Methods Microscopy

Embed Size (px)

Citation preview

  • 8/6/2019 Methods Microscopy

    1/40

  • 8/6/2019 Methods Microscopy

    2/40

    MICROSCOPIC ANATOMYMICROSCOPIC ANATOMY

  • 8/6/2019 Methods Microscopy

    3/40

    All living organisms are constructed from cells.

    Cells come in varied shapes and sizes

  • 8/6/2019 Methods Microscopy

    4/40

  • 8/6/2019 Methods Microscopy

    5/40

    Interaction of probe used (photons: light,phase contrast, polarizing & fluorescence

    microscopy; electron beams: EM), and tissue

    components produce image

    Considerations in microscopic analysis:that the probe being utilized must not be

    larger than the detail to be seen

    that the probe and object being

    investigated must interactit must be possible to observe and

    interpret this interaction

    Units for measuring microscopic dimensions:

    MICROSCOPY

  • 8/6/2019 Methods Microscopy

    6/40

    WHAT

    CAN

    WE

    SEE?

  • 8/6/2019 Methods Microscopy

    7/40

    Magnification increases the apparent size

    of the specimen; a property of both ocular &

    objective lenses

    Working distance is the distance between the

    specimen and the magnifying lens.

    Depth of field is a measure of the amount of

    a specimen that can be in focus. Highly

    sensitive video cameras enhance power of

    microscopes, and create digitized images

    that can be fed into computers for

    quantitative image analysis

    IMPORTANT TERMS IN MICROSCOPY

  • 8/6/2019 Methods Microscopy

    8/40

    Numerical aperture a measure of the size or

    angle of the cone of light delivered by the

    illuminating condenser lens to the object planeand of the cone of light emerging from the object.

    Resolving power a measure of linear distance of

    the smallest degree of separation at which 2

    details can still be distinguished from each other;dependent on quality of objective lens; R also

    varies according to the refractive index at the

    interface of the media used

    Generally resolution increases withmagnification, although there comes a point of

    diminishing returns where magnification is

    increased beyond added resolution gain.

  • 8/6/2019 Methods Microscopy

    9/40

  • 8/6/2019 Methods Microscopy

    10/40

    1.Fixation - prompt treatment of tissues infixatives

    for about12 hrs. (depending on tissue size)

    prevent autolysis by enzymes or bacteria, and

    preserve their morphologic and molecular

    composition.

    To render the structural components insoluble,

    chemicals that precipitate the proteins are used.

    The best fixatives are those that produce fine

    precipitates, e.g. buffered isotonic solution of 4%formaldehyde and glutaraldehyde react with

    amine groups (NH2) of proteins, or cross-link

    with protein

    PREPARATION OF TISSUES FOR MICROSCOPIC

    EXAMINATION (MICROTECHNIQUE)

  • 8/6/2019 Methods Microscopy

    11/40

    Gross distortions without basis in the structure

    of the living cell are termedfixation artifacts.

    Examples of artifacts: swelling and shrinkage of tissue

    components due to poor fixation,

    dehydration and/or embedding techniques;

    wrinkles, tears, air bubbles due to poor

    sectioning technique;

    dust and stain precipitate in section

    resulting from use of old stain solutions, use

    of improperly filtered or unfiltered stain

    solutions, mistakes made during

    preparation of the stain, or poor staining

    technique.

  • 8/6/2019 Methods Microscopy

    12/40

    Folded artifact Cracked tissue artifact

    Knife mark + folded artifact

  • 8/6/2019 Methods Microscopy

    13/40

    2.Dehydration and Clearing

    Bathing of tissue in graded concentrations of

    organic solvents (70-100% ethanol) to replace

    tissue water within 6-24 hrs.

    Ethanol is then replaced with solvent miscible

    in embedding medium (xylene, benzene,

    toluene) WHY? Most fixatives are water soluble, most

    embedding media are non-polar and are not

    miscible with water.

    Dehydration moves the tissue from a polar(water-based) medium to a non-polar medium

    (e.g. toluene) that is miscible with the

    embedding medium.

  • 8/6/2019 Methods Microscopy

    14/40

    3.Embedding in melted paraffin at 58-600C, or

    plastic resin at room temperature

    Tissue will be sectioned, and needs to be

    durable enough to withstand the sectioning

    process.

    Embedding in wax or plastic immobilizes

    structural components of tissue. Holds themin place as sectioning is done.

    Embedding medium must penetrate all

    cellular/intercellular spaces to impart rigid

    consistency to tissue before sectioning Tissue shrinkage and artifacts may result

    from heat needed for paraffin embedding;

    virtually absent in resin embedding

  • 8/6/2019 Methods Microscopy

    15/40

    4.Sectioning by

    microtome to a

    thickness of1-10 Qm Sections are then

    floated in warm water

    and transferred to

    glass slides Allows histologist to:

    see internal

    structure of tissue.

    stains or specific

    markers such as antibodies to more easily

    infiltrate the tissues.

    light to pass through tissue making structure

    visible.

  • 8/6/2019 Methods Microscopy

    16/40

    Images from thin sections

    are 2-D; living tissues are 3-D

    In order to understand the

    architecture of an organ,

    sections made in different

    planes should be studied

    How different3-dimensional

    structures may appear when

    thin-sectioned. A: Different

    sections through a hollow

    ball and a hollow tube. B: A

    section through a single

    coiled tube may appear as

    sections of many separate

    tubes. C: Sections through a

    solid ball (above) and

    sections through a solid

    cylinder (below).

  • 8/6/2019 Methods Microscopy

    17/40

  • 8/6/2019 Methods Microscopy

    18/40

    5.Staining to differentiate the

    colorless tissue elements as

    certain cellular elementstake up more stain than

    others, producing a contrast

    that allows observation of

    structure not visible in

    unstained tissue. It may alsoreveal differences in

    chemical nature of regions

    of the tissue.

    6.Mounting stained sections

    are placed on a slide in a

    gummy medium that

    hardens. The preparation is

    then covered with a thin

    cover glass.

  • 8/6/2019 Methods Microscopy

    19/40

    1.Light microscopes-

    compound, dissecting,

    brightfield, and phase-

    contrast Best resolution is 0.2

    m.

    Maximum

    magnifications are

    between 1000X and

    1250X.

    Anton vanLeeuwenhook

    (1632-1723)

    TYPES OF

    MICROSCOPES

  • 8/6/2019 Methods Microscopy

    20/40

    COMPOUND MICROSCOPE DISSECTING MICROSCOPE

  • 8/6/2019 Methods Microscopy

    21/40

    Compound

    microscopes

    bring small

    objects "closer"

    to the observer

    by increasing the

    magnification of

    the sample. Since the sample

    is the same

    distance from the

    viewer, a "virtual

    image" is formed

    as the light

    passes through

    the magnifying

    lenses.

  • 8/6/2019 Methods Microscopy

    22/40

  • 8/6/2019 Methods Microscopy

    23/40

    Bright-field

    Nomarski differential-

    phase-contrast

    phase-contrast

    dark-field

    Phase contrast microcopy- uses a lens system that changes

    light speed as it passes through structures with different

    refractive indices The phase of the light is altered by its passage through

    the cell, and small phase differences can be made

    visible by exploiting interference effects

    Phase-contrast and differential interference optics

    produce 3-D images of transparent living cells, tissues

  • 8/6/2019 Methods Microscopy

    24/40

    Contrast in

    LightMicroscopy

    Contrast inPhase

    Microscopy

  • 8/6/2019 Methods Microscopy

    25/40

    2.Fluorescence microscopy uses strong UV light

    source that irradiate substances dyed with

    fluorescent stains, e.g.

    acridine-orange

    These appear as

    brilliant, shiny particles

    on a dark background;

    useful for identifying& localizing NA in cells

    Fluorescence spectros-

    copy analyzes light

    emitted by fluorescent

    compounds in a micro-spectrophotometer

    This permits highly

    sensitive assays of

    cellular substances such as catecholamines

  • 8/6/2019 Methods Microscopy

    26/40

  • 8/6/2019 Methods Microscopy

    27/40

    Theconfocal

    microscope

    produces

    opticalsections by

    excluding

    out-of-focus

    light

  • 8/6/2019 Methods Microscopy

    28/40

    3.Polarizing

    microscopy

    birefrigentsubstances rotate

    direction of

    polarized light

    emerging frompolarizing filters

    Useful for

    visualizing

    substances withrepetitive, oriented

    molecular

    structuresCollagen fibers,

    polarizing microscopy

    Compact

    bone

  • 8/6/2019 Methods Microscopy

    29/40

    4.Electron microscopy uses high energy electron

    beams (between 5,000 - 109 electron volts)

    focused through electromagnetic lenses. Interaction of electrons deflected by lenses

    beamed on tissue components permits high

    resolution (0.2 - 1 nm) and 400x greater

    magnification than light microscopes The increased resolution results from the shorter

    wavelength of the electron beam

    Disadvantages of EM: requirements of a vacuum-

    enclosed system, high voltage, mechanicalstability; special treatment & sample preparation

    make it highly complex and costly; requires the

    services of well-trained personnel

  • 8/6/2019 Methods Microscopy

    30/40

  • 8/6/2019 Methods Microscopy

    31/40

  • 8/6/2019 Methods Microscopy

    32/40

    TE

    M

    S

    E

    M

  • 8/6/2019 Methods Microscopy

    33/40

    Scanning vs. Transmission EM:

    In the TEM, the image is formed directly on the image

    plane.

    In the SEM, the image is formed indirectly by

    accumulation of information from the specimen point

    by point.

    There is no need to cut ultra thin sections becausethe beam of the SEM does not pass through the

    specimen.

    The resolution of the SEM is about100 Angstrom vs.

    4-5 Angstrom achieved by the transmission type.

    The SEM has great depth of field making it possibleto obtain 3-D images.

    TEM magnifications are commonly over 100,000X

    SEM displays images on high resolution TV monitors.

  • 8/6/2019 Methods Microscopy

    34/40

    SEM: T-lymphocyte, E.coli

    attacked by macrophage

    TEM: mitochondria &

    chloroplast

  • 8/6/2019 Methods Microscopy

    35/40

    Specimen Preparation for EM:

    Fixation in osmium tetroxide, osmium dichromate,

    acrolein and glutaldehyde. Since registration of color is not possible with the

    EM system, staining with colored dyes is not done in

    EM studies.

    Specimen is mounted on a copper grid covered withcarbon and/or plastic film

  • 8/6/2019 Methods Microscopy

    36/40

  • 8/6/2019 Methods Microscopy

    37/40

    Freeze-cleaving, Freeze-

    etching or Cryofracture

    methods

    Used with EM; replicas are

    made of surfaces of frozen

    aqueous materials at very

    low temperatures in vacuo The use of chemical

    fixatives, dehydrating and

    embedding agents are

    avoided by using a freezing

    microtome/cryostat whichpermit sections to be

    obtained without

    embedding

  • 8/6/2019 Methods Microscopy

    38/40

    chloroplast thylakoid membranes

    Freezing does not inactivate

    most enzymes, hinders

    diffusion of smallmolecules, eliminates

    dissolution of tissue lipids

    by solvents

    The tissue is impregnated

    with a 25% glycerolsolution

    before rapidfreezingin

    liquidnitrogenorFreon12 at

    1000C to1550C.

    Notentirelyfree ofartifacts;valuable inthe studyof

    membranes andtheir

    junctionalspecializations.

    vesicles

  • 8/6/2019 Methods Microscopy

    39/40

    QUESTIONS?

  • 8/6/2019 Methods Microscopy

    40/40

    REFERENCES

    Bloom and Fawcett. 12th ed. Textbook ofHistology

    de Fiore, Atlas ofHistology

    Junquiera, LC and Carneiro, J. Basic Histology: Text and Atlas, 12th ed

    Ross, M. et. al., 4th ed. BasicHistology: Text and Atlas

    Kuehnel,W. Color Atlas of Cytology, Histology and Microscopic

    Anatomy

    Young, B.Wheaters FunctionalHistology: A Text and Color Atlas

    http://www.northland.cc.mn.us/biology/AP2Online/Nervous/Default.htm

    http://highered.mcgraw-

    hill.com/sites/0072495855/student_view0/histology_atlas.html

    http://www.meddean.luc.edu/lumen/meded/Histo/frames/histo_frames.html

    http://projects.galter.northwestern.edu/rhodin/

    http://www.siumed.edu/~dking2/index.htm

    http://www.bu.edu/histology/m/i_main00.htm

    http://www.histology-world.com/

    http://www.mhhe.com/biosci/ap/histology_mh/start_histology.html