cable surfaces using a surface disinfectant (see list of Brainlab approved disin-fectants). Follow disinfectant manufacturer's recommendations.NOTE: Use a dry cloth to remove any surface disinfectant residue.
5. Carefully clean the surface of the connectors, ensuring that no liquids enter the connector.
MICROSCOPE INTERFACE CABLES
Hardware User Guide Rev. 1.0 Microscope Navigation 17
Cleaning
18 Hardware User Guide Rev. 1.0 Microscope Navigation
3 LEICA MICROSCOPES3.1 Connection to Curve 1.0
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve 1.0 system according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
The compatible Leica microscopes are:• M525 (OH4/OHC4, F50/F40, C50/C40)• M530 (OHX, OH6, ARVEO)• M720 (OH5/OHC5)
Leica M525/M530/M720 (SAI)
Figure 3
NOTE: It is recommended to use an HD-SDI connection if available.
LEICA MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 19
3.2 Connection to Curve 1.1 and 1.2 and BuzzNavigation (Ceiling-Mounted)
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on thenavigation systems according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
NOTE: Secure any loose cables that hang from the Buzz Navigation (Ceiling-Mounted)Connection Unit to the rail of the ceiling supply unit (CSU) with the hook and loop strap.
The compatible Leica microscopes are:• M525 (OH4/OHC4, F50/F40, C50/C40)• M530 (OHX, OH6, ARVEO)• M720 (OH5/OHC5)
Leica M525/M530/M720 (SAI)
Figure 4
NOTE: It is recommended to use an HD-SDI connection if available.
Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)
20 Hardware User Guide Rev. 1.0 Microscope Navigation
3.3 Connection to Curve Ceiling-Mounted
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve Ceiling-Mounted system according to the following diagrams.NOTE: Secure any loose cables that hang from the Curve Ceiling-Mounted Connection Unit tothe rail of the ceiling supply unit (CSU) with the hook and loop strap.
The compatible Leica microscopes are:• M525 (OH4/OHC4, F50/F40, C50/C40)• M530 (OHX, OH6, ARVEO)• M720 (OH5/OHC5)
Leica M525/M530/M720 (SAI)
Figure 5
LEICA MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 21
3.4 Connection to Curve Navigation
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve Navigation system according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
NOTE: The Curve Navigation system is only compatible with Microscope Navigation 1.6software and above.
The compatible Leica microscopes are:• M525 (OH4/OHC4, F50/F40, C50/C40)• M530 (OHX, OH6, ARVEO)• M720 (OH5/OHC5)
Leica M525/M530/M720 (SAI)
Figure 6
NOTE: It is recommended to use an HD-SDI connection if available.
Connection to Curve Navigation
22 Hardware User Guide Rev. 1.0 Microscope Navigation
3.5 Connection to Kick 1.x and Kick 2.x
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theKick system according to the following diagrams.NOTE: Secure any loose cables that hang from Kick to the base of the system with the hook andloop strap provided. This strap fits through the hole located on the foot of the system.
The compatible Leica microscopes are:• M525 (OH4/OHC4, F50/F40, C50/C40)• M530 (OHX, OH6, ARVEO)• M720 (OH5/OHC5)
Leica M525/M530/M720 (SAI)
Figure 7
LEICA MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 23
Connection to Kick 1.x and Kick 2.x
24 Hardware User Guide Rev. 1.0 Microscope Navigation
4 HAAG-STREITMICROSCOPES
4.1 Connection to Curve 1.0
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve 1.0 system according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop straps provided. These straps fit through the hole located on the foot of the system.
Haag-Streit Hi-R 1000/Hi-R 700
Figure 8
NOTE: Image injection (HUD) is not available for Haag-Streit Hi-R 700.
NOTE: It is recommended to use an HD-SDI connection if available.
HAAG-STREIT MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 25
4.2 Connection to Curve 1.1 and 1.2 and BuzzNavigation (Ceiling-Mounted)
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on thenavigation systems according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
NOTE: Secure any loose cables that hang from the Buzz Navigation (Ceiling-Mounted)Connection Unit to the rail of the ceiling supply unit (CSU) with the hook and loop strap.
Haag-Streit Hi-R 1000/ Hi-R 700
Figure 9
NOTE: Image injection (HUD) is not available for Haag-Streit Hi-R 700.
NOTE: It is recommended to use an HD-SDI connection if available.
Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)
26 Hardware User Guide Rev. 1.0 Microscope Navigation
4.3 Connection to Curve Ceiling-Mounted
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve Ceiling-Mounted system according to the following diagrams.NOTE: Secure any loose cables that hang from the Curve Ceiling-Mounted Connection Unit tothe rail of the ceiling supply unit (CSU) with the hook and loop strap.
Haag-Streit Hi-R 1000/Hi-R 700
Figure 10
NOTE: Image injection (HUD) is not available for Haag-Streit Hi-R 700.
HAAG-STREIT MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 27
4.4 Connection to Kick 1.x and Kick 2.x
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theKick system according to the following diagrams.NOTE: Secure any loose cables that hang from Kick to the base of the system with the hook andloop strap provided. This strap fits through the hole located on the foot of the system.
Haag-Streit Hi-R 1000/Hi-R 700
Figure 11
NOTE: Image injection (HUD) is not available for Haag-Streit Hi-R 700.
Connection to Kick 1.x and Kick 2.x
28 Hardware User Guide Rev. 1.0 Microscope Navigation
5 ZEISS MICROSCOPES5.1 Connection to Curve 1.0
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve 1.0 system according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop straps provided. This strap fits through the hole located on the foot of the system.
Zeiss OPMI Pentero/Pentero C/PENTERO 800/PENTERO 900, KINEVO 900 (Legacy ODU Interface)
Figure 12
NOTE: It is recommended to use an HD-SDI connection if available.
ZEISS MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 29
5.2 Connection to Curve 1.1 and 1.2 and BuzzNavigation (Ceiling-Mounted)
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on thenavigation systems according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
NOTE: Secure any loose cables that hang from the Buzz Navigation (Ceiling-Mounted)Connection Unit to the rail of the ceiling supply unit (CSU) with the hook and loop strap.
Zeiss OPMI Pentero/Pentero 900
Figure 13
Zeiss KINEVO 900 (Advanced LAN Interface), TIVATO 700 (Advanced LAN Interface)
Figure 14
NOTE: Image injection (HUD) is not available for TIVATO 700.
Connection to Curve 1.1 and 1.2 and Buzz Navigation (Ceiling-Mounted)
30 Hardware User Guide Rev. 1.0 Microscope Navigation
5.3 Connection to Curve Ceiling-Mounted
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve Ceiling-Mounted system according to the following diagrams.NOTE: Secure any loose cables that hang from the Curve Ceiling-Mounted Connection Unit tothe rail of the ceiling supply unit (CSU) with the hook and loop strap.
Zeiss OPMI Pentero/PENTERO 800/PENTERO 900, KINEVO 900 (Legacy ODU Interface)
Figure 15
Zeiss TIVATO 700
Figure 16
ZEISS MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 31
5.4 Connection to Curve Navigation
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theCurve Navigation system according to the following diagrams.NOTE: Secure any loose cables that hang from Curve to the base of the system with the hookand loop strap provided. This strap fits through the hole located on the foot of the system.
NOTE: The Curve Navigation system is only compatible with Microscope Navigation 1.6software and above.
ZEISS Pentero, PENTERO 800/ PENTERO 900
Figure 17
NOTE: Only HD-SDI can be integrated.
ZEISS KINEVO 900 (Advanced LAN Interface), TIVATO 700 (Advanced LAN Interface)
Figure 18
NOTE: Image injection (HUD) is not available for TIVATO 700.
Connection to Curve Navigation
32 Hardware User Guide Rev. 1.0 Microscope Navigation
5.5 Connection to Kick 1.x and Kick 2.x
How to Connect
Connect the supplied cables between the microscope ports and the corresponding ports on theKick system according to the following diagrams.NOTE: Secure any loose cables that hang from Kick to the base of the system with the hook andloop strap provided. This strap fits through the hole located on the foot of the system.
Zeiss OPMI Pentero
Figure 19
Zeiss OPMI PENTERO 800/PENTERO 900
① ②
Figure 20
No. Component
① Adapter cable SVHS to 2BNC
② FGA 2B-4St/S-VHS-St 10m
NOTE: The Zeiss LEMO S-VIDEO cable (000000-0457-982) is provided by Zeiss.
ZEISS MICROSCOPES
Hardware User Guide Rev. 1.0 Microscope Navigation 33
Connection to Kick 1.x and Kick 2.x
34 Hardware User Guide Rev. 1.0 Microscope Navigation
INDEXD
disposal instructions..................................................................... 6
Mmaintenance...............................................................................17microscopes
Haag-Streit, connection to navigation system........................ 25Leica, connection to navigation system..................................19Microscope Adapter Set......................................................... 13Zeiss, connection to navigation system..................................29