Microelectrode Recording

Using the NEXDRIVE by ign

Page 2: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Microelectrode Recording

• Adjust NEXDRIVE base• Assemble the NEXDRIVE• Insert the cannulae• Insert the microelectrode• Additional tracks:

– Perform a parallel track

– Using the offset positioner

• Place the DBS lead

Page 3: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Adjust NEXDRIVE base

• Loosen locking screws• Adjust Z-stage to set

drive to target depth reported by IGS

• Tighten locking screws

Loosen

Loosen

Lock

Lock

Turn

Page 4: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

NEXDRIVE Assembly

• Remove alignment adapter

• Insert Center multi-lumen insert

• Attach micro-drive

Note: be sure drive is set to starting depth

Multi-lumen Insert

Page 5: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Cannulae insertion

• Insert stylet through passing cannula

• Attach collet to passing cannula

• Pass cannula/collet combo into drive

• Remove Stylet• Insert microelectrode

cannulaNote: be sure drive is set to starting depth

Page 6: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Microelectrode insertion

• Retract microelectrode within sheath

• Pass microelectrode through cannula

• Advance electrode• Connect recording

system cables

Note: be sure drive is set to starting depth

Page 7: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Microelectrode Recording

Page 8: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Parallel Tracts

• Retract drive to starting point

• Retract and remove microelectrode & collet

• Remove cannula & re-insert through offset

Page 9: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Offset Positioner

• Retract drive to starting point• Remove microelectrode• Remove cannulae• Remove drive and drive base• Remove center multi-lumen insert• Choose direction for offset• Insert offset positioner• Re-assemble drive and cannula

Page 10: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Possible Tracks

Center Positioner

3mm offset at “B”3mm

offset at “G”

Center Positioner

at 45°

Page 11: Microelectrode Recording Using the NEXDRIVE by ign

BY IG N

guiding the way

Lead Placement

• Set drive to desired target depth

• Remove microelectrode and spacer cannula

• Calibrate lead with calibration tube

• Remove calibration tube and insert lead